IAEA International Atomic Energy Agency RTC on RADIATION PROTECTION OF PATIENTS FOR RADIOGRAPHERS Accra, Ghana, July 2011 Image Quality and Patient Dose
Nov 11, 2014
IAEAInternational Atomic Energy Agency
RTC on RADIATION PROTECTION OF PATIENTS FOR RADIOGRAPHERS
Accra, Ghana, July 2011
Image Quality and Patient Dose
IAEA 2
Overview
• To become familiar with the factors that determine the image clarity and the way the image quality can be improved
IAEA 3
Imaging quality
• Efficient diagnosis requires • acceptable noise
• good image contrast
• sufficient spatial resolution
• These factors are linked
• “Objective” measurement of quality is difficult
IAEA 4
Factors affecting image quality
Image quality
Contrast
Distortion & artifact Noise
Blur or Unsharpness
IAEA 5
Image contrast
Low Contrast
Medium Contrast
High Contrast
Image contrast refers to the fractional difference in optical density of brightness between two regions of an image
IAEA 6
Some factors influencing contrast
• Radiographic or subject contrast• Tissue thickness
• Tissue density
• Tissue electron density
• Effective atomic number Z
• X Ray energy in kVp
• X Ray spectrum (filtration)
• Scatter rejection• Collimator
• Grid
• …
• Image contrast• The radiographic contrast
plus :
• Film characteristics
• Screen characteristics
• Windowing level of CT and DSA
IAEA 7
• Peak voltage value has an influence on the beam hardness (beam quality)
• It has to be related to medical question• What is the anatomical structure to be
investigated?
• What is the contrast level needed?
• For a thorax examination : 110 - 120 kV is suitable to visualize the lung structure
• However only 65 kV is necessary to see bone structure
Technique factors (1)
IAEA 8
Technique factors (2)
• The higher the energy, the greater the penetrating power of X Rays
• At very high energy levels, the difference between bone and soft tissue decreases and both become equally transparent
• Image contrast can be enhanced by choosing a lower kVp so that photoelectric interactions are increased
• Higher kVp is required when the contrast is high (chest)
IAEA 9
Technique factors (3)
• The mAs controls the quantity of X Rays (intensity or number of X Rays)
• X Ray intensity is directly proportional to the mAs
• Over or under-exposure can be controlled by adjusting the mAs
• If the film is too “white”, increasing the mAs will bring up the intensity and optical density
IAEA 10
Receptor contrast
• The film as receptor has a major role to play in altering the image contrast
• There are high contrast and high sensitivity films
• The characteristic curve of the film describes the intrinsic properties of the receptor (base + fog, sensitivity, mean gradient, maximum optical density)
• N.B.: Film processing has a pronounced effect on fog and contrast
IAEA 11
Image Contrast
• Difference in signal – pixel value, film density
High Low
IAEA 12
Video monitor
• The video monitor is commonly used in fluoroscopy and digital imaging • The display on the monitor adds flexibility in the choice
of image contrast
• The dynamic range of the monitor is limited (limitation in displaying wide range of exposures)
• Increased flexibility in displaying image contrast is achieved by adjustment of the window level or grey levels of a digital image
IAEA 13
Blur or lack of sharpness
• The boundaries of an organ or lesion may be very sharp but the image shows a lack of sharpness
• Different factors may be responsible for such a degree of “fuzziness” or blurring
• The radiologist viewing the image might express an opinion that the image lacks “detail” or “resolution” (subjective reaction of the viewer to the degree of sharpness present in the image)
IAEA 14
Resolution
• Smallest distance that two objects can be separated and still appear distinct
• Example of limits• Film/screen: 0.01 mm
• CT: 0.5 mm
• Other definition: “Point-spread” function• Characteristic of a “point” object
• Point object expected to be point in image
• Blurring due to imperfections of imaging system
IAEA 15
Factors affecting image sharpness
Image Unsharpness
GeometricUnsharpness
Object Unsharpness
Motion Unsharpness
Subject Unsharpness
IAEA 16
Resolution and Focal Spot Size
Penumbra
Appearance of imageMore blur Less blur
IAEA 17
Measuring Resolution
Line pair test object
One line pair
{
IAEA 18
Geometric blur
• If the focal spot is infinitesimally small, the blur is minimized because of minimal geometric bluntness
• As the focal spot increases, the blur in the image increases
Small focal spot Large focal spot
IAEA 19
Geometric blur
• Another cause of lack of geometric sharpness is the distance of the receptor from the object
• Moving the receptor away from the object results in an increased lack of sharpness
• N.B.: The smaller the focal size and closer the contact between the object and the film (or receptor), the better the image quality as a result of a reduction in the geometric sharpness
IAEA 20
Lack of sharpness in the subject
• Not all structures in the body have well-defined and separate boundaries (superimposition essentially present in most situations)
• The organs do not have square or rectangular boundaries
• The fidelity with which details in the object are required to be imaged is an essential requirement of any imaging system
• The absence of sharpness, in the subject/object is reflected in the image
IAEA 21
Lack of sharpness due to motion (1)
• Common and understandable blur in medical imaging
• Patient movement :• uncooperative child
• organ contraction or relaxation
• heart beating, breathing etc.
