Low-dose three-dimensional x-ray imaging of bacterial cells: supporting information Matthias Bartels *1 , Marius Priebe 1 , Robin N. Wilke 1 , Sven P. Kr¨ uger 1 , Klaus Giewekemeyer 1 , Sebastian Kalbfleisch 1 , Christian Olendrowitz 1 , Michael Sprung 2 and Tim Salditt *1 1 Institut f¨ ur R¨ontgenphysik, Georg-August-Universit¨ at G¨ottingen, Friedrich-Hund37073 G¨ottingen, Germany 2 HASYLAB at DESY, Notkestr. 85, 22607 Hamburg, Germany Email: Matthias Bartels * - [email protected]; Tim Salditt - [email protected]; * Corresponding author 1
7
Embed
Low-dose three-dimensional x-ray imaging of bacterial cells ...10.1186/2192...ux of 2:4 1011 counts per second at 70 mA ring current, as measured by a pixel detector (Pilatus, Dectris)
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Low-dose three-dimensional x-ray imaging of bacterial cells:supporting information
Matthias Bartels∗1, Marius Priebe1, Robin N. Wilke1, Sven P. Kruger1, Klaus Giewekemeyer1,Sebastian Kalbfleisch1, Christian Olendrowitz1, Michael Sprung2 and Tim Salditt∗1
The experiment was carried out at the holographic imaging end-station of the coherence beamline P10 of
PETRA III. The source in low β configuration consists of a 5 m long undulator with a period of 29 mm,
with a source size of 36µm× 6µm (1σ, horz.× vert.). The undulator beam was monochomatized by a
Si(111) double crystal to 13.8 keV and then focused by two Kirkpatrick-Baez (KB) mirrors polished to
fixed elliptical shape positioned at a distance of 87.7 m behind the source, as described in Kalbfleisch et
al. (2011); Salditt et al. (2011). The mirror alignment and beam profile in the focal plane was controlled
by scanning horizontally and vertically aligned planar x-ray waveguides through the beam, yielding typical
focus sizes of Dhorz = 370 nm and Dvert = 200 nm (best value: Dvert = 120 nm) full width at half maximum
(FWHM), respectively, and a total flux of 2.4× 1011 counts per second at 70 mA ring current, as measured
by a pixel detector (Pilatus, Dectris) positioned at 5.29 m in the widened far-field of the KB beam. The
waveguide (WG) system was positioned in the focal plane of the KB mirror, using a miniaturized fully
motorized goniometer with optical encoders (Attocube), three translations in xyz, and two rotations along
two directions orthogonal to the optical axis. Alignment of the waveguide as well as the sample was facilitated
by use of two on-axis optical microscopes, one directed downstream and one upstream with the beam. The
focal planes of both microscopes coincide with the KB focal plane in their respective reference positions,
and allow accurate determination of the (defocus) distance z1 between waveguide exit and sample plane
(Kalbfleisch et al. 2011; Salditt et al. 2011). The sample stage is equipped with an air-bearing rotation
(Micos) for ultra-high precision turns suitable for nano-tomography. On top of the rotation, a group of
xyz piezos (Physik Instrumente) is used for aligning the sample in the axis of rotation. Additional xyz
stages (Micos) below the rotation are used for aligning the rotation axis in the x-ray beam and for distance
variation between the waveguide and the sample. The detectors were placed at a distance of z1 +z2 = 5.29 m
behind the sample, see Fig. 1 of main manuscript. To record the empty beam far-field distribution needed to
reconstruct the near-field intensity distribution of the waveguide system, a single photon counting detector
with a rather large pixel size of 172 µm (Pilatus, Dectris), and correspondingly large angular range was
used. This detector also provides a particularly large dynamic range for waveguide alignment and KB flux
calibration. The projected holograms of the biological cells needed finer sampling, and were recorded with
a single photon counting detector of 55 µm pixel size (Maxipix) (Ponchut et al. 2007) without readout noise
or dark current. This single photon characteristic is essential for phase reconstruction algorithms taking into
account the noise distribution.
