1 SCIENTIFIC REPORTS | 6:35845 | DOI: 10.1038/srep35845 www.nature.com/scientificreports Microchip-based ultrafast serodiagnostic assay for tuberculosis Vigneshwaran Mani 1 , Bhairav Paleja 1 , Karima Larbi 1 , Pavanish Kumar 1 , Jo Ann Tay 1 , Jie Yee Siew 1 , Fatih Inci 2 , ShuQi Wang 3,4,5 , Cynthia Chee 6 , Yee Tang Wang 6 , Utkan Demirci 2 , Gennaro De Libero 1,7 & Amit Singhal 1 Access to point-of-care (POC), rapid, inexpensive, sensitive, and instrument-free tests for the diagnosis of tuberculosis (TB) remains a major challenge. Here, we report a simple and low-cost microchip-based TB ELISA (MTBE) platform for the detection of anti-mycobacterial IgG in plasma samples in less than 15 minutes. The MTBE employs a flow-less, magnet-actuated, bead-based ELISA for simultaneous detection of IgG responses against multiple mycobacterial antigens. Anti-trehalose 6,6-dimycolate (TDM) IgG responses were the strongest predictor for differentiating active tuberculosis (ATB) from healthy controls (HC) and latent tuberculosis infections (LTBI). The TDM-based MTBE demonstrated superior sensitivity compared to sputum microscopy (72% vs. 56%) with 80% and 63% positivity among smear-positive and smear-negative confirmedATB samples, respectively. Receiver operating characteristic analysis indicated good accuracy for differentiating ATB from HC (AUC = 0.77). Thus, TDM-based MTBE can be potentially used as a screening device for rapid diagnosis of active TB at the POC. e management and control of tuberculosis (TB) still remains a significant threat to public health 1 , partly due to the absence of cost-effective, sensitive, and rapid diagnostic tests 2,3 . Currently, sputum smear microscopy is the most commonly used point-of-care (POC) method for TB diagnosis in endemic countries, despite its poor sen- sitivity (30–60%) 4 . Although “gold standard” bacterial culture does provide the required sensitivity, the test takes several weeks and requires well-equipped laboratories and trained staff 5 . Such a long turn-around time oſten results in delayed diagnosis, continued transmission, and the risk of developing drug resistance 6 . Serological tests based on the detection of antibodies against mycobacterial protein antigens in the form of lateral flow devices or standard ELISAs have been extensively used for the diagnosis of TB 7 . However, these tests have demonstrated poor sensitivity (1–60%) and specificity (53–99%) compared with standard culture methods 8 , performing no better than sputum smear microscopy, and have failed to improve patient outcomes. As such, the World Health Organization (WHO) has recommended against their usage 7 . Endorsement by the WHO of nucleic acid amplification-based TB diagnostic tests, such as the automated GeneXpert MTB/RIF system (Cepheid), INNO-LiPA Rif TB kit (Innogenetics), and Genotype MTBDRplus assay (Hain Lifescience) has helped to fill this gap. However, their implementation in POC has been severely restricted by high maintenance costs and the need for sophisticated instrumentation, trained personnel, and uninterrupted electrical supply 9 . us, there is an urgent need for the development of a simple, sensitive, and portable assay for the early stage detection of TB at the POC. An ideal test must meet minimum specifications outlined by the WHO, such as short assay time (<3 h), minimal sample preparation steps, maintenance-free instrumentation, low-cost (<$10 per test), and envi- ronmentally acceptable waste disposability 10,11 . 1 Singapore Immunology Network, Agency for Science, Technology and Research (A*STAR), Singapore. 2 Demirci Bio- Acoustic-MEMS in Medicine (BAMM) Laboratory, Stanford University School of Medicine, Department of Radiology, Canary Center at Stanford for Cancer Early Detection, Palo Alto, California, USA. 3 State Key Laboratory for Diagnosis and Treatment of Infectious Diseases, First Affliliated Hospital, College of Medicine, Zhejiang University, Hangzhou, China. 4 Collaborative Innovation Center for Diagnosis and Treatment of Infectious Diseases, Hangzhou, China. 5 Institute for Translational Medicine, Zhejiang University, China. 6 Tuberculosis Control Unit (TBCU), Tan Tock Seng Hospital, Singapore. 7 Department of Biomedicine, University of Basel, Switzerland. Correspondence and requests for materials should be addressed to G.D.L. (email: [email protected]) or A.S. (email: amit_singhal@ immunol.a-star.edu.sg) Received: 31 May 2016 Accepted: 26 September 2016 Published: 24 October 2016 OPEN
11

Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

Feb 28, 2021

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

1Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

www.nature.com/scientificreports

Microchip-based ultrafast serodiagnostic assay for tuberculosisVigneshwaran Mani1, Bhairav Paleja1, Karima Larbi1, Pavanish Kumar1, Jo Ann Tay1, Jie Yee Siew1, Fatih Inci2, ShuQi Wang3,4,5, Cynthia Chee6, Yee Tang Wang6, Utkan Demirci2, Gennaro De Libero1,7 & Amit Singhal1

Access to point-of-care (POC), rapid, inexpensive, sensitive, and instrument-free tests for the diagnosis of tuberculosis (TB) remains a major challenge. Here, we report a simple and low-cost microchip-based TB ELISA (MTBE) platform for the detection of anti-mycobacterial IgG in plasma samples in less than 15 minutes. The MTBE employs a flow-less, magnet-actuated, bead-based ELISA for simultaneous detection of IgG responses against multiple mycobacterial antigens. Anti-trehalose 6,6′-dimycolate (TDM) IgG responses were the strongest predictor for differentiating active tuberculosis (ATB) from healthy controls (HC) and latent tuberculosis infections (LTBI). The TDM-based MTBE demonstrated superior sensitivity compared to sputum microscopy (72% vs. 56%) with 80% and 63% positivity among smear-positive and smear-negative confirmed ATB samples, respectively. Receiver operating characteristic analysis indicated good accuracy for differentiating ATB from HC (AUC = 0.77). Thus, TDM-based MTBE can be potentially used as a screening device for rapid diagnosis of active TB at the POC.

The management and control of tuberculosis (TB) still remains a significant threat to public health1, partly due to the absence of cost-effective, sensitive, and rapid diagnostic tests2,3. Currently, sputum smear microscopy is the most commonly used point-of-care (POC) method for TB diagnosis in endemic countries, despite its poor sen-sitivity (30–60%)4. Although “gold standard” bacterial culture does provide the required sensitivity, the test takes several weeks and requires well-equipped laboratories and trained staff 5. Such a long turn-around time often results in delayed diagnosis, continued transmission, and the risk of developing drug resistance6.

