DEVICE SPECIFICATIONS NI 6368 X Series Data Acquisition: 2 MS/s/ch, 16 AI, 48 DIO, 4 AO The following specifications are typical at 25 °C, unless otherwise noted. For more information about the NI 6368, refer to the X Series User Manual available from ni.com/ manuals. Analog Input Number of channels 16 differential ADC resolution 16 bits DNL No missing codes guaranteed INL Refer to the AI Absolute Accuracy section. Sample rate Single channel maximum 2.00 MS/s Minimum No minimum Timing resolution 10 ns Timing accuracy 50 ppm of sample rate Input coupling DC Input range ±1 V, ±2 V, ±5 V, ±10 V Maximum working voltage for all analog inputs Positive input (AI+) ±11 V for all ranges, Measurement Category I Negative input (AI-) ±11 V for all ranges, Measurement Category I Caution Do not use for measurements within Categories II, III, and IV. CMRR (at 60 Hz) 75 dB Bandwidth 1 MHz THD -80 dBFS
18
Embed
NI 6368 Device Specifications - National Instruments · DEVICE SPECIFICATIONS NI 6368 X Series Data Acquisition: 2 MS/s/ch, 16 AI, 48 DIO, 4 AO ... Maximum working voltage for all
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
DEVICE SPECIFICATIONS
NI 6368X Series Data Acquisition: 2 MS/s/ch, 16 AI, 48 DIO, 4 AO
The following specifications are typical at 25 °C, unless otherwise noted. For moreinformation about the NI 6368, refer to the X Series User Manual available from ni.com/manuals.
Analog InputNumber of channels 16 differential
ADC resolution 16 bits
DNL No missing codes guaranteed
INL Refer to the AI Absolute Accuracy section.
Sample rate
Single channel maximum 2.00 MS/s
Minimum No minimum
Timing resolution 10 ns
Timing accuracy 50 ppm of sample rate
Input coupling DC
Input range ±1 V, ±2 V, ±5 V, ±10 V
Maximum working voltage for all analog inputs
Positive input (AI+) ±11 V for all ranges, Measurement Category I
Negative input (AI-) ±11 V for all ranges, Measurement Category I
Caution Do not use for measurements within Categories II, III, and IV.
NoiseUncertainty = RandomNoise ⋅ 310, 000 for a coverage factor of 3 σ and averaging
10,000 points.
AI Absolute Accuracy ExampleAbsolute accuracy at full scale on the analog input channels is determined using the followingassumptions:• TempChangeFromLastExternalCal = 10 °C• TempChangeFromLastInternalCal = 1 °C• number_of_readings = 10,000• CoverageFactor = 3 σ
For example, on the 10 V range, the absolute accuracy at full scale is as follows:GainError = 114 ppm + 8 ppm · 1 + 5 ppm · 10 = 172 ppmOffsetError = 15 ppm + 35 ppm ·1 + 46 ppm = 96 ppm
Noise Uncertainty = 252 μV ⋅ 310, 000 = 7.6 µV
AbsoluteAccuracy = 10 V · (GainError) + 10 V · (OffsetError) + NoiseUncertainty =2688 µV
Analog OutputNumber of channels 4
DAC resolution 16 bits
DNL ±1 LSB, max
Monotonicity 16 bit guaranteed
Accuracy Refer to the AO Absolute Accuracy section.
Output FIFO size 8,191 samples shared among channels used
Data transfers DMA (scatter-gather), programmed I/O
AO waveform modes Non-periodic waveform, periodic waveformregeneration mode from onboard FIFO,periodic waveform regeneration from hostbuffer including dynamic update
Figure 1. Analog Output External Reference Bandwidth
–90
–80
–70
–60
–50
–40
–30
–20
–10
0
10
100 1 k 10 k 100 k 1 M
Frequency (Hz)
Nor
mal
ized
AO
Am
plitu
deA
ttenu
atio
n (d
B)
DAC Output CODE (HEX)
FFFF
BFFF
8FFF
83FF
80FF
803F
800F
8003
AO Absolute AccuracyAbsolute accuracy at full-scale numbers is valid immediately following self calibration andassumes the device is operating within 10 °C of the last external calibration.
