Top Banner
Instruction Manual Model TM20 PH/ORP Analyzer Microprocessor based pH/GRP X YOKOGAWA . 1M i286Ei-oiE-H
56

Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Jun 23, 2018

Download

Documents

ngokiet
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

InstructionManual

Model TM20

PH/ORP AnalyzerMicroprocessor based

pH/GRP X

YOKOGAWA .1M i286Ei-oiE-H

Page 2: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

a:PAGE

1 .

SECCN TI

=00. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

2.2-1.2-2.

3.3-1.3-1-1 .3-1-2.3-2.3-2-1 .3-2-2.

4.4-1.4-2.4-2-1 .4-2-2.4-2-3.4-2-4.4-2-5.4-2-6.4-2-7.4-3.

5.5-1.5-1 -1 .5-1-2.5-2.5-2-1 .5-2-2.5-3.5-3-1.5-3-2.5-3-3.5-3-4.

6.6-1.6-1-1 .6-1-2.6-1-3.6-2.6-3.6-4.6-5.6-6.6-7.6-8.6-9.6-10.

7.

-'

SP~CJTIOOS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1(;er s~if ica tioo. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Fuonal speification............................................3rnS'OO Ar WIG.............................................. 4Irtallation. . . . . . . . . .. . .. . . . . . . . . . . . . . . . . . . . . . . . . . ~ . . . . . .. . . . . . . . . . .4Irtallation site....................................................4fwtig rnthoo...................................................... 4Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5Syste ronfigiatior.................................................5Wiring diagr....................................................... 6

CO NA AN FUCTONS........................................9Canert na......................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9fuctons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9r1Sting mooe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 0Wash nxe........................................................... 11Se t nxe............................................................ 1 2calibratior rne.................................................... 13~am rne. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 4Display messages (basic operation)................................. .14.Display messages (eror coes)...................................... 15Intrt identification........................................... 17OPERON. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 8st: up............................................................ 18Checg of wiring.................................................. 1 8Oprating the anlyzer.............................................. 18~iibra tiona . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Calibration of the mesuing loop for pH mesuents.............. .18Calibration of the mesuing loop for ORP measuremts............. .19Description of the first levl opeation moes..................... .20~suing moe...................................................... 20Wash rne........................................................... 225et moe........................................................... .23c:libra tion rne...................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

REGUTION OF TH ANyZE.............................. . ..... .28Entering the prog moe............................ . . . . . . . . ~ . . . . . .28Entering with passwrd protection.................................. .28Entering without passwrd protection............................... .29Chging the passwrd numr....................................... .29Selection of the progg menues................................. 29Alar rnu.......................................................... 30mA outp.t me.u..................................................... .34Response tim menu.................................................. 37llash tir rnu........................ . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38Sta tale rnu................................................. 4 a~ce check rnu.................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 iCOnfiguation menu of set moe...................................... 42List of display messages during progarg......................... 44

FACTRY SETINGS (DEFAULT VAS).. . . . . . . . . ... . . . . . . . . . . . . . . . . . . . . . .48

Page 3: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

1. INOOClON

The m:el 'I0 Anlyser is a I1croprossor-based, use-proarle intrt forJ:th pH an ORP mesuements, which allows p:erul catational fuctions to becained with o:ntrol functions, al, etc.The anlyser is factory intialised an ready for use. It prvides:

Exemly high reliability by in-line an autmtic ser chec.Autmtic electre cleag cotroller.Maual or autmtic tentue c:tion based on ptl000.Proprtiona on/off-ctrl fuction.Autmtic lifer retion.IIIDLDII-fuction duing mainten.Self -diagontic fuctions.Non vola tile mary (lasts for over 1 0 yeas).

2. SPæIICATIONS.2-1. (;erl Spifications.

A. r-suring rage pHORP

-1 to +15 pH-1500 to +1500 mV

B. Trssion signl 0-20 ID or 4-20 ID(ma. load 500 Ohm)

c. T.ssion range pH : . spa: mi. 2 pH, rn.16 pH. zero: -1 to 15 pH

: . spa: mi. 100 mV, ma. 3000 mV. zero: -1500 to +1500 mV

OR

D. Additional function oftranssion signl The ID output ca be held duing cleag or

maintence (fixed value or lat vaue)E. Contact outputs

Contact capacity

-status-Action-Alarg

250 V AC, 5 Ai ma. 100 W250 V OC, 5 Ai ma. 50 WPotential-free o:ntact outpt, fail sae (SPS)(o:ntact is on at p:er off an duingmaintence )

: Potential-free oontact outpt, fail safe (SP)(o:ntact is on at p:er off an duing failo:ndi tion)

Potential-free o:ntact output (SPST)(o:ntact is on duing wash cycle an 9ff atp:er off)2 inepeent, -ptential-free o:tact,outputs (SPIJ)High an/or low al for each outptDirect (fail sae) or reverse for each outputContinuou or pusating for each outptFor a p.sating outpt a selection betw afixed duty cycle (se fig. 2-1)

Maintence o:ntact

Fail o:ntact

Wash o:ntact

Al cotacts

r se~nt

continuous~"~"""~"'

Î

~o

pulsaling(fixed duty)

Dead timeDead zone

0,1-2,050,01- 0.2 pH '"

,- 20mVl-100s0-100 %dead zone

50 % PeriFixed duly

pHlmV_ Figue 2-1.Process alar

Page 4: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

F. Pcer supply.Vol tage

Consumtion

G. Climtic conditionsAmient temratuestorage temratueHudi tyPrtection

H. Display rnthcx

I. HosingMaterial

Cale entriesFor riel 'l0PG

For riel 'l0NG

TesColour

Dimions

Weight

J. Keys

or a proportionally a:ntrolle: duty

cycle (see fig. 2-2)

100% r setplnt~ on

oN

95Y';

ir on~ ~ ~oN

tlme_

i~

g

1-100 s0-100%0-1000/.0-100%

0.1- 2.0 s0.01- 0.2 pH or

1- 20 mYRange torpropotio functio 0.1- 10 pH or

10 -100 mY

PeriMin. dutMax. dutFlxed dutDead limeDead zon

pHlmV_Figue 2-2Prportonally contrlled duty cycle

24 VD (-40/+20%)24 VAC (-20/+20%), 50/60 Hz110 VAC (-20/+20%), 50/60 Hz220 VAC (-20/+20%), 50/60 Hzma. 8.5 VA.

-10 to +55 °c (-14 to 131°F)-30 to +70 °c (-20 to 158 OF)0-90% R.H.Rain an dutprof to IP 65 (NE 4)

8-dgi ts Liquid Crstal Display(12 nm chcter height)

Cast ali.um with chemælresistive lacqer (case)Polycaronate (winow)Polyade (glan)E.P.D.M. (holder for magnetic cxer)PVC (nagnetic a:er)

1 x PG16 glan for sesor cales(a grt is supplied)4 x PG13.5 glands for trssionan supply cales

: 1/2-inch NP (fanle thead)

: 4 nm '2 with exerl ea coection

a. Botton: RA 6003 (!TSS gree)b. Coer: RA 7032 (beige)

Anlyser: 300 x 160 x 135 mn(13 x 6.3 x 5.3 inch)Packging: 330 x 200 x 170 ra(13 x 7.9 x 6.7 inch)

appro. 4.5 kg (10 lb)

5 ree-relays, activate: with magneticccer (they fuction with interlswitches)5 internl p.sh b.ttons (in paallelwith the ree-relays).3 interl DIP-switches.1 maanetic coder 2

Page 5: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

2-2. F\ctonal SpifiætionA. Input

B. Asyrtr p:tentiadj ustale rage pH

ORP

e. Slopeadjustale rage (pH ony)

D. Tetue a:tion(pH only)

E. PefonnceDisplay pH

ORPID outputTe. read-outTe. ca.

Differtial amlifier with twhigh imce ints (~1 0 ,:1 Ohm)Eqp:tential sleldig

: +/-2 pH (at 25°C):' +/-120 mV

80-105% of theoretica value

: maual : -50 to +150 °c (-60 to +300 OF)autotic: -50 to +150 °c (-60 to +300 OF)

using a pt 1000 (R.T.D.)

c:pH/tJT for the temtue CXfficient ofpros liquds (+/-1 pH/10°C)(e.g. toiler feeter LipH/D-T = -0.35/100e)

.1.000.020.10.1

Resp:nse ti (90 %) O. 5s

1.000.020.20.2

")

Page 6: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

3. rnSTTION AN WI3-1. Irtalation

3-1-1. Intallation site.The 'I0 anlyser is of a raintight constrction an ca be intalled outdoors.

H:ever, it shod be intalled as nea as IXssible to the sers avoidig longcale lengt betwee sers an anyzer.Selec an intalation site where:_ mecal vibtion an shoc are neglect.le¡_ no relaylp:er switche are in the direc envirnmt;_ the anlyzer is not rrunted in diec suight an severe wether coitions._ maintence actvities are IXssible (no corrsive abnspheres)

The amient tantue an humdity mut be with the limts of thspeifications (see prgraph 2 -1 ) .

3-1-2. M:ting methoo.The 'I0 anlyzer ca be intalled either on a 50 nm veical pipe or wall

suface. M:unting hae is supplied with the intrt if speified in the salesorder .

3-1-2-1. Pipe M:ting.The Anlyser ca be m:unted on a

50 rr pipe. Se fig. 3-1.

Ii/~

Fig. 3-1. Pipe lbuntig.

3-1-2-2. Wall M:tig.Fig. 3-2 shaHS haH wall m:untig is acleved, using the wall m:tig lie.

t.R////////'rn///L

10

ø

, '., -,,

,,,,

,

(~....I i, ', ', ,, ', ,

\..,'

,,

... D ,i

II

I

)00

Figue 3-2. Wall lbunting.

4

Page 7: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

3-1-2-3. Optional hoo for sunight.If speified (/H) a hoo for sunight

is delivered with the anlyzer. Fig. 3-3shows the mountig.

/~

Figue 3-3.Ho for sunight.

3-2. Wiring.(;erlly, trssion of signls representing the mesued pH or ORP value of a

liquid is by anlog, low voltages with ver high imce levels. Therefore,disturbing influences mut be rninsed.Before coiiectig sensor cales to the anyser inpit the following cotions mutbe considered:- the senr cales are (preferably) not roted in trck together with high pocales an cales with signls of switched high voltage or high. energy;

- the stada sensor cables are used;- the anyser is mounted as nea as posible to the senrs (short senr cales).

3-2-1. System configuation

TM20

INPUT OUTUT

POWER SUPPLY§ 24,110,220 V AC

0) 24 V OC

CURRENT OU~20 mA04-20 mA

~---dJ

MANTNACE · ~ ...

td c(

~ c(b ~5E -t

a1

2.5 or 10 mPROESS VALUECOACT OUS PO

PROCESS VALUE (1) __'

I PROCESS VALUE (2) : __:.

AUXIUARY CONACT OUS SPD WASH ~ ...

FAILURE ALARM. ...Figue 3-4.System configution.

.CONTACT CLOSED IN POWER OFFCONDITON

5

Page 8: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

3-2-2. Wiring diagrThe loction of the terls an the exerl wiring is sha, in fig. 3-5.

Fig. 3-6 shows the interl connecons an statu of the contact outputs at p:eroff condition.The opeating statu of the contact outputs is given in fig. 3-7.NJ: When using all contact outputs (alar, maintece, fail an wash) then

nnticore cale mut be applied.

