L e a f P o r o m e t e r Operator’s Manual Version 6.0
70

Leaf Porometer v6

Mar 09, 2016

Download

Documents

Decagon Devices

Leaf Porometer Manual
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Leaf Porometer v6

L e a f P o r o m e t e r

Operator’s Manual

Version 6.0

Page 2: Leaf Porometer v6

Decagon Devices, Inc.2365 NE Hopkins Court

Pullman, WA 99163Tel: (509) 332-2756Fax: (509) 332-5158www.decagon.com

Trademarks:“Leaf Porometer” is a registered

trademark of Decagon Devices, Inc.

© 2008 Decagon Devices, Inc. All rights reserved.

Page 3: Leaf Porometer v6

Table of Contents1. Introduction . . . . . . . . . . . . . . . . . . .1Customer Service and Tech Support . . . . . . . . . . . . . . . . 1Contact Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Seller’s Liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Repair Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Repair Costs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

2. About the Leaf Porometer . . . . . . 4Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Overview of the Leaf Porometer . . . . . . . . . . . . . . . . . . . 5Components of the Leaf Porometer . . . . . . . . . . . . . . . . . 5Keyboard Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6The External Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Setting the Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Calibration Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Difference Between Manual Mode and Auto Mode . . . . . 9Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

3. The Menus . . . . . . . . . . . . . . . . . . . 11The Main Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Measuring Small Leaves & Needles . . . . . . . . . . . . . . . . . 13Taking a Measurement in Auto Mode . . . . . . . . . . . . . . . 13Making a Measurement in Manual Mode . . . . . . . . . . . . 15Saving Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Averaging Readings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Measurement Troubleshooting . . . . . . . . . . . . . . . . . . . . 18Measurement Precautions. . . . . . . . . . . . . . . . . . . . . . . . . 19

Page 4: Leaf Porometer v6

The Data Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22The Configuration Menu . . . . . . . . . . . . . . . . . . . . . . . . . 25

4. Leaf Porometer Utility . . . . . . . . 30Downloading Your Data . . . . . . . . . . . . . . . . . . . . . . . . . 31How Saved Data Are Organized . . . . . . . . . . . . . . . . . . . 33Erasing Your Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Setting the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . 34Calibration Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34Retrieving Leaf Porometer Information . . . . . . . . . . . . . 35The Menus and Their Functions . . . . . . . . . . . . . . . . . . . 36

5. Leaf Porometer Theory . . . . . . . 42Stomatal Conductance . . . . . . . . . . . . . . . . . . . . . . . . . . . 42How the Leaf Porometer Works . . . . . . . . . . . . . . . . . . . 42

6. Care and Maintenance . . . . . . . .51Changing the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Cleaning the Sensor Block . . . . . . . . . . . . . . . . . . . . . . . . 51General Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

7. Troubleshooting . . . . . . . . . . . . . 55

Further Reading . . . . . . . . . . . . . . . 59

CE Compliance. . . . . . . . . . . . . . . . . 60

Appendix A. . . . . . . . . . . . . . . . . . . .61

Page 5: Leaf Porometer v6

Leaf Porometer Operator’s Manual1. Introduction

1. IntroductionThank you for purchasing Decagon Devices’ Leaf Porome-ter. The Leaf Porometer is a device used to measure sto-matal conductance in leaves. This manual is designed tohelp you understand the functions of your Leaf Porometer,make high-quality stomatal conductance measurements, andhelp you get the most out of your Leaf Porometer.

Customer Service and Tech SupportWhen contacting us, via fax or e-mail, include your LeafPorometer’s serial number (this can be found on the insideof the battery cover), the firmware version you are currentlyusing, your name, address, phone and fax number, and adescription of your problem.

Contact InformationPHONE: United States/Canada (toll free): 1-800-755-2751.Outside of the U.S./Canada: 1-509-332-2756.FAX: 1-509-332-5158E-MAIL: [email protected].

WarrantyThe Leaf Porometer is covered by a standard, one-year war-ranty on parts and labor, and is activated upon the arrival ofthe instrument at your location.

1

Page 6: Leaf Porometer v6

Leaf Porometer Operator’s Manual1. Introduction

Seller’s LiabilitySeller warrants new equipment of its own manufactureagainst defective workmanship and materials for a period ofone year from the date of receipt of equipment (the resultsof ordinary wear and tear, neglect, misuse, accident andexcessive deterioration due to corrosion from any cause arenot considered to be a defect); but Seller’s liability for defec-tive parts shall, in no event, exceed the furnishing ofreplacement parts f.o.b. the factory where originally manu-factured. Material and equipment covered hereby which isnot manufactured by the Seller shall be covered only by thewarranty of its manufacturer. Seller shall not be liable toBuyer for loss, damage, or injuries to persons (includingdeath), or to property or things of whatsoever kind (includ-ing, but not without limitation, loss of anticipated profits),occasioned by or arising out of the installation, operation,use, misuse, nonuse, repair, or replacement of said materialand equipment, or out of the use of any method or processfor which the same may be employed. The use of thisequipment constitute’s Buyer’s acceptance of the terms setforth in this warranty. There are no understandings, repre-sentations, or warranties of any kind, express, implied, stat-utory or otherwise, (including, but not without limitation,the implied warranties of merchantibility and fitness for aparticular purpose), not expressly set forth herein.

Repair InstructionsIf your Leaf Porometer should ever require a repair, callDecagon at (509) 332-2756 or 1-800-755-2751 (UnitedStates and Canada). We will ask you for your address, phonenumber, your Leaf Porometer’s serial number, and your cur-rent firmware version. For non-warranty repairs, we willalso ask for a method of payment.

2

Page 7: Leaf Porometer v6

Leaf Porometer Operator’s Manual1. Introduction

Before shipping your instrument to Decagon, pleasecontact Decagon to obtain a Request MaintenanceAuthorization Number (RMA). This will allow Decagon’srepair staff to keep track of your porometer. Once you haveacquired an RMA, send your Leaf Porometer to Decagon.Please include a document listing the complete shippingaddress, name, and department of the person responsiblefor the instrument, as well as a description of the problem.This will better help our technicians and shipping depart-ment to expedite repair on your Leaf Porometer, and ship itback to you.

Please pack your Leaf Porometer carefully. Ship it back inthe carrying case, preferably inside a cardboard box. Ship to:

Decagon Devices Inc.2365 NE Hopkins Court

Pullman, WA 99163

Repair CostsManufacturer’s defects and instruments under warranty willbe repaired at no cost. For non-warranty repairs, costs forparts, labor, and shipping will be billed to you.