Serological tests based on the detection of antibodies against mycobacterial protein antigens in the form of lateral flow devices or standard ELISAs have been extensively used for the diagnosis of TB7. However, these tests have demonstrated poor sensitivity (1–60%) and specificity (53–99%) compared with standard culture methods8, performing no better than sputum smear microscopy, and have failed to improve patient outcomes. As such, the World Health Organization (WHO) has recommended against their usage7. Endorsement by the WHO of nucleic acid amplification-based TB diagnostic tests, such as the automated GeneXpert MTB/RIF system (Cepheid), INNO-LiPA Rif TB kit (Innogenetics), and Genotype MTBDRplus assay (Hain Lifescience) has helped to fill this gap. However, their implementation in POC has been severely restricted by high maintenance costs and the need for sophisticated instrumentation, trained personnel, and uninterrupted electrical supply9. Thus, there is an urgent need for the development of a simple, sensitive, and portable assay for the early stage detection of TB at the POC. An ideal test must meet minimum specifications outlined by the WHO, such as short assay time (< 3 h), minimal sample preparation steps, maintenance-free instrumentation, low-cost (< $10 per test), and envi-ronmentally acceptable waste disposability10,11.

1Singapore Immunology Network, Agency for Science, Technology and Research (A*STAR), Singapore. 2Demirci Bio-Acoustic-MEMS in Medicine (BAMM) Laboratory, Stanford University School of Medicine, Department of Radiology, Canary Center at Stanford for Cancer Early Detection, Palo Alto, California, USA. 3State Key Laboratory for Diagnosis and Treatment of Infectious Diseases, First Affliliated Hospital, College of Medicine, Zhejiang University, Hangzhou, China. 4Collaborative Innovation Center for Diagnosis and Treatment of Infectious Diseases, Hangzhou, China. 5Institute for Translational Medicine, Zhejiang University, China. 6Tuberculosis Control Unit (TBCU), Tan Tock Seng Hospital, Singapore. 7Department of Biomedicine, University of Basel, Switzerland. Correspondence and requests for materials should be addressed to G.D.L. (email: [email protected]) or A.S. (email: [email protected])

Received: 31 May 2016

Accepted: 26 September 2016

Published: 24 October 2016

OPEN

Page 2: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

www.nature.com/scientificreports/

2Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

Advances in microscale and nanoscale technologies offer feasible approaches for the development of min-iaturised POC devices12,13. Microscale technologies allow integration and automation of multistep assays, such as ELISA14, thus enabling sample processing, target capture, and detection into a single integrated device, which speeds up the whole assay. In particular, magnetic beads (MB) have been exploited extensively in microfluidic-based ELISAs due to their uniform size, high surface-to-volume ratio, faster reaction kinetics, and ease of manipulation, providing better sensitivity with shorter assay time compared to conventional flat sur-faces15,16. Furthermore, with the use of an external magnet, MBs can be actuated/manipulated17–20 through a series of stationary reagents for bio-detection in automated assays21–24. This provides a simple ‘sample-in and answer-out’ based system, which is highly desirable for diagnosis at the POC.

We present herein the development of a microchip TB ELISA (MTBE), capable of detecting IgG responses against multiple antigens from plasma samples of active TB (ATB) patients in a rapid and miniaturised detection system. The MTBE utilises a Mycobacterium tuberculosis (Mtb) surface glycolipid i.e. trehalose 6,6′ -dimycolate (TDM) and two purified proteins, 38 kDa glycolipoprotein and antigen 85A (Ag85A), as antigens based on their known immunogenicity and their application in TB serodiagnosis25–32. The MTBE relies on the actuation of antigen-coated MBs through sequentially organised reagents for capturing Mtb antigen-specific IgG from the plasma, followed by labelling and colorimetric detection. We showed that MTBEs featuring detection of anti-TDM IgG response could reliably differentiate ATB patients from healthy controls (HC). Furthermore, the test requires less than 15 min from sample addition to detection, which is within the minimum specifications required for POC TB testing10.

ResultsCharacterisation of TDM-coated magnetic beads. We first standardized the coating of TDM on mag-netic beads. Since high lipid concentrations lead to the formation of large bead aggregates33. We assessed different TDM concentrations per surface area on MBs (0.16 μ g/cm2, 0.41 μ g/cm2, and 0.65 μ g/cm2) for optimal TDM coating. The resultant TDM-coated MBs were then tested for size distribution, amount of bound TDM, and the ability to detect anti-TDM IgG using a pooled plasma sample of ATB individuals.

Dynamic light scattering (DLS) showed uniform size distribution of MBs irrespective of TDM surface concen-trations (Supplementary Fig. S1). A polydispersity index (PI) of < 0.21 was obtained for all the MB preparations, indicating the presence of monodisperse MBs (based on the criteria, PI < 0.3 for monodisperse beads34). The mean diameter of the TDM-coated MBs obtained by DLS ranged from 4.2–4.6 μ m, which was well within the range of uncoated nascent MBs (dia. 4.5 μ m). Furthermore, large multi-bead aggregates were not visible by light microscopy, indicating the monodisperse nature of the TDM-coated MBs (Supplementary Fig. S2).

To assess the amount of TDM adsorbed onto the MB surface, TDM bound to each of the MB preparations was extracted for thin layer chromatography (TLC) analysis. TLC analysis indicated that the recovery of bound TDM (%) from MBs prepared using surface TDM concentrations of 0.16 μ g/cm2 and 0.41 μ g/cm2 were similar, but recovery from MBs using concentrations of 0.65 μ g/cm2 was lower (Supplementary Fig. S3). As higher lipid losses due to nonspecific adsorption to the reaction vials were significant in small batch preparations (0.2 mL), we repeated the TDM recovery process using MBs prepared in a large batch (1.6 mL). In the large batch preparation, MBs with a surface TDM concentration of 0.41 μ g/cm2 showed 54% recovery of bound TDM. At this concen-tration and based on the molecular weight of TDM (~2636 g/mol) and the total number of beads used in the preparation, the number of TDM molecules per bead was estimated at ~3.2 × 107. This is in concordance with the theoretical estimate of 3.4 × 107 of TDM molecules per bead based on the surface area of the bead (6.4 × 1011 m2, diameter 4.5 μ m) and the molecular exclusion area of TDM (~180 °A2/molecule), indicating a uniform monolayer of TDM on the bead surface.

