Preface
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page3
Preface
ThismanualisintendedfortechniciansinvolvedwiththecommissioningandserviceoftheEnrafgaugeswiththeoptionalHCUboardorICU_HPIboardinstalled.ThedescriptionfortheanalogleveloutputsettingsarealsoapplicablefortheoptionalICU_HPOboardinthe973SmartRadarLT.
Adescriptionprecedingthetechnicalproceduresgivesthetechnicalinformationnecessarytounderstanditsfunctioning.Itisrecommendedtoreadthisdescriptionpriortoperforminganyoftheprocedures.
Safetyandpreventionofdamage
RefertothechapterSafetyintheinstructionmanualoftheapplicableinstrument(servo/radargaugeorindicator)fordetailedsafetyinstructions.
"Warnings","Cautions",and"Notes"havebeenusedthroughoutthismanualtobringspecialmatterstotheimmediateattentionofthereader.
• AWarningconcernsdangertothesafetyofthetechnicianoruser;• ACautiondrawsattentiontoanactionwhichmaydamagetheequipment;• ANotepointsoutastatementdeservingmoreemphasisthanthegeneraltext,butdoesnotdeservea
"Warning"ora"Caution".
Thesequenceofstepsinaproceduremayalsobeimportantfromthepointofviewofpersonalsafetyandpreventionofdamage;itisthereforeadvisednottochangethesequenceofprocedurestepsoralteraprocedure.
Legalaspects
Thecommissioningandtroubleshootingtotheinstrumentmayonlybeconductedbyqualifiedengineers,trainedbyEnrafandwithknowledgeofsafetyregulationsforworkinginhazardousareas.
TheinformationinthismanualisthecopyrightpropertyofEnrafB.V.,Netherlands.EnrafB.V.disclaimsanyresponsibilityforpersonalinjuryordamagetoequipmentcausedby:
• Deviationfromanyoftheprescribedprocedures;• Executionofactivitiesthatarenotprescribed;• Neglectofthegeneralsafetyprecautionsforhandlingtoolsanduseofelectricity.
ECdeclarationofconformity
TheEnrafinstrument,inwhichtheoptionalHCU,ICU_HPIorICU_HPOboardisinstalled,isinconformitywiththeprotectionrequirementsofECCouncilDirective93/68/EEC.RefertoCEdeclarationofconformitydeliveredwiththeinstrumentortotheinstallationguideoftheinstrument.
Additionalinformation
PleasedonothesitatetocontactEnraforitsrepresentativeifyourequireanyadditionalinformation.
Page 4
Tableofcontents
Page4
Tableofcontents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61.1 HCUboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61.2 ICU_HPIboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61.3 Overviewoffunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71.4 HCUcompatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4.1 Replacementoraddingoptionalboardin854ATGor854XTGservogauges . . . . . . . . . . . 81.4.2 Replacementoraddingoptionalboardin873SmartRadar . . . . . . . . . . . . . . . . . . . . . . . . . . . 81.4.3 Replacementoraddingoptionalboardin877FDI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.5 Optionalfunctionsinthismanual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Spottemperaturemeasurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92.2 Commissioningofspottemperaturemeasurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.1 Spottemperaturesettings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102.2.2 Additionaltemperaturesettingsfor877FDI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122.3.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122.3.2 Dataitems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132.4.1 Temperatureerrorrequest(itemEM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132.4.2 Temperaturestatusrequest(itemMQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132.4.3 Temperaturepointervalue(itemVV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3 Averagetemperaturemeasurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153.2 Commissioningofaveragetemperaturemeasurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.2.1 Averagetemperaturesettings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163.2.2 Additionaltemperaturesettingsfor877FDI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203.3.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203.3.2 Dataitems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.4 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213.4.1 Temperatureerrorrequest(itemEM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213.4.2 Temperaturestatusrequest(itemMQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213.4.3 Temperaturepointer(itemsVPandVV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4 Waterbottommeasurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.1.1 Waterprobeversions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234.1.2 Principleofmeasurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.2 Commissioningmeasurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244.2.1 Commissioningmeasurementswithroofsuspendedprobes . . . . . . . . . . . . . . . . . . . . . . . . 254.2.2 Commissioningmeasurementsforsidemountedprobes . . . . . . . . . . . . . . . . . . . . . . . . . . . 264.2.3 Waterprobesettings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284.3.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284.3.2 Waterabove/belowprobewarning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284.3.3 Dataitems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.4 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 294.4.1 Watererrorrequest(itemEW) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 294.4.2 HARTcommunicationerrors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 294.4.3 Waterlevel(ullage)statusbytes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 294.4.4 Waterbottompointer(itemsVPandVV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Page 5
Tableofcontents
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page5
5 Analogleveloutput . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 315.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1.1 HCUanalogoutputcharacteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 315.1.2 ICU_HPOanalogoutputcharacteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.2 Commissioningofanalogleveloutput . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335.2.1 Settingofanalogoutput973SmartRadarLTto“active” . . . . . . . . . . . . . . . . . . . . . . . . . . . . 335.2.2 Analogoutputsettings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 345.2.3 Calibrateanalogoutputcircuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 355.2.4 Verifyoutputcurrent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 385.3.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 385.3.2 Dataitems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.4 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 395.4.1 Analogoutputerrorrequest(itemEA) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 395.4.2 Analogoutputstatusrequest(itemAQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 395.4.3 Analogoutputpointer(itemsVPandVV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 405.4.4 Communicationboard(ICU_HPO)actualstatus(itemOA) . . . . . . . . . . . . . . . . . . . . . . . . . 405.4.5 Communicationboard(ICU_HPO)lastencounteredfatalerror(itemOF) . . . . . . . . . . . . . . 41
AppendixA ASCIItable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42AppendixB Configurationprocedurefor964WaterScout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43AppendixC Configurationprocedurefor765/766VITOwaterprobe . . . . . . . . . . . . . . . . . . . . . . . . . 46AppendixD Configurationofsidemountedwaterprobes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47AppendixE Partnumbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49AppendixF Relateddocuments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Page 6
Introduction
Page6
1 Introduction
1.1 HCUboard
TheoptionalHCUboardisusedinEnrafservogauges854ATG,854XTG,Enraf873SmartRadargaugeandEnrafIndicator877FDItointerfaceoptionalequipmentas:
spottemperatureelement; averagetemperatureelement; waterbottomprobe; pressuretransmitters.
andtoprovideforananalogleveloutput.
ThisboardinterfacestheEnraf762VITOInterfacefor:
averagetemperaturemeasurementwithEnrafVITOtemperatureprobe764; averagetemperaturemeasurementwithEnraf864MTT; waterbottommeasurementwithEnraf765VITOwaterprobe; averagetemperaturemeasurementwithintegratedwaterprobewithEnraf766VITOprobe.
viaHART input1.®
TheHART input2oftheHCUboardcanbeusedtoconnectpressuretransmitterforHIMS andHTGor® ™
vapourpressuremeasurement,ortoconnectasidemountedwaterprobe.
Notes:1 HTG(HydrostaticTankGauging)onlyapplicablewith877FieldDisplay&Interface.
HIMSisavailablewiththelevelgauges(servoandradar).
2 Standard,the762VITOInterfaceisconnectedtoHARTinput1andotherHART transmittersare®
connectedtoHARTinput2oftheHCUoptionboard.ItmustbeverifiedthatthemaximumvaluesforcurrentandpoweroftheHCUoptionboardHARTinput2circuitarenotexceedingthemaximumvaluesoftheconnectedHART transmitters.®
IfthevaluesofHARTinput2circuitaretoohigh,thenconnecttheotherHART transmitterstoHART®
input1andthe762VITOInterfacetoHARTinput2(onlypossiblewhenHARTinput1isavailable!).
ProducttemperaturecanalsobemeasuredwithaPt100spotelement(RTD)bytheoptionalboard.Theoptionforanalogleveloutput(4-20mA)isavailableindifferentcombinationswiththetemperature,waterbottomandpressuremeasurement.
1.2 ICU_HPIboard
TheoptionalICU_HPIboardisusedinEnraf973SmartRadarLTtointerfaceoptionalequipmentas:
spottemperatureelement averagetemperatureelement waterbottomprobe pressuretransmitters
Note:Analogleveloutputwiththe973SmartRadarLTisprovidedbytheoptionalICU_HPOboard.
