8285 p. 1/10 www.burkert.com Modular analysis transmitter for pH/ORP, conductivity measurement The 8285 modular process analysis system is designed to measure and process liquid analy- sis parameters. The base unit contents the power supply, signal outputs, binary inputs and the front with graphic display with blacklight. Three slots are available, which depending on the applications can be occupied with modules for pH, conductivity or also with a module with additional outputs. The hygienic, polished stainless steel enclosure version allows application in the field of bio- technology, food processing, and in the phar- maceutical industry. Further applications in the chemical industry, environmental engineering, water and waste-water treatment, and for ap- plication in power plants are also possible. Technical data - Base unit Type 8221 Conductivity sensor Type 8200 pH sensor Type 8201 Enamel pH sensor Type 2031 On/Off Diaphragm valve • High flexibility provided by modular concept for several measuring parameters • Outstanding features for highly demanding applications • Simple and intuitive user interface supported by a large high-resolution graphic display • Compatible with most common pH and conductivity sensor types Type 8285 can be combined with... General data Mounting Wall, post/pipe, panel mounting, sealed against panel Materials Housing, cover Vision panel / Screws / Glands Stainless steel, polished 1.4305 Polycarbonate / Stainless steel / PA Weight Approx. 3.2 kg + approx. 150 g per module Display 1) LC graphic display, white blacklight 240 x 160 pixels resolution; German, English, French, Italian, Spanish, Swedish languages Keypad NAMUR keypad, individual keys, no double assign- ments [meas] [menu][ ] [ ] [ ] [ ] [enter] [softkey1] [softkey2], NAMUR LEDs red and green Logbook Storage capacity standard Extended logbook (option Item no. 558 083) Recording of function activations, appearance and disappearance of warning and failure messages, with date and time Approx. 50 entries, without SMARTMEDIA ® Card read on display, recording on SMARTMEDIA ® card > 50 000 entries, depending on free memory of SMARTMEDIA ® card Measurement recorder (option Item no. 558 083) Recording medium Recording capacity Recording Recording method Time base Zoom function 2-channel measurement recorder with marking of events (failure, maintenance request, function check, limit values) SMARTMEDIA ® card > 50 000 entries, depending on free memory of SMARTMEDIA ® card Process variables and span selectable Snapshot, min/max or mean value, average 10 s...10 h/pixel 10 fold zoom in the event of high rate of change Sensor monitor Direct display of measured values from sensor for validation KI recorder (option Item no. 558 074) Adaptive representation of process flow with monitoring and signaling of critical process parameters 1) Caution! Never expose the display to direct sun light! Only operate the display within the temperature range of 0 up to 50°C max.
10
Embed
Modular analysis transmitter for pH/ORP, conductivity ... · Sensocheck® Automatic monitoring of glass and reference electrode, message can be switched off Sensoface® provides information
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
• High flexibility provided by modular concept for several measuring parameters
• Outstanding features for highly demanding applications
• Simple and intuitive user interface supported by a large high-resolution graphic display
• Compatible with most common pH and conductivity sensor typesType 8285 can be combined with...
General data
Mounting Wall, post/pipe, panel mounting, sealed against panel
Materials
Housing, cover Vision panel / Screws / Glands
Stainless steel, polished 1.4305Polycarbonate / Stainless steel / PA
Weight Approx. 3.2 kg + approx. 150 g per module
Display1) LC graphic display, white blacklight240 x 160 pixels resolution; German, English, French, Italian, Spanish, Swedish languages
Keypad NAMUR keypad, individual keys, no double assign-ments [meas] [menu][ ] [ ] [ ] [ ] [enter] [softkey1] [softkey2], NAMUR LEDs red and green
Logbook
Storage capacity standard
Extended logbook (option Item no. 558 083)
Recording of function activations, appearance and disappearance of warning and failure messages, with date and timeApprox. 50 entries, without SMARTMEDIA® Card read on display, recording on SMARTMEDIA® card> 50 000 entries, depending on free memory of SMARTMEDIA® card
Measurement recorder
(option Item no. 558 083)
Recording medium Recording capacity
Recording Recording method Time base Zoom function
2-channel measurement recorder with marking of events (failure, maintenance request, function check, limit values)
SMARTMEDIA® card> 50 000 entries, depending on free memory of SMARTMEDIA® cardProcess variables and span selectableSnapshot, min/max or mean value, average10 s...10 h/pixel10 fold zoom in the event of high rate of change
Sensor monitor Direct display of measured values from sensor for validation
KI recorder
(option Item no. 558 074)
Adaptive representation of process flow with monitoring and signaling of critical process parameters
1) Caution! Never expose the display to direct sun light! Only operate the display within the temperature range of 0 up to 50°C max.
