30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table of Contents Title Page Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter Structure Rotary Switch Functional Buttons The Effectiveness of Functional Buttons Display Symbols Measurement Operation A. DC Voltage Measurement B. AC Voltage Measurement C. Measuring Resistance D. Testing Diodes E. Testing for Continuity F. Temperature Measurement G. AC Current Measurement Sleep Mode Specifications A. General Specifications B. Environmental Restriction Accuracy Specifications A. AC Voltage B. DC Voltage C. Resistance D. Continuity Test E. Diode Test F. Temperature G. AC Current Maintenance A. General Service B. Replacing the Battery
32

Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Feb 05, 2018

Download

Documents

dangdien
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

30

1

30302929282828272727

262626

25232220191716151513121110986543

Model UT201/202: OPERATING MANUAL

Table of Contents

Title Page

OverviewUnpacking InspectionSafety InformationRules For Safe OperationInternational Electrical SymbolsThe Meter StructureRotary SwitchFunctional ButtonsThe Effectiveness of Functional ButtonsDisplay SymbolsMeasurement Operation A. DC Voltage Measurement B. AC Voltage Measurement C. Measuring Resistance D. Testing Diodes E. Testing for Continuity F. Temperature Measurement G. AC Current MeasurementSleep ModeSpecifications A. General Specifications B. Environmental RestrictionAccuracy Specifications A. AC Voltage B. DC Voltage C. Resistance D. Continuity Test E. Diode Test F. Temperature G. AC CurrentMaintenance A. General Service B. Replacing the Battery

Page 2: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Table of Contents

Title Page

OverviewUnpacking InspectionSafety InformationRules For Safe OperationInternational Electrical SymbolsThe Meter StructureRotary SwitchFunctional ButtonsThe Effectiveness of Functional ButtonsDisplay SymbolsMeasurement Operation A. DC Voltage Measurement B. AC Voltage Measurement C. Measuring Resistance D. Testing Diodes E. Testing for Continuity F. Temperature Measurement G. AC Current MeasurementSleep ModeSpecifications A. General Specifications B. Environmental RestrictionAccuracy Specifications A. AC Voltage B. DC Voltage C. Resistance D. Continuity Test E. Diode Test F. Temperature G. AC CurrentMaintenance A. General Service B. Replacing the Battery

Model UT201/202: OPERATING MANUAL

2

Page 3: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Overview

This Operating Manual covers information on safetyand cautions. Please read the relevant informationcarefully and observe all the Warnings and Notes strictly.

WarningTo avoid electric shock or personal injury, read the“Safety Information” and “Rules for Safe Operation”carefully before using the Meter.

Digital Multimeter Model UT201/202 (hereafter referredto as “the Meter”) are 3 1/2 digits with steady operations,fashionable structure and highly reliable measuringinstrument. The Meter uses large scale of integrated circuitwith double integrated A/D converter as its core and hasfull range overload protection.

The Meter can measure AC/DC Voltage, AC Current,Resistance, Temperature( ), Diodes, Continuityand so on.

oF/ oC

Model UT201/202: OPERATING MANUAL

3

Page 4: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

In the event you find any missing or damage, pleasecontact your dealer immediately.

Unpacking Inspection

Open the package case and take out the Meter. Checkthe following items carefully to see any missing or damagedpart:

Point Contact Temperature Probe(UT202 only)

4 1.5V Battery (AAA) 2 pieces

Model UT201/202: OPERATING MANUAL

4

Item123

DescriptionEnglish Operating ManualTest Lead

Qty1 piece1 pair

1 piece

Page 5: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Safety Information

This Meter complies with the standards IEC61010: inpollution degree 2, overvoltage category (CAT. II 600V,CAT. III 300V) and double insulation.

CAT. II: Local level, appliance, PORTABLE EQUIPMENTetc., with smaller transient overvoltages than CAT. III.CAT. III: Distribution level, fixed installation, with smallertransient overvoltages than CAT. IV

Use the Meter only as specified in this operating manual,otherwise the protection provided by the Meter may beimpaired.

In this manual, a Warning identifies conditions andactions that pose hazards to the user, or may damagethe Meter or the equipment under test.

