• Introduction – Why digital? – Why dual energy? • Experimental setup • Image acquisition • Image processing and results A silicon microstrip system A silicon microstrip system with the RX64DTH ASIC for with the RX64DTH ASIC for dual energy radiology dual energy radiology
43
Embed
Introduction –Why digital? –Why dual energy? Experimental setup Image acquisition Image processing and results A silicon microstrip system with the RX64DTH.
Introduction: why digital ? Digital radiography has well known advantages over conventional screen-film systems –Enhance detecting efficiency w.r.t. screen-film –Image analysis –Easy data transfer
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
• Introduction– Why digital?– Why dual energy?
• Experimental setup• Image acquisition• Image processing and results
A silicon microstrip system A silicon microstrip system with the RX64DTH ASIC for with the RX64DTH ASIC for
dual energy radiologydual energy radiology
1) University of Eastern Piedmont and INFN, Alessandria, Italy L. Ramello;
2) University and INFN, Torino, Italy P. Giubellino, A. Marzari-Chiesa, F. Prino;
3) University and INFN, Ferrara, Italy; M. Gambaccini, A. Taibi, A. Tuffanelli, A. Sarnelli;
4) University and INFN, Bologna, Italy G. Baldazzi, D. Bollini;
5) AGH Univ. of Science and Technology, Cracow, Poland W. Dabrowski, P. Grybos, K. Swientek, P. Wiacek;
6) University of Antwerp, Antwerp, Belgium P. Van Espen;
7) Univ. de los Andes, Colombia C. Avila, J. Lopez Gaitan, J.C. Sanabria;
8) CEADEN, Havana, Cuba A.E. Cabal, C. Ceballos, A. Diaz Garcia, L. Bolaños;
9) CINVESTAV, Mexico City, Mexico L.M. Montano;
The CollaborationThe Collaboration
Introduction: why digital ?Introduction: why digital ?• Digital radiography has well known advantages over
Introduction: why dual Introduction: why dual energy ?energy ?
Example 1: dual energy Example 1: dual energy mammographymammography
Example 1: dual energy Example 1: dual energy mammographymammography
E 15-20 keV:Signal from cancer tissue deteriorated by the adipose tissue signal
E 30-40 keVCancer tissue not visible, image allows to map glandular and adipose tissues
Example 2: angiographyExample 2: angiography•Angiography = X-ray examination of blood vessels
determine if the vessels are diseased, narrowed or blocked Injection of a contrast medium (Iodine) which absorbs X-ray
differently from surrounding tissues
•Coronary angiographyIodine must be injected into the heart or very close to itA catheter is inserted into the femoral artery and managed up
to the heart→Long fluoroscopy exposure time to guide the catheter→Invasive examination
•Why not to inject iodine in a peripheral vein?Because lower iodine concentration would be obtained,
requiring longer exposures and larger doses to obtain a good image
But, if the image contrast could be enhanced in some way…
Example 2: angiography at Example 2: angiography at the iodine K-edge (II)the iodine K-edge (II)
Iodine injected in patient vessels acts as radio-opaque contrast medium
Dramatic change of iodine absorption coeff. at K-edge energy (33 keV)
Subtraction of 2 images taken with photons of 2 energies (below and above the K-edge)→ in the resulting image only the iodine signal remains and all other materials are canceled
Experimental setupExperimental setup• To implement dual energy imaging we need:
• a dichromatic beam• a position- and energy-sensitive detector
Bragg Diffraction on Highly Oriented Pyrolitic Grafite Crystal
W anode tube
Double slit collimator
Two spatially separated beams with different energies E-E and E+E obtained in 2 separate beams
More on the dichromatic beamMore on the dichromatic beam
incidentspectraat 3 energysettings …
… spectra after 3 cm plexiglass
(measured with HPGe detector)
• Fully parallel signal processing for all channels• Binary architecture for readout electronics
1 bit information (yes/no) is extracted from each stripThreshold scans needed to extract analog information
• Counts integrated over the measurement period transmitted to DAQ
data, control
Silicon strip detector Integrated circuit
100 m
current pulses
X-rays
PC
N. I. I/O cards PCI-DIO-N. I. I/O cards PCI-DIO-96 96
and DAQCard-DIO-24and DAQCard-DIO-24
Experimental setup: Single Photon Experimental setup: Single Photon Counting SystemCounting System
Experimental setup: PCBExperimental setup: PCB
detectorpitch adapter
ASICs
PCB:- One 400 strip detector- Pitch adapter- 6 RX64 chips
384 equipped channels- connector to DAQ card
2 protoype detectors:a) 6 x Single threshold RX64b) 6 x Dual threshold RX64
Detecting systemDetecting system
Chip RX64 → counts incident photons on each strip of the detector
4 cm
6.4 mm10 strip = 1 mm
micro-bondings
Silicon microstrip detectoreach strip is an independent detector which gives an electric signal when an X-ray photon crosses it and interacts with a silicon atom
Knowing from which strip the electric signal comes from,the position of the incoming X-ray phonton is reconstructed.