Foradescriptionofthesefunctions,refertosection1.1.TheICU_HPIboardhasthesameinputfunctionsastheHCUboard;theonlydifferenceistheshapeoftheboard,whichmakesittofitinthe973SmartRadarLT.
Page 7
Introduction
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page7
1.3 Overviewoffunctions
TheHCU/ICU_HPIoptionboardhasthefollowinghardwarechannels:
SPOTinput: forspottemperatureelementPt100 HARTinput1: for762VITOInterface HARTinput2: forHART pressuretransmittersand/orexternalwaterbottomprobe®
4-20mAoutput: foranalogleveloutput(onlywithHCUoptionboard)
Thehardwarecombinations,togetherwithsoftwareemulationsgivesintotal8differentmodelsaccordingtothetablebelow:
Salescodeforoption
FunctionwithHCUandICU_HPIoptionboard Emulationmode
B SpottemperaturePt100 TPU-2/HSU
C VITOtemperatureand/orwaterprobe HPU
J VITOtemperatureand/orwaterprobe+HARTdevice(s) HPU
U SpottemperaturePt100+HARTdevice(s) HSU
V Analogleveloutput MPU
W Analogleveloutput+VITOtemperatureand/orwaterprobe HCU
X Analogleveloutput+VITOtemperatureprobe MPU
Y HCUAnalogleveloutput+SpottemperaturePt100+VITOtemperatureand/orwaterprobe+HARTdevice(s)
Notes:1 Optioncodes:V,WandXarenotavailablewiththe973SmartRadarLT
(analogleveloutputisprovidedbytheICU_HPOoptionboard).2 PleasenotethatwithoptioncodeYinthe973SmartRadarLT,theanalogleveloutputisprovidedby
theICU_HPOoptionboard.3 OptioncodeYisnotavailableinthe854XTGservogaugeand877FDI.4 OptioncodeUisonlyavailableinthe854XTGservogaugeand877FDIwithoutconnectionfor977TSI5 HARTdevicescanbe:- HART pressuretransmittersforHIMS/HTGconfigurationorvapourpressure®
measurement;- HART waterbottomsensorse.g.Sidemountedwaterprobe.®
6 WithsalescodeY(allHCUfunctions),thespottemperaturemeasurementisdisablediftheVITOtemperature(andwater)probeispresent.
1.4 HCUcompatibility
TheHCUboardcanbeusedtoreplaceanexistingoptionalboardinthe854,873and877gauges.However,itrequiressomecheckingifmoreboardsintheinstrumentmustbeupdatedorexternalequipmentmustbereplaced.
Page 8
Introduction
Page8
TheHCUboardcanbeusedtoreplacethefollowingoptionalboards: TPU(-2)board (restrictedtoPt100RTD’sonly) MPUboard (foraveragetemperaturemeasurement,restrictedtoMTTorVITOprobesonly) HPUboard (foraveragetemperaturemeasurement,restrictedtoMTTorVITOprobesonly) HSUboard (fortemperaturemeasurement,restrictedtoPt100RTD’sonly)
Note:ReadcarefullytheinstallationguideoftheHCUboardbeforestartingtoreplacetheexistingoptionboard.
1.4.1 Replacementoraddingoptionalboardin854ATGor854XTGservogauges
TheHCUoptionalboardrequiresanXPU-2boardinthe854ATGand854XTGservogauges.IfthereisnoXPU-2boardinstalled,thentheinstalledXPU(orXPU-1)boardmustbereplacedbyanXPU-2board.
Notes:1 WhentheHCUboardisusedtoreplaceanoptionalboardwithaveragetemperaturemeasurement,
pleasenotethefollowing:The863MRTwith862MIRorthe864MTTwith862MITtemperatureconnectioncannotbemaintained. The864MTTcanbeconnectedtothe762VITOInterfacewhichinterfacestotheHCUboard. The863MRTcannotbeconnectedtothe762VITOInterface.
2 TheoptionaldatatransmissionchannelontheXPU-2(i.s.channelfor977TSIorRS-232C/RS-485channel)isonlypossiblewhenbackplane-2isinstalled.Backplane-2isstandardinstalledininstruments 854ATGwithseriesnumber:854-20-400andhigher; 854XTGwithseriesnumber:894-02-001andhigher.
1.4.2 Replacementoraddingoptionalboardin873SmartRadar
TheHCUoptionalboardcanbeinstalledinthe873SmartRadarwithoutanyproblem.RefertotheNote1insection1.4.1.
1.4.3 Replacementoraddingoptionalboardin877FDI
TheHCUoptionalboardcanbeinstalledinan877FDIwhenthereisanXPUorXPU-1board.However,beinformedthatwaterbottommeasurementisnotpossiblewithVITOprobeandSidemountedwaterprobe.Ifthisfunctionisrequired,thentheinstalledXPU(orXPU-1)boardmustbereplacedbyanXPU-2board.RefertoNote1insection1.4.1.
Note:TheoptionaldatatransmissionchannelontheXPU-2(i.s.channelfor977TSIorRS-232C/RS-485channel)isonlypossiblewhenbackplane-2isinstalled.Backplane-2isstandardinstalledinindicators877FDIwithseriesnumber:877-17-001andhigher;
1.5 Optionalfunctionsinthismanual
Thisinstructionmanualdescribestheoptionalfunctionsfor: spottemperaturemeasurement averagetemperaturemeasurement waterbottommeasurement analogleveloutput
TheoptionalfunctionsforHIMS,HTGandvapourpressuremeasurementaredescribedintheinstructionmanual:HIMS/HTGandvapourpressure(P3)measurement.
Page 9
Spottemperaturemeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page9
2 Spottemperaturemeasurement
2.1 Introduction
AstandardPt100spotelement(RTD)canbemeasuredanditstemperatureisinterpretbytheHCUandICU_HPIboardasproducttemperature.
Thetemperaturemeasurementisintrinsicallysafe,certifiedEExiaIIB.
PrincipleofoperationThetemperaturemeasurementisbasedona3-wireRTDconfiguration.OnlytheICU_HPIboardhastheoptionfora4-wireRTDconfiguration.
Theresistancemeasurementfor3-wireconfigurationisperformedinthefollowingsteps:
Thespotelementismeasured,includingthecableresistance Thecableresistanceismeasuredseparately,andsubtractedfromtheelementresistancetoavoid
influenceofthecablingonthetemperaturemeasurement Next,thesystemconstantisdetermined.Thisisdonebymeasuringanaccuratereferenceresistance
(R )±0.01%+2ppm/°C(-40÷85°C)ref InordertochecktheR ,atestresistor(R )ismeasured.Thisresistorhasthesameaccuracyastheref test
referenceresistor
Thesefourmeasurementsareperformedwithandwithoutcurrenttocompensateforanythermooroffsetvoltageinthemeasuringcircuit.
Thetemperaturedataisdisplayed(whenadisplayispresent)asproducttemperatureandtransmittedtothehost.
Page 10
Spottemperaturemeasurement
Page10
2.2 Commissioningofspottemperaturemeasurement
ThecommissioningoftheoptionalspottemperaturepartoftheHCUorICU_HPIboardshouldbeperformedafterthebasiccommissioningofthelevelgaugeorindicator.
2.2.1 Spottemperaturesettings
Item Name Description
W2= Protectionlevel2 Enterprotectionlevel2.
TD= Temperaturedimension Onecharacter;either‘C’or‘F’.DefaultsettoC.C : degreesCelsiusF : degreesFahrenheit
Bothdimensionshavethesameformat.
MN= Elementconfiguration Twodigits(default:16)whichmustbeseteitherto:03 for3wireRTDconfigurationor04 for4wireRTDconfiguration(ICU_HPIboardonly).
Note:Ifa4wireRTDconfigurationisused,strapST4ontheICU_HPIboardmustbeplacedtothecorrectposition(refertofig.2.1).
MO= RTDposition FormataccordingtoitemLD.MOrepresentsheightoftheRTDelementwithrespecttothezeropointofthelevelgauge.Ifthemeasuredleveldropsbelowthisheightthefollowingtwoactionstakesplace:1) Thecorrespondingstatusbit(bit3)issetinstatusbyte1
ofitemAP(temperaturerequest);2) Thetemperaturestatusontheinstrumentdisplay(ifapplicable)
issetto‘LR’.
