Top Banner
IAEA International Atomic Energy Agency RTC on RADIATION PROTECTION OF PATIENTS FOR RADIOGRAPHERS Accra, Ghana, July 2011 Image Quality and Patient Dose
60
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Lecture 05 Image Quality and Patient Dose

IAEAInternational Atomic Energy Agency

RTC on RADIATION PROTECTION OF PATIENTS FOR RADIOGRAPHERS

Accra, Ghana, July 2011

Image Quality and Patient Dose

Page 2: Lecture 05 Image Quality and Patient Dose

IAEA 2

Overview

• To become familiar with the factors that determine the image clarity and the way the image quality can be improved

Page 3: Lecture 05 Image Quality and Patient Dose

IAEA 3

Imaging quality

• Efficient diagnosis requires • acceptable noise

• good image contrast

• sufficient spatial resolution

• These factors are linked

• “Objective” measurement of quality is difficult

Page 4: Lecture 05 Image Quality and Patient Dose

IAEA 4

Factors affecting image quality

Image quality

Contrast

Distortion & artifact Noise

Blur or Unsharpness

Page 5: Lecture 05 Image Quality and Patient Dose

IAEA 5

Image contrast

Low Contrast

Medium Contrast

High Contrast

Image contrast refers to the fractional difference in optical density of brightness between two regions of an image

Page 6: Lecture 05 Image Quality and Patient Dose

IAEA 6

Some factors influencing contrast

• Radiographic or subject contrast• Tissue thickness

• Tissue density

• Tissue electron density

• Effective atomic number Z

• X Ray energy in kVp

• X Ray spectrum (filtration)

• Scatter rejection• Collimator

• Grid

• …

• Image contrast• The radiographic contrast

plus :

• Film characteristics

• Screen characteristics

• Windowing level of CT and DSA

Page 7: Lecture 05 Image Quality and Patient Dose

IAEA 7

• Peak voltage value has an influence on the beam hardness (beam quality)

• It has to be related to medical question• What is the anatomical structure to be

investigated?

• What is the contrast level needed?

• For a thorax examination : 110 - 120 kV is suitable to visualize the lung structure

• However only 65 kV is necessary to see bone structure

Technique factors (1)

Page 8: Lecture 05 Image Quality and Patient Dose

IAEA 8

Technique factors (2)

• The higher the energy, the greater the penetrating power of X Rays

• At very high energy levels, the difference between bone and soft tissue decreases and both become equally transparent

• Image contrast can be enhanced by choosing a lower kVp so that photoelectric interactions are increased

• Higher kVp is required when the contrast is high (chest)

Page 9: Lecture 05 Image Quality and Patient Dose

IAEA 9

Technique factors (3)

• The mAs controls the quantity of X Rays (intensity or number of X Rays)

• X Ray intensity is directly proportional to the mAs

• Over or under-exposure can be controlled by adjusting the mAs

• If the film is too “white”, increasing the mAs will bring up the intensity and optical density

Page 10: Lecture 05 Image Quality and Patient Dose

IAEA 10

Receptor contrast

• The film as receptor has a major role to play in altering the image contrast

• There are high contrast and high sensitivity films

• The characteristic curve of the film describes the intrinsic properties of the receptor (base + fog, sensitivity, mean gradient, maximum optical density)

• N.B.: Film processing has a pronounced effect on fog and contrast

Page 11: Lecture 05 Image Quality and Patient Dose

IAEA 11

Image Contrast

• Difference in signal – pixel value, film density

High Low

Page 12: Lecture 05 Image Quality and Patient Dose

IAEA 12

Video monitor

• The video monitor is commonly used in fluoroscopy and digital imaging • The display on the monitor adds flexibility in the choice

of image contrast

• The dynamic range of the monitor is limited (limitation in displaying wide range of exposures)

• Increased flexibility in displaying image contrast is achieved by adjustment of the window level or grey levels of a digital image

Page 13: Lecture 05 Image Quality and Patient Dose

IAEA 13

Blur or lack of sharpness

• The boundaries of an organ or lesion may be very sharp but the image shows a lack of sharpness

• Different factors may be responsible for such a degree of “fuzziness” or blurring

• The radiologist viewing the image might express an opinion that the image lacks “detail” or “resolution” (subjective reaction of the viewer to the degree of sharpness present in the image)

