Top Banner
Document No.: M-W3449AE-2.0 ANRITSU CORPORATION MX283087A TRX Sweep Calibration Operation Manual Remote Control Second Edition For safety and warning information, please read this manual before attempting to use the equipment. Additional safety and warning information is provided within the MS2830A Signal Analyzer Operation Manual (Mainframe operation) and MX283087A TRX Sweep Calibration Operation Manual (Operation). Please also refer to these documents before using the equipment. Keep this manual with the equipment.
86

MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Dec 10, 2018

Download

Documents

doandung
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Document No.: M-W3449AE-2.0

ANRITSU CORPORATION

MX283087A TRX Sweep Calibration

Operation Manual Remote Control

Second Edition

• For safety and warning information, please read this manual before attempting to use the equipment.

• Additional safety and warning information is provided within the MS2830A Signal Analyzer Operation Manual (Mainframe operation) and MX283087A TRX Sweep Calibration Operation Manual (Operation). Please also refer to these documents before using the equipment.

• Keep this manual with the equipment.

Page 2: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

ii

Safety Symbols To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Corporation uses the following safety symbols to indicate safety-related information. Ensure that you clearly understand the meanings of the symbols BEFORE using the equipment. Some or all of the following symbols may be used on all Anritsu equipment. In addition, there may be other labels attached to products that are not shown in the diagrams in this manual.

Symbols used in manual This indicates a very dangerous procedure that could result in serious injury or death if not performed properly.

This indicates a hazardous procedure that could result in serious injury or death if not performed properly. This indicates a hazardous procedure or danger that could result in light-to-severe injury, or loss related to equipment malfunction, if proper precautions are not taken.

Safety Symbols Used on Equipment and in Manual The following safety symbols are used inside or on the equipment near operation locations to provide information about safety items and operation precautions. Ensure that you clearly understand the meanings of the symbols and take the necessary precautions BEFORE using the equipment.

This indicates a prohibited operation. The prohibited operation is indicated symbolically in or near the barred circle.

This indicates an obligatory safety precaution. The obligatory operation is

indicated symbolically in or near the circle. This indicates a warning or caution. The contents are indicated symbolically in or

near the triangle. This indicates a note. The contents are described in the box. These indicate that the marked part should be recycled.

MX283087A TRX Sweep Calibration Operation Manual Remote Control 15 October 2010 (First Edition) 29 May 2013 (Second Edition) Copyright © 2010-2013, ANRITSU CORPORATION. All rights reserved. No part of this manual may be reproduced without the prior written permission of the publisher. The contents of this manual may be changed without prior notice. Printed in Japan

DANGER

WARNING

CAUTION

Page 3: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

iii

Notes On Export Management This product and its manuals may require an Export License/Approval by the Government of the product's country of origin for re-export from your country. Before re-exporting the product or manuals, please contact us to confirm whether they are export-controlled items or not. When you dispose of export-controlled items, the products/manuals need to be broken/shredded so as not to be unlawfully used for military purpose.

Page 4: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

iv

Page 5: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

I

About This Manual About this document

This operation manual is for MX283087A TRX Sweep Calibration (Remote Control).

MS2830A Signal Analyzer Operation Manual (Mainframe: Operation) MS2690A/MS2691A/MS2692A and MS2830A Signal Analyzer Operation Manual (Mainframe: Remote Control)

MX283087A TRX Sweep Calibration (Operation)

MX283087A TRX Sweep Calibration (Remote Control)

• Mainframe: Operation

• Mainframe: Remote Control

These describe basic operations, maintenance procedures, common functions and common remote functions of the signal analyzer.

• MX283087A TRX Sweep Calibration Operation Manual (Operation)

This document describes the operation of MX283087A TRX Sweep Calibration.

• MX283087A TRX Sweep Calibration Operation Manual (Remote Control:

This document)

This document describes the remote control of MX283087A TRX Sweep Calibration Operation Manual.

Page 6: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

II

Table of Contents

About This Manual........................................ I

Chapter 1 Overview .................................... 1-1 1.1 Product Overview ...................................................... 1-2

1.2 Basic Flow of Control ................................................. 1-3

1.3 Native Mode ............................................................... 1-11

1.4 Setting Numeric Program Data .................................. 1-13

Chapter 2 SCPI Device Message Details .. 2-1 2.1 Config Function Device Messages ............................ 2-3

2.2 Application Common Device Messages .................... 2-6

Chapter 3 SCPI Status Register ................ 3-1 3.1 Reading Measurement Status ................................... 3-2

3.2 STATus:QUEStionable Register ................................ 3-3

3.3 STATus:OPERation Register .................................... 3-13

Appendix A Programming Sample ........... A-1

Page 7: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

1

III

2

3

Appendix

Page 8: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

IV.

Page 9: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 1 Overview

1-1

1

Overview

This chapter outlines the remote control operation of the MX283087A TRX Sweep Calibration (hereinafter referred to as“MX283087A”).

1.1 Product Overview ......................................................... 1-2 1.1.1 Interface ............................................................ 1-2 1.1.2 Controlled Application....................................... 1-2

1.2 Basic Flow of Control .................................................... 1-3 1.2.1 Default Setting .................................................. 1-4 1.2.2 Setting Rx Signal Waveform ............................ 1-6 1.2.3 Setting Parameters ........................................... 1-7 1.2.4 Measuring ......................................................... 1-9

1.3 Native Mode ................................................................ 1-11 1.4 Setting Numeric Program Data ................................... 1-13

Page 10: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 1 Overview

1-2

1.1 Product Overview The MX283087A can be controlled from an external controller (PC) by remote control commands using the MS2830A Signal Analyzer (hereafter referred to as “this instrument”). The remote control commands are defined using the SCPI format and original format.

1.1.1 Interface This instrument has GPIB, Ethernet, and USB interfaces for remote control. Only one of these can be used at once.

The interface is automatically determined when a command is received at the start of communication. This instrument enters the remote state after the interface has been determined. on the front panel goes off in Local status and lights up in Remote status.

Refer to the MS2830A Signal Analyzer manual (Mainframe, Remote Control) for more details of the basic operations for remote control, such as interface setting.

1.1.2 Controlled Application Two types of the remote control commands can be used with this instrument: commands that are commonly applied to this instrument itself or all the applications (hereinafter, referred to as, “Common commands”), and the other commands unique to the applications. Common commands can be executed at any time, regardless of the currently controlled application. However, commands unique to an application can be executed at the controlled application. If it is executed at another application, the command is not executed or an error occurs.

In this instrument, multiple applications can be loaded at the same time. Only one application resource can be executed per piece of hardware at one time. The MX283087A measures by using the resource of RF Input/Output. Thus, the MX283087A cannot be executed at the same time with another application using the same resource such as the signal analyzer function. In order to execute a function unique to the application by using the remote control, you need to select this application while it has been running. Unlike other applications, the MX283087A uses the resources of the MS2830A Option 020/021 Vector Signal Generator (hereafter “Option 020”). Note that the MX283087A and Option 020 cannot be used at the same time.

Page 11: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

1.2 Basic Flow of Control

1-3

1

Overview

1.2 Basic Flow of Control This section describes the basic remote control command programming operations for measuring signals.

Figure 1.2-1 shows the flow of the basic control process.

End

Setting Rx Signal Waveform

Initial Setting

Measurement Condition Changed

Start

Setting Parameters

Execution of TRX/RX Measurement

Figure 1.2-1 Basic Flow of Control

(1) Initial Setting Initialize the communication interface and the parameters, set the communication mode, load applications, and select applications.

1.2.1 Default Setting

(2) Setting Rx Signal Waveform Set signal waveform pattern for Rx test.

1.2.2 Setting Rx Signal Waveform

(3) Setting Parameters Specify parameters applied to the TRX/RX measurement, such as the measurement mode, measurement frequency list, and measurement frequency level.

1.2.3 Setting Parameters

(4) Execution of TRX/RX Measurement Execute measurement end processing.

1.2.4 Measuring

Page 12: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 1 Overview

1-4

1.2.1 Default Setting As part of the initial settings, perform the preparations for using the measuring instrument and the application. The following actions are included in the initial settings:

(1) Initializing Communication Interface The remote control interface to be used is initialized so sending and receiving of commands can start. Refer to the operation manual of the interface used, for details about the remote control interface.

(2) Setting Language Mode and Response Format The language mode and the response format used to communicate are set. Refer to MS2690A/MS2691A/MS2692A and MS2830A Signal Analyzer manual (Mainframe Remote Control) for details.

(3) Starting the application The application is loaded.

(4) Selecting Application The target application is selected.

(5) Initialization All parameters and statuses are reset at initialization.

Page 13: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

1.2 Basic Flow of Control

1-5

1

Overview

Start

Starting application

End

Setting language mode and response format

Initializing communication interface

INST CONFIG

SYST:LANG SCPI

SYST:RES:MODE A

SYST:APPL:LOAD TRXSC

Selecting application

INST:SYST TRXSC

Initialization

*RST

*CLS

Figure 1.2.1-1 Initialization Flow and Command Example

Page 14: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 1 Overview

1-6

1.2.2 Setting Rx Signal Waveform Set up the waveform pattern of the Rx test signal. The waveform pattern set up by the Signal Generator application is used as this pattern. Switch to this application to set up the waveform pattern. For details of the control of Signal Generator, refer to MS2830A Signal Analyzer Vector Signal Generator Operation Manual Remote Control.

(1) Switching the Application Switch to the Signal Generator application.

(2) Setting Rx Signal Waveform Set up the waveform pattern of the Rx signal. This setup can be omitted if an unmodulated wave is used as the Rx signal.

(3) Enabling or Disabling the Modulation Function Enable (turn on) or disable (turn off) modulation.

(4) Switching the Application Switch the MX283087A application.

Start

End

Switching application

INST:SG

Selecting waveform (if necessary)

MMEM:LOAD:WAV <package>,<pattern>

RAD:ARB:WAV <package>,<pattern>

Selecting modulation function (if necessary)

OUTP:MOD ON

Swithcing application

INST:SYST TRXSC

Figure 1.2.2-1 Rx Signal Waveform Setup Flow and Command

Example

Page 15: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

1.2 Basic Flow of Control

1-7

1

Overview

1.2.3 Setting Parameters Specify the measurement mode, common parameters, TX parameters, RX parameters, and other parameters as required. The parameters can be set in any order, unless otherwise specified.

