Laser Based Techniques for Ultra Trace Isotope Production, Spectroscopy and Detection KLAUS D. A. WENDT 1, * , KLAUS BLAUM 1 , CHRISTOPHER GEPPERT 1 , PETER MU ¨ LLER 1 , WILFRIED NO ¨ RTERSHA ¨ USER 1 , ANNETTE SCHMITT 1 , PHILIPP SCHUMANN 1 , NORBERT TRAUTMANN 2 and BRUCE A. BUSHAW 3 1 Institut fu ¨r Physik, Johannes Gutenberg Universita ¨t, D-55099, Mainz, Germany e-mail: [email protected]2 Institut fu ¨r Kernchemie, Johannes Gutenberg Universita ¨t, D-55099, Mainz, Germany 3 Pacific Northwest National Laboratory, Richland, WA 99352, USA Abstract. A variety of research activities in the field of fundamental and applied nuclear physics has evolved in the last years using resonantly tuned radiation from powerful lasers. The technique of resonance ionization spectroscopy has delivered outstanding results and found broad acceptance in the last years as a particularly efficient and highly selective method for rare and exotic radioisotope studies. It is used for production, spectroscopy and detection of these species and provides complete isobaric, high isotopic and even some isomeric selection, which altogether is needed for on-line investigation of short lived species far off stability as well as for ultra trace determination. Good overall efficiency pushes the experimental limits of detection in elemental trace analysis down to below 10 6 atoms per sample, and additionally isotopic selectivity as high as 3 10 12 has been demonstrated. The widespread potential of resonance ionization techniques is discussed, focusing on the experimental arrangements for applications in selective on-line isotope production, spectroscopy of rare radioisotopes and ultra trace determination of radiotoxic isotopes like 238 Pu to 244 Pu, 135,137 Cs, 89,90 Sr or 41 Ca in environmental, technical and biomedical samples. Key Words: laser spectroscopy, laser development, nuclear structure, radioisotopes, resonance ionization, ultra trace detection. 1. Introduction As discussed extensively during this meeting on BLaser Methods In The Study of Nuclei, Atoms and Molecules,^ the atomic nucleus with its electromagnetic properties, determined by spin, nuclear moments and charge radius interacts significantly with the electronic shell. Thus atomic spectroscopy can reveal a detailed view into behavior, structure and stability of nuclear matter in its ground * Author for correspondence. Hyperfine Interactions (2005) 162:147–157 DOI 10.1007/s10751-005-9219-8 # Springer 2006
11
Embed
Laser Based Techniques for Ultra Trace Isotope Production, Spectroscopy and Detection
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Laser Based Techniques for Ultra Trace Isotope
Production, Spectroscopy and Detection
KLAUS D. A. WENDT1,*, KLAUS BLAUM1, CHRISTOPHER GEPPERT1,
PETER MULLER1, WILFRIED NORTERSHAUSER1, ANNETTE SCHMITT1,
PHILIPP SCHUMANN1, NORBERT TRAUTMANN2
and BRUCE A. BUSHAW3
1Institut fur Physik, Johannes Gutenberg Universitat, D-55099, Mainz, Germany
e-mail: [email protected] fur Kernchemie, Johannes Gutenberg Universitat, D-55099, Mainz, Germany3Pacific Northwest National Laboratory, Richland, WA 99352, USA
Abstract. A variety of research activities in the field of fundamental and applied nuclear
physics has evolved in the last years using resonantly tuned radiation from powerful lasers. The
technique of resonance ionization spectroscopy has delivered outstanding results and found broad
acceptance in the last years as a particularly efficient and highly selective method for rare and
exotic radioisotope studies. It is used for production, spectroscopy and detection of these species
and provides complete isobaric, high isotopic and even some isomeric selection, which
altogether is needed for on-line investigation of short lived species far off stability as well as
for ultra trace determination. Good overall efficiency pushes the experimental limits of detection
in elemental trace analysis down to below 106 atoms per sample, and additionally isotopic
selectivity as high as 3 � 1012 has been demonstrated. The widespread potential of resonance
ionization techniques is discussed, focusing on the experimental arrangements for applications in
selective on-line isotope production, spectroscopy of rare radioisotopes and ultra trace
determination of radiotoxic isotopes like 238Pu to 244Pu, 135,137Cs, 89,90Sr or 41Ca in environmental,
Figure 1. Comparison of different possible excitation schemes for resonance ionization. Typical
cross-sections for the individual excitation and ionization steps are indicated.
