Top Banner
Newer methods for measuring Intraocular Pressure Pradeep Ramulu MD MPH PhD Wilmer Nursing Conference June 8 th , 2012
22

Newer methods for measuring Intraocular Pressure

Dec 18, 2021

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Newer methods for measuring Intraocular Pressure

Newer methods for measuring Intraocular Pressure

Pradeep Ramulu MD MPH PhD

Wilmer Nursing Conference

June 8th, 2012

Page 2: Newer methods for measuring Intraocular Pressure

Financial Disclosures

I have no financial interest in any of the products discussed.

No unlabeled drugs will be discussed.

Page 3: Newer methods for measuring Intraocular Pressure

What is Intraocular Pressure?

“Tension” of the eye

Determined by balance b/w aqueous production

& outflow

Expressed in millimeters of mercury (mm Hg)

Page 4: Newer methods for measuring Intraocular Pressure

Elevated IOP induces strain on the optic nerve as it exits the eye

Page 5: Newer methods for measuring Intraocular Pressure

Strain affects the blood supply and astrocytes in the laminar beams

Page 6: Newer methods for measuring Intraocular Pressure

The end result is glaucoma

Normal Glaucoma

Page 7: Newer methods for measuring Intraocular Pressure

Applanation tonometry is the current standard for IOP measurement

Measures force necessary to flatten 3.06 mm2 of cornea

Goldmann applanator is slit-lamp mounted

Hand-held version: Perkins

Page 8: Newer methods for measuring Intraocular Pressure

Even Goldmann tonometry has limitations

Deviations in thick or thin clear corneas

Error in swollen/scarred corneas unknown

Real gold standard for IOP is cannulation of the

anterior chamber

Page 9: Newer methods for measuring Intraocular Pressure

Goldmann tonometer pushes cornea with adjustable force until it flattens

Page 10: Newer methods for measuring Intraocular Pressure

A beam splitter in the tonometer tip indicates when the force is correct

Force adjusted with dial

Endpoint = touching semicircles

created by split beam

Touching semicircles correct amount of

flattened cornea

Page 11: Newer methods for measuring Intraocular Pressure

Handheld applanators use the same principle

Patient need not be sitting

No need for slit lamp

Still need fluorescein

Page 12: Newer methods for measuring Intraocular Pressure

Applanation tonometry is not suitable for some patients

Irregular corneal surface

Blinkers/squeezers

Eyelid swelling/sensitivity

Page 13: Newer methods for measuring Intraocular Pressure

Several alternatives to traditional applanation tonometry

Noncontact tonometry

Tonopen

Pneumotonometer

Rebound tonometry

Page 14: Newer methods for measuring Intraocular Pressure

Non-contact tonometry relies on an “air puff”

No chance of spreading

infection

No need for skilled operator

Eliminates the needs for cleaning/disposables

Page 15: Newer methods for measuring Intraocular Pressure

Limitations to non-contact tonometry

Can be some patient discomfort

Prone to errors at extremes of IOP

Does not solve many of the issues with

Goldmann tonometry

Page 16: Newer methods for measuring Intraocular Pressure

Tonopen is also an applanation

Applanates a very small section of the cornea

“Works” in corneal irregularity

Averages several readings

Portable

Page 17: Newer methods for measuring Intraocular Pressure

Tonopen still has limitations

Need to poke the eye at least 5 times

Requires anesthesia

Accuracy unknown

Shared instruments often work poorly

Page 18: Newer methods for measuring Intraocular Pressure

Rebound tonometry works by a different mechanism

Lightweight probe bounced against cornea

Faster probe with higher IOP

Shorter probe/cornea contact time with higher

IOP

Page 19: Newer methods for measuring Intraocular Pressure

Rebound tonometer now available for clinic and research use

iCare Tonometer

Page 20: Newer methods for measuring Intraocular Pressure

Several distinct advantages to iCare

No need for anesthetic drop

No need to touch eye or head

High success of getting clinic IOP in children

Page 21: Newer methods for measuring Intraocular Pressure

iCare incorporation presents new challenges

Can avoid IOP portion of EUAs, but…..

Need to document ON appearance outside OR

Need to refract/check axial lengths outside OR

…..Not everyone is spared an EUA

Page 22: Newer methods for measuring Intraocular Pressure

iCare accuracy isn’t perfect

About 15% in whom IOP is off by 3 mm Hg

Validation studies only in simple patients

Doesn’t work well in supine patients