(1) Measurement mode

(2) Common parameters

• Number of sequences

• Number of segments

• Segment length

(3) TX parameters

• Measurement ratio

• Trigger on/off

• Trigger timeout

• Trigger level

• Analysis bandwidth

• Filter

• Frequency list

• Level list

(4) RX parameters

• Frequency list

• Level list

(5) RX trigger sequence parameters

• Number of short burst segments

• Number of time offset segments

• Short burst level

Page 16: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 1 Overview

1-8

Start

End

Measurement mode

SET:TSC:MODE TRX

Common parameters

SET:TSC:FREQ:STEP:COUN 2

SET:TSC:POW:STEP:COUN 3

SET:TSC:POW:STEP:LENG 10

TX parameter (set when in TRX mode)

SET:TSC:RAT 0.8

SET:TSC:TRIG ON

SET:TSC:TIM 10

SET:TSC:TRIG:LEV –30

SET:TSC:FREQ:SPAN 5MHZ

SET:TSC:FILT:TYPE OFF

SET:TSC:TX:FREQ:STEP 869.04MHz,893.97MHz

SET:TSC:TX:POW:STEP 30,20,10

RX parameter

SET:TSC:RX:FREQ:STEP 824.04MHz,848.97MHz

SET:TSC:RX:POW:STEP –10,-40,-60

TX trigger sequence parameter (Rx mode)

SET:TSC:RTS:SBUR:STEP:COUN 2

SET:TSC:RTS:TOFF:STEP:COUN 10

SET:TSC:RTS:SBUR:POW -20

Figure 1.2.3-1 Parameter Setup Flow and Command Example

Page 17: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

1.2 Basic Flow of Control

1-9

1

Overview

1.2.4 Measuring Perform measurement according to the specified settings in each measurement mode.

■TRx measurement mode

(1) Starting Measurement Start TRx measurement and wait for a trigger signal.

(2) Checking Trigger Wait State Make sure that the MX283087A can receive a trigger signal. Next, start test signal output from the DUT.

(3) Reading Measurement Results

Start

End

Starting Measurement

INIT:TSC

Checking Trigger Wait Status

ARM:TSC?

Reading Measurement Results

STAT:ERR?

FETC:TSC?

Figure 1.2.4-1 Measurement Flow (TRX Measurement Mode) and

Command Example

Page 18: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 1 Overview

1-10

■Rx measurement mode

(1) Starting Measurement Start Rx signal transmission.

(2) Ending Measurement Make sure that Rx signal transmission has finished.

Start

End

Starting Measurement

INIT:TSC

Ending Measurement

STAT:ERR?

Figure 1.2.4-2 Measurement Flow (RX Measurement Mode) and

Command Example

Page 19: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

1.3 Native Mode

1-11

1

Overview

1.3 Native Mode This instrument defines the syntax/format types of the remote control commands as “Language mode”. The language mode has two modes: SCPI and Native.

(1) SCPI mode The SCPI mode processes commands conforming to the syntax/format defined in SCPI (ver1999.0). For programming, you can use the character string in long/short form format and can omit angled bracket ( [ ] ) definition character strings. On the Configuration screen, the SCPI mode is automatically set after transmitting SYST:LANG SCPI.

(2) Native mode The Native mode processes commands that are in this instrument’s own definition type. Unless otherwise specified, the character string of a command header is fixed. If application commands are defined only by the SCPI mode, character strings converted according to the conversion rules are the commands in the Native mode.

Note: In the Native mode, you cannot use STATus:QUEStionable and STATus:OPERation registers. Neither can you use them by converting to Native mode according to the conversion rules.

On the Configuration screen, the Native mode is automatically set after transmitting SYST:LANG NAT.

Figure 1.3-1 SCPI Mode and Native Mode

SCPI Mode

Command Definition

AAAAaa:BBBBbb[:CCCCcc]:D|E <n>

Programming examples:

AAAAaa:BBBBbb:CCCCcc:D 0

AAAA:BBBB:CCCC:D 0

AAAA:BBBB:D 0

AAAA:BBBB:CCCC:E 0

Native Mode (initial state)

Command Definition (Original)

VWXYZ1 <n>

Programming examples:

VWXYZ 0

Command Definition (Converted from SCPI)

AAAA:BBBB:D <n>

Programming examples:

AAAA:BBBB:D 0

SYST:LANG SCPI

SYST:LANG NAT

AAAA:BBBB:E 0

Page 20: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 1 Overview

1-12

A part of this application is only defined as SCPI mode commands. You need to follow the conversion rules below in order to control this application by using the Native mode.

The below 1 to 5 indicates the conversion rules.

Conversion Rules

1. Move the numeric parameters in the SCPI mode program header to the head of the arguments. Omit a numeric parameter, which only has one value.

2. Use the first node if multiple ones can be selected.

3. Delete layers that can be deleted.

4. Alter all the long forms into the short ones.

5. Omit the colon (“:”) at the head of the command.

Example 1

To convert :CALCulate:MARKer[1]|2[:SET]:CENTer into a Native mode command.

1. Put a numeric parameter of the program header at the head of the

argument. :CALCulate:MARKer[1]|2[:SET]:CENTer

:CALCulate:MARKer[:SET]:CENTer <integer>

(A numeric value (1 or 2) is put <integer>.)

2. Delete layers that can be deleted.

:CALCulate:MARKer[:SET]:CENTer <integer>

↓ :CALCulate:MARKer:CENTer <integer>

3. Alter all the long forms into the short ones.

:CALCulate:MARKer:CENTer <integer>

↓ :CALC:MARK:CENT <integer>

4. Omit the colon (“:”) at the head of the command.

:CALC:MARK:CENT <integer>

↓ CALC:MARK:CENT <integer>

Page 21: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

1.4 Setting Numeric Program Data

1-13

1

Overview

1.4 Setting Numeric Program Data The following character programs can be used for setting numeric program data (numeric parameter) and are applicable only when using the SCPI mode.

(1) DEFault

After DEFault has been set to numeric program data, the target parameter is set to the default value.

(2) MINimum

After MINimum has been set to numeric program data, the target parameter is set to the minimum value.

(3) Maximum

After MAXimum has been set to numeric program data, the target parameter is set to the maximum value.

In the MX283087A, DEFault, MINimum, and MAXimum can be used for the following parameters.

<numeric_value>

<integer>

Page 22: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 1 Overview

1-14.

Page 23: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-1

2

SC

PI D

evice Message D

etails

This chapter describes the detailed specifications of SCPI remote control commands for executing the functions of the MX 283087A. The device messages are listed according to function. Refer to the MS2690A/MS2691A/MS2692A and MS2830A Signal Analyzer Operation Manual (Mainframe Remote Control) for detailed specifications of the IEEE488.2 common device messages and application common device messages.

2.1 Config Function Device Messages ............................................................................................. 2-3 :SYSTem:APPLication:LOAD <apl_name> ............................................................................... 2-4 :SYSTem:APPLication:UNLoad <apl_name> ........................................................................... 2-5 :SYSTem:APPLication:VERSion? <apl_name> ........................................................................ 2-5

2.2 Application Common Device Messages ..................................................................................... 2-6 :INSTrument[:SELect] SIGANA|SPECT|SG|CONFIG|TRXSC .................................................. 2-8 :INSTrument[:SELect]? .............................................................................................................. 2-9 :INSTrument:SYSTem SIGANA|SPECT|SG|CONFig|TRXSC,[ACTive]|INACtive|MINimum . 2-10 :INSTrument:SYSTem? ........................................................................................................... 2-11 :SETup:TSCalibration:TX:POWer:STEP[:LEVels]

<pow_1>,<pow_2>,<pow_3>,<pow_4>,<pow_5>,<pow_6>,<pow_7>,<pow_8>,

<pow_9>,<pow_10>,<pow_11>,<pow_12>,<pow_13>,<pow_14>,<pow_15>,

<pow_16>,<pow_17>,<pow_18>,<pow_19>,<pow_20>,<pow_21>,<pow_22>,

<pow_23>,<pow_24>,<pow_25>,<pow_26>,<pow_27>,<pow_28>,<pow_29>,

<pow_30>,<pow_31>,<pow_32>,<pow_33>,<pow_34>,<pow_35>,<pow_36>,

<pow_37>,<pow_38>,<pow_39>,<pow_40>, ….. ,<pow_80> .................................... 2-12 :SETup:TSCalibration:TX:POWer:STEP[:LEVels]? ................................................................. 2-13 :SETup:TSCalibration:RX:POWer:STEP[:LEVels]

<pow_1>,<pow_2>,<pow_3>,<pow_4>,<pow_5>,<pow_6>,<pow_7>,<pow_8>,

<pow_9>,<pow_10>,<pow_11>,<pow_12>,<pow_13>,<pow_14>,<pow_15>,

<pow_16>,<pow_17>,<pow_18>,<pow_19>,<pow_20>,<pow_21>,<pow_22>,

<pow_23>,<pow_24>,<pow_25>,<pow_26>,<pow_27>,<pow_28>,<pow_29>,

<pow_30>,<pow_31>,<pow_32>,<pow_33>,<pow_34>,<pow_35>,<pow_36>,

<pow_37>,<pow_38>,<pow_39>,<pow_40>,….,<pow_80> ....................................... 2-14 :SETup:TSCalibration:RX:POWer:STEP[:LEVels]? ................................................................ 2-15 :SETup:TSCalibration:RX:POWer:OFFSet:STATe ON|OFF|1|0 ............................................. 2-16 :SETup:TSCalibration:RX:POWer:OFFSet:STATe? ............................................................... 2-16 :SETup:TSCalibration:RX:POWer:OFFSet

<offset1>,<offset2>,<offset3>,<offset4>,<offset5>,<offset6>,<offset7>,<offset8>,

<offset9>,<offset10>,<offset11>,<offset12>,<offset13>,<offset14>,<offset15>,

<offset16>,<offset17>,<offset18>,<offset19>,<offset20> ............................................ 2-17 :SETup:TSCalibration:RX:POWer:OFFSet? ............................................................................ 2-18 :SETup:TSCalibration:RX:POWer:OFFSet:ERRor? ................................................................ 2-19 :SETup:TSCalibration:TX:FREQuency:STEP[:VALues]

<freq_1>,<freq_2>,<freq_3>,<freq_4>,<freq_5>,<freq_6>,<freq_7>,<freq_8>,

Page 24: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-2

<freq_9>,<freq_10>,<freq_11>,<freq_12>,<freq_13>,<freq_14>,<freq_15>,

<freq_16>,<freq_17>,<freq_18>,<freq_19>,<freq_20> ............................................... 2-20 :SETup:TSCalibration:TX:FREQuency:STEP[:VALues]? ........................................................ 2-21 :SETup:TSCalibration:RX:FREQuency:STEP[:VALues]