8285
p. 2/10
Technical data - Base unit (continued)
General data (continued)
Device self-test Test of RAM, FLASH, EEPROM, display and keypad, record for QM documentation to ISO 9000
Clock Real-time clock with date; Power reserve: approx. 1 year (lithium battery)
Data retention
in case of power failureParameters and factory settings > 10 years (EEPROM)
Logbook, statistics, records > 1 year (lithium battery)
Measurement recorder SMARTMEDIA® card
Module slots 3
Electrical connection Terminals via 5 x M20 x 1.5 cable glands
Connection cable Single wires and flexible leads up to 2.5 mm2 (AWG 14)
Ground wire: 2.5 mm2, screw M4 (EN 61010-1, 6.5.1.2)
Electrical data
Power supply
Overvoltage category Protection class Pollution degree
24 (–15 %)...230 (+15%) V AC/DC; approx. 10 VA/10 WIII2 (EN 61010-1)
Protection against
electrical shock
Protective connection according to EN 61010-1, 6.5.1
Binary input OK 1
Function
Switching voltage
Galv. separated (OPTO coupler); Vi ≤ 30 V, floating, galvanic isolation up to 60 Vswitches the device to HOLD mode (function check)
0...2 V AC/DC inactive, 10...30 V AC/DC active(invertible)
* User-defined1) To IEC 746 Part 1, at nominal operating conditions
Technical data - 8285 pH/ORP module
pH/ORP input
Measurement range pH value ORP value rH value Adm. voltage ORP + pH Adm. cable capacitance Glass electrode input1)
Ref. electrode input1)
Signal deviation1) 2)
(Display)
simultaneous pH and ORP measurement with several types of electrodes: - Input for pH/ORP glass electrode (type 8200)
0.1°C0.2 % meas. value + 0.5 K (< 1 K with NTC > 100°C)
Electrical data (continued)
Binary input OK 2
Function
Switching voltage
Galv. separated (OPTO coupler); Vi ≤ 30 V, floating, galvanic isolation up to 60 VSTART/STOP KI recorder, switch over to second parameter set0...2 V AC/DC inactive, 10...30 V AC/DC active(invertible)
Current output I1
Load monitoring Overrange*
Signal deviation1)
Current source
0/4...20 mA (22 mA), max. 10 V, galvanic isolation up to 60 V (galvanically connected with output I2)
Error message if load is exceeded22 mA in the case of a message< 0.25% current value + 0.05 mA0,00...22.00 mA
Current output I2
Load monitoring Overrange*
Signal deviation1)
Current source
0/4 ... 20 mA (22 mA), max. 10 V, galvanic isolation up to 60 V (galvanically connected with output I1)
Error message if load is exceeded22 mA in the case of a message< 0.25% current value + 0.05 mA0,00 ... 22.00 mA
Switching contacts*
Loadability Application*
4 relay contacts K1 ... K4, floating, galvanic isolation up to 60 VK1, K2, K3 are connected on one sideDC: < 30 V / < 500 mA, < 10 WK1 - K3, user definable for NAMUR mainte-nance request /function check, limit values, parameter set 2 active, rinsing contact, USP contact, K4 permanently set as alarm contact (NAMUR failure)
ORP* Automatic conversion to standard hydrogen electrode SHE when type of reference electrode is entered
Sensor standardization
ORP* Zero adjustable from –200...+200 mV
Sensor standardization
pH
Drift check*
Calimatic® buffer sets*
Nom. zero*
Nom. slope (25°C)*
Uiso*
1-/2-/3-point calibration (best fit line)
Operating modes:•Calimatic® automatic buffer recognition•Input of individual buffer values•Product calibration•Data entry of premeasured electrodesFine / standard / coarse•Fixed buffer sets:1- Knick/Mettler-Toledo; 2- Merck/Riedel;3- DIN 19267; 4- NIST Standard;5- Technical buffers to NIST; 6- Hamilton buffers•Manually enterable buffer set with max. three buffer tables (additional function Item no. 558 075)
pH 0 ... 14; calibration range ΔpH = ± 125 ... 61 mV/pH; calibration range 80...103 %-1000...+1000 mV
Calibration record Recording of: Zero point, slope, Uiso, response time, calibration process with date and time
8285
p. 3/10
Ambient temperature
Operation Storage
-20 to +55°C-20 to 70°C (limited through the electrode)
Relative humidity 10 ...95%, non condensated
Protection class IP65 / NEMA 4X
Standard EMC
Emitted interference Immunity to interference Lightning protection
NAMUR NE 21, EN 61326 VDE 0843 Part 20 /01.98,EN 61326/A1 VDE 0843 Part 20/A1 /05.99Class BIndustryEN 61000-4-5, Installation Class 2
