1 Part I – Basic Principles of HPLC Part I – Basic Principles of HPLC Chromatography and HPLC Various HPLC modes and applications Normal phase Reversed phase Reversed phase ion pairing Ion exchange (IC) SEC (GPC/GFC) Chiral separation HPLC system modes Isocratic elution Gradient elution
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
1
Part I – Basic Principles of HPLCPart I – Basic Principles of HPLC
Chromatography and HPLCVarious HPLC modes and applications
Normal phaseReversed phaseReversed phase ion pairing Ion exchange (IC)SEC (GPC/GFC)Chiral separation
HPLC system modes Isocratic elution Gradient elution
2
What is chromatography?What is chromatography?
Chromatography is one of the separation technique.
The main purpose of chromatography is
to separateseparate and quantifyquantify the target sample in the matrix.
3Why mixed sample Why mixed sample can be separated? can be separated?
river bed
direction of flow
4
What is the difference?What is the difference?
Interaction is different by gravityStrong
Weak
5How can the separation How can the separation be carried out?be carried out?
Separation can be carried out by the column.
packing material, 3-5um
column
6Interaction between packing Interaction between packing
material and samplematerial and sample
Due to difference interaction between packing material and sample, separation can be done.
packing material
sample A
sample B
7
colum
n
mixed sample
Separation can be done Separation can be done by the columnby the column
8
What kind of chromatography?What kind of chromatography?
High Performance Liquid chromatography (HPLC)
Gas Chromatography (GC) Thin-Layer Chromatography (TLC) Capillary Electrophoresis(CE)
9
Advantage of ChromatographyAdvantage of Chromatography
Simultaneous Analysis
High Resolution
High Sensitivity (ppm-ppb)
Small Injection Volume (1-100uL)
10
Advantage of HPLCAdvantage of HPLC
Moderate analysis condition - no need to vaporize the sample like
GC
Easy to fractionate the sample and purify
Good repeatability (C.V. < 1%)
11
Flow Diagram of HPLCFlow Diagram of HPLC
pumpinjector
column
ovendetector
12
Normal Phase ModeNormal Phase Mode
Petroleum etherPetroleum ether
CaCOCaCO33
Chlorophyll'sChlorophyll's
ChromatoChromatograph
ColorsColors
13
Effect of stationary phaseEffect of stationary phase
C18 (ODS)
Strong
C8
sample
sample
sample
C4
Medium
Weak
14
Elution orderElution order
SECColumn
15Relationship between Relationship between
MW and RTMW and RT
Molecu
lar Weigh
t (LogM
W)
Exclusion limit
Permeation limit
Time
16
Isocratic Elution SystemIsocratic Elution System
pump injector
column
ovendetector
Single SolventSingle Solvent
17
Gradient Elution SystemGradient Elution System
pumpinjector
column
ovendetector
pump
A
B
Time
B con
centration
18
Isocratic Elution ModeIsocratic Elution Mode
Long Time AnalysisLong Time Analysis
Bad SeparationBad Separation
MeOH / H2O = 6 / 4
MeOH / H2O = 8 / 2
( column : ODS type )
19
Gradient Elution ModeGradient Elution Mode
95%
30% MeO
H con
centration
20
Part II Instrumentation of HPLCPart II Instrumentation of HPLC
The number of maintenance parts is less. So this design is suitable for routine analysis.
26Dual plunger Dual plunger
with parallel flow line with parallel flow line
check valve
plunger head
Very low pressure fluctuation Refractive index detector Conductivity detectorElectrochemical detectorMS detector
The number of maintenance parts is more.
27
Single plunger Single plunger
check valve
High pressure-fluctuation Low sensitivity analysis using UV / PDA and Fluorescence detector
The number of maintenance parts is minimized.
28
HPLC InjectorsHPLC Injectors
Manual injector
Auto injector
29
6 port valve system6 port valve system
column
load position inject position
pump pump column
sample loop
30
Manual InjectorManual Injector
Rheodyne Manual InjectorRheodyne Manual Injector
31
Manual InjectorManual Injector
[LOAD][LOAD] [INJECT][INJECT]
Column
Pump
Column
Pump
32
Injection MethodInjection Method
Partial Injection MethodPartial Injection Methodbetter to inject less than half volume of
sample loop
Loop Injection MethodLoop Injection Methodbetter to inject more than 3 times
volume of sample loop
33
Dispersion and Dilution EffectDispersion and Dilution Effect
loop
pump
column
drain
drainpump
column
loop
drain
drain
Sample ZoneDilution ZoneDispersion Zone
Less than half volume Less than half volume More than half volume More than half volume
34
Dispersion and Dilution EffectDispersion and Dilution Effect
drain
drain
pump
column
loop
Sample ZoneDilution Zone
Less than 3 timesLess than 3 times More than 3 timesMore than 3 times
drain
drain
pump
column
loop
35
INJECT / LOAD positionINJECT / LOAD position
[INJECT ][INJECT ] position position
[LOAD][LOAD] position position
36
CautionCaution
Do not use pointed or beveled needle tip. Must use square end type.
Do not use more than pH 10 solution.Must change rotor seal.
37Column Temperature Column Temperature
Control DevicesControl Devices
Column Oven (heat /or cool) Heated Column Jacket
Aluminum block
Insulated Column Jacket Water bath
Column temperature control devices are functioning to keep the column temperature constant. The temperature fluctuation of column will influence retention time reproducibility.