<freq_1>,<freq_2>,<freq_3>,<freq_4>,<freq_5>,<freq_6>,<freq_7>,<freq_8>,

<freq_9>,<freq_10>,<freq_11>,<freq_12>,<freq_13>,<freq_14>,<freq_15>,

<freq_16>,<freq_17>,<freq_18>,<freq_19>,<freq_20> ............................................... 2-22 :SETup:TSCalibration:RX:FREQuency:STEP[:VALues]? ....................................................... 2-23 :SETup:TSCalibration:POWer:STEP:COUNt <integer> .......................................................... 2-24 :SETup:TSCalibration:POWer:STEP:COUNt? ........................................................................ 2-24 :SETup:TSCalibration:POWer:STEP:LENGth <integer> ......................................................... 2-25 :SETup:TSCalibration:POWer:STEP:LENGth? ....................................................................... 2-25 :SETup:TSCalibration:FREQuency:STEP:COUNt <integer> .................................................. 2-26 :SETup:TSCalibration:FREQuency:STEP:COUNt? ................................................................ 2-26 :SETup:TSCalibration:TIMeout <integer> ................................................................................ 2-27 :SETup:TSCalibration:TIMeout? .............................................................................................. 2-28 :SETup:TSCalibration:RATio <real> ........................................................................................ 2-29 :SETup:TSCalibration:RATio? ................................................................................................. 2-30 :SETup:TSCalibration:OFFSet <real> ..................................................................................... 2-31 :SETup:TSCalibration:OFFSet? .............................................................................................. 2-32 :SETup:TSCalibration:MODE TRX|RX .................................................................................... 2-33 :SETup:TSCalibration:MODE? ................................................................................................ 2-33 :SETup:TSCalibration:RTSequence:SBURst:STEP:COUNt <integer> ................................... 2-34 :SETup:TSCalibration:RTSequence:SBURst:STEP:COUNt? ................................................. 2-34 :SETup:TSCalibration:RTSequence:TOFFset:STEP:COUNt <integer> ................................. 2-35 :SETup:TSCalibration:RTSequence:TOFFset:STEP:COUNt? ................................................ 2-35 :SETup:TSCalibration:RTSequence:SBURst:POWer <pow> ................................................. 2-36 :SETup:TSCalibration:RTSequence:SBURst:POWer? ........................................................... 2-36 :INITiate[:IMMediate] ................................................................................................................ 2-37 :INITiate:TSCalibration ............................................................................................................. 2-37 :ARM:TSCalibration? ............................................................................................................... 2-38 :FETCh:TSCalibration[1]? ........................................................................................................ 2-39 :SETup:TSCalibration:FREQuency:SPAN <freq> ................................................................... 2-40 :SETup:TSCalibration:FREQuency:SPAN? ............................................................................. 2-40 :SETup:TSCalibration:FILTer:TYPE OFF|RNYQuist ............................................................... 2-41 :SETup:TSCalibration:FILTer:TYPE? ...................................................................................... 2-41 :SETup:TSCalibration:TRIGger[:STATe] ON|OFF|1|0 ............................................................ 2-42 :SETup:TSCalibration:TRIGger[:STATe]? ............................................................................... 2-43 :SETup:TSCalibration:TRIGger:LEVel <level> ........................................................................ 2-43 :SETup:TSCalibration:TRIGger:LEVel? ................................................................................... 2-44

Page 25: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-3

2

SC

PI D

evice Message D

etails

2.1 Config Function Device Messages

Config Function device messages available in this application are shown in Table 2.1-1.

Table 2.1-1 Config Function Device Messages

Function Device Messages

Load Application :SYSTem:APPLication:LOAD <apl_name>

Unload Application

:SYSTem:APPLication:UNLoad <apl_name>

Software Version Query

:SYSTem:APPLication:VERSion? <apl_name>

Page 26: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-4

:SYSTem:APPLication:LOAD <apl_name> Load Application

Function

This command loads an application. Command

:SYSTem:APPLication:LOAD <apl_name>

Parameter

<apl_name> Target application name SIGANA Signal Analyzer SPECT Spectrum Analyzer SG Signal Generator TRXSC TRX Sweep Calibration

Detail

This function loads an installed application and registers it to the Application Switch menu. For parameters to specify an application other than the standard ones, refer to the operation manual (remote control) for each application.

Example of Use

To load the Signal Analyzer application. SYST:APPL:LOAD SIGANA

Page 27: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.1 Config Function Device Messages

2-5

2

SC

PI D

evice Message D

etails

:SYSTem:APPLication:UNLoad <apl_name> Unload Application

Function

This command exits an application. Command

:SYSTem:APPLication:UNLoad <apl_name>

Parameter

<apl_name> Target application name SIGANA Signal Analyzer SPECT Spectrum Analyzer SG Signal Generator TRXSC TRX Sweep Calibration

Detail

This function exits the active application and deletes it from the Application Switch menu. For parameters to specify an application other than the standard ones, refer to the operation manual (remote control) for each application

Example of Use

To exit the Signal Analyzer application. SYST:APPL:UNL SIGANA

:SYSTem:APPLication:VERSion? <apl_name> Software Version Query

Function

This command queries the application version.

Query

:SYSTem:APPLication:VERSion? <apl_name>

Response

version

Parameter

<apl_name> Target application name SIGANA Signal Analyzer SPECT Spectrum Analyzer PNOISE Phase Noise TRXSC TRX Sweep Calibration

Example of Use

To query the version of the Signal Analyzer application. SYST:APPL:VERS? SIGANA

Page 28: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-6

2.2 Application Common Device Messages Application common device messages available in this application are shown in Table 2.2-1.

Table 2.2-1 Application Common Device Messages

Function Device Messages

Application Switch

:INSTrument[:SELect] SIGANA|SPECT|SG|CONFIG|TRXSC

:INSTrument[:SELect]?

Application Status

:INSTrument:SYSTem SIGANA|SPECT|SG|CONFig|TRXSC,[ACTive]|INACtive|MINimum

:INSTrument:SYSTem?

Edit Segment Tx Power Levels

:SETup:TSCalibration:TX:POWer:STEP[:LEVels] <pow_n>

:SETup:TSCalibration:TX:POWer:STEP[:LEVels]?

Edit Segment DUT Rx Power

:SETup:TSCalibration:RX:POWer:STEP[:LEVels] <pow_n>

:SETup:TSCalibration:RX:POWer:STEP[:LEVels]?

RX Level Offset :SETup:TSCalibration:RX:POWer:OFFSet:STATe ON|OFF|1|0

:SETup:TSCalibration:RX:POWer:OFFSet:STATe?

RX Level Offset Value

:SETup:TSCalibration:RX:POWer:OFFSet <offset_n>

:SETup:TSCalibration:RX:POWer:OFFSet?

RX Level Unleveled State

:SETup:TSCalibration:RX:POWer:OFFSet:ERRor? <integrity_indicator>,<sequence>,<segment>

Edit Sequence DUT Tx Frequency

:SETup:TSCalibration:TX:FREQuency:STEP[:VALues]

:SETup:TSCalibration:TX:FREQuency:STEP[:VALues]?

Edit Sequence DUT Rx Frequency

:SETup:TSCalibration:RX:FREQuency:STEP[:VALues]<freq_n>

:SETup:TSCalibration:RX:FREQuency:STEP[:VALues]?

Segment DUT Tx Power Count

:SETup:TSCalibration:POWer:STEP:COUNt <integer>

:SETup:TSCalibration:POWer:STEP:COUNt?

Segment DUT Tx Power Step Time

:SETup:TSCalibration:POWer:STEP:LENGth <integer>

:SETup:TSCalibration:POWer:STEP:LENGth?

Sequence Frequency Count

:SETup:TSCalibration:FREQuency:STEP:COUNt <integer>

:SETup:TSCalibration:FREQuency:STEP:COUNt?

Measure TRX Sweep Calibration Timeout

:SETup:TSCalibration:TIMeout <integer>

:SETup:TSCalibration:TIMeout?

Measure TRX Sweep Calibration Ratio

:SETup:TSCalibration:RATio <real>

:SETup:TSCalibration:RATio?

Page 29: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-7

2

SC

PI D

evice Message D

etails

Table 2.2-1 Application Common Device Messages (Cont’d)

Function Device Messages

TRX Sweep Calibration Time Offset

:SETup:TSCalibration:OFFSet <real>

:SETup:TSCalibration:OFFSet?

TRX Sweep Calibration Mode Select

:SETup:TSCalibration:MODE TRX|RX

:SETup:TSCalibration:MODE?

Short Burst Segment Count

:SETup:TSCalibration:RTSequence:SBURst:STEP:COUNt <integer>

:SETup:TSCalibration:RTSequence:SBURst:STEP:COUNt?

Time Offset Segment Count

:SETup:TSCalibration:RTSequence:TOFFset:STEP:COUNt <integer>

:SETup:TSCalibration:RTSequence:TOFFset:STEP:COUNt?

Short Burst Level :SETup:TSCalibration:RTSequence:SBURst:POWer <pow>

:SETup:TSCalibration:RTSequence:SBURst:POWer?

Initiate :INITiate[:IMMediate]

TRX Sweep Calibration Initiate

:INITiate:TSCalibration

TRX Sweep Calibration Arm Query

:ARM:TSCalibration?

TRX Sweep Calibration Fetch to Read TX Power Results

:FETCh:TSCalibration[1]?

Span Frequency :SETup:TSCalibration:FREQuency:SPAN <freq>

:SETup:TSCalibration:FREQuency:SPAN?

TRX Sweep Calibration Filter Type

:SETup:TSCalibration:FILTer:TYPE OFF|RNYQuist

:SETup:TSCalibration:FILTer:TYPE?

TRX Sweep Calibration Trigger Switch

:SETup:TSCalibration:TRIGger[:STATe] ON|OFF|1|0

:SETup:TSCalibration:TRIGger[:STATe]?

TRX Sweep Calibration Trigger Level

:SETup:TSCalibration:TRIGger:LEVel <level>

:SETup:TSCalibration:TRIGger:LEVel?

Page 30: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-8

:INSTrument[:SELect] SIGANA|SPECT|SG|CONFIG|TRXSC Application Switch

Function

This command switches the target application for operation/control.

Command

:INSTrument[:SELect] <apl_name>

Parameter

<apl_name> Target application name SIGANA Signal Analyzer SPECT Spectrum Analyzer SG Signal Generator CONFIG Config TRXSC TRX Sweep Calibration

Details

This function is used to switch the operation/control target application. An application whose status is CURRENT is a remote control target.

Example of Use

To switch the operation target application to the Signal Analyzer. INST SIGANA

Page 31: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-9

2

SC

PI D

evice Message D

etails

:INSTrument[:SELect]? Application Switch Query

Function

This command queries the reading for the target application for operation/control.

Query

:INSTrument[:SELect]?

Response

<apl_name>

Parameter

<apl_name> Target application name SIGANA Signal Analyzer SPECT Spectrum Analyzer SG Signal Generator CONFIG Config TRXSC TRX Sweep Calibration

Details

This function is used to switch the operation/control target application. An application whose status is CURRENT is a remote control target.

Example of Use

To query the operation target application. INST?

> TRXSC

Page 32: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-10

:INSTrument:SYSTem

SIGANA|SPECT|SG|CONFig|TRXSC,[ACTive]|INACtive|MINimum Application Status

Function

This command sets the status of the application.

Command

:INSTrument:SYSTem <apl_name>,<window>

Parameter

<apl_name> Target application name SIGANA Signal Analyzer SPECT Spectrum Analyzer SG Signal Generator CONFIG Config TRXSC TRX Sweep Calibration

<window> Application window status ACTive Operation enabled (displayed on top screen) INACtive Inactive MINimum Minimized

When omitted Same as ACT Details

This function is used to switch the operation/control target application.

Example of Use

To switch the window status of the Spectrum Analyzer to Active. INST:SYST SPECT,ACT

Page 33: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-11

2

SC

PI D

evice Message D

etails

:INSTrument:SYSTem? Application Status Query

Function

This command queries the status of the application.

