IOL POWER CALCULATION

NORMAL & POST LASIK EYES

INDOREDRISHTI.WORDPRESS.COM

Page 2: Iol power calculation normal and post lasik eyes

DR DINESH MITTAL DR SONALEE MITTAL

DRISHTI EYE HOSP VIJAYNAGAR INDORE

Page 3: Iol power calculation normal and post lasik eyes
Page 4: Iol power calculation normal and post lasik eyes
Page 5: Iol power calculation normal and post lasik eyes
Page 6: Iol power calculation normal and post lasik eyes
Page 7: Iol power calculation normal and post lasik eyes
Page 8: Iol power calculation normal and post lasik eyes

1 Cataract surgery has evolved into a refractive procedure with the goal of eliminating or significantly reducing the need for spectacle dependence.•2. One must consider not only the astigmatism induced by the cataract incision itself, but also the correction of preexisting astigmatism.•3. Incision length, depth, and distance from visual axis all affect astigmatism.

Page 9: Iol power calculation normal and post lasik eyes

•4. LRIs are commonly used to correct preexisting astigmatism, and•published nomograms are helpful in tailoring a surgical approach.•5. Toric IOLs and excimer laser ablation are alternative approaches to correcting astigmatism in the patient for refractive cataract surgery.•No single approach is best suited for all patients.

Page 10: Iol power calculation normal and post lasik eyes

INTRAOCULAR LENS CALCULATIONS•Choosing the appropriate IOL power is a major determinant of patient satisfaction with cataract surgery. •3 main factors •accurate measurements (biometry), •selecting calculations (formulas),• and assessing the patient’s needs to determine postoperative refractive target (clinical considerations).

Page 11: Iol power calculation normal and post lasik eyes
Page 12: Iol power calculation normal and post lasik eyes

BIOMETRY•At minimum, 2 measurements reqd to calculate implant power:•axial length & corneal curvature (keratometry) •Precise measurements critical •error of 0.3 mm in axial length will result in a 1-D error in IOL power.

Page 13: Iol power calculation normal and post lasik eyes

Axial Length•Axial length has been obtained utilizing A-SCAN ULTRASOUND. •This measurement is determined by calculation;• an ultrasound pulse is applied and the transit time through the eye is measured. Using estimated velocities of ultrasound waves through various media (ie, cornea, aqueous, lens, and vitreous) the distance travelled through the eye is calculated.

Page 14: Iol power calculation normal and post lasik eyes

• The instrument should have an oscilloscope screen to differentiate a good measurement from a poor one. Characteristic echo peaks or spikes should be observed when the probe is aligned properly • These include the following: a tall peak for the cornea,

tall peaks for the anterior and posterior lens capsule, tall peak for the retina, mod-erate peak for the sclera, and moderate-to-low peaks for orbital fat. If these spikes are not well seen, then the probe may be misaligned.

A SCAN ULTRASOUND

Page 15: Iol power calculation normal and post lasik eyes

•The contact A-scan technique must be performed carefully, as com-pressing the cornea will result in a shorter-than-expected measurement and multiple measurements should be taken and averaged. • If several are taken and differ by a significant amount, they should be discarded until consistent readings can be obtained.

Page 16: Iol power calculation normal and post lasik eyes

• It is also prudent to measure both eyes for comparison•the machine should be regularly calibrated, checking measurements against an eye of known axial length.

Page 17: Iol power calculation normal and post lasik eyes
Page 18: Iol power calculation normal and post lasik eyes

Immersion techniqueThe immersion technique may more accurately represent the true axial length because there is no corneal compression.

In this technique, the patient lies in the supine position and a scleral shell is placed on the eye and filled with Goniosol. The ultrasound probe is placed in this solution and the beam is aligned with the macula by having the patient look at the probe tip fixation light.

Although the immersion method may be strongly advocated by some users, applanation A-scan is the more commonly used method

Page 19: Iol power calculation normal and post lasik eyes

IMMERSION TECHNIQUE SHOWING PROBE IN OSSOINIG SHELL

Page 20: Iol power calculation normal and post lasik eyes

Optical biometry: IOLMaster• In the last decade, the technique of optical coherence biometry was in-troduced by Haigis, which utilizes light rather than ultrasound to measure the length of the eye. The first device introduced was the IOLMaster (Zeiss), based on the principle of partial coherence interferometry using a 780-nm multimode laser diode.

Page 21: Iol power calculation normal and post lasik eyes
Page 22: Iol power calculation normal and post lasik eyes

•Measurements taken without contact to eye, thus eliminating variability due to an examiner technique. •The distance measured lies between the anterior surface of tear film & retinal pigmented pigment epithelium (rather than the anterior surface of the cornea and internal limiting membrane ), which may be more physiologically accurate (refractive rather than anatomic axial length).

