Various Ionization Techniques used In Mass Spectroscopy Presented to Ritu Mam Presented By Pema Chodon 1 – M.Pharma Dept. Of Pharmaceutics Al – Ameen College Of Pharmacy Bangalore Re-Edited by Suraj C. 1 st M.Pharm AACP Ionization Techniques In Mass Spectroscopy Mass Spectrometer ION SOURCE Since the mass analyzer utilizes only gaseous ions i.e., starting point of mass spectrometric analysis is formation of gaseous analyte ions. Non –Volatile solids are first converted in to gases and from the gaseous sample the ions are produced in a Box like enclosure called Ion Source. Function Produces ion without mass discrimination of the sample. Accelerates ions into the mass analyzer. Classification of Ion Source: On the basis of the nature of the substance and the method by which ions are generated the ion sources are classified as Ion Source Mass Analyzer Ion collection System Data Handling System Vacuum System Inlet System 0
19

Ionizaion Techniques - Mass Spectroscopy

Jun 02, 2015

Download

Health & Medicine

Suraj Choudhary

Several Ionization techniques.
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Ionizaion Techniques - Mass Spectroscopy

Various Ionization Techniques used In Mass Spectroscopy

Presented to Ritu Mam

Presented By Pema Chodon

1 – M.Pharma Dept. Of Pharmaceutics

Al – Ameen College Of Pharmacy Bangalore

Re-Edited by Suraj C.

1st M.Pharm AACP

Ionization Techniques In Mass Spectroscopy

Mass Spectrometer

ION SOURCE • Since the mass analyzer utilizes only gaseous ions i.e., starting point of mass spectrometric analysis

is formation of gaseous analyte ions. • Non –Volatile solids are first converted in to gases and from the gaseous sample the ions are

produced in a Box like enclosure called Ion Source.

Function Produces ion without mass discrimination of the sample. Accelerates ions into the mass analyzer.

Classification of Ion Source:

On the basis of the nature of the substance and the method by which ions are generated the ion sources are classified as

Ion Source Mass Analyzer

Ion collection System

Data Handling System

Vacuum System

Inlet System

0

Page 2: Ionizaion Techniques - Mass Spectroscopy

Gas Phase Sources Electron Impact Ionization (EI) Chemical Ionization (CI) Field Ionizations (FI)

Desorption Sources

Field Desorption (FD) Electrospray Ionization (ESI) Matrix assisted desorption/Ionisation (MALDI) Plasma desorption (PD) Fast Atom Bombardment (FAB) Thermospray Ionization (TS) Secondary Ion Mass Spectrometry (SIMS)

Gas Phase Ionization Methods

1. Electron Impact Ionization

INTRODUCTION • Electron impact (EI) is the classical ionization method in mass spectrometry.

• It is the most widely used and highly developed method.

• It is also known as Electron bombardment or Electron Ionization.

CONSTRUCTION & WORKING:

• Electron impact ionization source consists of a ionizing chamber which is maintained at a pressure

of 0.005 torr and temperature of 200 ± 0.25 degrees.

• Electron gun is located perpendicular to chamber.

• Electrons are emitted from a glowing filament (tungsten or rhenium) by thermionic emission and

accelerated by a potential of 70 V applied between the filament and anode.

• These electrons are drawn in the ionization chamber through positively charged slits.

• The number of electrons is controlled by filament temperature and energy of energy is controlled

by filament potential.

1

Page 3: Ionizaion Techniques - Mass Spectroscopy

• The sample is brought to a temperature high enough to produce molecular vapors.

• The gaseous Neutral molecules then pass through the molecular leaks and enter the ionization

chamber (which is maintained at a pressure of 0.005 torr and a temperature of 200 ± 0.250 C).

MECHANSIM:

• The gaseous sample and the electrons collide at right angles in the chamber and ions are formed by

exchange of energy during these collisions between electron beam and sample molecules.

M Analyte molecule e- Electrons M.+ Molecular ions

• In this example, 20eV is transferred to a molecule following it’s collision with a 70eV electron.

