Top Banner

of 14

Lab 7 Transducers

Jun 02, 2018

Download

Documents

27189
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
  • 8/10/2019 Lab 7 Transducers

    1/14

    Transducer sensitivity and linearity

    Introduction

    In this experiment, we will use three types of transducers: photonic, inductive proximity, and capacitive

    proximity. These transducers function to measure small linear displacements. Each type is explainedbelow.

    Photonic transducer:

    This transducer emits a steady light source through optical fibers, which will hit the object. This light

    will reflect off the object and come back to be received by a photocell detector. The longer it takes for

    light to bounce back off the object, the larger the linear displacement between the object and the

    transducer. The graph below shows the voltage output as a function of displacement. At very close

    distances, the receiving fibres can capture most of the reflected light, and therefore the sensitivity is

    highest (the slope of the graph is largest). This linearity only lasts for a certain range; after a certain

    displacement, the voltage output decreases dramatically due to the inverse square law. Reflected light

    intensity also depends on the surface material of the object.

    Inductive proximity sensor:

    An important property of this device relies on the reluctance Rm of a magnetic circuit. This reluctance

    determines how much flux is produced for a given current in a coil. Therefore, this transducer uses an E-

    shaped piece with a coil wrapped around, as seen in the following diagram. With an AC source, the coil

    produces an alternating magnetic flux, which permeates through the air gap and the plate which is to be

    moved. When this plate does move, the magnetic flux will change and therefore the current will change

    (since the reluctance remains constant). The change in current is outputted as a voltage by a control box.

  • 8/10/2019 Lab 7 Transducers

    2/14

    Given the following formula:

    =

    I(t) = L Imax cos ( t) and =2

    N= number of turns of wire

    Capacitive proximity sensor:

    The capacitance of a capacitor depends on d, the separation of the plates; A the overlapping area; and

    the dielectric constant Ke . To measure small linear displacements between the plates, A and K are kept

    constant.

    As the capacitance changes with displacement, a simple circuit can be set up to output voltages that

    depend on C.

    Objectives:

    In the first part, we will calibrate the voltage outputs simultaneously for each transducer for many small

    displacements. Once our data is recorded, we will graph our results and analyze the response curves.From these graphs we can identify the three transducers respective linear ranges, their corresponding

    linearity, and the sensitivity which they produce. We will then compare these values for each transducer

    to determine which is most suitable for a particular measurement range. Knowing this is essential for

    any future experiments requiring the most accurate measurements for small linear displacements.

  • 8/10/2019 Lab 7 Transducers

    3/14

    Equipment and procedure

    - One X-Y sliding table equipped with micrometers. A flatplateis mountedon it.

    - One capacitive displacement transducerBC10-G30-Y0X

    - One inductivedisplacementtransducerBI10-G30-Y1- One Photonicsystem. Please be aware the fiber opticmustnot be bent.

    - All transducersare mountedon a fixed support.

    - Power supply.

    - One DVM

    - Three signal conditioners (Photonic: 3B11-01, Capacitive: 3B31-03, Inductive: 3B41-03).

    - LabVIEW virtual digital voltmeters.

    Procedure

    1. First check the lab setup. Make sure that the transducers are properly connected to their

    respectivecontrolboxes and power supplies. Record the model and gain of each signal

    conditionerused to amplify the signal of each transducer.

    2. Use LabVIEW to record the measurements. The program has already been created for you

    and is available on WebCT. Copy the file Lab7.vi to your desktop. Also download the excel file

    named Lab7.xls. This excel file containsthe tableyou will use to record the data.

    3. Turn on thepower supply and thephotoniccontroller(the switch is on the back of thebox).

    Use the digital voltmeter to measure and record the supply voltage. You will need this value to

    calculatethe static sensitivity of the Inductive and Capacitive transducers.

