Integrated PET/MRI

Ciprian Catana, MD, PhD

[email protected]

Page 2: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Outline

PET/MRI:

Brief history and current state-of-the-art

Methodological advances enabling new

science:

What MR can do for PET

What PET can do for MR

Potential research and clinical

applications

Page 3: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Representative paths in tissue of a 2 MeV electron

(Raylman et al. Int J Rad Onc Biol Phys, 1997; 37(5))

“Physics” motivation for simultaneous PET/MRI -

positron range error reduction in magnetic fields

e+

e+ e–

511 KEV

PHOTON

511 KEV

PHOTON

ANNIHILATION

ERROR

DUE TO

POSITRON

RANGE

q – particle charge; V – positron velocity; B –

magnetic field; Et – component of the positron

kinetic energy (MeV) perpendicular to the

magnetic field;

mp – rest mass of the positron

2 ) 2 (

334 . 0 t t p E E m

B R + =

= B x V q F Lor

*In-plane PET scanner spatial

resolution might improve at high

magnetic fields (i.e. B>7T), for high

energy positron emitters (i.e. 82Rb,

Et=3.15 MeV) if it is not dominated by

other factors (i.e. crystal size, non-

colinearity, etc.)

Page 4: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

PET effects on the MR:

No ferromagnetic components allowed;

Disturb homogeneity of the B0 field;

RF interference with the MR Tx/Rx coils;

Susceptibility artifacts and eddy currents. …

MR effects on the PET:

PMTs very sensitive to magnetic fields;

RFI, heating, vibrations, etc. …

General considerations:

Space constraints inside the MR;

Cost !

Integrating PET and MRI does not mean placing

an existing PET scanner inside an MR scanner.

Page 5: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Integrated PET/MR scanners have been

developed for small animal and human imaging.

(Wehrl HF, Judenhofer MS, Wiehr S, Pichler BJ,

Eur J Nucl Med Mol Imaging 36 (Suppl 1): S56-58; 2009, updated)

2011

Page 6: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

PET-MR(-CT) Scanners Available for Human Use

GE PET/CT Patient Transport GE MR

Philips Ingenuity TF PET/MR Siemens Biograph mMR

Page 7: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Outline

Integrated PET/MRI:

Brief history and current state-of-the-art

Methodological advances enabling new

science:

What MR can do for PET

What PET can do for MR

Potential research and clinical

applications

Page 8: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

What MR can do for PET

0.E+00

1.E-03

2.E-03

3.E-03

4.E-03

5.E-03

6.E-03

0 200 400 600Time (sec)

Counts (a.u.)

3. Partial volume effects corr.

2. Motion correction

4. Arterial input function estimation

1. Attenuation correction

Page 9: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Annihilation photons can interact with the

subject before reaching the detectors.

2 1

x

μ )(

2

xDep = xep =1

DxDx

coinc eeeppp === )(

21.

DeCorrAtten =..

xeIxI = )0()(

Uniform cylinder before and after AC (Ø=20 cm, µ=0.096 cm-1)

x D

Page 10: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

10

Estimating the tissue linear attenuation

correction factors from MR is difficult.

A number of factors have to be considered for implementing an accurate

MR-based attenuation correction method.

Separating the bone from air-filled cavities is the most challenging task using

conventional MR sequences.

MR Scaled CT Segmented CT

Page 11: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

11

TE = 2.46 ms TE = 0.07 ms

Ultra-short echo time (UTE) sequences

can be used for bone imaging.

Page 12: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

12

MR-based PET attenuation correction

using DUTE data

ATTENUATION MAPS

CTsegmented (top) and DUTEsegmented

(bottom)

RECONSTRUCTED PET IMAGES

Attenuation correction factors derived from the

CTsegmented (top) and DUTEsegmented (bottom).

(C. Catana et al, J Nucl Med, 2010)

Page 13: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

13

Skull/soft tissue/air segmentation can be

achieved from DUTE and MPRAGE data.

Segmented CT

Atlas-T1-DUTE

Clare Poynton

Page 14: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

14

Atlas-based methods allow generation of

continuous-valued attenuation maps.