Query

:INSTrument:SYSTem? <apl_name>

Response

<status>,<window>

Parameter

<apl_name> Target application name SIGANA Signal Analyzer SPECT Spectrum Analyzer SG Signal Generator CONFIG Config TRXSC TRX Sweep Calibration

<status> Application status CURR Executed and targeted for operation RUN Executed but not targeted for operation IDLE Loaded but not executed UNL Not loaded

<window> Application window status ACT Operation enabled (displayed on top screen) INAC Inactive MIN Minimized NON Not displayed

Details

This function is used to switch the operation/control target application.

Example of Use

To query the reading for the status of the Spectrum Analyzer. INST:SYST? SPECT > CURR,ACT

Page 34: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-12

:SETup:TSCalibration:TX:POWer:STEP[:LEVels]

<pow_1>,<pow_2>,<pow_3>,<pow_4>,<pow_5>,<pow_6>,<pow_7>,<pow_8

>,<pow_9>,<pow_10>,<pow_11>,<pow_12>,<pow_13>,<pow_14>,<pow_15

>,<pow_16>,<pow_17>,<pow_18>,<pow_19>,<pow_20>,<pow_21>,<pow_2

2>,<pow_23>,<pow_24>,<pow_25>,<pow_26>,<pow_27>,<pow_28>,<pow_

29>,<pow_30>,<pow_31>,<pow_32>,<pow_33>,<pow_34>,<pow_35>,<pow

_36>,<pow_37>,<pow_38>,<pow_39>,<pow_40>, ….. ,<pow_80> Edit Segment Tx Power Levels

Function

This command sets signal levels of every segment for TX test at TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:TX:POWer:STEP[:LEVels] <pow_n>

Parameter

<pow_n> Tx Power of segment n n 1 to 80 Range –30.0 dBm to 30.0 dBm Resolution 0.1 dBm

Suffix code DBM

Default 30.0, 28.0, 26.0, 24.0, 22.0, 20.0, 18.0, 16.0, 14.0, 12.0, 10.0, 8.0, 6.0, 4.0, 2.0, 0.0, –2.0, –4.0, –6.0, –8.0, –10.0, –12.0, –14.0, –16.0, –18.0, –20.0, –22.0, –24.0, –26.0, –28.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0, –30.0

Details

The second and the subsequent segments can be omitted

Example of Use

To set input level of every segment at TRX Adjustment measurement. SET:TSC:TX:POW:STEP

30.0,28.0,26.0,24.0,22.0,20.0,18.0,16.0,14.0,12.0,10.0,8

.0,6.0,4.0,2.0,0.0,-2.0,-4.0,-6.0,-8.0,-10.0,-12.0,-14.0

,-16.0,-18.0,-20.0,-22.0,-24.0,-26.0,-28.0,-30.0,-30.0,-

30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0

Page 35: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-13

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:TX:POWer:STEP[:LEVels]? Edit Segment Tx Power Query

Function

This command queries power levels of every segment for TX test at TRX Sweep Calibration measurement.

Query

:SETup:TSCalibration:TX:POWer:STEP[:LEVels]?

Response

<pow_1>,<pow_2>,<pow_3>,..,<pow_80>

Parameter

<pow_n> Tx Power of segment n n 1 to 80 Range –30.0 dBm to 30.0 dBm Resolution 0.1 dBm Suffix code None, Value is returned in dBm units.

Example of Use

To query DUT input level of every segment at TRX Sweep Calibration measurement. SET:TSC:TX:POW:STEP?

>

30.0,28.0,26.0,24.0,22.0,20.0,18.0,16.0,14.0,12.0,10.0,8

.0,6.0,4.0,2.0,0.0,-2.0,-4.0,-6.0,-8.0,-10.0,12.0,-14.0,

-16.0,-18.0,-20.0,-22.0,-24.0,-26.0,-28.0,-30.0,-30.0,-3

0.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.

0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,

-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-3

0.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.

0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,-30.0,

-30.0

Page 36: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-14

:SETup:TSCalibration:RX:POWer:STEP[:LEVels]

<pow_1>,<pow_2>,<pow_3>,<pow_4>,<pow_5>,<pow_6>,<pow_7>,<pow_8

>,<pow_9>,<pow_10>,<pow_11>,<pow_12>,<pow_13>,<pow_14>,<pow_15

>,<pow_16>,<pow_17>,<pow_18>,<pow_19>,<pow_20>,<pow_21>,<pow_2

2>,<pow_23>,<pow_24>,<pow_25>,<pow_26>,<pow_27>,<pow_28>,<pow_

29>,<pow_30>,<pow_31>,<pow_32>,<pow_33>,<pow_34>,<pow_35>,<pow

_36>,<pow_37>,<pow_38>,<pow_39>,<pow_40>,….,<pow_80> Edit Segment DUT Rx Power

Function

This command sets signal levels of every segment for RX test at TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:RX:POWer:STEP[:LEVels] <pow_n>

Parameter

<pow_n> Rx Power of segment n n 1 to 80 Range –120.0 dBm to –5.0 dBm Resolution 0.1 dBm

Suffix code DBM, DM

Default –15.0, –20.0, –25.0, –30.0, –35.0, –40.0, –45.0, –50.0, –55.0, –60.0, –65.0, –70.0, –75.0, –80.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0, –85.0

Details

The second and the subsequent segments can be omitted

Example of Use

To set DUT input level of every segment at TRX Sweep Calibration measurement. SET:TSC:RX:POW:STEP

-120,-115,-110,-105,-100,-95,-90,-85,-80,-75,-70,-65,-60

,-55,-50,-45,-40,-35,-30

Page 37: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-15

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:RX:POWer:STEP[:LEVels]? Edit Segment DUT Rx Power Query

Function

This command queries DUT input level of every segment at TRX Sweep Calibration measurement (during Rx test).

Query

:SETup:TSCalibration:RX:POWer:STEP[:LEVels]?

Response

<pow_1>,<pow_2>,<pow_3>,..,<pow_80>

Parameter

<pow_n> DUT Rx Power of segment n n 1 to 80 Range –120.0 dBm to –5.0 dBm Resolution 0.1 dBm Suffix code None, Value is returned in dBm units.

Example of Use

To query DUT input level of every segment at TRX Sweep Calibration measurement. SET:TSC:RX:POW:STEP?

>

–15.0,–20.0,–25.0,–30.0,–35.0,–40.0,–45.0,–50.0,–55.0,–6

0.0,–65.0,–70.0,–75.0,–80.0,–85.0,–85.0,–85.0,–85.0,–85.

0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,

–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–8

5.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.

0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,

–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–8

5.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.0,–85.

0,–85.0,–85.0,–85.0,–85.0,–85.0

Page 38: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-16

:SETup:TSCalibration:RX:POWer:OFFSet:STATe ON|OFF|1|0 RX Level Offset

Function

This command enables(On)/disables(Off) the RX Level Offset function.

Command

:SETup:TSCalibration:RX:POWer:OFFSet:STATe <switch>

Parameter

<switch> Enables/disables the RX Level Offset function OFF|0 Disabled ON|1 Enabled Default OFF

Example of Use

To enable the RX Level Offset function. SET:TSC:RX:POW:OFFS:STAT ON

:SETup:TSCalibration:RX:POWer:OFFSet:STATe? RX Level Offset Query

Function

This command queries the state of the RX Level Offset function.

Query

:SETup:TSCalibration:RX:POWer:OFFSet:STATe?

Response

<switch>

Parameter

<switch> State of the RX Level Offset function (Whether enabled or disabled)

0 Disabled 1 Enabled

Example of Use

To query the state of RX Level Offset function. SET:TSC:RX:POW:OFFS:STAT?

> 1

Page 39: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-17

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:RX:POWer:OFFSet

<offset1>,<offset2>,<offset3>,<offset4>,<offset5>,<offset6>,<offset7>,<offset

8>,<offset9>,<offset10>,<offset11>,<offset12>,<offset13>,<offset14>,<offset1

5>,<offset16>,<offset17>,<offset18>,<offset19>,<offset20> RX Level Offset Value

Function

This command sets the offset value for signal levels of every segment at TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:RX:POWer:OFFSet <Offset_n>

Parameter

<offset_n> Output level offset value for each RX frequency n 1 to 20 Range –100.00 dB to +100.00 dB Resolution 0.01 dB

Default 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00, 0.00

Details

Each <offset_n> corresponds to the frequency <freq_n> set by the :SETup:TSCalibration:RX:FREQuency:STEP[:VALues] command.

Example of Use

To set the Output level offset value for each RX frequency. SET:TSC:RX:POW:OFFS

10.00,10.00,10.00,10.00,10.00,10.00,10.00,10.00,10.00,10

.00,10.00,10.00,10.00,10.00,10.00,10.00,10.00,10.00,10.0

0,10.00

Page 40: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-18

:SETup:TSCalibration:RX:POWer:OFFSet? RX Level Offset Value Query

Function

This command queries the the offset value for signal levels of every segment at TRX Sweep Calibration measurement.

Query

:SETup:TSCalibration:RX:POWer:OFFSet?

Response

<offset_1>,<offset_2>,..,<offset_20>

Parameter

<offset_n> Output level offset value for each RX frequency n 1 to 20 Range –100.00 dB to +100.00 dB Resolution 0.01 dB

Details

Each <offset[n]> corresponds to the frequency <freq_[n]> set by the :SETup:TSCalibration:RX:FREQuency:STEP[:VALues] command.

Example of Use

To query the Output level offset value for each RX frequency. SET:TSC:RX:POW:OFFS?

>

10.00,10.00,10.00,10.00,10.00,10.00,10.00,10.00,10.00,10

.00,10.00,10.00,10.00,10.00,10.00,10.00,10.00,10.00,10.0

0,10.00

Page 41: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-19

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:RX:POWer:OFFSet:ERRor? RX Level Unleveled State Query

Function

This command checks whether the Segment DUT Rx Power setting falls within the range between the upper and lower limits due to the RX Level Offset setting.

Query

:SETup:TSCalibration:RX:POWer:OFFSet:ERRor?

Response

<integrity_indicator>,<sequence>,<segment>

Parameter

<integrity_indicator> indicator 0 Not unleveled 1 Unleveled

<sequence> First sequence number including unleveled segment

Range 0 to 20 0 if there is no unleveled segment

<segment> First unleveled segment number Range 0 to 80

0 if there is no unleveled segment

Details

If the Segment DUT Rx Power setting is out of the upper and lower limits though the RX Level Offset is set, then the value is rounded within the upper and lower limits before outputting.

Example of Use

To query the RX Level Unleveled State. SET:TSC:RX:POW:OFFS:ERR?

> 0,0,0

Page 42: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-20

:SETup:TSCalibration:TX:FREQuency:STEP[:VALues]

<freq_1>,<freq_2>,<freq_3>,<freq_4>,<freq_5>,<freq_6>,<freq_7>,<freq_8>,

<freq_9>,<freq_10>,<freq_11>,<freq_12>,<freq_13>,<freq_14>,<freq_15>,<f

req_16>,<freq_17>,<freq_18>,<freq_19>,<freq_20> Edit Sequence DUT Tx Frequency

Function

This command sets DUT Tx frequency of every segment at TRX Sweep Calibration measurement (during Tx test).

