Keep Control: PULPO 2 ESO‘s new housekeeping unit Christoph Geimer, Claudio Cumani, Nicolas Haddad, Javier Reyes, Javier Valenzuela, Bernhard Lopez PULPO 1 BACK PANEL PULPO 1 FRONT PANEL • 6 temperature sensors PT100 • 3 independent heater circuits • Shutter control • Vacuum monitoring • 1 alarm output • simple RS 232 interface • 33 units already build, very low failure rate • Standard housekeeping unit at all optical ESO instruments THE BEST THING WE HAD TO OFFER - PULPO 1 NEW CHALLENGES - OmegaCAM • PULPO 1 functionality plus: • much more sensors (at least • more alarm outputs • data logging • increased safety features
Keep Control: PULPO 2 ESO‘s new housekeeping unit. Christoph Geimer, Claudio Cumani, Nicolas Haddad, Javier Reyes, Javier Valenzuela, Bernhard Lopez. THE BEST THING WE HAD TO OFFER - PULPO 1. 6 temperature sensors PT100 3 independent heater circuits Shutter control Vacuum monitoring - PowerPoint PPT Presentation
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Keep Control: PULPO 2ESO‘s new housekeeping unit
Christoph Geimer, Claudio Cumani, Nicolas Haddad, Javier Reyes, Javier Valenzuela, Bernhard Lopez
PULPO 1 BACK PANELPULPO 1 FRONT PANEL
• 6 temperature sensors PT100• 3 independent heater circuits• Shutter control• Vacuum monitoring• 1 alarm output• simple RS 232 interface• 33 units already build, very low failure rate• Standard housekeeping unit at all optical ESO instruments
THE BEST THING WE HAD TO OFFER - PULPO 1
NEW CHALLENGES - OmegaCAM
• PULPO 1 functionality plus:• much more sensors (at least • more alarm outputs • data logging• increased safety features
“Rule with your heartand live with your conscience”
- QUEEN
Interfacing to the world
• 29 PT100 temperature sensors by default, up to 125 via external multiplexers• digital I/O control and power supply for external MUX board• Vacuum monitoring (Balzers or Edwards gauges)•seperated alarm outputs for vacuum and temperature (2 for each)• fiber interface to SLCU• additional standard RS 232 interface• Ethernet connection for log data download and • OPTO isolated shutter interface ( 1output, 4 inputs,2spares)• 8 independent heater circuits (72W, +-0.1K accuracy) • data logging capabilities with 65000 records• self recovery
“In the quiet of the night - let our candle always burn”
- QUEEN
PULPO 2
Housekeeping the universePULPO 2 in action
Instrument connection point
Temperature Alarm
Vacuum Alarm
SLCU