1

-,i

r

r

iii

iIIiIL -i

I

IIIL__ - --1

I

,L_________..IIIII

SUPPL Y

M"jnltn"nct SENSORALARMS F"il mA-OUT pH/ORP

Wash

1112 17 ~

14 13

Figue 3-5Lotion of the tenls an exerl wiring

alarm 1

maint.

wash

C3 .,¡-081 =:082 ,083 --~

~1~ =l 1~1.041 --042 ..

alarm 2

,if- .

mA-out

o 15 ~ glass Imela~

o 16 ~ Sen018 . 1- ~~~~

012 0-~'11 J iemp,"c -+ sensor,014 i. f- liQuid earth for

:0-7: 0- sens monnoring

fO 3ì .. relerencesensOf

Figue 3-6.Internl connections (contact outputsat poer off condition).

I-

PROCESS ALARM CONTACTS-DIRECT REVE RSE COI'IONACTION ACTION

~~~~.= ~11~~~C ~ POER

731831NC-- 731nlNO --OFF

711811 NO-- 711811 NC~ POR ON

~~li~c:= nl&2iC NO ALARM

7JI83INO-- norm opl""

~::~'i~~ 71181 HC-- . POR

~~~'I~O ::ON

7JI8JlNC-- AI AR"I

AUXILIARY ALARM CONTACTSSTATUS CONOlTlON

j~~O~ 51 NO-- POR52C _ug ~O := OFF

31 NO--~i~O =:

POWERJ2C ---- ON" NO-- NOR"'AL

ac ---- OPERATION

~~O~ FAIL

COl'1Ct

L1NO-, ""t.NAN:'2C~ MC

51 NO-- væ,.51C~ ..ODE

Figue 3-7.status of contact outputs

6

Page 9: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

3-2-2-1. Connecting the senr cales.1. Lose the tw screws on the front of the

anlyzer to rere the frot coer.2. Thead the cales thgh the gla and

græmet as sh~ in fig. 3-8. Be sure toplug any unuse: holes in the gramt. PULthe cales into the inter assenly ancorrecy coec each condctr to thecorresp:nding . te. Se also thewiring diagr (pagrph 3-2-2).NJ: When the ser imce che is

reqe: then connec liqud eato terl 1 4.If not, coect liqud ea to ter-mil 1 8 (terl 14 is pa of themesuin circuit).

3. Scew the gla nut fi:y on th glanl:y to keep out rristu. Hove, if thegla nut is screwe on too tight, thecales may be dage:.

3-2-2-2. Connecing the cales to the juncton 00.'Ie junction b: an connecting cale are

not reqire: if the senr cales arE"connece: to the anlyzer dily. The ma.lengt of the senr cales is 10m.Interconnec the terls of the junction b:an the anlyzer using the coectig cale.For the wiring proure, see pagrph 3-2-2-1.ID1. When using a connecting cale be sure to

strip off the inulation. It is imrtattht the coctive layer is reme: franthe p:lyethylene for a distace of atleast 5 nt.

2. The autantic imce check of theeleces is distube by long calelengt. Therefore the lengt of theconnecting cale may not exce 10m.

Figue 3-8.Senr cales mounting

~

I "...~._".\ lai

Figue 3-9.Junction l:x an connecg cale.

3-2-2-3. Connecing the p: supply/grun cale.'Is cale cotain thee condctrs (including grou) an its outer diamter

should be 9 to 12 nt. It is used to supply the po to the anlyzer an to coecthe interl grun circut to grun.Use a gr cale with resistace 100 Ohm or less. Coect the cale to the anlyzeras sho in the wiring diagr (se pagrph 3-2-2). For the wiring procure, referto pagrph 3-2-2-1.

3-2-2-4. Connecing the outpt signl cale.'Is cale is use: to trt outpt signls fran the anlyzer to a recrder or

reciver. Use a 2-cndctor shielded cale with an outer diamter of 9 to 12 mm.Connect the condctors of the cale to terls 61 (+) an 62(-) an shield to thegroun tennal ( 3) .For the wiring procure, refer to paagraph 3-2-2-1.

3-2-2-5. Connecting the cale for the process alar contacts.This cale is use: for the alar. Use a 4-cnductor cale with an outer diamter

of 9 to 12 rn. Connect the cale to the anlyzer as indcated in the wiring diagr.For the wiring procedure, refer to paagrph 3-2-2-1.

Page 10: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

3-2-2-6. Cbnnecting the cale for the washig device*.Ths cale is usoo to trsmt the wash sign for activating/de-activating a

Cieag device. Use a 2-cndctor shielded cale with an outer diameter of 9 to12 ir. Cbnnec the cale to tens 51 an 52 an the shield to the grou terl(3) .For the wiring procure, refer to pagrph 3-2-2-1.

3-2-2-7. Cbnnecting the cale for failure alar contact output*.

'!s cale is usoo for a failur alar. Cbntact is closoo at pcer off an at afail a:tion. Use a 2-coouctor cale with an outer diater of 9 to 12 nu. Cbnnectthe cale to the terls 31 an 32 of the anlyzer.For the wiring proooure. refer to paagrph 3-2-2-1.

3-2-2-8. Cbnnecting the cale for the maintence contact output*.Ths cale is usoo for activatig an indcator, shCMing tht the instrent is in

maintence noe or pcer lost. The cotact is ope in nonnl coition, closed,during maintence or pc off.For the wiring procure, refer to pagrpg 3-2-2-1.N:: Nonnlly the fail an maintence rontact outputs caot be closoo at the sa

tim. 'Ierefore, when l:th are closed there is a poer dow condition.

* Using all rontact-outpts it is necessa to apply multirore cale with a diamterof 9 to 12 nt.

8

Page 11: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

4. roNE NA AN FUClONS.4-1. Gonent naes. LC~i

L (êJ ~fI=:1. Robust enclosure.

Cast alumium lxy with unrealepolycaoonate window.IP 65 (ND 4) protection, also duringcalibration an maintence.

2. Informtion display.An easy to read Liquid Crstal Displayindicates all relevat inormtion (8alphaumeric chacters, 12 nu height)

c:

3. rata plate.For basic inormtions.

o

4. Tet plate.Gives inormtion a.ut connections. Figue 4-1.

Ccnents.

o

5. Hessage tale.The cced messages shaw on the display are exlained in the messages tale.

6. Suspeted paameter tale.The error cces sho on the display are exlained in the suspeted paamtertale.

7. Dip switches.Thee intemal dip switches for selection of the following paamters:- pH or OR function- password protection (yes or no)- autantic retu to mesuring mce (yes or no).

8. Oprating or function keys.These are ree-relays which ca be activated by a magnetic caer. With thelive opeating keys all settings of the 'l0 ca be perform.The setting is meu driven in an interactive way.

9. Swi tches .

These switches are paallel to the function keys. Switching ca be done maually,when coer is remed.

1 o. l-lagnetic canmer.This magnetic caer enles an opeator to activate the function keys. Itguantees IP 65 protection also during calibtion an progarg.

11. Input terls.

12. o.tput termls.13 . SUly termls.

4-2. fuctions.Using the magnetic cader 5 ree-relays ca l: activated. These ree-relays

duplicate 5 internl switches and have a dual function:1. First level opera tion for selection of:

a. l-1esuring m:eb. Hash mcec. Set m:ed. stada mcee. Progaring rre

2. Second level progaming for recnfiguation of the anlyzer.9

Page 12: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

4-2-1. Mesuring moe.DJing measuremt or after activatig the key 1I0ISPIAIl next inormtion ca be

displayed.l': An exle is given only. In pagraph 4-2-6 all display messages are listed.

Eror code

mesured value

process temratue

mA output signl

imce glasselecode

I. limt exceed

imce ref.elece .

I PH 88.88 I- start pasi tion10

Page 13: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

r

4-2-2. Wash moe.The anlyzer has a built-in wash cycle tim which ca be used if electre

Cieag is peform using jet spray. The tim stat autatically or maually.depeent on a pre-adjusted condition (see chapter "Prarg").When is progan for autantically, then washig sta contiuously after apre-adjusted tim is elapsed.

. When is prog for maually, then washig stats after actvating the key "WAH".

The output signl (contact .outpt, te.ls 51 - 52) ca be used to contrl asolenoid vave or spray pum.OJing the wash peicx an recer tim, the anlyzer is in the maintence mce (IDoutput signl of the anlyzer is "HOLD" at a fixed or last measued value, or"LIV" ) .

PH 88.88 mesured value

autantic

,,\\\ II.. WAH/11111 I I

i

/~wai ts for

"-

\ "confintion

-wash cycle sta

washig tim an measued value

washig is stopp

888.888 recer tim and mesued value

PH 88.88 sta ¡;sition

11

Page 14: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

4-2-3. SET mooeActivatig the "SET" key the opeator ca chec the set values of relevat

paters.NJ :1. A bling display inicates tht a value ca be chged using the "up" an "IX"

keys (see paagrph 5-3, "Description of the prim moes").2. Sa item of the listed inicctions in

the diagram belaw may be antted. This

depe on the selected paamters in the "Prg" moe (see pagrph 6,"Recnfiguation of the anlyzer".)

I PH 88. 88 ~ mesued valuei

activate II SE"

-/11 i ISE. . i r /'- waits for confirtion\ 1\'-

activate IIEN".. , i r I~/- ./-: AlL 88.88 setpint alar 1/ i I 'I I' 1\' ..

activate IIEN"

__ AlL 88.88-= A2H 88.88 - setpint ala 2activate "EN".. A2H 88.88 _-: 'I 888 C:: maual temature cation--Ii' If I I I \ I 1\"..

activate IIEN"

__ 'I 888 C.,_2 PH 88.88 - procss calibation/,., i'i 1\11' ,..

activate IIENII

--~ mA simator

actvate 'IEN"offset voltage

activate IIEN"

AP 888SL 8888--

--I ¡ II \ \ ,..-.

slope

activate 'IEN"

isotherl point of intersecon

activate "EN"ITP 888A1L 888 --\ , \ , ,..

---- sta of the cycle

J I f i I

L PH 88 .88 ~ stat pJsition

Page 15: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

4-2-4. Calibration mceAfter activating the key "STANAR" the anlyzer goes into the calibtion ine

(see paagraph 5-3-4).NJ :1. Duing calibation the nitence contact is closed.2. When the anlyzer is set for ORP mesuemt, then the function of the key

"STANAR" is switched off.

'I OFF

activate "EN"

ST. 1

mesued value

waits for confintion

auto. tem. ca.(ma. tem. ca. switched off,see pagrph 5-3-4).

ready for 1 st calibtion point

reme senrs fran proess, cleathem an irse in bufer 1

8.88 8.88 stailizing tim an mesued valueST.l 88.88 -- autantic accptace*

activate "EN

actvate "EN"ST.l EN - calibtion 1 is finshed**

ST.2

activate "EN"rnSER

activate "EN"

ready for 2nd calibration póint*

clea senrs an i.se thanin bufer 2

8.88 8.88 - stailizing tim an measued valueST.2 88.88 - autantic acceptace*

activate "EN"- calculated slope*-- calcuted offset voltage*

put senrs back in procss

PH 88.88 sta J;si tion

* An eror code may occu (see paragrph 5-3-4).** When 1 point calibration is reqired, activate "RE". '!e display incates the

calculated offset voltage (see pagraph 5-3-4).13

Page 16: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

4-2-5. Progg m:eThe key "PR" is used for reciguation of the anlyzer (see paagraph 6).

After actvatig the key "~" the display incates "I'" (bling) an theopetor ha the access to the prog m:e.rm: If ordered, the anyzer is proided with a passwrd protecion. Without the

correct paswrd it is not possible to enter the prog ine.