3

Page 8: Leaf Porometer v6

Leaf Porometer Operator’s Manual2. About the Leaf Porometer

2. About the Leaf Porometer

The Leaf Porometer is a battery-operated, menu-drivendevice used to measure stomatal conductance of leaves. Sto-matal conductance is a function of the density, size, anddegree of opening, of stomata, which are pores in plantsthat open to the outside air. The Leaf Porometer measuresstomatal conductance by putting the conductance of a leafin series with two known conductance elements, and com-paring the humidity measurements between them. It candisplay information in three selectable units:

• mmol/m2s (millimoles per meter squared seconds)

• m2s/mol (meters squared seconds per mole)• s/m (seconds per meter)

The first is a unit of conductance, while the other two areunits of resistance. To change between them, use the arrowkeys when taking a reading (see “Making a Measurement,”pp. 16). They can also be altered by using the Units screenunder the Configuration Menu.

SpecificationsOperating Environments: 5 - 40o C; 0 - 90% relativehumidity (non-condensing). NOTE: At temperatures less than10 °C, Automatic mode may experience errors, despite correctionsthat have been made. (See Chapter 3 more information aboutAuto mode.)

4

Page 9: Leaf Porometer v6

Leaf Porometer Operator’s Manual2. About the Leaf Porometer

Accuracy: 10%Sample Chamber Aperture: 6.35mm (0.25 in)Measurement Range: 0 to 1000 mmol/m2s1

Microcontroller Dimensions: 15.8 x 9.5 x 3.3 cm (6.2 x3.75 x 1.3 in)Data Storage: 4095 measurementsData Retrieval: Direct via RS-232Keypad: Six-key, menu-drivenClock: 24-hour, +/- one minute per monthInterface Cable: RS-232 serial cable (included)Power Supply: Four type “AA” batteries (included)

Overview of the Leaf PorometerThe Leaf Porometer features an easy-to-use, six-button,menu-driven interface. There are three primary menus: theMain Menu, Data Menu, and Configuration Menu (See“Menus,” Chapter 3). You can navigate between each ofthese by pressing the Menu button.

Components of the Leaf PorometerThe Leaf Porometer and its components come to you in afoam-padded carrying case. You should find the followingitems inside when your Leaf Porometer arrives:

• Leaf Porometer controller• RS-232 cable• Operator’s manual, including a CD-ROM with the Leaf Porometer Utility• External sensor head• Replacement leaf pad/outer leaf pad (see pp. 8)

5

Page 10: Leaf Porometer v6

Leaf Porometer Operator’s Manual2. About the Leaf Porometer

Keyboard Functions

Here is a brief overview of each key’s function:

Power: This key activates and deactivatesthe Leaf Porometer. (NOTE: The device willautomatically turn off if not used for more than10 minutes.)

6

Page 11: Leaf Porometer v6

Leaf Porometer Operator’s Manual2. About the Leaf Porometer

Menu: This key cycles between the threemenus at the top of the screen. NOTE:Pressing Menu while changing a setting will cancelany operations.

Arrow Keys: These keys navigate withinmenus and sub-menus, and modifynumerical values in sub-menus, as well asselect different units. Holding down anarrow key allows you to scroll between dif-ferent options quickly.

Escape: This key backs out of sub-menusand can be used to cancel certain actions.

Enter: This key is used to make selectionswithin menus and sub-menus, and alsobegins taking measurements.

The External SensorThe Leaf Porometer cannot take measurements without theaid of the external sensor. This sensor is responsible forgathering the information (vapor pressure, humidity, etc.)that the Leaf Porometer uses to calculate stomatal conduc-tance in a leaf. The following illustration shows the externalcomponents of the sensor head.

7

Page 12: Leaf Porometer v6

Leaf Porometer Operator’s Manual2. About the Leaf Porometer

Setting the TimeYou should make sure that the system time on the porome-ter is correctly set. This will make sure the timestamps ondownloaded readings are accurate (see “How Saved Data isOrganized,” Chapter 4). You can do this manually in theporometer (see “Time” in Chapter 3), or by using the LeafPorometer Utility (see “Calibration Number” in Chapter 4).

Calibration NumberEach sensor head has its own unique calibration number, afive-digit number, found on the sensor cable and calibrationcertificate. This number must be correctly entered for eachsensor used in order for it to take accurate readings. Youcan set this manually in the porometer (see “Cal Num (Cali-bration Number)” in Chapter 3), or by using the LeafPorometer Utility (see “Date/Time Formats for DataFiles,” Chapter 4).

8

Page 13: Leaf Porometer v6

Leaf Porometer Operator’s Manual2. About the Leaf Porometer

Difference Between Manual Modeand Auto ModeWhen the porometer is in Manual Mode (see pg 15 forinstructions on how to configure the porometer to ManualMode), the stomatal conductance is measured directly fromthe theory presented in Chapter 5 and displayed continu-ously. The real time conductance measurement can be savedat any time during a Manual Mode measurement. For thistheory to accurately predict the stomatal conductance usingthe steady state diffusion technique, true steady state condi-tions must exist in the diffusion path. Our tests have shownthat it can take 10 minutes or more for the steady state con-ditions to develop in the diffusion tube, which is muchlonger than anyone wants to wait for a stomatal conduc-tance measurement. Of greater importance, leaving the leafclip on the leaf surface for more than a few minutes cancause a stomatal conductance to change in reaction to thepresence of the clip.

Auto Mode makes an accurate stomatal conductance mea-surement in 30 seconds. To do this, the porometer takes thefirst 30 seconds of stomatal conductance data and applies a“look ahead” algorithm to predict the final stomatal con-ductance reading that would be achieved if unlimited timewere allowed for true steady state conditions to occur. AutoMode has several advantages over Manual Mode and otherporometers. First, the 30 second Auto Mode measurementduration ensures that stomatal conductance doesn’t changedue to the presence of the leaf clip. Second, Auto Modetakes only 30 seconds per measurement, ensuring that yourmeasurements are completed in a timely manner. Finally,Auto Mode sets a definite endpoint for the measurement,meaning that the user doesn’t have to pick an arbitrary end-

9

Page 14: Leaf Porometer v6

Leaf Porometer Operator’s Manual2. About the Leaf Porometer

point, as is common with other diffusion porometers andwith Manual Mode measurements.

For the porometer to “detect” the presence of a transpiringleaf and begin the Auto Mode measurement, the measuredstomatal conductance must reach 5 mmol m-2s-1. So, if youare making measurements on leaves with extremely lowconductances, it may be necessary to make the measure-ments in Manual Mode.

PrecautionsBefore using the porometer, remember:

• Avoid leaves that are wet or heavy with dew; nevertake readings in the rain.

• Do not touch the Teflon disk, or get your fingersnear it, during a reading. This will cause inaccuratereadings.

• If you blow into the sensor, it will take a few min-utes for the stomatal conductance reading to returnto normal.

• Clean the sensor periodically to keep it free of dirtand pollen that can build up during use and affectreadings. (See “Cleaning the Sensor Block,” inChapter 7.)

• Avoid chemical fumes. Fumes can be extremelyharmful to the sensor (i.e. alcohol, gasoline, foam).

10

Page 15: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

3. The MenusThe Leaf Porometer features three main menus, designedfor easy navigability and use.

The Main MenuThe Leaf Porometer displays the Main Menu when firstpowered on. This menu is used for taking measurements.

