Top Banner
User’s Manual R Model GD410 Gas Calorie Meter IM 11T03C01-01E IM 11T03C01-01E 3rd Edition (YK)
95

Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

Jul 12, 2018

Download

Documents

truongxuyen
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

User’sManual

R

Model GD410Gas Calorie Meter

IM 11T03C01-01E

IM 11T03C01-01E3rd Edition (YK)

Page 2: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

i

IM 11T03C01-01E

IntroductionThe Model GD410 Gas Calorie Meter is the optimal system for the city gas industry where calo-rific values are measured in order to monitor and control the calories of LNG mixed in with such gases as air or butane.Before use, read this manual thoroughly in order to fully realize the capabilities of the meter. The topics and information that require special attention when handling the product are given in the text of this manual along with cautionary notes - WARNING or CAUTION - which reflect the importance of the information.For safety reasons or to avoid possible damage to your equipment, strictly adhere to every cautionary note that appears in this manual. Cautionary notes, such as a WARNING, that also appear on the product itself are indicated together with the alert symbol shown below:

The manual is applied to the following product.Detector: GD300S ( Intrinsically Safe, Explosionproof) (Style S2)

Converter: GD410G (Genenal Purpose) (Style S3)

GD410S (Explosionproof) (Style S3)

Throughout this user’s manual, you will find several different types of symbols are used to identify different sections of text. This section describes these icons.

CAUTION Identifies instructions that must be observed in order to prevent the software or hardware

from being damaged or the system from becoming faulty.

WARNING Identifies instructions that must be observed in order to avoid physical injury and electric

shock or death of the operator.

IMPORTANT Identifies important information required to understand operations or functions.

TIP Identifies additional information.

SEE ALSO

Identifies a source to be referred to.

Clicking a reference displayed in green can call up its source, while clicking a reference displayed in black cannot.

3rd Edition : Jul. 10, 2015-00Media No. IM 11T03C01-01E (CD) 3rd Edition : Jul. 2015 (YK)All Rights Reserved Copyright © 2008, Yokogawa Electric Corporation

Page 3: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

ii

IM 11T03C01-01E

l Cross-CheckingtheSpecificationsUpon delivery of the purchased product, unpack it carefully and make sure it is completely free from damage that may have occurred during transport. It must be shipped in strict conformance to the purchaser's specifications. By way of precaution, confirm that the equipment is the exact model you ordered. Also check that all accessory components are included, when confirming the specifications, refer to the model and suffix codes indicated on the nameplate on the equipment. For a description of the model and suffix codes, refer to Chapter1,''Specifications.''

l Consideration of Operation ParametersThe GD410 meter operates with the same parameters set when it was delivered (default data), when it is put into operation under these conditions.Before starting measurement, check whether or not the default data meets your operating condi-tions. If necessary, reset the parameters to suit your operating requirements. To check the de-faults, make use of the sheet, "Records of GD410's Operation Parameter Settings," in the back of this manual. It is advisable that, if any of the operation parameter settings have been changed, the new data be noted in this record.

l Information Covered in This ManualThis manual covers all of the information for handling the GD410o converter and GD300S detec-tor, including instructions on installation, operation, setting of operation parameters, inspection and maintenance.

CAUTIONModification and adaptation of the instrument’s parts, wiring and construction are prohibited.

3rd Edition : Jul. 10, 2015-00

Page 4: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

iii

IM 11T03C01-01E

[NotationalConventionsSpecifictoThisManual]The following notational conventions apply to the representations of operation keys, information shown on the display, and information indicated on the product itself when they are discussed specifically in the text of this manual.

l Operation KeysIndicated with brackets [ ] as: [YES] Key, and 【 】 as 【YES】 key

l Information Shown on DisplayIndicated with braces 『 』as:

『HOLD』, meaning the status『YES』, meaning the indicator on an operation key『CALIB』, meaning a message『205』 (lit) or 『205』 (blinking), meaning the data item shown along with its state

l Information Indicated on ProductIndicated with angle brackets < > as: <n> (lit) or <o> (unlit), meaning the contact output indicator lamp along with its state<MEASURE> mode, meaning the mode of measurement

l Information on the State of BlinkingIndicated in shaded typography as:

, in contrast to (lit) (blinking)

3rd Edition : Jul. 10, 2015-00

Page 5: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

iv

IM 11T03C01-01E

u After-sales Warranty

n Do not modify the product.

n During the warranty period, for repair under warranty carry or send the prod-ucttothelocalsalesrepresentativeorserviceoffice.Yokogawawillreplaceor repair any damaged parts and return the product to you. Before returning a product for repair under warranty, provide us with the model name and serial number and a description of the problem. Any diagrams or data explaining the problem would also be appreciated.l If we replace the product with a new one, we won’t provide you with a repair report.

l Yokogawa warrants the product for the period stated in the pre-purchase quotation Yokogawa shall conduct defined warranty service based on its standard. When the custom-er site is located outside of the service area, a fee for dispatching the maintenance engineer will be charged to the customer.

n In the following cases, customer will be charged repair fee regardless of war-ranty period.

• Failure of components which are out of scope of warranty stated in instruction manual.

• Failure caused by usage of software, hardware or auxiliary equipment, which Yokogawa Electric did not supply.

• Failure due to improper or insufficient maintenance by user.

• Failure due to modification, misuse or outside-of-specifications operation which Yokogawa does not authorize.

• Failure due to power supply (voltage, frequency) being outside specifications or abnormal.

• Failure caused by any usage out of scope of recommended usage.

• Any damage from fire, earthquake, storms and floods, lightning, disturbances, riots, warfare, radiation and other natural changes.

nYokogawadoesnotwarrantconformancewiththespecificapplicationattheusersite.Yokogawawillnotbeardirect/indirectresponsibilityfordamageduetoaspecificapplication.

nYokogawaElectricwillnotbearresponsibilitywhentheuserconfigurestheproduct into systems or resells the product.

nMaintenanceserviceandsupplyingrepairpartswillbecoveredforfiveyearsafter the production ends. For repair for this product, please contact the near-estsalesofficedescribedinthisinstructionmanual.

3rd Edition : Jul. 10, 2015-00

Page 6: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

v

IM 11T03C01-01E

Model GD410 Gas Density Meter

CONTENTS

IM 11T03C01-01E 3rd Edition

Introduction ..............................................................................................................iu After-sales Warranty ...................................................................................................... iv

1. SPECIFICATIONS ..................................................................................... 1-11.1 SpecificationofGD410GasCalorieMeter ..................................................... 1-1

1.1.1 General Specifications ....................................................................... 1-1

1.1.2 GD300S Detector .............................................................................. 1-3

1.1.3 Pressure Transmitters (Specify with the option code.) ...................... 1-3

1.1.4 GD410G or GD410S Converter.........................................................1-4

1.1.5 GD350G Flow rate Setter ..................................................................1-5

1.2 ModelandSuffixCodes ...................................................................................1-61.2.1 Calorimetric Converter (General purpose) ........................................ 1-6

1.2.2 Calorimetric Converter (Explosionproof) ........................................... 1-6

1.2.3 Calorimetric Detector (Intrinsically Safe, Explosionproof) ................. 1-6

1.2.4 Flow rate Setter (Non-Explosionproof) .............................................. 1-7

1.2.5 Hardware for Connection with External Cables (For Explosionproof use) .... 1-7

1.2.6 Two-core, Double-Shielded Cable ..................................................... 1-7

1.2.7 Pressure-Reducing Valves for Gas Cylinders ................................... 1-7

1.3 External Views and Dimension ........................................................................ 1-81.3.1 GD410G Converter (General purpose) .............................................1-8

1.3.2 Pipe- and Wall-Mounting Hardware (Optional) .................................. 1-9

1.3.3 GD410S Converter (Explosionproof) ...............................................1-10

1.3.4 GD300S Detecor (Intrisically Safe, Explosionproof) ....................... 1-11

1.3.5 Detector Unit (Intrinsically Safe, Explosionproof) ............................1-12

1.3.6 Flow rate Setter ................................................................................1-13

2. INSTALLATION, WIRING AND PIPING .................................................... 2-12.1 Installing the Detector ......................................................................................2-1

2.1.1 Selecting the Location ........................................................................2-1

2.1.2 Mounting the Detector........................................................................2-2

2.2 Installing the Converter ....................................................................................2-32.2.1 Selecting the Location ........................................................................2-3

2.2.2 Mounting the Converter .....................................................................2-4

2.3 Piping .................................................................................................................2-6

3rd Edition : Jul. 10, 2015-00

Page 7: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

vi

IM 11T03C01-01E

2.4 Wiring .................................................................................................................2-72.4.1 Wiring Procedure ...............................................................................2-7

2.4.2 Cable Wired to Power Supply ..........................................................2-10

2.4.3 Cables Wired to Outputs ..................................................................2-10

2.4.4 Cables Wired to Contact I/Os .......................................................... 2-11

2.4.5 Cable Wired to GD300S Detector ...................................................2-12

2.4.6 Wiring to Model EJA310 pressure transmitter .................................2-13

2.4.7 Cables Wired to the Ground ............................................................2-14

3. OPERATION .............................................................................................. 3-13.1 Preparation for Operation ................................................................................ 3-1

3.1.1 Inspecting Installation, Piping and Wiring Workmanship .................. 3-1

3.1.2 Display on Operation Panel and Operation Keys ..............................3-2

3.1.3 Supplying Power ................................................................................ 3-3

3.1.4 Basic Key Operation .......................................................................... 3-4

3.1.5 Checking the Setting Parameters and Changes in Their Default Values ...... 3-6

3.1.6 Calibrating the Meter for Correct Readings ....................................... 3-6

3.1.7 Checking the Meter for Performance ................................................. 3-7

3.1.8 The state of the contact output .......................................................... 3-7

3.2 Normal Operation ..............................................................................................3-83.2.1 Starting Operation ..............................................................................3-8

3.2.2 Corrective Actions in Case of Failure .................................................3-8

3.2.3 Inspection and Maintenance .............................................................. 3-9

3.3 Shutdown and Restart ......................................................................................3-93.3.1 Measures for Shutdown ..................................................................... 3-9

3.3.2 Measures for Restarting .................................................................... 3-9

4. PARAMETER SETTING ............................................................................ 4-14.1 Summary of Setting Operations ...................................................................... 4-1

4.1.1 Measurement, Operation, Setting and Service ................................. 4-1

4.1.2 Key Operations .................................................................................. 4-1

4.1.3 Points to Be Noted When Making Settings ........................................4-2

4.2 Setting Parameters ...........................................................................................4-34.2.1 Setting Parameters at Measurement Level ....................................... 4-3

4.2.2 Setting Parameters at Operation Level ............................................. 4-3

4.2.3 Setting Parameters at Setting Level .................................................. 4-4

4.2.4 Setting Parameters at Service Level ................................................. 4-4

4.3 Parameter Setting .............................................................................................4-64.3.1 Message Display/Selection Mode ..................................................... 4-6

4.3.2 Parameter Setting at Operation Level ...............................................4-8

4.3.3 Parameter Setting at Setting Level ..................................................4-10

4.3.4 Parameter Setting at Service Level .................................................4-14

3rd Edition : Jul. 10, 2015-00

Page 8: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

vii

IM 11T03C01-01E

5. CALIBRATION PROCEDURE .................................................................. 5-15.1 Basic Calibration Procedure ............................................................................ 5-15.2 Preparation for Calibration .............................................................................. 5-1

5.2.1 Setting Type of Calibration and Checking Valve Operation ...............5-1

5.2.2 Setting Calibration Data .....................................................................5-3

5.3 Calibration .........................................................................................................5-45.3.1 Semi-automatic Calibration ...............................................................5-4

5.3.2 Manual Calibration .............................................................................5-5

5.3.3 Automatic Calibratio ...........................................................................5-6

6. INSPECTION AND MAINTENANCE ........................................................ 6-16.1 Routine Inspection and Maintenance ............................................................. 6-1

6.1.1 Checking Readings and Calibrating the Analyzer ............................. 6-1

6.1.2 Checking the Flow rate of Sample Gas ............................................. 6-1

6.1.3 Periodic Replacement of the Detector's O-rings ............................... 6-1

6.1.4 Replacing the Fuse ............................................................................6-2

6.1.5 Cleaning .............................................................................................6-2

6.2 Inspection In Case of Failure ........................................................................... 6-36.2.1 Inspecting the Analyzer in an Alarm Status ....................................... 6-3

6.2.2 Inspecting the Analyzer in a FAIL Status ........................................... 6-4

Operating Data Values of GD410 Gas Calorie Meter ............................Append-1Customer Maintenance Parts List .......................................... CMPL 11T3B1-01ECustomer Maintenance Parts List .......................................... CMPL 11T3B1-02ECustomer Maintenance Parts List .......................................... CMPL 11T3B1-03ERevision Information ...............................................................................................i

3rd Edition : Jul. 10, 2015-00

Page 9: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

Blank Page

Page 10: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-1

IM 11T03C01-01E

1. SPECIFICATIONS

1.1 SpecificationofGD410GasCalorieMeter1.1.1 GeneralSpecifications

In principle, this instrument is a gas density meter. For this reason, the GD410 can be used as a gas calorie meter only in the case that the density of the gas to be measured has the corelation with its calorific value.

l System Components(1) GD300S detector: Rain proof for outdoor use

Intrinsically safe and explosionproof construction (Exd [ia] IIB+H2T5)

(2) GD410G general purpose converter: Rain proof for outdoor use or GD410S explosionproof converter: Rain proof for outdoor use, explosionproof construction (Exd IIB+H2T6)(3) EJA310 Pressure transmitter (optional):

Rain proof for outdoor use, explosionproof construction(Exdo II CT4)

IMPORTANTWhen the detector or converter is installed outside, use a hood or a cabinet to protect it from sunlight.

l CharacteristicsRepeatability: ±0.5% of full scale or ±1 g/Nm3, whichever the greater

Linearity: ±1.0% of full scale (against the output range or the calibration range, whichever the greater).

Zero and span drift: ±3 g/Nm3/month

Response: 5 sec for 90% response (from the time when the reading begins to change) for a gas flow rate of 0.6 l/min.

l Measuring RangesCalorific value 0-133 MJ/Nm3, minimum range: calorific value equivalent to 0.1 kg/Nm3

Density: 0-6 kg/Nm3, minimum range: 0.1 kg/Nm3

l Output SignalsOutput 1: 4-20 mA DC

Isolated from inputs; load resistance: 600 W maximum

(Load Resistance of 250-550 W required when in the BRAIN communication mode.)

Output 2: 4-20 mA DC

Isolated from inputs; load resistance: 600 W maximum

3rd Edition : Jul. 10, 2015-00

Page 11: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-2

IM 11T03C01-01E

l Power SupplyPower supply Voltage: 100 to 240 V AC;

Rated voltage range: 100 to 240 V AC

Allowable voltage range: 85 to 264 V AC

Rated frequency: 50 or 60 Hz

Allowable frequency range: 47 to 63 Hz

Power supply Voltage 24 V DC;Rated voltage range: 24 V DC

Allowable voltage range: 21.6 to 26.4 V DC

l Dielectric strength (Between converter and power supply)For 100 to 240 V AC: 1500 V AC, one minuteFor 24 V DC: 500 V AC, one minute

l Isolation resistance (Between converter and power supply)500 V DC, 10 MΩ or highe

l Power ConsumptionApproximately 12 W.

l Requirements for Sample GasMeasurable type of gas: all gases, except for highly corrosive onesFlow rate: 0.1 to 1 l/min.Temperature: -10 to 50 °C (non-condensing)Pressure: 588.5 kPa (abs.) maximum

3rd Edition : Jul. 10, 2015-00

Page 12: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-3

IM 11T03C01-01E

1.1.2 GD300S Detector

l Material exposed to gasSUS 316 stainless steel, NBR(Acrylonitrile Butadiene Rubber) and Fluorine-contained Rubbe

l Ambient conditionTemperature: -10 to 60 °CHumidity: 5 to 95 %RH

l Installation: Pipe-mounted

l Finish:Cover: equivalent to Munsell 0.6GY3.1/2.0Case: equivalent to Munsell 2.5Y8.4/1.2

l Weight: approximately 7 kg (including mounting hardware)

1.1.3 Pressure Transmitters (Specify with the option code.)These transmitters are designed to compensate for pressure so that density in abnormal state can be measured. Choose the desired type of transmitter by specifying the option code of the detector.

lOptioncode''/KU'':detectorunit The detector unit is designed so that the GD300S detector and the pressure transmitter

(EJA310) are integrated on a single plate. This allows the space where the pressure trans-mitter is normally mounted to be used effectively for other purposes.

lOptioncode''/KUHT'':detectorunit The detector unit (described in the above item, "option code "/KU") is equipped with TIIS

cerified EJA TAG.