Next, we tested the ability of TDM-coated MBs to detect an anti-TDM IgG response in pooled plasma of ATB individuals using a flow cytometry-based MB immunoassay. Bead preparations with a surface TDM concen-tration of 0.41 μ g/cm2 displayed the highest levels of anti-TDM IgG antibodies in comparison to MBs prepared using 0.16 and 0.65 μ g/cm2 of TDM (Supplementary Fig. S4). Since temperature variation and reagent storage may influence the performance of an ELISA, we then tested the stability of TDM-coated MBs (0.41 μ g/cm2). The beads maintained 95% activity after 5 months of storage at room temperature and 74% activity after 10 months with a relative variation of 65–125% over time (Supplementary Fig. S5). Based on the aforementioned analysis, we selected the MB preparation with a surface TDM concentration of 0.41 μ g/cm2 for further use.

Comparison of magnetic bead (MB) ELISA with conventional plate ELISA format for anti-TDM response detection. TDM-coated MBs (0.41 μ g/cm2) were used to perform a MB ELISA for the detection of anti-TDM IgG in the plasma, and the resultant IgG levels were compared with a conventional plate ELISA for-mat (where TDM is coated on the plate surface). Both formats showed significantly elevated levels of anti-TDM IgG in the plasma of ATB patients compared to HC individuals (Fig. 1A,B). We further tested the specificity of the anti-TDM IgG antibodies using a competitive plate assay, where 10% free trehalose was used to block the antigen-antibody reaction. Out of the 22 ATB plasma samples, 12 showed 50% or more inhibition in IgG binding to TDM using free trehalose, suggesting plasma anti-TDM antibody affinity towards the trehalose moiety of TDM (Supplementary Fig. S6). In the remaining 10 ATB samples, no inhibition by free trehalose was observed, indicating binding of anti-TDM antibodies to epitopes nearby the glycosidic bond between trehalose and mycolic acid35,36. To benchmark the MB ELISA with a standard plate ELISA, we assessed anti-TDM IgG in the plasma of the same individuals using the two methods. Both methods showed sensitivity > 68% (Supplementary Table S1) and a good correlation (R2 = 0.96, Fig. 1C). The overall assay time for the MB ELISA was ~50 min, substantially faster than the conventional plate ELISA (5 h).

Page 3: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

www.nature.com/scientificreports/

3Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

Design of microchip TB ELISA (MTBE) assay. We then translated the MB ELISA assay onto a micro-chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different reactions/assays can be performed. Each channel consists of connected chambers, which are filled with spa-tially arranged, stationary, aqueous reagents separated by immiscible oil. The MTBE is carried out by actuating the antigen-coated MBs in each of these chambers using the magnet underneath. Each chamber performs dif-ferent functions; the first is used for IgG capture, the second for binding of biotin-labelled secondary anti-IgG antibody, and the third for binding of streptavidin polymeric enzyme, with alternate chambers for washing (Fig. 2A). Finally, the MB-bound polymeric enzyme induces TMB oxidation in the latter chamber, generating a blue-coloured substrate (Fig. 2B). The reaction was stopped and the optical density (OD) was measured at 450 nm. In the first three channels of the MTBE, we detected IgG against TDM, 38 kDa, and Ag85A antigens, respectively. The fourth and fifth channels were used to measure the total IgG response against TDM combined with each protein antigen. The sixth channel was used as negative control with no antigen added. The entire MTBE process requires ~15 min from plasma sample addition to colorimetric detection (Fig. 2B).

Simultaneous detection of IgG antibodies against multiple Mtb antigens using MTBE. We then used MTBE to detect IgG against three Mtb antigens (TDM, 38 kDa, and Ag85A) and their combinations in 146 plasma samples (65 ATB, 40 LTBI, and 41 HC). ATB samples were considered positive if the values were higher than the cut-off values obtained from the HC and LTBI samples31. The cut-off value for each antigen was set at 75% specificity in order to compare their relative sensitivity to a set specificity. Using the specificity set criteria,

Figure 1. Performance of TDM based magnetic bead (MB) ELISA with conventional plate bound assay. Comparison of the distribution of anti-TDM IgG responses in the plasma of active TB (ATB) and healthy control (HC) individuals determined using (A) MB ELISA where plasma was diluted 125-fold in 5% BSA buffer. (B) Conventional plate ELISA where plasma was diluted 2500-fold in 5% BSA. N = 40, ATB = 19 and HC = 21 (***P = 0.0003 for MB ELISA; ***P = 0.0007 for plate ELISA). (C) Correlation of anti-TDM plasma IgG responses using the MB ELISA and a conventional plate ELISA.

Page 4: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

www.nature.com/scientificreports/

4Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

88% (57/65) of the ATB plasma samples were positive for IgG against at least one of the three antigens tested. Only 40% (26/65) of the ATB plasma samples were positive for IgG against any two of the antigens tested individually, and only 28% (18/65) of the ATB plasma samples were IgG positive for all three antigens. These findings suggest a heterogeneous IgG response against the three antigens. Compared to HC plasma samples, ATB samples had significantly higher levels of IgG against all three antigens (Fig. 3A–C), with the highest sensitivity of 72% for TDM (Table 1), indicating greater reliability of the anti-glycolipid humoral response compared to the anti-protein response in differentiating ATB from HC. In addition, receiver operating characteristic (ROC) curve analysis demonstrated a better performance of TDM-based MTBE compared to 38 kDa and Ag85A in discriminating

Figure 2. Illustration of MTBE for simultaneous detection of IgG response against multiple Mtb antigens for TB diagnosis. (A) Schematic representation of a microchip employing magnet-actuated MB ELISA for the simultaneous detection of glycolipid, protein, and a mixture of glycolipid + protein-specific IgG antibodies in the plasma of ATB, latent TB infection (LTBI) and HC individuals. MTBE is performed by simultaneous actuation of antigen-coated MBs in each chamber through sequentially arranged reagents using six magnets below for incubation/washing. The overall time of the test from sample addition to detection is 15 min. (B) Photographs of MTBE before (left) and after 15 min (right) from addition of plasma sample.

Page 5: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

www.nature.com/scientificreports/

5Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

ATB from HC individuals (area under the curve (AUC) of 0.77 vs. 0.69 and 0.74, respectively) (Fig. 3D, Table 1). These results indicate that the IgG humoral immune response to TDM is a promising immunological marker for ATB detection. Comparison of the TDM MTBE with a classical plate ELISA and MB ELISA showed good corre-lation (R2 = 0.90–0.94; Supplementary Fig. S7A,B), suggesting that the TDM MTBE is as efficient as a benchtop ELISA with the important advantage of being performed in ~15 min without the requirement for sophisticated instrumentation.