Page 14: Lecture 05 Image Quality and Patient Dose

IAEA 14

Resolution

• Smallest distance that two objects can be separated and still appear distinct

• Example of limits• Film/screen: 0.01 mm

• CT: 0.5 mm

• Other definition: “Point-spread” function• Characteristic of a “point” object

• Point object expected to be point in image

• Blurring due to imperfections of imaging system

Page 15: Lecture 05 Image Quality and Patient Dose

IAEA 15

Factors affecting image sharpness

Image Unsharpness

GeometricUnsharpness

Object Unsharpness

Motion Unsharpness

Subject Unsharpness

Page 16: Lecture 05 Image Quality and Patient Dose

IAEA 16

Resolution and Focal Spot Size

Penumbra

Appearance of imageMore blur Less blur

Page 17: Lecture 05 Image Quality and Patient Dose

IAEA 17

Measuring Resolution

Line pair test object

One line pair

{

Page 18: Lecture 05 Image Quality and Patient Dose

IAEA 18

Geometric blur

• If the focal spot is infinitesimally small, the blur is minimized because of minimal geometric bluntness

• As the focal spot increases, the blur in the image increases

Small focal spot Large focal spot

Page 19: Lecture 05 Image Quality and Patient Dose

IAEA 19

Geometric blur

• Another cause of lack of geometric sharpness is the distance of the receptor from the object

• Moving the receptor away from the object results in an increased lack of sharpness

• N.B.: The smaller the focal size and closer the contact between the object and the film (or receptor), the better the image quality as a result of a reduction in the geometric sharpness

Page 20: Lecture 05 Image Quality and Patient Dose

IAEA 20

Lack of sharpness in the subject

• Not all structures in the body have well-defined and separate boundaries (superimposition essentially present in most situations)

• The organs do not have square or rectangular boundaries

• The fidelity with which details in the object are required to be imaged is an essential requirement of any imaging system

• The absence of sharpness, in the subject/object is reflected in the image

Page 21: Lecture 05 Image Quality and Patient Dose

IAEA 21

Lack of sharpness due to motion (1)

• Common and understandable blur in medical imaging

• Patient movement :• uncooperative child

• organ contraction or relaxation

• heart beating, breathing etc.

• Voluntary motion can be controlled by keeping examination time short and asking the patient to remain still during the examination

Page 22: Lecture 05 Image Quality and Patient Dose

IAEA 22

Lack of sharpness due to motion (2)

• Shorter exposure times are achieved by the use of fast intensifying screens

• N.B.: Faster screens result in loss of details (receptor sharpness)

• Further, the use of shorter exposure time has to be compensated with increased mA to achieve a good image

• This often implies use of large focal spot (geometric sharpness)

Page 23: Lecture 05 Image Quality and Patient Dose

IAEA 23

Distortion and artifacts

• Unequal magnification of various anatomical structures

• Inability to give an accurate impression of the real size, shape and relative positions

• Grid artifact (grid visualized on the film)

• Light spot simulating microcalcifications (dust on the screen)

• Bad film screen contact, bad patient positioning (breast)

Page 24: Lecture 05 Image Quality and Patient Dose

IAEA 24

Noise

• Defined as uncertainty or imprecision of the recording of a signal

• Impressionist painting: precision of object increases with number of dots

• X Ray imaging: when recorded with small number of X- photons has high degree of uncertainty,more photons give less noise

• Other sources of noise:• Grains in radiographic film• Large grains in intensifying screens• Electronic noise of detector or amplifier

Page 25: Lecture 05 Image Quality and Patient Dose

IAEA 25

Noise

Decreasing radiation intensity

Increasing noise

Page 26: Lecture 05 Image Quality and Patient Dose

IAEA 26

Contrast & Noise

No

ise Contrast

Page 27: Lecture 05 Image Quality and Patient Dose

IAEA 27

Radiography Issues

• Correct positioning• Improves diagnosis and reduces retakes

• PRE-exposure collimation• Minimises unnecessary tissue dose

• With CR/DR, there is a temptation to post-exposure (electronically) collimate – RESIST THIS!!