3rd Edition : Jul. 10, 2015-00

Page 13: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-4

IM 11T03C01-01E

1.1.4 GD410G or GD410S Converter

l DisplayReading:

Digital (6 digits maximum; resolution: 0.1 g/m3)

Data items shown:

calorific value (MJ/Nm3), density (kg/Nm3 or kg/m3), gas temperature (°C) and gas pressure (kPa or MPa)

Alarm indications: abnormal calorific value, abnormal physical density, abnormal corrected density, abnormal range of pressure input, and abnormal values of calibration

Parameters for calibration: time of calibration, settling time, starting time of calibration and calibration cycle

Self-diagnostic indications: sensor oscillation shut down, abnormal oscillation frequency of sensor, failure in temperature detection of sensor, failure in A/D conversion stage, and memory failure

Alarm settings: high/low limits the contact state can be set at either “normally open (NO)” or “normally closed (NC)”.

lContactOutputs/InputContact Outputs

Contact capacity: 250 V AC at 3A or 30 V DC at 3A Types of signals are Maintenance, Fail, Hi/Lo alarms, and a signal for opening/closing either the zero or span calibration solenoid valve. The contact state can be set at either ”normally open (NO)” or “normally closed (NC).” Fail contact is set onlt at NC

Contact Inputs

Dry contacts for starting calibration upon closure

l CalibrationManual (one touch), Semi automatic, Automatic calibration

l Communication Protocol: BRAIN communication

Data items that can be transmitted by the hand-held terminal are numerical data such as density, calories, temperature, pressure, and Hi/Lo alarm set points and self-diagnostic parameters.

l Ambient ConditionsTemperature: -10 to 55 °C

Humidity: 5 to 95 %RH

l Fuse250 V 1A time lag type authorized VDE/SEMKO 100 to 240 V AC Model250 V 2A time lag type authorized VDE/SEMKO 24V DC Model

3rd Edition : Jul. 10, 2015-00

Page 14: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-5

IM 11T03C01-01E

l InstallationNon-explosionproof models: Pipe-, panel- or wall-mounted

Explosionproof models: Pipe- or panel-mounted

l FinishModel GD410G (general purpose):

Front cover: equivalent to Munsell 0.6GY3.1/2.0

Case: equivalent to Munsell 2.5Y8.4/1.2

Model GD410S (explosionproof):

Equivalent to Munsell 0.6GY3.1/2.0

l WeightModel GD410G (general purpose): approx. 3 kg (including mounting hardware)Model GD410S (explosionproof): approx. 15 kg (including mounting hardware)

1.1.5 GD350G Flow rate SetterModel GD350G-1- o for indoor installation:

Construction: Indoor panel mount

Material: Case SECC-P10/10

Main parts SUS304, SUS316

Tube SUS316

Finish: equivalent to Munsell 0.6GY3.1/2.0

Weight: approximately 2.5 kg

Flow rate: 0.1 to 1 l/min.

Installation: pipe- or rack-mounting hardware supplied as standard

Model GD350G-2- o for outdoor installation:

Construction: dust- and drain-proof

Material: Case Aluminum alloy

Main parts SUS304, SUS316

Tube SUS316

Finish: Cover equivalent to Munsell 0.6GY3.1/2.0

Case equivalent to Munsell 2.5Y8.4/2.0

Weight: approximately 5 kg

Flow rate: 0.1 to 1 l/min.

Installation: pipe-or rack-mounting hardware supplied as standard

3rd Edition : Jul. 10, 2015-00

Page 15: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-6

IM 11T03C01-01E

1.2 ModelandSuffixCodes1.2.1 Calorimetric Converter (General purpose)

Suffix Code Option Code Specifications

-N -10 -11

-N-J-E

/PI/W/PA/H

General purpose converterAlways ''-N''

100 to 240VAC24V DCAlways ''-N"

Model CodeGD410G

General purposestandardsPower supply

Measured objectNameplate

Option Pipe-mounting hardwareWall-mounting hardwarePanel-mounting hardwareHood

JapaneseEnglish

T1.1.ai

Style:S3

1.2.2 Calorimetric Converter (Explosionproof)

Suffix Code Option Code Specifications

-J -10 -11

-N-J-E

/PI/PA/G

/H

Explosionproof converterTIIS Certified

100 to 240VAC24V DCAlways ''-N"

Model CodeGD410S

Explosionproof standardsPower supply

Measured objectNameplate

Option Pipe-mounting hardwarePanel-mounting hardwareExplosionproof cable gland*-Type quantity (1 to 6) in the fieldHood

JapaneseEnglish

* When using the converter in hazardous area, always attach the Yokogawa-assured explosionproof cable gland (option cord:/G). Spectify the number of cable glands for converter, see the table in Subsection 1.3.3.

Style:S3

1.2.3 Calorimetric Detector (Intrinsically Safe, Explosionproof)

Suffix Code Option Code Specifications

-J-E

/PI/KUHT/KU/G1

Explosion-proof detectorTIIS Certified

Model CodeGD300S

Explosionproof standardsNameplate

Option Pipe-mounting hardwareDetector unit with TIIS certified EJA TAG*1Detector unit *1Cable gland attached *2

JapaneseEnglish

-J

T1.3.ai*1: System containing the option code “/KU” and “/KUHT” are shipped as a unit into which both the detector and pressure transmitter are integrated. When using a separate pressure transmitter, make sure it is equivalent to the Model EJA310 pressure transmitter. See the GS 01C22D01-00E general specifications leaflet for more information. In addition, select the transimitter with the pressure range of less than 588.5 kPa abs. Moreover, “/KU” and “/KUHT” cannot be selected with “/PI” because they are panel mounting type.*2: When using the detector in a hazardous area, always attach the Yokogawa-assured specified explosionproof cable gland(option code:/G1).

Style:S2

3rd Edition : Jul. 10, 2015-00

Page 16: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-7

IM 11T03C01-01E

1.2.4 Flow rate Setter (Non-Explosionproof)Model code Suffix Code Option Code Specifications

GD350GDust- and rainproof casingFitting

Flowrate setter

-1-2

-J-A

Model w/o casing: for indoor installationModel w/ casing: for outdoor installation

Rc1/81/8NPT(F)

1.2.5 HardwareforConnectionwithExternalCables(For Explosionproof use)

Part No.L9811LL

G9601AN

DescriptionExplosionproof cable gland

Explosionproof cable gland

SpecificationsUsed for the converter(Cable outer diameter: φ8 to φ16mm)Used for the detector(Cable outer diameter: φ10 to φ13.5 mm)

T1.5.ai

1.2.6 Two-core, Double-Shielded Cable

GDW

Length

SpecificationsModel code Suffix CodeTwo core double shield cable (O.D. φ10.9 mm), both ends finished with cable pins.Length in meters, 500 meter maximum- L

T1.6.ai

1.2.7 Pressure-Reducing Valves for Gas CylindersPart No.L9850BA

L9850BB

DescriptionPressure-reducing valve for gas cylindersPressure-reducing valve for gas cylinders

SpecificationsModel with clockwise screw*

Model with counterclockwise screw*

*: For cylinders containing less than 5% of flammable gas (by volume), use a valve with a clockwise screw; otherwise use a valve with a counterclockwise screw.

T1.7.ai

3rd Edition : Jul. 10, 2015-00

Page 17: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-8

IM 11T03C01-01E

1.3 External Views and Dimension1.3.1 GD410G Converter (General purpose)

Unit:mm(in.)

144(

5.7)

144 (5.7) 23 112 (4.4)

80 (3.1)

80 (3

.1)

36 36

36

(1.4

)

A B C

D FE

20 (0

.8)

Four holes, 6 mm (0.24) in dia.,8 mm deep M6

Cable inlet (21 mm in dia.)Compatible with a PG13.5 cable gland

Grounding terminal (4 mm screws)

(0.9)

38

(1.4

)

(1.4)(1.4)

Weight: approximately 3 kg 6.6pounds)

184 (7.2)

Hood(Option code:/H)

220 (8.7)

72 (2

.8)

Cable gland Connection

A, B ・Pressure transmitter・Analog output ・Contact・DetectorC・Contact OutputD, E・Power SupplyF

3rd Edition : Jul. 10, 2015-00

Page 18: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-9

IM 11T03C01-01E

1.3.2 Pipe-andWall-MountingHardware(Optional)

F1.2.aiMounting pipe of 50 mm nominal size(60.5 mm in outer dia.)

Optionalhardware

Four holes,6 mm (0.2) in dia.

Unit: mm (in.)Weight: approximately 3 kg (6.6 pounds)

188 (7.4)

100

174

(6.9

)

200

(7.9

)

50 (2

)

• Hardware for Pipe Mounting: GD410G-N- -N- /PI

F1.3.aiThree holes,10 mm (0.4) in dia.

135 (5.3)

70 (2.8)

100 (4)

13 (0.51)

224

(8.8

)

224

(8.8

)

35 (1.4

)15 (0.6

)

Optionalhardware

Four holes,6 mm (0.2) in dia.

• Hardware for Wall Mounting: GD410G-N- -N- /W

+2 0

F1.4.ai

Unit :mm

23

(0.9)

139

178 (7)

139

100

(4)

+2 0 +0.0

8 0

12 mm (0.5) maximum(panel thickness)

Optionalhardware

Four holes,6 mm (0.2) in dia.

Dimensions of panel cutout

Weight: approximately 3 kg (6.6 pounds)

• Hardware for Panel Mounting: GD410G-N- -N- /PA

(5.4

7

)

+0.08 0(5.47 )

3rd Edition : Jul. 10, 2015-00

Page 19: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-10

IM 11T03C01-01E

1.3.3 GD410S Converter (Explosionproof)

Unit: mm (in.)Weight: approximately 15 kg (33.1 pounds)

A

A B

D E

C

F

Panel Cutout

182 (7.2)

140

(5.5

)

Ø198 1

Approx. 209 (8.2)

Approx. 173 (6.8)

2-M8, depth 16 mm, one on each side 36 (1.4)

56

5439

24

56

15 (0.6)

(259

) (10

.2)

HoodOption code:/H

(300) (11.8)

(Used to fix thepipe-mounting hardware)

App

rox.

242

(9.5

)

36 (1.4)

Mounting pipeof JIS 50 A nominal size(60.5 mm (2.4) in outer dia.)

Pipe-mounting hardware

Approx. 200 (7.9)

Approx. 182 (7.2)

Approx. 205.2 (8.1)

140

(5.5

)

Cable glands not included

4-M8, depth20 mm

Side view from arrow A direction

Approx. 212 (8.3) in dia.

Six wiring holes, G3/4 threadedGrounding terminal(5 mm (0.2) screw)

(2.2)(2.2)

(2.1

)(1

.5)

(0.9

)

4 holes with 10mm in dia,

Cable gland Connection

・Power SupplyA

・Contact OutputB, C

D ・Detector

E, F ・Pressure from Transmitter, ・Analog Output ・Contact Input

The screw hole for panel installationon the converter is M8 and 20mm deoth.The bolts attached to the product areM8 screw and 20mm in length.

3rd Edition : Jul. 10, 2015-00

Page 20: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-11

IM 11T03C01-01E

1.3.4 GD300S Detecor (Intrisically Safe, Explosionproof)

Approx. 253

Appr

ox. 2

63Ap

prox

. 150

(Ref

eren

ce

dimen

sion)

Approx. 216

Connect to converter

Class A grounding

G 3/4 Wire pipePrepare by customer

Sealing fitting hardware

Unit:mmWeight:Approx.7kg(including mounting hardware)

(Note) Option code (/G1) Explosion-proof cable gland G9601AN Cable O.D.: φ10 to φ16

When using the detector in a hazardous area,be sure to complement class A grounding working in a non-hazardous area.

• Hardware for Pipe Mounting: Model GD300S-J-/PI/G1

Prepare by customer

F1.6.ai

Explosion-proof cable gland (Note)

3rd Edition : Jul. 10, 2015-00

Page 21: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-12

IM 11T03C01-01E

1.3.5 Detector Unit (Intrinsically Safe, Explosionproof)

Approx. 250

Appr

ox.1

50Ap

prox

.350

300

(Ref

eren

ce d

imen

sion

)320350

Pressure transmitter

Detector

(Note 2)

(Note 1)

Gas inletRc1/4

4-φ12 holesGas outletRc1/4

The detector unit (option code: /KU or /KUHT) is shipped with a unit which both the detector and pressure transmitter are integrated into.

Explosion-proof cable gland

Connect to converter Class A grounding

G 3/4 Wire pipePrepare

by customer

Sealing fitting hardware

Unit:mmWeight: Approx.15kg(including mounting hardware)

(Note1) Option code (/G1) Explosion-proof cable gland G9601AN Cable O.D. : φ10 to φ16

When using the detector in a hazardous area, be sure to complement class A grounding working in a non-hazardous area.

Prepare by customer

(Note2) Explosion-proof cable gland G9601AM Cable O.D. : φ8-- to φ12

Connect to converter

Explosion-proof cable gland

• GD300S-J-/KU or /KUHT

F1.7.ai

3rd Edition : Jul. 10, 2015-00

Page 22: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<1.Specifications> 1-13

IM 11T03C01-01E

1.3.6 Flow rate Setter

50A Hardware for pipe mounting

Switching cockSS1

Flowmeter with needle valve FM1 (0.1 to 1 l /min)

4 - φ9 holes 70170

40

70 (φ60.5)

A25 40 40 40

B C

FM1

Diagram of internal piping

SS1 COM

D

280

260

Piping material: O.D.1/8" SUS316

Symbol

ABCD

User-interfaceNameSPANZERO

SAMPLEOUT

User-interfacePipe sizeRc 1/8Rc 1/8Rc 1/8Rc 1/8

User-interfacePipe size

1/8NPT (F)1/8NPT (F)1/8NPT (F)1/8NPT (F)

Unit: mmWeight:Approx. 2.5kg

GD350G-1-J GD350G-1-A

Model GD350G-1- for indoor installation:

F1.8.ai

Approx. 195

FM1

2B pipe(φ60.5)

Pipe mountingRack mounting

Internal layout

56 202

(K9410UA)Cable inlet

(Rc 1/8) or 1/8NPT(F)

150

40 40 40

75

260 150 23833

340

158

126.

5

SS1

A B C D

Rack mounting hardware(K9410UC)

4-φ6 holesHardware for pipe mounting

Switching cockSS1

Flowmeter with needle valve FM1 (0.1 to 1 l /min)

Diagram of internal piping

Piping material: O.D.1/8" SUS316

Symbol

ABCD

User-interfaceNameSPANZERO

SAMPLEOUT

GD350G-2-JUser-

interfacePipe sizeRc 1/8Rc 1/8Rc 1/8Rc 1/8

GD350G--2-AUser-

interfacePipe size

1/8NPT (F)1/8NPT (F)1/8NPT (F)1/8NPT (F)

Unit: mmWeight:Approx. 5kg

Model GD350G-2- for outdoor installation:

F1.9.ai

3rd Edition : Jul. 10, 2015-00

Page 23: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

Blank Page

Page 24: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-1

IM 11T03C01-01E

2. INSTALLATION, WIRING AND PIPINGThe Model GD410 Gas Calorie Meter is thoroughly inspected at the factory and carefully packed to ensure the equipment does not suffer any damage during transportation. The package should also be handled with care when unpacking to prevent the equipment from undergoing severe mechanical shock. After unpacking, visually check the equipment to ensure that it is free from any damage.Although the detector has no controls on it, it may need to be accessed for inspection or for other reasons. Install the detector in a location as close as possible to where the gas is sampled and where maintenance can be easily carried out. The converter has a display with controls on it; thus, you should install it so that the keys are positioned directly in front of you when workingNote that there must be a clearance of at least 400 mm at the back of the converter, GD410S, since cables are wired to the back by removing the screwed rear cover.

2.1 Installing the DetectorThe GD300S detector is an intrinsically safe, explosionproof instrument (Exd [ia]IIB+H2T5). Install the detector in a location where the following conditions are satisfied.

2.1.1 Selecting the Location

l Explosionproof constructionBefore installing the detector in an explosion-hazardous area, ensure that the area conforms to the explosion proof code noted above.

l No corrosive gasesCorrosive gases are not desirable because they may damage the electrical components in the detector.

l Slight mechanical vibrationAlthough the detector is vibration-resistant, vibration may loosen the connections of the external wiring.

l No exposure to direct sunlightExposure to direct sunlight may raise the temperature in the detector to abnormal levels, and should therefore be avoided. Note that such abnormal temperature levels can also result from heat irradiated from high-temperature equipment around the converter.

lHumiditymaintainedbetween5%and95%RHAvoid choosing a location that is likely to be exposed to abnormally high or low humidity over a prolonged period. It is recommended that the converter be used at a humidity between 25% and 85 % RH.

l No exposure to rain water Even though the detector is rainproof, whenever possible install it where it is protected from water splashes. The reasoning for this is that the detector cover may need to be removed for maintenance or for other reasons.

3rd Edition : Jul. 10, 2015-00

Page 25: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-2

IM 11T03C01-01E

2.1.2 Mounting the Detector The detector is designed for pipe mounting of horizontal position. Measuring accuracy can be worse if you don’t horizontally install it.

Approx.253

Appr

ox. 2

63Ap

prox

. 150

(Ref

eren

ce d

imen

sion)

Approx.216

Explosion-proof cable gland (Note)

Connect to converter

Class A grounding

G 3/4 Wire pipePrepare by customer

Sealing fitting hardware

Unit: mmWeight: Approx.7kg(including mounting hardware)

(Note) Option code (/G1) Explosion-proof cable gland G9601AN Cable O.D.:φ10 to φ16

When using the detector in a hazardous area,be sure to complement class A grounding working in a non-hazardous area.

• Hardware for Pipe Mounting:Model GD300S-J-/P1/G1

Prepare by customer

F2.1.ai

Figure2.1 Pipe Mounting

CAUTIONDo not have piping when carring or installing the detector connected to pipes, or the leakage from piping might occur.

