LCP Distal Ulna Plate. For capital and subcapital fractures of the ulna. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.
28

LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

Oct 05, 2018

Download

Documents

phungxuyen
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate. For capital and subcapital fractures of the ulna.

Surgical Technique

This publication is not intended for distribution in the USA.

Instruments and implants approved by the AO Foundation.

Page 2: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

Image intensifier control

This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction by a surgeon experienced in handling these products is highly recommended.

Processing, Reprocessing, Care and MaintenanceFor general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to:http://emea.depuysynthes.com/hcp/reprocessing-care-maintenanceFor general information about reprocessing, care and maintenance of Synthes reusable devices, instrument trays and cases, as well as processing of Synthes non-sterile implants, please consult the Important Information leaflet (SE_023827) or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance

Page 3: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 1

Table of contents

Introduction LCP Distal Ulna Plate 2

Indications 4

Clinical Cases 6

Surgical Technique 8

Implant Removal 18

Product Information Implants 19

Instruments 20

Bibliography 21

MRI Information 22

Page 4: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

2

1

3 3

2

1

2 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

LCP Distal Ulna Plate. For capital and subcapital fractures of the ulna.

Introduction

1 Pointed hooks and locking screw in the head Pointed hooks to grip the styloid process and to act

as reference point for plate application. Intercrossing locking screws securely hold the ulnar head. The shaft contains LCP combi-holes.

2 Angular stability The head accepts 2.0 mm locking screws. The shaft

accepts either 2.0 mm locking or cortical screws.

3 Oblong hole The oblong hole accepts 2.0 cortex screws for ulnar length

adjustment.

4 Anatomically precontoured The slim plate design, low screw/plate profile, rounded

edges and polished surface help limit the irritation of overlying soft tissues.

Page 5: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

4

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 3

Locking screwsX01.876–900 Locking Screw B 2.0 mm, self-tapping

Cortex screwsX01.356–381 Cortex Screw B 2.0 mm, self-tapping

Page 6: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

4 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

Indications

Fractures of the distal ulna which result in instability of the distal radioulnar joint

Fractures of the ulna head where the articular surface is either displaced, rotated or tilted.

displaced articular surface rotated articular surface

tilted articular surface

Page 7: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 5

Comminuted extra-articular fractures of the ulnar neck threatening stable congruency of the distal radioulnar joint.

Note: Not all fractures of the distal ulna require internal fixation. Many simple ulnar styloid fractures demand nothing more than symptomatic treatment.

Page 8: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

6 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

24 year old male patient.DRUJ and long-term painful instability of the fracture of the base of the ulna styloid. Retaining of the TFCC by osteotomy and osteosynthesis using the distal ulna plate.

Pre-op

Post-op

Clinical Cases

Page 9: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 7

83 year old female patient, fall.Stabilization of subcapital ulna and concomitant ulna styloid fracture with LCP Distal Ulna Plate.

LCP Distal Volar Column Plate for fixation of distal radius fracture.

Pre-op

Post-op (8 months)

Page 10: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

8 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

The distal ulna is an essential component of the distal radio-ulnar joint, which provides rotation to the forearm. The distal ulnar surface is also an important platform for stability of the carpus and, beyond it, the hand. Unstable fractures of the distal ulna therefore threaten both movement and stability of the wrist.

The size and shape of the distal ulna combined with the overlying mobile soft tissues, make application of standard implants difficult. The LCP Distal Ulna Plate has been spe cifically designed for use in fractures of the distal ulna.

Surgical Technique

Surgical Technique

Page 11: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 9

Gently retract the dorsal sensory branch of the ulnar nerve.

1Surgical approach

The distal ulna is subcutaneous. A bare area exists between the tendons of flexor carpi ulnaris and extensor carpi ulnaris. This bare area continues on towards the ulnar styloid and represents the optimum site for application of the LCP Distal Ulna Plate.

Make a longitudinal skin incision over the palpable ulna. Take great care to avoid the dorsal sensory branch of the ulnar nerve, which crosses the bone at this level to supply the dorsal skin of the hand.

Once the distal shaft of the ulna is visible, subperiosteal dissection will allow the fracture fragments to be visualized and reduced.

Page 12: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

10 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

Surgical Technique

2Contour plate (optional)

Instrument

347.901 Pliers, flat-nosed, pointed, for Plates 1.0 to 2.4

If necessary contour the plate with flat-nosed pliers.

Notes – The plate holes have been designed to accept some de-

gree of deformation. The undercut helps to ensure that the threaded holes will not be distorted with typical con-touring. Significant distortion of the threaded holes will reduce locking effectiveness.

– If possible, the plate should not be cut since the resulting sharp edges can irritate the overlying soft tissues.

Precaution: The plate features pointed hooks which should be handled with care.

Page 13: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 11

3Reduce fracture and position plate

Instrument

292.622/623 Guide Wire B 1.1 mm with threaded tip with trocar/ with trocar tip, length 150 mm, Stainless Steel

Expose and clean the fracture. Secure the pointed hooks of the LCP Distal Ulna Plate around the tip of the ulnar styloid as a reference guide.