2
Specimen preparation
Cells of the Deinococcus radiodurans wild-type strain were cultivated from freeze-dried cultures (DSM
No. 20539 by the German Collection of Microorganisms and Cell Cultures) for one day at 37◦ C on petri
dishes covered with nutrient medium (corynebacterium agar: 10 g/l casein peptone, 5 g/L yeast extract,
5 g/l glucose, 5 g/l NaCl, 15 g/l agar). Prior to preparation the actively growing cells were washed off
the culturing medium with ca. 1.5 ml buffer solution (2 g/l KH2PO4, 0.36 g/L Na2HPO4·2H2O, pH 7.2).
After placing a droplet of cell suspension onto the substrate, a Si3N4-foil (Silson) of 1 µm thickness and
5 × 5 mm2 lateral dimensions, the cells were allowed to adhere for 60 seconds. The remaining buffer was
blotted, the frame cryoplunged into liquid ethane to prevent crystallization (Dubochet et al. 1988) and
afterwards lyophilized in a home-built freeze-drier.
Raw data processing
For all projections N = 15 intensity distributions I(n)φ were collected with 40 second dwell time for each
measurement. To correct for lateral drift the N intensity distributions I(n)φ were aligned with respect to each
other by cross-correlation methods with sub-pixel accuracy (Guizar-Sicairos et al. 2008) (upsampling factor
100). The aligned intensity distributions I(n)φ,cc where then summed up yielding Iφ =
∑n I
(n)φ,cc. Fig. 1 shows
the improvement in data quality achieved by this method. The aligned data shows more fringes in comparison
to the sum of the non-aligned intensity distributions. To correct for unhomogenous illumination the intensity
distribution I0(x, y) of the empty beam was recorded during an exposure time of 40 minutes. The normalized
intensity distribution was then calculated as Iφ(x, y) = Iφ/I0 · 40 min/10 min. Residual low frequency-
variations were removed effectively by applying a Gaussian highpass filter and a subsequent correction
through division by the mean intensity fluctuations in the regions outside the cellular area independently
for the vertical and horizontal direction (Giewekemeyer et al. 2011). We find that applying a median filter
using a 3-by-3 neighborhood prior to phase reconstruction yields superior results.
Automated generation of the support area
For tomographic datasets it is crucial to determine the support area S ⊂ R2 for all projections automatically
and accurately. As a first step the same rectangular support is chosen for all projections as an initial
guess, as shown in Fig. 2 (a) for the projection corresponding to θ = 0◦. Using 150 iterations of the phase
reconstruction algorithm an approximate reconstruction of the sample is obtained (see Fig. 2 (b)). By
3
application of a threshold (−0.08 rad) to the reconstructed phase a binary image is generated. Application
of subsequent morphological operations (erosion and dilation) yield a suitable support area (see Fig. 2(c)).
Using this support area one obtains a phase reconstruction (see Fig. 2(d)) which can be used to refine the
support area by repeating the steps described above, similar to the Shrink-Wrap algorithm suitable for CDI
experiments (Marchesini et al. 2003).
Treatment of noise in iterative reconstruction
Experimental noise in the diffraction data can significantly disturb the results of iterative reconstruction
schemes (Williams et al. 2007) which are based on the implicit assumption that the measured intensity is
a noise-free representation of the diffracted intensity. To prevent possible overfitting of the noise in the
normalized intensity distribution a modified projector applied to the image intensity is used as described in
(Giewekemeyer et al. 2011). The detection plane intensity of the updated χ′n at position (x, y) then reads
|χ′n(x, y))|2 :=
(1− D
d
)I(x, y) +
D
d|χn(x, y)|2 (1)
if d > D, where d is defined by the the misfit
d2(|χn|2) := 1/N∑(x,y)
(|χn(x, y)|2 − I(x, y)
)2(2)
between the reconstructed and measured holographic intensity distribution. Here χ(x, y) = Dzeff [χ(x, y)]
denotes the near-field propagated object wave and N denotes the number of points (x, y), or pixels, at which
I was measured. The algorithm stops as soon as d ≤ D. An optimum value for the parameter D depends on
the noise statistics of the the experimental data. For the present analyis a median filter was applied prior
to phase reconstruction. A natural choice for the threshold parameter D is the standard deviation σ of the
normalized intensity distribution outside the cellular area (Giewekemeyer et al. 2011). The smallest D, for
which convergence in the sense defined above could be achieved within a reasonable amount of time (about
2000 iterations on average), was found to be in the interval of D = [0.9, 1.1]σ for all projections.