A similar trend was observed for differentiating ATB from LTBI individuals. Anti-TDM IgG levels provided the highest discrimination sensitivity (71%) when compared to the responses against the 38 kDa and Ag85A

Figure 3. Distribution of IgG responses determined by MTBE against independent antigens in the plasma of ATB and HC individuals. Antigens included (A) TDM, (B) 38 kDa, (C) Ag85A. The plasma in the MTBE was diluted 200-fold in 5% BSA. N = 106, ATB = 65, and HC = 41 (**P = 0.0012, ****P < 0.0001) (D) ROC curves for plasma IgG assays for individual antigens for differentiating ATB from HC individuals.

Active TB vs. Healthy controls

Antigen Sensitivity (%) Specificity (%)Positive predictive

value (PPV) %Negative predictive

value (NPV) %ROC, AUC

TDM 72 76 82 63 0.77

38 kDa 48 76 76 48 0.69

Ag85A 52 76 77 50 0.74

TDM + 38 kDa 65 76 81 57 0.77

TDM + Ag85A 71 76 82 62 0.80

Table 1. Evaluation of serodiagnostic potential of each antigen and their combinations.

Page 6: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

www.nature.com/scientificreports/

6Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

protein antigens (Supplementary Fig. S8A–C and Table S2). ROC curves further confirmed the superior accu-racy of the TDM-based MTBE (AUC, 0.75) compared with the protein antigen-based assay in differentiating ATB from LTBI individuals (Supplementary Table S2). Indeed, the IgG response against TDM was the strongest predictor for discriminating ATB from HC and LTBI according to a generalised linear model (using logistic regression analysis37).

MTBE IgG responses against a combination of antigens. Next, we investigated whether the collective measurement of the IgG response against two antigen types might further increase MTBE sensitivity. Therefore, we mixed TDM-coated MBs with either one of the protein antigen-coated MBs and performed the MTBE assay as before. The IgG response against the combination of MBs was able to differentiate ATB from HC individuals (Supplementary Fig. S9). However, the sensitivity obtained with combined MBs (65–71%) was no higher than that obtained by measuring response to TDM-coated MBs alone. This finding was confirmed by ROC curve anal-ysis (Table 1). Similarly, the MTBE based on the combined MBs did not show any improvement over TDM MBs alone in discriminating ATB from LTBI individuals (Supplementary Fig. S10 and Table S2).

We next tested ATB, LTB, and HC plasma samples for the preference of relative IgG responses against each of the three antigens and their combinations. We observed diverse and highly variable antibody patterns, with anti-TDM IgG responses (Normalized OD intensities, z-scores) more clearly differentiating ATB patients from LTBI and HC individuals (Fig. 4).

Effect of the sputum bacillary load on the IgG antibody response. We next compared the anti-TDM IgG levels according to the sputum bacterial burden. The TDM-based MTBE was positive in 80% of sputum smear-positive ATB samples and in 63% of smear-negative ATB samples. As before, the sensitivity of the TDM-based assay was higher than that of the protein-specific assays (Table 2). When ATB samples were stratified according to their smear grade, all smear grade 2+ and 4+ samples were positive for TDM-based IgG responses, with smear grade 2+ samples showing significantly higher anti-TDM IgG responses compared with smear-negative samples (Fig. 5A). Only a few samples from patients with smear grade 3+ were positive for TDM-based IgG responses. In addition, culture-positive ATB samples showed higher TDM-specific IgG responses than culture-negative samples (Fig. 5B). No association of bacillary load with an IgG response to the protein antigens was detected (Supplementary Fig. S11). TDM-specific responses were detected in 77% of culture-positive ATB samples and in 61% of culture-negative ATB samples, and provided a superior correlation with bacterial load compared with antibody responses against the other two tested protein antigens (Table 2). The

Figure 4. Patterns of reactivity of plasma to antigens in ATB, HC, and LTBI individuals. The heat map shows the reactivity of plasma to individual antigens, and their combinations. Each column represents the response observed in one plasma sample and each row depicts the response to different antigens or their combinations. Normalised OD values (row z-scores) are visualised as a colour spectrum as shown. The heat map was generated using R statistical computing software, using z-score = (x− μ )/σ , where x is an individual’s OD response, μ is mean of OD response from all individuals (N = 146) for each antigen and σ is the standard deviation. N = 146; ATB = 65; LTBI = 40; HC = 41.

Antigen

Smear-positive culture confirmed

cases (%)

Smear-negative culture confirmed

cases (%)

Culture-positive confirmed cases

(%)

Culture-negative confirmed cases

(%)

TDM 80 63 77 61

38 kDa 46 52 45 69

Ag85A 49 59 51 69

TDM + 38 kDa 66 67 68 61

TDM + Ag85A 77 67 72 77

Table 2. Percentage of IgG-positive samples classified according to sputum smear test and culture test.

Page 7: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

www.nature.com/scientificreports/

7Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

parameters of the MTBE described here comply with the minimum specifications required for TB POC testing as defined by the WHO, indicating a possible application of the TDM-based MTBE for the early detection of ATB patients (Table 3)10,38.

We next evaluated the possibility of combining MTBE and sputum microscopy using a two-stage screening process i.e. sputum microscopy samples with negative results followed up using the TDM-based MTBE. This combination enhanced the sensitivity of ATB detection from 56% (sputum microscopy alone) to 85% (sputum microscopy in combination with the MTBE), thus demonstrating that the MTBE could be used alongside smear microscopy to improve the accuracy of TB screening.

DiscussionDespite the identification of Mycobacterium tuberculosis as the cause of TB more than a century ago, the diagno-sis of TB in resource-limited settings continues to be a major challenge. TB diagnosis in endemic countries still depends primarily on sputum microscopy and culture, which have a number of limitations in terms of sensitivity, specificity, and turn-around time. Here, we present a simple and ultrafast miniaturised ELISA capable of detecting IgG responses against multiple antigens that can be used for the diagnosis of TB. The MTBE provides a simple sample addition (sample-in) to colorimetric detection (answer-out)-based system to differentiate active TB from healthy controls/latent TB infections with results available in less than fifteen minutes.

