N U C L E O N I X S Y S T E M S P V T . L T D I N S T R U C T I O N M A N U A L NUCLEONIX SYSTEMS PRIVATE LIMITED Plot No: 162A&B, PHASEII, I.D.A.Cherlapally, Hyderabad-500051, INDIA Phone: +91-40-27263701, 040-27262146, 040-68888777, Fax: 040 – 27262146 Mobile No: 7331104480, 7331104481, E-mail: [email protected]PRE-AMPLIFIER PA320
13
Embed
NUCLEONIX SYSTEMS PRIVATE LIMITEDnucleonix.com/Manuals/PA320_Man.pdf · NUCLEONIX SYSTEMS PRIVATE LIMITED Plot No: 162A&B, ... serial number of the unit, date of supply etc., ...
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
NUCLEONIX
SYSTEM
S PVT.
LTD
I N S T R U C T I O N
M A N U A L
NUCLEONIX SYSTEMS PRIVATE LIMITED Plot No: 162A&B, PHASEII, I.D.A.Cherlapally, Hyderabad-500051, INDIA
In all correspondance regarding the instrument, please mention the type, serial number of the unit, date of supply etc., of the unit.
o FEATURES
Compact solid state circuit design Suitable for wide range of x-ray Proportional and gas flow type of
detectors.
Rise time 100 nano sec.
It is a charge sensitive pre-amplifier uses N-channel FET, in the inputwith charge collection capacitor in the feedback of the operationalamplifier.
CHAPTER - I
INTRODUCTIONCharge sensitive Pre-amplifier PA320 is designed for use with gas filled proportional detectors, X-ray proportional detectors, BF-3 & He3 neutron detectors etc. Users will have the choice to change the load resistors & charge sensitive feedback capacitor according to the gain & sensitivity requirement for a specific application. Charge sensitive pre amplifier consists of an operational amplifier with capacitive feedback & uses an N-FET in the input stage. Output driver transistor is capable of driving a 100 ohms impedance coaxial cable.
Detector output is also brought out from the second stage, for using it for data acquisition purpose. This unit is provided with built in comparator with threshold adjustment to give TTL output for counting applications.
NUCLEONIX
SYSTEM
S PVT.
LTD
: 10 M Ohms
: 100 Ohms
: 1000mv / pico coulombs for first stage. Followed by additional voltage gain stage.
: Better than 100 nano seconds.
: 20 micro sec. (approx).
: Operational amplifier with FET in the front end.
Detector Load Resistor
Output impedance
Charge Sensitivity
Rise Time
Decay Time Constant
Charge Sensitive Device
Power Supply Inputs
Output(s) available:
Dimensions :
: +12V@ 20mA -12V@ 10mA
CHAPTER - II
SPECIFICATIONS
:(a) Linear signal (amplified detector output)(b) TTL output (after comparator
: 155mm L X 95mmW X 50mmH.
CHAPTER - III
INSTALLATION
A. CONTROLS / CONNECTORS:
EHT (MHV socket) : High Voltage connector, at which the EHT supply is provided
to connect to the X-ray proportional counter / gas proportional
/ He-3 / BF3 / Neutron detector for detector biasing.
Output (Linear) : BNC Co-axial connector at which output from the Pre-
Amplifieris generated for connecting to shaping amplifier.
Output (TTL) : This is a TTL output after discriminator provided to cut-off noise.
LV (socket) : Five pin circular connector at which the LV supply voltages
are provided from the NIM Standard instrumentation bin & power
supply unit. Power up the PA320 is fed.
Detector (MHV socket) : MHV connector, to which detector is connected preferably
through a short RG 59 cable. Cable end has to match with
the appropriate connector on the detector.
B. EXTERNAL CONNECTIONS:
Having connected the Pre-amplifier to the detector as explained earlier, proceed to make
the connection as follows:
1. Using the HV cable (MHV to MHV) connect the OUTPUT from High Voltage Unit, type
HV502 /HV503 of NUCLEONIX make or its equivalent to the HV connector (MHV
socket) on PA320.
2. Using Co-axial signal cable connect from the OUTPUT BNC connector on PA320 to the
INPUT of any Pulse shaping Amplifier typically LA520 or AF521or Spectroscopy amplifier.
3. Using LV - I/O, 5 core cable connect to the "LV" connector on PA320 to the LV connector
on Instrumentation bin & power supply. If your instrumentation bin has a 9-pin D-connector,
then you wire up one end with 9 pin D-connector & the other end with 5 pin I/O connector
to mate with corresponding connector on pre-amplifier.
CHAPTER – IV
BLOCK DIAGRAM & ITS DESCRIPTION
C1
Cc
R1
HV
1M 0.01/22KV
0.001/2KV
10M To Detector
anode
C7 RL
Non Inv gain stage
OUTPUT
MHV socket
MHV socket
FET input
TTL
R6
Op amp
Driver
+
P1
20K
+ -
-
Charge Integrator PZ circuit
Op amp
Op amp
Mono
Comparator Mono
OUTPUT
LINEAR
Block diagram of Charge sensitive pre amplifier of PA320 of Nucleonix make
Block diagram description: It can be seen from the block diagram, HV input to the detector is filtered further by an RC filter of R1 (1meg ohm) & capacitor C1 (0.01uf, 2KV). Detector such as an X-ray proportional detector, gas proportional detector or a Neutron detector BF3 or HE-3 IS connected at the MHV /SHV socket provided on the pre amp unit, thru appropriate co-axial cable. Signal from this is ac coupled & applied to the gate input of FET. Followed by this is a charge integrating op.amp with a feedback capacitor of 1pf which integrates the charge collated to give voltage pulse output. Followed by charge integrator output, there is a transistor driver, PZ- circuit and a non inverting op.amp stage, which provides further voltage gain required.
Trimpot (P1) on second stage op.amp can be used to adjust gain.Signal from the output of this second stage is fed to a comparator which helps in cutting of any noise signal. Output of this comparator is fed to a mono stable to get a fixed width TTL output signal. Both Linear output from the second stage & TTL outputs have been brought out for external use & processing or further amplification & acquisition.
Rl4.
HV TPI
SIZE,55Xll9mm CHARGE SENSITIVE PRE A�LIFIER 07 /05/2016 NSPL_PCB_0542
NOTE (2): To Change charge sensitivity gain,One has to change capacitor (C7) value. By default it is 1pf. One can increase it to 2.2pf or 3.3pf etc. Increasing the value decrease inversely charge sensitivity.
NOTE (3):Further Signal Voltage can be manipulated by P1 – trimpot which is in the op.amp (U2) feedback. Manipulation of P1 alters gain in the second stage.
Contact any front office "Business Executive"- He/She will take your details and connect you to concerned product engineer for any technical clarifications. Best thing is to email your technical queries and obtain the reply, rather than on telephone.
You can also contact General Manager or Director (Tech) if required.
e) Marketing ManagerOn business matters for all your marketing services / techno commercial requirements aboutNucleonix Products contact:
Contact General Manager for all sales / servicing and technical information including customer support related issues, on the delays, gaps & lapses by our staff. Contact G.M. regarding field installations & field servicing jobs schedule etc.