Page 28: Lecture 05 Image Quality and Patient Dose

IAEA 28

Summary

• Different technical and physical factors may

influence the image quality by impairing the

detection capability of the anatomical structures

useful for diagnosis (increasing the image

unsharpness)

• Some factors depend on the receptor, some others are

more related to the radiographic technique

Page 29: Lecture 05 Image Quality and Patient Dose

IAEAInternational Atomic Energy Agency

Patient dose assessment

Page 30: Lecture 05 Image Quality and Patient Dose

IAEA 30

Overview

• To become familiar with the patient dose assessment and dosimetry instrument characteristics.

Page 31: Lecture 05 Image Quality and Patient Dose

IAEAInternational Atomic Energy Agency

Parameters influencing patient exposure

Page 32: Lecture 05 Image Quality and Patient Dose

IAEA 32

Essential parameters influencing patient exposure

Tube voltageTube currentEffective filtration

Exposure time

Field size

Kerma rate[mGy/min]

[min]Kerma[Gy]

[m2]

Area exposureproduct [Gy m2 ]

}}

}

Page 33: Lecture 05 Image Quality and Patient Dose

IAEA 33

Factors in conventional radiography: beam, collimation

• Beam energy• Depending on peak kV and filtration• Regulations require minimum total filtration to absorb

lower energy photons• Added filtration reduces dose• Goal should be use of highest kV resulting in acceptable

image contrast

• Collimation• Area exposed should be limited to area of CLINICAL

interest to lower dose• Additional benefit is less scatter, better contrast

Page 34: Lecture 05 Image Quality and Patient Dose

IAEA 34

Factors in conventional radiography: grid,patient size

• Grids• Reduce the amount of scatter reaching image receptor• But at the cost of increased patient dose• Typically 2-5 times: “Bucky factor” or grid ratio

• Patient size• Thickness, volume irradiated…and dose increases with

patient size• Except for breast (compression), no control• Technique charts with suggested exposure factor for

various examinations and patient thickness helpful to avoid retakes

• Use of AEC exposure

Page 35: Lecture 05 Image Quality and Patient Dose

IAEA 35

Factors affecting dose in fluoroscopy

• Beam energy and filtration• Collimation• Source-to-skin distance

• Inverse square law: maintain max distance from patient

• Patient-to-image intensifier• Minimizing patient-to- I I will lower dose• But slightly decrease image quality by increased scatter

• Image magnification • Geometric and electronic magnification increase dose

• Grid• If small sized patient (les scatter) perhaps without grid

• Beam-on time!

Page 36: Lecture 05 Image Quality and Patient Dose

IAEA 36

Factors affecting dose in CT

• Beam energy and filtration• 120-140 kV; shaped filters

• Collimation or section thickness• Post-patient collimator will reduce slice thickness imaged

but not the irradiated thickness

• Number and spacing of adjacent sections

• Image quality and noise• Like all modalities: dose increase=>noise decreases

Page 37: Lecture 05 Image Quality and Patient Dose

IAEA 37

Factors affecting dose in spiral CT

• Factors for conventional CT also valid

• Scan pitch• Ratio of couch travel in one rotation divided by

slice thickness

• If pitch = 1, doses are comparable to conventional CT

• Dose proportional to 1/pitch

Page 38: Lecture 05 Image Quality and Patient Dose

IAEAInternational Atomic Energy Agency

Patient dosimetry methods

Page 39: Lecture 05 Image Quality and Patient Dose

IAEA 39

Patient dosimetry

• Radiography: entrance surface dose ESD• Output factors

• Dose – area product (DAP)

• Fluoroscopy: Dose - Area Product (DAP)

• CT:• Computed Tomography Dose Index (CTDI)

• Dose – length product (DLP)

Page 40: Lecture 05 Image Quality and Patient Dose

IAEA 40

From ESD to organ and effective dose

• Except for invasive methods, no organ doses can be measured

• The only way in radiography: measure the Entrance Surface Dose (ESD)

• Use mathematical models to estimate internal dose. • Mathematical models based on Monte Carlo

simulations• Dose to the organ tabulated as a fraction of the

entrance dose for different projections• Since filtration, field size and orientation play a role:

long lists of tables (See NRPB R262 and NRPB SR262)

Page 41: Lecture 05 Image Quality and Patient Dose

IAEA 41

From DAP to organ and effective dose

• In fluoroscopy: moving field, measurement of Dose-Area Product (DAP)