4-2-6. Display mesages (æsic opetion).In the diagram belC7 all massages tht may incated duing æsic opetion ar

listed. In addtion, an exlation of the inication is given.

r8.88 8.88

I A1H 88.88

I AlL 88.88

A2H 88.88

A2 88.88

~ 888

E 88.88.88"

I EROR

I ERORS

I HIGH II HIGH.HIGH I

I HIGH.LO r

I IM. CH iI INSET II I' 88.88 II LOW Ii ro ro Ii in 88.88 II mV 8888 I

I PH 88.88 I

I RE ii SE ii SL 888.8 I

ti an mæsued vaue are displayed (ca be displayed in washan calibrtion m:e)

setpint (high) of ala 1setpint (lC7) of al 1setpoint (high) of ala 2

setpoint (lC7) of alan 2

asymtr potential (offset) in mV

eror cce (see suspeed pater tale, pagraph 4-2-7)

an eror is fou (activate "DISPIAY" for iniæ.tion of the cxe)

erors are fou (activate "DISPY" for incation of the coes)

setpoint of high al is exceed

setpoints of both high alar ar exceed

setpint of both high an lC7 ala is exceeed

imce chec is peformediner sers in the l:fer solutionisother point of intesection

setpoint of iC7 al is exedsetpint of both iC7 alar are exceed

in outpt signmesured OR vaue (for OR anlyzer)

mesured pH value (for pH anlyzer)

retur elecodes in proess

key "SEt' is activated (anlyzer waits for confi.tion)

slope (pecetage of theoretical value)

to b? continued on page 15

Page 17: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Display messages (ront. of page 14).

(sr. 1

(sr.i 88.88

Csr.i EN

(sr. 2

(sr.2 88.88

( STAR

r TA888.8 °c

I TA88.8 of

I 'I88. 8 °c

I 0088.8 of

i 'I OFF

'I ON

I 'I °C888.8 -I

I'I °F888. 8

I * WA *

JWA

IWAHI

I ZGE 888

I ZGE HIGH

I ZGE IJ

I ZR 88.88

IZR HIGH

calibration of 1 st p:int ca be staed

selected b.ffer solution of calibtion FCint 1

calibrtion of 1st p:int is finshed

caibration of 2n p:int ca be staed

seeced bufer solution of caibtion p:int 2

key "Sl" is activated (anyzer wats for ronfintion)pross tantu (in °C)pross temtue (in OF)maually set tentue (in °C)

maually set tematue (in OF)

maual temtue cxtion is switched off (autatic ten.cx tion )maual temtue cxtion is switch on (no autatic ten.cantion)temtue to be set (in °C)

tentue to be set (in OF)

after wash peiod (recver tie is staed)

key "WAH" is actvated (anyzer wats for ronfinntion)

wash cycle is staed

imance glass electre (in MJhm)

imance glas elecode (high lit is exed)

imce glass elecode (law lii t is exed)imce referen elecode (in kOhm)imce refer elecode (high limt is exer).

4-2-7. Display mesages (eror coes).Duing mesuement, calibration of the mesuing loop, washg, etc. an eror cee

ca l: inicated on the display. en page 16 a cxlete list of all cees is give.Ths list includes an exlantion of the err, in whch m:e the err cee may

app an the action to solve the problem.If an eror cat be solve, please oontact your locl seice station or sales

agent.

15

Page 18: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

EROR amES

aJDE DISæImON MJDE AClON REFOIN Acr KE EXUITION

E01 ResfOnse or half value WAH A,B,C,D,E WAH p:. 5-3-2 note 1t:l chec failed SlANAR F,G,L RE pa. 5-3-4

E02 Asyme:~ potenua.1 SE I p:. ~_jnj note Lout of limts SlAN A,B p:. 5-3-4E03 , Bufer 1 to close to sr G p:. ~-j-4 note L

roffer 2E04 Slope out of limts SlAN A,B RE p:. 5-3-4 note 2

E05 Teratue out of ME D,E RE note 3lim ts

E06 Teratue of b.ffer SlAN D,E RE p:. 5-3-4 note 2solution out of limits

E07 Prcxss value out of ME B,C,E,F ,G note 3lim ts

E08 Lmce of glass ME A,B,C,E note jelectre out of limtsE09 Lmce of reference ME A,B,C,E note 3electrode out of limtsE10 EERO write failure MA . K note jE12 Reference voltage ME K note 2,3

MA . KE14 A/D convertor failure ME K note 2,3

MA . KE17 Spa rn output less PR I note 2

th 2 pH or 200 mV (ml out)E18 Gain eror of MEUR K note L,j

temperature IC MA . KE19 Overange AID ME K note L,j

convertor MA . KE21 RA selfdiagnose SlARTIG K note 2

failure (low block) UPE22 RA selfdiagnose i SlARTI K , note 2

failure (high block) UP i

N: :1. Progam will continue. After retug to mesuring m:e a I'fail" al is

given.2. Pram is stopp (vaue not accpted).3. "Fail" alan is given.

AClION FuINSA Clea senors thoroughly.B Oleck senrs. Replace them, if necessa (re-sta æ-libtion).C Oleck if sensor tips are imersed in the pross liquid.o Olec tempeture of the solution (see senr spifiæ-tions).E Oleck wiring.F Oleck isolation resistace of senrs an cales.G Oleck l:fer solutions (old solutions or a dirty beer is used). The use b.ffer

solution is not progamed.H Oleck pH value of prcxss liquid (e.g. by using a lab. meter).I Re-adjust value.R Disapp automtiæ-lly at next diagnosis (or staing up). If not, please

contact your locl serice station of Yogogawa Electrofact or Sales Agent.

16

Page 19: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

4-3. Intnt identification.As delivered, the 'I0 anlyzer is thoroughly tested at the factory an caletely

adjusted aCaJrdg users i reqirements. Inonntion is inicated on the data plate.If not, the anlyzer setting is based on a default statu (basic factory settig).

If you have any problem, please aJntact yo neaest serice station of YokogwaElectrofact or Sales Agent. When enquing arut the anlyzer, provide the moel naan serial num.

The anlyzer has a wide range of aJnfigution which ca be peorm by the useaccrdig to his s~ific 'reqemt.--------------------------------------------------------------------------------------WAInaJrrect use an setting of the anlyzer ca rest in a coed functionig. Ifso, chec the prog values an cae then with the default values or thechgings you have intentionally made.A calete list of default values is given in pagrph 6, "Reciguation of theanlyzer" .

17

Page 20: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

5. OPERTION5-1. sta Up5-1-1. O1eckig of wiring.

Cbnfinn tht all wires are correcy connected over the entire systan (see

paagrph 3-2).

5-1-2. Optig the anlyzer.Supply the correct p: to the anyzer an plug in.

5-1-3. The magnetic cc.It is not necsa to ope the case of the 'I0 anyzer for caibtion,

recnfiguation, etc. All halings are pefonned by a magnetic cader. '!sccer is place in a click-in hole on the front. For activating a key(ree-relay), the ccer has to be place in the middle of the gree field.

5-2. Calibation.The measuring loop for both pH an ORP mesuranents mut be caibted. A

description:is given in the pagrphs belCM.

5-2-1. Calibration of the mesuring loop for pH meaants.Basically there are thee method:

1. bufer solution calibration2. pross calibration3. input of electre chacteristics.

5-2-1-1. Bufer solution calibtion.Buffer solution calibration (or IIlifer

check") is pefonn using tw bufer solution.The 1 st solution is for calibtion of theoffset voltage (asyitr lXtential of the pairof electrodes). Geerlly, a value nea to theisothenl lXint of intesection is chosen(nornlly 7 pH).

The 2nd buffer solution is for calibrationof the seni ti vi ty of the electrodes (slope)and usually, a buffer solution of 4 pH is coosen.This solution has a high staility.

When bufer solutions accrding to the NBstad are chosen, then the 'I0 anlyzerrecses these solutions an caibration isperfonn. The NB stas are listed intale 5-1 (ST to ST5).ID: The procure to ca out bufer ,

solution calibration is describ in: .pagraph 5-3-4.

When no solution accordg to the NBstada is used, then the value of thesolution ca be entered in the marry (see

paagrph 6, IIRecnfigliration of theanlyzer II).

I

gt:'

~~

I

~

~

"00

.300

.200

.100~"E

t0

2 L_pH-100

- 200

12 "

-300

-LOO

Figue 5-1.Bufer solution calibtion

ST ST2 ST3 ST4 ST5 TE .1.67 4.00 6.98 9.46 13.42 o "c1.67 4.00 6.95 9 .JQ. 13.21 5 "C1..67 4.00 6~92 ~.33 13~9.Q -10-ocJ .67 4.00 6.90 9.28 12.81 -1S"oC1.68 4.00 6.88 9.23 i 2J~j 20-oC '1.68 4.01 6.87 9.18 12.45 25 "c1.68 4.02 6.85 9.14 12.29 30 DCi.69 4.02 tj.~ s4 9~io 12~lj -3S- °c~- ,

-4ö'c1.69 4.04 6.84 9.07 11 ~~81.70 4.05 6.83 9.04 11 . a4 -'4Š-õC_.- -_.-

J..71 4.06 6.83 ~.Ql 11 .71 50 DC1.72 4.08 6.83 9.99 11.57 sSOë'

-1.12 1.09 6.84 9.96 n.45 -60-õC1.73 4.11 6.84 9.94 11 .33 65 DC1.74 4.13 6.85 8.92 11 .22 70 DC1.75 4.15 6.86 8.90 11.11 75 DC1.77 4.16 6.86 8.89 11 .01 80 DC1.78 4.18 6.87 8.88 10.91 85 DC1.79 4.21 6.88 8.85 10.82 90 DC1 .81 4.23 6.89 8.83 10.73 -95" DC

1.83 4.25 6.90 8.81 10:65 100-oc'

Tale 5-1.NB stas 18

Page 21: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

5-2-1-2. Prss calibtion.Procss caibtion is made by using a sale of the process strea. Afterdeteng the pH value of ths samle accuately (e.g. by mea of the poet pHM:ter, Mcel PH81 of Yokogwa Electrofact),the 'I0 anlyzer ca be set in accxmìancewi th ths vaue., '!e prcure to ca out prosscalibtion is desib in the llSE" moe,paagrph 5-3-4.NJ :1. Both th poet pH-mter an the anlyzer

mut be cated for tenratue.2. The tim betw tag the sale, measuing

the pH value with the poet pH Meter anadjustig the TM0 anyzer must be smllin order to prevent abrption of air (withCD,-) or desorption of gases. (e.g. SOt).

3. The electre-system of both the poketpH Mete an the 'I0 anlyzer mut beeq. Espeially the reference electrodesare ver imrtt.To prevent different diffusion potentials ofNaT, the used electrolyte concetrtionmust be eql too.

5-2-1-3. Input of electrode chaacteristics.The calibration IIInput of elece

chcteristics II is not a generl method.

Howeer, santim the user prefers to anlysethe chacteristics of a pair of elecesin the la.ratory intead of in the field.The adjusbnts of offset voltage (asyitrpotential), slope an isotheil point ofthe TM0 anlyzer are set acrording to thelabratory results (see "SE" moe, paagrph5-3-3) .

LOa

200

100

a12 14

-100

-200

-300

all.r ~~s cali- -+

-400

Figue 5-2.Procss calibation

.LOO

'300

a. J.T.P - tared ion

b. iietr y po~hal

'700

:; .100

E

ta

-100

2 4_pH 14

- 200

-30017

-400

Figue 5-3.Input of electrode chacteistics

5-2-2. Calibation of the mesuring loop for ORP measuemts.Beically there are tw methods:

1. bufer solution caibtion2. pros calibtion.