• Since the ionization energy of most of the organic molecules is 15eV an electron is expelled to

produce a radical cation with 5eV excess energy.

• The positive ions formed in the chamber are drawn out by a small potential difference (usually

5eV) between the large repeller plate (positively charged) and first accelerating plate (negatively

charged).

• Strong electrostatic field (400 – 4000 V) is applied between the first and second accelerating plates

accelerates the ions according to their masses (m1, m2, m3 etc) to their final velocities.

• The ions emerge from the final accelerating slit as a collimated ribbon of ions.

• The energy and velocity of ions are given by :-

zV = ½ (m1v1) = ½ (m2v2) = ½ (m3v3) where: z = charge of the ion

V = accelerating potential v = velocity of ion

2

Page 4: Ionizaion Techniques - Mass Spectroscopy

ADVANTAGES Gives molecular mass and also the fragmentation pattern of the sample.

Extensive fragmentation and consequent large number of peaks gives structural information.

Gives reproducible mass spectra.

DISADVANTAGES

Sample must be thermally stable and volatile.

A small amount of sample is ionized (1 in 1000 molecules).

Unstable molecular ion fragments are formed so readily that are absent from mass spectrum.

NOTE: 70eV, the de Broglie wavelength of an electron matches with the length of typical bonds in

organic molecules (0.14 nm) and energy transferred to organic molecules is maximized at this

wavelength.

2. Chemical Ionization

INTRODUCTION

• In chemical ionization, the ionization of the analyte is achieved by interaction of it’s molecules

with ions of a reagent gas in the chamber or source.

CONSTRUCTION & WORKING:

• Chemical ionization is carried out in an instrument similar to electron impact ion source with

some modifications such as:-

Addition of a vacuum pump.

Narrowing of exit slit to mass analyzer to maintain reagent gas pressure of about 1 torr in

the ionization chamber.

Providing a gas inlet.

• It is a two part process.

• In the first step

A reagent gas is ionized by Electron Impact ionization in the source.

3

Page 5: Ionizaion Techniques - Mass Spectroscopy

The primary ions of reagent gas react with additional gas to produce stabilized reagent ions.

• In the second step, the reagent ions interact with sample molecules to form molecular ions.

• In this technique the sample is diluted with a large excess of reagent gas so that the probability of

ionizing collisions between sample molecules and the electrons is negligibly small and primary ions

are formed entirely from interaction with reagent gas ions.

• Gases commonly used as reagent are low molecular weight compounds such as Methane, tertiary

Isobutane, Ammonia, Nitrous oxide, oxygen and hydrogen etc.

TYPES OF CI:

• Depending upon the type of ions formed CI is categorized as:-

1. Positive Chemical Ionization

2. Negative Chemical Ionization

1. Positive Chemical Ionization

In this technique positive ions of the sample are produced.

In positive chemical ionization, gases such as Methane, Ammonia, Isobutane etc are used.

For example,

Ammonia is used as reagent gas.

First ammonia radical cations are generated by electron impact and this react with

neutral ammonia to form ammonium cation (reactive species of ammonia CI).

NH3 NH3

.+ + 2 e- NH3

.+ NH4+ + NH2

NH4

+ reacts with the sample molecules by proton transfer or Adduct formation to produce sample ions.

M + NH4

+ [M + H]+ + NH3 Proton transfer M + NH4

+ [M + NH4]+ Adduct formation When Methane is used as Reagent gas. Methane is ionized by electron impact:

CH4 + e- CH4

+ + 2e- Primary ions react with additional reagent gas molecules to produce stabilized

reagent ions: CH4

+ + CH4 CH5+ + CH3

CH3 + CH4 C2H5

+ + H2 The reagent ions then react with the sample molecules to ionize the sample

molecules:

CH5+ + MH CH4 + MH2

+ (Proton transfer) CH3

+ + MH CH4 + M+ (Anhydride abstraction) CH4

+ + MH CH4 + MH+ (Charge transfer)

e-

4

Page 6: Ionizaion Techniques - Mass Spectroscopy

2. Negative Chemical Ionization Negative chemical ionization is counterpart of Positive chemical ionization.

In this technique, negative ions of the sample are formed.