    4. The LabVIEW program allows you to record three voltage signals simultaneously. Specify

    the minimum and maximum limits as -1 and +10 respectively, and the channel numbers

    corresponding to the connections of your transducers (drop the arrow, click Browse, press the

    CTRL key and click on the 3 channels in the following order: 10,12,14). After this, you should

    see in the Physical channels box display DEV1/ai10, DEV1/ai12, DEV1/ai14. If you dont

    see that repeattheprevious step.

    5. Using the X-axis micrometer, adjust the flatplate position until the micrometer reads zero.

    This plate position should be close to, but not touching,the transducers.

    6. Move away from theplatein incrementscorresponding to the values given in the Lab 7 spread-sheet. The resolution of the micrometer is 0.01 mm. At each displacement, run the LabVIEW

    program once and record the voltage readings in your table. You will take approximately 70

    readings. The number of readings is the minimum needed to fully resolve the response curves of

    each transducer.

    7. When you finish collecting the data turn off the power supply, the photoniccontroller and the

    backplane.

  • 8/10/2019 Lab 7 Transducers

    4/14

    Results and Discussion:

    Inductive: 3B41-03 Photonic: 3B11-01 Capacitive: 3B31-03 Gain=1.0 for all three transducers, supply voltage = 5.0446 V

    Figure 1. Response curve of three transducers; photonic, capacitive, and inductive sensor

    -1.0

    0.0

    1.0

    2.0

    3.0

    4.0

    5.0

    6.0

    7.0

    8.0

    9.0

    10.0

    0.0 2.0 4.0 6.0 8.0 10.0

    output(volts)

    displacement (mm)

    Response Curve of Three Transducers

    Photonic

    Capacitive

    Inductive

  • 8/10/2019 Lab 7 Transducers

    5/14

    1) Photonic Sensor

    Table 1. Range of readings with high sensitivity and an approximate linear response

    Figure 2. Calibration plot for photonic readings with linear best fit line (Sensitivity = 9.947)

    Displacement

    (mm)

    Photonic reading

    (volts)0.2 0.942383

    0.225 1.235352

    0.25 1.28418

    0.275 1.713867

    0.3 1.972656

    0.325 2.006836

    0.35 2.485352

    0.375 2.807617

    0.4 2.856445

    0.475 3.520508

    0.55 4.560547

    0.625 5.253906

    0.7 5.776367

    V = 9.947*d - 1.0628

    R = 0.9949

    0

    1

    2

    3

    4

    5

    6

    7

    0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8

    Output(volts)

    Displacement (mm)

    Calibration Plot for Photonic Reading

  • 8/10/2019 Lab 7 Transducers

    6/14

    Displacement(mm)

    Points on the lineof best fit (volts)

    Residual(volts)

    0.2 0.9266 0.015783

    0.225 1.175275 0.060077

    0.25 1.42395 -0.13977

    0.275 1.672625 0.041242

    0.3 1.9213 0.051356

    0.325 2.169975 -0.163139

    0.35 2.41865 0.066702

    0.375 2.667325 0.140292

    0.4 2.916 -0.059555

    0.475 3.662025 -0.141517

    0.55 4.40805 0.152497

    0.625 5.154075 0.099831

    0.7 5.9001 -0.123733

    Table 2. Points on the line of best fit and residual values of photonic readings

    Residual values for corresponding displacements can be identified through two ways. 1) By

    subtracting points on the line of best fit from photonic reading 2) Using data analysis on excel

    Figure 3. Residual data for photonic readings with straight line connecting terminal points

    Displacement(mm)

    Residual(volts)

    0.2 0.015783

    0.225 0.060077

    0.25 -0.13977

    0.275 0.041242

    0.3 0.051356

    0.325 -0.163139

    0.35 0.066702

    0.375 0.140292

    0.4 -0.059555

    0.475 -0.141517

    0.55 0.152497

    0.625 0.099831

    0.7 -0.123733

    -0.2

    -0.15

    -0.1

    -0.05

    0

    0.05

    0.1

    0.15

    0.2

    0.2 0.3 0.4 0.5 0.6 0.7

    Deviation(volts)

    Displacement (mm)

    Residual Data for Photonic Reading

  • 8/10/2019 Lab 7 Transducers

    7/14

    The lower and upper accuracy limits of the response curve are -0.163139 volts and 0.152497

    volts, respectively, or -2.82% and 2.64% full scale. The full scale value of the response curve is

    taken at 0.7mm, therefore full scale value is 5.776367 volts.