The sensitivity of TDM-based MTBE (72%) or 38 kDa-based MTBE (46%) is in close agreement to classical plate bound TDM or 38 kDa ELISAs (69% and 47%, respectively)7. Moreover, the MTBE test performs well in comparison to other commercial tests that use 38 kDa and TDM as antigens, such as Pathozyme Myco-G (Omega diagnostics, UK) (10–85% sensitivity) and the TB glycolipid assay (Kyowa Medex, Japan) (59–90% sensitivity), respectively39. The higher sensitivity of TDM-based MTBE compared to sputum microscopy (72% vs. 56%) will allow the rapid screening and diagnosis of individuals with clinically active TB. This enhanced TDM response among active TB individuals is unlikely due to the prior BCG vaccination, since BCG vaccination is mandatory in Singapore and so most likely all subjects enrolled in our study had received it. Noteworthy, it has been suggested that the serological diagnosis using anti-TBGL (Tuberculosis glycolipid antigens) antibody is not influenced by prior BCG vaccination40. In addition, the MTBE offers the advantages of being faster, simpler, and cheaper, thus suitable for initial screening of potential TB patients who can later be verified with PCR-based assays.

The MTBE platform has three important technological implementations compared with a conventional ELISA. First, the working principle of MTBE i.e. MB actuation through chambers prefilled with stationary rea-gents circumvents the need for additional equipment, such as expensive pumps, valves, and sample metering that usually require challenging microscale fabrication techniques. MTBE requires simple non-lithographic fab-rication methods, and no sophisticated tools are needed to perform the test. Thus, the cost of the test, including reagents and device fabrication, remains extremely low (< U.S. $10). Second, contrary to the conventional ELISA, the MTBE approach utilises polymeric horseradish peroxidase (HRP) labels as opposed to single HRP-labelled secondary antibodies. The higher ratio of HRP to polymer in the labelling step (400 HRP/polymer) ampli-fies the colorimetric signal and enables detection of low IgG titre samples. Third, the bead-based ELISA assay employed in the microchip has specific advantages over conventional flat surfaces. The high local concentra-tion of antigens bound on the bead surface promotes efficient antigen-antibody binding compared to flat plate surfaces41, where binding is facilitated by slow diffusion, and requires several hours to reach saturation. This translates to faster reaction times for each step in the MTBE, significantly reducing the overall time of the assay to ~15 minutes using plasma sample matrices, but still performing as efficiently as the standard classical ELISA. Moreover, the plasma-based testing presented herein is easy to perform because of its simple sample collection (whole blood, venous/finger prick) compared to sputum sample (deep cough, several attempts required). This is

Figure 5. Distribution of anti-TDM IgG response among classified ATB samples. (A) Samples were classified based on AFB sputum smear grade. N = 62; -ve = 28; 1+ = 11; 2+ = 11; 3+ = 8 and 4+ = 4 (**P = 0.002 *P = 0.013). (B) Samples were classified according to culture test results. Culture positive, N = 47; culture negative N = 13 (*P = 0.0312).

Page 8: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

www.nature.com/scientificreports/

8Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

particularly relevant for sample collection from children for whom sputum collection is challenging and inva-sive. Furthermore, the current MTBE assay has many advantages over commercial immunochromatographic (IC) tests, which are widely used for TB diagnosis in low-income high burden countries. Although IC tests are user-friendly, rapid, and affordable, they lack the sensitivity of a classical ELISA (53%, 95% CI 42–64%39) and the results are qualitative, relying on a subjective interpretation of aggregated gold nanoparticle band inten-sity. Furthermore, IC tests are only able to provide binary reports (yes/no) for a single antigen, and analysing responses against multiple antigens is complex. In contrast, the MTBE device represents a simple and robust plat-form, which is as sensitive as a bench-top ELISA and provides an accurate numerical interpretation of responses against multiple biomarkers.

The MTBE platform is a flexible technology that can be adapted to diagnosis of other diseases. Simultaneous use of multiple rows could even be used to detect multiple biomarkers for several diseases at the same time. The detection mode is not restricted to a colorimetric readout, and can be translated to more sensitive electro-chemical42, electrochemiluminescent43, and plasmonic readouts44. The current prototype has the applicability to be further developed into a fully automated device for operation in resource-limited settings. We envisage that the automation of magnet actuation and the integration of a portable colorimetric sensor into a single device presented here will provide a single-step miniaturised assay for TB detection. Such a device will be truly useful in POC TB diagnosis for screening patients and TB contacts, where immediate treatment decisions are of high clinical significance45,46.

In conclusion, the MTBE platform is the first step towards the development of a POC test for TB and can be easily implemented in combination with sputum microscopy to speed up TB diagnosis in triage and community settings.

MethodsChemicals and Materials. Tosyl-activated superparamagnetic microbeads M-450 (MBs, Dynabeads, Cat. no. 14013, 4.5 μ m diameter, 4 × 108 beads/mL, 30 mg/mL) were from Invitrogen. TDM from Mycobacterium bovis (Cat. no. T3034), and fatty acid-free bovine serum albumin (BSA, Cat. no. A7030) were from Sigma. Goat F(ab’)2 anti-human IgG (H + L) labelled with biotin was from Southern Biotech. Nunc PolySorp 96-well plates,

FeaturesWHO minimum

specifications Microchip TB ELISA (Current assay)Sputum microscopy

(Conventional technique) Cepheid’s GeneXpert

Type of assay NA Microchip-based Mtb-specific glycolipid immune response detection

Direct acid-fast bacilli (mycobacteria) detection

Polymerase chain reaction based detection

Biomarkers NAAntibody response to trehalose

dimycolate (Mtb lipid), to 38 kDa, and to antigen 85A (two Mtb proteins)

Whole MtbAutomated system that

performs detection of Mtb rpoB gene and associated

mutations

Sample matrix NA Plasma/Serum/Blood Sputum Sputum

Time to results 3 h max; desirable 15 min 15 min 2 h 2 hr

Sample preparation3 steps maximum. -Can be

performed in BSL-1 Ability to use small volumes

2-step assay, collect the sample & add to the chip followed by detection -Can be

performed in BSL-1

1-step assay, sputum samples are stained and examined

under a microscope -Can be performed in BSL-1

-2 step assay -Decontamination of sputum,

followed by addition of sample to cartridge

Laboratory infrastructure NA No No Yes

Throughput 20 tests per day by a single operator 24–32 tests/day by a single operator > 20 tests per day by a single

operatorUp to 16 tests per day by a single operator using

GeneXpert IV

Waste disposableEnvironmentally acceptable disposable as simple burying

or burningChip burying may be possible Slides can be buried NR

Instrumentation

-Maintenance free -Robust in tropical conditions -Acceptable

replacement cost -Must fit in backpack, be shock-resistant

and work from battery

-Single disposable chip per test - potential device does not require

complex instrumentation, thus low replacement cost -potential prototype

device can be easily fitted inside a backpack

-Single disposable slides -Yes, requires only a

microscope -Not possible to be carried on a backpack

-Maintenance required -Associated replacement cost

-Bulky instrumentation

Possibility of POC application at resource limited settings

Yes YesNo (GeneXpert can be

adopted only at community level as it needs basic

laboratory infrastructure)