Why silicon detectors?Why silicon detectors?Main characteristics of silicon detectors:• speed of the order of 10 ns• spatial resolution of the order of 10 m•small amount of material
0.003 X0 for a typical 300 m thickness
• excellent mechanical properties• good resolution in the deposited energy
3.6 eV of deposited energy needed to create a pair of charges, vs. 30 eV in a gas detector
Silicon sensor diodeSilicon sensor diode•The impinging ionizing particles generate electron-hole pairs •The impinging photons which interact in the detector volume create an electron (via Photoelectric, Compton or Pair Production)
•The electron ionizes the surrounding atoms generating electron-hole pairs
• Electron and holes drift to the electrodes under the effect of the electric field present in the detector volume. •The electron-hole current in the detector induces a signal at the electrodes on the detector faces.Metal contact
n+-type implant
n-type bulk
Charged particle -V
+V
electron
hole
P+-type implant
photon
photoelectron
ReversebiasE
Why reverse biased diode?Why reverse biased diode?•The amount of charge deposited in the silicon detector is very small
≈5500 electrons are produced by a 20 keV photons making photoelectric effect in the silicon
Forward-biased junction: the signal would be masked by the fluctuations of the current which the applied field makes flow even in high resistivity, hyper-pure silicon.Reverse-biased junction: allows to obtain the necessary electric field and only a very small “dark” current also at room temperature.
-V
+V
depleted region
Increasing the polarization voltage, it is possible to extend the depletion layer down to the backplane.
To have full efficiency, the polarization voltage must be high enough to deplete the full detector thickness (typically 300 m)
junction
NOT GOOD
Silicon Microstrips detectorsSilicon Microstrips detectors• A micro-strip detector is a silicon detector segmented in long, narrow elements.
•Each strip is an independent p-n reverse-biased junction• Provides the measurement of one coordinate of the particle’s crossing point with high precision (down to 1 m).