5-2-2-1. Bufer solution calibtion.The caibration for ORP measuemts nee 1 point calibation only. '!s

rontrasts with the 2 points pH calibration.Calibation ca be peform by using the Redox bufer solution quydrone orhydroqone (see "SE" moe, paagrph 5-3-3).The measured potential difference with reference to a satuted caanl eleccxewill be (at 20°C):+50 :t5 mV for quinhydrone.-50 :tS mV for hydroqinone.The solutions must be prepaed freshly at pH 7.

~

19

Page 22: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

5-2-2-2. Prss calibration.When mesuremts are made in a pross liquid, then inaccuacies of the senr

system (e.g. as a resut of aging) ca be zeroe by asytr p:tential adjustmt.'!is reqres an accuate knCMledge of the ORP value (in mV an at a fixedtemtue of the procs value.Wl: For accuate detention of the ORP value of a procss liquid, the poket

pH/OR Meter, m:e1 PH83 of Yokowa Electrofact is reoed.

'!e asytr potential adjustemt ca also be used for chging the display reaingof the anlyzer (e.g. if a more convenent reag is reqed at a predeteredstag point an tenencies are imrtt only or if anther reference electrode isused th ordered in th intrctions).'!e proede for setting is given in the "SE" m:e, see paagraph 5-3-3.

5-3. Decription of the first level opeation moes.

5-3-1. Mesuing mce.

5-3-1-1. Flow cha.IÃing measuemt or repetedy

activating the key "DISPLAY" will resultin ~ display reading of imrtt paamters.An exle is given only.A calete list of display messages isgiven in paagrph 4-2-6.In paagraph 5-3-1 -2 an exlantion ofthe display messages is given.

";.,;' - ';1~1"';'" I';; ,

LOW

1 if "I.vont

J

HIGH

I IMP CHECK\ i I

I ZGE Hi:H

I ZGE LOW

ZRE HIGH

?n

Page 23: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

5-3-1-2. ~lantion of the display messages in the measuing mQe~In the diagr below all relevat messages are listed. Depeem: on the entered

progan, san messages may be ignred (e. g. when no imce chec is r~ed,then the inications "ZGE 888" an "ZR 888" will not be displayer).

STATUS DISPLAY

PH88.88mV888.8

IM. CHHIGH an/orI.ERER

EB8.88.88

'!888 °c'!888 OFTA888 °cTA888 OF

0UlJ CU ID 88.88

IMEDANCE GLS ZGE 888ELDE ZGE HIGHZGE w,

IMEDANCE RE. ZR 888ELDE ZR HIGH

SESITI SL 888.8

OFFET VOLTAGE M? 888

RrpH or ORP (mV) value is displayed (depeentoonfigution of the anlyzer)

imce check of senrs is peonnhigh an/or ICM ala of setpint isexeeererr message

After indicating "ER" an activatingthe key "DISPLAY" ths message inicatesthe eror coe (see tale on front)

The first 2 chacters refer to the selectertemture casation.'! = maual tematue cation (the

incated value is maually set value)TA = autantic temtue cantion

(the incater value is thetematue of the process).

°C¡OF = tematue in degees Cetigrde or. degees Faheneit.

¡

The value of the outpt cuent is indicaterin ID.

Iice of pH/OR electrode is inicaterin MJhr or set high/low limt is exeeed.

Imce of the reference electrode isincated or set high limt is exceeed.Slope of the electrooe system is inicatedin pecetage of theoretical slope (80 - 1 05 %)

Asyrtr potential of the electrode systemis indcated in mV (:! 120 mV).

N: :When a key is not activated, then the last selecter pater will be

displayed during 1 minute. After this tim the anlyzer retu to the normlstatus.

OO

DIAGDIS

TEER'I

21

Page 24: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

5-3-2. Wash moe

5-3-2-1. Flow chaThe anlyse includes a tim for autatic

oontrlling of a jet spray cleag device.'!e tir has a contact outpt (terls 51an 52) which, in tu ca be used to contrla solenid vave or dosing pu.

To have an optii cleag, the user ca set:1. interl betw cleag cycles2. lengt of th cleaing cycle3. lengt of the reltion tii.'!e settigs are entered in the "Prog" moe(see paagrph 6, "Renfigution of theanyzer) .

Ne to the autatic cleag, the user castat the wash cycle maually by activatingthe key llWAH". The lengt of the cleagcycle an relaxtion tim will be asprogai

5-3-2-2. staing the wash cycle maually.

~~mE~~RETRN NEX DOWN UP EHê

DISPLAY KE DISPLA RE" //PH 88.88 activate "WAH" :: WASH :: no acton, waits for confinntion

j-// ,,-

I

I

,\ ¡'I / I

: WA - activate "EN" WAG output contact is closed anI

i i I \ "- anlyzer in the maintence moe i

WA 8.88 8.88 ti of wash cycle an measuredvalue are displayed

ii

8.88 8.88 WA washig cycle is finshed an contactoutput is de-eergised i

¡

¡

,

WA 8.88 8.88 relaxtion tim an measued valueI

are displayed

8.88 8.88 PH 88.88 relaxtion tim is passed. Anlyzerretu to the measuing moe(maintence contact is ope)

22

Page 25: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

5-3-3. Set mce.5-3-3-1. Flow ch

In the set mce the opeator ca chgesettings which are disclosed by the intrtenginee. The selection is made in the "Conf.set" meu of the "Progam" mce (see paagraph6, "Renfiguation of the anlyzer").

Possible settigs are:1. Alan setpints if pross alar are

used.

2. Tapea tue if maual tematuecction is use:

3. pH (mV) value for procss calibation(see pagrph 5-2)

4. ID outpt if a "mil" output isset, the fixed ID outputca be chaged (ID sin-lator for recivingeqpnt)

5. Asyitr potential entering of electrcxechacteristics

6. Slope entering of electrcxechacteristics

7. Isothenl pointof intersection

entering of electrcxechacteristics

Geeral inonntionWhen the inication of the display is bliag,then the inicated value ca be chage: usingthe "up" an "OO" keys. After activating thekey "il" a fixe: display inication willshow the last entered value. Autatically thenex paamter will be incated (bling) onthe display an the procede ca be repete.

'!e seqence as sho in the flow cha andescrib in paagrph 5-3-3-2 will not benecssaily the sam as the seqence of thedisplay incations. Irrelevat settings areignore: or are not entered in the "Progam"m:e.Continuously activating the key ".E" willresult in an indication of your reqedi:ter an setting ca be mae.

/

23

Page 26: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

5-3-3-2. Description of the settigs.

DISPLAY KE DISPL RD 1

\ l i /activate "SET"

. ,PH 88.88 : SET - no action, waits for confintion

/ I' , -.

,J I Iii,. lf i'"

:. SE:: activate "mI" A1L:88.88 . setpoint of alar 1 ("t" = lC1 ala)~ -/1 I \ " 1\ anlyzer goes to maintence moe

. \ III1"up" or "OOWN" is chaged andA1L88.88 AlL xx.xx setpint of alar 1

ii' ,,\ activate "EN" setpint of alar 2 will be incated

.\ II/Ii

"up" or "OO" setpint of alar 2 is chaged ("H"-A2W 88.88 A2H xx.xx ='1/ , i ' activate "EN" high alar) an maual tanatue

will be indicated

, 1111

:TM888.8 °c "up or "IX" 'I :i.x °c maual temtue is chaged in °c'" " activate "EN" (or OF) an pH (or mV) value will beincated

," ii/i ..

:PW88.88 "up" or "IX" PH xx.xx pH (or mV) for process calibation is~, l I activate "EN" chged (to value of pross samle,

see pagraph 5-2-1-2) an mA outptwill be indicated. If E02 is indicatedsee note 1.

)Ì: 88.8 "up" or "OOWN" mA xx.x Value of fixed mA output signl isi,.''' activate "EN" chged (mA simlator for coecting

eqpnt). mA output retur to "WID"vaue an offset voltage (asytrpotential) will be inicated

:AP: 888 "up" or "IX" AP :i offset voltage for e-iectrcxe chaacte-/1 i i' activate "EN ristics is chaged (see paagraph

5-2-1-3 an note 2) an slope willbe indicated

~ \ \ II,

:SL: 888.8 "up" or "IX" SL :i. x slope for electrode chaacteristics"'\ activate "EN" is chaged (see pagrph 5-2-1-3 annote 2) an isothenl point ofintersection (IT) will be inicated

"t "":IT; 888 "up" or "OOWN" IT :i IT for electrode chaacteristics isIII \ activate "EN" chged (see pagrph 5-2-1-3 an

note 2) an setting is finshed.Display will inicate "AlL 888"(sta position)

-A1L-.888 activate "RE" PH 88.88 anlyzer is retued to measuring""",\\\ moe

N: :1. When the display indicates E02 88.88, the set pH (or mY) is out of acceptace

li.ts. Using the "up" or "OOWN" key the value ca be chaged.2. Never use these settings for buffer solution calibration. These settings are for

for entering the inputs of electrode chaacteristics only (pagraph 5-2-1-3).24

Page 27: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

5-3-4. Calibration (or sta) moe.In ths moe buffer solution calibration is pefonn using the key "STANAR"

(see paagraph 5-2-1-1 ). Next to the 'I0 anlyzer ths calibtion reqires a pairof eleces an bufer solutions of which the pH value is Ja (accucy 0,01 pH)as a function of tentue.5-3-4-1. Flow cha.

Before stag the prce, san remksnnt be made:1. The 'l0 anlyzer measues the voltage outpt

of the pair of el-et.e5 an caes it withthe theoretical value of the pro: bufersolutions when ideal elect.es are used.

2. The 'l0 anlyzer indicates the pH value atreference temabie (25°C) an the opeatornnt cae it with the value on the box orbottle of the bufer solution.If the inicated value differs fra the actlvalue, then the operator ca call anotherbufer solution by activatig the key "NE"until the correc bufer solution value isinicated.ru: An incorrect inication of the bufer

solution is often caused by:- old buffer solution- high asyretr voltage of the pair of

elect.es- dirty electrodes- residue of old buffer solutions mied'with the actul ones.

Therefore it is likely tht the anlyzerwill not accept the pefonn calibrationan an err message will be indicated.Activate IlREII an repet calibationafter cleaing the electres thoroughlyan prepaing new bufer solutions.

3. The pH value of 5 buffer solutions (to NBstad) are stored in the anlyzer as afunction of temrature. If a temratuecator is used during calibtion, theanlyzer recses the value easily. However,if maual temrabie cction is appliedthe actl tematue nnt be entered in the'I0 maually.

4. D.ing caibration the anlyzer is in themaintence moe. Maintence contact is closedan mA outpt is "HOW".

Prepation.Nessa eq pnt:1. 3 vessels of 200 ml, e.g. plastic calibation

dishes as supplied with the flow fittings.2. Bufer ¡:wders, talets or "ready to use"

buffer solution according to NB stadad.3. Deinerlised water to prepae solutions

an to flush the sensors.4. Thereter if no tematue ccpetor is

used.5. So suitale cleaing liquids (e.g. 5 %

hydrocloric acid) to clea the electrodesprior to calibration

Tl- ON

Q1::-:, "..::: I~ . £OIL'''''~_..____ ---~ ....i'=..' i

____ --_..-..( cot s:;.'."~sn r:.".o=::

~Tri~'",-"

.. .......

'l

~~__S2______; E012:;'-. .

AI' ~:;!:'~ I ,

SiO i

.~.

-------! COJ3::S~

-- -- - - EOl ;:;:;.9

'0 _ ___ - -! con:;~'''JSL :J?~'3

The calibation procedure is describ in paragraph 5-3-4-2. 25

Page 28: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

5-3-4-2. Calibration proure.Pour 50 to 100 nù of freshly prepaed b.fer solution into the beers an sta

calibtion by actvating the key IIsrANARII.