Command

:SETup:TSCalibration:TX:FREQuency:STEP[:VALues] <freq_n>

Parameter

<freq_n> DUT Tx Frequency of sequence n

n 1 to 20 Range 400 MHz to 3500 MHz Resolution 1 Hz

Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ

Hz is used when omitted. Default 1853.0MHz, 1853.2MHz, 1853.4MHz,

1853.6MHz, 1853.8MHz, 1854.0MHz, 1854.2MHz, 1854.4MHz, 1854.6MHz, 1854.8MHz, 1855.0MHz, 1855.2MHz, 1855.4MHz, 1855.6MHz, 1855.8MHz, 1856.0MHz, 1856.2MHz, 1856.4MHz, 1856.6MHz, 1856.8MHz

Details

The second and the subsequent segments can be omitted

Example of Use

To set DUT Tx frequency of every segment at TRX Sweep Calibration measurement SET:TSC:TX:FREQ:STEP

400MHZ,450MHZ,500MHZ,550MHZ,600MHZ,650MHZ,700MHZ,750MHZ,

800MHZ,850MHZ,900MHZ,950MHZ,1000MHZ,1050MHZ,1100MHZ,1150

MHZ,1200MHZ,1250MHZ,1300MHZ,1350MHZ

Page 43: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-21

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:TX:FREQuency:STEP[:VALues]? Edit Sequence DUT Tx Frequency Query

Function

This command queries DUT Tx frequency of every segment at TRX Sweep Calibration measurement (during Tx test).

Query :SETup:TSCalibration:TX:FREQuency:STEP[:VALues]?

Response

<freq_1>,<freq_2>,..,<freq_20>

Parameter

<freq_n> DUT Tx Frequency of sequence n

n 1 to 20 Range 400 MHz to 3500 MHz Resolution 1 Hz Suffix code None, Value is returned in Hz units.

Example of Use

To query DUT Tx frequency of every segment at TRX Sweep Calibration measurement SET:TSC:TX:FREQ:STEP?

>

400000000,450000000,500000000,550000000,600000000,650000

000,700000000,750000000,800000000,850000000,900000000,95

0000000,1000000000,1050000000,1100000000,1150000000,1200

000000,1250000000,1300000000,1350000000

Page 44: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-22

:SETup:TSCalibration:RX:FREQuency:STEP[:VALues]

<freq_1>,<freq_2>,<freq_3>,<freq_4>,<freq_5>,<freq_6>,<freq_7>,<freq_8>,

<freq_9>,<freq_10>,<freq_11>,<freq_12>,<freq_13>,<freq_14>,<freq_15>,<f

req_16>,<freq_17>,<freq_18>,<freq_19>,<freq_20> Edit Sequence DUT Rx Frequency

Function

This command sets DUT Rx frequency of every segment at TRX Sweep Calibration measurement (during Rx test).

Command

:SETup:TSCalibration:RX:FREQuency:STEP[:VALues] <freq_n>

Parameter

<freq_n> DUT Rx Frequency of sequence n n 1 to 20 Range 400 MHz to 3500 MHz Resolution 1 Hz

Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ

Hz is used when omitted. Default 1933.0MHz, 1933.2MHz, 1933.4MHz,

1933.6MHz, 1933.8MHz, 1934.0MHz, 1934.2MHz, 1934.4MHz, 1934.6MHz, 1934.8MHz, 1935.0MHz, 1935.2MHz, 1935.4MHz, 1935.6MHz, 1935.8MHz, 1936.0MHz, 1936.2MHz, 1936.4MHz, 1936.6MHz, 1936.8MHz

Details

The second and the subsequent segments can be omitted

Example of Use

To set DUT Rx frequency of every segment at TRX Sweep Calibration measurement SET:TSC:RX:FREQ:STEP

400MHZ,450MHZ,500MHZ,550MHZ,600MHZ,650MHZ,700MHZ,750MHZ,

800MHZ,850MHZ,900MHZ,950MHZ,1000MHZ,1050MHZ,1100MHZ,1150

MHZ,1200MHZ,1250MHZ,1300MHZ,1350MHZ

Page 45: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-23

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:RX:FREQuency:STEP[:VALues]? Edit Sequence DUT Rx Frequency Query

Function

This command queries DUT Rx frequency of every segment at TRX Sweep Calibration measurement (during Rx test).

Query :SETup:TSCalibration:RX:FREQuency:STEP[:VALues]?

Response

<freq_1>,<freq_2>,..,<freq_20>

Parameter

<freq_n> DUT Rx Frequency of sequence n

n 1 to 20 Range 400 MHz to 3500 MHz Resolution 1 Hz Suffix code None, Value is returned in Hz units.

Example of Use

To query DUT Rx frequency of every segment at TRX Sweep Calibration measurement SET:TSC:RX:FREQ:STEP?

>

400000000,450000000,500000000,550000000,600000000,650000

000,700000000,750000000,800000000,850000000,900000000,95

0000000,1000000000,1050000000,1100000000,1150000000,1200

000000,1250000000,1300000000,1350000000

Page 46: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-24

:SETup:TSCalibration:POWer:STEP:COUNt <integer> Segment DUT Tx Power Count

Function

This command sets number of segments at TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:POWer:STEP:COUNt <integer>

Parameter

<integer> Measured number of segments

Range 1 to 80 Resolution 1 Suffix code None Default 40

Example of Use

To set number of segments to 40 at TRX Sweep Calibration measurement. SET:TSC:POW:STEP:COUN 40

:SETup:TSCalibration:POWer:STEP:COUNt? Segment DUT Tx Power Count Query

Function

This command queries number of segments at TRX Sweep Calibration measurement.

Query

:SETup:TSCalibration:POWer:STEP:COUNt?

Response

<integer>

Parameter

<integer> Measured number of segments Range 1 to 80 Resolution 1 Suffix code None

Example of Use

To query number of segments at TRX Sweep Calibration measurement. SET:TSC:POW:STEP:COUN?

> 40

Page 47: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-25

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:POWer:STEP:LENGth <integer> Segment DUT Tx Power Step Time

Function

This command sets the duration of segments at TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:POWer:STEP:LENGth <integer> <t>

Parameter

<integer> Segment Duration 10 10 ms 20 20 ms (Default)

Example of Use

To set the duration of segments to 10 ms at TRX Sweep Calibration measurement . SET:TSC:POW:STEP:LENG 10

:SETup:TSCalibration:POWer:STEP:LENGth? Segment DUT Tx Power Step Time Query

Function

This command queries the duration of segments at TRX Sweep Calibration measurement.

Query

:SETup:TSCalibration:POWer:STEP:LENGth?

Response

<integer>

Parameter

<integer> Segment Duration

Range 10, 20 Suffix code None, Value is returned in ms units.

Example of Use

To query the duration of segments at TRX Sweep Calibration measurement SET:TSC:POW:STEP:LENG?

> 10

Page 48: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-26

:SETup:TSCalibration:FREQuency:STEP:COUNt <integer> Sequence Frequency Count

Function

This command sets number of sequences for TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:FREQuency:STEP:COUNt <integer>

Parameter

<integer> Measured number of sequences

Range 1 to 20 Resolution 1 Suffix code None Default 20

Example of Use

To set number of sequences to 20 for TRX Sweep Calibration measurement. SET:TSC:FREQ:STEP:COUN 20

:SETup:TSCalibration:FREQuency:STEP:COUNt? Sequence Frequency Count Query

Function

This command queries number of sequences for TRX Sweep Calibration measurement.

Query

:SETup:TSCalibration:FREQuency:STEP:COUNt?

Response

<integer>

Parameter

<integer> Measured number of sequences

Range 1 to 20 Resolution 1 Suffix code None

Example of Use

To query number of sequences for TRX Sweep Calibration measurement. SET:TSC:FREQ:STEP:COUN?

> 20

Page 49: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-27

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:TIMeout <integer> Measure TRX Sweep Calibration Timeout

Function

This command sets the timeout from start of measurement to trigger detection at TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:TIMeout <integer>

Parameter

<integer> The timeout from start of measurement to trigger detection at TRX Sweep Calibration measurement

Range 1 s to 30 s Resolution 1 s

Suffix code NS, US, MS, S

S is used when omitted. Default 5 s

Example of Use

To set the timeout from start of measurement to trigger detection to 10 s at TRX Sweep Calibration measurement. SET:TSC:TIM 10

Page 50: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-28

:SETup:TSCalibration:TIMeout? Measure TRX Sweep Calibration Timeout Query

Function

This command queries the timeout from start of measurement to trigger detection at TRX Sweep Calibration measurement.

Query

:SETup:TSCalibration:TIMeout?

Response

<integer>

Parameter

<integer> The timeout from start of measurement to trigger detection at TRX Sweep Calibration measurement

Range 1 s to 30 s Resolution 1 s Suffix code None, Value is returned in S units.

Example of Use

To query the timeout from start of measurement to trigger detection at TRX Sweep Calibration measurement. SET:TSC:TIM?

> 10

Page 51: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-29

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:RATio <real> Measure TRX Sweep Calibration Ratio

Function

This command sets the power measurement range of each segment at TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:RATio <real>

Parameter

<real> The power measurement range of each segment at TRX Sweep Calibration measurement

Range 0.20 to 0.90 Resolution 0.01 (ratio of which 1.00 is one segment) Suffix code None Default 0.50

Details

This changes the power measurement range of each segment at TRx Sweep Calibration measurement. The range is centered in the middle of the segment.

Example of Use

To set the power measurement range of each segment to 0.60 at TRX Sweep Calibration measurement. SET:TSC:RAT 0.60

Page 52: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-30

:SETup:TSCalibration:RATio? TRX Sweep Calibration Ratio Query

Function

This command queries the power measurement range of each segment at TRX Sweep Calibration measurement.

Query

:SETup:TSCalibration:RATio?

Response

<real>

Parameter

<real> The power measurement range of each segment at TRX Sweep Calibration measurement

Range 0.20 to 0.90 Resolution 0.01 (ratio of which 1.00 is one segment) Suffix code None

Example of Use

To query the power measurement range of each segment at TRX Sweep Calibration measurement. SET:TSC:RAT?

> 0.50

Page 53: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-31

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:OFFSet <real> TRX Sweep Calibration Time Offset

Function

This command sets the time offset between the measurement trigger and the measurement segment boundary for TX measurement time.

Command

:SETup:TSCalibration:OFFSet <real>

Parameter

<real> The time offset as the ratio of the segment duration

Range 0.00 to 0.05 Resolution 0.01 (ratio of which 1.00 is one segment) Suffix code None Default 0.00

Details

When the TRX Sweep Calibration measurement is executed, the TX measurement is delayed by time offset.

Example of Use

To set the time offset to 0.05 at TRX Sweep Calibration measurement. SET:TSC:OFFS 0.05

Page 54: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-32

:SETup:TSCalibration:OFFSet? TRX Sweep Calibration Time Offset Query

Function

This command queries the time offset between the measurement trigger and the measurement segment boundary for TX measurement.