MPRAGE

David Izquierdo

Page 15: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

15

The attenuation caused by the RF coils has to

be accounted for in an integrated scanner.

1.E+01

1.E+02

1.E+03

1.E+04

1.E+05

1.E+06

-60 -50 -40 -30 -20 -10 0 10 20Relative change (%)

no coil 3 mm X3 mm Y 3 deg Z

-5

5

-50

50

3 mm trans.

along Y axis

3°rot.

about Z axis

3 mm trans.

along X axis

Coil attenuation

ignored

Relative change images and histograms

BrainPET

gantry

Patient

table

RF coil

BrainPET prototype inside the 3T MR

scanner (MGH installation)

Page 16: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

16

Motion is difficult to avoid in long PET

studies.

Spatial resolution “loss”

image blurring/artifacts;

attenuation/emission data mismatch;

inaccurate quantification.

Page 17: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

17

Various motion correction approaches have

been investigated for neuroPET studies.

1. Eliminate head motion

2. Inter-frame correction

3. Event-by-event correction

Page 18: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

18 MR before and after MC

( Andre J.W. van der Kouwe et al, Magnetic

Resonance in Medicine 2006; 56: 1019-1032 )

Navigators introduced in standard MR sequences

provide high temporal resolution motion estimates.

Page 19: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

19

Motion correction algorithm for dynamic studies

on the BrainPET prototype.

(C. Catana et al, JNM, 2011)

Page 20: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

20

MR-derived motion estimates can be used

to retrospectively correct the PET data.

• Healthy volunteer;

• ~5 mCi 18FDG;

• Simultaneous MR-PET study;

• PET MC applied in LOR space.

PET data before (left) and after (right) MR-assisted motion correction

(C. Catana et al, JNM, 2011)

Motion estimates derived from EPI

MR series every 3 seconds

Page 21: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

In and out pressure

from ventilator

Gel with

background activity

Inflatable balloon

Hot spheres attached to balloon (10mm diameter)

Vent

Methods: Phantom Study in PET-MR

PET External

trigger

MR

“Beating Phantom”

Phantom

Guerin and El Fakhri, Med Phys 2011

Page 22: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Methods: Tagged MR

GRE sequence GRE sequence with tagging

Tagging patterns provide additional motion information.

Courtesy G. El Fakhri, MIPI, MGH

Page 23: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Phantom Results: Reconstruction

Sphere 2:

Diameter 10mm

Resolution 2mm each

Max motion ~3.0mm

(small motion) MR motion corrected

Uncorrected Gated 90

0

C=1.2

σ=19%

C=1.7

σ=48%

C=2.0

σ=20%

Chun, El Fakhri, JNM 2011

Page 24: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

• Motion Correction with Primate in simultaneous PET-MR

Gated tagged MR Gated PET

Primate Results

Chun, El Fakhri, JNM 2011

Page 25: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Primate Results

Chun et al. JNM 2012, 53(8):1284-1291

Page 26: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

26

Partial volume effects (PVE) affect PET

data quantification.

0

10

20

30

40

50

60

70

80

90

100

original 3 mm 5 mm 7 mm

To account for PVEs, information about the size of the structures of interest and

the spatially variant point spread function of the PET scanner is needed.

Page 27: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

27

The high resolution morphological MRI

data can be used for PET PVE correction.

Automated brain structures segmentation from the MPRAGE data.

Page 28: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

28

The high resolution morphological MRI

data can be used for PET PVE correction.

Morphological MR (ME-MPRAGE)

Original PET

Recovery coefficients

PET after regional PVE correction

Spencer Bowen, MGH

Page 29: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

29

Dynamic PET frames

FDG AIF

The non-invasive estimation of the radiotracer

input function can be improved using MR data. 0.E+00

1.E-03

2.E-03

3.E-03

4.E-03

5.E-03

6.E-03

0 200 400 600Time (sec)

Counts (a.u.)

50 100 150 200 250

50

100

150

200

250

50 100 150 200 250

20

40

60

80

100

120

140

TOF MR

ME-MPRAGE

Daniel Chonde, MGH

Page 30: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

30

MR-based

attenuation corr.

