Irradiator Technology HR T D Human Resources Training & Development Self-Shielded Irradiator Safety July 2011- Slide 1 of 55 Irradiator Technology Course (H-315) Safety Inspections & Licensing
Irradiator Technology HRTDHuman ResourcesTraining & Development
Self-ShieldedIrradiator
Safety
July 2011- Slide 1 of 55Irradiator Technology Course (H-315)
Safety Inspections& Licensing
Self-Shielded Irradiators
ANSI definition of a Category I irradiator
Irradiator in which sealed sources are
July 2011- Slide 2 of 55Irradiator Technology Course (H-315)
completely contained in a dry container constructed of solid materials, the sealed sources are shielded at all times, and human access to the sealed sources and the volumes undergoing irradiation is not physically possible in its designed configuration.
Self-Shielded Irradiators
Are
July 2011- Slide 3 of 55Irradiator Technology Course (H-315)
Inherently Safe
Self-Shielded Irradiators
• May Contain Thousands of Curies
• Designed to Be Placed in Unshielded R
July 2011- Slide 4 of 55Irradiator Technology Course (H-315)
Rooms
• Very Low Doses to Authorized Users
Examples of Self-Shielded Irradiators
July 2011- Slide 5 of 55Irradiator Technology Course (H-315)
Gammacell 40Exactor
Gammacell 1000IBL-437C
Examples of Self-Shielded Irradiators
July 2011- Slide 6 of 55Irradiator Technology Course (H-315)
J. L. Shepherd Mark IGammacell 220
Examples of Irradiators (Calibrators)
July 2011- Slide 7 of 55Irradiator Technology Course (H-315)
J. L. Shepherd 89-400 FEMA CDV-794
Example of a Movable Source Irradiator
• Sources are mounted on shielded operating rods
• Sources moved up from the “shielded” to the “irradiate” position
July 2011- Slide 8 of 55Irradiator Technology Course (H-315)
• Door is interlocked to prevent opening when source is in sample chamber
• Bypassing or failure of interlock can result in high radiation levels
J. L. Shepherd Mark I
Example of Self-Shielded Irradiator With Fixed (Non-Movable) Sources
• Sources are fixed and housed within the radiation shield.
• The shield contains a rotor that is operated by an electric drive.
July 2011- Slide 9 of 55Irradiator Technology Course (H-315)
The sample chamber is contained within the rotor and has an electric turntable. During operation, the rotor turns 180 degrees and the sample chamber is exposed to radiation. Best Theratronics Ltd.Gammacell 1000
Uses
• Irradiate Blood Products In Order To Prevent Graft-Versus-Host Disease (GVHD) in Immune Compromised Patients– Donor (Graft) T-cells may perceive the
R i i t (H t) f i d tt k
July 2011- Slide 10 of 55Irradiator Technology Course (H-315)
Recipient (Host) as foreign and attack • Irradiate Small Animals (e.g., for stem cell
research)• Irradiate Biological Samples• Sterilize insects
Self-Shielded Irradiators
• Typically located at– Blood Banks– Hospitals
U i iti
July 2011- Slide 11 of 55Irradiator Technology Course (H-315)
– Universities– Research Labs
Self-Shielded Irradiator Safety Inspections
• Generally, straight forward and uneventful• Typically located Indoors at a medical facility or
research lab• Majority of licensee’s program is within one room
July 2011- Slide 12 of 55Irradiator Technology Course (H-315)
• Not a lot of complicated procedures or regulations
• Emergency response actions are common sense (i.e., leave area and control access)
• Can typically be completed within several hours
Self-Shielded Irradiator Safety Inspections
• NRC Inspection Procedure – 87122, “Irradiator Programs”
• Determine if licensed activities are being conducted in a manner that will protect the health
July 2011- Slide 13 of 55Irradiator Technology Course (H-315)
and safety of workers and the general public• Determine if licensed activities are being
conducted in accordance with regulatory requirements
Inspection Preparations
• Review Docket File– Incidents, Follow-up Items, Inspection and
Enforcement History, and Violations for Closeout
• Review SSDR
July 2011- Slide 14 of 55Irradiator Technology Course (H-315)
– Maximum Source Activity– Maximum External Dose Rates– Details of Device Operation
• Review NMED Events• Be aware that J. L. Shepherd Mark I and 81-22
irradiators are subject to enhanced monitoring requirements
Inspection Focus Areas
• 1) Security & control of licensed materials• 2) Shielding• 3) Comprehensive safety measures• 4) Radiation dosimetry program
July 2011- Slide 15 of 55Irradiator Technology Course (H-315)
4) Radiation dosimetry program• 5) Radiation instrumentation and surveys• 6) Radiation safety training and practices• 7) Management oversight
Gather Information on Scope of Program
• Type of Irradiator(s)?• Irradiators IAW material authorizations in the
license?• No. of curies (or Bequerels;o)
July 2011- Slide 16 of 55Irradiator Technology Course (H-315)
• Irradiator Uses IAW Material License?• Frequency of use?• RSO is the same as the license? • Number of Authorized Users?