• K-edge subtraction imaging with contrast medium Cancel background structures by subtracting 2 images taken at energies just
below and above the K-edge of the contrast medium Suited for angiography at iodine (gadolinium) K-edge
- Cancel background structures to enhance vessel visibility Possible application in mammography (study vascularization extent)
- Hypervascularity characterizes most malignant formations • Dual energy projection algorythm
Make the contrast between 2 chosen materials vanish by measuring the logarithmic transmission of the incident beam at two energies and using a projection algorithm [Lehmann et al., Med. Phys. 8 (1981) 659]
Suited for dual energy mammography– remove contrast between the two normal tissues (glandular
and adipose), enhancing the contrast of the pathology– Single exposure dual-energy mammography reduces
radiation dose and motion artifacts
Dual energy imagingDual energy imaging
X-ray tube with dual energy output
Phantom
Detector box with 2 collimators
1.1. X-ray tube + mosaic crystal and 2 collimators to provide dual-energy output X-ray tube + mosaic crystal and 2 collimators to provide dual-energy output
2.2. Detector box with two detectors aligned with two collimatorsDetector box with two detectors aligned with two collimators
3.3. Step wedge phantom made of PMMA + Al Step wedge phantom made of PMMA + Al with 4 iodine solution filled with 4 iodine solution filled cavities of 1 or 2 mm diametercavities of 1 or 2 mm diameter
Angiography setupAngiography setup
15
10
5
0
pixe
ls
3002001000pixels
-0.8
-0.6
-0.4
-0.2
0.0
log
con
tegg
i
0 50 100 150 200 250 300 350
-1,0
-0,8
-0,6
-0,4
-0,2
0,0
0,2 Conc(I) = 370 mg/ml Measurement Simulation
ln[c
ount
(E=3
5.5K
ev)]
- ln[
coun
t(E=3
1.5K
ev)]
Strip Number
15
10
5
0
pixe
ls
3002001000pixels
161412108642
Con
tegg
i ( x
103 )
0 50 100 150 200 250 300 350
0,0
0,2
0,4
0,6
0,8
1,0 Conc(I) = 370 mg/ml E = 31.5 KeV
Measurement Simulation
Coun
ts /
Max
.Cou
nts
Strip Number
E = 31.5 keVE = 31.5 keV
15
10
5
0
pixe
ls
3002001000pixels
654321C
onte
ggi (
x103 )
0 50 100 150 200 250 300 350
0,0
0,2
0,4
0,6
0,8
1,0
Strip Number
Measurement Simulation
Conc(I) = 370 mg/ml E = 35.5 KeV
Coun
ts /
Max
.Cou
nts
E = 35.5 keV
5.3125.351 lnln NCNC logarithmic subtraction
Phantom structure not
visible in final image
Angiographic test results (I)Angiographic test results (I)
Possible decrease of iodine concentration keeping the same rad. dose
Angiographic test results (II)Angiographic test results (II)
Results with a second Results with a second phantomphantom
140 140
120 120
100 100
80 80
60 60
40 40
20 20
0 0
300
um p
ixel
300
300
200
200
100
100
0
0
100 um pixel
140 140
120 120
100 100
80 80
60 60
40 40
20 20
0 0
300
um p
ixel
300
300
200
200
100
100
0
0
100 um pixel
140 140
120 120
100 100
80 80
60 60
40 40
20 20
300
um p
ixel
300
300
200
200
100
100
0
0
100 um pixel
PhantomDigital Subtraction
Angiography
Dual Energy Angiography
smaller cavity (=0.4 mm) visible in DEA and not in DSA
Iodine conc. = 95 mg/ml
Dual energy projection Dual energy projection algorithmalgorithm
The mass attenuation coefficient μ of any material at a given energy E is expressed as a combination of the coefficients of any two suitable materials and :
E
aEaE
21
The logarithmic attenuation M = μξtξ of the material of thickness tξ is measured at two different energies: low (El) and high (Eh):
lhlh
lhhl
lhlh
hllh
hhh
lll
EEEEEMEMA
EEEEEMEM
A
EAEAM
EAEAM
2
1
21
21
A1 and A2 represent the thicknesses of the two base materials which would provide the same X-ray attenuation as material ξ.