MI= Switchhysteresis FormataccordingtoitemLD.Default:+000.1000(m).ItemMIisaswitchhysteresisaroundtheRTDpositionitemMO.
TU= Temperaturestatusconversion Onecharacter;default‘T’.Thischaracteristransmittedinthetemperaturestatusbytetothehostincaseof:- levelbelowRTDposition(itemMO)- invalidlevelreadingIfrequired,thisitemcanbesettoanothercharacter.
EX Exit Exitprotectionlevel.
Page 11
Spottemperaturemeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page11
Figure2.1 Strapfor3-or4-wireconfigurationonICU_HPIboard
2.2.2 Additionaltemperaturesettingsfor877FDI
Item Name Description
W2= Protectionlevel2 Enterprotectionlevel2.
OB= Optionalboard*) ThreeASCIIcharacters.ThisiteminformstheXPU(andXPU-1)whatoptionboardisinstalled.Ithastobesetto‘HPU’.
EG= Enabletemperature OneASCIIcharacter;either‘D’or‘E’.Withthisitemcanbetransmission*) selectedifthe877FDIshallansweronatemperaturerequest
(C-record),addressedtotheconnectedlevelgauge.E : transmissionoftemperaturerecordenabledD : transmissionoftemperaturerecorddisabled
Note:WhenitemEGissetto‘E’,makesuretherelatedlevelgaugehasnotemperatureoptioninstalled.
TF= Temperaturesourceselection OneASCIIcharacter;either‘I’or‘E’.Thisitemselectsifthereisatemperaturemeasurementinsidetheindicator(Internal),orthetemperaturehastobefetchedfromtheEnraffieldbus(External).
I : Internal(HCUoptionalboardinstalled)E : External
IM= Indicatormode OneASCIIcharacter.WithXPU-2,the877FDIcanbesetinstandalonetemperaturegaugemode.ThenitemIMhastobesetto‘T’.
EX Exit Exitprotectionlevel.
*) ThisitemisnotimplementedintheXPU-2.
Page 12
Spottemperaturemeasurement
Page12
2.3 Operation
2.3.1 Display
Foroperationofthedisplayandtheinformationonit(ifapplicable),refertotheinstructionmanualoftherelatedlevelgaugeorindicator.Below,onlyanoverviewisgivenwhichdisplayformatsgivesinformationaboutthespottemperaturemeasurement.
Displayformat Displayedinformation
AD
ProductlevelandtemperatureProducttemperatureandstatus
2.3.2 Dataitems
Thetablebelowlistsanumberofdataitems.Theycontainmeasureddata,verificationdataanderrordata.Theverificationdatacanbeusedtochecktheresultsofcertainstepsinthemeasuringsequence.
Thetemperaturestatusindicatesthevalidityofthemeasureddata.Theerrordataprovideslowlevelerrorinformationaboutthetemperaturemeasurement(refertosection2.4).
Item Description
AP Measuredtemperature(dimensionaccordingtoitemTD);
MU Measuredtestresistance(floatingpointformat);
MQ
MeasureddatathevalueisprecededbyfourcharactersfromthetemperaturestatusitemMQ.
mustbe:166.5Ω±0.03%.Temperaturestatusrequest(refertosection2.4.2).
VV Verificationdata Refertodescriptionatsection2.4.3.
EMFHH0
Diagnosticdata ErrorHCU/ICU_HPIrequest(refertosection2.4.1).FatalHCU/ICU_HPIerrorsLastfatalHCU/ICU_HPIerror
Page 13
Spottemperaturemeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page13
2.4 Troubleshooting
2.4.1 Temperatureerrorrequest(itemEM)
ThisitemcontainsthemostrecenttemperatureerrorencounteredbytheoptionalHCU/ICU_HPIboard.
xx00 Noerror,valueatinitialisationxx10 Countererror Hardwareerror.Resetinstrument;iferrorpersist,replaceHCUorICU_HPIboardxx15 UTItime-outxx29 R_cableerror ResistanceofB-andC-wiretoohigh(loopresistance:≤24Ω)xx30 Connectionerror CheckthewiringconnectionsofRTDtoinstrumentxx34 R error Hardwareerror.Resetinstrument;iferrorpersist,replaceHCUorICU_HPIboardtest
xx70 Wrongelementconfigurationspecified CheckitemMN(mustbesetto‘03’or‘04’)xx98 Notemperaturesourcesfound NoRTD(andnoVITOtemperatureprobe)connectedxx99 Spotoptionprintnotinstalled ChangeHCUorICU_HPIboardforthecorrecttype
xx: 28forHSU/TPU-2emulation30forHCUemulation
2.4.2 Temperaturestatusrequest(itemMQ)
Thetemperaturestatusrequestitemcontainsfourstatusbytes(Byte0,Byte1,Byte2,Byte3)fromtheoptionalHCU/ICU_HPIboard.Fordecoding,refertotheASCIItableinappendixA.
Statusbyte0:always‘0‘
Statusbyte1: Statusbyte2:bit 0: GeneralHCU/HPItemperaturefail bit 0: Lastvalidlevelused 1: Failinspottemperaturereading 1: Manuallevelused 2: 1 2: Leveltime-out 3: ProductlevelbelowRTDposition 3: 0 4: 0 4: 0 5: 0 5: Temperatureoutofrange
6: 1 6: 17: 0 7: 0
Statusbyte3:bit 0: Nopreviousstorecommand1-5: 0
6: 17: 0
Note:Onlythebitswhicharesetto‘1’haveanactivestatus.
Page 14
Spottemperaturemeasurement
Page14
2.4.3 Temperaturepointervalue(itemVV)
Bymeansofthevaluepointer(itemVP)avectorcanbeloadedtotheHCUorICU_HPIoptionboard.Next,withitemVV,theselecteddataisreturned.ItemVPconsistsof4positions,inthemiddleseparatedbya‘.’or‘,’:vw.xy(orvw,xy).Thevaluesforthevaluepointerarelistedinthetabletogetherwiththeobtaineddata.
vw , xy Selecteddata Example/Dimension
00 , 00 HCU/ICU_HPIEmulation&Function VV=HCUHCAOST--PR
00 , 01 Configurationbootcode VV=HCUCONFIG:3F
00 , 02 HCU/ICU_HPIhardwareversion VV=HWVERSION:0000 , 03 Bootcodesoftwareversion VV=BOOTSWVERS:01
01 , 00 Measuredspotresistance (floatingpointformat)01 , 01 Calculatedspottemperature (floatingpointformat)01 , 02 Measuredcableresistance (floatingpointformat)01 , 50 Temperatureupdatecounter
Emulation: MPU,HSU,HPU,HCUFunction: HC: HARTchannelinstalled
AO: analogoutputST: spottemperatureMT: VITOaveragetemperatureWS: externalwaterbottomprobeWT: VITOwaterbottomprobePR: pressuretransmitters
Salescodeoption: B bootcode: 01U 25Y 3F
ΩaccordingtoitemTD
Ω
Item Name Description
VP= Temperaturevaluepointer Temperaturevaluepointer;format:2digits,separator,2digits (refertotableabove).
Example: VP=01.00:valuepointerloadedtorequestthemeasuredspotresistance.
VV Temperaturepointervalue Temperaturepointervalue.ThisitemholdsthevaluerequestedbyitemVP(refertotableabove).Example: therequestedmeasuredspotresistance
(standardfloatingpointformat).
Page 15
Averagetemperaturemeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page15
Figure3.1 VITOtemperatureprobes&MTT
3 Averagetemperaturemeasurement
3.1 Introduction
AnEnrafaveragetemperatureprobecanbe:
• a764VITOtemperatureprobe;• a766VITOwaterandtemperatureprobe;• a864MTT.
Dependingonthetypeof764VITOtemperatureprobe,thereare15or16spotelements(thermocouples)equallyspacedoverthemeasuringrange.
Thetemperaturemeasurementisbasedonthermocouples(spots)andonePt100referenceresistor.
Themeasuringrangeisdefinedas:distancefromreferenceresistortopositionhighestelement.
Refertofigure3.1.