At the top of the screen is the stomatal conductance num-ber at left, and the currently selected unit of measurement atright. Below are two columns of numbers. The left is thesensor closest to the leaf, and the right is the sensor farthestfrom the leaf. Each column has the temperature of that sen-sor, and its percent humidity. The Leaf Porometer usesthese to calculate the stomatal conductance value (see“How the Leaf Porometer Works” in Chapter 5).

Making a MeasurementThe following sections discuss making a measurement inboth automatic and manual modes.

11

Page 16: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

Before beginning a reading, make sure that the Leaf Porom-eter is turned on, and that the external sensor is clean andproperly connected.

NOTE: Make sure you are holding the sensor in the correct manner.Holding the external sensor correctly is extremely important. If you donot, you can expose the Teflon disk to extra humidity and cause inac-curate readings. The pictures below illustrate the correct way to hold theexternal sensor:

Correct Incorrect

Once you are holding the sensor properly, place it on theleaf as shown below.

NOTE: It is important to consider if you will be measuring adaxialconductance (on the top of the leaf) or abaxial (the bottom). Since thestomates of most leaves are on the bottom of the leaf, most measure-ments will be abaxial:

12

Page 17: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

Abaxial (Bottom) Adaxial (Top)

Measuring Small Leaves & NeedlesWhen measuring small needles or leaves, they should beinserted into the sensor as shown below. This is because sin-gle needles or small leaves (including blades of grass) maynot adequately cover the apature of the sensor.

Taking a Measurement in AutoModeThere are two modes in which to make measurements,Automatic and Manual. The following illustrates how tomake a measurement in Auto mode:

13

Page 18: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

Caution: Before using, allow the sensor to equilibrateto the ambient temperature.

1. Select Auto mode. (Please see the “Manual Mode” sec-tion later in this Chapter for instructions on how toselect an Auto mode from the Configuration Menu.)

2. Return to the Main Menu. You will be prompted topress Enter to begin a reading.

3. After pressing Enter, the following screen will be displayed:

4. The reading will begin once the sensor head is placedon a leaf, and a conductance value greater than ambientis detected. A progress bar will also indicate the readingis underway.

If the screen below appears when you press Enter, open thesensor and gently wave it in the air. You should not blowinto the sensor, as this will throw off readings for a fewminutes.

14

Page 19: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

5. The screen will resemble the following. NOTE: In Automode, a timer will appear in the upper right hand corner of thescreen that indicates how many seconds remain.

6. Once the time limit has expired, the porometer will dis-play the final conductance. You may save or discardyour data. See “Saving Data” on the next page forinstructions.

Making a Measurement in ManualMode

Caution: Before using, allow the sensor head to equili-brate to the ambient temperature.

1. Select Manual mode from the Configuration Menu.

15

Page 20: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

2. Return to the Main menu and press Enter.

3. Place the sensor on a leaf. Unlike Automatic mode,there is no time limit for taking readings, and theelapsed time is displayed directly below the system time.

4. You may save your data at any point during the readingby pressing Enter. See below for instructions on how tosave your data.

Saving DataIn Automatic mode, you can save data as soon as the timerhas expired; in Manual mode, you save your data withoutinterrupting your reading. To save your data in either mode,1. The following screen will appear (you have to press

Enter in Manual mode to access it):

16

Page 21: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

2. You have three options:1) Select “Save” and press Enter to save your infor-

mation as-is. 2) Select “Discard” and press Enter to return to the

main screen and start over with a new reading3) Select “Annotate” and press Enter to give your

data a file name.

3. How to annotate a reading: Selecting “Annotate” willcause the following screen to appear:

You will see a series of lines, with a set of arrows aboveand below the first. If you have annotated a pre-viousreading, the old annotation will remain to be changed ordeleted. Use the arrow keys to select a letter or numberand press Enter to move to the next character.

17

Page 22: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

4. Repeat this until you reach the last character. If you wishto stop before the end, you will still have to press Enteruntil you reach the end of the row. (NOTE: HoldingEnter or Escape will allow you to scroll left and right through theSample ID quickly.)

Averaging ReadingsA leaf ’s stomatal conductance depends on the amount ofshade and sunlight it receives, the age of the leaf, and itsposition on the plant. To obtain a plant’s stomatal conduc-tance average, you must take a variety of readings from dif-ferent positions on the plant. The average is proportional tothe type of readings taken. For example, if you know yourplant receives 70% sunlight, you might consider takingseven readings from the sunny part, and three from the partin the shade.

You can also obtain the average readings for a field ofplants. For example, if you had a field of corn, you mightdecide to take a reading from the third leaf from the top oneach stalk of corn on that particular plot. These readingstogether would give you the average for that field.

Measurement TroubleshootingIf the following screen appears before starting a reading:

18

Page 23: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

Remove the sensor from the leaf and wait until the bar onthe right drops below the AMB> marker (this may take awhile), at which time the “Place sensor on leaf ” screen willreappear. Or, to reset more quickly, hold the sensor firmly,open the sensor head, and wave in the air lightly to equalizethe humidity. Do not blow into the sensor; this can causeinaccurate readings when the measurement begins. Oncethe ambient humidity has been reset, you will be promptedto place the sensor head on the leaf again.

If the stomatal conductance reaches a value beyond the rec-ommended range of the instrument(0 to 1000 mmol/m2s),an “approximate” symbol will appear to the left of the read-ing. If the stomatal conductance reaches a value greater thanor equal to 6000 mmol/m2s, an “infinity” symbol is shownin place of the measurement. Most often, the excessivelyhigh values are caused by water on the leaf surface. If youconsistently are getting high readings, check your leaf sur-face for the presence of water.

Measurement Precautions

General Precautions• Before using the sensor, allow it to equilibrate to the

ambient temperature. This may take 10 minutes orlonger to complete; if the temperature of the porome-ter is very different from the sampling environment,e.g. from a cool air conditioned room to a hot out-door environment.

• The first reading in Auto mode after the sensor hasbeen removed from a leaf for more than five minutes

19

Page 24: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

will always read low. Discard this reading and beginrecording with the second.

• Avoid leaves that are wet or heavy with dew; nevertake readings in the rain.

• Do not cover the white Teflon disk with your finger,or place your fingers near it, when taking a measure-ment; this will cause inaccurate readings for severalminutes.

• If you blow into the sensor head, it will take a fewminutes for the stomatal conductance reading toreturn to normal.

• Clean the sensor periodically to keep it free of dirtand pollen that can build up during use and affectreadings.

Chemical EffectsChemical vapors can interfere with the polymer layers usedto form the humidity sensors. These fumes can diffusethrough the sensor and throw off the sensor’s accuracy. It isa good idea, therefore, to never bring the sensor into con-tact with any sort of chemical vapor (i.e. glue, alcohol, gaso-line, foam).

Environmental EffectsWhen making a measurement, the following effects must beaccounted for.