• Voluntary motion can be controlled by keeping examination time short and asking the patient to remain still during the examination
IAEA 22
Lack of sharpness due to motion (2)
• Shorter exposure times are achieved by the use of fast intensifying screens
• N.B.: Faster screens result in loss of details (receptor sharpness)
• Further, the use of shorter exposure time has to be compensated with increased mA to achieve a good image
• This often implies use of large focal spot (geometric sharpness)
IAEA 23
Distortion and artifacts
• Unequal magnification of various anatomical structures
• Inability to give an accurate impression of the real size, shape and relative positions
• Grid artifact (grid visualized on the film)
• Light spot simulating microcalcifications (dust on the screen)
• Bad film screen contact, bad patient positioning (breast)
IAEA 24
Noise
• Defined as uncertainty or imprecision of the recording of a signal
• Impressionist painting: precision of object increases with number of dots
• X Ray imaging: when recorded with small number of X- photons has high degree of uncertainty,more photons give less noise
• Other sources of noise:• Grains in radiographic film• Large grains in intensifying screens• Electronic noise of detector or amplifier
IAEA 25
Noise
Decreasing radiation intensity
Increasing noise
IAEA 26
Contrast & Noise
No
ise Contrast
IAEA 27
Radiography Issues
• Correct positioning• Improves diagnosis and reduces retakes
• PRE-exposure collimation• Minimises unnecessary tissue dose
• With CR/DR, there is a temptation to post-exposure (electronically) collimate – RESIST THIS!!
IAEA 28
Summary
• Different technical and physical factors may
influence the image quality by impairing the
detection capability of the anatomical structures
useful for diagnosis (increasing the image
unsharpness)
• Some factors depend on the receptor, some others are
more related to the radiographic technique
IAEAInternational Atomic Energy Agency
Patient dose assessment
IAEA 30
Overview
• To become familiar with the patient dose assessment and dosimetry instrument characteristics.
IAEAInternational Atomic Energy Agency
Parameters influencing patient exposure
IAEA 32
Essential parameters influencing patient exposure
Tube voltageTube currentEffective filtration
Exposure time
Field size
Kerma rate[mGy/min]
[min]Kerma[Gy]
[m2]
Area exposureproduct [Gy m2 ]
}}
}
IAEA 33
Factors in conventional radiography: beam, collimation
• Beam energy• Depending on peak kV and filtration• Regulations require minimum total filtration to absorb
lower energy photons• Added filtration reduces dose• Goal should be use of highest kV resulting in acceptable
image contrast
• Collimation• Area exposed should be limited to area of CLINICAL
interest to lower dose• Additional benefit is less scatter, better contrast
IAEA 34
Factors in conventional radiography: grid,patient size
• Grids• Reduce the amount of scatter reaching image receptor• But at the cost of increased patient dose• Typically 2-5 times: “Bucky factor” or grid ratio
• Patient size• Thickness, volume irradiated…and dose increases with
patient size• Except for breast (compression), no control• Technique charts with suggested exposure factor for
various examinations and patient thickness helpful to avoid retakes
• Use of AEC exposure
IAEA 35
Factors affecting dose in fluoroscopy
• Beam energy and filtration• Collimation• Source-to-skin distance
• Inverse square law: maintain max distance from patient
• Patient-to-image intensifier• Minimizing patient-to- I I will lower dose• But slightly decrease image quality by increased scatter
• Image magnification • Geometric and electronic magnification increase dose
• Grid• If small sized patient (les scatter) perhaps without grid
• Beam-on time!