150 K. D. A. WENDT ET AL.
with Nk(t) being the level population, Jk(t) the photon density of the laser
field interconnecting level k with level k + 1, �k,kj1 the spontaneous transition
rate from level k to level kj1, given directly by the Einstein A-factor �k,kj1 =
Ak,kj1 and gk comprising the spontaneous decay rates from level k into lower
states, which are not further contributing, as well as the ionization process itself.
If all the atomic parameters of the excitation and ionization scheme are known,
the profile and intensity of the ionization signal can be determined with the use
of the well known Lorentzian frequency dependence of the individual optical
cross sections.
For a precise calculation the spatial and temporal laser profile, the spectral
frequency distribution as well as the three-dimensional velocity distribution of the
atomic beam must be properly taken into account. This leads to tedious numerical
integrations and convolutions but delivers precise theoretical predictions.
This simple approach is just not valid for the case of multi-step excitation
with narrow bandwidth continuous laser excitation, where the coherences
between the atomic levels involved, induced by the different laser fields, must
be properly considered. This can be done by applying the density matrix
formalism, which leads to rather complex asymmetric line shapes in the multi-
dimensional space of the detunings of the individual laser frequencies [15].
Obviously high efficiency in resonance ionization would involve either
(A) pulsed atomization well synchronized to the laser pulses, e.g., by pulsed
release or desorption of atoms from a target or beam source,
(B) the use of ultimately high repetition rate lasers together with continuous
atomization, or alternatively
(C) the combination of low repetition rate lasers with a buffer gas cell, as e.g.,
realized in an ion guide laser ion source [16].
The first process (A) is affected by unspecific background, i.e., ionized
molecules and clusters resulting from the pulsed heating or desorption process.
Thus, it has been dominantly applied to on-line studies on artificially produced
radioisotopes of e.g., refractory elements, which are not easily accessible by
other techniques, e.g., at experiments at ISOLDE/CERN [17–19].
In the second approach (B) the laser duty cycle directly affects the achievable
overall efficiency and must be chosen in the high kHz range. Furthermore the
location of resonance ionization is usually placed inside a widely closed hot
cavity or transfer tube to form a so-called resonance-ionization-laser-ion-source
(RILIS) [20–23]. The atoms have the chance to pass through the laser beam very
often and in this way the overall ionization probability is strongly enhanced up to
values of õ20%. Again background, stemming mainly from the unavoidable
competing process of surface ionization on the hot cavity walls leads to sig-
nificant interferences and limitations, which must be suppressed. For this task
combinations of RILIS with repelling electrodes and the localization of the
LASER BASED TECHNIQUES FOR ULTRA TRACE ISOTOPE 151
resonance ionization region inside a quadrupole ion guide without any wall
contact are developed. Filled with buffer gas, this latter device can also serve for
cooling and bunching of the resulting ion beam to enhance temporal and spatial
beam quality [14].
In approach (C) the atoms are stored in a noble gas buffer gas cell for many
ms with the interaction volume being illuminated by laser light. Laser repetition
rates of õ100 Hz are sufficient for efficient resonant ionization. The ions are
guided to a nozzle by electric fields, flushed out by a gas jet, separated from the
buffer gas by skimmers, and transferred to subsequent experiments or detection.
The overall efficiency here is about 10j4. Very recently this arrangement has
enabled the first spectroscopic studies on fermium with about 1010 atoms of255Fm [24].