• In similar way organ doses calculated by Monte Carlo modelling

• Based on mathematical model• Conversion coefficients estimated as organ doses

per unit dose-area product• Again numerous factors are to be taken into

account such as projection, filtration, …• Once organ doses are obtained, effective dose is

calculated following ICRP103

Page 42: Lecture 05 Image Quality and Patient Dose

IAEAInternational Atomic Energy Agency

Dose measurements: how to measure dose indicators ESD, DAP,CTDI…

Page 43: Lecture 05 Image Quality and Patient Dose

IAEA 43

Measurements of Radiation Output

X Ray tube

Filter

Ion. chamber

Lead slabTable top

SDD

Phantom (PEP)

Page 44: Lecture 05 Image Quality and Patient Dose

IAEA 44

Measurements of Radiation Output

• Operating conditions

• Consistency check

• The output as a function of kVp

• The output as a function of mA

• The output as a function of exposure time

Page 45: Lecture 05 Image Quality and Patient Dose

IAEA 45

Measurement of entrance surface dose

TLD, solid statedosimeter or ion chamber

Includes backscatter (~30%)

Page 46: Lecture 05 Image Quality and Patient Dose

IAEA 46

Dose Area Product (DAP)

Transmission ionization chamber

Page 47: Lecture 05 Image Quality and Patient Dose

IAEA 47

Dose Area Product (DAP)

0.5 m1 m

2 m

Air Kerma:Area:Areaexposure product

40*103 Gy2.5*10-3 m2

100 Gy m2

10*103 Gy10*10-3 m2

100 Gy m2

2.5*103 Gy40*10-3 m2

100 Gy m2

Page 48: Lecture 05 Image Quality and Patient Dose

IAEA 48

Calibration of a Dose Area Product (DAP)

Film cassette

10 cm 10 cm

Ionizationchamber

Page 49: Lecture 05 Image Quality and Patient Dose

IAEA 49

Levels of Dosimetry

• Level 1 - published tables

• Level 2 - Monte Carlo tables using known data

• Level 3 - direct measurement of skin dose

• Level 4 - humanoid phantom measurements with TLD

Page 50: Lecture 05 Image Quality and Patient Dose

IAEA 50

Level 1 - Published Dose Tables

• ICRP, NCRP and various books have tables of “typical” doses for various x-rays

• Tables show organ doses and sometimes effective dose

• The data is usually old, from x-rays made with slower film/screen systems than in current use

• Still useful as a guide

Page 51: Lecture 05 Image Quality and Patient Dose

IAEA 51

Level 2 - Monte Carlo Systems

• Provide organ doses, and effective dose

• Use calculated data but with user entry of various parameters :• HVL, kVp

• skin dose or free-in-air exposure at skin distance

• FSD, field size and position

Page 52: Lecture 05 Image Quality and Patient Dose

IAEA 52

Monte Carlo Systems

• Various computer programs and lookup tables, eg. TISSDOSE, XDOSE, PCXMC

• Most users do not know the actual values for input variables, so often must use assumed values

• HVL, kVp, FSD, field size easy to assume – kVp/mAs and field size not always easy to assume

Page 53: Lecture 05 Image Quality and Patient Dose

IAEA 53

Typical Radiology Doses (Melbourne Data)

Procedure ESD (mGy) Eff. Dose (mSv)

Abdomen (AP) 2.5 0.35

Chest (PA) 0.15 0.023

Lumbar spine AP/Lat 3.2/4.0 0.23/0.11

Pelvis (AP) 1.7 0.29

Thoracic spine (lat.) 2.6 0.097

Mammography (4 views) 4.4 0.44

Page 54: Lecture 05 Image Quality and Patient Dose

IAEA 54

XDose

Page 55: Lecture 05 Image Quality and Patient Dose

IAEA 55

Page 56: Lecture 05 Image Quality and Patient Dose

IAEA 56

PCXMC

Page 57: Lecture 05 Image Quality and Patient Dose

IAEA 57

Page 58: Lecture 05 Image Quality and Patient Dose

IAEA 58

ImPACT CT Dose

Page 59: Lecture 05 Image Quality and Patient Dose

IAEA 59

Page 60: Lecture 05 Image Quality and Patient Dose

IAEA 60

Summary

• In this lesson we learned the factors influencing patient dose, and how to have access to an estimation of the detriment through measurement of entrance dose, dose area product or specific CT dosimetry methods.