Oxygen and Hydrogen are used as reagent gasses.

This method is used for ionization of highly electronegative samples.

The negative ions are formed by following reactions :-

A. Resonance electron capture

M + e- M-

B. Dissociative electron capture RCl + e- R + Cl-

H2O + e- H + OH-

The ion molecule reaction occurring between negative ion formed in the chamber source

and the sample molecule include:-

Charge transfer.

Hydride transfer.

Anion- Molecule adduct formation.

ADVANTAGES

Used for high molecular weight compounds.

Used for samples which undergo rapid fragmentation in EI.

LIMITATIONS

Not suitable for thermally unstable and non-volatile samples.

Relative less sensitive then EI ionization.

Samples must be diluted with large excess of reagent gas to prevent primary interaction

between the electrons and sample molecules.

3. Atmospheric Pressure Chemical Ionization INTRODUCTION:

• It is a variant of chemical ionization and is carried out using an ion source similar to ESI.

• APCI produces ions using a reagent gas generated from solvent vapour.

CONSTRUCTION & WORKING:

• The solvent - a mixture of methanol, acetonitrile and water at 0.5 ml/min - is supplied to the APCI

probe by a pump (either from HPLC or LC).

• Liquid spray is produced by passing co-axial nebuliser gas (nitrogen).

• The solvent spray is vaporised by a heating.

• Once formed, the vapour is emanating from a corona pin held at 3 kV.

5

Page 7: Ionizaion Techniques - Mass Spectroscopy

• The electric field is sufficiently strong to ionize solvent vapour by either removal (positive ion

mode) or donation (negative ion mode) of an electron.

• Ion/molecule reactions then result in the formation of a reactive species.

• For example, with Methanol:

Positive In AP-CI

Negative Ion AP-CI

• Acid-base reaction then takes place between the sample and reagent gas, resulting in protonation (positive

ion mode) or deprotonation (negative ion mode) of the sample molecule (M).

Positive In AP-CI

Negative Ion AP-CI

• The sample ions are then accelerated out of the atmospheric pressure source and into the mass analyzer by

application of a small voltage (typically 20-70 V) to the skimmer cone.

• The pressure differential between source and analyzer regions is maintained by the presence of an area of

intermediate vacuum.

• During the ionization process itself, little energy is transferred to the sample molecule, and fragmentation

is minimal.

6

Page 8: Ionizaion Techniques - Mass Spectroscopy

• Upon acceleration of the sample ions through the hot solvent vapour, however, collisional activation and

subsequent fragmentation is common.

APPLICATIONS

APCI is suitable for the analysis of organic compounds with medium - high polarity.

Since positive ionization is dependent on protonation, molecules containing basic functional

groups such as amino, amide esters, aldehyde/ketone and hydroxyl can be analyzed.

Negative ionization depends upon deprotonation, molecules containing acidic functional

groups are analyzed by this method.

Can be used as LC/MS interface.

4. Field Ionization INTRODUCTION:

• FI is used to produce ions from volatile compounds that do not give molecular ions by EI.

• It produces molecular ions with little or no fragmentation.

• Application of very strong electric field induces emission of electrons.

CONSTRUCTION & WORKING:

• In this technique, sample molecules in vapour phase is brought between two closely spaced

electrodes in the presence of high electric field (107 - 108 V/Cm), it experiences electrostatic force.

• If the metal surface (anode) has proper geometry (a sharp tip, cluster of tips or a thin wire) and is

under vacuum (10-6 torr), this force is sufficient to remove electrons from the sample molecule

without imparting much excess energy.

• The electric field is produced by applying high voltage (20 KV) to these specially formed emitters

(made up of thin tungsten wire).