Technical data - 8285 pH/ORP module (continued)
Statistics Recording of:Zero, slope, Uiso, response time, glass and reference impedance with date and time of the last three calibrations and the First Calibration
Sensocheck® Automatic monitoring of glass and reference electrode, message can be switched off
Sensoface® provides information on the sensor condition:Zero/slope, response time, calibration interval, Sensocheck®, CalCheck® (can be
disabled)
CalCheck® Monitoring of electrode calibration range during measurement
Sensor network
diagram
Graphical representation of current sensor parameters in a network diagram on the display: Slope, zero, reference impedance, glass impedance, response time, cal timer, deviation from calibration range (CalCheck®)
Sensor monitor Direct display of measured values from sensor for validation pH input / ORP input / glass el. impedance / ref. el. impedance / RTD / temperature
KI recorder
(option Item no. 558 074)
Adaptive representation of process flow with monitoring and signaling of critical process parameters
Adaptive calibration
timer*
Automatic adjustment of calibration interval (Sensoface® signal), depending on measured values
ServiceScope®*
(option Item no. 558 076)
Monitoring the inputs for overdriveRepresentation on display
Tolerance adjustment
(option Item no. 558 077)
Tolerant calibration/adjustment, tolerance limits adjustable, graphical recording of zero point and slope of the last 40 calibra-tions
* User-defined1) To IEC 746 Part 1, at nominal operating conditions2) ± 1 count, plus sensor error3) At 20°C, doubles every 10 K
- Linear characteristic 00.00 ... 19.99 %/K (reference temp user-defined)
- NLF nat. waters to EN 27888W)
- Ultrapure water with NaCl traces (0 ... 120°C)W)
- Ultrapure water with HCl traces (0 ... 120°C)W)
- Ultrapure water with NH3 traces (0 ... 120°C)W)
- Ultrapure water with NaOH traces (0 ... 120°C)W)
W) for all waters: Reference temp 25°C
Concentration
determination*
(option Item no. 558 080)
for the substances: HNO3 0...28 % by wt -20...+50°C 35...96 % by wt -20...+50°CHCl 0...18 % by wt -20...+50°C 22...39 % by wt -20...+50°CH2SO4
4) 0...30 % by wt -17.8...+110°C 32...84 % by wt -17.8...+115.6°C 92...99 % by wt -17.8...+115.6°CNaOH5) 0...14 % by wt 0...+100°C 18...50 % by wt 0...+100°CNaCl 0...26 % by wt 0...+60°CUser-defined concentration chart (5x5x5 values)
Sensor monitoring* Sensocheck®; Polarization and cable capacitance
Sensoface® provides information on the sensor condition
Sensor
standardization*
Adm. cell constant Calibration record
Operating modes- Autom. calibration with KCl or NaCl solution- Manual: Entry of conductivity- Product calibration / adjustment to vessel- Entry of cell constant with simultaneous display of conductivity and temperature 0.0050 ... 199.99 cm-1
Recording of: Cell constant, calibration method, with date and time
Output curves* Linear; Trilinear; Function (logarithmic); As desired via chart
USP function Water monitoring in the pharmaceutical industry (USP) with possibility to enter a limit value (%) Output via relay contact (K1...K3, BASE) possible
* User-defined1) To IEC 746 Part 1, at nominal operating conditions
2) ± 1 count, plus sensor error 3) c = 0.0050 ... 199.99 cm-1
4) measurement limits at 27°C 5) measurement limits at 25°C
Environment and Standard data - common to base unit, pH/ORP, conductivity modules
(option Item no. 558 079)
8285
p. 4/10
pH, ORPMohm, kohm
Temp InterfaceRTD
Module8285 pH/ORP
2 EL / 4 EL
RTD
µS, mS/cm% by wt
Module8285 El.COND
Temp Interface
Moduleslot SMARTMEDIA®
Basic unit8285
OK 1 OK 2
ContactsK1, K2, K3
AlarmK4
I 2
mA
I 1VariPower®
0(4) ... 20 mA
Moduleslot
Moduleslot
mA
0(4) ... 20 mA
Gla
ss
Ref
eren
ce
OR
P
Module8285 OUTPUT
Interface
I 4
mA
I 3
mA
Electronicalcontacts
Binary input
Binary input
System overview
A modular concept: base unit, measuring module, additional functions
The Type 8285 is an expandable modular process analysis system. The base unit provides three slots which can be equipped by the user with any
combination of measuring or communication modules. The software capabilities can be expanded by additional functions (options).