In simple fractures of the ulnar neck, apply the plate then to the subcutaneous border of the distal ulna, securing points of fixation in both the head and the shaft.

Page 14: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

12 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

Surgical Technique

Note: It might be necessary to temporarily stabilize the fracture with a trans-styloid guide wire B 1.1 mm. The wire should be inserted between the distal hooks of the tempo-rarily applied plate.

Precaution: The head of the distal ulna is often fragile. Cau-tion should be exercised if using pointed reduction forceps, since the force of this instrument may cause further com-minution of the ulnar head. Much of the reduction will be performed indirectly.

Complete exposure of the ulnar head should not be pefor- med because this will detach essential soft tissue stabilizers.

Page 15: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 13

4Fix plate distally

Instruments

310.507 Drill Bit B 1.5 mm with Marking, Length 96/82 mm, 2-fluted, for Mini Quick Coupling

323.034 LCP Drill Sleeve 2.0, with Scale, for Drill Bits B 1.5 mm with marking

319.005 Depth Gauge for Screws B 2.0 and 2.4 mm, measuring range up to 40 mm

313.842 Screwdriver Shaft Stardrive 2.0, short,or self-holding, for Mini Quick Coupling313.843 Screwdriver Shaft Stardrive 2.0, long, self-holding, for Mini Quick Coupling

311.012 Handle with mini Quick Coupling for 313.842/843

Secure the drill sleeve in the desired hole. Predrill the hole with the drill bit B 1.5 mm through the drill sleeve and measure the screw length directly from gauge. Remove the drill sleeve.

Page 16: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

14 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

Surgical Technique

Alternatively, screw length may be measured with the depth gauge.

Introduce the measured length of the 2.0 mm locking screw with the screwdriver until seated.

Page 17: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 15

5aLength adjustment

Instruments

313.842 Screwdriver Shaft Stardrive 2.0, short,or self-holding, for Mini Quick Coupling 313.843 Screwdriver Shaft Stardrive 2.0, long, self-holding, for Mini Quick Coupling

311.012 Handle with mini Quick Coupling for 313.842/843

In fractures which require length adjustment, place one or two 2.0 mm locking screws in the ulnar head to securely fix the implant distally. Then place a 2.0 mm cortical screw in the oblong hole of the shaft, and obtain the correct length of reduction. Use surrounding holes to stabilize the fracture securely, using a combination of cortical and locking screws as dictated by bone quality.

5Length Adjustment and Final Fixation

Multiple options for screw insertion in the distal portion of the plate allow a wide range of fracture patterns to be se-curely stabilized.

Page 18: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

16 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

Surgical Technique

5cFixation: between the arms of the distal hooks

Instruments

314.667 Screwdriver Shaft 1.5, cruciform, with Holding Sleeve, length 66 mm, with Mini Quick Coupling

310.507 Drill Bit B 1.5 mm with Marking, Length 96/82 mm, 2-fluted, for Mini Quick Coupling

In fractures where it is necessary to stabilize the tip of the ulnar styloid process, the most distal plate hole is left empty. Remove the Guide Wire B 1.1 mm, which was used for pre-liminary fixation (see note step 3).

5bFixation: most distal hole

In the case of unstable fractures of the base of the ulnar styloid, a 2.0 mm locking screw can be applied through the most distal hole in the Distal Ulnar Plate. This screw does not need to reach the far cortex because it is a locking screw.

For the insertion of locking screws see step 4 on page 13.

Page 19: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 17

6Confirm accurate joint reconstruction

Confirm accurate joint reconstruction, screw placement and screw length from different angles – AP, lateral and multiple oblique views.

Check that no screws enter either the distal radioulnar or ulnocarpal joints by using an image intensifier.

AP-view

Overdrill the near fragment with a drill bit B 1.5 mm. Apply a cortical screw B 1.5 mm in lag mode between the arms of the distal hooks.

Page 20: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

18 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

Implant Removal

8Implant Removal

Instruments

311.012 Handle with Mini Quick Coupling

314.676 Screwdriver Shaft Stardrive 2.0, with Holding Sleeve, for Mini Quick Coupling

314.667 Screwdriver Shaft 1.5, cruciform, with Holding Sleeve, length 66 mm, with Mini Quick Coupling

To remove locking screws, first unlock all screws from the plate; then remove the screws completely from the bone. The last screw removed should be a non-locking screw on the shaft. This guarantees the plate not to spin when locking screws are removed.

Surgical Technique

7Postoperative treatment

Remove the surgical drain on the first post-operative day. Rest the forearm in a sugar tong splint for two weeks after surgery and begin supervised rehabilitation concentrating on both flexion/extension of the wrist and forearm rotation. Wear a moulded thermoplastic wrist gauntlet in between rehabilitation sessions for 3 – 4 weeks after surgery.