Location of Use, Comprehensive Safety Measures, & General Access Controls*
• Location(s) of use IAW license?• Licensee evaluated floor loading?• Sources are adequately protected from fire,
hazardous chemicals, industrial hazards, the
July 2011- Slide 17 of 55Irradiator Technology Course (H-315)
elements• Licensee limits unescorted access* to only
authorized and trained personnel?– Physical barriers and key control?– Trained Personnel?
* Specific access controls are evaluated in detail during an IC inspection
Shielding
• Perform independent radiation surveys to evaluate adequacy of shielding– Irradiator dose rates IAW SSDR?– Irradiator room dose rates?
July 2011- Slide 18 of 55Irradiator Technology Course (H-315)
– Areas adjacent to irradiator?
Personnel Monitoring
• Licensee issues NVLAP Dosimetry to AUs• A review of dosimetry results shows that radiation
exposures are ALARA• OR
July 2011- Slide 19 of 55Irradiator Technology Course (H-315)
• Licensee maintains a prospective evaluation that demonstrates that unmonitored individuals are not likely to receive, in one year, a radiation dose in excess of 10% of the allowable limits in 10 CFR 20
• The licensee’s prospective evaluation is reasonable
Personnel Monitoring and Public Dose
• Licensee has properly investigated any unusual dosimetry results?
• Licensee advises workers who are required to be monitored, and whose exposure exceeds 100
f th i di ti ?
July 2011- Slide 20 of 55Irradiator Technology Course (H-315)
mrem, of their radiation exposures?• Members of the public will not exceed 1 mSv (100
mrem) in one year, AND dose in an unrestricted area does not exceed 2 mrem in any one hour?
• Licensee should maintain documentation demonstrating compliance with the public dose limit?
Radiation Monitoring Instrumentation
• Licensee possesses or has access to properly calibrated radiation survey instruments
• Due to self-shielding design features, it is not required for a licensee to possess a survey i t t H th d t
July 2011- Slide 21 of 55Irradiator Technology Course (H-315)
instrument. However, they may need one to – Verify source integrity– Verify occupational dose w/I regulatory limits– Verify public dose w/I regulatory limits
• It is a good practice to possess a survey meter and use an area radiation monitor
J. L. Shepherd Mark I or 81-22 Irradiators
• An interlock failed on a J. L. Shepherd Mark I irradiator that allowed the shielded door to be opened when the sources were positioned in the sample
July 2011- Slide 22 of 55Irradiator Technology Course (H-315)
were positioned in the sample chamber
• This could subject the user to substantial radiation dose
J. L. Shepherd Mark I or 81-22 Irradiators
• NRC Issued an Order to Amend All J. L. Shepherd Mark I or Model 81-22 licensees to add additional monitoring requirements
July 2011- Slide 23 of 55Irradiator Technology Course (H-315)
Additional Monitoring Requirements for J. L. Shepherd Mark I or 81-22 Irradiators
• Irradiator cannot be used unless licensee provides a calibrated and operable radiation survey meter OR a room monitor
• Requires operator to visually verify that the source position is in safe condition before opening door
July 2011- Slide 24 of 55Irradiator Technology Course (H-315)
• Each room monitor (if used)– Operable at all times when irradiator is in use– Activates visual and audible alarms when radiation
levels exceed 2 mR/h– Properly located to detect radiation escaping from door– Visible to AU when next to the irradiator
Additional Monitoring Requirements for J. L. Shepherd Mark I or 81-22 Irradiators
• If a room monitor is not installed, a survey meter shall be used to– Check radiation level at door when door is closed– Check door for increase in radiation level when
July 2011- Slide 25 of 55Irradiator Technology Course (H-315)
door is opened• Operator shall position body to minimize dose• If abnormal radiation or malfunction occurs,
immediately notify regulatory authority• Do not attempt or authorize repair unless authorized
by license
Training
• Authorized users have been adequately trained in and are knowledgeable:– Device operation– Use of radiation survey meters
July 2011- Slide 26 of 55Irradiator Technology Course (H-315)
– Techniques to limit their exposure– Operating and emergency procedures– Access control procedures, and– Their responsibility to control access to the
device.