Additional functions must be ordered separately. They are supplied with a device-specific TAN for function release.
Additional functions
Activation via device
specific TAN
SMARTMEDIA® card
Data recording
Measuring modules
- pH / ORP / Temp
- Contacting conductivity / Temp
Communication modules
- OUT (additional switching and currents outputs)
- Further on request
3 module slots
for free combination of measuring
and communication modules
8285
p. 5/10
Modules
The modules: universally interchangeable
Various user-defined plug-in measuring modules for measurement and control functions can be
combined depending on the measuring task. They also facilitate subsequent trouble-free expansion
or modification.
Communication module for functional expansions: the OUT module for the expansion of the output
options is available.
If necessary, it can also record several measurement parameters in any combination with one device;
pH/pH, Cond/Cond etc. or e.g. simultaneaous pH and conductivity measurement.
Combined evaluation - i.e. the calculation of several measuring parameters e.g. for differential meas-
urement or quasi-redundant measuring systems. Up to 3 measuring modules can be combined.
Plug & Play - the modules are simply clicked into place in the slots provided, in any order. The modules are automatically recognized. Very straight for-
ward retro-fitting or modification; hot-swap technology.
The functions
Progress in perfection
In addition to excellent features such as the universally used VariPower® 20 to 265 V AC/DC power supply, the time and event controlled 2-channel
measurement recorder, the Sensocheck® sensor monitoring and the Calcheck® monitoring of the measured value distance between calibration, the Typ
8285 system can be further expanded with pioneering functions such as:
1. Early alarm detection with the KI recorder.
The KI recorder follows the course of the process and releases a message in the event of
abnormalities. The monitoring is always carried out for the primary measured variable, e.g.
pH or conductivity and parallel to that for the temperature. The visualization is graphical with
the process and limit value variation for both variable.
2. Checking of batch processes using the KI recorder.
The KI recorder records the course of a batch (self-teaching function). All further batches are
then monitored for deviations from the saved course.
3. Sensor network diagram for pH measurement.
Graphical representation of the current sensor parameters on the display in a network dia-
gram with slope, zero point, reference impedance, glass impedance, response time, calibra-
tion timer, deviation from calibration range.
4. ServiceScope®.
monitors whether the pH input signal lies within the input control range. Moreover, the rep-
resentation of the noise level over the time allows the distinction to be made between indi-
vidual disturbances, periodic and broadband disturbances wich is helpful for troubleshooting.
In this way, it is possible to check whether regularly recurring disturbances, e.g. large con-
sumers, which are regularly switched on or off can be simply detected. An error message is
generated if the noise level exceeds the failure limit.
Alarm message
pH
t
Toleranceband
KI recorder
Sensor network diagram
100Slope
Zero point
Reference impedance
Glass impedanceResponse time
Calibration timer
CalCheck50
0
t
Interferencelevel
Failure
ServiceScope®
8285
p. 6/10
SMARTMEDIA®
For parameters and the recording of data
• 5 parameter sets can be filed and loaded into the device
• A parameter set contains all parameter data, facilitating rapid exchange and speedy complete parameter
setting. Simple return to factory settings.
• Almost unlimited expansion of the measurement recorder
• Parameter sets can be transferred from one device to another. This removes the need for the tire some
repetition of inputting parameters.
