P-DUKE Technology Co., Ltd. www.pduke.com 2021.06.04 Page 1 Medical Automation Datacom IPC Industry Measurement Telecom Automobile Boat Charger PV Railway PART NUMBER STRUCTURE M S D 30 U S 12 B - T Application Package Dimension Output Input Output Output Protection Connector Code Code Power Voltage Quantity Voltage Type Options (W) (VAC) (VDC) Medical S:Encapsulated U: Universal S:Single 3P3:3.3 B:CLASS II T: Terminal Block Application D:Din rail type 85 ~ 264 05:5 M: Molex 7P5:7.5 D: Pin Type 09:9 : JST 12:12 121:12 15:15 151:15 18:18 24:24 28:28 36:36 48:48 53:53
10
Embed
Medical Automation Datacom IPC Industry Measurement ... Datasheet.pdfCON2: VHR-2N CON2: 09-50-8021 Crimp terminals Crimp terminals Wire dimension range 18 ~ 14AWG CON1: SVH-21T-P1.1
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
OUTPUT SPECIFICATIONS Parameter Conditions Min. Typ. Max. Unit
Output power 30 Watts Initial set voltage accuracy 230VAC and Full Load -1.0 +1.0 % Line regulation Low Line to High Line at Full Load -0.2 +0.2 % Load regulation No Load to Full Load 3.3Vout, 5Vout
Others -0.7 -0.5
+0.7 +0.5
% 10% Load to 90% Load 3.3Vout, 5Vout
Others -0.6 -0.4
+0.6 +0.4
Voltage adjustability -10 +10 % Minimum load 0 % Ripple and noise Measured by 20MHz bandwidth
mVp-p
With a 10μF/25V 1206 X7R MLCC
3.3Vout, 5Vout, 7.5Vout, 9Vout
50 50
With a 1μF/50V 1206 X7R MLCC
12Vout, 15Vout, 18Vout 24Vout, 28Vout, 36Vout
50 50
With a 0.1μF/100V 1206 X7R MLCC 48Vout, 53Vout 50 Temperature coefficient -0.02 +0.02 %/℃ Transient response Load step form 50 ~ 75% change at 2.5A/μs Peak deviation
Recovery time
500
3
%Vout μs
Over voltage protection % of Vout(nom); Automatics recovery 125 140 % Over load protection % of Iout rated; Hiccup mode 140 % Short circuit protection Continuous, automatics recovery
GENERAL SPECIFICATIONS
Parameter Conditions Min. Typ. Max. Unit Isolation voltage 1 minute (2MOPP insulation) Input to Output 4000 VAC Isolation resistance 500VDC 0.1 GΩ Switching frequency 230VAC and Full Load 30 45 60 kHz Safety approvals IEC/ EN/ ANSI/AAMI ES 60601-1
IEC/ EN/ UL 62368-1 UL:E360199
UL:E193009 CB:UL(Demko)
Weight Connector type Pin type
119g (4.20oz) 106g (3.74oz)
MTBF MIL-HDBK-217F, Full load 3.341 x 106 hrs
ENVIRONMENTAL SPECIFICATIONS
Parameter Conditions Min. Typ. Max. Unit Operating ambient temperature Natural convection With derating -40 +85 ℃ Storage temperature range -40 +85 ℃ Operating altitude 5000 m Shock IEC60068-2-27 Vibration IEC60068-2-6 Relative humidity Non-condensing 5% to 95% RH
EMI EN55011, EN55032, EN60601-1-2 and FCC Part 18 / 15
Conducted Radiated
Class B Class B
Harmonic currents EN61000-3-2 Full Load Class A Voltage flicker EN61000-3-3 EMS EN55024 and EN60601-1-2 ESD EN61000-4-2 Perf. Criteria A Radiated immunity EN61000-4-3 20 V/m Perf. Criteria A Fast transient EN61000-4-4 ± 2kV Perf. Criteria A Surge EN61000-4-5 DM ± 1kV Perf. Criteria A Conducted immunity EN61000-4-6 20 Vr.m.s Perf. Criteria A Power frequency magnetic field EN61000-4-8 30 A/m Perf. Criteria A Dip and interruptions EN61000-4-11
All dimensions in inch[mm] Pad size(lead free recommended) Through hole 1.2.3.4.5: Φ0.051[1.30] Top view pad 1.2.3.4.5: Φ0.064[1.63] Bottom view pad 1.2.3.4.5: Φ0.102[2.60]
MECHANICAL DRAWING MSD –D Pin Type PIN CONNECTION
PIN SINGLE 1 Neutral 2 Line 3 +Vout 4 -Vout 5 Trim
1. All dimensions in inch [mm] 2. Tolerance :x.xx±0.02 [x.x±0.5]
OUTPUT VOLTAGE ADJUSTMENT It allows the user to increase or decrease the output voltage of the module. This is accomplished by connecting an external resistor between the Trim pin and either the +Vout or -Vout pins. With an external resistor between the Trim and -Output pin, the output voltage increases. With an external resistor between the Trim and +Output pin, the output voltage decreases. The external Trim resistor needs to be at least 1/16W of rated power.
Trim Up Equation
H
KLV
LGR
up,o
U
Trim Down Equation
H
VV
GLVR
down,oo
down,oD
Trim Constants
Module G H K L
MSD30US3P3B 5100 2050 0.8 2.5
MSD30US05B 7500 2000 2.5 2.5
MSD30US7P5B 22000 2000 5 2.5
MSD30US09B 33000 2000 6.5 2.5
MSD30US12B 51000 2000 9.5 2.5
MSD30US15B 68000 2000 12.5 2.5
MSD30US18B 91000 2000 15.5 2.5
MSD30US24B 130000 2000 21.5 2.5
MSD30US28B 160000 2000 25.5 2.5
MSD30US36B 220000 2000 33.5 2.5
MSD30US48B 620000 2000 45.5 2.5
MSD30US53B 680000 2000 50.5 2.5
EXTERNAL OUTPUT TRIMMING Output can be externally trimmed by using the method shown below.