Intra-Operative Brain Function Monitors

Thomas Jan

Page 2: Intra-Operative Brain Function Monitors Thomas Jan.

Goals of General Anesthesia

• Anxiolysis• Analgesia• Hypnosis• Muscle Relaxation• Suppression of

Somatic/Autonomic Responses

Page 3: Intra-Operative Brain Function Monitors Thomas Jan.

Intra-Operative Awareness• Patient is not given enough general anesthesia and can recall events

during surgery• Incidence is between 2/1000 to 10/1000 per year

– Incidence is greater in OB, Trauma, and Cardiac cases• IOA heavily depends on interpatient pharmokinetic and

pharmodynamic variability

Page 4: Intra-Operative Brain Function Monitors Thomas Jan.

Intra-Operative Awareness Causes

• Equipment failure• Inadequate anesthesia• Patient co-morbidities

Page 5: Intra-Operative Brain Function Monitors Thomas Jan.

Intra-Operative Awareness Consequences

• Anxiety, anger, depression, irritability, anguish• Post-op traumatic neurosis• Recurrent nightmares• Insomnia• Difficulty concentrating• Exaggerated startle response• Unnatural preoccupation with death• 35% of patients had informed their anesthesia

care provider about their awareness and recall.

Page 6: Intra-Operative Brain Function Monitors Thomas Jan.

Monitoring Anesthetic Depth

• Indirect– Clinical signs– Respiratory Sinus Arrhythmia

• Direct– Isolated Forearm Technique– Auditory Evoked Potentials– Facial EMG– Electroencephalogram, Brain Function Monitors

Page 7: Intra-Operative Brain Function Monitors Thomas Jan.

Electroencephalogram

• Processed EEG signal– BIS Monitor, State Entropy, Narcotrend, Patient

State Analyzer, SNAP, Danameter• Goal: Targets brain activity and measures

depth of anesthesia

Page 8: Intra-Operative Brain Function Monitors Thomas Jan.
Page 9: Intra-Operative Brain Function Monitors Thomas Jan.
Page 10: Intra-Operative Brain Function Monitors Thomas Jan.
Page 11: Intra-Operative Brain Function Monitors Thomas Jan.

BIS Monitors decrease awareness

• Multi-center study in 2004; 2500 patients; Australia

• BIS decreases awareness in high risk patients– 2 cases of awareness in BIS

group

– 11 cases of awareness in control group

• NNT 138 patients

• Cost of prevention $2200 per one case of awareness

Page 12: Intra-Operative Brain Function Monitors Thomas Jan.

Can the BIS value go too low?• Prospective Observational Study;

1064 patients; Florida teaching hospital– BIS < 45 was one of the predictors of

mortality after non-cardiac surgery (RR 1.244)

– Mortality 5.5% all patients; 10.5% in > 65 years of age

• Predictors: – Patient co-morbidities

– Cumulative Deep Hypnotic Time

– Intraoperative SBP<80 mm

Page 13: Intra-Operative Brain Function Monitors Thomas Jan.

Brain Function MonitorsADVANTAGES• Monitor of sedation and hypnosis• Prediction of brain activity• Less anesthetic used• Easier wakeups

DISADVANTAGES• Does not measure movement• Does not predict hemodynamic response• Physiologic interference• Interference from other equipment

Page 14: Intra-Operative Brain Function Monitors Thomas Jan.

Conclusion• Studies have shown that brain function monitors can

decrease intra-operative awareness• These devices are best used in high risk situations: Trauma,

Obstetrics, Cardiac Surgery, previous history of awareness and explicit recall

• Could assist in delivery of multi-modal anesthetics in complex cases

• Remaining problems:– Opioid adjuncts have no effect on BIS, but lower hypnotic agent

requirements

– False sense of security

• Understand all the monitors you use in the OR!!!

Page 15: Intra-Operative Brain Function Monitors Thomas Jan.

References• S Muravchick, "Defining and measuring the anesthetic state," in The

Anesthetic Plan: From Physiologic Principles to Clinical Strategies (St Louis: Mosby-Year Book, 1991) 31.

• W H Liu et al, "Incidence of awareness with recall during general anaesthesia," Anaesthesia 46 (June 1991) 435-437

• N Moerman, B Bonke, J Oosting, "Awareness and recall during general anesthesia. Facts and feelings," Anesthesiology 79 (September 1993) 454-464.

• Different Conditions that could result in the bispectral index indicating an incorrect hypnotic state, Dashabe, Ashraf A, Anesth Analg 101(3);Sept 2005:765-73

• Anesthetic management and one-year mortality after non-cardiac surgery, Terri G. Monk et al, Anesth Analg 2005;100:4-10

• The Incidence of Awareness During Anesthesia: A multicenter United States Study, P.S. Sebel et al, Anesth Analg 2004;99:833-9