INTRAOCULAR INTRAOCULAR PRESSUREPRESSURE

OPTOM FASLU MUHAMMED

Page 2: Iop

S.C.C.O. 05/02/23 2

I. DEFINITIONI. DEFINITION A. the tissue pressure of the ocular

contents B. about 15 mm Hg but does fluctuate

(15.5 +/- 2.57) C. normal range of pressures: 10.5 -

20.5

Page 3: Iop

S.C.C.O. 05/02/23 3

II. HOW IS IOP PRODUCEDII. HOW IS IOP PRODUCED A. Aqueous produced by the ciliary

processes into the posterior chamber ---> flows to anterior chamber and vitreole chamber ---> bulk of fluid flows out the angle (trabecular outflow) of the anterior chamber (flow rate = 2.5 microliters/min)

Page 4: Iop

S.C.C.O. 05/02/23 4

Page 5: Iop

S.C.C.O. 05/02/23 5

Page 6: Iop

S.C.C.O. 05/02/23 6

II. HOW IS IOP PRODUCEDII. HOW IS IOP PRODUCED– 1. resistance to flow at the lens-iris interface– 2. resistance to flow at the angle of the

anterior chamber a. pressure in the eye increases until the force

pushing fluid out of the eye (i.e., the eye pressure) results in the same amount of fluid leaving the eye as is produced and enters the eye

B. Aqueous exchanged with the retina, lens, ciliary body, iris and cornea

Page 7: Iop

S.C.C.O. 05/02/23 7

Page 8: Iop

S.C.C.O. 05/02/23 8

Page 9: Iop

S.C.C.O. 05/02/23 9

Page 10: Iop

S.C.C.O. 05/02/23 10

III.FACTORS THAT III.FACTORS THAT INFLUENCE IOPINFLUENCE IOP A. Long Term

– 1. Genetics - relatives of individuals with open-angle glaucoma are more likely to have high IOP

– 2. Age - IOP increases with increasing age

Page 11: Iop

S.C.C.O. 05/02/23 11

– 3. Sex - IOP's equal in the age range 20 to 40, after menopause women have higher IOP's

– 4. Race - African-Americans have a higher incidence of glaucoma than whites

Page 12: Iop

S.C.C.O. 05/02/23 12

B. Short Term– 1. Diurnal variation - 3 to 6 mm Hg change

in 24 hr period; > 10 mm Hg change is pathogenic a. Change probably related to aqueous

production and not drainage

Page 13: Iop

S.C.C.O. 05/02/23 13

Page 14: Iop

S.C.C.O. 05/02/23 14

Page 15: Iop

S.C.C.O. 05/02/23 15

III. FACTORS THAT III. FACTORS THAT INFLUENCE IOPINFLUENCE IOP

– 2. Sitting - going from a sitting to a lying position results in an increase in IOP which is even greater in glaucoma patients

– 3. Total Body Inversion - causes an increase in IOP by as much as 15 mm Hg

Page 16: Iop

S.C.C.O. 05/02/23 16

III. FACTORS THAT III. FACTORS THAT INFLUENCE IOPINFLUENCE IOP

– 4. Blinking - raises IOP briefly– 5. Exercise - decreases IOP– 6. Blepharospasm - increases IOP– 7. Coughing - increases IOP– 8. Blood pressure - some people believe there

is a link between blood pressure and IOP but no clear evidence

– 9. General anesthesia - decrease IOP– 10. Alcohol - decreases IOP

Page 17: Iop

S.C.C.O. 05/02/23 17

III. FACTORS THAT III. FACTORS THAT INFLUENCE IOPINFLUENCE IOP

– 11. Cannabis - decreases IOP– 12. Tobacco - increases IOP– 13. Cholinergic Stimulating Agents (i.e.,

pilocarpine and echothiophate) - decrease IOP by increasing the aqueous outflow

– 14. Adrenergic Stimulating Agents (i.e., epinephrine, propine, iopidine, alphagan) - lower IOP by enhancing aqueous outflow

Page 18: Iop

S.C.C.O. 05/02/23 18

III. FACTORS THAT III. FACTORS THAT INFLUENCE IOPINFLUENCE IOP

– 15. Adrenergic Blocking Agents (i.e., timolol and betaxolol) - decrease IOP by decreasing aqueous production

– 16. Carbonic anhydrase inhibitors (i.e., diamox, trusopt, azopt) - decrease aqueous production

– 17. Prostaglandins (i.e., xalatan, rescula, travatan, lumigan) - increase uveoscleral outflow

Page 19: Iop

S.C.C.O. 05/02/23 19

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP A. Manometry

– 1. Cannulate the anterior chamber and directly measure the pressure

– 2. Can not be done on humans– 3. The original method used to measure

IOP

Page 20: Iop

S.C.C.O. 05/02/23 20

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP B. Tonometry

– 1. Indentation a. the older of the 2 methods to measure IOP in

humans b. involves measuring the indentation of the

cornea resulting from a given weight c. the Schiotz tonometer is an indentation

tonometer d. the weight of the tonometer displaces fluid in

the eye and thus affects the IOP measurement

Page 21: Iop

S.C.C.O. 05/02/23 21

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP– 2. Applanation

a. only flattens a small portion of the cornea so does not displace a large amount of aqueous.

