Legacy™4 Fixture-Mount & IQity Impression Technique™ Technical Bulletin Fixture-mount top designed to fail at approximately 112 Ncm Index fixture-mount flat to labial or buccal Place implant into osteotomy Vial labels are color-coded with two tear-off labels Each implant is packaged in a sterile vial Open vial and release implant assembly onto a sterile surface Cap labels are color-coded by platform diameter Remove implant assembly with hex tool Hex will then be aligned for proper abutment placement Place the fixture-mount on a stable, sterile surface with the pronged end facing up If, in the unlikely event, the fixture-mount top becomes loose during insertion, remove it with the hex tool. Implant insertion may be continued via one of the Legacy4 Alternative Insertion Options Insert the appropriate pointed end of the Legacy™4 Fixture- Mount Tool [FMT] into the pronged end of the fixture- mount and press completely down to ensure the prongs are correctly positioned Legacy4 Implant is mounted on a revolutionary two-piece color-coded fixture-mount. The top fails at excessive torques to protect the interface during insertion and detaches in impression for the accuracy of an open-tray transfer with the simplicity of a closed-tray. The fixture-mount may be shortened at the line to function as a final preparable abutment. A healing collar is included in each vial.