• The device parameters can be completely saved on a SMARTMEDIA® card and then archived directly or
on a PC
• Extended logbook
SMARTMEDIA® is a registered trademark of the Toshiba Corp. Japan
Everything on a card
The SMARTMEDIA® card (measuring only 4.5 by 3.7 cm and only1 mm thick) is an extremely compact, very widely used memory expansion medium
available to the 8285 system. The SMARTMEDIA® card is already a global standard in such diverse fields of digital data processing as MP3 players
and digital cameras. This means that the SMARTMEDIA® card can be connected via the very inexpensive, commercially available adapter to RS232C,
USB etc. or directly via a PCMCIA adapter to any PC.
For software updates and software functions
• Contains the complete 8285 software
• Software functions can be installed at a later date and disconnected via transaction numbers (TAN)
• Software updates to keep 8285 upgraded with the latest software – also on site
Simple operation:
Simply insert the SMARTMEDIA® card into the small slot
on the rear of the front door.
8285
p. 7/10
Display and menu structure
Transflective LC graphic display
(240 x 160 pixels) white backlighting,
high resolution and high contrast
Measurement display
User interface
with plaintext menus as recommended
by NAMUR.
Menu texts can be switched to German,
English, French, Italian, Swedish, and
Spanish.
Intuitively acquirable menu logic, based
on Windows standard.
Secondary displays
2 Softkeys
with context-sensitive functions
Red LED
signals failure (On) or maintenance
request/function check (flashing)
according to NE 44
Green LED
Voltage supply okay5 self-sealing cable glands
M20 x 1.5
for entry of voltage supply and signal lines
Control panel
3 function keys (menu, meas, enter)
and 4 arrow keys for menu selection
and data entries
4 captive screws
for opening the transmitter
(Caution! Make sure that the gasket
between FRONT and BASE is properly
seated and clean!)
menu
10.54mS/cm
7.00 pH
TC man
man 25.1 °C Ausg I1 4.37 mA
meas
CalibrationIn the calibration menu,
you select the calibration
mode and execute the
calibration.
MaintenanceIn order to carry out mainte-
nance on the sensor, you can
freeze the output current call
up the current sensor data
or activate a manual cur-
rent source function for test
purposes.
Parameter settingParameter setting is divided
into three levels: Viewing
level, Operator level and
Administrator level.
DiagnosticsIn the diagnostics menu, you
receive information about
existing messages, the sen-
sor condition and about the
instrument itself. You can also
display the logbook here.
Viewing levelIn the viewing level, you can view all
parameters but not change them.
Operator levelHere you can set the parameters which
are authorized in the Administrator level.
Only the parameters which may be
changed in every day operation appear.
Administrator levelAll device parameters can be set here.
It is also possible to define which pa-
rameters are accessible in the Opera-
tor level using the marker setting.
The menu key can be used to access the
individual submenus via the function keys.
8285
p. 8/10
Dimensions [mm]
163
159.5147.5
39.5
14445.5 72
5.07
27
981M6 internal (4x)
141
Rear view Control panel cutout
4 91
Control panel cutout
Contour of enclosure front
Contour of rear unit
Front view Side view
r15 ± 3
312
72
Ø 6.5 (4x)
27
Ø 30 ... 65Wall mounting Post/pipe mounting
Ø 30 ... 40 mm
Ø 30 ... 65 mm
Ø 40 ... 62 mm
Ø 62 ... 65 mm
X
M6x50
X
Xwithout nut
M6x70
Ø 30 ... 65 mmfor vertical or horizontal mounting
150
8285
p. 9/10
Ordering chart for modular analysis transmitter Type 8285
Ordering chart for accessories for transmitter Type 8285
Description
Ite
m n
o.
Additional functions: SMARTMEDIA® card not necessary
KI recorder (pH only) 558 074
Additional sets of buffer solutions (pH only) 558 075
Servivescope (pH only) 558 076
Tolerance band recorder (pH only) 558 077
Current output curve freely programmable 558 078
Temperature compensation ultra-pure water (conductivity only) 558 079
SMARTMEDIA® is a registered trademark of the Toshiba Corp. JapanCalCheck®, Calimatic®, Sensocheck®, Sensoface®, ServiceScope®, Varipower® are registered trademark of the Knick GmbH & Co. KG, Germany
8285
p. 10/10
Interconnection possibilities with other Bürkert devices
To find your nearest Bürkert facility, click on the orange box → www.burkert .com
In case of special application conditions,please consult for advice.
We reserve the right to make technical changes without notice. 0704/0_EU-en_00895021