• A Thismodelhasthereferenceresistorlocatedclosetothelowerendoftheprobe.Thereare16spots,equallydividedoverthemeasuringrange;
• B Here,thereferenceresistorisplacedat1m(3ft3)fromthelowerendoftheprobe.Onespotislocatedclosetothelowerendoftheprobetomeasuretemperatureatlowlevels.Withthismodel,thereare15spots,equallydividedoverthemeasuringrange;
• C Thisisa766VITOtemperatureandwaterprobe.Fortemperaturemeasurement,thismodelisequaltoB;
• D Thismodelof766VITOtemperatureandwatersensorhas16spots,equallydividedoverthemeasuringrange.Hence,thereisnospotatthelowerendofthewaterprobe.
UsingtheEnrafaveragetemperatureprobes(764VITOtemperatureprobe,766VITOwaterandtemperatureprobeorthe864MTT)a762VITOInterfaceisrequired.This762VITOInterfaceispoweredbythelevelgauge(or877FDI)overthesametwowiresastheHART communicationtakesplacetothelevelgauge.®
Inthelevelgauge,or877FDI,thetemperatureiscalculated.Thetemperatureofallspotsimmersedintheliquid,areaveragedforproducttemperaturecalculation.Allspotsabovetheliquidsurfaceareusedtocalculatetheaveragegastemperature.
Withthetemperature(andwater)probemodels,whichhaveaspotbelowthereferenceresistor,thatspottemperatureisnotusedintheaverageproducttemperaturecalculations.Itstemperatureisonlyusedwhentheproductleveldropsbelow1m(locationofreferenceresistor)toprovidetheuserwithatemperaturevalue.However,withstatus:‘levelexceedslowestspotelement’;itemMQ,bit3ofbyte1isset).
Page 16
Averagetemperaturemeasurement
Page16
3.2 Commissioningofaveragetemperaturemeasurement
Thewiringsequenceofallbluethermocouplewiresisonlymeasuredonceatstart-upandtheinformationisstoredintheHCU/ICU_HPImemory.Thetimerequiredtocompleteall15thermocouplemeasurementsafterpoweringuptheinstrumentisapproximately2minutes.
ThecommissioningoftheoptionalaveragetemperaturepartoftheHCUorICU_HPIboardshouldbeperformedafterthebasiccommissioningofthelevelgaugeorindicator.
3.2.1 Averagetemperaturesettings
Beforeprogrammingthetemperaturedata,thesensitivelengthandtheoffsetofthetemperatureprobemustbeknown.
Thesensitivelengthisequaltothemeasuringrange.Standard,themeasuringrangesare:
for764VITOtemperatureprobe(modelcode:....764B<ol><he>) :overalllength-700mm; for864MTT :overalllength-700mm; forotherVITOtemperature(andwater)probe :overalllength-1700mm.
Other(smaller)measuringranges(sensitivelengths)arepossible;checkonidentificationlabeloftheVITOprobe(orMTT)forthepositionofthehighestelement(last4digits;units:cm).
Theoffsetisthedistancefromthelowestelementtothetank’szeropoint(datumplate).The764VITOtemperaturesensorand766waterandtemperaturesensorcanhavethelowestelementondifferentplaces.Thisisidentifiedontheinstrumentlabel:
....764A<ol><he> lowestelementat0.065mfromlowerendofprobe; ....764B<ol><he> lowestelementat0.065mfromlowerendofprobe; .A..766B<ol><he> lowestelementat0.065mfromlowerendofprobe; .A..766A<ol><he> lowestelementat1mfromlowerendofprobe.
Withthe864MTT,thelowestelementisat0.065mfromlowerendofprobe.
Note:Inthemodelcodegivenaboveadot‘.’representsanycharacter,thenotation<ol>means:overalllength(4positions)andthenotation<he>means:positionofhighestelement(4positions).Thenotation<sl>means:sensitivelength(4positions)andisonlyusedwith864MTT.
ThepositionofthelowestelementcanalsobeverifiedbyrequestingtheVITOdevicetype&s/wversionbymeansofitemsVPandVV,providedyouknowifa764VITOsensorisconnected,ora864MTT.
Refertothetablebelowandtosection3.4.3howtoobtaindatawiththetemperaturepointeritems:
ItemVP ItemVV sensormodelcode
VP=03.10 VV:764PTTC10 ....764A<ol><he>
VP=03.10 VV:766PTTCW10 .A..766B<ol><he>
VP=03.10 VV:864PTTC10 ....764B<ol><he>VV:864PTTCW10 .A..766A<ol><he>VV:864PTTC10 ....864.<ol><sl>
Page 17
Averagetemperaturemeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page17
Figure3.2 Tankandtemperatureprobedata
Page 18
Averagetemperaturemeasurement
Page18
Refertofigure3.2.
Item Name Description
W2= Protectionlevel2 Enterprotectionlevel2.
TD= Temperaturedimension Onecharacter;either‘C’or‘F’.DefaultsettoC.C: degreesCelsiusF: degreesFahrenheit
Bothdimensionshavethesameformat.
MK= Sensitivelengthtemperature FormataccordingtoitemLD(default:+030.0000).probe Refertothedescriptioninsection3.2.1aboutthesensitivelength
(measuringlength)ofthetemperatureprobe.EnterthatvalueinitemMK.
MO= Temperatureelementoffset FormataccordingtoitemLD(default:+000.0000).ItemMOrepresentsthedistancefromtankzero(datumplate)tothelowestelement(spot)inthesensor.Thelowestelementinthesensordependsonthetypeoftemperaturesensor.Refertotheinstrumentlabelandfigure3.2.ThelocationofthelowestelementcanalsobefoundbyrequestingitemVV(withVP=03.10).Refertosection3.4.3.
MI= Switchhysteresis FormataccordingtoitemLD(default:+000.1000m).ThedistanceMIisahysteresisaroundtheswitchingpointofthetemperatureelements.
MP= Productimmersiondepth FormataccordingtoitemLD(default:+000.5000m).ThedistanceMPspecifiestheminimumrequiredliquidlevelaboveathermocouplebeforeitistakenintheaverageproducttemperaturecalculation.
MG= Gasimmersiondepth FormataccordingtoitemLD(default:+000.5000mfor854/873/973and+000.1000mfor877).ThedistanceMGspecifiestheminimumrequireddistancebelowathermocouplebeforeitistakenintheaveragegastemperaturecalculation.
TU= Temperaturestatusconversion Onecharacter(default:‘T’).Thisitemisusedtoindicatethetemperaturestatuscharacterincaseofareducedornotguaranteedtemperatureaccuracy.Thisisthecasewhenthetemperaturestatusis:
- outofspecifiedtemperaturerange- exceedingdifferentialtemperaturerange- lastvalidlevelused- manuallevelused- levelbelowlowesttemperatureelement
Ifrequired,thisitemcanbesettoanothercharacter.
EX Exit Exitprotectionlevel2.
Page 19
Averagetemperaturemeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page19
Withthe764VITOtemperatureprobesandalsowithan864MTT,itmightberequiredtodisablethelowestspotelementfromtheaveragetemperaturecalculation.ThatcanbedonebysettingthefirstpositioninitemMWto‘F’.Infact,withitemMWanyspottemperatureelement(ormorespots)canbedisabledfromaveragetemperaturecalculation.
Item Name Description
W2= Protectionlevel2 Enterprotectionlevel2
MW= MTTwiringconnection Sixteenhexa-decimalcharacters;default:0000000000000000.Whenone(ormore)oftheelementsmustbedisabledfromaveragetemperaturecalculation,setthedesiredpositionto‘F’.Forinstance,todisablelowestspotelement:
MW=F000000000000000Note:
WhenenteringitemMWwiththePortableEnrafTerminal,thedisplaywillbeclearedafterenteringthe15 character.th
Continuewithenteringthelastcharacterandpressreturn.All16enteredcharacterswillberepeatedatthetoplineofthePET.
EX Exit Exitprotectionlevel2
3.2.2 Additionaltemperaturesettingsfor877FDI
Item Name Description
W2= Protectionlevel2 Enterprotectionlevel2
OB= Optionboard*) ThreeASCIIcharacters.ThisiteminformstheXPUwhatoptionboardisinstalled.Thatcanbe:
MPU : forHCUoptionboardinMPUemulationmodeHPU : forHCUoptionboardinHPUemulationmode
EG= Enabletemperature OneASCIIcharacter;either‘D’or‘E’.Withthisitemcanbetransmission*) selectedifthe877FDIshallansweronatemperaturerequest
(C-record),addressedtotheconnectedlevelgauge.E : transmissionoftemperaturerecordenabledD : transmissionoftemperaturerecorddisabled
Note:WhenitemEGissetto‘E’,makesuretherelatedlevelgaugehasnotemperatureoptioninstalled.