Cost Below $10 per test < U.S. $10 -Low cost of chip -Use of low volume of reagents -Easy to perform

$1 per test -low cost of slides -Microscopes are widely

available

$17620 instrument $10–15 per cartridge Needs skilled

personnel

Sensitivity (Adult) Smear positive 95% Smear negative 60–80%

Smear positive 80% Smear negative 63% (ATB vs. HC) 34–60% Smear positive 98% Smear

negative 68% (ATB vs H)

Sensitivity (children) 80% NP 10–20% 67%

Specificity 95% 75% (ATB vs. HC)Needs to be confirmed with

other methods, cannot differentiate Mtb from other

mycobacterial strains.98%

Table 3. Comparison of minimum specifications recommended for POC TB testing10. NA: not applicable, NP: not performed, NR: not reported.

Page 9: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

www.nature.com/scientificreports/

9Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

1-step ultra TMB ELISA, and Pierce streptavidin poly-HRP were obtained from Thermo Scientific. Mycobacterial recombinant 38 kDa and Ag85A proteins were obtained from Mybiosource (CA, USA). Neodymium disc mag-nets (diameter 5 mm, thickness 2 mm) were from AliExpress Global Retail. Unless otherwise specified, all exper-iments were performed using PBS buffer without Ca+2 and Mg+2 ions.

Human subjects and sample collection. Blood samples of ATB, HC, and LTBI individuals were col-lected at the Tuberculosis Control Unit, Tan Tock Seng Hospital, Singapore after taking informed consent. Blood plasma was separated in a BSL3 facility, and immediately stored at − 80 °C. Of the samples collected, plasma from 65 ATB, 41 HC, and 40 LTBI individuals were randomly selected and stratified based on clinical data, such as the interferon-gamma release assay, sputum smear microscopy, and culture tests. The study was performed in accordance with relevant regulations and was approved by National Health Group domain specific review board (NHG DSRB No. 2010/00566). Sputum samples were stratified based on the AFB smear grade, which was performed according to the American Thoracic Society (ATS) with negative (-ve) representing 0 AFB/100 fields; 1+ representing 1–9 AFB/100 fields; 2+ representing 1–9 AFB/10 fields; 3+ representing 1–10 AFB/field and 4+ representing > 10 AFB/field.

Preparation of lipid and protein-coated MBs. For TDM coating, 0.8 mL of MB stock (4 × 108 beads/mL) were taken in a glass tube with a screw cap, and magnetically washed sequentially with 70% and 100% eth-anol. The MBs were then air dried, and 84 μ g TDM in 2.4 mL of solvent (9:1 hexane: ethanol) was added to the dried MBs. The MB dispersion was then sonicated for approximately 1 h in a water bath until the solvent evapo-rated to dryness. Control beads (beads without TDM) were similarly sonicated with TDM-free solvent, until the solvent evaporated. This step was followed by chemical and physical blocking, whereby dried TDM-coated MBs were sonicated for 2 min with 1.6 mL of 0.1% BSA/PBS buffer, and subsequently sonicated for another 2 min after addition of 6.4 mL of 0.2 M Tris buffer, pH 8.0. The bead dispersion was then mixed in a slow tilt rotor for 24 h at room temperature and washed with 10 mL of 0.1% BSA four times before reconstituting in 1.6 mL of 0.1% BSA/PBS buffer, and stored at 4 °C until further use.

For protein coating, 38 kDa and Ag85A proteins were covalently linked to MBs according to the manufac-turer’s protocol (Invitrogen). Briefly, 0.2 mL of stock MBs were washed with 1 mL of 0.1 M borate buffer, pH 9.5. Then, 40 μ g of protein per mg MBs were added and allowed to mix in a slow tilt rotor for 30 min. In the next step, BSA was added to final concentration of 0.1% and allowed to mix for additional 24 h. The MBs were washed twice with 0.1% BSA/ PBS, chemically and physically blocked with 0.2 M Tris and 0.1% BSA buffer overnight. MBs were washed twice with 0.1% BSA/ PBS, reconstituted in 0.4 mL of 0.1% BSA, and stored at 4 °C until further use.

Microchip device fabrication. The microchip was fabricated using a non-lithographic technique as previously reported47,48. The device consisted of two clear poly(methyl methacrylate) (PMMA) laser cut sheets assembled and bonded. The templates were designed using AutoCAD software. The designs were then supplied to vendor (Ying Kwang Acrylic Trading, Singapore) for laser cutting services. The top template was laser cut on 3-mm thick PMMA, whereas the bottom template was cut on 1.5-mm thick PMMA. The two tem-plates were bonded using spray adhesive (3M, Super 75). The adhesive was allowed to bond and dry for 20 min before the microchip was ready to use. The approximate dimensions of the six-channel microchip device were 95 mm × 70 mm × 5 mm (length × breadth × height).

MTBE principle. The microchip device consists of six channels with each channel featuring seven circular chambers interconnected alternately with six rhombus chambers. The circular chambers contain 70 μ L of aque-ous reagents, e.g. lipid/protein-coated MBs, wash buffer, detection tracer antibody, polymeric enzyme labels and colorimetric substrate, whereas the rhombus chambers contain 70 μ L of immiscible silicone oil. The oil provides a barrier between the aqueous reagents, and permits manually assisted magnetic actuation of beads between the reagents. Plasma and reagents were diluted in 5% fatty acid free BSA in PBS buffer. In channels 1, 2, and 3, an equal mixture of antigen-coated beads (5 μ L, 106 beads ) and BSA-coated beads (5 μ L, 106 beads ) were used, whereas channel 4 and 5 contained equal mixtures of TDM-coated MBs (5 μ L) and 38 kDa- or Ag85A-coated MBs (5 μ L); channel 6 contained BSA-coated MBs (10 μ L, 2 × 106 beads) as a control. In the first step, beads were reacted with 60 μ L of plasma (1:200) for 3 min. After the capture of specific IgGs, beads were magnetically actu-ated to the wash chamber (30 sec) to remove nonspecific plasma proteins. Beads were then actuated to the biotin anti-hIgG (500 ng/mL) chamber for biotin-antibody labelling for 2 min. After brief washing (30 sec), the beads were then actuated to the streptavidin poly-HRP (500 ng/mL) chamber for an additional 2 min for enzyme label-ling. After another wash (30 sec), beads were actuated to the one-step ultra TMB substrate solution and incubated for 5 min. The colorimetric reaction was stopped using an equal volume of 2M H2SO4 and the resultant solution was immediately transferred to a 96-well plate for absorbance measurement at 450 nm using a microplate reader (PerkinElmer Envision 2104 multilabel reader). Absorbance values for the test antigen beads (Channel 1–5) were subtracted from the BSA beads (Channel 6), and the difference correlated to the amount of lipid/protein-specific antibodies present in the plasma. The total time of the microchip ELISA assay is ~15 min.

