Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006
17

Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Dec 31, 2015

Download

Documents

Rosanna Robbins
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Luminosity Monitoring For PrimEx

Dan DalePrimEx Collaboration Meeting

December 16, 2006

Page 2: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Conclusion 1: What we have nowLuminosity Error Budget

________________________________________effect error(%)Uncorrected leakage current 0.5 (1.4%/3) TAC statistics 0.2 (0.5 per T channel)Electron counting systematics 0.8 (clock versus out of time)Electron counting statistics negligibleTAC reproducibility 0.4HYCAL collimation of TAC 0.25 (worst case, no photon collimator)Beam position/collimator effect 0.2Absorption in TAC runs not yet evaluated_______________________________________________________________

1.1%

Page 3: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Beam leakage from other Halls

If not corrected: ~1.4% effect on 1/3 of runs = 0.5% error

Page 4: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

TAC Statistics

• +- 0.5% per T channel

• If we combine all 0.2%

Electron Counting Systematics

0.8% discrepancy between CLOCK and OUT-OF-TIME methods

Electron Counting Statistics

Negligible.

Page 5: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Dependence of Tagging Ratios on Beam Intensity

Page 6: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Dependence of Tagging Ratios on Beam Intensity

Page 7: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Effect of Photon Beam Collimator

Page 8: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Effect of Photon Beam Collimator

Page 9: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Effect of Collimator Position on Tagging Ratios

Page 10: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Effect of Collimator Position on Tagging Ratios

Page 11: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Effect of Beam Missteering on Tagging Ratios

Run 4338: centeredRun 4340: 5mm upRun 4341: 5.7 mm X 0.22mm YRun 4342: 8.62 X Run 4343: 6 mm X

Page 12: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Short Term Stability of Tagging Ratios – Consecutive Runs

Page 13: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Short Term Stability of Tagging Ratios – Consecutive Runs

Page 14: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Long Term Stability of Tagging Ratios

Page 15: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Long Term Stability of Tagging Ratios

Page 16: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Dependence of Relative Tagging Ratios on Rate

Page 17: Luminosity Monitoring For PrimEx Dan Dale PrimEx Collaboration Meeting December 16, 2006.

Conclusion 2: What we can ultimately do.

Luminosity Error Budget________________________________________effect error(%)Corrected leakage current 0.3 (1.4%/3/2) TAC statistics 0.2 (0.5 per T channel)electron counting systematics 0.8 (clock versus out of time)Electron counting statistics negligibleTAC reproducibility 0.4HYCAL collimation of TAC 0.25 (worst case, no photon collimator)Beam position/collimator effect 0.2Absorption in TAC runs not yet evaluated_______________________________________________________________

1.0%