• In order to achieve high potential gradients necessary to effect ionization, the anode is activated by

growing carbon microneedles or whiskers.

7

Page 9: Ionizaion Techniques - Mass Spectroscopy

• These whiskers are 10 micro meters in length and greater than 1µm in diameters.

• These whiskers are capable of removing valence electrons from the organic molecules by quantum

mechanical tunneling mechanism.

• As concentration of sample molecules is high at the anode ion-molecule reactions often occur which

results in formation of protonated species ( M+H )+.

• Thus both M+ and (M+H) + is observed in FI spectrum.

• These cations are accelerated out of the source and their mass is analyzed by analyzer.

ADVANTAGES

As fragmentation is less, abundance of molecular ions (M+) is enhanced, hence this

method is useful for relative molecular mass and empirical formula determination.

DISADVANTAGES

Not suitable for thermally unstable and non volatile samples.

Sensitivity is les than EI ion source.

No structural information is produced as very little fragmentation occurs.

5. Field desorption

INTRODUCTION:

• In field desorption method, a multitipped emitter (made up of tungsten wire with carbon or silicon

whiskers grown on its surface) similar to that used in FI is used.

CONSTRUCTION & WORKING

• The electrode is mounted on a probe that can be removed from the sample compartment and

coated with the solution of the sample.

• The sample solution is deposited on the tip of the emitter whiskers either by

dipping the emitter into analyte solution or

using a microsyringe.

• The probe is then reinserted into the sample compartment which is similar to CI or EI unit.

• Then the sample is ionized by applying a high voltage to the emitter.

8

Page 10: Ionizaion Techniques - Mass Spectroscopy

NOTE: In some cases it is necessary to heat the emitter by passing a current through the wire to

evaporate the sample.

• Ionization takes place by quantum mechanical tunneling mechanism, which involves transfer of

ions from the sample molecule to the anode (emitter).

• This results in formation of positive ions which are radical ions (M+) and cations attached species

such as (M+Na)+.

• (M+Na)+ are produced during desorption by attachment of trace alkali metal ions present in

analyte.

ADVANTAGES

Works well for small organic molecules, low molecular weight polymers and petrochemical

fractions.

DISADVANTAGES

Sensitive to alkali metal contamination.

Sample must be soluble in a solvent.

Not suitable for thermally unstable and non volatile samples.

Structural information is not obtained as very little fragmentation occurs.

6. Electrospray ionization INTRODUCTION:

• Electrospray ionization is a technique used in mass spectrometry to produce ions from

macromolecules such as proteins, polypeptides and oligonucleotides having molecular weights of

10,000 Da or more.

CONSTRUCTION & WORKING:

• The method generates ions from solution of a sample by creating fine spray of charged droplets.

• A solution of sample is pumped through a fine, charged stainless steel capillary needle at a rate

of few microlitres/minute.

9

Page 11: Ionizaion Techniques - Mass Spectroscopy

• The needle is maintained at a high electric field (several kilovolts) with respect to cylindrical

electrode.

• The liquid pushes itself out of the capillary as a mist or aerosol of fine charged droplets.

• In the set of aerosol droplets is produced by a process involving formation of Taylor cone and a jet

from the tip of this cone.

• These charged droplets are then passed through desolvating capillary where the solvent is

evaporated in the vacuum and attachment of charge to the analyte molecules takes place.

• Desolvating capillary uses warm nitrogen as nebulising gas.

• The desolvating capillary is maintained under high pressure.

• As the droplets evaporate the analyte molecules comes closer together.

• These molecules become unstable as the similarly charged molecules comes closer together and the

droplets explode once again. This is referred as Coulombic fission.

• The process repeats itself until the analyte is free from solvent and is lone ion.

• The ion then moves to the mass analyzer.

NOTE: In electrospray process, the ions observed are quassimolecular ions that are ionized by

addition of a proton (hydrogen ion) to give (M+H)+ or other cations such as sodium ion (M+Na)+ or

removal of hydrogen ion (M-H).

