  • Status of Ionizing radiation Metrology at NIM of China 1

    Ionizing Radiation Metrology at NIM

    Zhang Jian

    National Institute of Metrology

  • Status of Ionizing radiation Metrology at NIM of China 2

    Outline

    Status and future of ionizing radiation metrology

    at NIM

    General introduction

    Highlights in radioactivity, dosimetry and neutron

    Future work

  • Status of Ionizing radiation Metrology at NIM of China 3

    Outline

    Status and future of ionizing radiation metrology

    at NIM

    General introduction

    Highlights in radioactivity, dosimetry and neutron

    Future work

  • Status of Ionizing radiation Metrology at NIM of China 4

    Organization

    National Institute of Metrology(NIM) is established in

    1955

    Two Sites in Beijing: He Ping Li and Changping

    和平里院区

    昌平院区

  • Status of Ionizing radiation Metrology at NIM of China 5

    Organisation

    • Three section: Radioactivity, Dosimetry, Neutron

    • Staff: Around 26 employees in this division

    New Building for LINAC, CT and monoenergetic X-rays

    Ionizing radiation building in He Ping Li Building for electron

    accelerator for high-dose dosimetry

  • Status of Ionizing radiation Metrology at NIM of China 6

    Overview of ionizing radiation metrology

    Primary standards: 18

    Secondary standards: 15

    Comparison: 34

    CMCs:195

  • Status of Ionizing radiation Metrology at NIM of China 7

    The Standards in Radioactivity Lab 放射性活度基标准装置

  • Status of Ionizing radiation Metrology at NIM of China 8

    Primary standards for dosimetry

    辐射剂量基准装置

    Air kerma standard for (10-60)kV X-rays

    Air kerma standard for 60Co γ-rays

    Absorbed to graphite of 60Coγ-rays by cavity chamber Water Calorimeter

    Air kerma standard for (60-250)kV X-rays Air kerma standard for 137Cs γ-rays

  • Status of Ionizing radiation Metrology at NIM of China 9

    Fricke dosimetry Graphite calorimeter

    Radiation facility and TLD system Alanine/EPR dosimetry

    Standard for Diagnosis x-rays

    Primary and secondary standard for dosimetry 辐射剂量基标准装置

    http://124.42.64.140/nimlab.web/Common/frmMaxPic.aspx?picId=17&Type=2

  • Standards for neutron 中子基准装置

    Primary standard of neutron source strength

    primary standard of 14 MeV neutron absorbed dose

    Neutron reference radiation field

  • Status of Ionizing radiation Metrology at NIM of China 11

    Outline

    Status and future of ionizing radiation metrology

    at NIM

    General introduction

    Highlights in radioactivity, dosimetry and neutron

    Future work

    Collaboration between NRC and NIM

    Review of the collaboration

    Proposal on future collaboration

  • Status of Ionizing radiation Metrology at NIM of China 12

    Absolute measurement of radon radioactivity 氡活度绝对测量

    • An absolute measurement system of radioactivity for condensed radon has being developed based on defined angled method.

    • Dimensions parameters will be measured and uncertainties will be evaluated recently.

    冷头

    射线范围

    radioactivity

  • Status of Ionizing radiation Metrology at NIM of China 13

    Radon-222 and its progenies metrology 氡-222及其子体计量

    Main room

    Fan

    MM

    Mixing room

    Entry room Radiumsourcesroom

    1500

    1600

    650

    1350

    850

    1650

    3000

    2000

    15

    20

    25

    30

    35

    0

    64

    64

    56

    56

    48

    48

    40

    40

    32

    32

    2416

    16

    Te

    mp

    era

    ture

    Time (h)

    8

    0

    75

    80

    85

    90

    95

    100

    24

    8

    Hu

    mid

    ity (

    %)

    Time (h)(o

    C)

    0 8 16 24 32 40 480

    200

    400

    600

    EECunatt. (Bq/m3)

    EECsum (Bq/m3)

    fp

    Time (hour)

    0

    2000

    4000

    6000

    0.0

    0.1

    0.2

    0.3

    0.4

    f p

    EE

    Csu

    m(B

    q/m

    3)

