Top Banner
Protection from MR hazard Dr Mohit Goel JRIII, 10/7/14
28
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: MR hazard, protection

Protection from MR hazard

Dr Mohit GoelJRIII, 10/7/14

Page 2: MR hazard, protection

American Society for Testing and Materials (ASTM) published a standard in 2005 for the marking of devices brought into the MR ENVIRONMENT, which includes new safety definitions.

Definitions from ASTM International standard

Page 3: MR hazard, protection
Page 4: MR hazard, protection

Static magnetic field biological effectsmechanical effects

Gradient Field induced currents (PNS)auditory damage

RF Field RF power depositionBurns

Safety issues centre around the 3 electromagnetic fields and issues indirectly related

Page 5: MR hazard, protection

Static magnetic field

1.5 T is 30,000 x the strength of earth’s magnetic field

Measured in Gauss or Tesla (10,000G equivalent to 1T)

Page 6: MR hazard, protection

Static Magnetic Fields

1. Biological effects (potential risk)- exposure to static magnetic fields of up to 4T are not thoughtto be harmful

Biological effects relevant to clinical imaging- distorted ECG (magnetohydrodynamic effect)- consider prudency with pregnancy

2. Mechanical effects (very real risk)- translational or attractive forces on metallic objects when brought into the field

A superconducting magnet is always switched on!

Page 7: MR hazard, protection

Static Magnetic FieldBiological effects

Magnetohydrodynamic effect – augmented T wave

Outside field

0.5 Tesla

1.5T

Page 8: MR hazard, protection

Magnetohydrodynamic effect seen as augmentation of T-wave

• Caused by the effect of the static magnetic field on moving blood (systole) as a conducting fluid.

• The gradient and RF fields also affect the configuration of the ECG

• Morphological ECG changes are therefore difficult to detect and diagnose, but rhythm is usually recognised

• Any concern regarding rhythm, remove patient from scanner and perform 12 lead ECG

Page 9: MR hazard, protection

Static Magnetic FieldPregnancy

Patients• 1st trimester –avoid MR where possible• 2nd and 3rd trimester – decision made on a risk versus benefit

determination. For example if it avoids the patient being subjected to x-rays.

Health Care Workers• May enter MR scanning room regardless of trimester• Should not remain in the room when scanner is operational,

avoiding exposure to gradient and radiofrequency fields

Page 10: MR hazard, protection

Static Magnetic FieldMechanical effects

• Projectile or missile effect- the attractive forces exerted by the static magnetic field present the greatest potential for patient injury

- objects will be pulled out of hands, pockets etc, and fly into magnet which has caused injury and death.

• Effect on ferromagnetic implants - electro-mechanical eg pacemakers

- biomedical eg valves, stents

Page 11: MR hazard, protection

What is typically ferromagnetic?

EQUIPMENT PERSONAL ITEMS (leave outside)

• Oxygen cylinders Keys, pens• Wheelchairs Bleeps• Trolleys Mobile phones• IV stands Coins• Monitoring equipment Stethoscopes• Ventilators Scissors

It is easy to forget objects, particularly when responding to an emergency!

Page 12: MR hazard, protection

Never be complacent – accidents do happen!

Oxygen cylinder Infusion pump

Page 13: MR hazard, protection

Floor buffer

Page 14: MR hazard, protection

• Cerebral aneurysm clips• Metallic foreign body in the eye• Shrapnel, bullets (in critical area)

• Ocular implants (containing metal)

• Swan-Ganz

“Contraindications” to MR*

Implants & metal Electromechanical implants

• Pacemakers /ICD’s• Pacing wires• Cochlear implants• Neurostimulators• Hydrocephalus shunts

Any device electrically or mechanically activated* In some circumstances MR has been performed despite “contraindications”, especially in specialist centres. Work is in progress to make some devices safe. For example, some cerebral aneurysm clips may now be scanned.

