Top Banner
EMPIR 14SIP01 Workshop Dr. Karl Jousten, PTB, Vacuum Metrology 1 Jan 30 to Feb 1, 2017 ISO TS 20177 Procedures to measure and report outgassing rates Introduction 1. Purpose and scope 2. Structure of content
22

ISO TS 20177 Procedures to measure and report outgassing rates

Dec 18, 2021

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: ISO TS 20177 Procedures to measure and report outgassing rates

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology 1 Jan 30 to Feb 1, 2017

ISO TS 20177 Procedures to measure and

report outgassing rates

Introduction

1. Purpose and scope

2. Structure of content

Page 2: ISO TS 20177 Procedures to measure and report outgassing rates

2

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

qout ≤ 1·10-5 Pa m³/s (-> discussion later) Total outgassing rate and species outgassing rate For total outgassing rate: concept of nitrogen equivalent (-> discussion later) No direct line of sights required (-> discussion later) Methods described in TS shall be comparable Traceability a must for each method Relatively high uncertainties acceptable (more than 10% o.k.)

Scope/suppositions

Page 3: ISO TS 20177 Procedures to measure and report outgassing rates

3

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

Structure of Content

1. Scope2. Normative references3. Terms and definitions4. Symbols and abbreviated terms

Obligatory from ISO template

9. References10. Appendix

5. Measurement systems6. Measurement procedures7. Measurement uncertainties8. Reporting results

Flexible

Page 4: ISO TS 20177 Procedures to measure and report outgassing rates

4

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

5. Measurement systems

1. General recommendations for systems, chambers and pumps, gauges)

1. Recommendations for systems2. Chambers and pumps3. Vacuum gauges

2. Systems applying the throughput method

1. Continuous expansion system as flow comparator2. Throughput system with calulated conductance element

(one and two path design) 3. Throughput system with measured effective pumping speed4. Throughput system with modulated conductance

3. Accumulation systems

1. Basic accumulation system2. Accumulation system with gas analysis

Page 5: ISO TS 20177 Procedures to measure and report outgassing rates

5

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

6. Measurement procedures

1. Recommended sample preparation 2. Course and time period of measurement3. Particular measurement procedures

1. Procedure with continuous expansion system as flow comparator

2. Procedure with throughput system with calulated conductance element

3. Procedure with throughput system with measured effective pumping speed

4. Procedure with throughput system with modulated conductance

5. Procedure with accumulation systems

Page 6: ISO TS 20177 Procedures to measure and report outgassing rates

6

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

6. Measurement uncertainties

• Continuous expansion system as flow comparator• Throughput system with calulated conductance element • Throughput system with measured effective pumping speed• Throughput system with modulated conductance• Basic accumulation system• Accumulation system with gas analysis

Editorial: Numbering requested?

Page 7: ISO TS 20177 Procedures to measure and report outgassing rates

7

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

Measurement uncertainties: List

Method Uncertainty contribution to be

considered

5.2.1 Continuous expansion system

as flow comparator

throughput (flow in) qpV,i

temperature T

5.2.2 Throughput system with

calculated conductance element (one

and two path design)

measured pressure p1 by gauge

(uncertainty in calibration certificate,

long-term instability, non-linearity,

correction factor for other species than

nitrogen)

estimated or measured (by gauge)

pressure p2 (one path design only)

conductance C

Page 8: ISO TS 20177 Procedures to measure and report outgassing rates

8

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

Measurement uncertainties: List

Method Uncertainty contribution to be

considered

5.2.3 Throughput system with

measured effective pumping speed

flow from standard leakpossible change of CN2 due to unknown change with pressuremeasured pressure p1 by gauge (uncertainty in calibration certificate, long-term instability, non-linearity, correction factor for other species than nitrogen)measured pressure p0 by gaugePosition of gauge or QMSQMS sensitivities

Page 9: ISO TS 20177 Procedures to measure and report outgassing rates

9

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

Measurement uncertainties: List

Method Uncertainty contribution to be

considered

5.2.4 Throughput system with

modulated conductance

measurement with vacuum gauge

(uncertainty in calibration certificate,

long-term instability, non-linearity,

correction factor for other species than

nitrogen)

conductances

incomplete equilibrium

Page 10: ISO TS 20177 Procedures to measure and report outgassing rates

10

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

Measurement uncertainties: List

Method Uncertainty contribution to be considered

5.3.1 Basic accumulation system accumulation chamber and sample volume V

temperature T

pressure of accumulating gas p

time interval Δt

5.3.2 Accumulation system with

gas analysis system

accumulation chamber and sample volume V,

temperature T

pressure of accumulating gas p

time interval Δt

volume of optional chamber for calibration

gas V

pressure of calibration gas p

nonlinearity of the QMS

Page 11: ISO TS 20177 Procedures to measure and report outgassing rates

11

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

5.2.1 Continuous expansion system as flow comparator1 measurement chamber 2 pump chamber 3 by-pass (recommended) 4 sample chamber 5 load-lock 6 sample 7 QMS 8 total pressure gauge 9 pump systems 10 reservoir 11 CDG 12 connection to gas supply

Page 12: ISO TS 20177 Procedures to measure and report outgassing rates

12

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

5.2.2 Throughput system

with calculated

conductance (one-path

design)

1 measurement chamber 2

pump chamber 3 vacuum

gauge and QMS (optionally)