DISPIY KE RET REpH 88.88 activate "ST" srANAR no action, 'I0 wats for a:intion

STAN actvate "EN '! OFF if auto. tem. ca. is selectec orùy(anlyzer is in maintence lOe)

'I OFF activate ''N'' '! CN ths action will be necssa to entermaual tem. ca. duing caibtion

'I OFF actvate "EN" sr. 1 action for roth II'I OFFII an II'I ONII

an (if maua tem. ca. was selected'I rn IIsr. 111 was inicated direcly)sr. 1 activate IIEN 'I°C 888 maual temtue ca be set

or orINSERT no maual terature nnst be set

'I°C 888 lIuplI or Il00WNII '!°C xx maual tem. is set betwee 0 - 100 uC

activate IIENII an "INSET will be inicated

INSER inse sers in l:fer solution 1

INSERT activate IIEN" 8.88 8.88 stailizing tim (coter) an mesued.. vaue are inicated

x.xx 8.88 sr.l 8.88 selected bufer solution is incatedor orEOl /E06 eror rnsaage (see note 1)

ST.l 8.88 activate i 'EN" or sr.l EN correct l:fer vaue is inicated oractivate IlNEII sr.1 xx ca be selecec with key IlNEII an

sb2re£ acti~tig IIEN"ST.1 EN offset voltage is cacuated an slop

rem unchged. A selection of 1 or,- l.-int calibration ca be made.ST. 1 EN activate IlREII AP 8888 1 ¡;int calibration is seleced an

or calcuated offset voltage is incatedE02 or eror messaqe (se note 2)

AP 8888 activate IIENII RE 1 ¡;int calibration is finshed ananlyzer will retu to mesuring m:e(after actvatinq the key IIENII)

Sl.l EN activate "EN" sr.2 2 ¡;int caibtion is seeced.sr.2 activate "EN" '!°C 8.88 value of maual temtue ca be set

or (if selected before) orINSERT auto. tE!. cr. was selected

'I°C B.88 lIuplI or "OOII '!°C x.xx tematue of b.fer solution 2 is setactivate IIENII

INSER rine electrcàes an irse then inhiffpr ~llltion 2

NO: 1. Stailizing time is out of limts of preset response tie (coe EOl) ortemature is too high (a:e E06). Activate ''R1I an re-stacalibration procecure.

2. Offset voltage (asymtr ¡;tential) out of limts (err a:e E02).Activate ''R1I and re-sta calibration procure.

26

Page 29: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

DISPIAY KE RET REINSER activate "EN" 8.88 8.88 stailizing tim (counter) an mesured

vaue are incated

x.xx 8.88 Sl.2 8.88 selected h.fer solution is incatedor or

" EOl fE06 err mesaqe ( see note 3)ST.2 8.88 actvate "EN" or SL 888.8 correc h.fer solution is inicated

actvate "NE': Sl.2 x.xx or ca be selected with key ''N'' an

actvate "EN"stored ~ctvatig "EN"

ST.2 8.88 SL 888.8 calcuated slope is incatedor orE02lE03l-E94 err mesaqe ( se note 4)

SL 888.8 AP 888 caculated offset voltage (asytror ¡ntential) is incated orE02/E03fE04 eror mesaqe (se note 4)

AP 888 RE p.t sers back in pross

RE activate "RE" PH 8.88 calibation is caleted. Anlyzerretu to mesuring m:e.

ro: 3. Stailizing criterium is out of limts of preset resp:mse ti (coe E01 )or tematue is to high (coe E06). Activate ''R'' to re-stacalibtion in h.fer solution 2 (ST.1 EN" will be incated).

4. Asymetr ¡ntential is to high (a:e E02) or slope is out of limts (coeE04). Activate "RE" to re-sta caibrtion.When selected h.fer solutions are to close (coe E03) the anlyzer willretu to the sta of calibration 2 after actvating the key "EN"( "ST. 1 EN" will be incated).

27

--

Page 30: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

'\

6. Recigution of the anlyzer.The anyzer has a wide rage of configutions (over 100) an it is likely tht

yo will fin the configution for your s~ific application. However, there is arisk tht the user ca cxnfe the anlyzer by progg errs. Therefore, a listof factory settings (DEFAULTS) is given an we strngly advise to use ths list nexto the values you are settig.Tae note to the position of the deciml point an chec each deviation. Nev forgetto ch whether a settig is mae intentionaly or by mistae.

Flow cha.Ths flCM ch ShCM the cclete progg m:e with different meus (sec

level opeting). ID~~~~RETunN NEXT DOWN UP ENTER

~ CONF. SET ~,

ST.ABLES I WSHTIMER ~ RESP. TIME:

6-1. Entering the prog moe.Depeg on the position of DIP switch 1 (incate: on the front plate) the

anlyzer ca be set with or without password protection ("PAS ON" or "PASswOFF" ). As stada, a passwrd num is set (factory setting). 'I prevent entering inthe progg noe, ths num ca be chge: by an intrt enginee.6-1-1. Entering with passwrd protecon.After actvating the key "PR" the display will incate "PRIl (bling).

The int.t enginee has the accss to the prog noe when he kn thcorrec passwrd num (DIP switch 1 is set to "p~ 00") 0

i

DISPLAY iæ RET RE"1,.".,11",

PH 88.88 actvate "PR" ~PR ~ no action, th anyzer waits for-..1'',1"'11 ,..- confirtion

,\' , " I I l J , ,activate "EN" PASS 5000 the cxrrect pass word num must: PR :."i, ",dll"'- be set

PASS 5000 "up" or "rx" PAS 5004 .factry setting (ths num ca bechaged by the intnt engin, seepaagraph 6-1-3)

PASS 5004 activate "EN" AL yo have passed the soft wae key anyou have accss to the proang rre(the anl vzer enters the maintence m:e)

28

--

Page 31: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

6-1-2. Entering without passwrd protection.After activatig the key "PRcx" the intrt enginee has the accss to the

prog rre diectly (DIP switc 1 is set to "PAS OFF").

DISPIAY KE RET RE\ i i' lit I l III..

PH 88.88 activate "PR" -: PR :. no acton, the anlyzer waitsfor

""11 /I., '/I'" cxnfir tion

, TT -T i activate "EN" PAS 5004 stored passwrd num is incated: PR ~11/..11111\'- (factory setting)

PAS 5004 activate "EN" AL you have access to the progamg roe(the anlyzer enters the mateceroe)

6-1-3. Olging th passwrd num.The IXsition of DIP swich 1 is incated on the front plate of the anlyzer.

Before entering a new password num, check the IXsition of the switch 1 ("PASRDON" is factry set).

Wl: The IXsition of the DIP switch is checked ever 5 secnds. After chaging theIXsition of the switch yo have to wait for ma. 5 secs.

DISPIAY KE RET RE.-

PH 88.88set DIP switch 1 to "PAS OFF"

\ i II l 1 I l , ii..

PH 88.88 activate "PR" :- PR 2 no acton, anlyzer waits for-/111/1""' \ coirtion

;PR~ ':. activate "EN" ::PAs:5004 password num (factry settig) ca be

'i i'' r l l 11' ,-;'ØI,.II" chged

;PASS:5004 'up" or "i:" -'pÄ$ , new passwrd num is set-; : xx'" \11''

/,.,...1\

1..11.. new passwrd num is enter 1 )~PASS:XX activate "ENII AL an. .." I " anlyzer go into maintence roe

AL set DIP switch 1 to "PAS~ 00". Softwae key is acti VP..

1) Mae a clea note of the new entered passwrd num an store it saely.

6-2. Selection of the progg rrues.After entering the progarg roe (see pagraph 6-1) the intrent enginee

has the accss to the progg meues. Freqently activating the key "NE" willresult in an incation of meues acoording to the seence as listed in the digramon page 30.

ID :1. After maing a selection you ca enter the iru by activating the key "EN".2. After finshing the settings in a meu, you ca retu to the meu selection

moe by activating the key IRE" an you ca mae next selection.3. When you are entirely ready with settings in the different meues, then activating

the key "RE" twice will bring you to the mesuring m:e an is proggcapleted. 29

Ii

Page 32: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

DISPLAY PRCGAL all functions of the pross contact outputs

mA ou for 1. trsmssion signl (0 - 20 mA or 4 - 20 mA)2. measuring range3. "oow" function in maintence moe

¡ TE. em.

4. "BUR OUT" function during fail alar.

for 1- autcmtic or maual tanture cction2. tenrature reading in DC or OF3. c:fficient for addition tem. cap. of the

proess

REP. TIM for 1 . staility reqirement during calibration2. autcmtic or maual acceptace b.fer solution

value during calibration.

WSH TIM for 1 . autctic wash cycle (yes or no)2. interl of the wash cycle3. wash tim4. recover tim

ST. TAB for 1 . checkig the 5 progar b.fer solutions2. replacing one or more NB stads

IM. CH for 1. imce check (yes or no)2. imce check in measuring or maintence moe3. intervl of imce checks4. high limt of the imce of glass an reference

electrodesa:. SEr for defintion of the settings which ca re done in the

"SEr" mode an which you do not disclose to theo ator.

6-3. Alar menu.In the alar menu all functions of the procss alar contact outpts ca be

progar according to the speifications as describ in paagrph 2-1. In thedescription on next pages san exles are give. The flow ch shows the ccleterou tine.

USJ; ABTIONS:IIGH = setpoint high alarmDIR = direct action (fail safe).

Contact is closed at poer offan fail alar

= continuous alar= fixed duty cycle= p=.. i ode= dead tim= mi. duty

CDFIXPERIJD Mi

PISClDB

D MARNG

= setpoint ow a= reverse action

Contact is closed at po offan fail alar.

= pulsating alar= controlled duty cycle= dea zone= ma. duty= rage

I.RE

N: :1. The step by step instrctions on next pages are intened for the first ~ir user.

These instrctions illustrate how the display will appe when the settings arecorrectly perform in the configuation exles.

2. For those with previous exrience on the'I0 anlyzer, please read the manualto familiarize yourself with the advance features of this anlyzer.

Page 33: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

6-3-1. Flow ch of the al meu.

8.8 I

Ai

& Proceed OJ dlJCfibrdundlf AI foulinl

8.8 I

A I

31

Page 34: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

6-3-2. Exles.6-3-2-1. Exle 1: CN/OFF LI DIR/RE AL DEA ZO DEA TI

alar 1 ON HIGH RE CD O. 1 pH 1.0 s

alar 2 OFF - - - - -

DISPLAY KE RET IiAL actvate "EN" Al al 2 mut be disaled

Al activate "NE'.' A2 00 action, waits for confiition

A2 actvate "mr" A2 CN al 2 ca be disaled

A2 CN actvate "NE" A2 OFF alar 2 is disaled

A2 OFF activate "mr" Al 00 action, waits for confinntion

Al actvate "EN'! Al CN alar 1 is active

Al CN activate "mr A1 IJ high alar mut be set

Al I. activate "NE" Al HIGH high alar is selected

Al HIGH activate "mr" Al DIR reversed action is reqired

Al DIR activate "NE" A1 RE reversed action is selected

Al RE activate "mr'! A1 CD continuous alar is reqed

Al CD activate "EN" DB 0.1 dead zone of 0,1 pH is reqed

DB 0, 1 activate "EN or 1.0 dea tim of 1.0 sece is reed-

IY 1.0 activate "EN" A2 prg of exle 1 is f !nshed

6-3-2-2. Exle 2: Continuous alar of exle 1 is replace by pusatig al

ON /OFF LIM DIR/RE PI DEA ZO DEA TI.FIX Cl

PER. IXalar 1 CN HIGH RE 10 s 50 % 0.1 pH 2.0 s

alar 2 OFF - - - - - - -

DISPLAY KE RET REAl activate "EN" Al CX after activating "EN" severl ti

Al CX activate "NE" Al PI p.sating action is reqed

Al PIS activate "EN" Al FI fixed duty cycle is reqed

Al FIX activate "EN" PER 1 0 peiode of fixed duty cycle is 10 s

PER 1 0 activate "EN" IX 50 CN/OFF ratio of 50 %

DU 50 activate "EN'! DB 0.1 selected dead zone of exle 1

DB O. 1 activate "EN'! or 1.0 selected dead tim of exle 1

or 1.0 activate !'EN" or 2.0 dead ti is chaged to 2 s

or 2.0 activate "EN" A2 pramg for alar 2 ca be made orretur to prc:ar m:e (key "RE"). --

Page 35: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

6-3-2-3. Exle 3. Fixed duty cycle of exle 2 is replace by proportionallyalaAL LIM DIR/RE PIS .