Data analysis. The absorbance values for the microchip and conventional plate or MB-bound assay was measured at 450 nm, by subtracting the values of the control BSA-coated MBs from the test antigen-coated MBs. The cut-off point (horizontal dotted line on the figures) and the sensitivity of the assay were determined by main-taining a constant specificity of 75%. This limit was chosen to compare the sensitivity to multiple antigens at a constant specificity. The ROC curves were generated in GraphPad Prism 5 software, by plotting the true positive rate and the false positive rate. Positive predicting values were obtained from the ratio of true positive samples and the sum of true and false positive samples, whereas negative predicting values were obtained from the ratio

Page 10: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

www.nature.com/scientificreports/

1 0Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

of true negative samples and the sum of true and false negative samples. Significant differences in IgG responses between different populations were determined using the Mann-Whitney unpaired t-test (GraphPad Prism 5).

References1. Stewart, G. R., Robertson, B. D. & Young, D. B. Tuberculosis: a problem with persistence. Nat Rev Microbiol 1, 97–105 (2003).2. Pai, M., Kalantri, S. & Dheda, K. New tools and emerging technologies for the diagnosis of tuberculosis: part II. Active tuberculosis

and drug resistance. Expert Rev Mol Diagn 6, 423–432 (2006).3. Caminero, J. A., Sotgiu, G., Zumla, A. & Migliori, G. B. Best drug treatment for multidrug-resistant and extensively drug-resistant

tuberculosis. Lancet Infect Dis 10, 621–629 (2010).4. Steingart, K. R. et al. Sputum processing methods to improve the sensitivity of smear microscopy for tuberculosis: a systematic

review. Lancet Infect Dis 6, 664–674 (2006).5. Adler, H., Straub, C. & Frei, R. Comparison of BacT/ALERT 3D, Lowenstein-Jensen medium and Middlebrook 7H10/7H11 biplate

for recovering mycobacteria from clinical specimens. Eur J Clin Microbiol Infect Dis 24, 499–500 (2005).6. Storla, D. G., Yimer, S. & Bjune, G. A. A systematic review of delay in the diagnosis and treatment of tuberculosis. BMC Public Health

8, 15 (2008).7. Steingart, K. R. et al. Commercial serological antibody detection tests for the diagnosis of pulmonary tuberculosis: a systematic

review. PLoS Med 4, e202 (2007).8. Special Programme for Research and Training in Tropical Diseases, Laboratory-based evaluation of 19 commercially available rapid

diagnostic tests for tuberculosis. WHO Geneva (2008).9. Niemz, A. & Boyle, D. S. Nucleic acid testing for tuberculosis at the point-of-care in high-burden countries. Expert Rev Mol Diagn

12, 687–701 (2012).10. McNerney, R. & Daley, P. Towards a point-of-care test for active tuberculosis: obstacles and opportunities. Nat Rev Microbiol 9,

204–213 (2011).11. World Health Organization, Tuberculosis Diagnostics Technology and Market Landscape, 4th edition (2015).12. Chin, C. D., Linder, V. & Sia, S. K. Commercialization of microfluidic point-of-care diagnostic devices. Lab Chip 12, 2118–2134

(2012).13. Sackmann, E. K., Fulton, A. L. & Beebe, D. J. The present and future role of microfluidics in biomedical research. Nature 507,

181–189 (2014).14. Wang, S. et al. Integration of cell phone imaging with microchip ELISA to detect ovarian cancer HE4 biomarker in urine at the

point-of-care. Lab Chip 11, 3411–3418 (2011).15. Gijs, M. A., Lacharme, F. & Lehmann, U. Microfluidic applications of magnetic particles for biological analysis and catalysis. Chem

Rev 110, 1518–1563 (2010).16. Mani, V., Chikkaveeraiah, B. V. & Rusling, J. F. Magnetic particles in ultrasensitive biomarker protein measurements for cancer

detection and monitoring. Expert Opin Med Diagn 5, 381–391 (2011).17. Gijs, M. M. Magnetic bead handling on-chip: new opportunities for analytical applications. Microfluid Nanofluid 1, 22–40 (2004).18. Lehmann, U. et al. Two-dimensional magnetic manipulation of microdroplets on a chip as a platform for bioanalytical applications.

Sensors and Actuators B: Chemical 117, 457–463 (2006).19. Lehmann, U., Vandevyver, C., Parashar, V. K. & Gijs, M. A. Droplet-based DNA purification in a magnetic lab-on-a-chip. Angew

Chem Int Ed Engl 45, 3062–3067 (2006).20. van Reenen, A., de Jong, A. M., den Toonder, J. M. & Prins, M. W. Integrated lab-on-chip biosensing systems based on magnetic

particle actuation–a comprehensive review. Lab Chip 14, 1966–1986 (2014).21. Wang, S. et al. Micro-a-fluidics ELISA for rapid CD4 cell count at the point-of-care. Sci Rep 4, 3796 (2014).22. Berry, S. M., Regehr, K. J., Casavant, B. P. & Beebe, D. J. Automated operation of immiscible filtration assisted by surface tension

(IFAST) arrays for streamlined analyte isolation. J Lab Autom 18, 206–211 (2013).23. Sur, K. et al. Immiscible phase nucleic acid purification eliminates PCR inhibitors with a single pass of paramagnetic particles

through a hydrophobic liquid. J Mol Diagn 12, 620–628 (2010).24. den Dulk, R. C., Schmidt, K. A., Sabatte, G., Liebana, S. & Prins, M. W. Magneto-capillary valve for integrated purification and

enrichment of nucleic acids and proteins. Lab Chip 13, 106–118 (2013).25. He, H. et al. Rapid serodiagnosis of human mycobacteriosis by ELISA using cord factor (trehalose-6,6′ -dimycolate) purified from

Mycobacterium tuberculosis as antigen. FEMS Microbiol Immunol 3, 201–204 (1991).26. Kunnath-Velayudhan, S. et al. Dynamic antibody responses to the Mycobacterium tuberculosis proteome. Proceedings of the