Light. The amount of CO2 required for a plant changesdepending on the amount of sunlight available to it. Inbroad daylight, more CO2 is required, so the stomataremain open. At dusk, the stomata close, and in darknessthey are often closed completely. Readings taken on over-cast days, for example, generally have lower conductances

20

Page 25: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

than those taken on sunny ones. Additionally, leaves inshadow or in shade usually have lower conductances thanthose in greater light. Therefore, make sure to consider thelighting conditions you will be measuring in.

Temperature. A plant’s stomatal conductance is alsoaffected by temperature. For most plants, on a very hot or avery cold day, photosynthesis will slow down or stop, andthe stomata will narrow or will not open, to prevent internalwater loss. Therefore, be sure to consider the temperaturebefore conducting measurements.

Humidity. A factor that will cause the stomatal conduc-tance to change is the humidity at the leaf ’s surface. As theoutside humidity increases, the stomata will open. Porome-ters that dry the air around the stomata decrease conduc-tance; likewise, those that allow surface humidity to increasedo the opposite.

Carbon Dioxide (CO2). One purpose of the stomata is toregulate the leaf ’s usage of CO2, which is important to con-trolling the balance of humidity. An increase in CO2 willcause stomata to close, while a decrease will cause them toopen. Thus, it is important to consider the CO2 levels ofyour measuring environment. You, the Leaf Porometeroperator, can adversely affect readings by exhaling CO2directly on the plant you are measuring. This is particularimportant in growth chambers or greenhouse, where CO2from an operator’s breath can double or triple ambient lev-els.

21

Page 26: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

The Data Menu

In this menu, you can view any saved data, download datato a computer, and erase all data stored in the Porometer’smemory. This menu features three sub-menus.

ViewHere you can view measurement data stored on yourporometer. To do this:

1. Press Enter to select the view sub-menu.

2. For each measurement, you will see the time acquired,stomatal conductance/resistance value (depends on whichunit you have set as your default) and its annotation (if thatmeasurement was saved with an annotation).

22

Page 27: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

DownloadThis menu option sends the data saved in the porometer tothe terminal software running on your computer.

NOTE: Your Leaf Porometer comes with the Leaf Porometer Utilitysoftware for Microsoft Windows. The Utility makes downloading datato your computer very easy. In general, you should use the Leaf Porom-eter Utility to download data instead of this menu option on the Poro-meter. See Chapter 4 for more information.

If you are unable to use the Leaf Porometer Utility, you canuse terminal software to transfer the data from your porom-eter to your computer. The following steps should apply tomost terminal software programs:

1. Configure your terminal software with the following settings:• 9600 baud• 8 data bits, 1 stop bit, no parity• No software/hardware flow control• Append line feeds to incoming line ends • Echo typed characters locally.

2. Connect your Leaf Porometer to an available serial port on your computer using the included RS-232 serial cable.

3. Set your terminal software to capture received dataif you want to save the data.

4. Select “Download” from the Data Menu of the Leaf Porometer.

23

Page 28: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

EraseThis will erase all your stored data records in your LeafPorometer memory.

WARNING! Once this feature is activated, all data willbe permanently deleted from the Poromter, and cannotbe recovered!

To erase data:

1. Press Enter to select the Erase sub-menu. The follow-ing screen will appear.

2. Press Enter again to erase the memory, or ESC if you decide not to erase the data. “Erasing...” will be dis-played as data is being removed, which usually takes about 6 seconds.

3. When this is complete, you will be returned to the Data Menu.

24

Page 29: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

The Configuration Menu

This menu allows you to alter system settings, such as the current time, date, and operating mode, select theappropriate units for your readings, and view informationabout your Leaf Porometer. There is also a battery icon tothe right, indicating how much battery power is currentlyavailable. This menu features seven sub-menus:

DateTo change the current date,

1. Press Enter to select the Date display.2. The current date will appear in the center of the

screen, in day/month/year format. A pair of arrows will be present (above and below) the first number.

25

Page 30: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

3. Use the arrow keys to change this number. (Holding down the up/down arrow will allow you to scroll quickly between values.)

4. Press Enter to move to the next value, and repeat step #3. If you need to return to the previous number,

press Escape.5. When you have finished changing the last number,

press Enter and you will be returned to the Configura-tion Menu.

TimeTo change the current time,

1. Press Enter to select the Time display.2. The current time will appear in the center of the screen, in 24-hour format. A pair of arrows will be present (one above and below) the first number.

26

Page 31: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

3. Use the arrow keys to change this number. (Holding down the up/down arrow will allow you to scroll quickly between values.)

4. Press Enter to move to the next number, and repeat step #3. If you need to return to the previous number, press Escape.

5. When you have finished changing the last number, press Enter and you will be returned to the Configura-tion Menu.

ModeThe Mode Menu allows you to choose which type of mea-surement mode to use in the Main menu. Press Enter toselect a mode. See the end of Chapter 2 for the descriptionsof Manual and Auto Modes. Here is a description of how touse each:

• Auto mode. Takes readings automatically for 30 sec-onds. You must press Enter before placing the sensoron the leaf for Auto mode to work properly. In this mode, you are unable to save your informa-tion mid-reading. You must wait until the timer runsout after 30 seconds, then you may save your read-ing. Also, if initial conductance is less than 5 mmol/m2 s, you will see a screen prompting you to placethe sensor on a leaf. Conversely, if the initial con-ductance is too high, a warning screen will appear.

• Manual mode. Continuously measures conductanceand updates the screen. This mode is well-suited forlow conductances. The user can decide when equi-librium has been reached. Pressing Enter will giveyou the option to save your data at any point duringthe reading.

27

Page 32: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

UnitsData in the Leaf Porometer can be displayed in three units.These can be accessed on the Main Screen, as well as underthe Configuration Menu. When in the Configuration Menu,select “Units” and press Enter to select a unit.

About ScreenThis menu displays your Leaf Porometer’s serial numberand firmware version.

ContrastThis menu allows you to alter the screen contrast settings.To access it, use the Down arrow key to scroll to it andpress Enter. The following screen will appear:

28

Page 33: Leaf Porometer v6

Leaf Porometer Operator’s Manual3. The Menus

Use the arrow keys to alter the contrast level. Press Enter tosave or Escape to cancel adjustments.

Cal Num (Calibration Number)This screen will display the calibration number currentlystored in the porometer. This is a five-digit number that isfound on a sticker attached to the sensor cable and calibra-tion certificate. This can be entered manually in the porom-eter, or by using the Leaf Porometer Utility.

Press the Enter key to select the Cal Num menu. Use thearrow keys to choose a number, and press Enter to move tothe next space.

When you have selected the number, press Enter to returnto the Configuration menu.

29

Page 34: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

4. Leaf Porometer Utility

Leaf Porometer Utility main screen

The Leaf Porometer Utility is a program designed specifi-cally for interfacing with the Leaf Porometer. Use this pro-gram to download measurement data to your computer,erase the Leaf Porometer’s data, set the date and time, andsee information about your Leaf Porometer.