4,10 valves 5 by-pass

(optionally) 6 sample 7

sample chamber 8 load lock

9 pump systems

Page 13: ISO TS 20177 Procedures to measure and report outgassing rates

13

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

5.2.2 Throughput system

with calculated

conductance (two-path

design)

1 measurement chamber 2

pump chamber 3,4 vacuum

gauge and QMS (optionally)

5 by-pass 6 sample 7 sample

chamber 8 load lock 9 pump

systems 10 second path line

B 11 valve for path A

(upstream), 12 valve for path

B (downstream)

Page 14: ISO TS 20177 Procedures to measure and report outgassing rates

14

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

5.2.3 Throughput system with measured

effective pumping speed

1 measurement chamber 2 standard leak 3

vacuum gauge and QMS (optionally) 4 pump

systems 5 sample chamber 6 sample 7 load

lock

Page 15: ISO TS 20177 Procedures to measure and report outgassing rates

15

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

5.2.4 Throughput system with modulated conductance

1 measurement chamber 2 element with first conductance and

holder 3 element with second conductance 4 vacuum gauge and

QMS (optionally) 5 linear feedthrough 6 pump systems 7 sample 8

sample chamber 9 load lock

Page 16: ISO TS 20177 Procedures to measure and report outgassing rates

16

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

5.3.1 Basic accumulation system

1 measurement chamber of known volume 2 inert vacuum gauge

3 time measurement 4 pump systems 5 load lock 6 sample

Page 17: ISO TS 20177 Procedures to measure and report outgassing rates

17

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

5.3.2 Accumulation system

with gas analysis system

1 measurement chamber of

known volume 2 inert vacuum

gauges 3 time measurement 4

pump systems 5 load lock 6

sample 7 QMS 8 Ionization

vacuum gauge 9 volume for

calibration gas (may also be

connected to chamber 1) 10 to

gas reservoir

Page 18: ISO TS 20177 Procedures to measure and report outgassing rates

18

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

System Vapour

outgassing

Outgassing

is time

dependent

Identifies

species

Outgassing

rate > 10-9 Pa

m³ s-1

Outgassing

rate 10-9 Pa

m³ s-1

Accuracy Effort,

Experience

Sample is

chamber

5.2.1 Continuous

expansion system as

flow comparator

+ ++ ++ + - ++ Very high +

5.2.2 Throughput

system with

calculated

conductance element

(one path design)

+ ++ ++ + + + high +

5.2.2 Throughput

system with

calculated

conductance element

(two path design)

+ - ++ + ++ + high +

5.2.3 Throughput

system with

measured effective

pumping speed

+ ++ - + - -- low +

5.2.4 Throughput

system with

modulated

conductance

- - ++ + ++ + high +

5.3.1 Basic

accumulation system

-- - -- + ++ + low ++

5.2.3 Accumulation

system with gas

analysis system

-- - + + ++ + high +

Table 3 — Applicability and characteristic of the different measurement systems. It means: ++ well suited, + suitable, - less suitable , --- not suitable

Page 19: ISO TS 20177 Procedures to measure and report outgassing rates

19

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

Method Traceable quantity Unit Traceable device Note

5.2.1 Continuous expansion system as flow comparator

C m³ s-1 = 10³ L s-1

One or several conductance elements

Specific for gas species

pi Pa

Vacuum gauge for medium and rough vacuum, preferentially with gas independent indication

Pressure of gas species i in reservoir, measured by secondary standard

5.2.2 Throughput system with calculated conductance element (one and two path design)

C m³ s-1 = 10³ L s-1 Orifice or short duct Specific for gas

species

p PaVacuum gauge for high and ultrahigh vacuum

Calibrated for nitrogen

pi Pa Quadrupol mass spectrometer

Optional, sensitivity for gas species i , stability check by sensitivity nitrogen

Table 4 — Quantities, by which traceability of outgassing rate (either total or partial) to the SI is established in the different measurement systems

Page 20: ISO TS 20177 Procedures to measure and report outgassing rates

20

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

Method Traceable quantity Unit Traceable device Note

5.2.3 Throughput system with measured effective pumping speed

qpV Pa m³ s-1 Nitrogen standard leak To determine Ceff

p PaVacuum gauge for high and ultrahigh vacuum

Calibrated for nitrogen

qpV,i Pa m³ s-1 Standard leak for gas species i

Option, if QMS is used, to determine Ceff,i

5.2.4 Throughput system with modulated conductance

C m³ s-1 = 10³ L s-1

Two orifices or short ducts

Specific for gas species

p PaVacuum gauge for high and ultrahigh vacuum

Calibrated for nitrogen

pi Pa Quadrupol mass spectrometer

Optional, sensitivity for gas species i , stability check by sensitivity nitrogen

Page 21: ISO TS 20177 Procedures to measure and report outgassing rates

21

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

Method Traceable quantity Unit Traceable device Note

5.3.1 Basic accumulation system

V m³ Chamber volume

p PaVacuum gauge for high and medium vacuum

Calibrated for nitrogen

t s clock

Page 22: ISO TS 20177 Procedures to measure and report outgassing rates

22

ISO TS 20177 (WD)

EMPIR 14SIP01 Workshop – Dr. Karl Jousten, PTB, Vacuum Metrology Jan 30 to Feb 1, 2017

Method Traceable quantity Unit Traceable device Note

5.2.3 Accumulation system with gas analysis system

V m³ Chamber volume

p PaVacuum gauge for high and medium vacuum

Calibrated for nitrogen

t s clock

pi Pa Quadrupole mass spectrometer

Sensitivity for gas species i