2 advantages to the IOLMaster

Page 23: Iol power calculation normal and post lasik eyes

patient asked to focus on a small red fixation light, & examiner maneuvers focusing spot within the measurement reticule, sampling areas until the best peak pattern is obtained. 5 to 20 measurements obtained until the readings differ by less than 0.1 mm. Maximal axial length measured 40 mm.

IOLMASTER USE

Page 24: Iol power calculation normal and post lasik eyes

IOL MASTER DISADVANTAGE

The primary disadvantage of this optical device is that any significant axial opacity, such as a corneal scar, dense posterior subcapsular plaque, darkly brunescent cataract, or vitreous hemorrhage, will reduce the signal-to-noise ratio (SNR) to the point that reliable measurements are not possible.

Page 25: Iol power calculation normal and post lasik eyes

A-scans should be remeasured under the following conditions•1. Axial length is less than 22 mm or greater than 25 mm in either eye.•2. The difference between the 2 eyes is greater than 0.3 mm.•3. The measurements do not correlate with the patient’s refraction (ie, hyperopes should have shorter eyes, and myopes should have longer eyes).

Page 26: Iol power calculation normal and post lasik eyes

Lens power calculation formulas evolved over past 30 yrs, since first theoretical formula for iris-supported iol published by Fyodorov in 1967.

Page 27: Iol power calculation normal and post lasik eyes
Page 28: Iol power calculation normal and post lasik eyes

First Generation• Initial formulas were based largely on axial length,

but with the availability of posterior chamber implants, consideration had to be given to the distance from the cornea to the implant (anterior chamber depth). By studying large numbers of cases, linear regression techniques were used to determine a formula for predicting emmetropic implant power. The most widely used regression formula was developed by Sanders, Retzlaff , and Kraff in 1980 and is known as the SRK formula.

Page 29: Iol power calculation normal and post lasik eyes

First Generation•P = A – (2.5 × AL) – (0.9 × K)

•where P is the lens implant power (diopters),• AL is the axial length (mm), •K is average keratometry (diopters), • and A is the constant (no units), a theoretical value that relates the lens power to axial length and keratometry. It is specific to the design of the IOL and its intended position inside the eye. This number is specified by the IOL manufacturer.

Page 30: Iol power calculation normal and post lasik eyes

Second Generation• The SRK formula is a linear equation derived by fitting collected data to a straight line. However, the optical system is nonlinear and begins to produce significant error with short or long eyes. To improve accuracy, the formula was modified, taking into consideration variation in axial length. • This en-hancement is known as the SRK II formula in which the A-constant is defined at different axial lengths:

Page 31: Iol power calculation normal and post lasik eyes

Second Generation SRK II •A = A + 3 (AL < 20 mm)•A = A + 2 (20 mm < AL < 21 mm) •A = A + 1 (21 mm < AL < 22 mm)• A = A (22 mm < AL < 24.5 mm) •A = A – 0.5 (24.5 mm < AL)

Page 32: Iol power calculation normal and post lasik eyes

Third Generation• Holladay further refined the theoretical formulas by proposing a relationship between the steepness of the cornea and the position of the IOL. • Instead of factoring in anterior chamber depth, formula would calculate the distance from the cornea to iris plane and & distance from the iris plane to the IOL.

Page 33: Iol power calculation normal and post lasik eyes
Page 34: Iol power calculation normal and post lasik eyes

Third Generation• . This second variable was termed the surgeon factor, was specific to each lens, and had the ability to be personalized and adjust for any consistent bias in the surgeon’s results. Hoffer achieved the same effect via another approach in his Hoffer Q formula. Retzlaff followed suit to take into consideration not only the position of the implant, but also incorporated a correction for the retinal thickness, thus developing the SRK/T formula in 1990.

Page 35: Iol power calculation normal and post lasik eyes

Fourth Generation• Haigis presented the notion that geometry of each IOL model should be considered in determining which formula to use. The geometry for a particular IOL model is not the same at all powers; therefore, the Haigis formula utilizes 3 lens constants to address it•a0 constant moves the power prediction curve up or down a1 constant is tied to the anterior chamber depth a2 constant is tied to the measured axial length

Page 36: Iol power calculation normal and post lasik eyes

•Optimization of the Haigis formula requires collecting pre- and postop-erative data from over 200 patients in order to allow surgeon-specific optimization, which is available online.• In the late 1990s, the Holladay 2 IOL consultant software was introduced to improve upon predictability by incorporating additional data points. It requires 7 measurements including white-to-white corneal diameter, anterior chamber depth, lens thickness, patient’s age, preoperative refraction, keratometry, and axial length. Th is formula may be more precise in unusual eyes such as those that have undergone refractive surgery.