NOTE: Furthermore, multiple charged ions are often observed and these ions are even electron

species indicating that electrons have neither been added nor removed.

ADVANTAGES

Most important techniques for analysis of high molecular weight biomolecules such as

polypeptides, proteins, oligonucleotides and synthetic polymers.

Can be used along with LC and capillary electrophoresis.

10

Page 12: Ionizaion Techniques - Mass Spectroscopy

7. Matrix Assisted Laser Desorption (MALDI)

INTRODUCTION:

• Matrix assisted laser desorption is a technique in mass spectrometry for ionization of biomolecules

(polymers such as proteins, polypeptides and sugars) and synthetic polymers that are more fragile

and form fragments when ionized by conventional methods.

• It is most similar to ESI in both softness and ions produced.

A) Matrix

Matrix is used in MALDI to

Absorb the laser energy.

Prevent analyte agglomeration.

Protect analyte from being destroyed by direct laser beam.

Matrix consists of a crystallized molecules of which the most commonly used are :-

3,5 – dimethoxy – 4 – hydroxy cinnamic acid (sinapinic acid)

α – cyano – 4 – cinnamic acid (α – cyano or α – matrix)

2,5 – dihydroxy benzoic acid (DHB)

Preparation of Amtrix:

a) Solution of the matrix is made in a mixture of highly purified water and another organic

compound (acetonitrile or ethanol).

b) Triofluoro acetic acid (TFA) is also added.

c) If sinapinic acid is used as a matrix the solution is prepared by adding 20 mg/ml of

sinapinic acid, Water: acetonitrile: TFA (50:50:0.1)

d) Matrix solution is then mixed with the analyte to be investigated.

NOTE: The organic compound acetonitrile dissolves hydrophobic proteins present in the

sample while water dissolves hydrophilic proteins.

e) The solution is then spotted in a air tight chamber on the tip of the sample probe.

11

Page 13: Ionizaion Techniques - Mass Spectroscopy

f) With a vacuum pump the air is removed and vacuum is created which leads to

evaporation of the solvent leaving behind a layer of recrystalized matrix containing

analyte molecules.

Some of the more commonly used matrices are:

B) Laser

The solid mixture is then exposed to pulsed laser beam.

The matrix absorbs the laser energy and transfers some of this energy to the analyte

molecules which results in the sublimation of sample molecules as ions or the matrix after

UV MALDI Matrix List

Compound Abbreviation Mass (Da) Solvent Wavelength (nm) Applications

2,5-dihydroxy benzoic acid DHB

154

ACN, water, methanol, acetone, chloroform

337, 355, 266

peptides, nucleotides, oligonucleotides, oligosaccharides

3,5-dimethoxy-4-hydroxycinnamic acid

sinapic acid; sinapinic acid; SA

224

ACN, water, acetone, chloroform

337, 355, 266

peptides, proteins, lipids

4-hydroxy-3-methoxycinnamic acid

ferulic acid

194

ACN, water, propanol

337, 355, 266 proteins

α-cyano-4-hydroxycinnamic acid

CHCA

189

ACN, water, ethanol, acetone

337, 355 peptides, lipids, nucleotides

Picolinic Acid PA

123

Ethanol 266 oligonucleotides

3-hydroxypicolinic acid

HPA

139

Ethanol 337, 355 oligonucleotides

12

Page 14: Ionizaion Techniques - Mass Spectroscopy

absorbing the laser energy gets ionized and transfer part of this charge to the sample

molecules and ionize it.

Nitrogen or carbon lasers are most commonly used.

The ions produced in this process are quassimolecular ions that are ionized by addition of

proton (M+H)+ or a cation such as sodium (M+Na)+ or removal of a proton (M-H)-.

It generally produces singly charged ions in some cases doubly charged ions such as

(M+2H)2+ are also observed.

The chamber consists of two electrodes and the ions are produced between the electrodes.