    EE

    Cu

    na

    tt.(B

    q/m

    3)

    The schematic view and main dimensions of the NIM radon chamber. All values are given in millimeters

    The picture of radon chamber

    Monitoring data of temperature and humidity in the mode of keeping stable at set values

    The regulation of the attached and unattached ratio based on the control of radon concentration and aerosol concentration.

    radioactivity

  • Status of Ionizing radiation Metrology at NIM of China 14

    Concentration measurement of radon isotopes progenies by LSC

    ( ) ( )

    ( ) ( ) ( ) ( )

    ( ) ( ) ( ) ( )

    1 1 1( ) [ ]

    ( ) ( ) ( ) ( ) ( ) ( )

    A

    B A B

    C B C

    A B C

    tA A p

    Bt t tB B p A p

    B A

    Ct t tC C p B p

    C B

    t t tA p C B

    B A C A B A C B C A C B

    A t A t e

    A t A t e A t e e

    A t A t e A t e e

    A t e e e

    )()()()(

    )()(

    tt

    EF

    FpE

    tpFF

    tpEE

    FEF

    E

    eetAetAtA

    etAtA

    Membrane Sampling Membrane in Liquid scintillation vials

    Measured by Tricarb 3100

    Radon-222 progenies metrology 氡-222子体计量

    radioactivity

  • Status of Ionizing radiation Metrology at NIM of China 15

    87.8508

    1.71951.7195

    73.5943

    60.3600

    3.8667

    R 2.5750

    6.4733

    5.1733

    76.5000

    FEP efficiency simulation for HPGe Detector 高纯锗探测器全能峰效率曲线

    Radiography of a HPGe detector (CANBERRA GX6250)

    Scan by CT

    0 200 400 600 800 1000 1200 1400 1600

    5.0x10-4

    1.0x10-3

    1.5x10-3

    2.0x10-3

    2.5x10-3

    3.0x10-3

    FE

    P E

    ffcie

    ncy

    Energy/keV

    Experiment

    Calculation

    Detector parameters

    Energy/keV Exp-FEP-eff Cal-FEP-eff Relative

    deviation 59.5409 2.971E-03 2.976E-03 -0.17%

    88.0336 2.925E-03 2.947E-03 -0.76%

    121.7817 2.752E-03 2.793E-03 -1.50%

    244.6974 2.016E-03 2.045E-03 -1.43%

    344.2785 1.616E-03 1.619E-03 -0.19%

    356.1029 1.577E-03 1.579E-03 -0.15%

    411.1165 1.422E-03 1.427E-03 -0.36%

    443.965 1.334E-03 1.337E-03 -0.20%

    604.72 1.063E-03 1.072E-03 -0.86%

    661.657 1.005E-03 1.002E-03 0.27%

    778.9045 8.991E-04 8.931E-04 0.67%

    964.079 7.726E-04 7.665E-04 0.79%

    1173.228 6.768E-04 6.725E-04 0.64%

    1332.492 6.167E-04 6.169E-04 -0.03%

    1408.013 5.947E-04 5.879E-04 1.14%

    radioactivity

  • Activity Standardization and Comparison Plan for F-18 放射性核药物F-18活度测量及比对

    Standardization of medical nuclide F-18 and plan to take part in

    the comparison using the SIRTI, piloted by BIPM (estimate in 2018)

    SIRTI used in the comparison for F-18

    Primary standard for F-18 measurement

    radioactivity

  • Status of Ionizing radiation Metrology at NIM of China 17

    PET system of Inliview 3000 animal PET/SPECT/CT system

    Collaboration of NIM with

    Key Laboratory of Particle

    & Radiation Imaging

    (Tsinghua University) on

    nuclear medical image

    quality study

    Metrology of radioactivity

    of F-18 Nuclide (F-18 FDG)

    & study of image quality

    enhancement of pre-

    clinical and clinical PET

    systems

    phantom experiments were performed Transverse plane of reconstructed images for ultra-micro hot-rod phantom experiment (100 iterations, unit: mm).