Table 2. AO Absolute Accuracy
NominalRange
PositiveFull
Scale
NominalRange
NegativeFull
Scale
ResidualGainError
(ppm ofReading)
GainTempco(ppm/°C)
ReferenceTempco(ppm/°C)
ResidualOffsetError
(ppm ofRange)
OffsetTempco(ppm ofRange/
°C)
INLError(ppm
ofRange)
AbsoluteAccuracy
at FullScale(μV)
10 -10 129 17 5 65 1 64 3,256
5 -5 135 8 5 65 1 64 1,616
Note Accuracies listed are valid for up to two years from the device externalcalibration.
AO Absolute Accuracy EquationAbsoluteAccuracy = OutputValue ⋅ (GainError) + Range ⋅ (OffsetError)
Output selections 10 MHz Clock, frequency generator output;many internal signals
Debounce filter settings 90 ns, 5.12 μs, 2.56 ms, custom interval,disable; programmable high and lowtransitions; selectable per input
Bus InterfaceForm factor x1 PXI Express peripheral module,
specification rev 1.0 compliant
Slot compatibility x1 and x4 PXI Express or PXI Express hybridslots
DMA channels 8, can be used for analog input, analog output,digital input, digital output, counter/timer 0,counter/timer 1, counter/timer 2,counter/timer 3
All PXIe devices may be installed in PXI Express slots or PXI Express hybrid slots.
Power RequirementsCaution The protection provided by the device can be impaired if the device isused in a manner not described in the X Series User Manual.
1 Has a self-resetting fuse that opens when current exceeds this specification.
14 | ni.com | NI 6368 Device Specifications
Maximum Working VoltageMaximum working voltage refers to the signal voltage plus the common-mode voltage.
Channel to earth 11 V, Measurement Category I
Caution Do not use for measurements within Categories II, III, or IV.
Shock and VibrationOperational shock 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with IEC 60068-2-27.Test profile developed in accordance withMIL-PRF-28800F.)
Random vibration
Operating 5 to 500 Hz, 0.3 grms
Nonoperating 5 to 500 Hz, 2.4 grms(Tested in accordance with IEC 60068-2-64.Nonoperating test profile exceeds therequirements of MIL-PRF-28800F, Class 3.)
SafetyThis product is designed to meet the requirements of the following electrical equipment safetystandards for measurement, control, and laboratory use:• IEC 61010-1, EN 61010-1• UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or the OnlineProduct Certification section.
Electromagnetic CompatibilityThis product meets the requirements of the following EMC standards for electrical equipmentfor measurement, control, and laboratory use:• EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity• EN 55011 (CISPR 11): Group 1, Class A emissions• EN 55022 (CISPR 22): Class A emissions• EN 55024 (CISPR 24): Immunity• AS/NZS CISPR 11: Group 1, Class A emissions• AS/NZS CISPR 22: Class A emissions• FCC 47 CFR Part 15B: Class A emissions• ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended foruse in commercial, light-industrial, and heavy-industrial locations. In Europe,Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intendedfor use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medicalequipment that does not intentionally generate radio frequency energy for thetreatment of material or inspection/analysis purposes.
Note For EMC declarations and certifications, and additional information, refer tothe Online Product Certification section.
CE Compliance This product meets the essential requirements of applicable European Directives, as follows:• 2014/35/EU; Low-Voltage Directive (safety)• 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
16 | ni.com | NI 6368 Device Specifications
Online Product CertificationRefer to the product Declaration of Conformity (DoC) for additional regulatory complianceinformation. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in theCertification column.
Environmental ManagementNI is committed to designing and manufacturing products in an environmentally responsiblemanner. NI recognizes that eliminating certain hazardous substances from our products isbeneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impactweb page at ni.com/environment. This page contains the environmental regulations anddirectives with which NI complies, as well as other environmental information not included inthis document.
Waste Electrical and Electronic Equipment (WEEE)EU Customers At the end of the product life cycle, all NI products must bedisposed of according to local laws and regulations. For more information abouthow to recycle NI products in your region, visit ni.com/environment/weee.
电子信息产品污染控制管理办法(中国 RoHS)中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物
质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录
ni.com/environment/rohs_china。(For information about China RoHScompliance, go to ni.com/environment/rohs_china.)
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product andcompany names mentioned herein are trademarks or trade names of their respective companies. For patents covering NIproducts/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or theNational Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs)and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and otherimport/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATIONCONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained inthis manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forthin FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.