TF= Temperaturesourceselection OneASCIIcharacter;either‘I’or‘E’.Thisitemselectsifthereisatemperaturemeasurementinsidetheindicator(Internal)ortobefetchedfromtheEnraffieldbus(External).
I : Internal(HCUoptionboardinstalled)E : External
EX Exit Exitprotectionlevel2.
*)ThisitemisnotimplementedintheXPU-2.
Page 20
Averagetemperaturemeasurement
Page20
3.3 Operation
3.3.1 Display
Foroperationofthedisplayandtheinformationonit(ifapplicable),refertotheinstructionmanualoftherelatedlevelgaugeorindicator.Below,onlyanoverviewisgivenwhichdisplayformatsgivesinformationabouttheaveragetemperaturemeasurement.
Displayformat Displayedinformation
ACD
ProductlevelandaverageproducttemperatureAveragegastemperatureandstatusAverageproducttemperatureandstatus
3.3.2 Dataitems
Thetablebelowlistsanumberofdataitems.Theycontainmeasureddata,verificationdataanderrordata.Theverificationdatacanbeusedtochecktheresultsofcertainstepsinthemeasuringsequence.Thetemperaturestatusindicatesthevalidityofthemeasureddata.Theerrordataprovideslowlevelerrorinformationaboutthetemperaturemeasurement(refertosection3.4.).
Item Description
AGAPMQMU
U0-UFV0-VF
Measured/ Averagegastemperaturecalculateddata Averageproducttemperature
Temperaturestatusrequest(refertosection3.4.2)Measuredtestresistance(floatingpointformat);mustbe:166.5Ω±0.03%Calculatedspotpositions(seebelow)Spottemperatures(seebelow)
VVRW
Verificationdata Refertodescriptionatsection3.4.3Thermocouplewiringconnectionsequence
EMFHF0
Diagnosticdata ErrorHCU/ICU_HPIrequest(refertosection3.4.1)FatalHCU/ICU_HPIerrorsLastfatalHCU/ICU_HPIerror
ItemV0isthetemperatureofthelowestspotelement.ThecalculatedspotpositionitemU0correspondswithitemV0.Refertofigure3.2tolocatethelowestspot(L)witheachparticulartemperatureprobemodel.
Page 21
Averagetemperaturemeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page21
3.4 Troubleshooting
3.4.1 Temperatureerrorrequest(itemEM)
ThisitemcontainsthemostrecenttemperatureerrorencounteredbytheHCU/ICU_HPIboard.
xx00 Noerror,valueatinitializationxx11 NoreplyoninitialHARTcommands Communicationerror.Checkconnectionlevelgauge-762VITOInterface,orchange762
VITOInterfaceorHCU/ICU_HPIboardxx50 Pt100(orspot0)error BrownwireconnectedtoCN1A-1(ororangewireconnectedtoCN2B-1)notconnectedxx51 Spot1(orPt100)error BluewiretoCN2B-1(orbrownwiretoCN1A-1)notconnectedxx52 Spot2error BluewiretoCN2A-2notconnected: : :xx64 Spot14error BluewiretoCN5A-2notconnectedxx65 Spot15error YellowwiretoCN5A-1notconnectedxx70 Sub-systemerror Datafrom762VITOnotvalid.Resetinstrument;iferrorpersist,replace762VITOInterfacexx71 Wrongprobeconnection ThermocouplewiresofVITOsensorconnectedwrongly(blueandyellowwires)xx77 WrongPt100connection BrownandredwiresofVITOsensorconnectedwronglyxx80 R error Measurementontestresistorfailed.Resetinstrument;iferrorpersist,replace762VITOtest
xx84 Standarddeviationtoolarge EMCinfluence(checkcableshields)ordefective762VITOInterfacexx92 Nodataavailable Communicationerrorbetween762VITOInterfaceandlevelgauge.Checkwiring;replace
762VITOInterfaceorHCU/ICU_HPIboardxx98 Notemperaturesourcesfound NoVITOtemperatureprobe(andnoRTD)connected
xx: 22forMPUemulation Witherrorcodesxx50toxx65,connectornumbersarementioned.24forHPUemulation RefertoInstallationguide762VITOInterface&764,765or766VITO30forHCUemulation temperatureand/orwaterbottomprobeforlocationoftheseconnectors.
3.4.2 Temperaturestatusrequest(itemMQ)
Thetemperaturestatusrequestitemcontainsfourstatusbytes(Byte0,Byte1,Byte2,Byte3)fromtheoptionalHCU/ICU_HPIboard.Fordecoding,refertotheASCIItableinappendixA.
Statusbyte0:Containsthecharacters‘0’÷‘F’,indicatingthehighestimmersedspotelementofthetemperatureprobe.Atstartup:’I‘.
Statusbyte1: Statusbyte2:bit 0: Generaltemperaturefail bit 0: Lastvalidlevelused
1: Failinaverageproducttemperature 1: Manuallevelused2: Failinaveragegastemperature 2: Leveltimeout3: Levelexceedslowestspotelement 3: Devicenotcalibrated4: Levelexceedshighestspotelement 4: Exceedingdifferentialtemp.range5: Spotelementfail(oneormorespotsdefect) 5: Outofspecifiedtemperaturerange6: 1 6: 17: 0 7: 0
Statusbyte3:bit 0: Nopreviousstorecommand1-5: 0
6: 17: 0
Note:Onlythebitswhicharesetto‘1’haveanactivestatus.
3.4.3 Temperaturepointer(itemsVPandVV)
Bymeansofthevaluepointer(itemVP)avectorcanbeloadedtotheHCUorICU_HPIoptionboard.Next,withitemVV,theselecteddataisreturned.ItemVPconsistsof4positions,inthemiddleseparatedbya‘.’or‘,’:vw.xy(orvw,xy).Thevaluesforthevaluepointerarelistedinthetabletogetherwiththeobtaineddata.
Page 22
Averagetemperaturemeasurement
Page22
vw , xy Selecteddata Example/Dimension
00 , 00 HCU/ICU_HPIEmulation&Function VV=HCUHCAOMT--PR
00 , 01 Configurationbootcode VV=HCUCONFIG:3F
00 , 02 HCU/ICU_HPIhardwareversion VV=HWVERSION:0000 , 03 Bootcodesoftwareversion VV=BOOTSWVERS:01
02 , 00 CalculatedtemperatureofspotconnectedtoCN1A-1 (floatingpointformat)02 , 01 CalculatedtemperatureofspotconnectedtoCN2B-1 (floatingpointformat): : :02 , 15 CalculatedtemperatureofspotconnectedtoCN5A-1 (floatingpointformat)02 , 20 MeasuredPt100resistance (floatingpointformat)02 , 25 Standarddeviation (floatingpointformat)02 , 26 V (floatingpointformat)02 , 27 LeakcurrentI (floatingpointformat)02 , 28 LeakcurrentI (floatingpointformat)
03 , 01 ErrorcountersHARTaddresses3,4and5
03 , 09 DetectedHARTdeviceaddresses (5forVITOInterface)03 , 10 InitialdetectedVITOdevicetype&s/wversion
03 , 11 RealactivefunctionsofVITOInterface(refertodescriptionofVP=03,10)03 , 12 ActualexecutedHARTcommandsfrom762VITOInterface03 , 20 Averagingconstant762VITOInterface (floatingpointformat)
Emulation: MPU,HSU,HPU,HCUFunction: HC: HARTchannelinstalled
AO: analogoutputST: spottemperatureMT: VITOaveragetemperatureWS: externalwaterbottomprobeWT: VITOwaterbottomprobePR: pressuretransmitters
Salescodeoption: C bootcode: 0AJ 2EW 1AX 12Y 3F
test
1
2
(762VITOInterfacehasHARTaddress5)
Description: 764: 764VITOtemperatureprobewithPt100referenceresistorat1mandspotclosetoprobe’send
766: 766VITOwaterandtemperatureprobewithspotclosetoprobe’send
864: 764VITOtemperatureprobewithPt100referenceresistorandlowestspotclosetotheprobe’send.766VITOwaterandtemperatureprobewithPt100referenceresistorandlowestspotat1m864MTT.
