Top Banner
MULTITRIP – General Multi Use Accurate & Reliable – Proven Temprecord performance. Applications – Monitoring storage and transit of pharmaceuticals, perishable foods, temperature-sensitive Industrial products, etc… Applications also include food processing, monitoring temperature-sensitive Industrial equipment, tuning and validating HVAC systems… 8K non-volatile memory Calibrated temperature accuracy is 0.2°C / 0.35°F (Logger Case temperature range is -25°C to +60°C or -13°F to +140°F) 3-Point Calibration built into logger: • superior accuracy guarantee • no need for corrections or complex calculations to be applied to downloaded data • avoids human data entry errors, providing data integrity and peace of mind Calibrated and corrected at 3-Points to ISO/IEC 17025 Calibration Laboratory Standards • traceable Calibration Certificate included Temperature Units: °C or °F Display Resolution: 0.01°C or 0.01°F TRW software: • Hi-speed download • Serial RS-232 or USB connectivity • compatible with Windows XP, Vista and Windows 7 Compliance made easy with automated PDF report generation. Use speed-buttons from TRW software to : • save as *.TR and/or *.PDF files • export to MS Excel – part or all samples • email • print Save time with Auto Mode (for batch programming): • Start / Stop / Status display • download/read recorded data • save downloaded file as *.TR or *.PDF • print file • email file • reuse logger Fully welded food grade plastic case • FDA approved • recyclable Case protection rating: waterproof to IP67 or NEMA 6 TM
2

MULTITRIP – General Multi Use

Feb 21, 2022

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: MULTITRIP – General Multi Use

MULTITRIP – General Multi Use

➤ Accurate & Reliable – Proven Temprecord performance.

➤ Applications – Monitoring storage and transit ofpharmaceuticals, perishable foods, temperature-sensitiveIndustrial products, etc… Applications also include foodprocessing, monitoring temperature-sensitive Industrialequipment, tuning and validating HVAC systems…

➤ 8K non-volatile memory

➤ Calibrated temperature accuracy is 0.2°C / 0.35°F(Logger Case temperature range is -25°C to +60°Cor -13°F to +140°F)

➤ 3-Point Calibration built into logger:• superior accuracy guarantee• no need for corrections or complex calculations to be

applied to downloaded data• avoids human data entry errors, providing data integrity

and peace of mind

➤ Calibrated and corrected at 3-Points to ISO/IEC 17025Calibration Laboratory Standards• traceable Calibration Certificate included

➤ Temperature Units: °C or °F

➤ Display Resolution: 0.01°C or 0.01°F

➤ TRW software:• Hi-speed download• Serial RS-232 or USB connectivity• compatible with Windows XP, Vista and Windows 7

➤ Compliance made easy with automated PDFreport generation. Use speed-buttons from TRWsoftware to :• save as *.TR and/or *.PDF files• export to MS Excel – part or all samples• email

• print

➤ Save time with Auto Mode (for batchprogramming):• Start / Stop / Status display• download/read recorded data• save downloaded file as *.TR or *.PDF• print file• email file• reuse logger

➤ Fully welded food grade plastic case• FDA approved

• recyclable

➤ Case protection rating:waterproof to IP67 or

NEMA 6

TM

Page 2: MULTITRIP – General Multi Use

Multiple Use Calibrated Data LoggersUse the Multitrip data logger time and time again. After each use, download recorded data, reuse and the data logger is ready for its next application

Accuracy – Calibrated to 0.2°C / 0.35°FMultitrip data loggers go through a 3-Point calibration process during manufacturing. Traceable Calibration Certificates are included with each logger from our ISO/IEC 17025 accredited laboratory. PDF copies of Calibration Certificates are also downloadable from the web site at: www.temprecord.com/certificates.html

8K Non-volatile MemoryThe Multitrip data logger supports 8K of memory (7,935 samples). User-configurable sample period provides from 4.4 hours up to 2 years of data recording.

LED Status Indicators‘Within limits’ (Green) LED Indicator and ‘Out of limits’ (Red) LED Indicator allow instant confirmation of recorded data being within or outside of specified upper and lower temperature range.

Push Button Start/Stop or MarkThe Push Button (when enabled) allows User to ‘Start’ or ‘Stop’ the data logger anywhere without needing sophisticated software or hardware in the field. During recording, indication ‘Markers’ can be included for future reference by pressing the button for two seconds until the red LED flashes two times. If button disabled then the logger can be started or stopped using Temprecord for Windows (TRW) software – available FREE!

Four External Probe OptionsChoose from four different external probe options including Stainless Steel Straight probe, Bent probe, Teflon probe and Plastic Handle probe.

Push ButtonIntegrated Push Button feature allows User to ‘Start’ or ‘Stop’ data logger in the field. Also allows for the setting of indication ‘Markers’ during data recording for future reference. Functionality of the Push Button can be enabled or disabled through TRW software.

Retaining HoleFully moulded and ultrasonically welded retaining hole for easy securing of the data logger within different environments.

LED Status IndicatorsThe ‘Out of limits’ (Red) LED Indicator and ‘Within limits’ (Green) LED Indicator allow instant confirmation of recorded data being within the specified temperature range. Software setup allows the selection of range limits and the number of samples out of range to activate the alarm feature.

* Accuracy is calibrated with a traceable reference thermometer to 0.2˚C (0.35˚F) at temperatures -15˚C, 0˚C and +40˚C

Parameter SpecificationsModel Multitrip Green – Multi Use Temperature Data LoggerMemory 8K Non-volatileSample Capacity Approx. 7,935 samples minimum

Sample Period Programmable between 2 seconds and 36 hours in 2 second increments

Start Delay Programmable between 10 seconds and 36 hours

Start and Stop Logger Button (if enabled), Palm PDA/Reader, TRW Software/Reader

Recording Duration 4.4 hours to 2 yearsOperating Temperature Range of Case -25˚C ~ +60˚C, -13˚F ~ +140˚FOperating Temperature Range of Data Logger

-25˚C ~ +60˚C, -13˚F ~ +140˚F

Operating Temperature Range with STANDARD Probe

-50˚C ~ +110˚C, -58˚F ~ +230˚F

*Temperature Accuracy 0.2˚C / 0.35˚F over the range -20˚C / -4˚F to +50˚C /+122˚F else 0.5˚C / 0.9˚F

Display Resolution 0.01˚C or 0.01˚FWithin Limits Indicator One flash from Green LED Indicator every 10 secondsOut of Limits Indicator One flash from Red LED Indicator every 10 secondsBattery Type Manganese Dioxide Lithium (3.0V)Dimensions 129 x 73.5 x 14mm or 51/8 x 27/8 x 5/8 inchesWeight 81g or 2.85ozCase Colour Green

Case and Protection Rating IP67 or NEMA 6. Fully welded FDA approved food grade casing

Warranty Period 12 months on data logger from date of purchase 2 years on battery

Temprecord International Limited

PO Box 58 430, Botany, Auckland 2163, New Zealand.Phone: (09) 274 9825, Fax: (09) 273 4020Email: [email protected]: www.temprecord.com

60

50

40

30

20

10

0

-10

-20

-30

MULTITRIP GENERAL USE DATA LOGGERS

Case Protection RatingWaterproof to IP67 or NEMA 6 rated. Ultrasonically welded plastic case is durable in extreme conditions. FDA approved food grade casing.

˚C

Teflon Probe

Straight Probe

Plastic Handle Probe

Bent Probe

TM