3. Pulsed laser systems for resonance ionization
The discussion given here shall be limited to high repetition rate laser systems as
being typically used for resonance ionization: Until very recently these consisted
of two to three tunable dye lasers pumped by at least two powerful Cu-vapor
lasers (Cvl) [25]. Outstanding properties are the high repetition rate of 6 to 11
kHz and the broad tuning range through the use of a number of suitable dyes as
well as frequency doubling units, together with a well adapted short pulse length
of õ20 ns and spectral width of a few GHz. However, drawbacks of such a
system are its size, high maintenance efforts and costs of the pump lasers. This
has led to the development of novel, powerful and easy-to-handle solid state laser
systems, which are presently conquering the field [26]. Nowadays highly reliable
Q-switched and intracavity doubled Nd:YAG lasers with repetition rates of 1–25
kHz, power of 50 to 100 W at 532 nm and pulse lengths in the 100 ns regime are
commercially available. These are well suited for simultaneously pumping of up
to three titanium–sapphire (Ti:Sa) lasers as well as further solid state amplifier
stages. Each solid state laser provides an average power of up to 3 W in a tuning
range from õ700 to õ1,000 nm, line width around 2–5 GHz and pulse duration
of õ60 ns. For resonance ionization the three lasers must be synchronized, which
is achieved with Pockels cells used as intracavity Q-switches. Single pass
frequency doubling and tripling is possible and required to cover the blue and
ultraviolet spectral ranges.
As compiled in Figure 2, similar to dye lasers, Ti:Sa laser systems are
nowadays generally capable of resonantly ionizing the far majority of elements
in the periodic table. While the Cvl pumped dye lasers already have proven their
performance successfully on more than 20 elements for analytical and on-line
applications, excitation schemes on many remaining elements, in particular
concerning application of the rather novel Ti:Sa laser system must still be
searched and investigated spectroscopically in the near future. These activities
are in progress at a number of on-line facilities and laboratories worldwide [14].
152 K. D. A. WENDT ET AL.
In the frame of analytical applications on elemental species, pulsed laser RIMS
is frequently used for trace and ultra trace determination on long-lived radiotoxic
isotopes of actinides and technetium, where necessary suppression of isobars
excludes conventional mass spectrometric techniques. In combination with
thermal evaporation and simple and reliable time-of-flight mass spectrometers
pulsed laser resonant ionization significantly surpasses the specifications of
radiometric counting techniques in respect to detection limit and short measuring
time by far [27].
4. Resonant processes using narrow bandwidth continuous wave lasers
Resonance ionization using continuous wave lasers provides high isotopic se-
lectivity through utilization of the isotopic shift in high resolution spectroscopy.
In cases where naturally occurring isotope shifts are small and insufficient to
reach the required selectivity collinear geometry on fast atomic beams can be
applied by utilizing the mass dependent Doppler shift. With kinetic energies in
the range of several 10 keV, a large artificial isotope shift in the many GHz range
is introduced, which strongly enhances isotopic selectivity and even can serve for
tuning fixed frequency lasers into resonance to atomic transitions within the rest
frame of the fast moving atom. This strongly simplifies the laser equipment
needed. Furthermore, the conservation of the longitudinal energy spread during
acceleration leads to a significant reduction in the velocity spread dependent
Doppler width. The production of a fast atomic beam involves extensive
machinery: i.e., ionization in a conventional ion source, acceleration of the ions
to the necessary kinetic energy and subsequent efficient neutralization in a charge
Figure 2. Periodic system of the elements, the shaded boxes indicate the accessibility of the
individual element to resonance ionization using Ti:Sa lasers.
LASER BASED TECHNIQUES FOR ULTRA TRACE ISOTOPE 153
exchange medium. Fortunately these components are at least partly realized
within the components of a mass separator, which makes this technology ideally
suited for direct on-line use at exotic radioisotope production facilities of the
ISOL type. The charge exchange process furthermore provides significant
population of up to 10% in meta-stable atomic states, from which an efficient
resonance excitation into high lying states close to the ionization limit is possible
in a single excitation step.
Collinear RIMS was first applied for the determination of the ultra trace noble
gas isotope 3He in environmental samples [28], and later on at CERN for the
investigation of nuclear charge radii and nuclear moments within the isotopic
chain of Yb [29]. This technique was also used for a fast and sensitive laser based
determination of 89,90Sr in various environmental and technical samples [30].