Query

:SETup:TSCalibration:OFFSet?

Response

<real>

Parameter

<real> The time offset as the ratio of the segment duration

Range 0.00 to 0.05 Resolution 0.01 (ratio of which 1.00 is one segment) Suffix code None

Example of Use

To query the time offset at TRX Sweep Calibration measurement. SET:TSC:OFFS?

>0.05

Page 55: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-33

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:MODE TRX|RX TRX Sweep Calibration Mode Select

Function

This command switches the measurement mode between TRX and RX.

Command

:SETup:TSCalibration:MODE <MODE>

Parameter

<MODE> Measurement mode TRX Sets to TRX. RX Sets to RX.

Default TRX

Details

Set to RX if doing RX measurement using trigger sequence.

Example of Use

To set the RX Measurement SET:TSC:MOD RX

:SETup:TSCalibration:MODE? TRX Sweep Calibration Mode Select Query

Function

This command queries the measurement mode.

Query

:SETup:TSCalibration:MODE?

Response

<MODE>

Parameter

<MODE> Measurement mode TRX TRX measurement RX RX measurement

Example of Use

To query the Measurement Mode SET:TSC:MODE?

> RX

Page 56: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-34

:SETup:TSCalibration:RTSequence:SBURst:STEP:COUNt <integer> Short Burst Segment Count

Function

This command sets the Short Burst Segment counts of trigger sequence, in RX mode.

Command

:SETup:TSCalibration:RTSequence:SBURst:STEP:COUNt

<integer>

Parameter

<integer> Short Burst Segment count

Range 1 to 100 Resolution 1

Default 1 Details

Sets Short Burst Segment count. Available when RX mode is selected.

Example of Use

To set number of segment to 5 for Short Burst at RX Measurement Mode. SET:TSC:RTS:SBUR:STEP:COUN 5

:SETup:TSCalibration:RTSequence:SBURst:STEP:COUNt? Short Burst Segment Count Query

Function

This command queries the Short Burst Segment counts of trigger sequence, in RX mode.

Query

:SETup:TSCalibration:RTSequence:SBURst:STEP:COUNt?

Response

<integer>

Parameter

<integer> Short Burst Segment count

Range 1 to 100 Resolution 1

Example of Use

To query the number of segment for short burst SET:TSC:RTS:SBUR:STEP:COUN?

> 5

Page 57: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-35

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:RTSequence:TOFFset:STEP:COUNt <integer> Time Offset Segment Count

Function

This command sets the Time Offset Segment counts of trigger sequence, in RX mode.

Command

:SETup:TSCalibration:RTSequence:TOFFset:STEP:COUNt

<integer>

Parameter

<integer> Time Offset Segment count

Range 1 to 100 Resolution 1

Default 1 Details

Sets Time Offset Segment count. Available when RX mode is selected.

Example of Use

To set number of segment to 5 for Time Offset at RX Measurement Mode. SET:TSC:RTS:TOFF:STEP:COUN 5

:SETup:TSCalibration:RTSequence:TOFFset:STEP:COUNt? Time Offset Segment Count Query

Function

This command queries the Time Offset Segment counts of trigger sequence, in RX mode.

Query

:SETup:TSCalibration:RTSequence:TOFFset:STEP:COUNt?

Response

<integer>

Parameter

<integer> Time Offset Segment count

Range 1 to 100 Resolution 1

Example of Use

To query the number of segment for Time Offset SET:TSC:RTS:TOFF:STEP:COUN?

> 5

Page 58: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-36

:SETup:TSCalibration:RTSequence:SBURst:POWer <pow> Short Burst Level

Function

This command sets the Short Burst transmission power of trigger sequence, in RX mode.

Command

:SETup:TSCalibration:RTSequence:SBURst:POWer <pow>

Parameter

<pow> Rx Power of Short Burst

Range –120.0 dBm to –5.0 dBm Resolution 0.1 dBm

Suffix code DBM, DM

Default –15.0 dBm

Details

Sets the Short Burst power that the instrument transmits in RX mode.

Example of Use

To set the power of Short Burst to –20 dBm at RX Measurement Mode. SET:TSC:RTS:SBUR:POW –20

:SETup:TSCalibration:RTSequence:SBURst:POWer? Short Burst Level Query

Function

This command queries the Short Burst transmission power of trigger sequence, in RX mode.

Query

:SETup:TSCalibration:RTSequence:SBURst:POWer?

Response

<pow>

Parameter

<pow> Rx Power of Short Burst

Range –120.0 dBm to –5.0 dBm Resolution 0.1 dBm Suffix code None, Value is returned in dBm units.

Example of Use

To query the power of short burst SET:TSC:RTS:SBUR:POW?

> -20.0

Page 59: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-37

2

SC

PI D

evice Message D

etails

:INITiate[:IMMediate] Initiate

Function

This command starts the measurement.

Command

:INITiate[:IMMediate]

Details

In this application program, this command becomes the same operation as :INITiate:TSCalibration.

Example of Use

To start measurement INIT

:INITiate:TSCalibration TRX Sweep Calibration Initiate

Function

This command starts the measurement.

Command

:INITiate:TSCalibration

Details

Starts measurement according to the specified measurement mode.

Example of Use

To start measurement INIT:TSC

Page 60: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-38

:ARM:TSCalibration? TRX Sweep Calibration Arm Query

Function

This command queries the status of measurement preparation at TRX Sweep Calibration measurement.

Query

:ARM:TSCalibration?

Response

<status>

Parameter

<status> The status of measurement preparation at TRX Sweep Calibration measurement

0 Not Ready 1 Ready

Example of Use

To query the status of measurement preparation at TRX Sweep Calibration measurement. ARM:TSC?

> 1

Page 61: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-39

2

SC

PI D

evice Message D

etails

:FETCh:TSCalibration[1]? TRX Sweep Calibration Fetch to Read TX Power Results

Function

This command queries the result for TRX Sweep Calibration measurement.

Query

:FETCh:TSCalibration[1]?

Response

<level_n>

Parameter

<level_n> Power level of segment n

Range of n Multiplier of :SET:TSC:POW:STEP:COUN parameter and :SET:TSC:FREQ:STEP:COUN parameter

Resolution 0.01 dBm Suffix code None, Value is returned in dBm units.

Details

The power level of the tuning segment is excluded from the result.

Example of Use

To query the result for TRX Sweep Calibration measurement. FETC:TSC?

> 10.00,11.00,12.00,13.00,14.00,15.00

Page 62: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-40

:SETup:TSCalibration:FREQuency:SPAN <freq> Span Frequency

Function

This command sets the span frequency.

Command

:SETup:TSCalibration:FREQuency:SPAN <freq>

Parameter

<freq> Span frequency Range/Resolution 2500000, 5000000, 10000000, 25000000

Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ Hz is used when omitted.

Default 25 MHz

Example of Use

To set the span frequency to 5 MHz. SET:TSC:FREQ:SPAN 5MHZ

:SETup:TSCalibration:FREQuency:SPAN? Span Frequency Query

Function

This command queries the span frequency.

Query

:SETup:TSCalibration:FREQuency:SPAN?

Response

<freq>

Parameter

<freq> Span frequency

Range/Resolution 2500000, 5000000, 10000000, 25000000 Suffix code None, Value is returned in Hz units.

Example of Use

To query the span frequency SET:TSC:FREQ:SPAN?

> 5000000

Page 63: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-41

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:FILTer:TYPE OFF|RNYQuist TRX Sweep Calibration Filter Type

Function

This command sets filter type for TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:FILTer:TYPE <filter>

Parameter

<filter> Filter type OFF No filtering (Default) RNYQuist Root Nyquist filter

Details

When the setting of the frequency span is only 5 MHz, the route nyquist filter is effective.

When the setting of the frequency span is not 5 MHz, the setting of Route Nyquist filter is treated as no filtering though the parameter change is possible.

Example of Use

To set the filter type to Root Nyquist. SET:TSC:FILT:TYPE RNYQ

:SETup:TSCalibration:FILTer:TYPE? TRX Sweep Calibration Filter Type Query

Function

This command queries the filter type for TRX Sweep Calibration measurement.

Query

:SETup:TSCalibration:FILTer:TYPE?

Response

<filter>

Parameter

<filter> Filter type OFF No filtering RNYQ Root Nyquist filter

Example of Use

To query the filter type for TRX Sweep Calibration measurement. SET:TSC:FILT:TYPE?

> RNYQ

Page 64: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-42

:SETup:TSCalibration:TRIGger[:STATe] ON|OFF|1|0 TRX Sweep Calibration Trigger Switch

Function

This command sets the measurement trigger status to On/Off for TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:TRIGger[:STATe] <switch>

Parameter

<switch> The status of measurement trigger for TRX Sweep Calibration measurement

OFF|0 Trigger OFF ON|1 Trigger ON

Default ON

Details

When the trigger is OFF, the measurement starts immediately after initiate command. When the trigger is ON, the measurement waits for the TX signal power which is above the trigger level after initiate command.

Example of Use

To set the trigger to OFF for TRX Sweep Calibration measurement. SET:TSC:TRIG OFF

Page 65: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

2.2 Application Common Device Messages

2-43

2

SC

PI D

evice Message D

etails

:SETup:TSCalibration:TRIGger[:STATe]? TRX Sweep Calibration Trigger Switch Query

Function

This command queries the measurement trigger status for TRX Sweep Calibration measurement.

Query

:SETup:TSCalibration:TRIGger[:STATe]?

Response

<switch>

Parameter

<switch> The status of measurement trigger for TRX Sweep Calibration measurement

0 Trigger OFF 1 Trigger ON

Example of Use

To query the trigger status for TRX Sweep Calibration measurement. SET:TSC:TRIG?

> 1

:SETup:TSCalibration:TRIGger:LEVel <level> TRX Sweep Calibration Trigger Level

Function

This command sets the external trigger level for TRX Sweep Calibration measurement.

Command

:SETup:TSCalibration:TRIGger:LEVel <level>

Parameter

<level> The external trigger level for TRX Sweep Calibration measurement

Range –30 to –10 dB Resolution 1 dB

Suffix code DB

DB is used when omitted. Default –30 dB

Example of Use

To set the external trigger level to –10 dB for TRX Sweep Calibration measurement. SET:TSC:TRIG:LEV -10

Page 66: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 2 SCPI Device Message Details

2-44.

:SETup:TSCalibration:TRIGger:LEVel? TRX Sweep Calibration Trigger Level Query

Function

This command queries the external trigger level for TRX Sweep Calibration measurement.

Query

:SETup:TSCalibration:TRIGger:LEVel?

Response

<level>

Parameter

<level> The external trigger level for TRX Sweep Calibration measurement

Range –30 to –10 dB Resolution 1 dB Suffix code None, Value is returned in DB units.

Example of Use

To query the external trigger level for TRX Sweep Calibration measurement. SET:TSC:TRIG:LEV?

> -10

Page 67: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

3

SC

PI S

tatus R

egister

Chapter 3 SCPI Status Register

3-1

This chapter explains the SCPI commands used to read the state of

the application and the status register.