    Equation of the line is V= -0.279032*d +0.0715894 (linear line of residual plot)

    Displacement (mm)Point on the line

    (volts)Residual

    (volts) q

    0.2 0.015783 0.015783 1.11022E-16

    0.225 0.0088072 0.060077 0.0512698

    0.25 0.0018314 -0.13977 0.1416014

    0.275 -0.0051444 0.041242 0.0463864

    0.3 -0.0121202 0.051356 0.0634762

    0.325 -0.019096 -0.16314 0.144043

    0.35 -0.0260718 0.066702 0.0927738

    0.375 -0.0330476 0.140292 0.17333960.4 -0.0400234 -0.05955 0.0195316

    0.475 -0.0609508 -0.14152 0.0805662

    0.55 -0.0818782 0.152497 0.2343752

    0.625 -0.1028056 0.099831 0.2026366

    0.7 -0.123733 -0.12373 3.33067E-16

    Table 3. Points on the line of residual plot, residual and q values

    |

    |

    = 0.2343752

    =1 || 100% =10.2343752

    5.776367100% = 95.94%

  • 8/10/2019 Lab 7 Transducers

    8/14

    2) Capacitive Sensor

    Table 4. Range of capacitive readings with high sensitivity and an approximate linear response

    Figure 4. Calibration plot for capacitive readings with linear best fit line (Sensitivity = -0.3332)

    Displacement(mm)

    Capacitivereading (vol ts)

    2.90 0.86426

    3.05 0.82520

    3.20 0.78125

    3.35 0.75195

    3.50 0.67871

    3.65 0.59082

    3.80 0.54688

    3.95 0.54199

    4.10 0.47852

    4.25 0.41016

    4.40 0.35645

    4.55 0.31250

    4.70 0.27344

    4.85 0.24902

    V = -0.3332*d+ 1.8383R = 0.9924

    0.0

    0.1

    0.2

    0.3

    0.4

    0.5

    0.6

    0.7

    0.8

    0.9

    1.0

    2.5 3.0 3.5 4.0 4.5 5.0

    Output(volts)

    Displacement (mm)

    Calibration Plot for Capacitive Reading

  • 8/10/2019 Lab 7 Transducers

    9/14

    Displacement(mm)

    Point on the line ofbest fit (volts)

    Residual(volts)

    2.90 0.87202 -0.00776

    3.05 0.82204 0.00315

    3.20 0.77206 0.00919

    3.35 0.72208 0.02987

    3.50 0.6721 0.00661

    3.65 0.62212 -0.03130

    3.80 0.57214 -0.02527

    3.95 0.52216 0.01983

    4.10 0.47218 0.00634

    4.25 0.4222 -0.01204

    4.40 0.37222 -0.01578

    4.55 0.32224 -0.00974

    4.70 0.27226 0.00118

    4.85 0.22228 0.02674

    Table 5. Points on the line of best fit and residual values of capacitive readings

    Figure 5. Residual data for capacitive readings with straight line connecting terminal points

    Displacement(mm) Residual (volts)

    2.90 -0.00776

    3.05 0.003155

    3.20 0.00919

    3.35 0.029873

    3.50 0.006611

    3.65 -0.0313

    3.80 -0.02527

    3.95 0.019832

    4.10 0.006336

    4.25 -0.01204

    4.40 -0.01578

    4.55 -0.00974

    4.70 0.001177

    4.85 0.026743

    -0.04

    -0.03

    -0.02

    -0.01

    0

    0.01

    0.02

    0.03

    0.04

    2.9 3.1 3.3 3.5 3.7 3.9 4.1 4.3 4.5 4.7

    Deviation(volts)

    Displacement (mm)

    Residual Data for Capacitive Reading

  • 8/10/2019 Lab 7 Transducers

    10/14

    The lower and upper accuracy limits of the response curve are -0.0313 volts and 0.029873 volts,

    respectively, or -3.62% and 3.46% full scale. The full scale value of the response curve is taken

    at 2.9mm, therefore full scale value is 0.864258 volts.