A Note identifies the information that user should payattention to.

International electrical symbols used on the Meter andin this Operating Manual are explained on page 8.

Model UT201/202: OPERATING MANUAL

5

Page 6: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Rules For Safe Operation (1)

l Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors.l Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. Replace damaged test leads with identical model number or electrical specifications before using the Meter.l Do not apply more than the rated voltage, as marked on the Meter, between the terminals or between any terminal and grounding. If the value to be measured is unknown, use the maximum measurement position and reduce the range step by step until a satisfactory reading is obtained.l When measurement has been completed, disconnect the connection between the test leads and the circuit under test, remove the testing leads away from the input terminals of the Meter and turn the Meter power off.l The rotary switch should be placed in the right position and no any changeover of range shall be made during measurement is conducted to prevent damage of the Meter.l Do not carry out the measurement when the Meter’s back case and battery compartment are not closed to avoid electric shock.l Do not input higher than 600V between the Meter’s terminals and the grounding to avoid electric shock and damages to the Meter.

To avoid possible electric shock or personal injury,and to avoid possible damage to the Meter or to theequipment under test, adhere to the following rules:

Warning

Model UT201/202: OPERATING MANUAL

6

Page 7: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Rules For Safe Operation (2)l When the Meter working at an effective voltage over 60V in DC or 30V rms in AC, special care should be taken for there is danger of electric shock.l Use the proper terminals, function, and range for your measurements.l Do not use or store the Meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after dampened.l When using the test leads, keep your fingers behind the finger guards.l Disconnect circuit power and discharge all high -voltage capacitors before testing resistance, continuity and diode.l Replace the battery as soon as the battery indicator appears. With a low battery, the Meter might produce false readings that can lead to electric shock and personal injury.l When servicing the Meter, use only the same model number or identical electrical specifications replacement parts.l The internal circuit of the Meter shall not be altered at will to avoid damage of the Meter and any accident.l Soft cloth and mild detergent should be used to clean the surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident.l The Meter is suitable for indoor use.l Turn the Meter off when it is not in use and take out the battery when not using for a long time.l Constantly check the battery as it may leak when it has been using for some time, replace the battery as soon as leaking appears. A leaking battery will damage the Meter.

Model UT201/202: OPERATING MANUAL

7

Page 8: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

International Electrical Symbols

AC (Alternating Current).

DC (Direct Current).

AC or DC.

Grounding.

Double Insulated.

Continuity Test.

Diode.

Fuse.

Warning. Refer to the Operating Manual.

Conforms to Standards of European Union.

Capacitance Test

Deficiency of Built-In Battery

Model UT201/202: OPERATING MANUAL

8

Page 9: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

9

The Meter Structure (see figure 1)

( figure 1)

1. Input Terminals2. LCD Display3. Functional Buttons4. Rotary Switch5. Trigger: press the lever to open the transformer jaws. When the pressure on the lever is released, the jaws will close.6. Hand Guards: to protect user’s hand from touching the dangerous area.7. Transformer Jaws: designed to pick up the AC current flowing through the conductor. It could transfer current to voltage.

Model UT201/202: OPERATING MANUAL

Page 10: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Rotary SwitchBelow table indicated for information about the rotary switchpositions.

RotarySwitch

PositionFunction

AC/DC voltage measurement.: Diode test.

: Resistance measurement.Ω

Power is turned off.OFF

: Continuity test.Ω

Temperature measurement(UT202 only)

oCoF

AC current measurement range from 0.001A to 400.0AA

Model UT201/202: OPERATING MANUAL

10

Page 11: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Functional Buttons(1)Below table indicated for information about the functionalbutton operations.

AC voltage measurement.

l Press HOLD to enter and exit the Hold mode in any mode, the Meter beeps.l Press and hold HOLD button while turning on the Meter to display full icons.

HOLD

MAX Press MAX to start recording and updating of maximum values.