SYSTEM REQUIREMENTS: To use the Leaf Porome-ter Utility, you must meet the following minimum systemrequirements:

• Microsoft Windows 98 or NT 4 (SP 5) or better

30

Page 35: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

• Intel Pentium Pro or better processor• One available serial port or one available USB port(most models of USB-to-serial adapters supported)• Microsoft Excel 97 or better (for saving data as .xls files)•

You can install the Leaf Porometer Utility using theincluded CD-ROM (found in the inside cover of the LeafPorometer Operator’s Manual). You can also download andinstall the latest version of the Utility from http://www.decagon.com/home/downloads.php .

Downloading Your DataThis will transfer all saved measurement data on the LeafPorometer to your computer. Do the following:

1. Make sure the RS-232 cable is connected to the LeafPorometer, and to a serial communication (COM) porton your computer.

2. Open the Leaf Porometer Utility.

3. Choose the appropriate COM port from the “Use com-puter communication port” control on the main screen.

4. Click the “Download” button in the lower right-handcorner, or “Download Data” in the File Menu.

If you encounter an error message after clicking this button, please refer to the Troubleshooting section for instructions.

31

Page 36: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

5. The following screen will then appear:

Save Leaf Porometer data dialog

Name your data file, select where it will be saved to, and in what format. You may choose between the fol-lowing formats: • Microsoft Excel Workbook (*.xls);• Text (Tab delimited) (*.txt);• CSV (Comma delimited) (*.csv);• Raw data (Comma delimited) (*.txt)

6. Click “Save” to download your data to the specifiedlocation. The progress bar shows the status of thedownload process.

NOTE: You can cancel a download in progress using the cancel but-ton. If you cancel, no downloaded data are saved.

32

Page 37: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

How Saved Data Are OrganizedData are organized in the same way regardless of what for-mat it was saved in. The information will be divided intofour columns:

1. Measurement Time2. Stomatal conductance/resistance3. Temperature in °C or °F4. Sample ID (if saved with measurement)

NOTE: The utility formats measurement dates according to the Win-dows Local settings found in the Control Panel under “Regional andLanguage Options” (“Regional Settings” in Windows 98). You mayoverride this and set the format to day/month/year by going to theEdit Menu, click Preferences, Units tab, and by clicking “Use tem-plate” under Date/Time Format for Data Files.

Erasing Your DataThis will erase all data stored on your Leaf Porometer.

WARNING! Once this feature is activated, all data willbe permanently deleted from the Leaf Porometer, andcannot be recovered!

To erase your data, do the following:

1. Make sure that the RS-232 cable is connected to theLeaf Porometer, and to a COM port on your computer.

2. Choose the appropriate COM port from “Use com-puter communication port” control on the main screen.

33

Page 38: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

3. Click “Erase,” in the lower left-hand corner, or “EraseData...” in the File Menu. If you encounter an errormessage after clicking this button, please refer tothe Troubleshooting section for instructions.

4. The progress bar shows the status of the erase process.

Setting the Date and TimeYou can set the Leaf Porometer’s date and time by selecting“Set Porometer Date/Time” from the Actions Menu. Thiswill automatically synchronize the Leaf Porometer’s dateand time to your computer’s date and time.

Calibration Number

Sensor Calibration Number screen

You can use the Leaf Porometer Utility to enter the externalsensor’s calibration number. To do this, go to Actions > SetSensor Cal Num... . The above screen will appear. Enter thecurrent sensor’s calibration number (found on the sensorcable) then click OK. All five numbers must be entered,even if the first number is a zero! If the number is not cor-rect, an error message will be displayed.

34

Page 39: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

Retrieving Leaf Porometer Information Choose “View Porometer Information...” from the ActionsMenu to see information about your Leaf Poro-meter:

Leaf Porometer Information screen

If you encounter an error message after clicking thisbutton, please refer to the Troubleshooting section forinstructions.

This is a good way to find the serial number, firmware ver-sion, battery status, and number of stored measurementsfor your Leaf Porometer.

35

Page 40: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

The Menus and Their FunctionsThe four menus across the top of the screen allow you tochange program settings, and settings on the Leaf Poro-meter. Below is a brief overview of the options each menucontains.

FileDownload Data. Saves measurement data stored in your LeafPorometer as a data file on your computer (see instructionsabove).

Erase Data. Erases all data in your Leaf Porometer (seeinstructions above).

Exit. Quits the program.

EditPreferences... This consists of three sections:

36

Page 41: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

1. Data File

Preferences - Units screen

Measurement Units: If you select “Get units from LeafPorometer settings,” this will save all downloaded mea-surement data in the units used on the Leaf Porometer.You can also choose to override the set-tings in the LeafPorometer and save your measurement data in one ofthree units:• mmol/m2 s - stomatal conductance(millimoles per meter squared seconds);• m2 s/mol - stomatal resistance

(meters squared seconds per mole);• s/m - stomatal resistance

(seconds per meter).

37

Page 42: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

Temperature Units: This allows you to choose what unit oftemperature data files will be saved with, °C (Celsius) or °F(Fahrenheit).

Column Headers: This option sets column headers (i.e“Date/Time,” “Temperature”) for saved data files. Thisfeature is on by default. It does not apply to raw datafiles, which do not have column headers.

Date/Time Format: Here you can set the format the dateand time will be displayed in saved data files. By default,the program will use Windows Regional and LanguageOptions (accessible in the Control Panel in Windows),the system settings on your computer.

2.Communications

Preferences - Communications screen

Command Retries: If you encounter difficulty communi-cating with your Leaf Porometer, you can set the num-

38

Page 43: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

ber of times the computer should automatically try re-sending communications commands (up to 10).

Maximum Baud Rate: You can set the maximum baudrate for talking to your Leaf Porometer. Choose a lowerbaud rate if you are not getting reliable communicationsfrom your device.

Force Find all Communication Ports: This will detect allCOM ports on your computer, and should be used ifyour serial-to-USB adapter does not appear in the drop-down menu.

39

Page 44: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

3. Application

Preferences - Application screen

Automatic Internet Version Check: When selected, the util-ity will automatically check for utility updates for theLeaf Porometer Utility when you are connected to theInternet. You can also check for these manually byselecting “Check for Updates...” from the Help Menu.

Automatic Clock Synchronization: Automatically sets theLeaf Porometer’s time to your computer’s time, whenyou are connected to it. (See instructions above for set-ting this manually.)

Error Log File: This adds more troubleshooting mes-sages to the error log, which can be sent to Decagon.

40

Page 45: Leaf Porometer v6

Leaf Porometer Operator’s Manual4. The Leaf Porometer Utility

ActionsSet Porometer Date/Time. See the section on “Setting Dateand Time” above.

Set Sensor Cal Num... Here you can set the sensor head’s cali-bration number. See page 35. View Porometer Information... This shows the serial number,firmware version, the number of stored readings in the LeafPorometer, the firmware status, battery status, current dateand time, and the sensor calibration number.