Approx. 250

Appr

ox.1

50Ap

prox

.350

300

(Ref

eren

ce d

imen

sion)

320350

Pressure transmitter

Detector

(Note 2)

(Note 1)

Gas inletRc1/4

4-φ12 holesGas outlet

Rc1/4

The detector unit (option code:/KU or /KUHT) is shipped with a unit which both the detector and pressure transmitter are integrated into.

Explosion-proof cable gland

Connect to converter

Class A grounding

G 3/4 Wire pipePrepare

by customer

Sealing fitting hardware Unit: mmWeight: Approx.15kg(including mounting hardware)

(Note1) Option code (/G1) Explosion-proof cable gland G9601AN Cable O.D.: φ10 to φ16

When using the detector in a hazardous area, be sure to complement class A grounding working in a non-hazardous area.

Prepare by customer

(Note2) Explosion-proof cable gland G9601AM Cable O.D.: φ8 to φ12

Connect to converter

Explosion-proof cable gland

F2.2.ai

Figure 2.2 Detector Unit

3rd Edition : Jul. 10, 2015-00

Page 26: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-3

IM 11T03C01-01E

2.2 Installing the ConverterThe GD410 converter comes in either the general purpose Model GD410G or explosion proof Model GD410S (explosion proof code: Exd IIB+H2T6). Choose the Model GD410S if the con-verter is to be used in a hazardous area.

2.2.1 Selecting the Location

l Ease of operationSelect a location where you can easily view the readings on the display and work with the keys. Installing the converter closer to the detector will ease your maintenance work, including calibra-tion.

l Explosionproof constructionWhen installing the converter in an explosion-hazardous area, ensure that the area conforms to the explosionproof code noted above.

l No corrosive gasesCorrosive gases are not desirable because they may damage the electrical components in the detector.

l Slight mechanical vibrationAlthough the detector is vibration-resistant, vibration may loosen the connections of the external wiring.

l No exposure to direct sunlightExposure to direct sunlight may raise the temperature in the detector to abnormal levels, and should therefore be avoided. Note that such abnormal temperature levels can also result from heat irradiated from high-temperature equipment around the converter. If the converter is likely to receive direct sunlight and the temperature in the converter will likely exceed the operating limits, attach a hood (optional).

lHumiditymaintainedbetween5%and95%RH.Avoid choosing a location that is likely to be exposed to abnormally high or low humidity over a prolonged period. It is recommended that the converter be used at a humidity between 25% and 85 % RH.

l No exposure to rainwaterEven though the converter is rainproof, whenever possible install it where it is protected from water splashes. The reasoning for this is that the converter cover may need to be removed for maintenance or for other reasons

3rd Edition : Jul. 10, 2015-00

Page 27: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-4

IM 11T03C01-01E

2.2.2 Mounting the Converter(1) Mounting the Model GD410S Converter

Approx. 212 in dia.

Approx. 200 Approx. 209

App

rox.

242

Approx. 173 Approx. 36

200

140

182A

Side view from arrow A direction

Cable glands not included

2-M8 (one on each side)(Used to fix the pipe-mounting hardware)

Mounting pipe of JIS 50A nominal size(60.5 mm in outer dia.)

Pipe-mounting hardware

Unit: mm

Note: Cable gland L9811LL Outer diameter φ8 to φ16

182

140

φ198

±1

4-φ10 through-hole

Panel cutout

The screw hole for panel installationon the converter is M8 and 20mm depth.The bolts attached to the product areM8 screw and 20mm in length.

F2.3.aiFigure 2.3 Pipe Mounting

(2) Mounting the Model GD410G ConverterPipe Mounting

Figure 2.4 shows the mounting hardware and how to mount the converter.

Converter

Nuts (two)

Washers (two)

Stanchion (50A pipe)

U-bolt

Bracket forpipe moounting

Bracket

Bracket setscrew

Note: When attaching an awing hood, use the two bracket setscrews along the top.

F2.4.ai

Figure 2.4 Pipe Mounting

3rd Edition : Jul. 10, 2015-00

Page 28: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-5

IM 11T03C01-01E

Wall MountingFigure 2.5 shows how to mount the converter on a wall.

Note : The converter for wall mounting comes with the samemounting hardware as that of a pipe-mounting model. When installing the converter on a wall, use the bracket only.

Mounting hole(three places)

Converter

M8 bolt (not supplied)Use a bolt long enough to fit into the mounting hole.

Bracket

70

35

102

144

Center of the converter

3-M8 screw-hole or 3-φ10 through-hole

Unit : mm

Dimension of cutout for wall mounting F2.5.ai

Figure 2.5 Wall Mounting

Panel MountingFigure 2.6 shows how to mount the converter on a panel.

Insert the converter in the panel cutout before attaching the brackets.

Converter

Bracket

Panel

Setscrew (two)

139+20

139+20

Dimensions of cutout for panel mounting

F2.6.ai

Figure 2.6 Panel Mounting

3rd Edition : Jul. 10, 2015-00

Page 29: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-6

IM 11T03C01-01E

2.3 PipingThe piping connected to the meter comprises a line that feeds the gas under measurement to the detector, a line that returns (or releases to the atmosphere) the gas exhausted after measure-ment, and lines that carry the zero- and span-point calibration gases. The type of piping chosen depends on the composition of the gas under measurement, its pressure, the amount of dust mixed, and the response (dead time). It is advisable however to use JIS-based SUS stainless-steel piping of sizes from 6 mm and 4 mm in outer and inner diameters up to JIS 15A (1/2 in.).Refer to the following instructions when connecting the meter piping. l Figure 2.7, "System Configuration," shows an example of piping. Connect the lines securely

to ensure that there is no gas leakage in the system.

l Make sure the pressure of the sampled gas is no greater than 0.5 MPa when measured at the detector inlet. If the pressure is too high, use a pressure-reducing valve to regulate the pressure to a normal level. If the pressure is too low, it must be raised using a pump.

l If the gas under measurement contains dust, mist or moisture, such impurities must be removed. Install a filter, mist separator or dehumidifier to remove these impurities from the gas.

l In order to return the gas exhausted after measurement, the difference in pressure between the detector inlet and the point of returning the gas must be 0.5 kPa minimum. Choose pip-ing with a large inner diameter to minimize pressure loss in the return gas line.

l ALWAYS install a stop valve at the point where the gas is sampled (or returned).

l The pressure transmitter for compensating pressure is designed to detect the pressure inside the detector; install a pressure lead pipe as close to the detector as possible (prefer-ably, no more than 0.5 m away).

Sample gas line

Switching valveNote 1

P1

Note 2P2

Note 4

Filter

F2.7.ai

GD300Sintrinsically safe, explosionproof detectorExd [ia]IIB+H2T5

Note 1: P1 (inlet pressure) Max. 0.5 MPaNote 2: P1 (inlet pressure) - P2 (outlet pressure) 0.5 kPa (depending on the size and length of the pipe)Note 3: Flow rate must be 0.1 to 1.0 l/min.Note 4: The cylinder pressure must be reduced to P1 (inlet pressure).

Zero pointcalibration gas

Span point calibration gas Pressure regulator

for gas cylinder

Explosionproofpressure transmitter

Note4

FlowmeterNote 3

Flowrate setterGD350G

Exhaust-gas line

Detector unit(Option code: /KU or /KUHT) Pressure signal

(two-core shield cable)

GD410GGeneral purpose converter

equivalent to NEMA 4X

Sensor signal(two-core shield cable)

Class A grounding(Ground the detector in a non-hazardous area.)

or

GD410SExplosionproof converter

ExdIIB+H2T6

Class D grounding

85 to 264V AC or 24V DCOutput 1: 4-20mA DC(load resistance 600 Ω maximum)(Brain communication mode: load resistance 250 to 550 Ω)

Output 2: 4-20mA DC(load resistance 600 Ω maximum)

Contact input [dry] (start of calibration)

Hand-held terminal(for Brain communication)

Contact output [dry] (FAIL, ALARM, MAINTENANCE)250V AC 3A or 30V DC 3A

Contact [dry] for zero and span solenoid valves250V AC 3A or 30V DC 3A

Figure2.7 SystemConfiguration

3rd Edition : Jul. 10, 2015-00

Page 30: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-7

IM 11T03C01-01E

2.4 Wiring

WARNINGDangerHighVoltage!

Some parts of the meter’s internal assembly have high voltage. Inadvertent contact with those parts may result in electrical shock or injury. ALWAYS turn off the power to the meter before removing the rear or front cover.

This section explains how to wire the GD410 meter. Note that this document is limited to the basic system configuration only (detector, converter and pressure transmitter).For details on the wiring of instruments used to receive analog output signals or contact output signals, see their respective instruction manuals.

CAUTIONAfter wiring to the terminal block , do not push the remaining cables into the enclosure.

2.4.1 Wiring Procedure(1) The types of cables wired to the GD410o converter are:

・ Cable wired to power supply

・ Cable wired to detector input

・ Cables wired to output signals (two)—or one signal if BRAIN communication is used

・ Cables wired to contact outputs (five) -- provided as necessary

・ Cable wired to contact input -- provided as necessary

・ Cables wired to pressure transmitter

・ Cables wired to the ground

(2) The GD410o converter has six cable-inlets for external wiring. Wire the converter through cable glands. As a rule, choose cable glands as shown in Figure 2.8.

Cable gland Connection

A, B•Pressure transmitter•Analog output •Contact Input

C •Detector

D, E •Contact Output

F •Power Supply

Unit : mm (in.)Cable inlet (21 mm (0.9) in dia.)Compatible with a Pg13.5 cable gland

Grounding terminal (4 mm screw)

A B C

D E F

< GD410G >

F2.8.ai

3rd Edition : Jul. 10, 2015-00

Page 31: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-8

IM 11T03C01-01E

Cable gland Connection

A

•Pressure transmitter•Analog output •Contact Input

B, C•DetectorD•Contact Output

E, F

•Power Supply

Unit : mm (in.)Cable inlet (21 mm (0.9) in dia.)Compatible with a Pg13.5 cable gland

Grounding terminal (5 mm screw)

A B C

D E F

F2.8a.ai

< GD410S >

Figure 2.8 Choice of Cable Glands

(3) The following explains the general wiring procedure for each type of converter.<Wiring the GD410S converter>

CAUTION

Wait 5 minutes after power disconnection, before opening the enclosure.

CAUTIONThe rear cover and the metal plugs at the external-wiring connections are fixed onto the convert-er case with hex setscrews. Before removing the cover, be SURE to loosen the setscrews. If you rotate the cover or a plug without loosening the set screws, their threads may be severed, making the cover un-removable.When the rear cover or any of the cable glands or plugs are removed, handle them with utmost care so that the threads are not damaged. When re-attaching it on to the converter, clean the threads so they will not be damaged due to such foreign matters as dust.

・ Loosen the set screws and remove the rear cover by rotating it counter clockwise.

・ Guide the required external-wiring cables through the cable glands to the converter.

Exercise care so as to avoid connecting them with the wrong polarities.

・ After wiring, securely fasten the rear cover and cable glands.

<Wiring the GD410G converter>

・ Loosen the setscrews in the four corners and remove the front cover.

・ Remove the terminal cover. Guide the required external-wiring cables through the cable glands to the converter. Exercise care so as to avoid connecting them with the wrong polari-ties.

・ After wiring, reinstall the terminal cover and front cover in place.

3rd Edition : Jul. 10, 2015-00

Page 32: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-9

IM 11T03C01-01E

*1: Power supply Use a two-conductor cable (Note).*2: Connection to the detector a. Use a two-conductor shielded cable(Note). In addition, the go-and-return resistance must be 50Ω or less (for conductors with a cross sectional area of 1.25 mm2 and a length no greater than 1.4 km) b. Be sure to ground the two-conductor shielded cable at either end of the shield. If an unusual phenomenon occurs due to noise, increase the countermeasures against noise. (For example, ground the detector body or use a double -shielded cable. If a double-shielded cable is used, ground shields of each conductor on either side of the cable. Ground the end on the outer shield on the detector side to the case of the detector and connect that end en the converter side to terminal 13.) c.Terminal 13 is for the detector only.*3: Connection to the pressure transmitter a. Use a two-conductor shielded cable(Note). In addition, the go-and-return resistance must be 50Ω or less (For conductors with a cross section of area of 1.25 mm2 and a length no greater than 1.4 km). b. Be sure to ground the two-conductor shielded cable at either end of the shield.*4: Be sure to ground the case of the converter.*5: The contacts for the contact output are all dry contacts whose NO/NC state can be freely set except for the FAIL contact which is an NC contact and cannot be freely set. Their contact rating is 250 V AC, 3A or 30 VDC, 3A each.*6: If the detector is to be installed and used in a hazardous area, be sure to implement class A grounding work. In addition, the grounding point of class A grounding must be located in a non-hazardous area.*7: Use a two conductor shielded cable for any output signal(Note). Be sure to ground the two-conductor shielded cable at either end of the shield.*8: Terminal 26 is connected to the case-grounding terminal.*9: The contact input (contact for starting calibration) is a dry contact. Calibration starts when the contact is closed.*10: The voltage in parentheses is specified for a case when the converter is driven by DC power supply.*11: For wiring and grounding of the pressure transmitter, refer to the user’s manual attached to the pressure transmitter.

Note: The following table identifies the outer diameters of cables (suitable for the given cable glands) that can be used depending on the equipment connected.

GD410S Explosion-proofGD410G General purpose

MAINTENANCEcontact output *5

ALARM contact output*5

FAIL contact output*5

Contact output for span calibration solenoid valve *5

Contact output for zero calibration solenoid valve *5

Power supply, 85 to 264 V AC *1

(24 V DC*10)

Class D grounding *4

MAINT

ALM

FAIL

SPAN

ZERO

LN

Case groundingterminal

CONTIN

ANLGOUT1

ANLGOUT2

GD410G,S Converter

Contact input(calibration start contact)*9Isolated output signal, 4 to 20 mA DCLoad resistance: 600 Ω or less *7(250 to 550 Ω for Brain communication)

Isolated output signal, 4 to 20 mA DCLoad resistance: 600 Ω or less *7

Pressure transmitter

Case grounding terminal

GD300S detector

Class A grounding(grounding resistance 10 Ω or less)*6

Case grounding terminal

*2

*3

14

15

16

17

18

19

2021

22

23

2425

26

SNSRPWR

SNSRINP

DETINP

SHIELD

1

2

34

5

6

7

8

9

10

11

12

13

F2.9.ai

equipment connected

Converter

DetectorGD300S

∅8 to ∅12∅10 to ∅16

∅8 to ∅16

Pressure transmitterEJA310 Output signal Contact input

∅8 to ∅16

∅6 to ∅12

∅8 to ∅16

∅6 to ∅12

*11

G*81 2

( )

( )

Figure 2.9 Terminals on the Converter

3rd Edition : Jul. 10, 2015-00

Page 33: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-10

IM 11T03C01-01E

2.4.2 Cable Wired to Power SupplyThis cable supplies the power complying with the specifications to the GD410o converter. Use a two-conductor cable of at least 1.25 mm2 thick to connect the power source to the terminal block of the converter. Be SURE to install a double-pole power switch externally.

CAUTIONThe GD410S converter contains a power switch and a fuse. The switch cannot be operated how-ever, if the converter is installed in an explosion-hazardous area; it must be left placed in the ON position, For this reason, install a power switch externally. When turning the switch on, BE SURE to simultaneously shutdown the lines of both polarities.

Wire the cable as instructed below:(1) Use a cable 1.25 to 2.5 mm2 thick. Treat the end of the cable by stripping the core wires

back to a length of 7 mm (see Figure 2.10).(2) Wire the cable to the L and N terminals on the block.

Loosen the terminal screws, insert the stripped ends of the core wires into the terminals, and fasten the screws to fix the cable.

2.4.3 Cables Wired to OutputsThese cables are used to transmit 4-20 mA DC signals and carry out BRAIN communication. Use shielded cables of 0.75 mm2 minimum in thickness (or a two-conductor shielded cable for single output).Wire the cables as instructed below:(1) Use cables 0.75 to 2.5 mm2 thick. Treat the cable ends by stripping the core wires back 7

mm (see Figure 2.10). (2) Wire the cable for out put 1 to terminals 3 and 4 on the block and the cable for output 2 to

terminals 5 and 6 on the block. Exercise care so as to avoid connecting the cables with the wrong polarities. Note that only terminals 3 and 4 (output 1) are effective for BRAIN commu-nication. Loosen the terminal screws, insert the stripped ends of the core wires into the terminals, and fasten the screws to fix the cable. (tightening torque: 0.4 Nm)

(3) Ground the shielding wire at a given terminal on the converter; do not ground the other end of the cable.

7mm

Figure 2.10 Examples of Treatment on Cable End Wired to the Converter

3rd Edition : Jul. 10, 2015-00

Page 34: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-11

IM 11T03C01-01E

2.4.4 CablesWiredtoContactI/OsThe contact I/Os of the GD410o converter comprise the start-of-calibration contact (input), the FAIL, ALARM and MAINTENANCE contacts, and the contacts for operating the solenoid valves for span and zero calibrations. Choose the type of cables for a group of calibration-purpose con-tacts and for a group of other contacts separately.For wiring to the calibration contact output, use a four-conductor shielded cable 1.25 mm2 mini-mum in thickness. For wiring to other contacts, use an eight-conductor shielded cable of 0.13 to 1.25 mm2 in thickness, while choosing the number of core wires depending on the number of contacts used.Since the contact outputs are voltage-free, driving such devices as alarm lamps requires a sepa-rate power supply. For the contact input (start of calibration), provide a voltage-free signal.Wire the cables as instructed below:(1) Use cables 0.13 to 1.25 mm2 thick. Treat the cable ends by stripping back the core wires to

7 mm from the end (see Figure 2.10).(2) Wire the cables to their respective given terminals on the block, being careful not to mistake

one terminal for another. For terminal numbers assigned to the respective contacts I/Os, see Figure 2.9. Loosen the terminal screws, insert the stripped ends of the core wires into the terminals, and fasten the screws to fix the cables. The appropriate tightening torque is 0.4 Nm.

