Laerdal Compact Suction Unit Training Guide

Greater Sydney HEMS ASNSW

Page 2: Lcsu3 training guide

LCSU 3● Battery powered

portable suction unit, up to 45 mins constant use.

● 300ml canister, single use.

● Variable vacuum 50 – 500mmHg.

● Rechargeable and removable battery.

Page 3: Lcsu3 training guide

LCSU 3 top

Power on /off

Suction power mmHg

Battery low indicator

External power

Charging (out when fully charged)

Page 4: Lcsu3 training guide

LCSU 3 Side

Tubing attachment

Filter assembly

Graduated 300mL canister. Disposable

single use only

Suction control

Page 5: Lcsu3 training guide

USE

1.Attach tubing and Yankauer to canister inlet.

2.Turn on and check suction by occluding outlet – look for green light to 550mmHg.

3.Leave off until suction actually needed – (conserve battery and lessen noise)

4.Ensure unit is upright to avoid contact of fluid with filter.

Page 6: Lcsu3 training guide

StowagePrehospital Trauma Pack 2 (Red).The unit is self contained, including one canister, Yankauer Sucker and tubing length.

Page 7: Lcsu3 training guide

Cannister● Canister snaps

into place.

● Note, if filter becomes wet, the suction will cease to protect the internal motor.

● Always use in upright position to maximise available volume

Page 8: Lcsu3 training guide

Batteries● Use coin or similar to open housing.

● Remove Battery and disconnect cable.

● Replace with fully charged battery.

● Battery can be charged within unit.

● Device can run while being charged

● Charge duration 5 hours max

● Battery life – 45 mins maximum continuous use.

Page 9: Lcsu3 training guide

Important Notes●The suction will cease if the filter becomes wet, this

protects the motor.

●Avoid canister content from coming into contact with filter, eg by careless handling.

●The low battery warning indicates < 5 mins power.

● If suction fails consider moving to alternative, eg Vehicle, Aircraft, Bridge.

●The catheter tip 60mL syringe is available in the intubation kit as a last resort.

Page 10: Lcsu3 training guide

Usage●This device is to be utilised on all RSI if vehicle

suction is not practical or available.

●The 300mL canister may fill rapidly, source alternate suction if likely, by moving patient to vehicle.

●The canister cannot be emptied and re-used.

●The device will be checked on daily level 3 checks.

●The entire unit should be rotated and placed on charge, for level 1 checks and after use.

Page 11: Lcsu3 training guide

Specifications● Weight: 1.53 kg● Size: 18cmH x 26.7cmW x 7.4cmD● Disposable Canister Capacity: 300ml ● Integrated Filter: Internal Bacterial Filter/ Fluid Shut-off ● Patient Suction Tubing (non-sterile) ● Length: 0.9m ● Vacuum - Range: 50-550mmHg (6.7 - 73kPa) ● (+/-27.5 mmHg) ● Battery Operation Time: Up to 45 minutes ● Noise Level: < 70 dB(A) ● *Conditions may vary according to environmental factors, system configuration, power source,

age/maintenance of battery, and/or state of charge. ● Electrical Requirements: 100-240 V AC 47-63 Hz, 0.75A max; 12 V DC, 33 W max ● Battery: High Capacity Rechargeable NiMH ● Charging Time: Up to 5 hours ● Operating ● Temperature Range: 0 degrees C - 40 degrees C● Relative Humidity: 0-95% ● Atmospheric Pressure: 10.2 Psi (70 kPA) - 15.4 Psi (106 kPA) ● Storage and Transport ● Temperature Range: -40 degrees C - 70 degrees C● Relative Humidity: 0-95% ● Atmospheric Pressure: 7.3 Psi (50 kPA) - 15.4 Psi (106 kPA)