A typical set-up for high resolution RIMS on a collimated thermal atomic
beam is sketched in Figure 3. Multi-step excitation with narrow-band cw lasers is
applied in perpendicular geometry in combination with a quadrupole mass
spectrometer (QMS) or alternatively a magnetic sector field [31]. Due to its
limited experimental expenditure and its compactness, this technique is
especially well suited for trace analysis if compact and cheap diode lasers are
used. Further advantages compared to other RIMS versions are the extremely
high elemental and isotopic selectivity together with a good overall efficiency.
This technique has already been applied successfully for ultra trace analysis of90Sr [32], 135,137Cs [33] and 210Pb [34]. In all applications isotope-selective
Figure 3. Schematic sketch of the compact experimental set-up for diode laser based high
resolution RIMS at a quadrupole mass spectrometer for analytical applications requiring highest
isotopic selectivity.
154 K. D. A. WENDT ET AL.
stepwise excitation in an atomic beam delivered extremely high isotopic selec-
tivity of up to 3 � 1012, which has been demonstrated for the case of 41Ca [35].
A recent review of these techniques in comparison to AMS and atomic trap trace
analysis is found in [36].
5. Conclusions
The spectroscopic technique of resonance ionization on atomic species is
nowadays realized in various experimental arrangements, which are specifically
tailored to meet the demands of selective ionization at on-line facilities or those
of ultra trace determination of rare radioisotopes in various samples for different
applications. Laser systems, mass spectrometric components and ion detectors
are well adapted both for the on-line and shift work as well as for the routine-like
operation in dedicated environmental or biomedical analytical studies. In
analytical applications the technique is complementary in a number of aspects
to AMS and extends the capabilities of conventional mass spectrometry.
Outstanding features are the isobaric and, if required, additional high isotopic
selectivity, high ionization efficiency and the corresponding low detection limits.
Ultimate specifications reported include isobaric suppression of numerous orders
of magnitude, depending on the species under study, isotope selectivities of up to
3 � 1012, overall resonance ionization efficiencies of 20% and beyond and a
precision in isotope ratio measurements of õ1%. In analytic applications
detection limits below 106 atoms are routinely realized, e.g., for Pu isotopes.
The method is further improved by the ongoing development on the side of the
lasers. It is currently extended to cover the far majority of elements within the
periodic table for the purpose to further enhance the knowledge on rare and
exotic species.
References
1. Bemis C. E. Jr. and Carter H. J. (eds.), Lasers in Nuclear Physics, Harwood Academic Publ.,
New York, 1982.
2. Application of Lasers in Atomic Nuclear research, Proc. V. Int. Workshop, Poznan, Dubna
JINR Press, 2001.
3. Otten E. W., Nuclear radii and moments of unstable isotopes, In: D.A. Bromley, Treatise on
Heavy-Ion Science, Vol. 8, Plenum Press, London, 1989, p. 517–638.
4. Neugart R., Lasers in nuclear physics, Eur. Phys. J. A15 (2002), 35.
5. Kluge H.-J. and Nortershauser W., Laser for nuclear physics, Spectrochim. Acta B58 (2003),
1031.
6. Lethokov V. S. (ed.), Laser Photoionization Spectroscopy, Academic Press, Orlando, 1987.
7. Hurst G. S. and Payne M. G. (eds.), Principles and Applications of Resonance Ionisation
Spectroscopy, Adam Hilger Publ., Bristol, 1988.
8. Payne G., Deng L. and Thonnard N., Rev. Sci. Instrum. 65 (1994), 2433.
9. Wendt K., Eur. Mass Spectrom. 8 (2002), 273.
LASER BASED TECHNIQUES FOR ULTRA TRACE ISOTOPE 155
10. Montaser A. (ed.), Inductively Coupled Plasma Mass Spectrometry, Wiley, John & Sons,
New York, 1998.
11. Tuniz C., Bird J. R., Fink D. and Herzog G. F. (eds.), Accelerator Mass Spectrometry, CRC
Press LLC, Boca Raton, 1998.