Cable

Cable inlet

Tightening screw

Terminal

3rd Edition : Jul. 10, 2015-00

Page 35: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-12

IM 11T03C01-01E

2.4.5 Cable Wired to GD300S DetectorThe cable wired to the GD300S detector comes from the GD410o converter. Use a shielded ca-ble of 10 to 16 mm in finished outer diameter that suits the dedicated cable gland. Wire the cable through the cable gland, being careful not to connect the cable with the wrong polarities. Attach JIS M4-size (4-mm) crimp-on terminals to the ends of the core wires (see Figure 2.10). Connect the shielding wire to terminal 13 on the converter (see Figure 2.9).If a malfunction occurs and it is assumed to be due to noise, use a double-shielded cable (for example, the cable with Yokogawa's part number A1150WC). Connect the outer shielding wire to terminal 13 on the converter and the other end of the wire to the grounding terminal on the detector. Connect the inner shielding wires to the grounding terminal on the converter (see Figure 2.12).

Terminals for wiring to converter

Class A grounding terminal

Figure2.11 ConfigurationofDetectorTerminals

11 12 13

Outer shielding wire

Class A grounding terminal

Do not connect the inner shielding wires here.

Inner shieldingwires

Outer shieldingwire

GD300SDetector

GD410Converter

F2.12.ai

Class A grounding

Figure 2.12 Connection Between Detector and Converter When Using a Double-Shielded Cable

CAUTIONThe intrinsically safe grounding terminal, the terminal for Class A grounding in the terminal box of the GD300S, must be separately grounded with the Class A grounding or in equivalent grounding manners in a non-hazardous area.

3rd Edition : Jul. 10, 2015-00

Page 36: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-13

IM 11T03C01-01E

2.4.6 Wiring to Model EJA310 pressure transmitterWhen Wiring to model EJA310 pressure transmitter, connect by using cables from GD410o converter. Wire the cables through cable glands. Be sure not to misconnect polar. For finished external diameter, refer to Figure 2.9, "Terminals on the converter." The size of the terminal screw threads is M4. Use a crimp terminal suited for your cable terminal.

EJA310Pressure transmitter GD410

Converter

7 8 9 10

+ - +

F2.13.ai

Figure 2.13 Connection Between Detector and Pressure Transmitter

For wiring and grounding of the pressure transmitter, refer to the user’s manual attached to the pressure transmitter.

3rd Edition : Jul. 10, 2015-00

Page 37: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<2. Installation, Wiring and Piping> 2-14

IM 11T03C01-01E

2.4.7 Cables Wired to the GroundWhen using the GD300S detector in an area requiring explosion protection, BE SURE to ground it as per the Class A Grounding Standard using a conductor of at least 2 mm2 in nominal thick-ness. The Class A ground terminal must be used for joining between converter and the earth cable as shown in Figure 2.14. For sealing wiring ports of the converter, use the explosionproof cable gland or the sealing fitting hardware.Also BE SURE to ground the GD410o converter as per the Class D Grounding Standard. Use No. 26 internal wiring terminals or the grounding terminals of the converter. (The appropriate tightening torque for the internal wiring terminals is 0.4 Nm.) Case-grounding terminal can be used as protective ground terminal. When using the detector's grounding terminals, attach M4 crimp terminals to the ends of the grounding cable. When crimping the terminals, use tools that fit the terminals. Wire the earth cable with the explosionproof cable gland (in case of the GD410G, use the attached cable gland) when connecting the cable to the internal terminal No.26.

Earth cable

24 25 26L N G

Terminal

Earthing of internal wiring terminal

Ground terminalLock washer

Ground wiringcrimp terminal

Joining advice of Convertercase ground terminal

Converter case

Earthing of enclosure wiring terminal

Figure 2.14 Power and Ground Wiring

3rd Edition : Jul. 10, 2015-00

Page 38: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<3. Operation> 3-1

IM 11T03C01-01E

3. OPERATIONThis chapter explains how to operate the Model GD410 Gas Calorie Meter.

3.1 Preparation for Operation

CAUTION(1) DO NOT supply power to the meter when inspecting it.(2) Leave the power switch which is built in the meter, in the ON position.(3) DO NOT open the terminal box cover with the power turned on, if the meter is installed in an

explosion-hazardous area.

3.1.1 Inspecting Installation, Piping and Wiring Workmanship

Before turning the power on, inspect workmanship with regards to installation, piping and wiring as instructed below:l Examine how the GD410 gas calorie meter and its peripheral equipment are installed.

l Check the sample-gas line, which feeds the GD300S detector, for any gas leakage.

l Inspect the system to ensure that the GD410 meter and its peripheral equipment are wired correctly. After inspection, shut the terminal box cover securely.

l Place the power switch built in the GD410S converter (explosionproof) in the ON position.

3rd Edition : Jul. 10, 2015-00

Page 39: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<3. Operation> 3-2

IM 11T03C01-01E

3.1.2 Display on Operation Panel and Operation KeysFigures 3.1 and 3.2 illustrate the operation pannels of the Model GD410G and GD410S meters. The operation pannel has a display. It also has the seven operation keys for setting operation parameters and so on.

UNIT /MODEkg/m3 DENSkg/Nm3 DENS COMP

MJ/Nm3

MEASUREDISPLAYSEMI CALMANUAL CALVALVE

SUB MENU

RANGECAL DATAALARMSERVICE

YES NO MODE

ENT

*

E X A GD

contactsMAINTALARMCALFAIL

GD410G

ADJUST

Mark of operation level (Refer to 4.3.2)

State DisplayData Display

Message Display

Key operation Display

Operation Key

【YES】: When the blinking "Key Operation Display" is YES, the YES key is pushed. 【NO】: When the blinking “Key operation Display" is NO, the NO key is pushed. 【MODE】: When switching from the measure- ment mode to the operation mode, the mode key is pushed. When switching from modes other than the measurement to the measurement mode, the mode key is pushed. 【 】: This switch is used to select the input data the digit. 【 】: This switch selects the input data value. 【ENT】: When the input data is entered, this switch is pushed

Pointer(The unit of the measurement and the measurement mode (Or setting mode) light.)

Setting mode at measurement level display (Refer to 4.3.3)

Setting level switch key

Contact signal output display lamp (Refer to 3.1.7)

F3.1a.ai

Unit expression

Figure 3.1 Functional Description of GD410G Converter (Cover Open)

UNIT /MODEkg/m3 DENSkg/Nm3 DENS COMP

MJ/Nm3

MEASUREDISPLAYSEMI CALMANUAL CALVALVE

SUB MENU

RANGECAL DATAALARMSERVICE

CONTACTSMAINT ALARM CAL FAIL

EXAGD

YES NO MODE

ENT

*Mark of operation level (Refer to 4.3.2)

State DisplayData Display

Message Display

Key operation Display

Operation Key

【YES】: When the blinking "Key Operation Display" is YES, the YES key is pushed. 【NO】: When the blinking “Key operation Display" is NO, the NO key is pushed. 【MODE】: When switching from the measure- ment mode to the operation mode, the mode key is pushed. When switching from modes other than the measurement to the measurement mode, the mode key is pushed. 【 】: This switch is used to select the input data the digit. 【 】: This switch selects the input data value. 【ENT】: When the input data is entered, this switch is pushed

Pointer(The unit of the measurement and the measurement mode (Or setting mode) light.)

Setting mode at measurement level display (Refer to 4.3.3)

Setting level switch key

Contact signal output display lamp (Refer to 3.1.7)

F3.1b.ai

Unit expression

Figure 3.2 Functional Description of GD410S Converter (Cover Open)

3rd Edition : Jul. 10, 2015-00

Page 40: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<3. Operation> 3-3

IM 11T03C01-01E

3.1.3 Supplying PowerFirst make sure that when the GD410 meter is turned on, no control instruments connected to the meter will come into operation by the signals transmitted from the meter. Next, turn on the external power switch to start the meter. At that point, make sure the internal power switch is in the ON position if the meter is an explosionproof model.After power-on, the meter will begin to operate in the measurement mode. [MajorItemsShowninMeasurementMode]

l Measured Values

In the Message Selection mode under the DISP mode, choose the item you want to show.

Physical density X.XXXX KG/M3Corrected density X.XXXX KG/NM3Calorific value XXX.XXX MJ/NM3Measured gas temperature XXX.X °CMeasured gas pressure XXX.XX KPA or MPAAnalog output 1 XXX.X MA1%Analog output 2 XXX.X MA2%

Item No. of Digits and Position of Decimal Point

Message

T3.1.ai

l Types of Messages

・Measurement units ・Error number (shown if an error occurs) ・Alarm number (shown if an alarm occurs)l HOLD and FAIL Indications (shown regardless of whether the meter is in the measurement

mode or not)

・HOLD appears when the meter is in a hold state. ・FAIL appears if an error occurs.

3rd Edition : Jul. 10, 2015-00

Page 41: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<3. Operation> 3-4

IM 11T03C01-01E

3.1.4 Basic Key OperationThe basics of key operation are the selection between modes/levels, the selection and execution of the function/action, and entry of data values.1. Selecting Between Modes at the Operation Level (See Chapter 4 for details.)l You can enter the measurement mode by pressing the [MODE] key while the meter is in a

mode other than the measurement mode.

When switching between modes, note the following two points:(1) Press the [MODE] key once. The message field first shows a password prompt and then

changes to look like one shown in Figure 3.3 -(1).

In the display shown left, the 『YES』 and 『NO』fields blink. Pressing the 【YES】 key will changethe message to read 『START』, indicating the analyzer is now in the semi-automatic calibration mode.

When password is selected, the message field shows <passw>, prompting you to enter a passwords. Type 『331』 and press the【ENT】key. The message field changes to look like one shown left.

F3.2.ai

(1) (2)

Figure 3.3 Examples of Display at Operation Level

(2) In the display shown in Figure 3.3-(1), each press of the [NO] key changes the message field, enabling you to choose a different mode by its indication.

2. Selecting Between Modes at the Setting Level (See Chapter 4 for details.)(1) In the measurement mode, press the [*] key (to select between settings).

In the display shown left, the 『YES』 and 『NO』fields blink. Pressing the 【YES】 key will changethe message to read 『OUT1』, indicating the analyzer is now in the analog output setting mode.

When password is selected, the message field shows <passw>, prompting you to enter a passwords. Type 『331』 and press the【ENT】key. The message field changes to look like one shown left.

F3.3.ai

(1) (2)

Figure 3.4 Examples of Display at Setting Level

(2) In the display shown in 3.4-(1), each press of the [NO] key changes the message field, enabling you to choose a different mode by its indication.

3rd Edition : Jul. 10, 2015-00

Page 42: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<3. Operation> 3-5

IM 11T03C01-01E

3. Entering Data Values

IMPORTANTThe entered data are kept in merory even if you turn off the power.

The following explains how to make changes to existing data entries.In the display shown in Figure 3.4-(2), pressing the [YES] key causes the message field to default to [*CALRY]. Pressing the [YES] key once again changes the message to [*Z_CAL], the range of lower limit setting, and makes the indication of the lower limit blink.Assume the existing data entry of the range's lower limit is 100 MJ/Nm3. To change the value to 95 MJ/Nm3, follow the instructions given below:(1) Press the 【>】key so “1“ starts blinking.(2) Press the【 】> key until the blinking "1" changes to ''0".(3) Press the 【>】key until the ‘tens’ digit begins blinking. Then, press the 【 】> key until the

blinking ‘tens’ digit changes to “9".(4) Press the【>】key until the ‘ones‘ digit begins blinking. Then, press the【 】> key until the

blinking digit changes to ''5”.(5) Press the [ENT]key. This confirms the value “95".

Pressing the【ENT】key enters the value shown in the data field.

In the display shown in Figure 3.4 (2), pressing the 【YES】 key will change the display to look like this.

F3.4.ai

(1) (2)

Figure 3.5 Display During Data Setting

3rd Edition : Jul. 10, 2015-00

Page 43: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<3. Operation> 3-6

IM 11T03C01-01E

4. Entering a Password A password must be entered to proceed into the operation or setting level. When you attempt to enter that level, a password-entry message (password prompt) like one shown in Figure 3.6-(1) appears. The required password is "331".

Press the【MODE】key, then the【*】key changes the display to look like this.

Type the password (331) and then press the【ENT】key. If you type a wrong password,you cannot enter the level.

(1) (2)

Figure 3.6 Password Prompt and the Reading of a Password Typed

3.1.5 Checking the Setting Parameters and Changes in Their Default Values

Set up the relevant parameters to meet individual operating requirements. The GD410 meter is designed to prohibit parameter setting unless you enter the password (331).The analog output range is factory-set to "0-133.000''. At the operation level, set the parameter “Density/ Convert-to-Calorie Factor” under the ADJUST mode and the parameter “Calorie Unit: MJ/Nm3“ under the DISP mode. Then, at the setting level, set the analog output.In the case of critical parameters, confirm that they are given correct settings even if the values set upon shipment from the factory (default values) will apply without needing any changes.Methods of changing the default values are detailed in Chapter 4. Refer to this information before you start setting parameters.

3.1.6 Calibrating the Meter for Correct ReadingsFor details on the calibration of the meter for correct readings, see Chapter 5.

3rd Edition : Jul. 10, 2015-00

Page 44: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<3. Operation> 3-7

IM 11T03C01-01E

3.1.7 Checking the Meter for PerformanceWhen the required parameters have been completely set, bring into operation all equipment composing the measurement loop. Keep this equipment operating for a while. After making sure there is no problem with the equipment, go into normal operation.For reference, the following summarizes the setting functions (related to signals) of the GD410 meter that are used to obtain the optimum operating conditions.[FunctionsRelatedtotheOutputSignal]

The output signal can be held at a value immediately before calibration or at a desired value during calibration or at the setting/ service level. This is to prevent the output signal from adversely affecting the operation of the equipment configuring the measurement loop (default setting: the output signal is held at the value immediately before).[FunctionsRelatedtotheHigh/Low-LimitAlarmContactOutput]

Select the parameters for which the alarms are raised, and then set their respective high/low-limit set points. The meter transmits an alarm contact signal if any of these set points is exceeded. (Define the voltage-free contacts and the state of contacts at the service level).[FunctionsRelatedtotheFAILContactSignal]

If a FAIL state occurs, an appropriate message appears on the LCD display. For details on these functions, see Chapter 6.

3.1.8 The state of the contact output

Power SupplyOFFPower SupplyON

OPEN

NORMAL: CLOSEFAILURE: OPEN

CLOSE CLOSE CLOSE

The State of the contact output can be voluntarily set.(See Chapter 4 “Parameter Setting”.)

FAIL SPAN MAINT ALM

T3.2.ai

ZERO

CLOSE

3rd Edition : Jul. 10, 2015-00

Page 45: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<3. Operation> 3-8

IM 11T03C01-01E

3.2 Normal Operation In normal operation, there is no need for working with the GD410 meter except when calibrating it. Unless there is any failure found, carry out maintenance, inspection, and so on at the same time the meter is calibrated. The meter is designed to prohibit you from calibrating it or setting each parameter unless you have entered the password (331).

3.2.1 Starting OperationFollow the procedure below to start operation.

Completing these steps return to the <MEASURE> mode.

Turn on the external power switch.

Set the operating parameters.

Calibrate the analyzer for correct zero and span set points.

F3.5.ai

The following details the operations required in this procedure.(1) The meter, when turned on, indicates the calorific value. See Figure 4.1 for the operations

used to change to the required unit of calorie.(2) Set the relevant operating parameters by checking and modifying their default values in

order to meet individual operating requirements.(3) Supply the zero and span gases to check readings. If the readings contain errors, the meter

must be calibrated. For details on calibration, see Chapter 5.(4) When calibration is complete, the meter returns to the <MEASURE> mode.

3.2.2 Corrective Actions in Case of FailureIf the GD410 meter senses a failure, it sends out a FAIL contact output signal through terminals 18 and 19. In such a case, the output signal is held at either the value held immediately before or at the preset value for each setting.The message field indicates the type of error by the error code. Should a failure occur, check the nature of the failure and take corrective action without delay. Table 3.1 summarizes the numeric codes of errors that may occur in the measurement mode. For details on each type of error, see Table 6.2.

Table 3.1 Errors that May Occur in the Measurement Mode

Error No.Err. 01Err. 02Err. 03Err. 04Err. 05

Sensor shutdownFaulty oscillation frequency of sensorSensor's failure to detect temperatureFailure in A/D converterFailure in memory

Reset the power and contact service personnel.Reset the power and contact service personnel.Contact service personnel.Contact service personnel.Contact service personnel.