SELECT Press SELECT button to switch between and .oCoFΩ

Model UT201/202: OPERATING MANUAL

11

Page 12: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

The Effectiveness of Functional ButtonsNot every functional buttons can be used on every rotaryswitch positions. Below two tables describe which functionalbuttons can be used on which rotary switch positions

Model: UT201

Rotary SwitchPositions Functional Buttons

SELECT MAX HOLD

VV

N/AN/A

Ω N/A N/A

N/A

AA

2/20A N/A

200/400A N/A

Model: UT202

Rotary SwitchPositions Functional Buttons

SELECT MAX HOLD

VV

N/AN/A

Ω N/A

N/A

AA

2/20A N/A

200/400A N/A

oFoC(K-Type)

Model UT201/202: OPERATING MANUAL

12

Page 13: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Model UT201/202: OPERATING MANUAL

13

Display Symbols(1) (see figure 2)

No. Symbol Meaning

3

The battery is low.

Warning: To avoid false readings,which could lead to possible electricshock or personal injury, replacethe battery as soon as the batteryindicator appears.

4The Meter is in the auto range modein which the Meter automatically selectsthe range with the best resolution.

5 Test of diode.6 The continuity buzzer is on.7

Date hold is active.8

1 AC Indicator for AC voltage or current

2 DC Indicator for DC voltage

MAX Maximum reading displayed

oCoF9

Ω,kΩ,MΩ10

Page 14: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Model UT201/202: OPERATING MANUAL

14

No. Symbol Meaning

11 A Amperes (amps). The unit of current.

Display Symbols(2) (see figure 2)

mV, V Volts. The unit of voltage.mV: Millivolt. 1x10-3 or 0.001 volts

12

13 Indicates negative reading

14 OL The input value is too large for the selected range

Page 15: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Model UT201/202: OPERATING MANUAL

15

Measurement Operation(1)

WarningTo avoid harms to you or damages to the Meter fromeletric shock, do not attempt to measure voltageshigher than 600V AC/DC, although readings may beobtained.

red black

( figure 3)

The DC Voltage ranges are:200.0mV, 2.000V, 20.00V, 200.0V and 600V.

To measure DC voltage, connect the Meter as follows:

1. Insert the red test lead into the VΩ terminal and the black test lead into the COM terminal. 2. Set the rotary switch to V . 3. Connect the test leads across with the object being measured. The measured value shows on the display.

Notel In each range, the Meter has an input impedance of 10MΩ. This loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10kΩ, the error is negligible (0.1 or less).

A. DC Voltage Measurement (see figure 3)

Page 16: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Model UT201/202: OPERATING MANUAL

16

Measurement Operation(2)

l When DC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.

B. AC Voltage Measurement (see figure 4)

red black

( figure 4)

WarningTo avoid harms to you or damages to the Meter fromeletric shock, do not attempt to measure voltageshigher than 600V AC/DC, although readings may beobtained.

Page 17: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Measurement Operation(3)Notel In each range, the Meter has an input impedance of 10MΩ. This loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10kΩ, the error is negligible (0.1 or less).l When AC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.

17

Model UT201/202: OPERATING MANUAL

C.Measuring Resistance (see figure 5)

red black

( figure 5)

WarningTo avoid harms to you, do not attempt to input voltageshigher than 60V DC or 30V rms AC.

To avoid damages to the Meter or to the devicesunder test, disconnect circuit power and dischargeall the high-voltage capacitors before measuringresistance.

Page 18: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table
Page 19: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Measurement Operation(5)

D.Testing Diodes (see figure 6)

red black

( figure 6)

WarningTo avoid damages to the Meter or to the devices undertest, disconnect circuit power and discharge all thehigh-voltage capacitors before testing diodes.

Use the diode test to check diodes, transistors, and othersemiconductor devices. The diode test sends a current through the semicondutor junction, then measure thevoltage drop across the junction. A good silicon junctiondrops between 0.5V and 0.8V.

To test the diode out of a circuit, connect the Meter asfollows: 1. Insert the red test lead into the VΩ terminal and the black test lead into the COM terminal. 2. Set the rotary switch toΩ and press SELECT button to select measurement mode. 3. For forward voltage drop readings on any semiconductor component, place the red test lead on the component’s anode and place the black test lead on the component’s cathode.