PT: Pt100TC: thermocouplesW: waterprobe10: softwareversion1.0
accordingtoitemTDaccordingtoitemTD
accordingtoitemTDΩ
VAA
VV=0000:0000:0013
VV=----5---------VV=766PTTCW10model:....764A<ol><he>
model:.A..766B<ol><he>
model:....764B<ol><he>
model:.A..766A<ol><he>
model:....864.<ol><sl>
VV=766PTTC
Item Name Description
VP= Temperaturevaluepointer Temperaturevaluepointer;format:2digits,separator,2digits (refertotableabove).
Example: VP=02.20:valuepointerloadedtorequestthePt100elementresistance.
VV Temperaturepointervalue Temperaturepointervalue.ThisitemholdsthevaluerequestedbyitemVP(refertotableabove).Example: therequestedPt100elementresistance
(standardfloatingpointformat).
Page 23
Waterbottommeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page23
Figure4.1 VITOwaterprobes,Sidemountedwaterprobe&964WaterScout
4 Waterbottommeasurement
4.1 Introduction
Thewaterbottomprobeisbasedonacapacitivemeasurementprinciple.Theprobeisoneplateofthecapacitorandearthed,electricallyconductingpartsaswaterandthetankshell,istheotherplate.Achangeinwaterlevelisdetectedbythewaterbottomprobeasachangeincapacitance.
4.1.1 Waterprobeversions
Therearethreedifferentwaterprobeversionswhichcanbeinterfaced:• the765VITOwaterprobe;• the766VITOwaterandtemperatureprobe;• theexternalHART waterbottomprobe(Sidemountedwaterprobeor964WaterScout).®
The765VITOwaterprobeand766VITOwaterandtemperatureprobeareconnectedtothe762VITOInterface,whichisconnectedtotheHCUboardorICU_HPIboard.TheexternalwaterbottomprobeisconnectedtoHARTinput2(sameinputwherepressuretransmittersareconnected).
Refertofigure4.1.A) 765VITOwaterprobe
Canbedeliveredinthreesensitivelengthsforthewatermeasurement:0.5m,1.0mand2.0m.B) 766VITOwaterandtemperatureprobe
Thereisonesensitivelengthforthewatermeasurement:0.5m.C) 964WaterScout(previousmodel)
WiththeintroductionoftheVITOfamily,the964WaterScouthasbecomeobsolete.Itcan,however,beconnectedtotheHCU/ICU_HPIoptionboard
D) Sidemountedwaterprobe
Page 24
Water level PV PV LRVPV URV PV LRV
× sensitive length probe water probe bottom position
LW MX VRVT VR
× WP WB
Waterbottommeasurement
Page24
4.1.2 Principleofmeasurement
Whencorrectlycalibrated,thecapacitancevaluegiveninPVLRV(PrimaryVariableLowerRangeValue)isthecapacitancemeasuredbytheprobewhennowaterispresent(sensitivepartofprobecompletelyimmersedinproduct).ThecapacitancevaluegiveninPVURV(PrimaryVariableUpperRangeValue)isthecapacitancemeasuredbytheprobewhenthesensitivepartofprobecompletelyimmersedinwater.
Note:The765and766VITOwaterprobesusesitemsVRandVTtostoretheminimumandmaximumcapacitancevalues.
Themeasuredcapacitancebythewaterprobehasalinearfunctionwiththewaterlevel.Hence,thewaterleveliscalculatedas:
where: PV =themeasuredcapacitancebythewaterprobe;withVITOwatersensor,itemMXisusedPVLRV =theminimumcapacitance(0%water);withVITOwaterprobe,itemVRisusedPVURV =themaximumcapacitance(100%water);withVITOwaterprobe,itemVTisused
or,
where: LW =waterlevelWP =waterprobelengthWB =waterprobebottomposition
4.2 Commissioningmeasurements
Thepurposeofthesecalibrationmeasurementsistodeterminethewaterprobeoffsetandacontrolmeasurementforthewaterlevel.
Withthesemeasurementstheproductlevelisusedasahorizontallinebetweenthepointwherethetankzeropointislocatedandthepointwherethewaterprobeisinstalled.
Page 25
Waterbottommeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page25
Figure4.2 Calibrationmeasurementsforwaterprobe
4.2.1 Commissioningmeasurementswithroofsuspendedprobes
Ifthecommissioningtakesplacedirectlyafterinstallation,themeasurementsattheprobepositioninfigure4.2havealreadybeenmadeiftheinstructionsarefollowedfromtheinstallationguide.Otherwise,thatmeasurementshouldbemadenow,bypartlyremovingtheprobetoenableameasuringtapetobeloweredfromthenozzle.Thefollowingvaluesareobtained:
• Nozzleheight• Productlevelatwaterprobelocation• Waterlevelatwaterprobelocation
Fromthesemeasurements,calculatethe‘ullage’value‘A’:
A=Nozzleheight-productlevelatwaterprobelocation
Thenextstepistofetchtheproductlevelwithrespecttothetankzeropoint.Thiscanbedoneinseveralways:• Readthelevelfromthelevelgaugeifthathasrecentlybeenchecked(calibrated);• Takeamanual‘innage’measurementtothedatumplateofthetank;• Takeamanual‘ullage’measurementfromtheUpperReferencePointandcalculatethelevelby:level
=URP-‘ullage’.
Refertofigure4.2.Atthebottompartoftheprobethereisaninactivemetalpartwithaheightof25mm(1").Thatisrepresentedasdistance‘B’.Theoffset,whichhastobeprogrammedinitemWB,canbefoundfrom:
WB=(A+productlevelatreferencepoint)-(Nozzleheight-B)Note:
IfWBisanegativefigurethenitispossibletohaveanegativewaterlevelreading.
Calibratingthewaterprobe
RefertoAppendixCforinformationaboutcalibratingthewaterprobe.Whereismentionedthewaterdip,refertothemeasuredwaterlevelattheprobepositionasdescribedbefore.
Page 26
Waterbottommeasurement
Page26
Figure4.3 Calibrationmeasurementsforsidemountedprobe
4.2.2 Commissioningmeasurementsforsidemountedprobes
Fromtheprobeinstallationmustbeknown: Activelengthofprobe Totallengthofprobe Angleofinstallation(α) Height‘H’ofnozzle
Thewaterprobelength(WP)canbecalculatedas:
WP=activelengthofprobexcos(α)
Offset(WB)canbecalculatedas:
WB=H-(totallengthxcos(α))
Whenthewaterbottommeasurementhasnorelationtothetankzero,theoffsetWBcanbeusedaswaterprobebottomposition.
Ifthewaterbottommeasurementisrelatedtothetankzero,thenthewaterprobebottomposition(WB)canbedeterminedwhentheprobeisimmersedinsomewater.
Thereadingoftheprobeistaken;itemLW.Amanualwaterdipistakenfromthetankzeropoint(datumplate);notethisvalueas‘W’.CalculatethewaterprobebottompositionWBas:
WB=W-LW
Calibratingthewaterprobe
RefertoAppendixDforinformationaboutcalibratingthewaterprobe.Whereismentionedthewaterdip,refertothemeasuredwaterlevelattheprobepositionasdescribedabove.
Page 27
Waterbottommeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page27
4.2.3 Waterprobesettings
Item Name Description
W2= Protectionlevel2 Enterprotectionlevel2
WB= Waterprobebottomposition FormataccordingtoitemLD(default:+000.0000m).Thisitemgivestheoffsettothewaterprobezeropointinrelationwiththetankzeropoint.Refertosections4.2.1and4.2.2howtoobtainthisoffsetvalue.
WP= Waterprobelength FormataccordingtoitemLD(default:+001.0000m).Thewaterprobelength(togetherwiththewaterprobebottomposition)isrequiredtocalculatethewaterlevel.
WF= Waterlevelfieldcommunication Onecharacter:‘S’or‘D’.Onlyvalidwith854levelgauges.Thisitemselectsfromwhichsourcethewaterlevelismeasuredandtransmittedinthelevelrecord(B-record)tothesystem.