Nature of Failure Corrective Actions

T3.3.ai

3rd Edition : Jul. 10, 2015-00

Page 46: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<3. Operation> 3-9

IM 11T03C01-01E

3.2.3 Inspection and MaintenanceDue to the operating principle (polymodal self-excited oscillation circuitry), the sensor has only the slightest marginal drift, which is, in fact, generated by the detector itself or a result of contamination from dust, etc.. Therefore, the GD410 meter basically does not have to undergo any inspection or maintenance. Periodic inspection and maintenance are recommended however, to ensure optimum performance. For details on inspection and maintenance, see Chapter 6.

3.3 Shutdown and Restart3.3.1 Measures for Shutdown

Data set in the meter are retained even when the power is turned off. If the system needs to be shut down over a prolonged period, turn off the power. Any foreign matter that accumulates and hardens on the detector or any corrosive substance that condenses on the detector can be the cause of problems. Fully purge the sensor with fresh air (instrument air). Cleaning of the sensor is also recommended where it is appropriate; consult Yokogawa.

3.3.2 Measures for RestartingVisually check all component equipment of the meter before you resume measurement after a long-term shut down of the meter. Re-confirm the wiring and piping have tight connections .

3rd Edition : Jul. 10, 2015-00

Page 47: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

Blank Page

Page 48: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-1

IM 11T03C01-01E

4. PARAMETER SETTINGWhen using the GD410 analyzer, set data and select functions according to the measuring conditions. This chapter describes the procedures for setting various parameters.

4.1 Summary of Setting Operations4.1.1 Measurement, Operation, Setting and Service

Parameters can be set by selecting the appropriate mode. These modes are classified into four levels: measurement, operation, setting, and service.[MeasurementLevel]

This level is used to view various measured values, and includes a mode for selecting between items to be viewed in the measurement mode and another for selecting a parameter to be measured.[OperationLevel]

This is basically a level used to perform operations relating to routine inspection and/or maintenance such as calibration.[SettingLevel]

At this level, there are modes for setting data related to analog outputs, calibration data and alarm contact outputs.[ServiceLevel]

The GD410 analyzer has numerous functions. At this level, there are modes for selecting the functions necessary for system operation.

4.1.2 Key OperationsKeys can be operated in an "interactive" manner. Operate the keys according to the information in the data or message field, information pointed to by the pointer (mode indication), and for information in the key operation indicators. For basic key operations, see Subsection 3.1.4.[Interactions]

l Flashing pointerThis inquires whether you want to go to the mode being pointed at or move the pointer to the next mode. When the system indicates the modes for the setting or service level, the message is preceded by an asterisk (*). When you enter one of the modes, the pointer stops flashing and remains continuously lit.l Flashing key operation indicators

This prompts you to choose an indicator from the indicators flashing on the display and press the corresponding key.lFlashingnumeralsindatafield

This asks whether you want to change the flashing numeral or move to the next digit and have it flash. Press the appropriate key. If neither of these actions is required, press the [ENT] key.[AbortingtheSettingOperation]

Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement (<MEASURE>) mode to a mode at the operation level. It is also used to return to the measurement mode from other modes.

3rd Edition : Jul. 10, 2015-00

Page 49: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-2

IM 11T03C01-01E

4.1.3 Points to Be Noted When Making Settings(1) PasswordYou are not allowed to access any mode unless you enter the given password (331).A password prompt appears when:l at the operation level, the [MODE] key is pressed during the measurement mode,

orl at the setting level, the [*] key is pressed during the measurement mode.

(2) Related ParametersIf any change has been made to a particular parameter, check its relationship with parameters set in other modes so that there will not be any inconsistency.[ParameterstoCheckWhentheAnalogOutputSettingsAreChanged]

a. High/low-limit alarm set points and contact outputs

b. Fixed value for held output signal

[ParameterstoCheckWhenHigh/Low-limitAlarmSetpointsAreChanged]

a. Alarm actions (low-limit alarm or high-limit alarm operation)

b. Analog output settings

3rd Edition : Jul. 10, 2015-00

Page 50: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-3

IM 11T03C01-01E

4.2 Setting ParametersSubsections 4.2.1 to 4.2.4 show the setting parameters for each level.

4.2.1 Setting Parameters at Measurement LevelWhen turned on, the analyzer starts up in the measurement mode.

Table 4.1 Setting Parameters at Measurement Level

Read-onlyRead-onlyRead-onlyRead-only

Measurement mode Indication of measurement value Physical density Corrected density Calorific value Indication of message Unit Error symbol and number Alarm mark and number Indication of HOLD and FAIL HOLD FAIL

DISP mode Physical density Corrected density Calorific value Measured gas temperature Measured gas pressure Analog output 1 (%) Analog output 2 (%)

KG/M3KG/NM3MJ/NM3

Err. NOALM. NO

HOLDFAIL

DISP KG/M3

KG/NM3 MJ/NM3

°CKPA or MPA

MA1% MA2%

Mode/Setting parameter Display Data to Be Set (or Conditions)

Set the unit in CODE 20.

RemarksSee Figure 4.1 for key operation.The unit is the one set in the DISP mode.

The unit is the one set in the DISP mode.Appear if an error occurs.Appear if an alarm occurs.Appear even during other modes.Appear during a HOLD state.Appear if a failure occurs.

Selection of parameters to be measured in the measurement mode.

T4.1.ai

4.2.2 Setting Parameters at Operation LevelPress the [MODE] key and enter the password (331) to gain access to this level (see Figure 4.2 for key operation).

Table 4.2 Setting Parameters at Operation Level

Mode/Setting parameter Display Data to Be Set (or Conditions) Remarks

T4.2.ai

See Figure 5.3 for key operation.See Figure 5.4 for key operation.

Calibration mode Semi-automatic calibration Manual calibration Setting of zero-point calorific value Setting of zero-point density Setting of span-point calorific value Setting of span-point density

SEM.CALMAN.CAL*Z_CAL*Z_DNS*S_CAL*S_DNS

One-touch calibration

No setting is required for manual calibration.Control of zero-gas valveControl of span-gas valveSwitching between sample and calibration gases

Zero/span-gas valve operation

Zero-gas valve operation Span-gas valve operation Sample-gas valve operation

VALVE

See Figure 5.1 for key operation.

Density/calorie conversion factor setting Automatic zero-point adjustment Factor setting Zero point Factor K0 Factor K1

ADJUSTAT_ADJSET_KK0_ADJ

K0K1

The decimal point can be repositioned for this setting only.

Range: -99999 to 99999Range: -99999 to 99999Range: 0 to 99999

See Figure 4.3 for key operation.

(Floating-point)

3rd Edition : Jul. 10, 2015-00

Page 51: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-4

IM 11T03C01-01E

4.2.3 Setting Parameters at Setting LevelPress the [*] key and enter the password (331) to gain access to this level (see Figure 4.4 for key operation).

Table 4.3 Setting Parameters at Setting LevelMode/Setting parameter Display Data to Be Set (or Conditions) Remarks

T4.3.ai

Alarm-point settingSelect a parameter for which an alarm is set. Physical density Corrected density Calorific valueSet the high/low limits. Physical density Corrected density Calorific value

*ALARM

*DENS*C_DNS*CALRY

*L_DNS,*H_DNS*L_CP.D,*H_CP.D*L_CAL,*H_CAL

0.0000 to 6.00000.0000 to 6.00000.000 to 133.000

See Figure4.5 for key operation.

No setting is required if the parameter is temperature or pressure.

Analog output setting Output1 Set a parameter to be output.

Set the zero and span points. Physical density Corrected density Calorific value Output 2 (same as output 1)

Setting of calibration data Zero-point calorific value Zero-point density Span-point calorific value Span-point density Output hold value during calibration ・Disable: 0 ・Enable (value immediately before): 1 ・Enable(preset value): 2 Setting of preset value (Only if the preceding parameter is set to 2 (preset value))

*RANGE*OUT1*DENS

*C_DNS*CALRY*'TEMP*PRESS

*Z_DNS,*S_DNS*Z_CP.D,*S_CP.D*Z_CAL,*S_CAL

*OUT2

*CAL.DT*Z_CAL*Z_DNS*S_CAL*S_DNS*C_HLD

*PR.SET

Physical densityCorrected densityCalorific valueGas temperatureGas pressure 0.0000 to 6.00000.0000 to 6.00000.000 to 133.000

0.000 to 133.0000.0000 to 6.00000.000 to 133.0000.0000 to 6.0000

See Figure 5.2 for key operation.

See Figure 4.6 for key operation.

4.2.4 Setting Parameters at Service Level Press the [*] key and enter the password (331) to select the service level. When you select this level, the analyzer shows a Code No. promptly. Type the appropriate code number and press the [ENT] key.

3rd Edition : Jul. 10, 2015-00

Page 52: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-5

IM 11T03C01-01E

Table 4.4 Setting Parameters at Service Level

Display Data to Be Set (or Conditions) Remarks

T4.4.ai

Code No. Item

See Figure 4.8 for key operation.

If “preset value” is selected in the preceding parameter.

Setting of hold during maintenance Hold enable/ disable

Setting of the preset hold value

*M_HLD

*PR.SET

・Disable: 0・Enable (value immediately before): 1・Enable (preset value): 2-10.0 to 110.0

See Figure 4.10 for key operation.See Figure 4.11 for key operation.See Figure 4.12 for key operation.

If “fixed value” is selected in the preceding parameter.

Setting of output-smoothing constants Setting of slates of contact outputs Setting of pressure Correction Enalbe/disable/fixed value

Setting of fixed value for pressure correction

*SMOTH

*CNTCT

*P.COMP

*P.FIX

00 to 60 00 to 15

・Disable: 0・Enable (measured value ): 1・Enable (fixed value): 20.1 to 99999 kPa, 0.0001 to 9.9999MPa

Setting for measuring corrected density Setting of reference temperature Setting of reference pressureSetting of gas pressure Setting of zero-gas pressure Setting of span-gas pressure

*C.D.TMP*C.D.PRS

*Z_PRS*S_PRS

-20 to 80.00.1 to 999.99kPa, 0.0001 to 9.9999MPa

0.1 to 999.99kPa, 0.0001 to 9.9999MPa0.1 to 999.99kPa, 0.0001 to 9.9999MPa

・Disable: 0・Enable: 1

・Disable: 0・Enable: 1

Selection for automatic calibration

Selection for remote semi-automatic calibration

Setting of data for automatic/ semi- automatic calibration

*AUTO.C

*REMOT

*CAL.P Zero and span: 0Zero: 1Span: 200 to 5900 to 59

00.01.01 to 99.12.3100.00 to 23.59Hour: 0Day: 1Hour: 0 to 23Day: 0 to 255

Selection of calibration item

Calibration time Stabilization time Starting time of calibration Setting of year/month/day Setting of hour/minute Setting of unit of calibration interval

Calibration interval

*CAL.T*STAB.T

*Y_M_D*H_M

*CYCL.U

*CYCL.T

Selection of pressure unit *PRES.U ・kPa: 0・MPa: 1

00.01.01 to 99.12.3100.00 to 23.59

Setting of time on real-time clock Setting of year/month/day Setting of hour/minute

Showing/hiding of negative measured value (-)

*Y_M_D*H_M

*MINUS ・Show: 0・Hide: 1

Indication of calibration coefficients(read-only) Zero SpanIndication of oscillation frequency (read only) Frequency: F2 Frequency: F4 Frequency ratio

*C_K_Z*C_K_S

*F2.KHZ*F4.KHZ*F2/F4

Indication of software version (read-only)

Setting of high-resolution mode

*REV

*S_CYC

SX.XXXXX.XXXX

X.XXXXX.XXXXX.XXXX

・Normal resolution: 0・High resolution:1

See Figure 4.13 for key operation.

See Figure 4.14 for key operation.

See Figure 4.15 for key operation.

See Figure 4.16 for key operation.

See Figure 4.17 for key operation.

See Figure 4.18 for key operation.See Figure 4.19 for key operation.For leap years, the setting can cope with ‘February 29’ .See Figure 4.20 for key operation.

See Figure 4.21 for key operation.

See Figure 4.22 for key operation.

See Figure 4.23 for key operation.

See Figure 4.24 for key operation.

See Figure 4.9 for key operation.

If “preset value” is selected in the preceding parameter.

・Disable: 0・Enable (value immediately before):.1・Enable (preset value): 2-10.0 to 110.0

*E_HLD

*PR.SET

Setting of hold against errors Hold enable/ disable

Setting of the preset hold value

01

02

04 05 10

11

12

13

14

15

20

30

31

40

41

42

43

3rd Edition : Jul. 10, 2015-00

Page 53: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-6

IM 11T03C01-01E

4.3 Parameter Setting[AbortingtheSettingOperation]

Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement (<MEASURE>) mode to a mode at the operation level. It also used to return to the measurement mode from other modes.

4.3.1 MessageDisplay/SelectionMode Select one of the following seven parameters to be shown in the message field. Each press of the [NO] key cycles through the choices in the order shown below:1. Physical density 『X.XXXX KG/M3』2. Corrected density 『X.XXXX KG/NM3』3. Calorific value 『XXX.XXX MJ/NM3』4. Temperature 『XXX.X °C』 5. Pressure 『XXX.XX KPA』 or 『X.XXXX MPA』 (depends on the setting of CODE 20)6. Current for analog output 1 (%) 『XXX.X MA1%』7. Current for analog output 2(%) 『XXX.X MA2%』 Show the desired choice and press the [YES] key. That choice is set as a parameter to be measured in the measurement mode. Parameters 1 to 3 are configurable, while parameters 4 to 7 are read-only.

3rd Edition : Jul. 10, 2015-00

Page 54: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-7

IM 11T03C01-01E

【NO】

【YES】

【NO】

F4.1.ai

DISP Measurement Mode

【NO】

【NO】

【NO】

【NO】

【NO】

【NO】

【YES】* 1

【YES】* 1

【YES】* 1

【YES】

【YES】* 2

【YES】

【YES】* 1

Measurement Mode

X.XXXXKG/M3

(Physical density)

X.XXXXKG/NM3

(Corrected density)

XXX.XXXMJ/NM3

(Calorific value)

XXX.X°C

(Measured gas temperature)

XXX.XXKPA

(Measured gas pressure)

XXX.XMA1%

(Current for analog output1(%))

XXX.XMA2%

(Current for analog output2(%))

*1: Press the 【YES】 key to show the parameter in the measurement mode.*2: The message field shows “MPA” if MPa is choosen for the pressure unit.

Measurement Mode【NO】

Figure4.1 MessageDisplay/SelectionMode(DISPMode)

3rd Edition : Jul. 10, 2015-00

Page 55: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-8

IM 11T03C01-01E

4.3.2 Parameter Setting at Operation Level

(1) Operational LevelTo move to the operation mode, press the [MODE] key and then enter the password (331).Select a mode from the following four choices of the operation level. Each press of the [NO] key cycles through the choices in the order shown below: 1. Semi-automatic calibration mode 『SEM.CAL』 2. Manual calibration mode 『MAN.CAL』 3. Valve operation mode 『VALVE』 4. Factor setting mode 『ADJUST』 Press the [YES] key to enter the desired mode when it is indicated.

331

【ENT】

【NO】

【NO】

【NO】

【YES】

【YES】

【YES】

【YES】

【YES】

【YES】

【MODE】

Semi-automatic calibration mode

Manual calibration mode

Valve operation mode

Measurement Mode

*PASSW

SEM.CAL

MAN.CAL

VALVE

See the process in Figure 5.3.

See the process in Figure 5.4.

See the process in Figure 5.1.

F4.2.ai【NO】

【YES】 【YES】Density/calorie conversion factor setting modeADJUST

See the process in Figure 4.3.

Figure 4.2 Modes of Operation Level

3rd Edition : Jul. 10, 2015-00

Page 56: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-9

IM 11T03C01-01E

(2)Density/CalorieConversionFactorSettingMode[ADJUST]A caloric value is derived from the measured density value in the following equation, H=K1 x r +K0+K0_ADJWhere H is the caloric value, r is the density. K1, K0, and K0_ADJ are the conversion factors. Normally, K0 and K1 factors are automatically set (while the K0_ADJ factor is "0”) when the zero and span calibrations are performed with inputs of the density and the caloric value. After the calibration, an error may be generated between the actual caloric value of the gas of interest, Hm, and the reference caloric value, Hs, obtained from, for example, gas chromatograph. In this case the K0_ADJ factor is used to compensate for the error. Figure 4.2 illustrates how to set the factors.

・ As explained above, K0 and K1 factors are automatically set when the density and caloric values of the zero and span gases, respectively, are entered upon calibration. These factors can also be specified by direct input of the values in the [SET_K] mode.

・ When reference caloric value, Hs, is set for [VALUE] and actual measured value, Hm, is set for [CAL.SET] in the [AT_ADJ] mode, the compensation value for the error between Hs and Hm will be automatically set for [K0_ADJ]. Without setting values in the [AT_ADJ] mode, the same thing can also be achieved by direct setting of the compensation value for the error between Hs and Hm for [K0_ADJ]. If the compensation factor is set both through the [AT_ADJ] mode and by directly setting for [K0_ADJ], only the last-set factor will be valid.

a. Automatic zero-point adjustment [AT_ADJ]: 0 to133.000b. Factor setting [SET_K]:

・Zero point [K0_ADJ]: -99999 to 99999

・K0 [K0]: -99999 to 99999

・K1 [K1]: -99999 to 99999Note: The result of executing the above item is also reflected on [K0_ADJ] in item b.