Model UT201/202: OPERATING MANUAL

19

Page 20: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Measurement Operation(6)

Notel In a circuit, a good diode should still produce a forward voltage drop reading of 0.5V to 0.8; however, the reverse voltage drop reading can vary depending on the resistance of other pathways between the probe tips.l Connect the test leads to the proper terminals as said above to avoid error display.l The LCD will display OL indicating either open circuit or wrong polarity connection.l The unit of diode is volt (V), displaying the forward voltage drop readings.l To remove the objects being tested from the circuit when measuring can obtain a more accurate result.l When diode testing has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.

E.Testing for Continuity (see figure 7)

red black

( figure 7)

Model UT201/202: OPERATING MANUAL

20

Page 21: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Model UT201/202: OPERATING MANUAL

21

Measurement Operation(7)

WarningTo avoid damages to the Meter or to the devices undertest, disconnect circuit power and discharge all thehigh-voltage capacitors before measuring continuity.

To test for continuity, connect the Meter as follows: 1. Insert the red test lead into the VΩ terminal and the black test lead into the COM terminal. 2. Set the rotary switch to Ω and press SELECT button to select measurement mode. 3. The buzzer sounds if the resistance of a circuit under test is less than 50 Ω . 4. The buzzer may or may not sounds if the resistance of a circuit under test is between 50 Ω to 120 Ω . 5. The buzzer does not sound if the resistance of a circuit under test is higher than 120Ω.

Notel The buzzer beeps once when pressing any buttons at any rotary switch positions except at 2/20A positions if the button is valid. If the button is not valid, it does not beep. At 2/20A rotary switch position, the buzzer is set not to beep.l The buzzer beeps 5 times continuously on around 1 minute before entering the sleep mode. When it is just before entering the sleep mode, it will have one long beep to warn you.l The LCD displays OL indicating the circuit being tested is open.l When continuity testing has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.

Page 22: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Model UT201/202: OPERATING MANUAL

22

Measurement Operation(8)

F. Temperature Measurement (UT202 Only) (see figure 8)

The temperature measurement ranges are -40oC~1000oCand -40oF~1832oF.

To measure temperature, connect the Meter as follows: 1. Insert the red temperature probe into the VΩ terminal and the black temperature probe into the COM terminal. 2. Set the rotary switch to oFoC and press SELECT button to select oF oroC measurement mode. oC measurement mode is default. 3. Place the temperature probe to the object being measured. The measured value shows on the display.

red black

( figure 8)

Notel The Meter automatically displays the temperature value inside the Meter when there is no temperature probe connection.l When temperature measurement has been completed, disconnect the connection between the temperature probe and the circuit under test, and remove the temperature probe away from the input terminals of the Meter.

Page 23: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Model UT201/202: OPERATING MANUAL

23

Measurement Operation(9)

G. AC Current Measurement (see figure 9)

( figure 9)

WarningTo avoid electric shock, never measure current whilethe test leads are inserted into the input terminalsand disconnect test leads and tested circuit connection.

Never attempt an in-circuit current measuremnetwhere the open-circuit voltage between the circuitand the ground is greater than 600V

Use proper function and range for the measurement.The measuremnet ranges of current are: 2.000A, 20.00A,200.0A and 400A.

To measure current, do the following: 1. Set the rotary switch to 2/20A or 200/400 A 2. Press the lever to open the transformer jaws. 3. Center the conductor within the transformer jaw. The measured value shows on the display, it is a effective value of sine wave (mean value response).

Page 24: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Model UT201/202: OPERATING MANUAL

24

Measurement Operation(10)

Note:l To obtain accurate reading, measure only one conductor at each time.l When current measurement has been completed, disconnect the connection between the conductor under test and the jaw, and remove the conductor away from the transformer jaw of the Meter.

Page 25: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table
Page 26: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table
Page 27: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table
Page 28: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table
Page 29: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table
Page 30: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table
Page 31: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table
Page 32: Model UT201/202: OPERATING MANUAL · PDF file30 1 30 30 29 29 28 28 28 27 27 27 26 26 26 25 23 22 20 19 17 16 15 15 13 12 11 10 9 8 6 5 4 3 Model UT201/202: OPERATING MANUAL Table

Model UT201/202: OPERATING MANUAL

32