Radiological Postings
• Licensee uses appropriate radiological postings to alert personnel to the presence of radioactive material
• NOTE: For security purposes, to minimize unwarranted attention, the licensee may elect to remove Caution posting from outer door and move it to inside the
July 2011- Slide 27 of 55Irradiator Technology Course (H-315)
irradiator room• 10 CFR 20.1903 Posting Exemption – room or area is not
required to be posted with a Caution sign due to the presence of a sealed source, provided radiation is
Leak Tests & Inventory
• Sealed sources are leak tested are performed as prescribed by license condition.– 6 months when in use– 10 years when in storage
July 2011- Slide 28 of 55Irradiator Technology Course (H-315)
• Licensee conducts a physical inventory as prescribed by license condition– Typically, six-month frequency– Record: Manufacturer’s name, model no.,
radionuclide, quantity, and date of inventory
Irradiator Maintenance, Repair, or Alteration
• Determine if licensee has performed non-routine maintenance, and if so, verify that licensee is authorized to perform non-routine maintenance
• Licensee must have specific authorization to repair, remove, replace or alter:
July 2011- Slide 29 of 55Irradiator Technology Course (H-315)
p , , p– Electrical or mechanical systems that control
shielding or source movement– Shielding or sealed sources– Safety interlocks, or– Anything that affects the safe operation of the
irradiator
Irradiator Maintenance, Repair, or Alteration
• If the licensee’s authorization is limited to the performance of routine maintenance, verify that the licensee follows the manufacturer’s procedures for routine:
Cl i d l b i ti
July 2011- Slide 30 of 55Irradiator Technology Course (H-315)
– Cleaning and lubrication,– Mechanical or electrical maintenance, or– Other non-safety critical repairs
Operating and Emergency Procedures
• Include an analysis of each type of material to be irradiated to determine compatibility with the irradiator design, and any special safety proceduresI l d i t ti f d ti
July 2011- Slide 31 of 55Irradiator Technology Course (H-315)
• Include instructions for use and routine maintenance, according to manufacturer’s instructions
• Include instructions for maintaining security• Include steps to maintain ALARA• Include instructions for maintaining
accountability
Operating and Emergency Procedures
• Include steps to control access to a malfunctioning or damaged irradiator
• Include steps to take and whom to contact when an irradiator is damaged or malfunctions
July 2011- Slide 32 of 55Irradiator Technology Course (H-315)
• Include specific operating procedures for irradiators with movable sources
• Are maintained current and provided to all users • Posted in irradiator room, or instructions for
access to the procedures
Emergency Procedures
• Common Sense & Include the following:– Leave irradiator room to reduce exposure– Control access– Contact individual responsible for program
July 2011- Slide 33 of 55Irradiator Technology Course (H-315)
p p gand to initiate emergency response
– Telephone numbers for Responsible Individual, Manufacturer, Security, Fire Department, Police, NRC or State Regulator
• Survey outside room to determine if further restrictions are required
Operating and Emergency Procedures
• Licensees can change their operating and emergency procedures only if:– The changes are reviewed and approved and
documented by licensee management
July 2011- Slide 34 of 55Irradiator Technology Course (H-315)
– Affected staff are trained in the procedures before implementation
– Changes are consistent with license conditions and commitments in license application, and
– The changes do not degrade safety
Management Oversight
• Licensee performs annual review of radiation protection program content and implementation
• Licensee has procedures for reporting defects• RSO is knowledgeable and ensures activities are
July 2011- Slide 35 of 55Irradiator Technology Course (H-315)
conducted IAW approved procedures and regulations
• RSO has sufficient authority, organizational freedom, time, resources, and management prerogative to identify radiation safety problems; initiate, recommend, or provide C/A; stop unsafe operations; and verify implementation of C/A
Self-Shielded Irradiator Shipment
July 2011- Slide 36 of 55Irradiator Technology Course (H-315)
Loading into Cask Ready To Ship
Examples of Self-Shielded Irradiator NMED Events: 2000 - 2011
• Elevated extremity exposure (27.8 rem) received due to source transfer rod becoming disengaged during source reload of a J.L. Shepherd 81 series irradiator.