    Transverse plane, coronal plane and sagittal plane of reconstructed images for whole-body mouse 18F-FDG experiment (50 iterations):

    Profiles of point source reconstructed images

    Radioactivity Metrology usage in nuclear medicine 核医学放射性活度计量

    radioactivity

  • Status of Ionizing radiation Metrology at NIM of China 18

    Water Calorimeter for high energy photon absorbed dose 高能光子水吸收剂量基准装置

    In collaborating with NRC, NIM is developing a primary standard for absorbed dose to water at radiotherapy dose levels. The developing standard for water absorbed dose is based on a water calorimeter and is designed to operate in Co-60 and megavoltage photon beams from a linear accelerator.

    The absorbed dose to water of 10 MV X-ray in horizontal beam is absolute measured with the combined uncertainty of 0.35%.

    dosimetry

  • Status of Ionizing radiation Metrology at NIM of China 19

    Measurement of Energy response of image plates exposed to electrons From Linac

    成像板(IP)在医用加速器电子束下的能响

    Response of the Ips to photon at energy between 4MeV and

    15Mev

    dosimetry

  • Status of Ionizing radiation Metrology at NIM of China 20

    Alanine dosimeter(1Gy~100Gy) 丙氨酸剂量计的剂量测量范围扩展

    Vacuum connector Schematic diagram

    dosimetry

  • Status of Ionizing radiation Metrology at NIM of China 21

    Linearity

    cubic polynomial after natural logarithm Fitting a straight line

    0 20 40 60 80 100

    0.00

    0.05

    0.10

    0.15

    0.20

    0.25

    EP

    R in

    ten

    sity p

    er

    un

    it m

    ass(m

    g)

    Dose to water(Gy)

    Y = A + B * X

    Parameter Value Error

    ------------------------------------------------------------

    A 0.0056 2.08393E-4

    B 0.00237 4.09537E-6

    ------------------------------------------------------------

    R SD N P

    ------------------------------------------------------------

    0.99999 3.86209E-4 8

  • Status of Ionizing radiation Metrology at NIM of China 22

    Primary standard of air kerma for mammography 乳腺X射线空气比释动能基准装置

    The goal of mammography is the early detection of breast cancer.

    Primary standard of mammography

    dosimetry

  • Status of Ionizing radiation Metrology at NIM of China 23

    Primary standard of air kerma for (250-600)kV X-ray (250-600)kV X射线空气比释动能基准

    (250-450)kV X射线空气比释动能国家基准

    石墨空腔基准电离室

    X射线辐射装置

    电离室定位系统

    微弱电流测量系统

    控制与监测系统

    蒙特卡罗模拟

    电离室设计、加工

    电离室内部尺寸精密测量

    不确定度分析 量值复现 国际比对

    辐射防护仪表 环境监测仪表 治疗水平电离室 X射线探测器

    量值传递

    稳定性测试

    系统测试与检验

    软件的研制

    电容测试

    光管调节装置

    滤过系统

    辐射能谱的测量

    电流系统搭建

    辐射质的研究与建立

    6米导轨

    多维移动平台

    激光定位

    量杆及基准面

    视频监控

    辐射监测

    系统控制软件

    校准自动化软件

    漏电测试

    光阑、快门

    Primary standard: Graphite cavity chamber

    Reference radiation facility Main part of primary standard

    dosimetry

  • Status of Ionizing radiation Metrology at NIM of China 24

    Standard for Image quality control in X-ray diagnosis X射线诊断图像质控标准

    To develop technical performance standards for x-ray diagnosis, focusing particularly on image quality: Spatial Resolution, Contrast Sensitivity.

    Scan picture of HpGe detector Industrial CT equipment

    dosimetry

  • Status of Ionizing radiation Metrology at NIM of China 25

    To calibrate energy resolution, efficiency of X-rays detector To calibrate The Hard X-ray Modulation Telescope (HXMT):

    HXMT:the first high energy astrophysics satellite in China.