National Academy of Sciences 107, 14703–14708 (2010).27. Lim, J. H. et al. Purification and immunoreactivity of three components from the 30/32-kilodalton antigen 85 complex in

Mycobacterium tuberculosis. Infect Immun 67, 6187–6190 (1999).28. Maekura, R. et al. Clinical evaluation of anti-tuberculous glycolipid immunoglobulin G antibody assay for rapid serodiagnosis of

pulmonary tuberculosis. J Clin Microbiol 39, 3603–3608 (2001).29. He, X. Y. et al. Clinical evaluation of the recombinant 38 kDa protein of Mycobacterium tuberculosis. Scand J Infect Dis 40, 121–126

(2008).30. He, X.-Y. et al. Assessment of Five Antigens from Mycobacterium tuberculosis for Serodiagnosis of Tuberculosis. Clinical and

Vaccine Immunology: CVI 18, 565–570 (2011).31. Julian, E. et al. Serodiagnosis of tuberculosis: comparison of immunoglobulin A (IgA) response to sulfolipid I with IgG and IgM

responses to 2,3-diacyltrehalose, 2,3,6-triacyltrehalose, and cord factor antigens. J Clin Microbiol 40, 3782–3788 (2002).32. Mizusawa, M. et al. Increased synthesis of anti-tuberculous glycolipid immunoglobulin G (IgG) and IgA with cavity formation in

patients with pulmonary tuberculosis. Clin Vaccine Immunol 15, 544–548 (2008).33. Retzinger, G. S., Meredith, S. C., Takayama, K., Hunter, R. L. & Kezdy, F. J. The role of surface in the biological activities of trehalose

6,6′ -dimycolate. Surface properties and development of a model system. J Biol Chem 256, 8208–8216 (1981).34. Muller, R. H. & Heinemann, S. Fat emulsions for parenteral nutrition. I: Evaluation of microscopic and laser light scattering methods

for the determination of the physical stability. Clin Nutr 11, 223–236 (1992).35. Chan, C. E. et al. Novel phage display-derived mycolic acid-specific antibodies with potential for tuberculosis diagnosis. J Lipid Res

54, 2924–2932 (2013).36. Thanyani, S. T., Roberts, V., Siko, D. G., Vrey, P. & Verschoor, J. A. A novel application of affinity biosensor technology to detect

antibodies to mycolic acid in tuberculosis patients. J Immunol Methods 332, 61–72 (2008).37. Francisco, A. Biomarker-based Prediction Models: Design and Interpretation Principles. Bioinformatics and Biomarker Discovery:

“Omic” Data Analysus for Personalized Medicine (2010).38. Lemaire, J. F. & Casenghi, M. New diagnostics for tuberculosis: fulfilling patient needs first. J Int AIDS Soc 13, 40 (2010).39. Steingart, K. R. et al. Commercial serological tests for the diagnosis of active pulmonary and extrapulmonary tuberculosis: an

updated systematic review and meta-analysis. PLoS Med 8, e1001062 (2011).40. Nabeshima, S. et al. Serum antibody response to tuberculosis-associated glycolipid antigen after BCG vaccination in adults. Journal

of infection and chemotherapy: official journal of the Japan Society of Chemotherapy 11, 256–258 (2005).

Page 11: Microchip-based ultrafast serodiagnostic assay for tuberculosis...chip device for rapid detection of IgG in plasma samples. Each microchip has six channels in which six different Each

www.nature.com/scientificreports/

1 1Scientific RepoRts | 6:35845 | DOI: 10.1038/srep35845

41. Mani, V., Wasalathanthri, D. P., Joshi, A. A., Kumar, C. V. & Rusling, J. F. Highly efficient binding of paramagnetic beads bioconjugated with 100,000 or more antibodies to protein-coated surfaces. Anal Chem 84, 10485–10491 (2012).

42. Rusling, J. F., Bishop, G. W., Doan, N. & Papadimitrakopoulos, F. Nanomaterials and biomaterials in electrochemical arrays for protein detection. J Mater Chem B Mater Biol Med 2 (2014).

43. Mani, V., Kadimisetty, K., Malla, S., Joshi, A. A. & Rusling, J. F. Paper-based electrochemiluminescent screening for genotoxic activity in the environment. Environ Sci Technol 47, 1937–1944 (2013).

44. Tokel, O., Inci, F. & Demirci, U. Advances in plasmonic technologies for point of care applications. Chem Rev 114, 5728–5752 (2014).

45. Pai, N. P., Vadnais, C., Denkinger, C., Engel, N. & Pai, M. Point-of-care testing for infectious diseases: diversity, complexity, and barriers in low- and middle-income countries. PLoS Med 9, e1001306 (2012).

46. Pai, N. P. & Pai, M. Point-of-care diagnostics for HIV and tuberculosis: landscape, pipeline, and unmet needs. Discov Med 13, 35–45 (2012).

47. Moon, S. et al. Integrating microfluidics and lensless imaging for point-of-care testing. Biosens Bioelectron 24, 3208–3214 (2009).48. Moon, S. et al. Enumeration of CD4+ T-cells using a portable microchip count platform in Tanzanian HIV-infected patients. PLoS

One 6, e21409 (2011).

AcknowledgementsThis work was supported by A*STAR-CIMIT grant and BMRC-SERC diagnostic grant. We would like to thank Kerry McLaughlin of Insight Editing London for critical review of the manuscript, Fariyal Ahmed for preparing technology disclosure documents, Alessandra Nardin and Sebastien Bertin-Maghit for help in the IRB application. S.W. would like to acknowledge the support from the Ministry of Science and Technology of the People’s Republic of China (2016YFC1101302).

Author ContributionsA.S. conceived the project, and V.M. performed the experiments. B.P. and K.L. assisted with flow cytometry and TLC experiments. B.P., P.K. and A.S. helped with data analysis and maintaining the plasma sample inventory. J.A.T. and J.Y.S. process blood samples in BSL3. C.C. and Y.T.W. provided the blood samples for the study. F.I., S.W. and U.D. participated in analysis and discussion. G.D.L. and A.S. supervised the study. V.M., G.D.L. and A.S. analysed the results and wrote the manuscript.

Additional InformationSupplementary information accompanies this paper at http://www.nature.com/srepCompeting financial interests: V.M., A.S. and G.D.L. have filed a provisional PCT application for the MTBE assay.How to cite this article: Mani, V. et al. Microchip-based ultrafast serodiagnostic assay for tuberculosis. Sci. Rep. 6, 35845; doi: 10.1038/srep35845 (2016).

This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the article’s Creative Commons license,

unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/ © The Author(s) 2016