Compartment model

&

Kinetic analysis

Accurate tissue time activity curves

&

radiotracer AIF estimate

Improved PET data quantification

Partial volume

effects corr.

MR-based

motion corr.

Image-based AIF

estimation

What MR can do for PET

Page 31: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

31

What PET can do for MR

PET-guided MR imaging

In-vivo quantification of “smart” MR probes

Techniques cross-calibration and validation

Page 32: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

32

Fibrin Targeted Gd-based Contrast Agent

Fibrin-

specific

Peptide

Bis Gd

chelate

Bis Gd

chelate Linker

Improved fibrin affinity. High relaxivity: 4 Gd + protein binding.

No metabolism issues. Minimal Gd retention.

N H

H N

N H

H N

N H

H N

N H

O

O

O

O

O

O

O H N

O

N

O

N H

O

O N H

S

O H

S

O O N H 2 O

C l

O H

H N

O H

O H

O H N

H N N H

O

H N

N H

N N

N N

O

O

O

O

O

O

O

O G d

O H 2

N N

N N

O

O

O

O

O O

O

O

G d O H 2

O

O

N N

N N

O

O

O

O

O O

O

O

G d O H 2

O

N N

N N

O

O

O

O

O

O

O

O G d

O H 2

O

E P - 2 1 0 4 R

N a 5

Overoye-Chan et al., J Am Chem Soc 2008, 130:6025-39

Page 33: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

33

A B C

Image pre contrast

agent

Image post contrast

agent Arterial anatomy

(Ritika Uppal, Ilknur Ay, Peter Caravan)

Embolic Stroke Model

Page 34: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

34

Fibrin-targeted MR-PET Probe

Focused MR study

Whole-body PET “screening”

High sensitivity PET can guide the high

resolution MR study.

Uppal, Caravan, Radiology, 2011;258(3):812-20

Cu

Gd Gd

Gd

Peptide

Page 35: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

35

Relaxivity is sensitive to many molecular factors

that can be modulated to induce an MR signal.

Enzyme activation change property of contrast agent

(Louie, Meade et al, Nature Biotech, 2000)

A) GFP mRNA injection in R; B) EgadMe, β-gal injection in

R; C) β-gal staining

Page 36: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

36

pH-responsive “smart” MR probe

(Kalman, Sherry, Caravan et al, Inorg Chem 2007, 46,5260;

Ali, Sherry, Caravan et al, Chem Eur J 2008, 14:7250-8)

Low pH: more exchangeable protons (blue)

proton exchange rate ideal

High pH: fewer exchangeable protons (blue)

proton exchange too fast

Page 37: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

37

The MR signal depends on both relaxivity and

contrast agent concentration.

0

1

1

1

1][

1

TGdr

T+=

contrast agent

concentration

relaxivity

Two unknowns, one measurement

Page 38: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

38

Responsiv

e P

robe

Quantification

MR

PET

Dual

probe

(L. Frullano, C. Catana, T. Benner, A. Sherry, P.

Caravan. Angew. Chem. Int. Ed. 2010; 49:2382-4)

pH (electrode)

pH

(M

R-P

ET

) y = 1.01x

R2 = 0.95

6

6.5

7

7.5

8

6 6.5 7 7.5 8

A “smart” dual-modal MR-PET agent can

quantitatively and non-invasively measure pH.

Page 39: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

39

Cross-validate PET and MRI cerebral perfusion

measurements.

[O15]H2O PET and steady-state ASL MR

(F.Q.Ye et al. MRM 2000; 44:450-456)

(Y Ozsunar and AG Sorensen et al;

Topics in MRI 2000; 11(5): 259-272)

Bolus tracking technique

(L Ostergaard. JMRI 2005, 22:710-7)

Arterial spin labeling technique

Page 40: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

40

Validate and model the relationship between

OEF and BOLD signal during neuronal

activation.

(H. Ito et al. J Cerebral Blood Flow and Metab. 2005; 25: 371-377)

Page 41: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

41

MR-based

attenuation corr.

Improved PET data quantification

Partial volume

effects corr.

MR-based

motion corr.

