CSEM SA | Rue Jaquet-Droz 1 | CH-2002 Neuchâtel | T +41 32 720 5111 | F +41 32 720 5700 [email protected] www.csem.ch PULSEMON provides low-noise amplification chains and integrates high resolution analog to digital converters (ADC) as well as LED drivers and an acquisition sequencer. It operates with ultra-low power at both the circuit and system level. Integrated Pulse Oxymeter and Heart Rate Monitor Chip, architecture & implementation Photo-plethysmography and applications KEY FEATURES Integrated LED drivers with programmable current from 2 mA to 140 mA (up to 3 LEDs) Programmable amplification gain and cutoff frequency Multi-channel (multi-sensor) measurement (3 channels in parallel) Separate fully programmable sample rate for each ADC and each LED Low-power heart rate mode (<0.8 mW including LED current ) Ultra-fast <1 second data output capability from shutdown mode High-resolution receive channel with >14.5 noise-free bits Active ambient light removal APPLICATIONS Sport measurement devices (chest free watches) Wellness applications (long term measurements on body) Medical or paramedical vital sign monitoring applications Watch prototype PPG principle
1

Integrated Pulse Oxymeter and Heart Rate Monitor

Apr 27, 2022

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Integrated Pulse Oxymeter and Heart Rate Monitor

CSEM SA | Rue Jaquet-Droz 1 | CH-2002 Neuchâtel | T +41 32 720 5111 | F +41 32 720 5700

[email protected] www.csem.ch

PULSEMON provides low-noise amplification chains and integrates high resolution analog to digital converters (ADC) as well as LED drivers and an acquisition sequencer. It operates with ultra-low power at both the circuit and system level.

Integrated Pulse Oxymeter and Heart Rate Monitor

Chip, architecture & implementation

Photo-plethysmography and applications

KEY FEATURES

• Integrated LED drivers with programmable current from 2 mA to 140 mA (up to 3 LEDs)

• Programmable amplification gain and cutoff frequency

• Multi-channel (multi-sensor) measurement (3 channels in parallel)

• Separate fully programmable sample rate for each ADC and each LED

• Low-power heart rate mode (<0.8 mW including LED current )

• Ultra-fast <1 second data output capability from shutdown mode

• High-resolution receive channel with >14.5 noise-free bits

• Active ambient light removal

APPLICATIONS

• Sport measurement devices (chest free watches)

• Wellness applications (long term measurements on body)

• Medical or paramedical vital sign monitoring applications

Watch prototypePPG principle