HelpHelp... Accesses the Leaf Porometer Utility help file, whichprovides detailed information on how to use the program.

Send Feedback to Decagon... Provides the user with a way tosubmit feedback to Decagon.

Check for Utility Updates... Goes online and checks for utilityupdates.

Check for Firmware Updates... Goes online and checks forfirmware updates for the Leaf Porometer controller.

About. Displays the current program version and Decagoncontact information.

41

Page 46: Leaf Porometer v6

Leaf Porometer Operator’s Manual5. Leaf Porometer Theory

5. Leaf Porometer Theory

Stomatal ConductanceBy definition, stomatal conductance is the measure of therate of passage of carbon dioxide (CO2) or water vaporthrough the stomata of a leaf. Stomata are small pores onthe top and bottom of a leaf that are responsible for takingin and expelling CO2 and moisture from and to the outsideair. The Leaf Porometer measures the rate at which thishappens.

How the Leaf Porometer WorksThe Leaf Porometer measures the stomatal conductance ofleaves by putting the conductance of the leaf in series withtwo known conductance elements. By measuring thehumidity difference across one of the known conductanceelements, the water vapor flux is known. The conductanceof the leaf can be calculated from these variables.

We know the humidity at three places: inside the leaf, and atboth of the humidity sensors. The Leaf Porometereffectively calculates the resistance between the inside andoutside of the leaf: the stomatal conductance. Resistance ismeasured between the leaf and the first humidity sensor,and the first and second sensors. The following diagramschematically illustrates this:

42

Page 47: Leaf Porometer v6

Leaf Porometer Operator’s Manual5. Leaf Porometer Theory

The parameters listed above represent the following:Cleaf The mole fraction of vapor inside the leafC1 The mole fraction of vapor at node 1C2 The mole fraction of vapor at node 2gs Stomatal conductance of the leaf surfacegd1 Vapor conductance of the diffusion path between

leaf surface and node 1gd2 Vapor conductance of the diffusion path between

node 1 and node 2d1 Distance between the leaf surface and the first

humidity sensord2 Distance between the two humidity sensors

43

Page 48: Leaf Porometer v6

Leaf Porometer Operator’s Manual5. Leaf Porometer Theory

The variable of interest is gs or the stomatal conductance.First the vapor flux along the diffusion path will bedetermined using the relative humidity difference betweennodes 1 and 2 as follows:

(1)

Where the C's are related to relative humidity by

(2)

where hr is the relative humidity, es(Ta) is the saturated vapor pressure at air temperature, and Patm is atmospheric pressure. Saturated vapor pressure is calculated by the Tetens formula with appropriate coefficients for water vapor:

(3)

NOTE: T must be in degrees Celsius.

Next we must determine the value of gd2 by the equation

(4)

)( 212 CCgF dvapor −=

atm

asri p

TehC )(=

+=

97.240502.17exp611.0)(

TTTe as

22

ˆdD

g vapord

ρ=

44

Page 49: Leaf Porometer v6

Leaf Porometer Operator’s Manual5. Leaf Porometer Theory

where is the molar density of air and Dvapor is the diffusivity of water vapor. Both of these quantities are temperature and pressure dependent, however when multiplied together as

in equation (3) some of this dependency drops out:

Using these C and g values we can now solve equation (1)for the flux:

Now that the vapor flux has been determined, the stomatalconductance, gs , can be found. This requires someassumptions. First, we assume that the relative humiditywithin the leaf tissue is 1.0, so by equation

(9)

75.05

25

75.1

15.273)1012.2)(6.44(ˆ

)/(1012.2)3.101,15.273(

15.2733.101)3.101,15.273(),(

15.2733.101

6.44ˆ

=

=

=

=

TxD

Then

smxDIf

TP

DPTD

TP

vapor

ref

arefavapor

a

ρ

ρ

(5)

(6)

(7)

[ ])()(1ˆ2211

2asrasr

atm

vaporvapor TehTeh

PdD

F −

=

ρ

atm

asleaf p

TeC )(=

45

Page 50: Leaf Porometer v6

Leaf Porometer Operator’s Manual5. Leaf Porometer Theory

Next, we assume that all conductance values are in series sothat the flux is constant between any two nodes. We alsoassume that the temperature of the leaf is equal to thetemperature of the first humidity sensor. (The sensorblock head has been constructed from aluminum toeliminate the temperature difference.) This means that wecan write equation (1) for node 1 and the leaf node and thenset it equal to equation (8):

(10)

(11)

(12)

Solving for gs+d1:

(13)

)( 11 CCgF leafdsvapor −= +

[ ])1)((1111 ras

atmdsvapor hTeP

gF −

= +

[ ] [ ])()(ˆ1)1)(( 22112

111

asrasratm

rasatm

ds TehTehdD

PhTe

Pg

=−+ ρ

[ ]

)1)((

)()(ˆ

11

22112

1ras

asrasr

ds hTe

TehTehdD

g−

=+

ρ

(8)

46

Page 51: Leaf Porometer v6

Leaf Porometer Operator’s Manual5. Leaf Porometer Theory

Using the rule for series combination of conductance, wecan solve for gs:

(14)

Hence,

(15)

so:

(16) Therefore, gs is a function of the distances betweenhumidity sensors, temperature, and the two relativehumidity readings.

When the conductance is small, the humidities are nearlythe same, and the denominator of the denominator of eq. 6goes to zero, causing problems. Multiplying top and bottomby the denominator gives

11

111

ddss ggg−=

+

[ ] Dd

TehTehDdhTe

g asrasr

ras

s ρρ ˆ)()(ˆ)1)((1 1

2211

211 −−−

=

[ ]1

2211

211

)()()1)((

ˆ

dTehTehdhTe

Dg

asrasr

ras

vapors

−−−

[ ] 12211211

2211

)]()([)1)(()]()([ˆ

dTehTehdhTeTehTehD

gasrasrras

asrasrvapors −−−

−=

ρ

47

Page 52: Leaf Porometer v6

Leaf Porometer Operator’s Manual5. Leaf Porometer Theory

(17)NOTE: The resulting gs is in units of mol/m2s.

From the Leaf Porometer, the two distances are:

d1 = 3.35 mmd2 = 11.43 mm

Cooling of the evaporating surface. The equationsdeveloped so far assume that the evaporating surface

temperature is equal to the sensor 1 temperature. A morerealistic assumption is that it cools in proportion to the fluxof water vapor, which is proportional to the vapor pressuredifference between 1 and 2. The effect of this cooling is to

reduce the apparent humidity at the evaporating surface.The actual vapor pressure at the evaporating surface is

(18) Where s is the slope of the saturation vapor concentrationfunction. The temperature difference can be computedfrom an energy balance for the evaporating surface:

(19)

Where is the latent heat of vaporization, so then

)()( 11 aoaatmaso TTsPTee −−=

)()( 2121 CCgFTTgcH dvaporaoaHp −==−= λλ

48

Page 53: Leaf Porometer v6

Leaf Porometer Operator’s Manual5. Leaf Porometer Theory

(20)

Where

Is an apparent psychrometer constant and cp is the specificheat of air.