FI Cl PER D MI D MA RNG or

ala 1 CN lIGH RE Cl 10 5 % 90 % 2 pH 2.0 s

alar 2 OFF ,- - - - - - - -

DISPLY KE RET REAl activate "EN" Al FIX after activating "EN" seveal tims

Al FIX activate "NE" Al CI a proportionally controlled outpt is set

Al Cl activate "EN" PER 1 0 the peioo of the contrlled acton is 10 s

PER 1 0 activate "EN" DMI5 duty of 5 % for setpoint 'Value

DMIS activate "EN" D MA 95 duty of 95 % at en of control range

D MA 95 "up" or "OO" D MA 90 ID. duty is chged to 90 %

D MA 90 activate "EN" RNG 2.0 contrl range is 2 pH unts

RNG 2.0 activate "EN" A2 progg for A2 ca be made or retuto progam moe (kev "RE")

6-3-2-4. Application of a procs alar output (continuous alar).In the figues 6-1 (high alar) an 6-2 (low alar) the function of a continuous

alar is visualised. These functions ca be used e.g. for controlling a solenoidvalve (dosing). The dead tim (1J) an dead zone (DB) are set to prevet "rattling"of the relais contacts.

- High alar.,An increasing pH value will pass the setpoint value at tl. The relais contactoutpt will close at tl + IJ an remin closed until the pH value deceases to thesetpoint value - DB (at t2). At t2 + IJ the relais contact will ope again.

- I. alar.A deceasing pH value will pass the setpoint value attl. The relais cotactoutpt will close at tl + or an remins closed until the pH increases to thesetpoint value + DB (at t2). At t2 + IJ the relais contact will ope again.

tli- 0.1i

i

f- OJ.t2 tl

!-D11

setpJint + DBTi

¡- O.Ti

t2

setpJint - DB

p- t -tsetpoint setpint

DB = dead zone (in pH)or = dead tim (in secnds)

Figue 6-1.Oontinuous high alar function

DB = dea zone (in pH)or = dead tim (in secnd)

Figue 6-2.Oontinuous lCM alar function

33

Page 36: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

6-4. mA outpt menu.Basically, this meu is for proarg of:

1. the trssion signl an mesuring range2. the outp.t sign during maintence rre3. the output signl during a fail condtion.The calete routin is given in the flow cha.

6-4-1. Trssion signl an mesuing rangeAs delivered, the anl¥Zer is set for a tr-

mission signl being ordered (rage cce) anincated on the tex plate on the front.

'Ie trssion signl ca be set to 0 - 20 mAor 4 - 20 mA.

'Ie measuring range ca be set betw -1 to+ 15 pH (-1500 to +1500 mY) as speified in thegenral speification (see paagraph 2-1 ) .

Exle:The 'I 20 anlyzer has be ordered with:- mA output 0 - 20 mA- measuing range: 0 - 5 pH

These values mut be chaged to:- mA outp.t 4 - 20 mA- measuring range: 2 - 12 pH

The procedure is describ on page 35.

'f CH

E" ~\~I~\~flEnRH NE DOWN UP El~I

34

Page 37: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Select the mA output menu (see paagraph 6-2).

DISPu\Y KE RFT REmA out activate "EN" 0 - 20 mA mA output signl mut be chaged

o - 20 mA activate "NE" 4 - 20 mA confirm selection

4 - 20 mA activate "EN" PHL 00.00 beincale value measuing range

PH 00.00 "up" an "IX" PHL 02.00 reqre: beincale value. Cbnfin. selen.

PHL 02.00 activate · 'EN" PHH 05.00 enscae value measuing range

PHL 05.00 "up" an "rx " PHH 12.00 reqed enscale value. Cbnfinselecon

PHH 1 2.00 activate · 'EN" HOLD ON progarg is ccleted. Retu tomA outpt meu or nex settinq.

6-4-2. Outpt signl during maintence ioe.A selection ca be made betwee:

1. last mesure: outpt value remin constat ("HOLD LA")2. output value goes to a well defined value ("HOLD FI")3. output value is "live", also in maintence mce ("HOLD OFF").

Exle:Factory settingReqire: setting

"HOLD LA" function."HOLD FIX" fuction (fixe: outpt value: 23 mA).The pross canter reczes ths as a "BU UPSCA" value.

DISPu\Y KE RET REmA our activate "EN" 4 - 20 mA entere: in mA outpt meu

4 - 20 mA activate "EN" HOLD 00 after activating "EN" sevel timHOLD ON activate "EN" HOLD LA "HOLD ON" is entered an "mID "Ll"

must be chage:HOLD LAT activate "NE" HOLD FIX confin selecion

HO FIX activate "ENi HOLD 00.00 selec the reqired mA outpt valuemA

HOLD 00.00 "up" an "IX" HOLD 23.00 confin selectionmA mA

HOLD 23.00 activate "EN" BU OFF proaig caleted. Retu to mAmA outpt meu or nex setting.

6-4-3. mA output in fail condition.A selection ca be made betwee up scale ("BU UP"), dOl scale ("BU ra") or

"BU OFF" (no fail alan) functon.

DISPu\Y KE RET REmA our activate "EN" 4 - 20 mA entere: in the mA outp.t menu

4 - 20 mA activate "NE" BU OFF after activating "NE" seveal tim

BU OFF activate "EN" BUR OFF without fail alan

activate "NE" BU ON with fail alarBUR ON activate "EN-l BU up up scale alan (23.00 mAl

BU up activate "EN" 4 - 20 mA up scale alan is selected or

activate "NE" BU OO dOl scale alar (0 or 3.3 mAl isseleced. Cbnfirm selection an retuto mA outot menu.

-ic;

Page 38: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

6-4-4. Teratue cction menu.Basicaly, this meu is for prog of:1. the tanratue cofficient2. maual or autartic temtue cction3. display reaing in °c or OF'!e calete routie is give in the fla, chart.

6-4-4-1. Tetue cofficient.(;erlly, pH meters haVe a circuit for

catig the inluen of temature onthe mV output of a pa of eleccxes (socalled: NEST a:tions). Net to thscotion, the 'I0 anlyzer includes thepossibility for a:tig the tematuecoficient of the proess liqud itself.Since Yokogwa Electrfact is not inormedalt your speific reqents, the factorysetting is 0.00 pH/oc.For mesuremts in allised pue water,l:iler fee water, etc. the factry settingwill be -0,35 pH/oc.

It is easy to detene the tematuecofficient of you pross:1. Set tenratue cofficient to 0.002. Mesue pH value at nonnl procss tempe-

ratue: = pH(tl) at temratue tl.3. Tae a Sale of the procss liqud

an chage the tar ture to t2.Alia, the electrcxes to stailize (athenstatic bath is necessa).

4. Read the pH value at t2 = pH(T2).5. Calcuate the temratue cofficient

fran:

pH(tl) - pH(t2)Tc = -------------- x 10tl t2

F' æA~~~~~RERN NEX DOWN UP ENc§

~ME:ASURE (

¡PH BBBel

6-4-4-2. Teatue cation.The 'I0 anyzer is develope for autatic temtue cation using apt 1000 cator. .

If mesured without tanratue cator, maual temtue cation mut beset.

6-4-4-3. Display reading in °c or of.Basically, the 'I0 anlyzer is develope for readig in °C. lIever, display

reading in OF (mesuring m:e) is also possible.00: The tales of the b.ffer solutions (NB sta) an te:tue cofficient

are in °C.

Exle of settings in the temtue cation menu:Ordered setting : maual taiatue o:petion with a tem. cofficient = O.

Reqired setting: automtic temature cotion an a display reading in OOC.Teature cofficient: -0,2 pHIlO °C.

36

Page 39: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Enter the progg rre accrding to pagrph 6-1 an select the tematuecxtion meu (see pagrph 6-2).

DISPLAY KE RET RlTE. a: activate ·

'EN mEF 0.00 tem . cofficient ca be set

CX 0.00 "up" an "IX" mEF -0.20 reqed tem. cofficient

mEF -0.20 actvate "ENJ' TE . MA te. cofficient is entered antem. cc. ca be selected

TE . MA actvate 'NE" TE. NJl. auto. tem. ca. is set

TE. AUT. actvate "EN" TE. °c auto tem. ca. is entered an displayreadinq ca æ set

TE. °c actvate · 'EN" mEF -0.20 reaing in °c is entered an progar-miq is cxleted

-mEF -0.20 actvate "RE" TE. aM retued in the tem. cctionmenu (see pagrph 6-2).

6-5. Response tim meu.In ths meu the stailizing criterium for

the calibration proure ca be pro(for pH mesuements only).After imsing the electrcxes in a Wffersolution, the mesuremt has to be stailize.As soo as stale redig is achieved, thecalibtion proure ca be peormed.A selection ca be mae betwee:1. maual acceptace2. autatic accptace.

- Maual acceptace.The opeator waits for stale reaing in orderto confin the readig in the caibtion m:e(activate "EN").

- Autatic acceptace.Stale reading mut be perform within a pro-gramed response'tim. If not, the display willinicate the err iæsage EOl (slCM response).

'!e stailizing criterium is defined as

b.pH/6t (see fig. 6-3) whch item nust beseqentially progar.l "((EPiAl~E

:ia.

t1l,lSERT

l

Figue 6-3.Stailizing criterium.

-~L

FI CHT

.~~R~d 8N'~~I~Irr'~'i

=:: I

D.8 I

I ' ,IMIN

~99 !

TIME

i

tU: Factory setting is 0.01 pHIS s. Never chge this criterium too quick. A highbpH (e.g. 0.04 pH) an a short Ät will give a bad calibration and conseqently,

; l'nCcurate lT--surE'J1t. "),

Page 40: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Exle:1. chage fi; maual accptace to autantic acceptace2. iipH = 0.013. ct = 30 secnd4. ma. respoe tie = 10 nùutes.NJ: Ths exle is for those application reqing most accute measuaæt

an conseqently, irst accuate calibtion.

====================================================================================WAG

====================================================================================To acheve optin acccy of the calibration, the factory settings for autanticacceptace are caefully selected (DEFAULT values). fun' t chge these settingswithout gcx knledge of you procss reqemts.

Select the response ti meu according to paagrph 6-2.

DISPLAY KE RFT RDREP TI activate "EN" STAB MA menu is entered

STAB MA activate "NE" STAB AUT acceptace is chged fran ma. to auto.