Image-based AIF

estimation

MR and PET Can Help Each Other

PET-guided

MR imaging

Quantification of

MR probes

Techniques

cross-calibration

Multimodal quantitative MR-PET imaging

Page 42: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

42

Automated Data Processing and Analysis

Dan Chonde, MGH

Page 43: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

43

Outline

Integrated PET/MRI:

Brief history and current state-of-the-art

Methodological advances enabling new

science:

What MR can do for PET

What PET can do for MR

Potential research and clinical

applications

Page 44: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Potential Applications

How can MR-PET …

1) increase diagnostic accuracy?

2) improve patient experience?

3) advance scientific discovery?

How can MR-PET …

1) increase diagnostic accuracy?

oncology, cardiology, neurology

2) improve patient experience?

3) advance scientific discovery?

Page 45: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Most frequent site of

metastases*

PET/MRI is relevant for (favorable to PET/CT) …

Tumor entity Brain Lung Liver Bone Staging

category

Special objective/prognostic factor

Head and neck SCC − + − + T Extracapsular spread; bone infiltration

Non-small cell lung

cancer

+ + + + M Distant metastases

Breast cancer + + + + T/M Primary diagnosis and T-stage (benefit compared with

PET/CT; potential benefit over MRI mammography alone

is questionable); distant metastases

HCC − + + + T Pretransplantation evaluation

Colorectal

carcinoma

− + + − T/M Circumferential resection margin; liver metastases; tumor

regression rate to neoadjuvant therapy

Soft-tissue sarcoma − + + − T/M Tumor size and depth of infiltration defines T category;

muscular, neurovascular, and bone invasion

Primary bone

tumors

− + −

− T Presurgical evaluation (e.g., neurovascular invasion);

exact tumor size and response to neoadjuvant treatment

Melanoma + + + + M Exact number and location of metastases for presurgical

evaluation

Lymphoma M Extranodal dissemination; early therapy response

assessment

“Indications in which PET/MRI may be favorable

over PET/CT, depending on tumor entity”

*Frequency of metastatic spread (frequently [+], rare [−]) is according to AJCC Cancer Staging Manual, seventh edition; PET/CT and

PET/MRI are considered equally accurate for N-staging, and thus importance of N-staging is not discussed.

C. Buchbender et al. Oncologic PET/MRI, Part 1 & 2. JNM 2012; 53: 928-938 &1244-1252

Page 46: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Head & Neck MR-PET

More precise metabolic-anatomic allocation of the FDG-avid lesion

FDG-PET/CT

FDG-PET/MR

C. Buchbender et al.

J Nucl Med 2012; 53:928-

938

54-y-old man with

gingival SCC arising

from maxilla

Page 47: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Spencer Bowen, MGH

Accurate spatial registration

MRscan 1 PETscan 1 Fused MRscan 1-PETscan 1

MRscan 2 Fused MRscan 2-PETscan 1

Breast MR-PET

Page 48: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Liver PET/CT + MRI

P. Veit-Haibach, F.P. Kuhn, F. Wiesinger, G. Dalso and G. van Schulthess.

Magn Reson Mater Phy (2013) 26: 25-35

Increased lesion conspicuity

Page 49: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Pelvis MR-PET

DWI MRI

T1-weighted post-contrast MRI FDG PET PET/MRI fusion

ADC map

Collaboration with A. Guimaraes, MGH

Improved soft tissue discrimination and functional information

Page 50: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Whole-body PET/MRI

Total body bone scan with 219 MBq 18F-NaF of a

61-year-old female (76 kg), with 80 s per bed

position for PET acquisition and 105 min uptake

time.

Images courtesy of HZDR, Dresden

Data acquired using the Philips WB PET/MR

scanner

Magn Reson Mater Phy (2013) 26:5-23

Page 51: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

“Advantageous Features of Combined PET/MR

with Focus on Cardiac Applications”

C. Rischpler, S.G. Nekolla, I. Dregely and M. Schwaiger. Hybrid PET/MR Imaging of the

Heart: Potential, Initial Experiences, and Future Prospects. JNM 2013; 54:402-415

Page 52: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

52

Myocardial Perfusion MR-PET Study

13NH3

PET

Gd-DTPA

MR

Fused

MR-PET

C. Rischpler, S.G. Nekolla, I. Dregely and M. Schwaiger. Hybrid PET/MR Imaging of the

Heart: Potential, Initial Experiences, and Future Prospects. JNM 2013; 54:402-415

Page 53: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

53

Simultaneous 13NH3-PET/MR study

demonstrates stress-induced ischemia.