And

(21)

Substituting this into eq. 18 gives

(22)

The vapor pressure difference between the leaf surface andthe first humidity sensor then becomes

(23)

atmHp

daoa P

CCCCgcgTT

**)( 12121

21 γγ

λ ∆=

−=−=−

2

*d

Hp

ggc

λγ =

)]()([ 22111 asrasr TehTeh −=∆

11 *)( ∆−=

γsTee aso

∆−=− 1

1

111 )(*1)( r

asaso h

TesTeee

γ

49

Page 54: Leaf Porometer v6

Leaf Porometer Operator’s Manual5. Leaf Porometer Theory

The equation therefore becomes

(24)

Where q is

(25)

Both the saturation vapor pressure and its slope have strongtemperature dependence, but their ratio is a weak linearfunction of temperature, so the final form of the equationfor q gives a value which is fairly constant with temperature(though the temperature correction is made in code). Atroom temperature a value of q = 0.1 appears to work well.

112111

1

)1)((ˆ

ddhqTeD

gras

vapors ∆−−∆−

∆=

ρ

*103.4071.0

)(*)(*

4

11 γγγ atmasatm

atm

as PTx

TePsP

Tesq

−−===

50

Page 55: Leaf Porometer v6

Leaf Porometer Operator’s Manual7. Care and Maintenance

6. Care and Maintenance

Changing the BatteriesAlthough the Leaf Porometer has been designed to providean excellent battery lifespan, the batteries will eventuallyrequire changing. When this happens, a low-battery indica-tor will appear in the upper left hand corner of the screen.

1. Carefully turn over the Leaf Porometer and locate the battery cover.

2. Place your thumb on the grooves and push upward to loosen the cover.

3. Remove the old batteries and insert new ones. Be sure to orient the fresh batteries properly.

4. Use the Leaf Porometer or the Leaf Porometer Utility to update the time and date.

Cleaning the Sensor BlockIt is necessary to clean the Sensor Block periodically toremove dust and pollen that can build up during routineuse. Follow these instructions to clean the Sensor block.

λ

51

Page 56: Leaf Porometer v6

Leaf Porometer Operator’s Manual7. Care and Maintenance

1. Carefully remove the Teflon disk with a sharp objectsuch as pair of tweezers or point of a knife. The TeflonDisk should be handled around the edges as much aspossible, and always kept free of dust and moisture.Should the disk get wet, allow it to air dry for 24 hoursbefore continuing use of the instrument.

2. Hold the sensor head open and use a dry swab to cleanout the diffusion path. Swabs are included with your kit.Do not use a damp swab, alcohol or other organic liq-uids to clean the diffusion path! Direct contact withwater on the sensors as well as chemical vapors cancause damage!

3. Reinsert the Teflon Disk. Remember to handle aroundedges as much as possible. Use something dry to pressthe filter completely into its counter bore. If the TeflonDisk becomes damaged discard and replace. Fivereplacement disks are included, if additional disks areneed please contact Decagon Devices.

52

Page 57: Leaf Porometer v6

Leaf Porometer Operator’s Manual7. Care and Maintenance

Maintaining the Seals & Teflon DiskWhen you clean your sensor head, be sure to check the con-dition of the seals, and make sure that they are making fullcontact with the leaf. If the seals require replacement, aspare set has been included in your Leaf Porometer ship-ment. Also, the Teflon disk should be replaced if it be-comes excessively dirty. If you need more replacements ofeither component, contact Decagon.

Replacing the Seal. In the event a seal falls off, you cansimply replace it by peeling off the adhesive paper on theback and replacing the old seal. If the seal is very loose, it isbest to peel it off and replace it in the same manner. Whenyou replace a seal, look to see if the area you will be applyinghas dirt or soil particles, as these can prevent the adhesivefrom applying fully. You might try wiping the area with aslightly damp cloth before applying the new seal.

Replacing the Teflon filter disk. A spare set of Teflon fil-ter disks have also been included. To replace the disk, use asharp object and carefully remove the disk. Then, handlingthe disk around the edges (using tweezers is advised), insertthe spare.

Calibration. Yearly calibration of the Leaf Porometer ishighly recommended.

53

Page 58: Leaf Porometer v6

Leaf Porometer Operator’s Manual7. Care and Maintenance

General PrecautionsWhile the Leaf Porometer has been designed for quality anddurability, it should be kept clean and the batteries alwaysfresh to ensure proper working condition. Here are a fewthings to remember when using this device:

• Allow the sensor head to equilibrate to the ambient temperature before using.

• Keep the external sensor clean, as the accuracy of readings can decline if there is excess humidity or dirt present.

• Do not touch the Teflon disk. Touching it will offset your readings. Let it dry for 24 hours or until the ambient humidity reading returns to normal.

• While this device is water-resistant, do not use it in rain, and do not use it on leaves covered with dew or water droplets.

54

Page 59: Leaf Porometer v6

Leaf Porometer Operator’s Manual6. Troubleshooting the Leaf Porometer

7. Troubleshooting the Leaf Porometer

The following descriptions should help you if you encoun-ter any difficulty using the Leaf Porometer Utility.

PROBLEM: The Leaf Porometer Utility tells me thecommunication port I want to use is in use by anotherapplication, but I don't think any other programs arerunning.

SOLUTION: Some PDA synchronization software moni-tor serial communication ports. Disable Microsoft's Active-Sync or Palm's HotSync system software while using theserial port with the Leaf Porometer Utility.

PROBLEM: My USB to Serial adapter is not showingin the communication port picker.

SOLUTION: Enable “Force find all CommunicationPorts” in the Preferences Menu by going to the Edit Menu,clicking Preferences, then the Communication tab, andenabling the check box at the bottom of the screen.Enabling this option may find other serial ports that are notavailable for use by the Leaf Porometer Utility (for example,modems installed in your computer).

PROBLEM: Downloading data stops in the middlewith an error message saying the Utility lost connec-tion with the Leaf Porometer.

55

Page 60: Leaf Porometer v6

Leaf Porometer Operator’s Manual6. Troubleshooting the Leaf Porometer

SOLUTION: A noisy serial connection can disrupt theconnection between the Utility and the Leaf Porometer. Ifthis error happens regularly, you can try setting your baudrate lower or increasing the number of times a command issent to the Leaf Porometer. Choose the “Communications”tab in Preferences to alter this.

While the Leaf Porometer has been designed to be a reli-able, long-lasting piece of equipment, problems can occur.Following is a list of some of the most common problemsyou could encounter while using the Leaf Porometer. If youhave a problem not described here, or if these remedies can-not fix your problem, please contact Decagon at1-800-755-2751 (US/Canada) or at 1-509-332-2756 (else-where).