Extraction of mass density maps
As described earlier (Giewekemeyer et al. 2010) the reconstructed phase distribution ϕ(x, y) can be rescaled
into a projected electron density map σe(x, y) using the relation σe(x, y) = −ϕ/(λr0), where λ is the
wavelength of the incident radiation and r0 the classical electron radius. σe is further related to the mass
density σm as σm = A/Z · u · σe with the atomic mass unit u and A, Z denoting the atomic number and the
4
mass number of the imaged material, averaged over one resolution volume. Thus one can relate the phase of
the object transmission function at point r in the object plane uniquely with an effective area mass density
σm := 2u · σe via
σm(r) = −(
2u
λr0
)· ϕ(r). (3)
The absolute area mass density is then given as σm = A/(2Z) · σm. For most biological material the factor
A/(2Z) deviates from 1 by less than 10%.
Dose
The dose quantification in this experiment relies on the single photon counting capability of the pixel de-
tectors. Summing up the photons traversing the sample and registered on the detector is easily possible,
since the support of the sample can be directly inferred from the holographic image. Taking the quantum
efficiency of the detector (Maxipix with 500 µm Si (Ponchut et al. 2007)) at 13.8 keV photon energy, ab-
sorption in window materials, and residual air-scattering into account, the dose taken up by the sample is
actually higher by 40%. However, for for the relationship between imaging resolution and dose, this dose
as measured by the detector is the relevant quantity, so that we stick to this quantity. The total (detector)
dose applied during the tomographic scan was about 1.6 · 105 Gy, based on calculations presented in Howells
et al. (2009), corresponding to a dose applied to the sample of 2.2 · 105 Gy.
5
ReferencesDubochet J, Adrian M, Chang JJ et al (1988) Cryo-electron microscopy of vitrified specimens. Q Rev Biophys 21:129
Ponchut C, Clement J, Rigal J M (2007) Photon-counting X-ray imaging at kilohertz frame rates. Nucl Instrum MethA 576:109
Giewekemeyer K, Thibault P, Kalbfleisch S et al (2010) Quantitative biological imaging by ptychographic x-raydiffraction microscopy. PNAS 107:529
Giewekemeyer K, Kruger S P, Kalbfleisch S et al (2011) X-ray propagation microscopy of biological cells usingwaveguides as a quasipoint source. Phys Rev A 83:023804
Guizar-Sicairos M, Thurman S T, Fienup J R (2008) Efficient subpixel image registration algorithms. Opt Lett 33:156
Howells M R, Beetz T, Chapman H N et al (2009) An assessment of the resolution limitation due to radiation-damagein X-ray diffraction microscopy. J El Spec Rel Phen 170:4
Kalbfleisch S, Neubauer H, Kruger S P et al (2011) The Gottingen Holography Endstation of Beamline P10 atPETRA III/DESY. AIP Conf Proc 1365:96
Marchesini S, He H, Chapman H N et al (2003) X-ray image reconstruction from a diffraction pattern alone. PhysRev B 68:140101
Salditt T, Kalbfleisch S, Osterhoff M et al (2011) Partially coherent nano-focused x-ray radiation characterized byTalbot interferometry. Opt Express 19:9656
Williams G, Pfeifer M, Vartanyants I et al (2007) Effectiveness of iterative algorithms in recovering phase in thepresence of noise. Acta Cryst A 63:36
6
FiguresFigure 1
(a) Sum of the non-aligned intensity distributions, normalized by the empty beam. (b) Normalized intensity
distribution obtained by aliging N = 15 intensity distributions with respect to each other with sub-pixel
accuracy before summation.
0.9
1
1.1
0.9
1
1.1
(a) (b)
Figure 2
(a) Holographic intensity corresponding to θ = 0◦. The rectangular support (dashed rectangle) is used as
an initial guess for all projections. (b) Approximate reconstruction of the sample after 150 iterations of the
phase reconstruction algorithm. (c) Support area generated by application of a threshold and subsequent
morphological operations. (d) Phase reconstruction obtained by using the support area shown in (c).