13NH3-PET Stress LGE MRI Fused LGE MRI with 13N-PET Stress

Anterior/anterolateral ischemia (PET) without significant delayed contrast enhancement (MRI).

Left anterior descending artery disease was subsequently confirmed by catheterization.

Courtesy of J. Lau, R. Laforest, P. Woodard

Page 54: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

54

Simultaneous FDG-PET and DCE MR study

allows myocardial viability assessment.

MR FDG-PET Fused PET/MR Cine

Delayed contrast enhancement MR and FDG-PET images acquired in diastole.

Fused cine created from the PET list mode data binned into 8 phases fused with

simultaneously acquired free-breathing real time SSFP cardiac cine.

[J. Lau, R. Laforest, S. Sharma, J. McConathy, A. Priatna,

L. Amado, R. Gropler, P. Woodard. ISMRM 2013, Oral: #0573]

Page 55: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Potential Benefits

How can MR-PET …

1) increase diagnostic accuracy?

2) improve patient experience?

3) advance scientific discovery?

Page 56: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

MR-PET can improve patient experience

• Two exams in one session:

– Increased patient compliance

– Reduced need for sedation/anesthesia in pediatric

patients

– One pharmacological challenge for two exams

• Reduced radiation exposure:

– Pediatric patients, women of childbearing age,

chronic patients

Page 57: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

The radiation dose from CT vs FDG-PET is

significantly higher in children compared to adults.

(F.H. Fahey, “Dosimetry of Pediatric PET/CT”. J Nucl Med

2009; 50: 1483-1491)

(H. Jadvar et al, “PET and PET/CT in Pediatric

Oncology”. Semin Nucl Med 2007; 37:316-331)

Page 58: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:
Page 59: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

F.W. Hirsch, B. Sattler et al. University of Leipzig

15-yo boy with left testicular tumor with retroperitoneal,

supraclavicular, hepatic and lung metastases

Page 60: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Potential Benefits

How can MR-PET …

1) increase diagnostic accuracy?

2) improve patient experience?

3) advance scientific discovery?

Page 61: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

61

Simultaneous MR-PET opens new opportunities

for studying the brain.

(Figure from “Imaging of the Human Brain in Health and Disease”, J.E. Johnson ed.)

Page 62: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

Neuropsychiatric conditions contribute the most to the overall

burden of non-communicable disease, more than either

cardiovascular disease or cancer.

Prince M et al. No health without mental health. Lancet 2007; 370(9590): 859-77

Contribution by different non-communicable diseases to DALYs worldwide in 2005

DALY – disability-adjusted life-year (sum of the years lived with disability and years of life lost)

Page 63: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

PET and MRI provide complementary information about

the brain.

Hemodynamic parameters:

cerebral blood flow (H215O, 15O-

butanol, 13NH3,…), cerebral blood volume (11CO)

Substrate metabolism:

glucose (18F-FDG), oxygen (15O2)

Protein synthesis: 11C-methionine, 11C-leucine, 11C-

tyrosine

Amino acid transport: 18F-fluoroethyltyrosine, 18F-

fluorophenylalanine,…

Nucleosides and DNA synthesis: 18F-fluorothymidine,…

Neurotransmitter biochemistry:

precursors (18F-FDOPA, 11C-AMT,…), transporters (11C-methylphenidate, 11C-cocaine,…), receptors (11C-raclopride, 11C-nicotine, 18F-altanserin,…), enzyme activity (11C-deprenyl, 11C-donepezil,…)

Anatomy:

high resolution morphology,

angiography

Perfusion:

cerebral blood flow and blood volume,

mean transit time, time to peak,

relative vessel size and permeability,…

Water diffusion:

mean diffusivity, fractional anisotropy,

apparent diffusion coefficient, fiber

orientation

Brain function:

BOLD contrast, PWI

Chemical composition:

1H-MRS (NAA, Cr, Cho, Lac, mI) , 31P-

MRS

PET RADIOTRACERS MR TECHNIQUES

Page 64: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:
Page 65: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

65

Surface projections

show areas with

reduced metabolism

Collaboration with

Brad Dickerson,

Alexander Drzezga

MR-PET allows the assessment of

anatomical/functional/molecular changes in

dementia.