b. better accuracy than indentation c. the NCT and the Goldmann tonometers are

examples

Page 22: Iop

S.C.C.O. 05/02/23 22

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP C. Schiotz

Tonometer– 1. a plunger of a

known weight pushes on the cornea - thus result depends on ocular rigidity a. concept of ocular

rigidity developed by Friedenwald

Page 23: Iop

S.C.C.O. 05/02/23 23

Page 24: Iop

S.C.C.O. 05/02/23 24

Page 25: Iop

S.C.C.O. 05/02/23 25

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP– 2. Advantages of Schiotz tonometry

a. small and easily transported b. inexpensive (about $100) c. does not require electricity

Page 26: Iop

S.C.C.O. 05/02/23 26

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP– 3. Disadvantages of Schiotz tonometry

a. not extremely accurate - ocular rigidity dependent and instrument scale markings are not detailed

b. requires anesthetic for most patients c. assumes everyones epithelium is 0.05 mm thick d. technique can produce abrasions e. best if patient in a reclining position f. placing the tonometer on the eye changes the

IOP

Page 27: Iop

S.C.C.O. 05/02/23 27

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP D. Goldmann Tonometry

– 1. The most accurate method for IOP measurement

– 2. Readings within 1 - 2 mm of actual IOP– 3. Flattens a small portion of the cornea

Page 28: Iop

S.C.C.O. 05/02/23 28

Page 29: Iop

S.C.C.O. 05/02/23 29

Page 30: Iop

S.C.C.O. 05/02/23 30

Page 31: Iop

S.C.C.O. 05/02/23 31

Page 32: Iop

S.C.C.O. 05/02/23 32

Page 33: Iop

S.C.C.O. 05/02/23 33

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP– 4. Theory

a. the cornea is covered with a tear layer which exerts a surface tension (force in towards the cornea)

b. a probe applied to the cornea is acted against (a force pushing out from the cornea) by the corneal thickness and elasticity (the bending force)

Page 34: Iop

S.C.C.O. 05/02/23 34

Page 35: Iop

S.C.C.O. 05/02/23 35

c. if the area of the probe is of the proper size then the force from the surface tension will cancel the bending force

Page 36: Iop

S.C.C.O. 05/02/23 36

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP d. this leaves

– Pressure = Force / Area e. the area of the probe has a diameter of 3.06

mm

Page 37: Iop

S.C.C.O. 05/02/23 37

Page 38: Iop

S.C.C.O. 05/02/23 38

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP– 5. Procedure

a. instill fluorescein into the tear layer b. view fluorescein pattern with the blue light on

the slit lamp c. doubling prism in place to split the view in

half d. image of the split circle must be lined up e. pressure = the number on the drum times 10

in mm Hg.

Page 39: Iop

S.C.C.O. 05/02/23 39

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP– 6. Sources of Error

a. improper width or position of mires b. inappropriate fluorescein levels c. unusual corneal thickness

Page 40: Iop

S.C.C.O. 05/02/23 40

Page 41: Iop

S.C.C.O. 05/02/23 41

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP– 7. Advantages of Goldmann Tonometry

a. highly accurate and reliable (procedure does not influence IOP)

b. accepted norm for IOP measurement c. easy to perform d. not very expensive (about $1000)

Page 42: Iop

S.C.C.O. 05/02/23 42

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP– 8. Disadvantages of Goldmann Tonometry

a. requires anesthesia b. can result in an abrasion c. must sterilize instrument after each use d. not portable

Page 43: Iop

S.C.C.O. 05/02/23 43

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP E. Noncontact Tonometer

– 1. Achieves corneal flattening by an air jet of calibrated, increasing force

– 2. Corneal flattening is detected by a photo cell

– 3. From the known force of the air jet and the dimensions of the air jet the pressure is calculated

Page 44: Iop

S.C.C.O. 05/02/23 44

Page 45: Iop

S.C.C.O. 05/02/23 45

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP– 4. The higher the IOP the longer it takes to

flatten the cornea (i.e., if IOP = 17 mm Hg, flattening takes 10.5 msec; if IOP = 36 mm Hg, flattening takes 24 msec)

– 5. Advantages of NCT a. no anesthesia required b. fairly reliable for pressures in the normal range

(less reliable as the pressure increases) c. quick and easy to perform d. does not touch the eye (can use on diseased eyes)

Page 46: Iop

S.C.C.O. 05/02/23 46

IV. MEASUREMENT OF IOPIV. MEASUREMENT OF IOP– 6. Disadvantages of NCT

a. expensive (over $5000) b. not accurate at high pressures c. patient apprehension (puff) raises IOP d. not easily portable

Page 47: Iop

S.C.C.O. 05/02/23 47

V. IOP AND GLAUCOMAV. IOP AND GLAUCOMA A. If the IOP is high enough for a long

enough time, the patient will lose visual field in a characteristic pattern

B. If the angle is totally blocked the pressure rises to 55 or 60 mm Hg and field loss is seen in a few hours with total loss in a few days

Page 48: Iop

S.C.C.O. 05/02/23 48

V. IOP AND GLAUCOMAV. IOP AND GLAUCOMA C. In open angle glaucoma the pressure

is more often in the 20's or 30's– 1. this pressure can be tolerated for

months before damage occurs– 2. sometimes patients with normal

pressure develop glaucomatous changes

Page 49: Iop

S.C.C.O. 05/02/23 49

V. IOP AND GLAUCOMAV. IOP AND GLAUCOMA D. To diagnose glaucoma usually need:

– 1. high IOP– 2. visual field defects– 3. cupping at the disc

Page 50: Iop

S.C.C.O. 05/02/23 50