    National standard of X-rays fluence Hard X-ray Calibration Facility(HXCF)

    硬X射线标定装置

    Hard X-ray calibration facility HXMT satellite

    dosimetry

  • Status of Ionizing radiation Metrology at NIM of China 26

    Measurement of monoenergetic x-rays facility 单能X射线测量装置

    0 20 40 60 80 100 120 1400.55

    0.60

    0.65

    0.70

    0.75

    0.80

    0.85

    0.90

    0.95

    1.00

    1.05

    Eff

    icie

    ncy(%

    )

    Energy(keV)

    Experimental data

    MC simulation

    Result from CCD detector

    20 40 60 80 100 1200

    500

    1000

    1500

    2000

    2500

    3000

    3500

    4000

    4500

    5000

    5500

    Flu

    x c

    ts/s

    Energy (keV)

    Flux of monochromatic light

    0 100 200 300 400 500

    0

    20

    40

    60

    80

    100

    Effic

    ien

    cy(%

    )

    Energy(keV)

    Full energy peak efficiency

    detect efficiency

    Measured by GEM detector Spectrum of monoenergetic X-rays

    Efficiency in parallel beams Compare experiment and simulation

    dosimetry

  • Status of Ionizing radiation Metrology at NIM of China 27

    K fluorescence radiation facility K 荧光辐射装置

    Simulation module based on MCNP

    Yield of K radiation VS thickness of radiator

    Energy Spectrum of K fluorescence

    K fluorescence radiation facility

    dosimetry

  • Status of Ionizing radiation Metrology at NIM of China 28

    An irradiator with a 137Cs source had been installed in the neutron radiation field to verify the photon response of neutron dose meters or neutron detectors. The radioactivity:1.85×1011Bq (5Ci).

    The old long counter was replaced by a new one with BF3 detectors. One of the purposes for the new counter is used as a transfer standard for other neutron reference radiation field.

    Improving for the neutron reference radiation field 中子辐射场升级改造

    neutron

  • Status of Ionizing radiation Metrology at NIM of China 29

    Outline

    Status and future of ionizing radiation metrology

    at NIM

    General introduction

    Highlights in radioactivity, dosimetry and neutron

    Future work

    Collaboration between NRC and NIM

    Review of the collaboration

    Proposal on future collaboration

  • Status of Ionizing radiation Metrology at NIM of China 30

    Traceability system for Radiotherapy based on LINAC 基于医用加速器的放疗剂量量值体系的完善

    Establish primary standard

    K6 comparison

    Traceability system

    Update the Traceability system for Radiotherapy

    Research the dosimetry measurement method and dosimetry instrument used in clinical QA

    Improve and assess the test method for clinical LINAC

    Small field dosimetry

    Ongoing

  • Status of Ionizing radiation Metrology at NIM of China 31

    Standard for radiotherapy in low/medium energy X-ray 低、中能X射线放疗的国家标准

    For 10-400kV X-ray 1. From air kerma to absorbed dose to water 2. Measured by graphite calorimeter and

    water calorimeter

    New radiotherapy devices in Kilovoltage X-ray, such as brachytherapy with miniature X-ray tube. To research new technology for calibration

    Ongoing

  • Status of Ionizing radiation Metrology at NIM of China 32

    Metrology in Synchrotron X-rays 同步辐射X射线计量

    Calibration of energy response of detection elements for controlled nuclear fusion by Synchrotron X-rays.

    To measure the flux of synchrotron X-rays.

    On plan

    http://photo.blog.sina.com.cn/showpic.html

  • Status of Ionizing radiation Metrology at NIM of China 33

    Standards for radiation environmental monitor and radiation safety 超低本底剂量学、辐射环境监测、辐射安全

    Low background facility at NIM Low background laboratory at Jinping, Sichuan (On Plan)

    2、Standards of testing for radiation detection systems and radiation safety 3、Pulsed radiation dosimetry

    1、Standard for radiation environmental monitor

    pulsed radiation facility radiation detection systems

    On plan

  • Status of Ionizing radiation Metrology at NIM of China 34

    Neutron 中子能谱和中子注量堆

    • The neutron spectrum measurement

    • A set of BSS including 12 spheres of 3.0”, 3.5”, 4.0”, 4.5”, 5.0”, 6.0”, 7.0”, 8.0”, 9.0”, 9.5”, 10.0” and 12.0” made by Centronic was already sent to our laboratory and plan to be scaled in 2017. Then the BSS will be used to measure the neutron spectrum in some actual workplaces for the onsite calibration and the preparation of simulative field from 2016.