C
C90°
M1
R
1
M2
2
If a monochromatic beam of intensity I0 goes through material ξ which is partly replaced by another material ψ …
I0
I1 I2
ξψ
… then the vertexes of log. attenuation vectors M2 (material ξ) and M1 (mat. ξ + ψ) lie on a line R which is defined only by the properties of materials α, β, ξ and ψ. Projecting along direction C, orthogonal
to R, with the contrast cancellation angle :
… it is possible to cancel the contrast between materials ξ and ψ: both M1 and M2 will project to the same vector
A2
A1
Dual energy projection Dual energy projection algorithmalgorithmThe logarithmic attenuation M in a given pixel can be represented
as a vector having components A1 and A2 in the basis plane, the modulus will then be proportional to the gray level of that pixel
sincos 21 AAC
Mammographic phantomMammographic phantom• Three components: polyethylene (PE), PMMA and
water to simulate the attenuation coeff. (cm-1) of the adipose, glandular and cancerous tissues in the breast
S. Fabbri et al., Phys. Med. Biol. 47 (2002) 1-13
E _fat _gland _canc
20 .456 .802 .844
40 .215 .273 .281
E μ_PE μ_PMMA μ_water
20 .410 .680 .810
40 .225 .280 .270
300
pixel
350
300
250
200
150
100
50
0
pixe
l
Image processing (1)Image processing (1)
300pixel
350
300
250
200
150
100
50
0
pixe
l
Low thr. High thr.
Measured (raw)
300pixel
350
300
250
200
150
100
50
0
pixe
l
16 keV
300pixel
350
300
250
200
150
100
50
0
pixe
l
32 keVHE and LE imagesCorrect for:
1. pixels with huge n. of counts (bad counter conversion)
2. dead pixels3. X-ray beam fluctuations4. subtract high threshold
image from low threshold one
5. correct for spatial inhomogeneities of beam and detector extracted from flat-field profiles
300pixel
350
300
250
200
150
100
50
0
pixe
l
300pixel
350
300
250
200
150
100
50
0
pixe
l
16 – 32 keV 18 – 36 keV
300pixel
350
300
250
200
150
100
50
0
pixe
l
300pixel
350
300
250
200
150
100
50
0
pixe
l
1= PMMA 2=water3=PE 4=(water+PE)
Image processing (2)Image processing (2)
Low statistics due to:1) 2nd order harmonic2) dectecting efficiency
Simulation with MCNPSimulation with MCNP
1=detector2=PMMA3=water4=PE
MCNP-4C simulation with ENDF/B-VI library• Photons and electrons
tracked through the phantom and the detector (including the inactive region in front of the strips)
• Energy deposition in each strip recorded
• histogram of counts vs. strip number filled
Top View
Side View
Experiment vs. Simulation (1)Experiment vs. Simulation (1)RX64DTH 16 – 32 keV
300pixel
350
300
250
200
150
100
50
0
pixe
l
300pixel
350
300
250
200
150
100
50
0
pixe
l
simulation 16 – 32 keV
300pixel
350
300
250
200
150
100
50
0
pixe
l
300pixel
350
300
250
200
150
100
50
0
pixe
l
Experiment vs. Simulation Experiment vs. Simulation (1)(1)
3400
3200
3000
2800
2600
2400
2200
3002001000
Simul.16 keV Left Part Meas.16 keV Left Part
3400
3200
3000
2800
2600
2400
2200
3002001000
Simul.16 keV Right Part Meas.16 keV Right Part
3100
3000
2900
2800
2700
2600
3002001000
Simul.32 keV Left Part Meas.32 keV Left Part
3100
3000
2900
2800
2700
2600
3002001000
Simul.32 keV Right Part Meas.32 keV Right Part
Results (1): SNR vs. proj. Results (1): SNR vs. proj. angleangle100
Conclusion and OutlookConclusion and Outlook• We have developed a single photon counting silicon detector equipped
with the RX64DTH ASIC, with two selectable energy windows• The energy resolution of 0.8 keV (rms) is well adapted for dual energy
mammography and angiography• We have performed mammography imaging tests with a three-material
phantom– We have demonstrated the feasibility of contrast cancellation between two
materials, enhancing the visibility of small features in the third one• We have performed angiography imaging tests with 2 different phantoms
and iodine contrast medium– We have demonstrated the feasibility of logarithmic subtraction between two
images, enhancing contrast of small vessels also with lower iodate solution concentrations
• OUTLOOK: – Increase photon statistics at high energy, optimize exposure conditions– New detector materials, CZT?– Tests with a more realistic mammographic phantom