3.1 Reading Measurement Status ...................................... 3-2 :STATus:ERRor? .......................................................... 3-2

3.2 STATus:QUEStionable Register ................................... 3-3 :STATus:QUEStionable[:EVENt]? ................................ 3-5 :STATus:QUEStionable:CONDition? ........................... 3-5 :STATus:QUEStionable:ENABle <integer>.................. 3-6 :STATus:QUEStionable:ENABle? ................................ 3-6 :STATus:QUEStionable:NTRansition <integer> .......... 3-7 :STATus:QUEStionable:NTRansition? ......................... 3-7 :STATus:QUEStionable:PTRansition <integer> .......... 3-8 :STATus:QUEStionable:PTRansition? ......................... 3-8 :STATus:QUEStionable:MEASure[:EVENt]? ............... 3-9 :STATus:QUEStionable:MEASure:CONDition? ........... 3-9 :STATus:QUEStionable:MEASure:ENABle <integer> 3-10 :STATus:QUEStionable:MEASure:ENABle? ............. 3-10 :STATus:QUEStionable:MEASure:NTRansition

<integer> ........................................................ 3-11 :STATus:QUEStionable:MEASure:NTRansition? ...... 3-11 :STATus:QUEStionable:MEASure:PTRansition

<integer> ........................................................ 3-12 :STATus:QUEStionable:MEASure:PTRansition? ...... 3-12

3.3 STATus:OPERation Register ..................................... 3-13 :STATus:OPERation[:EVENt]? .................................. 3-14 :STATus:OPERation:CONDition? .............................. 3-14 :STATus:OPERation:ENABle <integer> .................... 3-15 :STATus:OPERation:ENABle? ................................... 3-15 :STATus:OPERation:NTRansition <integer> ............. 3-16 :STATus:OPERation:NTRansition? ........................... 3-16 :STATus:OPERation:PTRansition <integer> ............. 3-17 :STATus:OPERation:PTRansition? ............................ 3-17

Page 68: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 3 SCPI Status Register

3-2

3.1 Reading Measurement Status :STATus:ERRor? Measurement Status Error Query

Function

This command queries a measurement error.

Query

:STATus:ERRor?

Response

<status>

Parameter

<status> Measurement Status

Value = bit0 + bit1 + bit2 + bit3 + bit4 + bit5 + bit6 + bit7 + bit8 + bit9 + bit10 + bit11 + bit12 + bit13 + bit14 + bit15

bit0: 20 = 1 No measurement bit1: 21 = 2 Level Over bit2: 22 = 4 Timeout bit3: 23 = 8 (Not Used) bit4: 24 = 16 (Not Used) bit5: 25 = 32 (Not Used) bit6: 26 = 64 (Not Used) bit7: 27 = 128 (Not Used) bit8: 28 = 256 (Not Used) bit9: 29 = 512 (Not Used) bit10: 210 = 1024 (Not Used) bit11: 211 = 2048 (Not Used) bit12: 212 = 4096 (Not Used) bit13: 213 = 8192 (Not Used) bit14: 214 = 16384 (Not Used) bit15: 215 = 32768 (Not Used)

Range 0 to 65535 Details

0 is returned at normal termination.

Usage Example

To query a measurement error. STAT:ERR?

> 0

Page 69: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

3

SC

PI S

tatus R

egister

3.2 STATus:QUEStionable Register

3-3

3.2 STATus:QUEStionable Register The hierarchical structure of the QUEStionable Status register is described in Figures 3.2-1 and 3.2-2, and Tables 3.2-1 and 3.2-2.

VOLTage (NOT USED) DB0 CURRent (NOT USED) DB1 TIME (NOT USED) DB2 POWer (NOT USED) DB3 TEMPerature (NOT USED) DB4 FREQuency DB5 PHASe (NOT USED) DB6 MODulation (NOT USED) DB7 CALibration (NOT USED) DB8 MEASure DB9 NOT USED DB10NOT USED DB11NOT USED DB12INSTrument (NOT USED) DB13Command Warning (NOT USED) DB14NOT USED DB15

Status Byte Register DB3

Figure 3.2-1 QUEStionable Status Register

Table 3.2-1 Bit Definition of QUEStionable Status Register

Bit Definition

DB5 Reference Clock Unlock DB9 QUEStionable Measure Register Summary

NOT USED DB0 NOT USED DB1 NOT USED DB2 NOT USED DB3 NOT USED DB4 Level Over DB5 NOT USED DB6 NOT USED DB7 Signal Abnormal DB8 NOT USED DB9 NOT USED DB10NOT USED DB11NOT USED DB12NOT USED DB13NOT USED DB14NOT USED DB15

QUEStionable Status Register DB9

Figure 3.2-2 QUEStionable Measure Register

Table 3.2-2 Bit Definition of QUEStionable Measure Register

Bit Definition

DB2 Timeout DB5 Level Over

Page 70: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 3 SCPI Status Register

3-4

Table 3.2-3 lists the device messages for the QUEStionable Status register.

Table 3.2-3 Device Messages for QUEStionable Status Register

Function Device Message

Questionable Status Register Event

:STATus:QUEStionable[:EVENt]?

Questionable Status Register Condition

:STATus:QUEStionable:CONDition?

Questionable Status Register Enable

:STATus:QUEStionable:ENABle <integer>

:STATus:QUEStionable:ENABle?

Questionable Status Register Negative Transition

:STATus:QUEStionable:NTRansition <integer>

:STATus:QUEStionable:NTRansition?

Questionable Status Register Positive Transition

:STATus:QUEStionable:PTRansition <integer>

:STATus:QUEStionable:PTRansition?

Questionable Measure Register Event

:STATus:QUEStionable:MEASure[:EVENt]?

Questionable Measure Register Condition

:STATus:QUEStionable:MEASure:CONDition?

Questionable Measure Register Enable

:STATus:QUEStionable:MEASure:ENABle <integer>

:STATus:QUEStionable:MEASure:ENABle?

Questionable Measure Register Negative Transition

:STATus:QUEStionable:MEASure:NTRansition <integer>

:STATus:QUEStionable:MEASure:NTRansition?

Questionable Measure Register Positive Transition

:STATus:QUEStionable:MEASure:PTRansition <integer>

:STATus:QUEStionable:MEASure:PTRansition?

Page 71: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

3

SC

PI S

tatus R

egister

3.2 STATus:QUEStionable Register

3-5

:STATus:QUEStionable[:EVENt]? Questionable Status Register Event

Function

This command queries Event register of QUEStionable Status register.

Query

:STATus:QUEStionable[:EVENt]?

Response

<integer>

Parameter

<integer> Bit Sum Total of Event Register Resolution 1 Range 0 to 65535

Usage Example

To query event register of QUEStionable Status register. STAT:QUES?

> 0

:STATus:QUEStionable:CONDition? Questionable Status Register Condition

Function

This command queries Condition register of QUEStionable Status register.

Query

:STATus:QUEStionable:CONDition?

Response

<integer>

Parameter

<integer> Bit Sum Total of Condition Register Resolution 1 Range 0 to 65535

Usage Example

To query Condition register of QUEStionable Status register. STAT:QUES:COND?

> 0

Page 72: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 3 SCPI Status Register

3-6

:STATus:QUEStionable:ENABle <integer> Questionable Status Register Enable

Function

This command sets Event Enable register of QUEStionable Status register.

Command

:STATus:QUEStionable:ENABle <integer>

Parameter

<integer> Bit Sum Total of Event Enable Register Resolution 1 Range 0 to 65535

Usage Example

To set value of Event Enable register of QUEStionable Status register to 16. STAT:QUES:ENAB 16

:STATus:QUEStionable:ENABle? Questionable Status Register Enable Query

Function

This command queries Event Enable register of QUEStionable Status register.

Query

:STATus:QUEStionable:ENABle?

Response

<integer>

Parameter

<integer> Bit Sum Total of Event Enable Register Resolution 1 Range 0 to 65535

Usage Example

To query Event Enable register of QUEStionable Status register. STAT:QUES:ENAB?

> 16

Page 73: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

3

SC

PI S

tatus R

egister

3.2 STATus:QUEStionable Register

3-7

:STATus:QUEStionable:NTRansition <integer> Questionable Status Register Negative Transition

Function

This command sets transition filter (Negative Transition) of QUEStionable Status register.

Command

:STATus:QUEStionable:NTRansition <integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Negative Transition)

Resolution 1 Range 0 to 65535

Usage Example

To set transition filter (Negative Transition) of QUEStionable Status register to 16. STAT:QUES:NTR 16

:STATus:QUEStionable:NTRansition? Questionable Status Register Negative Transition Query

Function

This command queries transition filter (Negative Transition) of QUEStionable Status register.

Query

:STATus:QUEStionable:NTRansition?

Response

<integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Negative Transition) Resolution 1 Range 0 to 65535

Usage Example

To query transition filter (Negative Transition) of QUEStionable Status register. STAT:QUES:NTR?

> 16

Page 74: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 3 SCPI Status Register

3-8

:STATus:QUEStionable:PTRansition <integer> Questionable Status Register Positive Transition

Function

This command sets transition filter (Positive Transition) of QUEStionable Status register.

Command

:STATus:QUEStionable:PTRansition <integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Positive Transition)

Resolution 1 Range 0 to 65535

Usage Example

To set transition filter (Positive Transition) of QUEStionable Status register to 16. STAT:QUES:PTR 16

:STATus:QUEStionable:PTRansition? Questionable Status Register Positive Transition Query

Function

This command queries transition filter (Positive Transition) of QUEStionable Status register.

Query

:STATus:QUEStionable:PTRansition?

Response

<integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Positive Transition)

Resolution 1 Range 0 to 65535

Usage Example

To query transition filter (Positive Transition) of QUEStionable Status register. STAT:QUES:PTR?

> 16

Page 75: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

3

SC

PI S

tatus R

egister

3.2 STATus:QUEStionable Register

3-9

:STATus:QUEStionable:MEASure[:EVENt]? Questionable Measure Register Event

Function

This command queries Event register of QUEStionable Measure register.

Query

:STATus:QUEStionable:MEASure[:EVENt]?

Response

<integer>

Parameter

<integer> Bit Sum Total of Event Register Resolution 1 Range 0 to 65535

Usage Example

To query Event register of QUEStionable Measure register. STAT:QUES:MEAS?

> 0

:STATus:QUEStionable:MEASure:CONDition? Questionable Measure Register Condition

Function

This command queries Condition register of QUEStionable Measure register.

Query

:STATus:QUEStionable:MEASure:CONDition?

Response

<integer>

Parameter

<integer> Bit Sum Total of Condition Register Resolution 1 Range 0 to 65535

Usage Example

To query Condition register of QUEStionable Measure register. STAT:QUES:MEAS:COND?

> 0

Page 76: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 3 SCPI Status Register

3-10

:STATus:QUEStionable:MEASure:ENABle <integer> Questionable Measure Register Enable

Function

This command sets Event Enable register of QUEStionable Measure register.