    Equation of the line is V=0.017693846*d 0.059072153 (linear line of residual plot)

    Displacement(mm)

    Point on the line(volts)

    Residual(volts) q

    2.90 -0.00776 -0.00776 2.0004E-06

    3.05 -0.005105923 0.003155 0.008260923

    3.20 -0.002451846 0.00919 0.011641846

    3.35 0.000202231 0.029873 0.029670769

    3.50 0.002856308 0.006611 0.003754692

    3.65 0.005510385 -0.0313 0.036810385

    3.80 0.008164462 -0.02527 0.033429462

    3.95 0.010818539 0.019832 0.009013461

    4.10 0.013472616 0.006336 0.007136616

    4.25 0.016126693 -0.01204 0.028170693

    4.40 0.018780769 -0.01578 0.034555769

    4.55 0.021434846 -0.00974 0.031174846

    4.70 0.024088923 0.001177 0.022911923

    4.85 0.026743 0.026743 1E-10

    Table 6. Points on the line of residual plot, residual and q values

    || = 0.036810385 =1 || 100% =1

    0.036810385

    0.864258 100% = 95.74%

  • 8/10/2019 Lab 7 Transducers

    11/14

    3) Inductive Sensor

    Displacement

    (mm)

    Inductive reading

    (volts)5.45 1.01563

    5.60 1.25488

    5.75 1.53809

    5.90 1.85059

    6.00 2.06055

    6.50 2.92481

    Table 7. Range of inductive readings with high sensitivity and an approximate linear response

    Figure 6. Calibration plot for inductive readings with linear best fit line (Sensitivity = 1.845)

    V = 1.845*d - 9.0501

    R = 0.9985

    0.0

    0.5

    1.0

    1.5

    2.0

    2.5

    3.0

    3.5

    5.2 5.4 5.6 5.8 6.0 6.2 6.4 6.6

    Output(volts)

    Displacement (mm)

    Calibration Plot for Inductive Reading

  • 8/10/2019 Lab 7 Transducers

    12/14

    Displacement(mm)

    Point on the line ofbest fit (volts)

    Residual(volts)

    5.45 1.00515 0.01047

    5.60 1.2819 -0.02702

    5.75 1.55865 -0.02056

    5.90 1.8354 0.01519

    6.00 2.0199 0.04065

    6.50 2.9424 -0.01759

    Table 8. Points on the line of best fit and residual values of inductive readings

    Figure 7. Residual data for inductive readings with straight line connecting terminal points

    The lower and upper accuracy limits of the response curve are -0.027017 volts and 0.040647

    volts, respectively, or -0.92% and 1.39% full scale. The full scale value of the response curve is

    taken at 6.5mm, therefore full scale value is 2.92481 volts.

    Equation of the line is V=0.017693846*d 0.059072153 (linear line of residual plot)

    Displacement (mm)Residual

    (volts)

    5.45 0.010475

    5.60 -0.027017

    5.75 -0.020564

    5.90 0.015186

    6.00 0.0406476.50 -0.017595

    -0.04

    -0.03

    -0.02

    -0.01

    0

    0.01

    0.02

    0.03

    0.04

    0.05

    5.45 5.65 5.85 6.05 6.25 6.45Deviation(volts)

    Displacement (mm)