【NO】

【ENT】【ENT】

ADJUST【NO】

XXX.XXXVALVE

VALVE SEM.CAL

END

X.XXXXK0_ADJ

XXX.XXXCAL.SET

X.XXXXK0

X.XXXXK1

Measurement mode

【YES】

【NO】

【NO】

F4.3.ai

【YES】

*2

【ENT】

AT_ADJ SET_K

【YES】【YES】

*3

【ENT】【ENT】

*1: Using the 【 】, 【 】 and 【ENT】 keys, set a value (numeral position of the decimal point and sign). Each press of the 【 】 key moves through the fields of the numeral decimal point and sign in this order. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.*2: Set a value using the 【 】, 【 】 and 【ENT】 keys.*3 When the reading stabl1izes, press the 【ENT】 key to confirm the factor. This parameter does not accept input from the 【 】 and 【 】 keys.

*1

*1

*1

【NO】

Figure4.3 Density/CalorieConversionFactorSettingMode

3rd Edition : Jul. 10, 2015-00

Page 57: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-10

IM 11T03C01-01E

4.3.3 Parameter Setting at Setting LevelTo move to the setting level, Press the [*] key and then enter the password (331).The main task at the setting level is to set data values such as the measuring range values.Note that the function for which you have set data values at the setting level, will not work if that function is turned off at another level. For this reason, care must be taken when dealing with modes that relate to each other. The setting level has four levels/modes, as shown below, although this subsection explains only the analog output setting and alarm-point setting modes. The calibration data setting mode is discussed in Chapter 5.1. Analog output setting mode 『*RANGE』2. Calibration parameter setting mode 『*CAL.DT』3. Alarm-point setting mode if 『*ALARM』4. Service level. 『*SERVC』Select a mode from the above four choices. Each press of the [NO] key cycles through the choices in sequence. Press the [YES] key to enter the desired mode when it is indicated.

You can abort the setting operation in any of the modes by pressing the【MODE】key.Doing this will return the analyzer to the measurement mode.

331

【ENT】

【NO】

【NO】

【NO】

【YES】

【YES】

【ENT】

【ENT】

【ENT】

【YES】

【*】

Analog output setting mode

Calibration parameter setting mode

Alarm-point setting mode

Measurement Mode

*PASSW

*RANGE

*CAL.DT

*ALARM

See the process in Figure 4.5.

See the process in Figure 5.2.

See the process in Figure 4.6.

F4.4.ai

【YES】 【ENT】Service level

See the process in Figure 4.7 to 4.24.

*SERVC

【NO】

Figure 4.4 Process at the Setting Level

3rd Edition : Jul. 10, 2015-00

Page 58: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-11

IM 11T03C01-01E

1. Analog Output Setting Mode 『*RANGE』

The analog output has output 1 and output 2. Only output 1 can be used for communication purposes. Set a range appropriate for 4-20 mA DC output signals. The range has two set points: the Lower limit (zero point) of the range corresponding to the minimum (0%) of a given output signal and the upper limit (span point) of the range corresponding to the maximum (100%) of the given output signal.(1) Output1 『*OUT1』 a. Select a parameter to be output.

Physical density 『*DENS』

Corrected density 『*C_DNS』

Calorific value 『*CALRY』

Gas temperature 『*TEMP』

Gas pressure 『*PRESS』

b. Set the zero and span points. (This item is not applied if temperature or pressure are chosen as the parameter in the above item.)

Physical density 『*Z_DNS』, 『*S_DNS』: a value from 0.0000 to 6.0000 (kg/m3)

Corrected density 『*Z_CP.D』, 『*S_CP.D』: a value from 0.0000 to 6.0000 (kg/Nm3)

Calorific value 『*Z_CAL』, 『*S_CAL』: a value from 0.000 to 133.000 (MJ/Nm3)

(2) Output 2 『*OUT2』The parameters of Output 2 is the same as those of output 1. Select and set by reffering to the above.

3rd Edition : Jul. 10, 2015-00

Page 59: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-12

IM 11T03C01-01E

*1: You can abort the setting operation in any of the modes by pressing 【MODE】 key. Doing this will return the analyzer to the measurement mode.*2: The display shows the previous parameter first.*3: Each press of the【NO】key toggles between the parameters『OUT1』and『OUT2』

【NO】

【ENT】

【ENT】

*RANGE

【ENT】

【NO】

X.XXXX*Z_DNS

*SERVC *CAL.DT

*DENS X.XXXX*S_DNS

X.XXXX*Z_CP.D

X.XXXX*S_CP.D

XXX.XXX*Z_CAL

XXX.XXX*S_CAL

(Physical density Range)

(Corrected density Range)

(Calorific value Range)

*CALRY

【ENT】

【ENT】

【ENT】

【YES】

【YES】

【YES】

【NO】

【NO】

【NO】

*C_DNS

F4.5.ai

【YES】

*2

(Gas temperature)

(Gas pressure)

*TEMP

*PRESS

【NO】

【NO】

【YES】

【YES】

【NO】*OUT1 *OUT2

【YES】【YES】

【ENT】

331*PASSW

Measurement mode

【*】

*3

Figure 4.5 Process of the Analog Output Setting Mode

3rd Edition : Jul. 10, 2015-00

Page 60: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-13

IM 11T03C01-01E

2. Alarm-point Setting Mode 『*ALARM』Set the upper /lower limits for the measured values of the parameters noted below, for the purpose of raising an alarm.a. Select a parameter for which alarms are raised.

・Physical density 『*DENS』

・Corrected density 『*C_DNS』

・Calorific value 『*CALRY』

Press the [YES] key to enter the desired mode when it is indicated.b. Set the upper/lower limits.

・Physical density 『*L_DNS』, 『*H_DNS』: a value from 0.0000 to 6.0000 (kg/m3)

・Corrected density 『*L_CP.D』, 『*H_CP.D』: a value from 0.0000 to 6.0000 (kg/Nm3)

・Calorific Value 『*L_CAL』, 『*H_CAL』 :a value from 0.000 to 133.000 (MJ/Nm3)

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10.*2: The display shows the previous parameter first.

【NO】

【ENT】

【ENT】

*ALARM

【ENT】

【NO】

X.XXXX*L_DNS

*CAL.DT *SERVC

*DENS X.XXXX*H_DNS

X.XXXX*L_CP.D

X.XXXX*H_CP.D

XXX.XXX*L_CAL

XXX.XXX*H_CAL

(Physical density)

(Corrected density)

(Calorific value)

*CALRY

【ENT】

【ENT】

【ENT】

【YES】

【YES】

【YES】

【NO】

【NO】

【NO】

*C_DNS

F4.6.ai

【YES】

*1

*2

Figure 4.6 Process of the Alarm-point Setting Mode

3rd Edition : Jul. 10, 2015-00

Page 61: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-14

IM 11T03C01-01E

4.3.4 Parameter Setting at Service LevelTo move to the service level, select 『*SERVC』 at the setting level, and press the [YES] key. The main task at the setting level is to set data values such as a measuring range. To go into each mode of the service level, enter the appropriate code number.

CAUTION DO NOT enter codes other than those listed in Table 4.4. A change to the settings of any unlisted code can result in the failure of the analyzer to operate properly. If you have entered a wrong code number by mistake, press the [MODE] key to return to the measurement mode. You may exit the mode of that wrong code by pressing the [ENT] key, it you haven't made any change to the relevant data.

When『*SERVC』is shown on the display, press【YES】key. The message field changes to look like the one shown left.

Press the【ENT】key with code number 01 shown on the display, for example. The analyzer enters the mode for setting the parameters of that code.

Figure 4.7 Entry of Code Number at Service Level

The following explains the parameter setting procedures in the order of code numbers.

3rd Edition : Jul. 10, 2015-00

Page 62: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-15

IM 11T03C01-01E

CODE01:SettingofHoldduringMaintenance This setting determines whether or not the output signal is held during maintenance. As a value to be held, you can select either a value immediately before or a preset value.a. Hold enable/disable 『*M_HLD』

・Disable: 0

・Enable: 1

・Enable (preset value): 2

b. Setting of preset hold value 『*PR.SET』This item is available when the above item is set to "2".Configurable range: -10.0 to 110.0 (percent of analog output)

【ENT】

01*CODE

*SERVC

X*M_HLD

XXX.X*PR.SET

【ENT】

【ENT】

【YES】

F4.8.ai

*1

*2

*3

*4

*5

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10.*2: Enter the code number.*3: Set a value using the 【 】 and 【ENT】 keys.*4: Set a value using the【 】, 【 】 and 【ENT】 keys. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.*5: Setting 0 or 1 returns to the service level with the indication 『*SERVC』.

Figure4.8 ProcessofSettingtheHoldFunctionforDuringMaintenance

3rd Edition : Jul. 10, 2015-00

Page 63: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-16

IM 11T03C01-01E

CODE02:SettingtheHoldFunctionintheEventofErrorThis setting determines whether or not the output signal is held if an error occurs. You can select either the value immediately before the error occurence or a preset value as the value to be held.a. Hold enable/disable 『*E_HLD』

・Disable: 0

・Enable: 1

・Enable (preset value): 2

b. Setting of preset hold value 『*PR.SET』This item is available when the above item is set to "2".Configurable range: -10.0 to 110.0 (percent of analog output)

【ENT】

02*CODE

*SERVC

X*E_HLD

XXX.X*PR.SET

【ENT】

【ENT】

【YES】

F4.9.ai

*1*2

*3

*4

*5

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Set a value using the 【 】 and 【ENT】 keys.*4: Set a value using the【 】, 【 】 and 【ENT】 keys. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.*5: Setting 0 or 1 returns to the service level with the indication 『*SERVC』.

Figure4.9 ProcessofSettingtheHoldFunctionintheEventofError

CODE 04: Setting of Output-smoothing Constants 『*SMOTH』This setting defines the constants for output smoothing.The configurable range is 00 to 60.

【ENT】

04*CODE

*SERVC

XX*SMOTH

【ENT】

【YES】

F4.10.ai

*1*2

*3

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Set a value using the【 】, 【 】 and 【ENT】 keys. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.

Figure 4.10 Setting Output-smoothing Constants

3rd Edition : Jul. 10, 2015-00

Page 64: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-17

IM 11T03C01-01E

CODE 05: Setting of Contact Outputs States『*CNTCT』 This setting defines the states of the contact outputs.The configurable range is 00 to 15.

【ENT】

05*CODE

*SERVC

XX*CNTCT

【ENT】

【YES】

F4.11.ai

*1

*2

*3

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Set a value using the【 】, 【 】 and 【ENT】 keys. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.

Figure 4.11 Process of Setting the State of Contact Outputs

Each contact output takes either of the following two states (in table); depending on the value you set.

0 NO NO NO NO

1 NO NO NO NC

2 NO NO NC NO

3 NO NO NC NC

4 NO NC NO NO

5 NO NC NO NC

6 NO NC NC NO

7 NO NC NC NC

8 NC NO NO NO

9 NC NO NO NC

10 NC NO NC NO

11 NC NO NC NC

12 NC NC NO NO

13 NC NC NO NC

14 NC NC NC NO

15 NC NC NC NC

Value Zero contact SPAN contact MAINT contact ALM contact

T4.5.aiNO: means "normally open"NC: means "normally closed"

3rd Edition : Jul. 10, 2015-00

Page 65: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-18

IM 11T03C01-01E

CODE 10: Setting of Pressure CorrectionThis setting determines whether or not the measured density should be corrected by the gas pressure. The density can be corrected either by receiving a signal from a pressure transmitter or by using a fixed pressure setpoint.If you want the analyzer to correct the density using a pressure transmitter, you MUST connect a pressure transmitter to the converter and enter the pressure input and zero and span points with CODE 12 at the service level.If you do not want the analyzer to correct the density, the analyzer shows and outputs the measured density equivalent to a value at an atmospheric pressure of unity (101.33 kPa ABS).a. Enable/disable/select preset value 『*P.COMP』

・Disable: 0

・Enable: 1

・Enable (fixed value): 2

b. Setting of fixed value for pressure correction 『*P.FIX』This item is available when the above item is set to "2".

For kPa, 0.10 to 999.99

For MPa, 0.0001 to 9.9999

【ENT】

10*CODE

*SERVC

X*P.COMP

X.XXXX*P.FIX

【ENT】

【ENT】

【YES】

F4.12.ai

*1

*2

*3

*4

*5

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Set a value using the 【 】 and 【ENT】 keys,*4: Set a value using the【 】, 【 】 and 【ENT】 keys. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.*5: Setting 0 or 1 returns to the service level with the indication 『*SERVC』.*6: The position of the decimal point changes depending on the setting of CODE 20. If the code is set to “0” (kPa), the reading is XXX.XX; if the code is set to “1” (MPa), the reading is X.XXXX.

*6

Figure 4.12 Process of Setting the Pressure Correction

3rd Edition : Jul. 10, 2015-00

Page 66: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-19

IM 11T03C01-01E

CODE 11: Setting for Corrected Density MeasurementThis setting defines the reference temperature and pressure required to obtain the corrected density.

a. Setting of reference temperature 『*C.D.TMP』a value from -20.0 to 80.0 (ºC)

b. Setting of reference pressure 『*C.D.PRS』For kPa, a value from 0.1 to 999.99

For MPa, a value from 0.0001 to 9.9999

【ENT】

11*CODE

*SERVC

XX.X*C.D.TMP

X.XXXX*C.D.PRS

【ENT】

【ENT】

【YES】

F4.13.ai

*1

*2

*3

*4

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Set a value using the【 】, 【 】 and 【ENT】 keys. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.*4: The position of the decimal point changes depending on the setting of CODE 20. If the code is set to “0” (kPa), the reading is XXX.XX; if the code is set to “1” (MPa), the reading is X.XXXX.

*3

Figure 4.13 Process of Setting for Corrected Density Measurement

3rd Edition : Jul. 10, 2015-00

Page 67: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-20

IM 11T03C01-01E

CODE 12: Setting of Gas PressureThis setting defines the zero and span points of the pressure transmitter's measurement range.

a. Setting of zero-point gas pressure 『*Z_PRS』For kPa, a value from 0.1 to 999.99

For MPa, a value from 0.0001 to 9.9999

b. Setting of span-point gas pressure 『*S_PRS』For kPa, a value from 0.1 to 999.99

For MPa, a value from 0.0001 to 9.9999

The zero and span points set here are also used for the analog output.

【ENT】

*CODE12*SERVC

X.XXXX*Z_PRS

X.XXXX*S_PRS

【ENT】

【ENT】

【YES】

F4.14.ai

*1

*2

*3

*4

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Set a value using the【 】, 【 】 and 【ENT】 keys. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.*4: The position of the decimal point changes depending on the setting of CODE 20. If the code is set to “0” (kPa), the reading is XXX.XX; if the code is set to “1” (MPa), the reading is X.XXXX.

*3

*4

Figure 4.14 Process of Setting Gas Pressure

CAUTIONSet 『*Z_PRS』to the same zero-point gas pressure as the zero-point pressure of the pressure transmitter (set point value of 4 mA). Set 『*S_PRS』to the same span-point gas pressure as the span-point pressure of the pressure transmitter (set point value of 20 mA). If the zero- and span-point pressure setting values between the GD410 and the transmitter is different, measurement deviation will occur.

3rd Edition : Jul. 10, 2015-00

Page 68: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-21

IM 11T03C01-01E

CODE 13: Selection for Automatic Calibration 『*AUTO.C』This setting determines whether or not automatic calibration is carried out. If automatic calibration is to be carried out, you must also configure CODE 15 at the service level.

・Disable: 0

・Enable: 1

【ENT】

*CODE13*SERVC

X*AUTO.C

【ENT】

【YES】

F4.15.ai

*1

*2

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10.*2: Enter the code number.*3: Set a value using the【 】, 【 】 and 【ENT】 keys.

*3

.Figure4.15 ProcessofEnabling/DisablingAutomaticCalibration

CODE 14: Setting for Remote Semi-automatic Calibration 『*REMOT』This setting determines whether or not semi-automatic calibration is carried out remotely. If remote semi-automatic calibration is to be carried out, you must wire the contact input.

・Disable: 0

・Enable: 1

【ENT】

*CODE14*SERVC

X*REMOT

【ENT】

【YES】

F4.16.ai

*1

*2

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10.*2: Enter the code number.*3: Set a value using the【 】, 【 】 and 【ENT】 keys.

*3

.Figure4.16 ProcessofEnabling/DisablingRemoteSemi-automaticCalibration

CAUTIONWhen the contact input is closed, semi-automatic calibration starts. If semi-automatic calibration is desired again after the first semi-automatic calibration, open the contact input, and close it again.

3rd Edition : Jul. 10, 2015-00

Page 69: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-22

IM 11T03C01-01E

CODE15:SettingofDataforAutomatic/Semi-automaticCalibrationThis setting defines the data needed to carry out automatic/semi-automatic calibration.For more information about calibration, see Chapter 5, "Calibration Procedure".(a) Selection of calibration item『*CAL.P』

Zero and span: 0

Zero: 1

Span: 2

(b) Calibration time『*CAL.T』 a value from 00 to 59

(c) Stabilization time『*STAB.T』 a value from 00 to 59

(d) Starting time of calibration (1) Setting of year/month/day『*Y_M_D』

a value from 00.01.01 to 99.12.31

(2) Setting of hour/minute『*H_M』 a value from 00.00 to 23.59

(e) Setting of unit of calibration interval『*CYCL.U』

Hour: 0

Day: 1

(f) Calibration interval『*CYCL.T』 When the above item (e) is set to "0", a value is from 00 to 23 (as hour).