When the polymers form cations the cathode is placed right behind the sample and anode in

front of the sample.

The cations get attracted towards the negatively charged anode. This acceleration is used to

move the ion to the detector.

When the polymer forms anions the electrodes are interchanged.

i) Atmospheric pressure-matrix assisted laser desorption

• AP-MALDI is a variant of MALDI which is carried out at atmospheric pressure (760 torr).

• AP-MALDI is performed using an instrument similar to ESI source with spray replaced by a

sample probe or MALDI target.

• Main difference MALDI and AP- MALDI is the pressure at which ions are produced.

Lasers Used for MALDI

Laser Wavelength(nm) Reference

Nitrogen Laser 337 (Tanaka 1988)

CO2 10600 (Overberg 1991)

13

Page 15: Ionizaion Techniques - Mass Spectroscopy

• In MALDI ions are produced at (10 mtorr) while in AP- MALDI ions are formed at (760 torr)

atmospheric pressure as a result AP- MALDI provides better and faster cooling which makes it

softer ionization technique than MALDI .

APPLICATIONS

Used in proteomics

Estimation of DNA, RNA and oligosaccharides.

Used in analysis of lipids, phosphopeptides and synthetic polymers.

8. Plasma desorption

INTRODUCTION

• Plasma desorption produces molecular ions from the samples coated on a thin foil when a highly

energetic fission fragments from the Californium-252 “blast through” from the opposite side of

the foil.

• The fission of Californium-252 nucleus is highly exothermic and the energy released is carried

away by a wide range of fission fragments which are heavy atomic ion pairs.

• Ion pair fission fragments depart in opposite directions.

• Each fission of this radio active nucleus gives rise to two fragments traveling in opposite

directions (because necessity of momentum conversation).

• A typical pair of fission fragments is 142Ba18+ and 106TC22+, with kinetic energies roughly 79 and

104 MeV respectively.

• When such a high energy fission fragments passes through the sample foil, extremely rapid

localized heating occurs, producing a temperature in the range of 10000K.

• Consequently, the molecules in this plasma zone are desorbed, with the production of both

positive and negative ions.

• These ions are then accelerated out of the source in to the analyzer system.

9. Laser desorption

INTRODUCTION:

• Laser desorption methods involves interaction of pulsed laser beam with the sample to produce

both vaporization and ionization.

• Laser beam is usually of different wavelengths from far U.V to far IR depending upon the sample

to be analyzed.

REQUIREMENTS

Laser wavelength must be at absorption wavelength of the molecule.

14

Page 16: Ionizaion Techniques - Mass Spectroscopy

In order to avoid decomposition absorbed energy must be quickly dispersed in the molecules.

IONIZATION TECHNIQUE:

• Ionization is carried out by two techniques :-

Microprobe techniques

Laser beam is focused to a very small spot on the back side of a thin metal foil that

holds a thin film of sample.

Ions emerge out on the front side from a small cratered hole in the foil.

Bulk analysis techniques

The technique uses a less focused beam and larger samples.

The laser beam produces microplasma that consists of neutral fragments with

elementary and fragment ions.

The ions produced are largely protonated and deprotonated species that have a

unit charge.

ADVANTAGES

Used for larger biomolecules such as proteins and carbohydrates.

DISADVANTAGE

Laser pulse lasts only for a few micro seconds, suitable mass analyzers are limited to

time-of-flight and fourier transform spectrometers.

Molecules of molecular weight less than 1000 Da for biopolymers and 9000 Da for

synthetic polymers cannot be studied as they get decomposed.

10. Fast Atom Bombardment INTRODUCTION:

• It is an ionization technique in which the analyte and non-volatile liquid matrix mixture is

bombarded by a high energy beam of inert gas such as Argon or Xenon.

• This technique is used for ionization of polar high molecular weight compounds such as

polypeptides.

• Commonly used matrices include :-

Glycerol

Monothioglycerol

Carbowax

2,4 – dipentyl phenol

3 – nitrobenzyl alcohol (3 – NBA)

• These solvents easily dissolve organic compounds and do not evaporate in vacuum.