Command

:STATus:QUEStionable:MEASure:ENABle <integer>

Parameter

<integer> Bit Sum Total of Event Enable Register Resolution 1 Range 0 to 65535

Usage Example

To set a value of Event Enable register of QUEStionable Measure register to 16. STAT:QUES:MEAS:ENAB 16

:STATus:QUEStionable:MEASure:ENABle? Questionable Measure Register Enable Query

Function

This command queries Event Enable register of QUEStionable Measure register.

Query

:STATus:QUEStionable:MEASure:ENABle?

Response

<integer>

Parameter

<integer> Bit Sum Total of Event Enable Register Resolution 1 Range 0 to 65535

Usage Example

To query Event Enable register of QUEStionable Measure register. STAT:QUES:MEAS:ENAB?

> 16

Page 77: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

3

SC

PI S

tatus R

egister

3.2 STATus:QUEStionable Register

3-11

:STATus:QUEStionable:MEASure:NTRansition <integer> Questionable Measure Register Negative Transition

Function

This command sets transition filter (Negative Transition) of QUEStionable Measure register.

Command

:STATus:QUEStionable:MEASure:NTRansition <integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Negative Transition)

Resolution 1 Range 0 to 65535

Usage Example

To set transition filter (Negative Transition) of QUEStionable Measure register to 16. STAT:QUES:MEAS:NTR 16

:STATus:QUEStionable:MEASure:NTRansition? Questionable Measure Register Negative Transition Query

Function

This command queries transition filter (Negative Transition) of QUEStionable Measure register.

Query

:STATus:QUEStionable:MEASure:NTRansition?

Response

<integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Negative Transition)

Resolution 1 Range 0 to 65535

Usage Example

To query transition filter (Negative Transition) of QUEStionable Measure register. STAT:QUES:MEAS:NTR?

> 16

Page 78: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 3 SCPI Status Register

3-12

:STATus:QUEStionable:MEASure:PTRansition <integer> Questionable Measure Register Positive Transition

Function

This command sets transition filter (Positive Transition) of QUEStionable Measure register.

Command

:STATus:QUEStionable:MEASure:PTRansition <integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Positive Transition)

Resolution 1 Range 0 to 65535

Usage Example

To set transition filter (Positive Transition) of QUEStionable Measure register to 16. STAT:QUES:MEAS:PTR 16

:STATus:QUEStionable:MEASure:PTRansition? Questionable Measure Register Positive Transition Query

Function

This command queries transition filter (Positive Transition) of QUEStionable Measure register.

Query

:STATus:QUEStionable:MEASure:PTRansition?

Response

<integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Positive Transition)

Resolution 1 Range 0 to 65535

Usage Example

To query transition filter (Positive Transition) of QUEStionable Measure register. STAT:QUES:MEAS:PTR?

> 16

Page 79: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

3

SC

PI S

tatus R

egister

3.3 STATus:OPERation Register

3-13

3.3 STATus:OPERation Register The hierarchical structure of the OPERation Status register is described in Figure 3.3-1 and Table 3.3-1.

CALibrating DB0 SETTling DB1 RANGing (NOT USED) DB2 SWEeping (NOT USED) DB3 MEASuring DB4 Waiting for TRIG DB5 Waiting for ARM (NOT USED) DB6 CORRecting (NOT USED) DB7 FILE Operating DB8 NOT USED DB9 NOT USED DB10NOT USED DB11NOT USED DB12INSTrument (NOT USED) DB13PROGram (NOT USED) DB14NOT USED DB15

Status Byte Register DB7

Figure 3.3-1 OPERation Status Register

Table 3.3-1 Bit Definition for OPERation Status Register

Bit Definition

DB0 CAL Executed DB1 Warm-up displayed DB4 Capture executed (Always 1 at Continuous

measurement)

DB5 Waiting for trigger signal DB8 Operating on file

Table 3.3-2 lists the device messages for the OPERation status register.

Table 3.3-2 Device Messages for OPERation Status Register

Function Device Message

Operation Status Register Event :STATus:OPERation[:EVENt]?

Operation Status Register Condition :STATus:OPERation:CONDition?

Operation Status Register Enable :STATus:OPERation:ENABle <integer>

:STATus:OPERation:ENABle?

Operation Status Register Negative Transition

:STATus:OPERation:NTRansition <integer>

:STATus:OPERation:NTRansition?

Operation Status Register Positive Transition

:STATus:OPERation:PTRansition <integer>

:STATus:OPERation:PTRansition?

Page 80: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 3 SCPI Status Register

3-14

:STATus:OPERation[:EVENt]? Operation Status Register Event

Function

This command queries Event register of OPERation Status register.

Query

:STATus:OPERation[:EVENt]?

Response

<integer>

Parameter

<integer> Bit Sum Total of Event Register Resolution 1 Range 0 to 65535

Usage Example

To query Event register of OPERation Status register. STAT:OPER?

> 0

:STATus:OPERation:CONDition? Operation Status Register Condition

Function

This command queries Event register of OPERation Condition register.

Query

:STATus:OPERation:CONDition?

Response

<integer>

Parameter

<integer> Bit Sum Total of Condition Register Resolution 1 Range 0 to 65535

Usage Example

To query Event register of OPERation Condition register. STAT:OPER:COND?

> 0

Page 81: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

3

SC

PI S

tatus R

egister

3.3 STATus:OPERation Register

3-15

:STATus:OPERation:ENABle <integer> Operation Status Register Enable

Function

This command sets Event Enable register of OPERation Status register.

Command

:STATus:OPERation:ENABle <integer>

Parameter

<integer> Bit Sum Total of Event Enable Register Resolution 1 Range 0 to 65535

Usage Example

To set Event Enable register of OPERation Status register to 16. STAT:OPER:ENAB 16

:STATus:OPERation:ENABle? Operation Status Register Enable Query

Function

This command queries Event Enable register of OPERation Status register.

Query

:STATus:OPERation:ENABle?

Response

<integer>

Parameter

<integer> Bit Sum Total of Event Enable Register Resolution 1 Range 0 to 65535

Usage Example

To query Event Enable register of OPERation Status register. STAT:OPER:ENAB?

> 16

Page 82: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 3 SCPI Status Register

3-16

:STATus:OPERation:NTRansition <integer> Operation Status Register Negative Transition

Function

This command sets transition filter (Negative Transition) of OPERation Status register.

Command

:STATus:OPERation:NTRansition <integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Negative Transition)

Resolution 1 Range 0 to 65535

Usage Example

To set transition filter (Negative Transition) of OPERation Status register to 16. STAT:OPER:NTR 16

:STATus:OPERation:NTRansition? Operation Status Register Negative Transition Query

Function

This command queries transition filter (Negative Transition) of OPERation Status register.

Query

:STATus:OPERation:NTRansition?

Response

<integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Negative Transition)

Resolution 1 Range 0 to 65535

Usage Example

To query transition filter (Negative Transition) of OPERation Status register. STAT:OPER:NTR?

> 16

Page 83: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

3

SC

PI S

tatus R

egister

3.3 STATus:OPERation Register

3-17

:STATus:OPERation:PTRansition <integer> Operation Status Register Positive Transition

Function

This command sets transition filter (Positive Transition) of OPERation Status register.

Command

:STATus:OPERation:PTRansition <integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Positive Transition)

Resolution 1 Range 0 to 65535

Usage Example

To set transition filter (Positive Transition) of OPERation Status register to 16. STAT:OPER:PTR 16

:STATus:OPERation:PTRansition? Operation Status Register Positive Transition Query

Function

This command queries transition filter (Positive Transition) of OPERation Status register.

Query

:STATus:OPERation:PTRansition?

Response

<integer>

Parameter

<integer> Bit Sum Total of Transition Filter (Positive Transition)

Resolution 1 Range 0 to 65535

Usage Example

To query transition filter (Positive Transition) of OPERation Status register.

STAT:OPER:PTR?

> 16

Page 84: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Chapter 3 SCPI Status Register

3-18.

Page 85: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Appendix A Programming Sample

A-1

Ap

pen

dix

Appen

dix A

Select a waveform pattern for RX test on Signal Analyzer Application

Command Description

INST SG Switches the active application to SG MMEM:LOAD:WAV? “CDMA”, “TEST” Queries the status of waveform pattern. > 1 "TEST" waveform is available MMEM:LOAD:WAV “CDMA”, “TEST” Loads the waveform pattern from the hard

disk to the waveform memory. *OPC? Queries the operation completion status. > 1 The waveform patterns have been loaded. RAD:ARB:WAV “CDMA”, “TEST” Selects the waveform pattern to be played. RAD:ARB:WAV:REST Plays the waveform pattern

TRX Sweep Calibration Setting

Command Description

INST TRXSC Switches the active application to TRX Sweep Calibration.

SET:TSC:FREQ:SPAN 10MHZ Setting frequency span SET:TSC:TRIG:LEV –20DB Setting the trigger level SET:TSC:TX:FREQ:STEP 825.03MHZ, 837.00MHZ, 848.97MHZ

Setting the frequency list of each sequence for TX measurement

SET:TSC:RX:FREQ:STEP 870.03MHZ, 882.00MHZ, 893.97MHZ

Setting the frequency list of each sequence for RX measurement

SET:TSC:RX:POW:STEP –55, -40, -40, -30, -30, -25, -25, -22, -22

Setting the power levels list of each segment for RX measurement

SET:TSC:TX:POW:STEP 30, 26, 22, 18, 14, 10, 6, 2, -2

Setting the power levels list of each segment for TX measurement

SET:TSC:POW:STEP:COUN 9 Setting the number of segments SET:TSC:FREQ:STEP:COUN 3 Setting the number of frequency sequence SET:TSC:TIM 3 Setting the TRX Sweep Calibration's timeout SET:TSC:RAT 0.60 Setting the TX power measurement window

ratio

Page 86: MX283087A TRX Sweep Calibration Operation Manual Remote ... · Document No.: M-W3449AE-2.0 . ANRITSU CORPORATION . MX283087A TRX Sweep Calibration . Operation Manual . Remote Control

Appendix A Programming Sample

A-2.

Executing TRX measurement

Command Description

INIT:TSC Initiates the TRX Sweep Calibration measurement.

ARM:TSC? Queries the ready status of TRX Sweep Calibration.

> 0 Not ready ARM:TSC? Re-queries until responding '1'. > 1 Ready (Orders the DUT to begin transmission of its TX Frequency/Power Series.) STAT:ERR? Queries the measurement status > 1 Measurement in progress STAT:ERR? Re-queries until responding other than '1'. > 0 Normal termination FETC:TSC? Queries the TRX Sweep Calibration

measurement result. > 30.00, 28.00, 26.00, 24.00, 22.00, 20.00, 18.00, 16.00, 14.00, 30.01, 28.01, 26.01, 24.01, 22.01, 20.01, 18.01, 16.01, 14.01, 30.02, 28.02, 26.02, 24.02, 22.02, 20.02, 18.02, 16.02, 14.02

The results are returned to the time series without tuning segment. (The number of Tx Power measurement results is 27, as the number of segment is 9 and the number of sequence is 3.)