• Leaking sealed source capsule (contaminated weld leached over time)• Source rack on a J.L. Shepherd 81-16A irradiator failed to retract due to reduced
air pressure in the operating piston due to water in compressor supply line.• Eldorodo irradi. shutter mechanism failed to close due to a faulty solenoid valve.
July 2011- Slide 37 of 55Irradiator Technology Course (H-315)
• 3 examples of door to J.L. Shepherd Model 68A Mark I failed to open due to a mechanical failure
• Sources in a J.L. Shepherd Mark I irradiator stuck in up position during testing due to a timer failure
• Multiple examples of badges with elevated readings that were determined to be dose to the badge only.
• Shepherd Mark I door and source interlock failures• Type B shipping container that did not meet the required Certificate of Compliance
Self-Shielded Irradiator Licensing
July 2011- Slide 38 of 55Irradiator Technology Course (H-315)
Good NewsLicensing is Not Difficult
• Follow Guidance in
NUREG 1556 V l 5
July 2011- Slide 39 of 55Irradiator Technology Course (H-315)
• NUREG 1556, Vol. 5
• As always, the application is easier to review if the licensee uses the checklist
Ensure Licensee
• Provides Device and Source Make & Model Nos.• Specifies Device with an Approved SSDR• Includes a Total Possession Limit IAW SSDR
July 2011- Slide 40 of 55Irradiator Technology Course (H-315)
Financial Assurance
July 2011- Slide 41 of 55Irradiator Technology Course (H-315)
• Typically, self-shielded Irradiator Licensees do not need Financial Assurance
Training Requirement is Easy to Meet
July 2011- Slide 42 of 55Irradiator Technology Course (H-315)
Training Requirement is Easy To Meet
July 2011- Slide 43 of 55Irradiator Technology Course (H-315)
Irradiator Training
July 2011- Slide 44 of 55Irradiator Technology Course (H-315)
Facility & Equipment
• Use Standard Boiler Plate Response• “We will ensure that where each self-shielded
irradiator is located corresponds to the – Conditions of Use (In SSDR)
July 2011- Slide 45 of 55Irradiator Technology Course (H-315)
Limitations and Other Conditions of Use From SSDR
July 2011- Slide 46 of 55Irradiator Technology Course (H-315)
Licensee to Ensure
• The floor beneath the Self-Shielded Irradiator is adequate to support the weight
• Each area where a Self-Shielded Irradiator is located is equipped with an automatically
t d fi d t ti d t l t
July 2011- Slide 47 of 55Irradiator Technology Course (H-315)
operated fire detection and control system (sprinkler, chemical, or gas)
• OR• The location of the area or other controls ensure
low level radiation risk attributed to fires.