15

Page 17: Ionizaion Techniques - Mass Spectroscopy

CONSTRUCTION & WORKING:

• The bombarding beam consists of Xenon or Argon atoms of high translational energy.

• This beam is produced by first ionizing the Xenon (or Argon atoms with electrons to give Xenon

radical cations.

Xe + e - = Xe.+ +2e-

• The radical cations are then accelerated to 6 – 10 KeV to give radical cations of high translational

energy (Xe)++, which are then passed through a chamber containing Xenon atoms at a pressure of

10-5 torr.

• During this passage high energy cation obtain electrons from Xenon atoms to become high energy

atoms (Xe).

• The lower energy ions are removed by electrostatic deflector.

(Xe)++ Xe.+ + Xe

(Xe).+ + Xe (Xe) + Xe.+

MATRIX PREPARATION:

• The analyte is dissolved in the liquid matrix such as glycerol and applied as a thin layer on the

sample probe shaft.

• The mixture is bombarded with the high energy beam of Xenon atoms.

• Xenon ionizes the glycerol molecules to give glycerol ions.

• These ions react with the surrounding glycerol molecules to produce (G+H)+ as reactant ions.

• The sample molecules then undergo proton transfer or hydride transfer or ion-pair interaction

with reactant ions to give quassimolecular or psuedomolecular ions such as (M+H)+, (M-H)- or

(M+G+H)+.

16

Page 18: Ionizaion Techniques - Mass Spectroscopy

• These ions are then extracted from slit lens system designed to collect ions and directed to mass

analyzer. ADVANTAGES

Used for ionization of polar high molecular weight samples.

Provides rapid heating of samples and reduces sample fragmentation.

Rapid ionization.

DISADVANTAGES

Difficult to distinguish between low molecular weight compounds.

Compounds must be soluble in liquid matrix.

Not good for multiply charged compounds.

11. Secondary ion mass spectrometry INTRODUCTION:

• Secondary ion mass spectrometry is nearly identical to FAB except the primary ionizing beam is

an ion beam rather than a neutral atom beam.

• The Cesium or Argon ions are most commonly used.

CONSTRUCTION & WORKING:

• The source consists of a cylindrical grid and a vertically placed ion gun or filament.

• Argon or Cesium gas is ionized by heating the filament to produce monoenergetic noble gas

ions.

• The ion gun can produce an ion beam of diameter ranging from 0.1mm to 1mm.

• The ions are accelerated to a potential of 300 to 3000 eV.

• This ion beam is then bombarded on to the surface of the sample.

• This results in the formation of secondary sample ions by charge transfer interaction between

the sample molecules and the primary gas ions.

• The ions formed in the cylindrical grid are then extracted from one end and focused on the

target or mass analyzer by an electrostatic lens system.

ADVANTAGES

Higher sensitivity

Selection of Beam diameter permits for rapid transition from a small surface analysis with a

small beam to a large surface area.

17

Page 19: Ionizaion Techniques - Mass Spectroscopy

12. Thermal ionization or Surface ionization • Thermal surface ionization source is useful for inorganic solid materials.

• Samples are coated on a tungsten ribbon filament and then the filament is heated until the sample

is evaporates.

• As the sample evaporates it undergoes ionization.

• The probability of ionization is predictable and is a function of work function of :-

Ionization potential of the sample E1

Work function of the filament material Φ

Filament temperature T

• This can be summarized as follows

n+/n0 = exp[z(Φ – E1)/KT] Where, z = electronic charge

K = Boltzmann’s constant

n+ = Number of ions formed

n0 = Number of neutral species

REFERENCES 1. Principles of Instrumental analysis. Fifth Edition by Douglas. A. Skoog, F. James Holler

and Timothy A. Nieman. Page No. 499 – 511.

2 Instrumental Methods Of Analysis. Seventh Edition by Willard Meritt. Page No. 468 –

74.