STAB AUT activate "EN" dpH 0.01 auto. acceptace is entered an ~pH cabe set

dpH 0.01 . activate · 'EN" dT 5 L'pH is entered an t ca be set (in s)SE

dT 5 "UP" an "OOWNIl dT 30 ct is chaged. Confirm selection.SEe SEedT 25 activate "EN" TIME 5 ~t is entered an response ti ca beSEe MI chaaedTIM 5 "up" an 1l00"- TIM 1 0 response time is chged. ConfirmMI MI selectionTIM 10 activate "EN" STAB AUT progamg is a:pleted. Retu toMI respOnse time meuSTÃB AUT activate "REII REP. TI meu is selected (see paagraph 6-2)

6-6. Wash tim meu.Basically, the wash tim meu has two subes:

1. the wash cycle tier2. the halftim value senr chec.Both sules ca be pro in the wash tim meu.

6-6-1. The wash cycle ti.When washing irt be caied out autantically i then the interl tim betw

washing ("W. INII), the wash tim ("W. TIII) an the relaxtion or recer tim("R. TIII) must be pro.Nar :1. Progaming of the wash cycle timer is for jet cleag only. This methcx of

cleag disturbes the mesuremt.2. The progam wash tim an recer tim remin also for the maually stated

washig cycle.

6-6-2. The halftim value sensor check.Duing the recver time of the wash cycle a halftim sensor check ca be

perfonn .38

Page 41: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Exle:_ Washig is made with a 5 % hydrochloric acid._ Before staing the wash cycle the mesured pHvalue of the proess is pH 7.

_ After staing the wash cycle, the pH valuedeceases to pH 2 (acid on the ser).

FI æAT

-i .n"" ;,lJ(;l I

The'l 20 anlyzer aveges both values. D.ingthe recer tim the averged value nnst beachev with the proam halftim.'!e pro halftim value is depeent onthe applied ty of iæsuring electre (tyof glass), the temtue, the flow rate anfouling of the glass merane, etc. '!erefore,ths time nnst be detered by trial an eror.A goc guidline is to eqlize halfti anrecer tim.

or .." ::1

Ò,i*ir'

~.~

..cr '"

us ,

1"","''4 ø õ; I

~"I Will

-.'iU

:=.~ 1

'!e principle of halftim senr chec is shoin figue 6-4.

.. 'l"''''iIUl(-I1II

.. WAIT.,1i1Ii1i1

I,--- - --- -1-- -1

1

,I1

iIIII1 Iu~

l I W5H i"~(~ i

Figue 6-4. -.. tPrinciple of halftime value sensor chec.

rP: When the eror message EOl is incated, then the halfti value is exceed.This does not necssaily me tht replaceent of the electrcres is neeed.EOl may also inicate tht:1. the reseroir of the cleag liqud is anty2. tlie solenid valve has a leage3. the pH value of the process liquid is deceased during the wash cycle.

A. Progg of the wash cycle ti.Select the wash ti meu (see pagraph 6-2).

DISPLAY KE RET REWS TI activate "EN" wc OFF wash timer was switched off

wcr OFF activate "NE" ~ CN wash tim is switched on. Confirselection

\VC CN activate "EN" W. INl 0.0 interl = 1 0 hos. Confirm selection

HR or chaae usina keys "up" an "I:".

w.IN0.0 activate "EN" W.TI 0.1 wash time = O. 1 mi. Confirm selection

HR MI or chge using ''u'' an "OO".W.TIM 0.1 activate "ENH R.TI 0.2 reer tie = 0.2 min. Confirm se-

MI MI lection or chge using "up" an "ro"R. TL~ 0.2 activate EN" HCH OFF progaig of wash cycle is a:leted.MI Retur to wash timer menu (key

RE") or next setting.39

Page 42: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

B. Setting of the halfti value senr chec.

Select the wah tim meu (see pagrph 6-2).DISPLAY KE RET REWS TI activate "EN" wcr ON wash timer meu is selected

wcr ON activate ''N'' HCH OFF inication after activatig "NE"several tim (señr chec off)

HCH OFF actvate "NE" HCH ON coir selection (senr ch on)

HCH CN activate "EN" H.TI 0.1 chge halftim value (if reqred)MI MIH.TIO.l "up" an "IX" H. TIM 0.2 confinn settingMI MIH.TI 0.2 activate "EN" ~ON progamg is capleted. Retur toMI wash cycle tim mel) (key "RE").

6-7. Sta tales meu.As stada Yokogawa Elecofact entered

5 tales of lifer solutions (NB stads).In this meu you ca overiew the protales an/or enter your ow tale (s ) .The prog tale (s) will be erased.

A selecti~n ca be made betwee 2 sues:1. overiew of the lifer tales2. proamg of a new l:fer tale.

6-7 -1. Overiew of the l:fer tales.In this sule ("sr. vn")you ca

only overview the progar liffertales. After entering the sue(key "EN") tale "ST" is inicated.Wi th key "NE" the reqired tale ,cabe selected. After activating the key"EN" the values are inicated withinterls of 2 s.

6-7-2. Progamg of a new l:fer tale.In ths sule ("ST. PR" ) you ca

prog your ow l:fer tale.EXLE :For calibration of the mesuing systemyou always use the l:fer solution pH = 7of Yokogawa Electrfact an you wat toprogam the tale in the 'l0 anlyzer.Tale "ST3" is proar for a NBliffer solution of pH = 7. '!e values ofthis tale at vaious temratus are neato the values of the Yokogawa Electrofacttale of liffer solution pH = 7.Therefore, it is reced to ersetale "ST3".

FIW OIT

I ST TA~:'ES I

6I ~l ~ROG ; ~

c? "ISTI ¡;3,89~ ~sr5 ee,58

06 "é. ~~i :rvlF'N I

1m ?e,ee ';::;',~ Sf 53" 8eØ? $'1000 89,eei

I,lt'I 2MC

I

I 100 e 2, e:' J

.1:.-í

1100 ae,eu I~U

i

NJ :1. The 'IN20 anlyzer indicates a tale by displaying the value at 25 °C.2. The values of the Yokcaawa Electrofact liffer solutions at vaious tertures

are printed on the the l:xes of the liffer p::Mders.40

Page 43: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Selec the stadard tales meu (see paagraph 6-2).

DISPLAY KE RET RrST TAB activate "EN" Sf VI the stad tales meu is selected

Sf VI the progar values ca be checed(not chged)

S! VI activate "NE-ir Sf PRCX sue for proam the new values.Cbnfii: selection.

S! PR activate "EN" ST1 1.68 1 st tale of tuffer solution. O1ge to3m tale (key "NE")

ST1 1.68 activate "NE" Sf3 6.87 3m tale is selected. Coii selection

S!3 6.87 activate "EN" PR 6.87 value of the NB b.fer solution at 25 "C.'Is value nnst be chaged.

PR 6.87 "up" an "LOiT PR 6.98 value of the Yokogawa Electrofact J:fersolution at 25°C. Cbii: selection.

PR 6.98 activate · 'EN" 000 6.98 the tale of the NB bifer solution iserased an new value (at 25°C) is set

000 6.98 "up" an "LO" 000 7. 1 0 the value of the new tufer solution ato °C. Cbnfii selection.

000 7. 1 0 activate "EN 005 6.98 selection is confll. The value at. -5 "C-ca be set.

005 6.98 "up" an -"OO" 005 7.07 the value at 5 "C is set. Cbnfumselection.

005 7.07 activate "EN" 010 6.98 selection is confirm. Value at 10°Cca be set, etc. etc. to value at 100°C.

100 7.02 .. acti vateTTEN" ST VI values are entered and a check ca bepefoi:ed, if necessa

S! VI activate "EN" Sl 1.68 1 st tale is selected. O1ge to 3mtale (key "NE")

S!3 6.98 activate "EN" 000 7. 1 0 all values are inicated at 2 s interls

100 7.02 ST VI proarg is fineshed. Retu tosta tales meu (key "RE").

6-8. Imce chec meu.In this meu the dync in-line ser chec

of ooth the glass an the reference electrode cabe proar. A selection ca be ride whetherthe check is peform in the rrsuring noe orin the riintence noe.Since rrsuemt is t61raily distube, thechecks in the mesuring noe are pefoi:ed at(progarle) inteials. '!e featue of checgin the maintence noe (continuou) isintroduced to get arou ths difficulty.

The qulity of the senr is exressed as animance. Espeially for the reference serit is imrtt to kn tht the imce ismeasured betwee liqud ea an reference pin.Therefore you riy not progam the spifiedimce of the reference electrode (thsvalue is mesured in a satuated KCl solution).Geerally, the conductivity is much la,er

(resistace much higher) an therefore the valueto be adjusted must be must higher.

í ll-P. CJo!:'" J

ó

~6' ~';9..

r CHI( i ME.&5 ~~ CrlY. ..'NJj

oI

f J..'T õõ, I~ '" '"y ¡""N, ===-1

¡;J:I" E. ¡.fI~

Ur iG!H ¡,'i;r~'..

g-" - .. (~~;1C.v,.~""i '''E'''-'

IO.lP CM!CKI

p,.~__j",:",c,,( 41

.. _ ~ iI;' - :-;. = = j

Page 44: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Exle:As order, the 'I0 anlyzer is for pue water application (TC = 0.35 pH/10°C).For ths application you will proam an senr :ice check during mesuemt.

~:1. '!e :ice of the reference ser is in kOhm an of the mesuring senr (pH

an OR) in MJHm.2. '!e inicated :ices are ccted to the reference temtue (25°C).3. I. limts:

- reference senr:- glass ser- æP senr

o kOhm5 MJhmo lbhm.

Select the :ice check meu (see paagrph 6-2).

DISPLAY KE RET REIM. OI activate "EN" æK. OFF entered in the meu an im . chec is

switched off

æK. OFF activate "EN" æK. ON im. check is switched on. O:inselection.

æK. 00 activate "EN" æK. MA im. check is switched on. Select formesuring or maintenceßode.

æK. MA activate "NE" æK. ME mesuring noe is seected. tbi.selection.

æK. ME activate "EN" LIN 100 interl tim betwee is 1 UU ID.MI. 1 00

1. IN 1 00 activate "EN" ZR 500 :ice (high limt) of referencesensor. Ths value must be 2000 kOhm.

ZR 500 "up" an "i:" ZR 2000 high limt is 2000 kOhm. Cbin selecion

ZR 2000 activate "EN" ZGE 2000 imce (high limt) of glass ser(in MJhm)

ZGE 2000 activate "EN" æK 00 prarg is cxpleted. Retu to:ice check meu (key ''R'')

æK ON activate ''R i IM CH :ice check meu (see paagrph 6-2).¡

6-9. Configuation meu of set noe.In ths meu the intnent engineer ca enter the settings whch are prteced by

a pësswrd in the progam noe. Ths feature prevents an opetor to set values inthe set rre.It is pJssible to disale the cxlete set noe.As delivered, settings in the set noe ca be done by an opetor conform thesettings of a conventional intrt.

Exle:You have ordered a stada intrt with alan an the opetor ca chge thesetpJints of it.You wat to disable these settings.

NJ: "ON" = setting ca be done in the set noe."OFF"= setting must be done in the progam noe an is protected by the

pëssword .

42

Page 45: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

6-9-1. Flow ch of th o:nfiguation meu set noe.

~I~I~~~RETRN NEXT DOWN UP ~TV~ ~

A2 ON

TM ON

;~

AP ON AP OFF

I'~

SL ON

1 T P ON

43

Page 46: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

6-9-2. Decription of prog in the configution menu set m:e.

Select the coigution meu of the set m:e (se pagrph 6-2).DISPLA KE REULT RECX. SET actvate !'EN" Al ON entered in the meu. Chge settig

Al CN actvate llNEII Al OFF ala 1. Co in selecon.

Al OFF activate IIEN" AlL 0.00 al 1 is disaled for the opetor anseb:xÜnt (low) of al 1 ca be chged

AlL 0.00 "up" an "roII A1L 2.00 setpint ( low) of alan 1. CoruTnnselection.