S : fromthewaterprobe(VITOwatersensor,WaterScoutorsidemountedwaterprobe)
D : fromwaterdip(displacer)Select‘S’for854thegauges(default:D).
WH= Waterhighalarm FormataccordingtoitemLD(default:+000.0000m).Withthisitemawaterhighalarmsetpointcanbegiven.Whenactive,thewaterhighalarmconditionissignalledontheinstrumentdisplayindisplayformatK(ifdisplayisavailable).Thewateralarminformationisalsotransmittedinthewaterlevelrecord(M-record).
WL= Waterlowalarm FormataccordingtoitemLD(default:+000.0000m).Withthisitemawaterlowalarmsetpointcanbegiven.Whenactive,thewaterlowalarmconditionissignalledontheinstrumentdisplayindisplayformatK(ifdisplayisavailable).Thewateralarminformationisalsotransmittedinthewaterlevelrecord(M-record).
WS= Wateralarmhysteresis FormataccordingtoitemLD(default:+000.0100m).Thewateralarmhysteresispreventsthewateralarmtobeswitchedonandoffwhenthewaterlevelisaroundthewaterhighorwaterlowalarmsetpoint.
EX Exit Exitprotectionlevel.
Waterprobelength:The766VITOwaterandtemperatureprobehasonelengthforthewatermeasurement;theotherwaterprobescanhavedifferentlengths.Theactualwaterprobelengthcanbefoundfromtheidentificationcodeoftheprobe.Thetablebelowgivesanoverview:
765VITO 766VITOwaterand 964WaterScout Sidemountedwaterprobewaterprobe temperatureprobe
0.485m 0.485m 0.5m0.985m Notrequiredto 1.0m1.985m 2.0mspecifyinitemWP
activelengthxcos(α)(refertosection4.2.2)
Page 28
Waterbottommeasurement
Page28
4.3 Operation
4.3.1 Display
Foroperationofthedisplayandtheinformationonit,refertotheinstructionmanualoftheapplicablelevelgaugeor877FDI.Below,onlyanoverviewisgivenwhichdisplayformatgivesinformationaboutthewaterbottommeasurement.
Displayformat Displayedinformation
K Waterlevelandstatus
4.3.2 Waterabove/belowprobewarning
Automaticallyawarningisgivenwhenthewaterlevelistoolowortoohighfortheprobetomeasurecorrectly.
• Thewateraboveprobestatus(AP)isdisplayedandsignalledinthewaterlevelrecordwhenthewaterleveliswithin1%oftheprobelength(itemWP)fromthetopofthewaterprobe.
• Thewaterbelowprobestatus(BP)isdisplayedandsignalledinthewaterlevelrecordwhenthewaterleveliswithin1%oftheprobelength(itemWP)fromthebottomofthewaterprobe.
Aroundthesewarningsthereisahysteresisof½%oftheprobelength(itemWP).
4.3.3 Dataitems
Below,asummaryisgivenoftheavailabledataitemsanderrorcodes.Thereisoneoperationalcommandwiththewaterbottommeasurement.
Item Description
LWLHN4PVMX
Measured/ Waterlevel,precededbytwostatusbytes(refertosection4.4.3)calculateddata Waterullage,precededbytwostatusbytes(refertosection4.4.3)
SerialnumberexternalwaterprobeMeasuredcapacitancefromsidemountedwaterprobeMeasuredcapacitancefrom765/766VITOwaterprobe
VV Verificationdata Refertodescriptionatsection4.4.4
EWHEFHH0
Diagnosticdata Errorwaterrequest(refertosection4.4.1)HARTcommunicationerrors(refertosection4.4.2)FatalHCU/ICU_HPIerrorsLastfatalHCU/ICU_HPIerror
Operationalcommand:
SR StopHARTrequest(usedwithHARTcommunicator).
Page 29
Waterbottommeasurement
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page29
4.4 Troubleshooting
4.4.1 Watererrorrequest(itemEW)
DataitemEWcontainsthemostrecenterrorcodeconcerningthewaterprobemeasurement,encounteredbytheoptionalHCU/ICU_HPIboard.
ItemEWwithexternalwaterbottomprobe(Sidemountedwaterprobeor964WaterScout):xx00 Noerror,valueatinitializationxx11 NoreplyoninitialHARTcommands Checkconnectionlevelgauge-waterprobe;checkifHARTdeviceaddressofwaterprobe
is4;changewaterprobeelectronics,orHCU/ICU_HPIboardxx35 Wrongdimension DimensionofPVshouldbe:m(metres)ormm(millimetres)xx40 Devicemalfunction Defectivewaterprobe(electronics)xx98 Nowatersourcesfound Nowaterprobeconnected
ItemEWwith765and766VITOwaterprobe:xx00 Noerror,valueatinitializationxx01 C error Eitherdefective762VITOInterfaceoradefective(leaking)waterprobetestxx02 Linearityerror Shouldbewithin1%ofthetestcapacitance.Iftoolarge,change762VITOInterfacexx03 Nowaterprobeavailable Measuredcapacitanceis>4000pF;checkwaterprobewiring(greenwireandcoaxcable)to
762VITOInterfaceterminalsxx04 Subsystemerror Defective762VITOInterfacexx05 Capacitancetoolow Measuredcapacitance<50pF;checkwaterprobewiringorchange765/766VITOprobexx06 Capacitancetoohigh Measuredcapacitance>3000pF;checkwaterprobewiringorchange765/766VITOprobexx07 Capacitanceshortcut Eitherwaterprobeerrororinternal762VITOInterfaceerrorxx11 NoreplyoninitialHARTcommands Communicationerror.Checkconnectionlevelgauge-762VITOInterface,orchange762
VITOInterfaceorHCU/ICU_HPIboardxx79 Notcalibrated CheckitemsVRandVT(settozeroorareequaltoeachother)xx98 Nowatersourcesfound Nowaterprobeconnected
xx: 24forHPUemulation; 28forHSUemulation; 30forHCUemulation
4.4.2 HARTcommunicationerrors
ItemHEcontainsfourcounters.Thefirstthreecountersareusedbypressuretransmitters(ifpresent).Thelastcountercontainsthenumberoftimestheexternalwaterprobeisdetectedabsent.Theexternalwaterprobeisconsideredtobeabsentafterthreeconsecutiverequests,orafterreceivingthreeincorrect/invalidanswerrecords. DefaultvalueitemHE 000:000:000:000
4.4.3 Waterlevel(ullage)statusbytes
Thewaterlevelandullage,itemsLWandLH,areconstructedasfollows:s s lllllllll0 1
lllllllll :waterlevel(orullage)value;formataccordingtheleveldimensionLD.s s :twostatusbytes,bitcoded;containingthefollowinginformation:0 1
statusbytes statusbytes0 1
bit 0: generalprobefail bit 0: waterprobe(0=present;1=absent)1: lowwateralarm 1: 0(notused)2: highwateralarm 2: 0(notused)3: waterbelowprobewarning 3: 0(notused)4: wateraboveprobewarning 4: 0(notused)5: 0 5: 06: 1 6: 17: 0 7: 0
Note:Onlythebitswhicharesetto‘1’haveanactivestatus.