IAEA 36
Factors affecting dose in CT
• Beam energy and filtration• 120-140 kV; shaped filters
• Collimation or section thickness• Post-patient collimator will reduce slice thickness imaged
but not the irradiated thickness
• Number and spacing of adjacent sections
• Image quality and noise• Like all modalities: dose increase=>noise decreases
IAEA 37
Factors affecting dose in spiral CT
• Factors for conventional CT also valid
• Scan pitch• Ratio of couch travel in one rotation divided by
slice thickness
• If pitch = 1, doses are comparable to conventional CT
• Dose proportional to 1/pitch
IAEAInternational Atomic Energy Agency
Patient dosimetry methods
IAEA 39
Patient dosimetry
• Radiography: entrance surface dose ESD• Output factors
• Dose – area product (DAP)
• Fluoroscopy: Dose - Area Product (DAP)
• CT:• Computed Tomography Dose Index (CTDI)
• Dose – length product (DLP)
IAEA 40
From ESD to organ and effective dose
• Except for invasive methods, no organ doses can be measured
• The only way in radiography: measure the Entrance Surface Dose (ESD)
• Use mathematical models to estimate internal dose. • Mathematical models based on Monte Carlo
simulations• Dose to the organ tabulated as a fraction of the
entrance dose for different projections• Since filtration, field size and orientation play a role:
long lists of tables (See NRPB R262 and NRPB SR262)
IAEA 41
From DAP to organ and effective dose
• In fluoroscopy: moving field, measurement of Dose-Area Product (DAP)
• In similar way organ doses calculated by Monte Carlo modelling
• Based on mathematical model• Conversion coefficients estimated as organ doses
per unit dose-area product• Again numerous factors are to be taken into
account such as projection, filtration, …• Once organ doses are obtained, effective dose is
calculated following ICRP103
IAEAInternational Atomic Energy Agency
Dose measurements: how to measure dose indicators ESD, DAP,CTDI…
IAEA 43
Measurements of Radiation Output
X Ray tube
Filter
Ion. chamber
Lead slabTable top
SDD
Phantom (PEP)
IAEA 44
Measurements of Radiation Output
• Operating conditions
• Consistency check
• The output as a function of kVp
• The output as a function of mA
• The output as a function of exposure time
IAEA 45
Measurement of entrance surface dose
TLD, solid statedosimeter or ion chamber
Includes backscatter (~30%)
IAEA 46
Dose Area Product (DAP)
Transmission ionization chamber
IAEA 47
Dose Area Product (DAP)
0.5 m1 m
2 m
Air Kerma:Area:Areaexposure product
40*103 Gy2.5*10-3 m2
100 Gy m2
10*103 Gy10*10-3 m2
100 Gy m2
2.5*103 Gy40*10-3 m2
100 Gy m2
IAEA 48
Calibration of a Dose Area Product (DAP)
Film cassette
10 cm 10 cm
Ionizationchamber
IAEA 49
Levels of Dosimetry
• Level 1 - published tables
• Level 2 - Monte Carlo tables using known data
• Level 3 - direct measurement of skin dose
• Level 4 - humanoid phantom measurements with TLD
IAEA 50
Level 1 - Published Dose Tables
• ICRP, NCRP and various books have tables of “typical” doses for various x-rays
• Tables show organ doses and sometimes effective dose
• The data is usually old, from x-rays made with slower film/screen systems than in current use
• Still useful as a guide
IAEA 51
Level 2 - Monte Carlo Systems
• Provide organ doses, and effective dose
• Use calculated data but with user entry of various parameters :• HVL, kVp
• skin dose or free-in-air exposure at skin distance
• FSD, field size and position
IAEA 52
Monte Carlo Systems
• Various computer programs and lookup tables, eg. TISSDOSE, XDOSE, PCXMC
• Most users do not know the actual values for input variables, so often must use assumed values
• HVL, kVp, FSD, field size easy to assume – kVp/mAs and field size not always easy to assume
IAEA 53
Typical Radiology Doses (Melbourne Data)
Procedure ESD (mGy) Eff. Dose (mSv)
Abdomen (AP) 2.5 0.35
Chest (PA) 0.15 0.023
Lumbar spine AP/Lat 3.2/4.0 0.23/0.11
Pelvis (AP) 1.7 0.29
Thoracic spine (lat.) 2.6 0.097
Mammography (4 views) 4.4 0.44
IAEA 54
XDose
IAEA 55
IAEA 56
PCXMC
IAEA 57
IAEA 58
ImPACT CT Dose
IAEA 59
IAEA 60
Summary
• In this lesson we learned the factors influencing patient dose, and how to have access to an estimation of the detriment through measurement of entrance dose, dose area product or specific CT dosimetry methods.