Page 21: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 19

Implants

Plates

X42.531 LCP Distal Ulna Plate 2.0, with hooks, 7 holes, length 46 mm

X = 2 Stainless Steel X = 4 Pure Titanium

Locking screws

X01.876–900 Locking Screw B 2.0 mm, self-tapping

Cortex screws

X01.356–381 Cortex Screw B 2.0 mm, self-tapping

Optional

X00.806–824 Cortex Screw B 1.5 mm, self-tapping, length 6 mm

X = 2 Stainless Steel X = 4 Titanium Alloy

All implants are also available sterile.Add suffix S to article number.

Product Information

Page 22: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

20 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

Instruments

311.012 Handle with Mini Quick Coupling

314.676 Screwdriver Shaft Stardrive 2.0, with Holding Sleeve, for Mini Quick Coupling

323.034 LCP Drill Sleeve 2.0, mit Scale, for Drill Bits B 1.5 mm with marking

313.842 Screwdriver Shaft Stardrive 2.0, short, self-holding, for Mini Quick Coupling

313.843 Screwdriver Shaft Stardrive 2.0, long, self-holding, for Mini Quick Coupling

310.507 Drill Bit B 1.5 mm with Marking, Length 96/82 mm, 2-fluted, for Mini Quick Coupling

319.005 Depth Gauge for Screws B 2.0 and 2.4 mm, measuring range up to 40 mm

The above instruments are part of LCP Compact Hand

292.622/623 Guide Wire B 1.1 mm with threaded tip with trocar/with trocar tip, length 150 mm, Stainless Steel

Optional Instruments

347.901 Pliers, flat-nosed, pointed, for Plates 1.0 to 2.4

314.667 Screwdriver Shaft 1.5, cruciform, with Holding Sleeve, length 66 mm, with Mini Quick Coupling

Page 23: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

LCP Distal Ulna Plate Surgical Technique DePuy Synthes 21

Bibliography

Ring D, McCarty LP, Campbell D, Jupiter JB (2004) Condylar blade plate fixation of unstable fractures of the distal ulna associated with fracture of the distal radius.J Hand Surg [Am] 29(1):103–9

May MM, Lawton JN, Blazar PE (2002) Ulnar styloid fractures associated with distal radius fractures: incidence and implica-tions for distal radioulnar joint instability. J Hand Surg [Am] 27(6):965–71

Dennison DG (2007) Open Reduction and Internal Locked Fixation of Unstable Distal Ulna Fractures With Concomitant Distal Radius Fracture.J Hand Surg [Am] 32:801–805

Bibliography

Page 24: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

22 DePuy Synthes LCP Distal Ulna Plate Surgical Technique

MRI Information

Torque, Displacement and Image Artifacts according to ASTM F 2213-06, ASTM F 2052-06e1 and ASTM F2119-07Non-clinical testing of worst case scenario in a 3 T MRI system did not reveal any relevant torque or displacement of the construct for an experimentally measured local spatial gradient of the magnetic field of 3.69 T/m. The largest image artifact extended approximately 169 mm from the construct when scanned using the Gradient Echo (GE). Testing was conducted on a 3 T MRI system.

Radio-Frequency-(RF-)induced heating according to ASTM F2182-11aNon-clinical electromagnetic and thermal testing of worst case scenario lead to peak temperature rise of 9.5 °C with an average temperature rise of 6.6 °C (1.5 T) and a peak temperature rise of 5.9 °C (3 T) under MRI Conditions using RF Coils [whole body averaged specific absorption rate (SAR) of 2 W/kg for 6 minutes (1.5 T) and for 15 minutes (3 T)].

Precautions: The above mentioned test relies on non-clini-cal testing. The actual temperature rise in the patient will depend on a variety of factors beyond the SAR and time of RF application. Thus, it is recommended to pay particular attention to the following points: – It is recommended to thoroughly monitor patients under-

going MR scanning for perceived temperature and/or pain sensations.

– Patients with impaired thermo regulation or temperature sensation should be excluded from MR scanning proce-dures.

– Generally it is recommended to use a MR system with low field strength in the presence of conductive implants. The employed specific absorption rate (SAR) should be reduced as far as possible.

– Using the ventilation system may further contribute to reduce temperature increase in the body.

MRI Information

Page 25: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital
Page 26: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital
Page 27: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital
Page 28: LCP Distal Ulna Plate. For capital and subcapital fractures …synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes International... · LCP Distal Ulna Plate. For capital and subcapital

0123

Synthes GmbHEimattstrasse 34436 OberdorfSwitzerlandTel: +41 61 965 61 11Fax: +41 61 965 66 00www.depuysynthes.com

Not all products are currently available in all markets.

This publication is not intended for distribution in the USA.

All surgical techniques are available as PDF files at www.depuysynthes.com/ifu ©

DeP

uy S

ynth

es T

raum

a, a

div

isio

n of

Syn

thes

Gm

bH. 2

016.

A

ll rig

hts

rese

rved

. 03

6.0

00.

494

DSE

M/T

RM

/081

5/0

462

(1)

03/1

6