FDG

PET

MPRAGE

Fused

MR-PET

DTI

C. Catana et al. “PET and MR Imaging: The odd couple or a match

made in heaven?” JNM 2013; 54:1-10

Page 66: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

66

MR-PET could help us understand the mechanism

of action of therapeutic agents in GBM patients.

[18F]FDG [18F]FMISO [18F]FLT [18F]FET

PET

MRI

Fused

MRI-PET

(C. Catana et al. “PET/MRI for Neurological Applications”. JNM 2012; 53:1916-1925 )

Page 67: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

67 Sobesky et al Stroke 2005; 36: 980-985

Cross-calibration measures could elucidate the

mismatch-penumbra debate in ischemic stroke patients.

Page 68: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

68

0.0 0.1 0.2 0.3 0.4-0.1

0.0

0.1

0.2

0.3

ΔBPND

ΔB

OLD

(%

Sig

nal C

hanges)

r = 0.96

p = 0.001

Thalamus (L)

-0.4 -0.2 0.0 0.2 0.4 0.6-0.1

0.0

0.1

0.2

0.3

ΔBPND

ΔB

OLD

(%

Sig

nal C

hanges)

r = -0.28

p = 0.5

Putamen/NAc (R) (b)

(a)

2.3

3

Z

fMR

I

PE

T

Thalamus

Putamen/NAc

x = -14 x = -4

y = 14 y = 4 0.0 0.1 0.2 0.3 0.4

-0.1

0.0

0.1

0.2

0.3

ΔBPND

ΔB

OLD

(%

Sig

nal C

hanges)

r = 0.96

p = 0.001

Thalamus (L)

-0.4 -0.2 0.0 0.2 0.4 0.6-0.1

0.0

0.1

0.2

0.3

ΔBPND

ΔB

OLD

(%

Sig

nal C

hanges)

r = -0.28

p = 0.5

Putamen/NAc (R) (b)

(a)

2.3

3

Z

fMR

I

PE

T

Thalamus

Putamen/NAc

x = -14 x = -4

y = 14 y = 4

Wey, Gollub, Kong et al.

fMRI-PET Assessment of the Response to a

Physiological Challenge (i.e. experimental

pain)

Brain regions show fMRI-PET activation overlap

Page 69: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

FDG-PET/MRI Study of Dexmedetomidine

Sedation

pASL-MRI

Seun Akeju, MGH

FDG-PET

Baseline

After sedation

Baseline

After sedation

Page 70: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

70

Page 71: Integrated PET/MRI - AMOS Onlineamos3.aapm.org/abstracts/pdf/77-22654-312436-101917.pdf · scanner (MGH installation) 16 ... trigger MR “Beating Phantom ... Integrated PET/MRI:

71

Acknowledgements

Massachusetts General Hospital

Bruce Rosen

Daniel Chonde, Spencer Bowen, Kevin Chen, David Izquierdo

Jacob Hooker, Grae Arabasz, Shirley Hsu, Steve Carlin, Chris

Moseley

Alexander Guimaraes, Andre van der Kouwe, Elizabeth

Gerstner, Tracy Batchelor, Peter Caravan, Larry Wald, Brad

Dickerson, Alexander Drzezga, Jian Kong, Randy Gollub,

Darin Dougherty, Monica Wey, Joe Mandeville, Christin

Sander, …

Siemens Healthcare

Greg Sorensen

Larry Byars, Christian Michel, Matthias Schmand, …

Keith Heberlein, Michael Hamm, Thomas Benner, Matthias

Fenchel, …

University of California Davis

Simon Cherry (PhD advisor)