Page 15: MR hazard, protection

Implants

Cerebral clips• modern clips are considered safe (titanium, elgiloy)• obtain operation notes with serial number of clip• take patient consent

Foreign bodies• Maybe situated near vascular or nervous tissue• If in doubt – x-ray

Page 16: MR hazard, protection

Safety Checklist – comprehensive but concise

• Removal of accessories - watch, jewellery (except wedding rings), body piercing rings, hearing aids, glasses, false teeth, artificial limbs and prostheses

• Removal of clothes containing metal eg zips, bras

Important to know• Previous heart surgery?• Diabetic or epileptic?• Asthmatic or allergies? (in relation to contrast)• Tattoos or permanent eye liner

(iron oxide)?

Essential to know• Cardiac pacemaker?• Previous neurosurgery?• Implants or metal in the body?• Pregnant (prudent approach)?• Drug patch with foil backing?

Essential preparation

Page 17: MR hazard, protection

Preparation of patient• Remove watches / jewellery except gold wedding rings• Remove hearing aids, false teeth, glasses, prostheses• Remove all clothes except socks and underpants• Patient gowns - no pockets

- no metallic fastenings

- ¾ length sleeves for IV access

- wrap round for easy chest access• Brass changing room keys• Screen all accompanying personnel• Check any suspicious item with small bar magnet

Page 18: MR hazard, protection

Gradient fields - Induced Currents

Gradient fields induce an electric field and thus a current in the patient, potentially this can be of sufficient intensity in modern systems to produce a physiological response

- peripheral nerve stimulation (PNS)- cardiac stimulation is not considered possible

Page 19: MR hazard, protection

Hearing protection mandatory above 90dBtime averaged for:

• patients• staff remaining in the scanner room• relatives accompanying children or patients

Gradient fields – Auditory damage

Page 20: MR hazard, protection

Radiofrequency (RF) fields

Thermogenic effects - health & safety concern

Physiological tissue heating response

• most of the transmitted RF power is transferred into heatwithin the patient’s tissue

• all MR systems have safety thresholds to avoiddangerous levels

• Patients with compromised thermoregularory systems areat greatest risk

Page 21: MR hazard, protection

Specific Absorption Rate (SAR)

SAR is the RF power absorbed per unit mass of tissue (expressed in W/kg)

• complex function of numerous variables

• calculated by software from the average forward power passing into the RF transmitter coil and the body mass situated in the RF transmitting field

Therefore an accurate patient weight is vital

• SAR increases 9 fold from 0.5T to 1.5T

Page 22: MR hazard, protection

Radiofrequency (RF) fields

Potential for burns

• 1°, 2°, 3° burns have occurred in the past in patients undergoing MRI

• This is a result of excessive heat developing in the devices or objects

• ECG system is often the culprit• Interventional MRI poses greater risk

Page 23: MR hazard, protection

Prevention of burns

• Electrodes - carbon fibre studs- placed close together

• ECG leads - carbon fibre- fibre optic- high impedance- short as possible (plaited if

necessary)• All conductive leads should be placed in a linear fashion

coming out of bore of scanner• Avoid crossed limbs where possible

Page 24: MR hazard, protection

RF Burn from non-Carbon Electrode

Page 25: MR hazard, protection

QuenchCryogens maintain the magnetic field - helium

Quenching refers to the events that occur when the liquid cryogens that cool the magnet coils boil off rapidly, which results in helium escaping very rapidly from the cryogen bath. This means that the coils cease to be superconducting and become resistive. A quench will in general be accompanied by a loud bang or thundering or hissing or rushing sound with the cold gas expulsion.

Causes - physical - human error (accidental) - intervention (elective)

Effects - if no ventilation is present, pressure build up - asphyxiation/frostbite

Page 26: MR hazard, protection

Elective Quenching

The magnet should only be quenched in two situations:-

• If someone is trapped to the scanner by a ferromagnetic object and is injured and/or distressed (eg O2 cylinder, piece of equipment)

• If there is a fire in the immediate vicinity on order to reduce risk to the Fire Brigade

Page 27: MR hazard, protection

Action to be taken in the event of a Quench

• Evacuate the room as quickly as possible• Ensure the door is kept open during evacuation• Close door after evacuation• If trapped in room stay close to floor level• Seek the advice of a senior physicist immediately• Call scanner engineer

Page 28: MR hazard, protection

This could be you!

Thank you