When the above item (e) is set to "1", a value is from 00 to 255 (as day).

3rd Edition : Jul. 10, 2015-00

Page 70: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-23

IM 11T03C01-01E

X*CAL.P

(Selection of calibration item)

XX*CAL.T

(Setting of calibration time)

XX*STAB.T

(Setting of stabilization time)

XX.XX.XX*Y_M_D

(Setting of calibration starting time [year/month/day] )

XX.XX*H_M

(Setting of calibration starting time [hour/minute] )

X*CYCL.U

(Selection of calibration interval unit)

XXX*CYCL.T

(Selection of calibration interval)

【ENT】

*SERVC

【ENT】

【YES】

F4.17.ai

*1

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Set a value using the【 】, 【 】 and 【ENT】 keys. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.*3: The number of digits on LCD varies depending on the unit of the interval unit (parameter *CYCL.U). If the parameter is set to “0” (hour), the indication is two-digit number. If the parameter is set to “1” (day), the indication is three-digit number.

15*CODE *2

*2

*2

*2

*2

*2

*2

*2*3

【ENT】

【ENT】

【ENT】

【ENT】

【ENT】

【ENT】

Figure 4.17 Process of Setting Data for Automatic Semi-automatic Calibration

3rd Edition : Jul. 10, 2015-00

Page 71: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-24

IM 11T03C01-01E

CODE 20: Selection of Pressure Unit 『*PRES.U』This setting defines the pressure unit for the gas pressure transmitter.

・kPa: 0

・MPa: 1

【ENT】

*CODE20*SERVC

X*PRES.U

【ENT】

【YES】

F4.18.ai

*1

*2

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Set a value using the【 】, 【 】 and 【ENT】 keys.

*3

Figure 4.18 Process of Setting Pressure Unit keys.

CODE 30: Setting on Time on Real time ClockThis setting defines the data and time. a. Setting of year/month/day 『*Y_M_D』: configurable range = 00.01.01 to 99.12.31 b. Setting of hour/minute 『*H_M』: configurable range = 00.00 to 23.59

【ENT】

30*CODE

*SERVC

XX.XX.XX*Y_M_D

XX.XX*H_M

【ENT】

【ENT】

【YES】

F4.19.ai

*1

*2

*3

*3

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Set a value using the【 】, 【 】 and 【ENT】 keys. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.

Figure 4.19 The Process on Setting Time of Real-time Clock

3rd Edition : Jul. 10, 2015-00

Page 72: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-25

IM 11T03C01-01E

CODE31:Showing/HidingofNegativeMeasuredValues 『*MINUS』This setting determines whether negative measured values (-) should be shown or hidden.

・Show: 0

・Hide: 1

【ENT】

*CODE31*SERVC

X*MINUS

【ENT】

【YES】

F4.20.ai

*1

*2

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Set a value using the【 】, 【 】 and 【ENT】 keys.

*3

Figure4.20 ProcessofShowing/HidingNegativeMeasuredValues

CODE40:IndicationofCalibrationCoefficients(Read-only) This setting determines the calibration coefficients to be indicated.

a. Zero 『*C_K_Z』: indicated as SX.XXXX

b. Span 『*C_K_S』: indicated as X.XXXX

【ENT】

40*CODE

*SERVC

SX.XXXX*C_K_Z

X.XXXX*C_K_S

【ENT】

【ENT】

【YES】

F4.21.ai

*1

*2

*3

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: By pressing the 【ENT】 key, the next indication is displayed. This parameter does not accept the【 】 and 【 】 key input.

Figure4.21 ProcessofSettingtheCalibrationCoefficientstoBeIndicated

3rd Edition : Jul. 10, 2015-00

Page 73: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-26

IM 11T03C01-01E

CODE 41: Indication of Oscillation Frequency (Read-only)This setting determines the F2 and F4 oscillation frequencies coefficients to be indicated.

a. Indication of frequency F2 (kHz) 『*F2.KHZ』: indicated as X.XXXXX

b. Indication of frequency F4 (kHz) 『*F4.KHZ』: indicated as X.XXXXX

c. Indication of frequency ratio (kHz) 『*F2/F4』: indicated as X.XXXXX

【ENT】

41*CODE

*SERVC

X.XXXXX*F2.KHZ

X.XXXXX*F4.KHZ

【ENT】

【ENT】

【YES】

F4.22.ai

*1

*2

*3

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: By pressing the 【ENT】 key, the next indication is displayed. This parameter does not accept the【 】 and 【 】 key input.

【ENT】

X.XXXXX*F2/F4

*3

*3

Figure 4.22 Process of Setting the Oscillation Frequencies to Be Indicated

CODE 42: Indication of Software Version (Read-only)『*REV』 This setting allows the software version to be confirmed (indicated).

42*CODE

*SERVC

X.XX*REV

【ENT】

【ENT】

【YES】

F4.23.ai

*1

*2

*3

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Pressing the 【ENT】 key returns to the service level with the indication 『*SERVC』. This parameter does not accept the【 】 and 【 】 key input.

Figure 4.23 Process of Setting the Software Version to Be Indicated

3rd Edition : Jul. 10, 2015-00

Page 74: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<4. Parameter Setting> 4-27

IM 11T03C01-01E

CODE43:SettingofHigh-resolutionMode『*S_CYC』This setting selects between the normal mode and the high-resolution mode.

・Normal mode: 0

・High-resolution mode: 1

43*CODE

*SERVC

X*S_CYC

【ENT】

【ENT】

【YES】

F4.24.ai

*1

*2

*3

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Enter the code number.*3: Set a value by using the【 】, 【 】 and 【ENT】 keys.

Figure4.24 ProcessofSelectingtheHigh-resolutionMode

3rd Edition : Jul. 10, 2015-00

Page 75: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

Blank Page

Page 76: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<5. Calibration Procedure> 5-1

IM 11T03C01-01E

5. CALIBRATION PROCEDUREZero- and span-point calibration can be carried out using standard gases (zero and span gases). Three modes are available for calibration: automatic, semi-automatic (remote semi-automatic) and manual. Even if automatic calibration is enabled in the parameter setting, you can interrupt the system with either the semi-automatic or manual mode of calibration. The following explains the basic procedure and operation used in calibration.

5.1 Basic Calibration Procedure

Semi-automatic mode

Entry of Calibration data See the process in Figure 5.2

Entry of density/calorie conversion factor See the process in Figure 4.2

Selection of semi-automatic/manual calibration See the process in Figure 4.1

See the process in Figure 5.4

See the process in Figure 5.3.

Manual mode

F5.1.ai

5.2 Preparation for CalibrationPrepare for calibration during the initial operation of the analyzer, by carrying out the following.

・ At the operation level, set the parameters for semi-automatic and manual calibration. Also set the valve-operation parameter for switching between gases to check the valve perform-ance.

・ At the setting level, set the calibration data (zero- and span-point values, states of output during calibration etc.).

・ At the service level, determine whether automatic and remote semi-automatic calibrations are enabled or disabled in CODE 13 and 14. Set the calibration data (selection of calibration points, calibration time, stabilization time, etc.) in CODE 15.

5.2.1 Setting Type of Calibration and Checking Valve Operation

(1) Operations to Set the Semi-automatic and Manual CalibrationIn the measurement mode, press the [MODE] key, enter the password (331), and select the desired type of calibration.

・ To select semi-automatic calibration, show 『SEM.CAL』 and press the [YES] key.

・ To select manual calibration, show 『MAN.CAL』 and press the [YES] key.

For details on this process, see Figure 4.2.

3rd Edition : Jul. 10, 2015-00

Page 77: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<5. Calibration Procedure> 5-2

IM 11T03C01-01E

(2) OperationsforSettingtheAutomatic/RemoteSemi-automaticCalibration・ Selection for automatic calibration 『*AUTO.C』

At the service level, select CODE 13. ・ Disable: 0

・ Enable: 1

・ Selection for remote semi-automatic calibration 『*REMOT』 At the service level, select CODE 14.

・ Disable: 0

・ Enable: 1

(3) Valve CheckCheck the performance of valves according to the “valve operation flowchart” at the operation level.

【NO】VALVE

【NO】

XXX.XXXV_ZERO

MAN.CAL ADJUST

END

XXX.XXXZ_OPEN

XXX.XXXV_SPAN

XXX.XXXS_OPEN

(Control of zero-gas valve)【YES】

【YES】

【YES】

【YES】 or【NO】

【NO】

【NO】

【NO】

WAIT

【YES】

*2

Measurement mode

*3

【NO】

【NO】

*7 *7

*7*7

*2

*4

*4

*6

*5

*1

(Control of span-gas valve)

(Stabilization time)

*1: See Figure 4.2 “Modes of Operation level” on page 4-8.*2: Press the 【NO】 key to return to the parameter for sample gas.*3: Usingthe 【NO】 key,show 『V.ZERO』, 『V.SPAN』 and 『END』 in succession.*4: Using the 【YES】 key, open the zero-gas or span-gas value to read the measured value at the point.*5: Press the 【YES】 key to go into the measurement mode state after a given stabilization time elaspes; press the 【NO】 key to go into that state immediately.*6: This step is skipped if no valve was opened.*7: The display shows the same data item to be measured as the one in the measurement mode.

F5.2.ai

Figure 5.1 Mode for Valve Operation

3rd Edition : Jul. 10, 2015-00

Page 78: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<5. Calibration Procedure> 5-3

IM 11T03C01-01E

5.2.2 Setting Calibration DataTo set calibration data, press the [*] key in the measurement mode, enter the password (XXX), and select the parameter 『*CAL.DT』. See Figure 4.4 for the process at the setting level.a. Zero-point calorific value 『*Z_CAL』: a value from 000.000 to 133.000 (MJ/Nm3)b. Zero-point density 『*Z_DNS』: a value from 0.0000 to 6.0000 (kg/m3)c. Span-point calorific value 『*S_CAL』: a value from 000.000 to 133.000 (MJ/Nm3)d. Span-point density 『*S_DNS』: a value from 0.0000 to 6.0000 (kg/m3)e. Output during calibration 『*C_HLD』 ・Disable: 0 ・Enable: 1 ・Enable (preset value): 2f. Setting of preset value 『*PR.SET』: a value from -10.0 to 110.0

This setting can be available only if "2'' is set on『*C_HLD』.

 

【NO】

【ENT】

*RANGE【NO】

XXX.XXX*Z_CAL

*ALARM*CAL.DT

X.XXXX*Z_DNS

X*C_HLD

XXX.X*PR.SET

【ENT】

【ENT】

【ENT】

F5.3.ai

【YES】

*2

*3

*1: See “Figure 4.4 Process at the Setting Level” on page 4-10. *2: Set a value using the【 】, 【 】 and 【ENT】 keys. If you enter a value that exceeds the limits, the display shows 『*OVER』. Press the 【YES】 or 【NO】 key to re-set the value.*3: This step is skipped if the parameter 『*C_HLD』 is set to a value other than “2”.

*1

【ENT】

XXX.XXX*S_CAL

X.XXXX*S_DNS

【ENT】

*2

*2

*2

*2

*2

(Setting of preset value)

(Setting of Zero-point calorific value)

(Setting Zero-point density)

(Setting Span-point calorific value)

(Setting Span-point density)

(Setting Output during calibration)

Figure 5.2 Mode for Setting Calibration Data

3rd Edition : Jul. 10, 2015-00

Page 79: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<5. Calibration Procedure> 5-4

IM 11T03C01-01E

5.3 Calibration5.3.1 Semi-automatic Calibration

In this mode of calibration, you can carry out one-touch calibration. To use this feature, you must observe the following instructions.

・ The analyzer needs solenoid valves for controlling the calibration gases (zero and span gases). Connect the solenoid valves with utmost care to ensure that there is no gas leakage in the system.

・ Install the calibration gas cylinders in a location close to the GD300S detector. If the detector needs to be installed in an explosion-hazardous area, the solenoid valves being used must also be protected against explosion.

・ The contacts for controlling solenoid valves are built in the GD410S converter. It is the user's responsibility to supply power to drive the contacts. You must install a switch and a fuse along the power line. Note that the contacts are rated at 250 V AC/3A or 30 V DC/3 A.

【NO】ADJUST

【NO】

XX.XXXXZERO

MAN.CALSEM.CAL

XX.XXXXSPAN

XX.XXXXWAIT

【MODE】

F5.4.ai

【YES】

*2

*3

Measurement mode

START

【YES】

*1

*2

*1

*2

【NO】

*1: If a calibration data error 『ALM.10』 occurs, the display alternates between the 『WAIT』 and 『ALM.10』 indications. The 『ALM.10』 indication still remains on display even after the mode move to the measurement mode. This error massage does not disappear until normally ending in re-calibration or resetting the analyzer.*2: The values are operated and the calibration data are read at the time setpoint (calibration time) which is defined in the mode for setting automatics/semi-automatic calibration data in CODE 15. The display always shows the corrected density.*3: During automatic calibration, the display shows the same data items as those of the 『ZERO』 and subsequent indications.*4: When pressing the 【MODE】 key during 『ZERO』 or 『SPAN』 status, the 『WAIT』 appears on the display. When pressing the 【MODE】 key once again, the analyzer returns to the measurement mode. (This operation means you have cancelled automatic/semi-automatic calibration.)

Measurement mode

Figure 5.3 Semi-automatic (Automatic) Calibration Mode

3rd Edition : Jul. 10, 2015-00

Page 80: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<5. Calibration Procedure> 5-5

IM 11T03C01-01E

5.3.2 Manual CalibrationIn this mode of calibration, you visually make sure the reading has settled, and then confirm the reading manually (by pressing the [ENT] key). There are two ways to switch between the zero, span and sample gases: by using manually-operated valves or by using solenoid valves. When switching between the gases using solenoid valves, follow the instructions on valve installation in Subsection 5.3.1.

【NO】SEM.CAL

【NO】

XXX.XXXZ_CAL

VALVEMAN.CAL

X.XXXXXZ_DNS

X.XXXXXWAIT

【ENT】

F5.5.ai

【YES】

*1

ALM.10

ZERO

【YES】

Measurement mode

*2

*5

【NO】

XXX.XXXS_CAL

X.XXXXXS_DNS

【YES】

*3

*6SPAN

*4

X.XXXXXCAL.SET

X.XXXXXCAL.SET

【ENT】

【ENT】

【ENT】

【ENT】

【ENT】

【YES】 or 【NO】

*3

*3

*4

*3

【NO】

*1: See “Figure 4.2 Modes of Operational Level” on page 4-8.*2: If a calibration data error occurs, the display shows 『ALM.10』. Pressing the 【YES】 or 【NO】 key returns the display to the 『MAN.CAL』.*3: Set a desired value by using the 【 】, 【 】 and 【ENT】 keys.*4: When the reading settles, press the 【ENT】 key to confirm your calibration. This parameter does not accept input with the 【 】 and 【 】 keys. If you press the 【MODE】 key (for ESCAPE) after this parameter appears, the display shows 『WAIT』. The display always shows the corrected density.*5: Press the 【YES】 key to go into the measurement mode states after a given stabilization time elapses: Press the 【NO】 key to go into that state immediately.*6: In case of pressing the 【NO】 key after skipping zero calibration (by pressing 【NO】 key in the 『ZERO』 status), the analyzer returns to 『MAN.CAL』. In case of pressing the 【NO】 key after finishing zero calibration, the analyzer goes to 『WAIT』 status.

Figure 5.4 Manual Calibration Mode

3rd Edition : Jul. 10, 2015-00

Page 81: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<5. Calibration Procedure> 5-6

IM 11T03C01-01E

5.3.3 Automatic CalibratioThis mode of calibration takes effect when you select “enable” for automatic calibration in CODE 13 『*AUTO.C』 at the service level. Even during automatic calibration, you can interrupt the system with the semi-automatic or manual mode of calibration. One cycle of automatic calibration is skipped, however, if the time of automatic calibration arrives when semi-automatic calibration is in progress.

CAUTIONDuring measurement after the calibration ends, the hydrogen density value might exceed 100 percent on the display, under the state that the hydrogen gas is filled and pressurized in the tank.This cause is in purity shortage of the density of the H2 gas when calibration. The H2 purity in the tank rises along with pressurizing the H2 gas. When the purity of the H2 gas exceeds the purity used for calibration, the density display exceeds 100.0%. At this time, the followings are recommended. It is waited to reach the highest pressure in the tank. Under such a condition, when it reaches the highest pressure, only zero calibration gas (H2) side is made.The operation is follows. Get out from the calibration screen by pushing [NO] key against the question of 『SPAN』 after the

『CAL SET』 screen of a normal zero gas calibration procedure is done in [ENT] key.

3rd Edition : Jul. 10, 2015-00

Page 82: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<6. Inspection and Maintenance> 6-1

IM 11T03C01-01E

6. INSPECTION AND MAINTENANCE To ensure the GD410 analyzer remains highly accurate with excellent operating conditions, routine inspection and maintenance is essential. Refer to the check items in this chapter to determine which items to apply in your inspection and maintenance, and check them periodically.

6.1 Routine Inspection and Maintenance6.1.1 Checking Readings and Calibrating the Analyzer

Measure the standard gases every two to three months to check the output signal (Note: the interval varies depending on the operating conditions), If the output signal has errors, carried out zero- and span- point calibration. For details on calibration, see Chapter 5.