PROBLEM: The Leaf Porometer won’t turn on.

SOLUTION: Make sure that the batteries are inserted correctly, and/or ifthey have enough power to activate the machine.

PROBLEM:You cannot see your display very well.

SOLUTION:Turn off the Leaf Porometer and then turn it on again. Ifthis does not fix the problem, the Leaf Porometer features abuilt-in screen for changing screen contrast. See “Contrast”in Chapter 3 for details. You may be able to fix the contrast

56

Page 61: Leaf Porometer v6

Leaf Porometer Operator’s Manual6. Troubleshooting the Leaf Porometer

manually, which can be done using a terminal port program.Contact Decagon for information regarding this procedure.

PROBLEM:No information is being displayed.

SOLUTION:

Any conductivity that is lower that 5 mmol/m2s1 will not bedisplayed in Automatic mode. Therefore, check to see whatmode you’re in before beginning. It is advisable to use Man-ual mode if you know that your leaf has a conductivity thatis lower than the minimum threshold for Auto mode. Also,remember that you must press Enter in Automatic modebefore you begin taking a reading. You may also have gottenmoisture inside the sensor assembly or Teflon disk bytouching it with your finger. This will distort the sensor’sabilities to take accurate readings.

PROBLEM:You are not getting the conductivity reading youexpected.

SOLUTION:The inside of the sensor assembly may be dirty. Remove theTeflon disk and clean it thoroughly (see next Chapter). Theteflon disk also may either be saturated or dirty and mayneed to be dried out or replaced. Additionally, if your sensorwas exposed to any outgassing materials, it may requirerepairs. Please contact Decagon for further help.

57

Page 62: Leaf Porometer v6

Leaf Porometer Operator’s Manual6. Troubleshooting the Leaf Porometer

PROBLEM:The following screen appears:

SOLUTION:This screen means that your firmware was not loaded prop-erly, and that an update is needed. (The firmware for theLeaf Porometer is user-upgradeable.) To obtain this, visitDecagon’s website to see if new updates are available; if not,contact Decagon for the newest firmware version.

Note: You will lose any saved data records when you installthe new firmware. Therefore, download any data you wishto save before re-loading the new firmware. If this is not anoption, contact Decagon for further help.

To obtain new firmware for your Leaf Porometer, see thesection on “Check for Firmware Updates” in Chapter 4 formore information.

Note: See also Chapter 4 for troubleshooting the LeafPorometer Utility.

58

Page 63: Leaf Porometer v6

Leaf Porometer Operator’s ManualFurther Reading

Further ReadingFor more information on porometry, stomatal conductance,and how porometers work, consider reading:

Campbell, G. S and Norman, J.M. An Introduction to Envi-ronmental Biophysics. Springer-Verlag: New York,Heidelberg, Berlin. Second Ed. 1998. pp. 90-92

Jones, H.G. Plants and Microclimate. Cambridge UniversityPress: Cambridge. 1983.

Marshall, B. and Woodward, F.I. Instrumentation for Envi-ronmental Psychology. Cambridge University Press:Cambridge. 1985.

Pearcy, R.W. et al. Plant Physiological Ecology: Field-Methods and Instrumentation. Chapman and Hall:New York, London. 1989. Pp. 137-158.

59

Page 64: Leaf Porometer v6

Leaf Porometer Operator’s ManualLeaf Porometer CE Compliance

CE Compliance

DECLARATION OF CONFORMITY

Application of Council Directive: 89/336/EEC

Standards to Which Conformity is EN61326 : 1998Declared: EN55022 : 1998

Manufacturer’s Name: Decagon Devices, Inc.

Type of Equipment: Leaf Porometer

Model Number: SC-1

Year of First Manufacture: 2005

This is to certify that the Leaf Porometer, manufactured byDecagon Devices, Inc., a corporation based in Pullman,Washington, USA, meets or exceeds the standards for CEcompliance as per the Council Directives noted above. Allinstruments are built at the factory at Decagon and perti-nent testing documentation is freely available for verifica-tion.

60

Page 65: Leaf Porometer v6

Leaf Porometer Operator’s ManualLeaf Porometer CE Compliance

Appendix A

Send Feedback to DecagonDecagon Software makes it easy to send feedback, bugreports, and feature requests to Decagon or your DecagonDistributor. Choose “Send Feedback to Decagon...” fromthe help menu. This opens the window shown below.

61

Page 66: Leaf Porometer v6

Leaf Porometer Operator’s ManualLeaf Porometer CE Compliance

Enter your name, and other contact information. Tell uswhat type of feedback you are sending (General Feedback,Feature Suggestions, Bug Report, or Other). Indicate howyou want us to respond to your feedback (E-mail or tele-phone).

Use the description area to give details for your feedback. Ifyou are reporting a bug, it is very helpful for you to tell uswhat steps you took for the bug to happen and any errormessage you saw. By default, bug reports also include thesoftware error files.

You can send Decagon a file using this form too. This isuseful for sending data files that you have questions about.

If you work directly with a Decagon representative, puttheir E-mail address in the field at the bottom of the form.This sends the contents of the form to them. Your Deca-gon representative can follow-up with you directly.

62

Page 67: Leaf Porometer v6

Leaf Porometer Operator’s Manual

IndexAAbaxial conductance 12Annotating data 17Appendix A 61Arrow keys 7Automatic mode 27Averaging readings 18

BBatteries

changing 51type 5

Button descriptions 6

CCarbon dioxide 21CE compliance 60Comm ports, force find all 39Configuration menu 25Contrast

adjusting using the contrast menu 28

DData menu 22Date 25Downloading data 23

EE-mail 1

63

Page 68: Leaf Porometer v6

Leaf Porometer Operator’s Manual

Enter key 7Enviromental effects on readings 20Environmental effects on readings 21Erasing data 23Escape key 7External sensor

cleaning the sensor block 51correct hand position 12

FFax number 1Firmware update 58Further reading 60

GGeneral precautions 54

HHow the Leaf Porometer works 42Humidity 21

LLeaf porometer utility

actions menu 41auto clock synchronize 40command retires 38downloading data 31edit menu 36erasing data 33error log file 40help menu 41maximum baud rate 39measurement units for data files 37

64

Page 69: Leaf Porometer v6

Leaf Porometer Operator’s Manual

retreiving status information 35saved data organization 33setting date & time 34system requirements 30troubleshooting 55version checks 40

Light 20

MMain menu 11Manual mode 27Measurement

saving 17taking in automatic mode 13taking in manual mode 15

Menu key 7Mode

selecting 27

PPhone number 1Power key 6

RRepairs

instructions 2SSeller’s liability 2Send Feedback 61Specifications 4Stomatal conductance

explanation of 42

65

Page 70: Leaf Porometer v6

Leaf Porometer Operator’s Manual

TTech support 1Temperature 21

UUnits

explanations of 4

VViewing data 22

WWarranty 1

66