    • Rebuilding the thermo neutron fluence pile

    • The primary standard of thermal neutron fluence rate was established in 1970. It include six Am-Be neutron sources (strength (3.7 to 6.3) × 106s-1) and moderating medium constituting with olefin and graphite. The thermal neutron fluence field is in the center of moderator. The construction and shape of moderator was change and the thermal neutron fluence field is too small are main problems of the standard. We plan to rebuild a new thermo neutron flence pile in 2018.

    Ongoing

  • Status of Ionizing radiation Metrology at NIM of China 35

    Proposal on future collaboration

    • Enhance the measurement on radiotherapy dosimetry, such as electron calorimeter, HDR brachaytherapy, alanine dosimetry, MC simulation on FAC, and beta dosimetry.

    • Develop Monte Carlo simulation techniques, enhance the measuring capability in radon, nuclides based on Internal gas counting and hold bilateral comparisons in nuclear medicine nuclides.

  • Status of Ionizing radiation Metrology at NIM of China 36

    Absorbed Dose measurement for electron, and carbon ion therapy

    高能电子束和粒子束剂量学

  • Status of Ionizing radiation Metrology at NIM of China 37

    Monte carlo simulation based on EGSnrc code 基于EGSnrc的蒙卡建模

    • MC simulation for free air chamber, cavity chamber, linac, and related dosimetry modeling.

  • Status of Ionizing radiation Metrology at NIM of China 38

    Fricke and Alanine Dosimetry 硫酸亚铁剂量计和丙氨酸剂量计

    • Fricke dosimeter

    • Compare G (F3+) values in a Co-60 beam between NIM and NRC

    • To study electron beam G (F3+) values used to be develop an absolute dose to water standard for megavoltage electron beam .

    • To study method used Fricke dosimetry system to measure an absolute

    dose to water for HDR brachytherapy • Alanine

    • Research on how to improve the measurement uncertainty in 1Gy to 5Gy

    • Make smaller size used for small field dosimetry research

    • Study application in radiotherapy and Radiography

  • Status of Ionizing radiation Metrology at NIM of China 39

    Enhance measurement capability in radon 提升氡测量能力

    Develop absolute measurement system for radon radioactivity.

    Improve radon gas and radon progenies secondary standards

    Build other radon secondary standards used in radiation protection field,Such as radon in water and radon emission rate standard.

    Radon becomes an important factor leading to lung cancer in China.

    Radon potential map also reveals metrology demands in Canada

  • Status of Ionizing radiation Metrology at NIM of China 40

    Internal Gas Counting measurement 内充气计数测量

    Establish the standards for gas radioactivity, based on internal gas counting techniques by applying length compensation method.

    Investigate the method to calibrate the radioactive gas monitor systems in nuclear pow plants ( noble gas nuclides Kr-85, Xe-133 etc., particular H-3 in heavy water reactors) .

    Preliminary design of gas radioactivity measurement system

    gas circuit

    detectors electronic modules

  • Status of Ionizing radiation Metrology at NIM of China 41

    Bilateral comparisons in nuclear medicine nuclides 核医学核素活度双边比对

    Cyclotron to produce new type medicine nuclides.

    Enhance measurement technique based on DCC and TDCR to meet the requirement to absolutely standardize short life nuclides such as F-18, N-13, C-11 etc.

    Develop portable measuring system for such kind of nuclear medicine nuclides.

    Hold bilateral comparison to test and improve metrology methods.

    Diagnosis and treatment by nuclear medicine need precise measurement

  • Status of Ionizing radiation Metrology at NIM of China 42

    Thanks

    Hope to have more cooperation

    with all of you.