6.1.2 Checking the Flow rate of Sample GasThe sample gas should be supplied to the detector at 0.1 to 1.0 l/min. Periodically make sure the given flow rate is kept constant. When checking the flow rate, also make sure the piping is normal and there is no leakage of calibration gases.

6.1.3 Periodic Replacement of the Detector's O-ringsThe O-rings are made from NBR, a material that is not very susceptible to plastic deformation. Their sealing capabilities therefore are not readily impaired. Any deterioration in these O-rings however can result in not only gas leakage but reduced resistance to mechanical vibration, which in turn may lead to reading errors. For this reason, it is recommended that the O-rings be replaced at a fixed interval (every two to three years). Consult Yokogawa for the replacement of O-rings.

3rd Edition : Jul. 10, 2015-00

Page 83: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<6. Inspection and Maintenance> 6-2

IM 11T03C01-01E

6.1.4 Replacing the Fuse(1) For safety reasons, stop the power supply to the converter by switching the external

breaker.(2) Remove the fuse from the fuse holder. Turn the holder cap counter clockwise by 90 degrees

using a flat-blade screw driver that fits the holder cap. Then, pull out the fuse together with the cap.

(3) Check that the rating of the new fuse is correct, and put it in the fuse cap; insert the cap (and fuse) in the holder; then while pressing down, turn the cap clockwise by 90 degrees using the flat-blades crew driver.

(4) If the new fuse breaks too soon, the circuit may be abnormal. Contact Yokogawa in such a case.

Fuse Holder

Fuse Holder

General purpose type

Explosion proof type F6.1.ai

Figure6.1PositionoftheFuseHolder

6.1.5 CleaningUse a dry, clean, soft cloth when cleaning for inspection or maintenance. You can use a neutral detergent on the transparent window of the non-explosion-protected converter (weathering-resistant PVC sheet), when extremely dirty. But do not use any organic solvent.If the device is so dirty and/or badly scratched that it affects key operation or reduces visibility, replace the case cover (part No. : K9215JA).

3rd Edition : Jul. 10, 2015-00

Page 84: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<6. Inspection and Maintenance> 6-3

IM 11T03C01-01E

6.2 Inspection In Case of Failure6.2.1 Inspecting the Analyzer in an Alarm Status

The GD410 analyzer outputs an ALARM signal through terminals 16 and 17 if a failure is detected by the self-diagnostic function during measurement. In addition, the <ALARM> lamp on the operation panel comes on and the message field shows an alarm number. The analyzer outputs no contact signal, however, if the alarm number is 『ALM.09』. If a failure (alarm) is detected, take the corrective action given in Table 6.1.

Table 6.1 Corrective Actions when an Alarm Status Is Detected

ILLEGAL CALIB

Message ShownALM.01

ALM.02

ALM.04

ALM.07

ALM.08

ALM.09

ALM.10

Failure Type

High/low-limit alarm Against physical density

High/low-limit alarm against compensated density

High/low-limit alarm against calorific value

Error in pressure input range

Abnormal sample-gas temperature

Battery failure

Error in calibration (zero and span points)

Diagnostic Conditions

Physical density Low-limit-of-physical density alarm setpointPhysical density High-limit-of-physical densityCompensated density Low-limit-of-compensated-density alarm setpointCompensated density High-limit-of-compensated-density alarm setpointCalorific value Low-limit-of-calorific-value alarm setpointCalorific value High-limit-of-calorific-value alarm setpointZero point -3% pressureSpan point +5% pressurePressure 0.1 kPa-25 °C measured gas temperature 80 °C

-0.3000 zero-point calibration value 0.30000.5000 span-point calibration value 1.5000

Corrective Actions

Change the high/low limits.

Change the high/low limits.

Change the high/low limits.

Check the sample-gas pressure and the pressure range.Use the sample gas within the tolerance limits.Contact Yokogawa service personnel.

Redo calibration.

Message via Communication

<

>

<

OUT OF DENSITY

OUT OF COMP DENSITY

OUT OF CALORY

OUT OF GAS PRESS

OUT OF GAS TEMP

LOW BATTERY

<

<<

<

<

<<

<T6.1.ai

<

>

>

>

3rd Edition : Jul. 10, 2015-00

Page 85: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<6. Inspection and Maintenance> 6-4

IM 11T03C01-01E

6.2.2 Inspecting the Analyzer in a FAIL StatusIn GD410, the state of terminals 18 and 19 is changed from close to open if a FAIL status is detected by the self-diagnostic function during measurement. In addition, the <FAIL> lamp on the operation panel comes on and the message field shows an error number. If a FAIL status is detected, take the corrective action given in Table 6.2.

Table 6.2 Corrective Actions when a FAIL Status Is Detected

T6.2.ai

FAULT MEMORY

Message Shown

Err. 01

Err. 02

Err. 03

Err. 04

Err. 05

Failure Type

Sensor oscillation shutdown

Error in sensor’s oscillation frequency Error in sensor’s oscillation frequencyA/D converter failure Memory failure

Diagnostic Conditions

Failure to detect oscillation frequency from the detector Frequency F2: out of range from 1000 to 10000 HzFrequency F4: out of range from 4000 to 10000 HzSensors failure to detect temperatureFailure in A/D converter Erroneous data in EEPROM/EPROM/RAM

Corrective Action

Reset the power supply.Contact Yokogawa service personnel.Reset the power supply.Contact Yokogawa service personnel.

Contact Yokogawa service personnel.Contact Yokogawa service personnel.Contact Yokogawa service personnel.

Message via Communication

FAULT SENSOR

OVER FREQUENCY

FAULT TEMP

FAULT A/D

3rd Edition : Jul. 10, 2015-00

Page 86: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<Operating Data Values of GD410 Gas Calorie Meter Append-1

IM 11T03C01-01E

Operating Data Values of GD410 Gas Calorie Meter

Item

*SET_K under ADJUST*SET_K under ADJUST*SET_K under ADJUST*RANGE*RANGE*RANGE*RANGE*RANGE*RANGE*RANGE*RANGE*CAL.DT*CAL.DT*CAL.DT*CAL.DT*CAL.DT*CAL.DT*ALARM*ALARM*ALARM*ALARM*ALARM*ALARM*ALARM

K0_ADJ

K0

K1

*OUT1*OUT2*Z_DNS*S_DNS*Z_CP.D*S_CP.D*Z_CAL*S_CAL*Z_CAL*Z_DNS*S_CAL*S_DNS*C_HLD*PR.SET

*L_DNS*H_DNS*L_CP.D*H_CP.D*L_CAL*H_CAL

Density/calorie conversion factor(zero adjustment)Density/calorie conversion factor(zero point)Density/calorie conversion factor(span point)Selection of unit for measurement of output1Selection of unit for measurement of output2Zero point of physical density (4mA output)Span point of physical density (20mA output)Zero point of corrected density (4mA output)Span point of corrected density (20mA output)Zero point of calorific value (4mA output)Span point of calorific value (20mA output)Calorific value of calibration gas (zero gas)Density of calibration gas (zero gas)Calorific value of calibration gas (span gas)Density of calibration gas (span gas)State of analog out put during calibrationPreset value of analog out put during calibration Selection of unit for alarm output measurementLow-limit alarm against physical densityHigh-limit alarm against physical densityLow-limit alarm against corrected densityHigh-limit alarm against corrected density Low-limit alarm against calorific valueHigh-limit alarm against calorific value

-99999 to 99999

-99999 to 99999

-99999 to 99999

0.0000 to 6.00000.0000 to 6.00000.0000 to 6.00000.0000 to 6.00000.000 to 133.0000.000 to 133.0000.000 to 133.0000.0000 to 6.00000.000 to 133.0000.0000 to 6.0000

0 to 2-10.0 to 110.0

0.0000 to 6.00000.0000 to 6.00000.0000 to 6.00000.0000 to 6.00000.000 to 133.0000.000 to 133.000

0

36.669

18.658

*CALRY*C_DNS

06.0000

06.00000.000

133.00040.0000.178560.0001.7834

1-10.0

*CALRY0.00006.00000.00006.00000.000

133.000

Setting/Code Setting Range Default User's

Setpoint

T.ap1.ai

Description

3rd Edition : Jul. 10, 2015-00

Page 87: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

<Operating Data Values of GD410 Gas Calorie Meter Append-2

IM 11T03C01-01E

User's Setpoint

01

02

040510

11

12

1314

15

2030

31

40

43

*M_HLD*PR.SET*E_HLD*PR.SET*SMOTH*CNTCT*P.COMP*P.FIX*C.D.TMP*C.D.PRS*Z_PRS*S_PRS*AUTO.C*REMOT*CAL.P*CAL.T*STAB.T*Y_M_D

*H_M*CYCL.U*CYCL.T

*PRES.U*Y_M_D

*H_M*MINUS*C_K_Z

*C_K_S

*S_CYC

Analog output during maintenancePreset value during maintenanceSelection of output upon failurePreset value upon failureOutput-smoothing constantSetting of contact output statesSetting of pressure correctionPresent value for pressure correctionReference temperature for corrected densityReference pressure for corrected densitySetting of gas pressure (zero point)Setting of gas pressure (span point)Selection for automatic calibrationSelection for remote semi-automatic calibrationSelection of calibration itemCalibration timeStabilization timeStarting time of calibration

Setting of calibration interval unitCalibration interval

Selection of pressure unitSetting of interval clock (year/month/day)

Setting of internal clock (hour/minute)Showing/hiding of negative measured value(-)Indication of calibration coefficient (zero-point coefficient)Indication of calibration coefficient (span-point coefficient)Setting of computing interval

0 to 2-10.0 to 110.0

0 to 2-10.0 to 110.0

00 to 6000 to 150 to 2

0.1 to 999.99-20.0 to 80.00.1 to 999.990.1 to 999.990.1 to 999.99

0 or 10 or 10 to 2

0 to 590 to 59

00.01.01 to99.12.31

00.00 to 23.590 or 1

1 to 23(HOUR)1 to 255(DAY)

0 or 100.01.01 to99.12.31

00.00 to 23.590 or 1

-0.3000 to 0.3000

0.5000 to 1.50000 or 1

1-10.0

1-10.0

1151

101.330.0

101.3380.00

140.0000011

00.01.01

00.000

23300

0

0

Setting/Code Item Description Setting Range Default

T.ap2.ai

3rd Edition : Jul. 10, 2015-00

Page 88: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

CustomerMaintenanceParts List

Yokogawa Electric Corporation

All Rights Reserved, Copyright © 1997, Yokogawa Electric Corporation.Subject to change without notice.

Model GD300S, GD320SGas Density Meter, Gas Calorie Meter,H2 Purity Meter and Replacement Meter,Explosion Proof Detector

CMPL 11T3B1-01E1st Edition : July. 1997(YK)

3rd Edition : Apr. 2008 (YK)

Unit : mm

Item Part No. Qty Description

6 5 1 3 2 4

7, 8

9

123456789

F9301DYF9301DQG9303AMG9303LKG9339AAY9408ZUF9340ALY9401WLG9601AN

111122111

CoverCoverO-RingO-RingBracket AssyBoltScrewToothed Lock WasherCable Grand Assy

Page 89: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the
Page 90: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

CustomerMaintenanceParts List

All right reserved, Copyright © Copyright 1997, Yokogawa Electric Corporation. Subject to change without notice

GD400G, GD410G, GD420GGas Density Meter, Gas Calorie Meter, H2 Purity Meter and Replacement MeterNon-Explosion-proof Converter

CMPL 11T3B1-02E1st Edition : July 1997 (YK)

4th Edition : June 2013 (YK)

4

6 5

1

2

3

Item Part No. 1 A1109EF A1111EF 2 L9811FV 3 K9334CN 4 K9313DW 5 Y9101XA 6 Y9420LU

Qty Description 1 Fuse 1.0A (For 100 to 240V AC) 1 Fuse 2.0A (For 24V DC) 6 Cable Gland 6 Insert 1 Cover Assy (For GD400G, GD410G and GD420G) 4 O-Ring 4 Screw

Page 91: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the
Page 92: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

CustomerMaintenanceParts List

All right reserved, Copyright © Copyright 1997, Yokogawa Electric Corporation. Subject to change without notice

GD400S, GD410S, GD420SGas Density Meter, Gas Calorie Meter, Hydrogen Purity and Replacement Meter Explosion-proof Converter

CMPL 11T03B01-03E1st Edition : Apr. 1997 (YK)4th Edition : Apr. 2008 (YK)

Item Part No. 1 A1109EF A1111EF 2 L9817MH 3 K9292XL 4 L9800GT 5 K9214GZ 6 L9811LL 7 K9214BF

237

1

6

5

4

Qty Description 1 Fuse 1.0A (For 100 to 240V AC) 1 Fuse 2.0A (For 24V DC) 1 O-Ring 1 Screw 1 Screw 3 Plug 3 Cable Gland G3/4 1 Cover

Page 93: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the
Page 94: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

i

IM 11T03C01-01E

Revision Informationl Title : Model GD410 Gas Calorie Meter User’s Manuall Manual No. : IM 11T03C01-01E

Jul.2015/3rdEdition Page i Page 1-1 “Linearity”: Addition of the description. Page 1-2 “Requirements for Sample Gas”: Pressure value are changed. Page1-4 “Fuse”: Change of the type. Page1-6 Addition of the option code “/H” and style code are added. Page1-8, 1-10 Figures are changed. Page2-7 Addition of the "CAUTION". Page2-13, 2-14 Figures are changed. Page4-5, 4-22, and Appedix-2 “Calibration time” and “Stabilization time” are are changed. CMPL 11T3B1-02E 4th edition, CMPL 11T3B1-03E 4th edition.

July2008/2ndEdition Revised and Corrected over all, IM Format and Style renewal (As a result of changing Data Source

from paper to electric files and of applying Indesign software)

Outline of change (not described all in detail) The terms, “explosionproof“, “general purpose” and “gas density meter” etc., are used as standard.

Chapter “Introduction” :Chapter name is change. Several different types of symbols and their description are added. Description with CAUTION (page ii) is added. After-sales Warranty is newly added. “Display on Operation Panel and Operation keys” shifts to Subsection 3.1.2.

Chapter 1: Description with IMPORTANT (p1-1) is added. Section 1.1: Some specifications are changed, added or deleted. (e.g.: characteristics and measuring ranges in general, material and ambient

condition of GD300S; option "/KUHT" of pressure transmitter; Data items, communication, fuse and weight of GD410; GD350 (material, finish and weight) etc. )

Section 1.2: Subsection 1.2.3: Option /KUHT is added to and note marked as “*1” is deleted from

GD300S. Subsection 1.2.5: Part No. and specification are changed. Section 1.3: Explanations (e.g. change to G3/4, class A, parts No., dimensions etc.) are added or

corrected. Subsection 1.3.3: Panel cutout is added. Subsection 1.3.5: Drawings are changed. Subsection 1.3.6: Flow rate is corrected.

Chapter 2: Description with CAUTION (p2-2 and p2-12) Subsection 2.1.2: Drawings are changed in Figure 2.1 and 2.2. Subsection 2.2.2: Panel cutout is added in Figure 2.3. Section 2.3: Some explanations are changed in Figure 2.7. Section 2.4: GD410S drawing and its explanation are added in Figure 2.8. Subsection 2.4.1: Drawing of CAUTION label is added. Drawing (Wiring of Pressure transmitter to ground etc.) and explanations

(table contents, note etc.) are added or corrected in Figure 2.9. Subsection 2.4.4 : Some description are added and corrected. Drawing is added. Subsection 2.4.6: EJA 310 is wired to ground in Figure. Subsection 2.4.7 : All changed.

Chapter 3: Description with IMPORTANT (p3-5), instead of CAUTION, is added. An additional subsection is inserted to 3.1.2. After its insertion, the original subsection

No. is larger by one and the numbers of Figure are changed. Subsection 3.1.3: Table contents is changed

2nd Edition : Jul. 07, 2008-00

Page 95: Model GD410 Gas Density Meter - Yokogawa Electric · The Model GD410 Gas ... 6.2.1 Inspecting the Analyzer in an Alarm ... gas calorie meter only in the case that the density of the

ii

IM 11T03C01-01E

n If you want have more information about Yokogawa products, you can visit Yokogawa’s home page at the following web site. Home page: http://www.yokogawa.com/n Written by Environmental & Analytical Products PMK Group IA Div., Product Business Center Yokogawa Electric Corporation n Published by Yokogawa Electric Corporation 2-9-32 Nakacho, Musashino-shi, Tokyo 180-8750, JAPANn Printed by KOHOKU PUBLISHING & PRINTING INC.

Subsection 3.1.8 is added.

Chapter 4: Descriptions with CAUTION is added in CODE 12 and 14. All Figures are revised and corrected. Section 4.2: Parts of table contents are revised and corrected. Section 4.3: Rearrangement of subsections; Revised and corrected a lot over the section

Chapter 5: Explanations of Figures are revised and corrected. Subsection 5.3.3: Descriptions with CAUTION is added.

Chapter 6: Subsection 6.1.4 is added. Subsection 6.1.2: Flow rate is corrected to 0.1 to 1 l/min Subsection 6.2.1: Diagnostic conditions in ALM 07 is changed.

Appendix: A lot of values in Tables are corrected.

CMPL : Items and parts No. are changed.

Revision Information: newly added

Feb.1998/1stEdition Newly published

2nd Edition : Jul. 07, 2008-00