    Residual Data for Inductive Reading

  • 8/10/2019 Lab 7 Transducers

    13/14

    Displacement (mm) Point on the line (volts) Residual (volts) q

    5.45 0.010475001 0.010475 1.15E-09

    5.60 0.006465001 -0.02702 0.033482

    5.75 0.002455001 -0.02056 0.023019

    5.90 -0.001554999 0.015186 0.016741

    6.00 -0.004228332 0.040647 0.044875

    6.50 -0.017594999 -0.01759 1.5E-09

    Table 9. Points on the line of residual plot, residual and q values

    || = 0.044875 =1 || 100% =1

    0.044875

    2.92481100% = 98.47%

    From the readings generated from each transducer, usable range for the transducers which has a

    high sensitivity and an approximately linear response are identified. From the linear best fit line

    superimposed on the calibration plot of each transducer, the sensitivity of each transducer was

    identified. Sensitivity of photonic, capacitive, inductive transducer is 9.947, -0.3332, and 1.845

    V/mm, respectively. Therefore, it can be seen that inductive transducer has the highest

    sensitivity. Therefore, inductive transducer is most responsive to small changes in displacement

    as the equation for the sensitivity is, = = .Linearity of each transducer was identified using the equation:

    =1 || 100%, whereas ||were found by observing Table 3, Table 6,andTable 9. Linearity of photonic, capacitive, and inductive transducer is 95.94%, 95.74%, and

    98.47% respectively.

    Usable range of photonic, capacitive, and inductive transducer is 0.2mm to 0.7mm, 2.9mm to

    4.85mm, and 5.45mm to 6.50mm respectively. This means there are number of gaps in the range

    of the displacement given by the micrometer that are not within usable ranges of the transducers.

    These are: 0 to 0.2mm, 0.7mm to 2.9mm, 4.85mm to 5.45mm, and 6.50mm to 10mm

    To account for these gaps and have accurate measurements, there is one method I can think of.

    We introduce a new reference point from which to measure; it can be a metal plate at an arbitrary

    point. However this reference point must be within an existing usable range of one of the

    devices. Therefore we can measure its distance relative to the transducer. Then we measure the

    distance from the object to the transducer (object must be within usable range). Finally, we

  • 8/10/2019 Lab 7 Transducers

    14/14

    subtract these two distance values to determine the displacement of the object from the new

    reference plate. Thickness of the reference plate must be known.

    For all three transducers, our results showed very high linearity over their usable ranges.

    Although each device had its own unique usable range, we are confident they will output

    accurate measurements over that range. When it comes to high sensitivity and high linearity overtheir usable range, the inductive sensor was superior to the other two. However, the capacitive

    sensor had a larger usable range by 186% compared to that of the inductive sensor, with only a

    linearity of 3% lower. Therefore, determining which transducer is the best depends on the

    measurement range of the experiment, and the small changes in displacement that need to be

    measured (the resolution), which relies on the sensitivity.

    Conclusion

    There are three main objectives we have learned from this lab. The first is how each of the three

    types of transducers work. The second is how to perform a static calibration using a micrometer

    and setting up the virtual instruments in LabView. The third is how to do a statistical analysis of

    the response curves obtained from the calibration. In other words, how to determine linear range;

    calculate sensitivity and linearity from a residual plot using Microsoft Excel. Once weve

    obtained these values for each transducer, we can appropriately and effectively use these devices

    for future experiments. Knowing this information is essential for future experiments, where we

    must decide which transducer is most suitable for a particular displacement range. The most

    suitable choice of transducer will have the largest linear range over which we are measuring, the

    highest percentage linearity, and the highest sensitivity within that range. From our results, we

    can conclude that the inductive sensor has the highest sensitivity and linearity over its usable

    range. But most importantly, if we are measuring displacements outside its usable range, then

    another transducer, such as capacitive or photonic, would be much more accurate.

    Reference:

    Wheeler, Anthony J., and A. R. Ganji.Introduction to Engineering Experimentation. Upper

    Saddle River: Pearson/Prentice Hall, 2004. Print.