Need for Sulphur measurement in Multi- product pipelines Using PAC Antek High Speed Sulphur Analyser
23
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Need for sulphur measurement in multi product pipelines

Need for Sulphur measurement in Multi-product pipelinesNeed for Sulphur measurement in Multi-product pipelines

Using PAC Antek High Speed Sulphur Analyser

Page 2: Need for sulphur measurement in multi product pipelines

Moving Products

Different product batches are “pushed” through the system abutting each other. The product is kept in turbulent flow to minimize interfaces the mixing zone between batches.

Product Integrity During Movement

Pipelines use optimum sequencing of product grades to reduce the potential for contamination and/or

downgrading of the products transported.

Reformulated regular

gasoline

All premium grades

Conventional regular

gasoline

High sulphur Diesel

Kerosene/ Jet Fuel

Low sulphur diesel fuel

Reformulated regular

gasoline

Page 3: Need for sulphur measurement in multi product pipelines

Methods for interface detection• PAC High Speed Sulphur analysis (1 min.)• Other techniques ( 4-5 min.)

Page 4: Need for sulphur measurement in multi product pipelines

Analysis Times for Various Methods

EDXRF >10 Minutes

Lead Acetate Paper Tape 8+ minutes

PUVF 4-5 minutes

Gas Chromatography 5-6 minutes

ANTEK UVF (standard) 2 ½ minutes

ANTEK UVF HSS 1 minute or less

Page 5: Need for sulphur measurement in multi product pipelines

RAPID RESPONSE OF THE ANALYZERsulphur IN GASOLINE SAMPLE SWITCHING FROM 110 TO 220 AND 15 mg/kg sulphur

0153045607590

105120135150165180195210225240

1 11 21 31 41 51 61 71 81 91 101 111 121 131 141 151

Minutes (1 Analysis/min.)

Co

nc

en

tra

tio

n (

pp

m/w

t. S

)

96-97% RESPONSE IN 1 MINUTE OR LESS

Page 6: Need for sulphur measurement in multi product pipelines

RECOVERY TIME FOR GASOLINE

TRANSITION FROM 220 TO 15 mg/kg sulphur GASOLINE

0153045607590

105120135150165180195210225240

-3 -2 -1 0 1 2 3 4 5 6 7 8 9

Injections (minutes)

mg

/kg

90% RESPONSE IN 1 MINUTE OR LESS

Page 7: Need for sulphur measurement in multi product pipelines

Other instruments and technologies use averaging technique for Sulphur measurement

0

20

40

60

80

100

120

• T90 Response every 2.5mins or more 3 min 6 min

Page 8: Need for sulphur measurement in multi product pipelines

Comparison of 6200 HSS with other techniques

0

50

100

150

200

250

300

350

0 60 120 180 240 300 360 420 480 540 600 660 720 780

• Product change over• 6200HSS detection • Other instruments & techniques

Page 9: Need for sulphur measurement in multi product pipelines

Principle of measurement UV-Fluorescence According to ASTM D 5453

SO2 + uv SO2*SO2+eAmount of energy produced is directly

proportional to the number of sulphur atoms in the sample

R-S + O2 1000+°C SO2 + combustion products

Page 10: Need for sulphur measurement in multi product pipelines

Total sulphur Analytical Flow path

Pyro-UV-Fluorescence Method (ASTM 5453, ISO EN 20846 )

Pyro-Oxygen

Inlet (O2 )

SO2*

PMT

photon

Detector

Vent

FurnaceInjection Valve

SO2+ CO2

DCSUV-Lamp

Hydrocarbon Stream

Sample Conditioning

Carrier Gas (N2,)

R-S + O2 1050°C H2O + SO2 + CO2

4-20 mA

Heat traced

Page 11: Need for sulphur measurement in multi product pipelines

SIMPLE UV FLUORESCENCE DETECTOR

UV Lamp

PMT

Filter and Collimating lenses

UVExcitation

Light

Interference

Filter

Gas Stream from Pyro Tube

Voltage

Time

Theory of OperationDetect Sample

Page 12: Need for sulphur measurement in multi product pipelines

SIMPLE UV FLUORESCENCE DETECTOR

UV Lamp

PMT

SO2

SO2 in sample stream

Voltage

Time

Theory of Operation

Page 13: Need for sulphur measurement in multi product pipelines

SIMPLE UV FLUORESCENCE DETECTOR

UV Lamp

PMT

SO2

Emission Light

Voltage

Time

Theory of Operation

Page 14: Need for sulphur measurement in multi product pipelines

SIMPLE UV FLUORESCENCE DETECTOR

UV Lamp

PMTSO2

Voltage

Time

Theory of OperationDetect Sample

Page 15: Need for sulphur measurement in multi product pipelines

SIMPLE UV FLUORESCENCE DETECTOR

UV Lamp

PMT

SO2

Voltage

Time

Theory of OperationDetect Sample

Page 16: Need for sulphur measurement in multi product pipelines

SIMPLE UV FLUORESCENCE DETECTOR

UV Lamp

PMT

SO2

Voltage

Time

Theory of OperationDetect Sample

Page 17: Need for sulphur measurement in multi product pipelines

SIMPLE UV FLUORESCENCE DETECTOR

UV Lamp

PMT

Voltage

Time

Theory of OperationDetect Sample

Page 18: Need for sulphur measurement in multi product pipelines

Petroleum Product:Flow Through PipelineTYPICAL SEQUENCE

• High speed analysis can significantly reduce the transmix volumes & increase the operational profitability

High sulphur Diesel

Ultra Low sulphur Diesel

Premium Gasoline

Regular Gasoline

Jet Fuel

ANTEK P6200 HSS

REGULAR GASOLINE

REGULAR GASOLINE

REGULAR GASOLINE

DIESELFUEL

DIESELFUEL

PREMIUM GASOLINE

TRANSMIX

Compatible InterfacesTransmix: Interface material that must be processed

Page 19: Need for sulphur measurement in multi product pipelines

6200 Additional Features

• Specifications

– Detection Method: UV Fluorescence– ASTM Method: ASTM D-5453– Ranges: Adjustable– Lower Detection: 1 PPM– Accuracy: +/- 1% of Full-Scale– Repeatability: < +/- 1% of Full-Scale– Linearity: < +/- 1% of Full-Scale– Response Time T90: less than 1 minute

Page 20: Need for sulphur measurement in multi product pipelines

Advantages of HSS • Single Injection single result.• Reduced time lag in interface detection.• No guess work Exact Sulphur Values.• Reduced risk of off-spec products in tank.• Reduced use of transmix.• Increase operational profitability.• Complete automation of interface cut.

Page 21: Need for sulphur measurement in multi product pipelines

Cost benefit analysis using HSS

PAC Antek 6200 HSS

Other techniques

Time taken for analysis

1 mins 4 mins

Amount of interface required @ 800KL/hr

27KL 70KL

Cost of interface @Rs 10 /ltr

Rs. 2,70,000 Rs. 7,00,000

Saving per cut > Rs.4,00,000

Page 22: Need for sulphur measurement in multi product pipelines

Field Proven

• No other instrument known to be installed worldwide for multi-product pipelines yet

• In India, 2 instruments operational at BPCL and 3 shipped to HPCL

• Large installed base of sulphur analysers in refineries in India and worldwide

Page 23: Need for sulphur measurement in multi product pipelines

Thank You!

In India:Marsap Services Pvt. Ltd.Email: [email protected]: www.marsap.com