Standard License Condition Required for Shepherd Mark I or 81-22 Irradiators
• Irradiator cannot be used unless licensee provides a calibrated and operable radiation survey meter or room monitor
• Requires operator to visually verify that the source position is in safe condition before opening door
July 2011- Slide 48 of 55Irradiator Technology Course (H-315)
• Each room monitor (if used)– Operable at all times when irradiator is in use– Activate visual and audible alarms when radiation levels
exceed 2 mR/h– Properly located to detect radiation escaping from door– Visible to AU when next to the irradiator
Standard License Condition Required for Shepherd Mark I or 81-22 Irradiators
• If a room monitor is not installed, a survey meter shall be used to– Check radiation level at door when door is closed– Check door for increase in radiation level when door is
opened
July 2011- Slide 49 of 55Irradiator Technology Course (H-315)
p• Operator shall position body to minimize dose• If abnormal radiation or malfunction occurs, immediately
notify regulatory authority• Do not attempt or authorize repair unless authorized by
license
Operating and Emergency Procedures
• Include an analysis of each type of material to be irradiated to determine compatibility with the irradiator design, and any special safety proceduresI l d i t ti f d ti
July 2011- Slide 50 of 55Irradiator Technology Course (H-315)
• Include instructions for use and routine maintenance, according to manufacturer’s instructions
• Include instructions for maintaining security• Include steps to maintain ALARA• Include instructions for maintaining
accountability
Operating and Emergency Procedures
• Include steps to control access to a malfunctioning or damaged irradiator
• Include steps to take and whom to contact when an irradiator is damaged or malfunctions
July 2011- Slide 51 of 55Irradiator Technology Course (H-315)
• Include specific operating procedures for irradiators with movable sources
• Are maintained current and provided to all users • Posted in irradiator room, or instructions for
access to the procedures
Emergency Procedures
• Common Sense & Include the following:– Leave irradiator room to reduce exposure– Control access– Contact individual responsible for program
July 2011- Slide 52 of 55Irradiator Technology Course (H-315)
p p gand to initiate emergency response
– Telephone numbers for Responsible Individual, Manufacturer, Security, Fire Department, Police, NRC or State Regulator
• Survey outside room to determine if further restrictions are required
Operating and Emergency Procedures
• Licensees can change their operating and emergency procedures only if:– The changes are reviewed and approved and
documented by licensee management
July 2011- Slide 53 of 55Irradiator Technology Course (H-315)
– Affected staff are trained in the procedures before implementation
– Changes are consistent with license conditions and commitments in license application, and
– The changes do not degrade safety
Routine Maintenance and Lubrication
• Licensee should provide standard response:
• We will implement and maintain procedures for routine maintenance of our self-shielded
July 2011- Slide 54 of 55Irradiator Technology Course (H-315)
irradiators according to each manufacturer’s written recommendations.
• Note: Routine Maintenance Does Not Include Any Activities that Involves the Source, Source Drive Mechanism, or removing shielding or source and any other activity in which personnel could receive high doses
Non-Routine Maintenance
• Most Licensee’s Rely on the Manufacturer
• Licensee should provide a response such as
July 2011- Slide 55 of 55Irradiator Technology Course (H-315)
• We will have the self-shielded irradiator manufacturer or other person authorized by NRC or an Agreement State perform non-routine Maintenance.
Standard Responses For
• Instruments• Inventory• Dosimetry• Public Dose
July 2011- Slide 56 of 55Irradiator Technology Course (H-315)
• Leak Tests• Minimization of Contamination, and Waste
Management
/ColorImageDict > /JPEG2000ColorACSImageDict > /JPEG2000ColorImageDict > /AntiAliasGrayImages false /CropGrayImages true /GrayImageMinResolution 150 /GrayImageMinResolutionPolicy /OK /DownsampleGrayImages false /GrayImageDownsampleType /Bicubic /GrayImageResolution 150 /GrayImageDepth 8 /GrayImageMinDownsampleDepth 2 /GrayImageDownsampleThreshold 1.50000 /EncodeGrayImages true /GrayImageFilter /FlateEncode /AutoFilterGrayImages false /GrayImageAutoFilterStrategy /JPEG /GrayACSImageDict > /GrayImageDict > /JPEG2000GrayACSImageDict > /JPEG2000GrayImageDict > /AntiAliasMonoImages false /CropMonoImages true /MonoImageMinResolution 1200 /MonoImageMinResolutionPolicy /OK /DownsampleMonoImages false /MonoImageDownsampleType /Bicubic /MonoImageResolution 1200 /MonoImageDepth -1 /MonoImageDownsampleThreshold 1.50000 /EncodeMonoImages false /MonoImageFilter /CCITTFaxEncode /MonoImageDict > /AllowPSXObjects false /CheckCompliance [ /None ] /PDFX1aCheck false /PDFX3Check false /PDFXCompliantPDFOnly false /PDFXNoTrimBoxError true /PDFXTrimBoxToMediaBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXSetBleedBoxToMediaBox true /PDFXBleedBoxToTrimBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXOutputIntentProfile (None) /PDFXOutputConditionIdentifier () /PDFXOutputCondition () /PDFXRegistryName () /PDFXTrapped /False