Page 37: Iol power calculation normal and post lasik eyes

•One or more of the third-generation formulas (SRK/T, Holladay 1, and Hoffer Q) are generally programmed into A-scan biometers sold today. •The optical biometers are now incorporating the fourth-generation formulas (Haigis and Holladay 2).•Over time, some trends have emerged regarding which formulas to use in general categories of patients:

Page 38: Iol power calculation normal and post lasik eyes
Page 39: Iol power calculation normal and post lasik eyes

THEORETICAL FORMULAE BASED ON MATHEMATICAL PRINCIPLES REVOLVING AROUND THE SCHEMATIC EYEREGRESSION FORMULAE WORKING BACKWARDS ON POSTOPERATIVE OUTCOMES3RD AND 4TH GENERATION MIX OF BOTH

Page 40: Iol power calculation normal and post lasik eyes
Page 41: Iol power calculation normal and post lasik eyes
Page 42: Iol power calculation normal and post lasik eyes

IOL CALCULATION POST REFRACTIVE SURGERY

Page 43: Iol power calculation normal and post lasik eyes

INSTRUMENT ERRORINSTRUMENT ERROR IS BECAUSE KERATOMETERS READ AT 3.2 MM AND NOT AT THE CENTRE

Page 44: Iol power calculation normal and post lasik eyes

INDEX OF REFRACTION ERRORTHIS K READING ERROR OCCURS BECAUSE OF CHANGED RATIO OF RADIUS OF CURVATURE OF ANTERIOR AND POSTERIOR SURFACE OF CORNEA SECONDARY TO KERATO REFRACTIVE PROCEDURE

Page 45: Iol power calculation normal and post lasik eyes

UNOPERATED EYES IN UNOPERATED EYES THE RATIO BETWEEN ANTERIOR AND POSTERIOR SURFACE OF CORNEA IS 1.21

Page 46: Iol power calculation normal and post lasik eyes

MYOPIC OPERATED EYES

IN OPERATED EYE THIS RATIO CHANGES

Page 47: Iol power calculation normal and post lasik eyes

POST MYOPIA TREATED EYES KERATOMETRY OVER ESTIMATED

Page 48: Iol power calculation normal and post lasik eyes

POST HYPERMETROPIA TREATED EYES

Page 49: Iol power calculation normal and post lasik eyes

POST HYPEROPIA TREATED EYES KERATOMETRY UNDER ESTIMATED

Page 50: Iol power calculation normal and post lasik eyes

KERATOMETRY DEPENDENT & INDEPENDENT FORMULA

Page 51: Iol power calculation normal and post lasik eyes

HOW TO CALCULATE IOL POST KERATO REFRACTIVE SURGERY

Page 52: Iol power calculation normal and post lasik eyes

ARAMBERRI DOUBLE K METHOD

Page 53: Iol power calculation normal and post lasik eyes

PRE OP K READING FOR ELP

Page 54: Iol power calculation normal and post lasik eyes

HOW TO DETERMINE PRE OP K

Page 55: Iol power calculation normal and post lasik eyes

PRE OP K READING

Page 56: Iol power calculation normal and post lasik eyes

PRE OP K READING

Page 57: Iol power calculation normal and post lasik eyes

TRUE POST OP K READING FOR IOL POWER CALCULATION

Page 58: Iol power calculation normal and post lasik eyes

MEASUREMENT OF POST OP K READING

Page 59: Iol power calculation normal and post lasik eyes

POST OP K READING HISTORY DEPENDENT

Page 60: Iol power calculation normal and post lasik eyes

POST OP K READING HISTORY INDEPENDENT

Page 61: Iol power calculation normal and post lasik eyes

K CORRECTED METHODS

Page 62: Iol power calculation normal and post lasik eyes

K CORRECTED METHODS SHAMMAS

Page 63: Iol power calculation normal and post lasik eyes

K CORRECTED METHODS THEORETICAL

Page 64: Iol power calculation normal and post lasik eyes

K CORRECTED KOCH & WANG

Page 65: Iol power calculation normal and post lasik eyes

POST OP K READING TOMOGRAPHY METHODS

Page 66: Iol power calculation normal and post lasik eyes

TOMOGRAPHY METHODS

Page 67: Iol power calculation normal and post lasik eyes

HEIDELBERG RAYTRACING STUDY

FOUR FORMULAS ARE BESTHAIGISHOFFER QOLSEN 2RAYTRACING

Page 68: Iol power calculation normal and post lasik eyes

IOL CALCULATION BEST IN UNOPERATED EYES

Page 69: Iol power calculation normal and post lasik eyes

IOL CALCULATION BEST POST KERATOREFRACTIVE SURGERY

Page 70: Iol power calculation normal and post lasik eyes

ASCRS SITE FOR IOL POWER CALCULATION

Page 71: Iol power calculation normal and post lasik eyes