AlL 2.00 actvate "mr" A2 ON al 2 ca be disaled.

A2 CN activate "NE" A2 OFF alan 2 is disaled. Confin selection

A2 OFF activate "EN" A2H 14.00 setpoint (high) of al 2. CÒnfinnselecion or chge e.g. to 12 pH.

A2H 14.00 "up" an "OO" A2H 12.00 setpint (high) of al 2. Coinselection.

A2H 12.00 activate "EN" 'I ON maual temtue. You ca chge rrreitems (see flow cha on page 43).

'l CN activate "RE" CX. SE retued in the conf. meu (seepaagrph 6-2).

6-1 o. List of display messages duing progg.

NJ: Display messages which are not listed in ths tale are listed in paagraph4-2-6.

10 - 20 ID t outpt sign betwee 0 - 20 ID

i 4 - 20 ID I outpt signl betwee 4 - 20 ID

1888 888 I Teature an mesured value ar displayed in Stad tales meu

I Al i sta defing alar 1

I Al FIX Ialar 1 has a ¡:sating acton with fixed duty cycle

I Al OFFIalan 1 is disaloo (setpoint of ala 1 nnt be set in pr m:e)

IAl ONLalri 1 is enled (setpint of al 1 nnt be set in set m:e)

I Al mmIcontiuous alan of outpt 1

I Al CIIP-cntrlled outpt of al 1

i Al OIR I diect (fail sae) outpt of al 1

I Al HIGH I al 1 is a high ala

I A1 I. I alar 1 is a low alai

I Al PIS I alar 1 is a ¡:sating al

I Al REI

reverse (non-fail safe) outpt of al 1

I A2 I(see alai 1 incation )

44

Page 47: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Display messages (cot.)

AI1S I iæu for setting ala fuctionsAP OFF I offset voltage (asytr p:tential) must be set in proam m:e

AP ON I offset voltage (asyntr p:tential) mut be set in set rreBU OFF I no b.-out signl duing fail oond tion

BU ON I b.-o sigrl duing fail cotionBU OC li-out sign is 0 or 3.3 IDBU UP li-ot sign is 23 IDæK. OFF autatic senr imce chec is disaledæK. 00 autatic ser imce chec is enledæK. ME autatic senr im chec in mesuin m:e

æK. .M i. autatic senr chec in nainten m:etemtue cofficient (dpH/10 °C)CDEF 8.88

CD. SEl

D.MA 88

D.MI 88

DB 8.88

dpH 8.88

dT

Dr 8.8

I IX 88

IH.CH OFF

IH.CH ON

IH.'I 8.8

I IDLD88.88

IIDLD OFF

I IDLD ON

IIDLD FIX

IIDLD LA

IHR. 88.8

I I.IN 888

oonfiguation iæu of set m:e

I ma duty cycle of P-cntrolled alar

I mi duty cycle of P-cntrolled alar

dea zone (in pH unts)

rage (in pH unts) for acceptace of wffer vaue

I stale tim (in secnds) for accptace of b.fer vaue (display isalterting with sæ message)

L de tim (in se~). Display is altertig with sæ message.

I ¡:centage of ¡:icx ti duing whch alar is actveI halfti value check duing jetcleag disaledI hafti value check duing j etcleag enledI tim duing recer in whch haf vaue mut be reached (display isalternting with MI message)

I outp.t signl for IDLD FI (altertig with ID message)

live outpt duing naintence m:emId ID outp.t duing niatence rre

ID outp.t ru to a speified value in naintece m:e

ID outpt is fixed at last mesured value in maten rrehours (always altertig with anther rrssage)

intervl for autantic ser imce chec in mesuring m:e(alterting with MI message)

45

Page 48: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Display ræssages (cont.)

meu for progg ser ime checi IM. æælI IT OFF I

I'IToo II ID 00

I ID cu

I MI. 88

I mV OFF

I mV 00

I mVH 8888

I mYL 8888

I PASS 8888

I PER 88.8

I pH OFF

I pH ON

I PHH 88.88

I PHL 88.88

I PR. 88.88

I PRooI R.TI 8.8

I REP TI

I RNG 8.88

I sæ. 888

1 SL OFF

I SL 00

I sr. PRi sr. VII sr.88.88 I

I sr. TAB I

i srAB. ATJ I

I srAB. MA I

isotherl ¡:int of intersection must be set in proam m:e

isotherl ¡:int of interseon nust be set in set m:e

ID outpt siition in set m:e is enled

iæu for defing mA outpt signl in pro m:emiutes (always alterting with anther mesage)

proess caibration in set m:e is disaled

press calibration in set m:e is enledhigh lint of mesuring rage in mV

lCM limt of mesuring rage in mV

adjust passwrd mnnb

peicr tim of pusating alar (in send). Display is altertingwith sæ message.procss calibration in set ire is disaled

process calibration in set m:e is enled

high limt of mesuring rage in pH unts

lCM limt of mesuring rage in pH tits

value of b.fer solution (at 25 °C) ca be proprogam key is activated (anlyzer waits for confirmtion)

recer time after jetcleag (display is alterting withMI message)meu for progamg the response ti chec an accptace ofreadings during calibtioncotrl rage (in pH or mY) duing a P-cntrlled alar

seco (display is alatertig with anther mesage)

slope mut be set in the pro rreslope mut be set in the set m:e

sue for progg b.fer talessue for overiewing (or checg) the pro b.fer tales

set vaue (at 25 °C) of b.fer tale

meu for overiewing or reprog the hlfer tales

the analyzer detenes if inication during calibration is staleenughopeator ca decide if inication duing caibration is stale enugh

46

Page 49: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

Display ræssages (o::mt.)

I TE. !iI TE. MA

I TE. °c

I TE. of II Tr. ro II TI 8.8 II '! OFF II'!CN II w. IN 88.81

I W.TI 8.8 I

I WC OFF I

Iwcr 00 i

I WSH. TI I

autatic temture a:tion using a pt 1000 seor

m:ual temtue a:tion

display in °c

display in of.

meu for prog temtue o:tion

allowle Iæ. tim for reching stale criterium during caibration(display is alterting with MI message)tetue for m:ual temtue a:tion must be set in prom:etetue for maual temtue capetion must be set in set, ioeinter for autatic jetcleag in hos (display ~is alterntingwi th HR message) ,tim in miutes for jetclea9 (display is alterting with MImessage)autatic jetcleag is disaled

autatic jetcleag is enled

meu for proarg the wash cycle an haftim vaue senr check

4~

Page 50: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

7. Factory settigs (DEFAULT values).As stadard, the TM0 anlyzer is progar as listed in the paagraphs belCM. On

reqest or if ordered, the values ca be chaged.

7-1. DIP switches.

SW FACIY smIN REPASRD ON/OFF CN if set to "OFF" the opeator ca

. proam witht passing the sofwekey

PH/OR pH generlly, speified in range coe

AtJTIC ON/OFF 00 if no key is activated the anyzerretu after 10 -iinutes to themeasuring mce autantically.If set to "OFF" the anlyzer retuafter activating the key "RE".

7-2. Progam mce.

7-2-1. Passwrd menu.

PH 88.88 Ii

DIP switch "PASffRD"is set to "ON" and

passwrd num = 0

DIP switch "PASSW"is set to "OFF" an

passwrd num is 5004

"up" and "IX"

:j004:

DIP switch "PASffRDis set to "CN" anpasswrd numr is 5004(factory setting)

48

Page 51: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

7-2-2. Al meu.

I see pagraph 7-2-1

r-i

I

_J

LD :1. dea zone (DB)

7. duty (OO)

: 0.01 ~ DB ~ 0.20 pH1 ~ DB ~ 20 mVO. 1 ~ IJ ~ 20 s1 c: PER ~ 1 00 so ~ D.MI ~ 100 %o ~ D.MA ~ 100 %0.1 ~ RN ~ 10 pH

:10 ~ RNG ~ 1000 mV: 0 c: OO ~ 1 00 %

2. dead tim (IJ)3. peiode (PER) ,4. min. duty (D.il)5. rn. duty (D.MA)6. range (RN)

tn'E: Pram of duty cycle is limted by in on/off t:i of 0,3 s for alarcontact (4, 5,7).

49

Page 52: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

7-2-3. mA outp.t meu.

I PASRD I see pagrph 7-2-1I'",I

4 - 20 mA

HOLD OFF

BUR ON BU i:

LITS:1. low limt (pHL)

2. high limt (pHH)

3. hold output (HJID)

lU: pHH - pHL;: 2 pH

-1 ~ pHL ~ +15 pH-1500 ~ mVL ~ +1500 mV

-1 ~ pHH ~ +15 pH-1500 ~ pHH ~ +1500 mV

0.00 ~ Il ~ 23.00m

7-2-4. Teature cation menu.

I see pagraph 7-2-1

Tr MA

LIMTS:1. tematue cofficient : -1.00"- ilEF.-+l.00 pH/10 °c

NJ: When -Be or -BF is specified in range coe, then the tem. coefficient is setto -0.3 pH/10 °C.

i:,

Page 53: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

7-2-5. Resp:mse tim meu.

PASSWRDIIII11

see paagrph 7-2-1

LIMTS:1. range (dPH)2. tim (dT)3. respoe tim (TI)N': define TIM ~ dT

: 0.01 -c: 5 -c: 0.1 .c

dPH -c 0.05 pHdT -c 30 sTIM -c 10.0 II

7-2-6. Wash tim meu.

PASSWRDII11

see paagrph 7-2-1

LIMTS .1. wash tim (W.TI) 0.1 ~ W.TI ~ 1 0 min.2. recer tim (R.TI) 0.1 ~ R. TI ~ 10 II.

3. halftim (H. TIHE) 0.1 ~ H.TI ~ 1 0 II.

4. interl tim (W.IN) 0.1 ~ W.IN ~ 24 hous

si

Page 54: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

7-2-7. Sta tales meu.

see pagrph 7-2-1

005 13.21

6iI

o

tU:' '!e prog values are listed on page 18 of ths maua.

7-2-8. Imce chec meu.

PAS5iRD I see pagrph 7-2-1IIII

LIM :1. interl tim (I.IN)2. im. ref. senr (ZR)3. im. glass senr (ZGE)

2 ~ i.IN ~ 240 min.50 ~ ZRE ~ 2000 kOhm50 ~ ZGE ~ 2000 MJ

52

Page 55: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

!7-2-9. Coigution meu set m:e.

I PASSWRDIIii

(A1L) -1 .: A1L ~ + 15 pH

-1500 ~ AlL ~ +1500 mV

2. high alan (A2H) -1 ~ A2H ~ + 15 pH-1500 ~ A2H " +1500 mV

3. ma. tem ('I) -50 ~ 'I °c ~ + 150°C-58 ii 'I of ~ +302 of

4. offset voltage (AP) -120 ~ AP ~ +120 mV

5. slope (SL) +80 ~ SL ~ +105 %

6. isothernl ¡:int (IT) -1 ~ IT ~ + 15 pH

Page 56: Instruction pH/GRP X Manual PH/ORP Analyzer … · PH/ORP Analyzer Microprocessor based pH/GRP X ... Entering without passwrd protection ... Differtial amlifier with tw

7-3. Retrieving the factory settig.If the 'IN20 anlyzer doe not function correctly, e.g. by m!sprg, th the

factory settigs ca be retriev ver easily.Op the case an hale as follow:

. DISPLAY KES . - DISPI,.

pH 88.88 DISPIA/SF/~ DEFAUT

WAS/ST.

DFAT pH 88.88

Since the opetor has to activate seval keys at once, it is not p:ssible to havean untentionl chage of vaues.

54