Page 30
Waterbottommeasurement
Page30
4.4.4 Waterbottompointer(itemsVPandVV)
Bymeansofthevaluepointer(itemVP)avectorcanbeloadedtotheHCUorICU_HPIoptionboard.Next,withitemVV,theselecteddataisreturned.ItemVPconsistsof4positions,inthemiddleseparatedbya‘.’or‘,’:vw.xy(orvw,xy).Thevaluesforthevaluepointerarelistedinthetabletogetherwiththeobtaineddata.
vw , xy Selecteddata Example/Dimension
00 , 00 HCU/ICU_HPIEmulation&Function VV=HCUHCAOMTWTPR
00 , 01 Configurationbootcode VV=HCUCONFIG:3F
00 , 02 HCU/ICU_HPIhardwareversion VV=HWVERSION:0000 , 03 Bootcodesoftwareversion VV=BOOTSWVERS:01
03 , 01 ErrorcountersHARTaddresses3,4,5 VV=0000:0013:0000
03 , 09 DetectedHARTdeviceaddresses (seeVP03,01) VV=1---5---------03 , 10 InitialdetectedVITOdevicetype&s/wversion VV=766PTTCW10
03 , 11 RealactivefunctionsofVITOInterface(refertodescriptionofVP=03,10) VV=766PTTCW03 , 12 ActualexecutedHARTcommandsfrom762VITOInterface
04 , 00 C from762VITOInterface (floatingpointformat) pF04 , 01 C from762VITOInterface (floatingpointformat) pF
Emulation: MPU,HSU,HPU,HCUFunction: HC: HARTchannelinstalled
AO: analogoutputST: spottemperatureMT: VITOaveragetemperatureWS: externalwaterbottomprobeWT: VITOwaterbottomprobePR: pressuretransmitters
Salescodeoption: C bootcode: 0AJ 2EU 25W 1AY 3F
(externalwaterprobehasHARTaddress4and762VITOInterfacehasHARTaddress5)
Description: 765: VITOwaterbottomprobe766: VITOwaterandtemperatureprobewithspotcloseto model:.A..766B<ol><he>
probe’send864: 766VITOwaterandtemperatureprobewith model:.A..766A<ol><he>
referenceresistorandlowestspotat1mPT: Pt100TC: thermocouplesW: waterprobe10: softwareversion1.0
test
linearity
Item Name Description
VP= Waterbottomvaluepointer Waterbottomvaluepointer;format:2digits,separator,2digits (refertotableabove).
Example: VP=04.00:valuepointerloadedtorequestthemeasuredtestcapacitor.
VV Waterbottompointervalue Waterbottompointervalue.ThisitemholdsthevaluerequestedbyitemVP(refertotableabove).Example: therequestedmeasuredtestcapacitor
(standardfloatingpointformat).
Page 31
Analogleveloutput
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page31
5 Analogleveloutput
5.1 Introduction
TheanalogleveloutputcircuitisanoptionwhichcanbeinstalledontheHCUboard.TheHCUboardisanoptionboardforthe854servogauges,the873SmartRadargaugeanthe877FDI.Forthe973SmartRadarLT,theanalogleveloutputisavailablewiththeoptionalICU_HPOboard.
TheanalogoutputcircuitontheHCUandICU_HPOoptionalboardsaregalvanicallyseparatedfromthegaugeelectronics.
Thissection(analogleveloutput)describesthesettingsforanalogleveloutputforboththeHCUandICU_HPOboard.FortheHART communicationoftheICU_HPOboard,refertoinstructionmanual973SmartRadarLT®
HART communicationoutput.®
5.1.1 HCUanalogoutputcharacteristics
Theanalogoutputispassive,henceanexternalpowersupplyisrequired.Thevoltageoftheloopsupplyshouldbeatleast12Vdc,andmaynotexceed64Vdc(measuredattheterminalsofthegauge).
Theloopresistanceisthetotalresistancefromtheloop,andhenceincludessupplyoutputresistance,totalcableresistance(includingseriesresistor)andtheinputresistancefromreaderoranaloginput.
ToreducethepowerdissipationontheHCUboard,anexternalresistormustbewiredinserieswiththeloopincasetheloopsupplyvoltageishigherthan30Vdc(refertofigure5.1A).Thepowerratingoftheexternalresistancedependsonthevalue;werecommend:1W/kΩ.
5.1.2 ICU_HPOanalogoutputcharacteristics
Theanalogoutputcanbeseteitherpassiveoractive(refertosection5.2.1).Whensettoactive,thesupplyvoltageintheloopis:20Vdc(±5%).
Thevoltageoftheloopsupplyshouldbeatleast8Vdc,andmaynotexceed64Vdc(measuredattheterminalsofthegauge).Supplyvoltagesover30Vdcrequireanexternalseriesresistor(refertofigure5.1B).
Note:HART communicationrequiresaminimumloopresistanceof250Ω.®
ThemaximumloopresistanceforHART communicationis1100Ω.®
ThisimpliesthatforHART communicationtheminimumsupplyvoltageis14Vdcandthemaximumsupply®
voltageis36Vdcwitha250Ωresistor.
Withahigherexternalsupplyvoltage,theseriesresistormustbeincreased.WiththemaximumresistorforHART communication(1100Ω),themaximumloopsupplywillbe57Vdc.®
Thepowerratingoftheexternalresistancedependsonthevalue;werecommend:1W/kΩ.
Page 32
Analogleveloutput
Page32
Figure5.1 Supplyvoltageversusloopresistanceforanalogleveloutput
Page 33
CN2
CN1
ICU-HPO board
PassiveActive
CN1CN2 -
+
HART
4-20mA
Analogleveloutput
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page33
Figure5.2 Changingoverfrompassivetoactiveanalogoutputloopsupply
5.2 Commissioningofanalogleveloutput
5.2.1 Settingofanalogoutput973SmartRadarLTto“active”
SettingtheanalogoutputlooptoactiveorpassiveisdonebyselectingthecorrectconnectorontheICU_HPOboard.Standard,theSmartRadarLT’saredeliveredwiththeanalogoutputloopsetas“passive”.Hence,anexternalsupplyvoltageisrequired.
Whentheanalogloopshouldbesetto“active”,theconnectorontheICU_HPOboardneedstobechangedfromposition.Thatprocedureisdescribedbelow.
1 SwitchoffmainspowerfromtheSmartRadarLT;2 Openelectroniccompartmentcover;3 LocatetheICU_HPOboard(refertofigure5.2);4 LoosenthecableconnectorfromconnectorCN1(passive),andplugitintoCN2(active);5 Closeelectroniccompartmentcover;6 Applymainspower.
ThecommissioningoftheoptionalanalogleveloutputpartoftheHCU/ICU_HPOboardshouldbeperformedafterthebasiccommissioningofthelevelgaugeorindicator.
Page 34
Analogleveloutput
Page34
5.2.2 Analogoutputsettings
Item Name Description
W1= Protectionlevel1 Enterprotectionlevel1
AM= Analog4mAlevel FormataccordingtoitemLD(default:+000.0000m).Enterinthisitemthelevelvaluewhichmustcorrespondwiththecurrentof4mA
AN= Analog20mAlevel FormataccordingtoitemLD(default:+000.0000m).Enterinthisitemthelevelvaluewhichmustcorrespondwiththecurrentof20mANote:
ThevalueinitemAMcanbeconsideredas‘offset’;Thevalueof(AN-AM)canbeconsideredas‘span’.
AK= Analogoutputmode Onecharacter,defaultsetto:D(analogoutputdisabled).Viathisitemisselectedunderwhatconditionstheanalogoutputcurrentwilloperate;seetable:
ItemAK Validlevel Invalidlevel LevelfailTime-out/start-up
D Analogoutput disabled
O (overwrite I =3.5mA(duringstart-up)P mode) I =22mA(duringstart-up)
I =<AF>oo
o
Ifno<LV>L I =3.5mA I =3.5mAM I =22mA I =22mA
I =<AL> I =<LV>o oo
o
o
o
A I =3.5mAB I =22mA
I =<AL>oo
o
I I =3.5mAJ I =22mA
I =<AL>oo
o
<AF> : valuegivenbyitemAFinanalogoutputoverwritemode<AL> : actuallevelvalue<LV> : lastvalidlevelvalueI : outputcurrentoNote:
Thesetting‘D’shouldnotbeusedwiththe973SmartRadarLT.
EX Exit Exitprotectionlevel.
Note:Withthe973SmartRadarLTitispossibletowritevalueswiththeHARTcommunicatorforPVLRVandPVURV,whichwillbestoredlocallyontheICU_HPOboardandusedinsteadofresp.itemsAMandAN.
Page 35
Analogleveloutput
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page35
Figure5.3 Testloopforcalibratinganalogoutputcircuit
5.2.3 Calibrateanalogoutputcircuit
CalibrationoftheanalogoutputcircuitisnecessarywhentheXPUNOVRAMis(re)formattedandtheitemscouldnotbedownloadedfromalogfileorwhenanewHCU/ICU_HPOboardisinstalled.Ofcourse,calibrationcanalsobeperformedasaroutinetobesurehavingagoodsignal.
Connectapowersupply(15-24Vdc)andanaccuratemAmeter(accuracybetterthan10µA)totheinstrument(refertofigure5.3).
Note:The973SmartRadarLTcanbesettoactive.Thennoexternalpowersupplyisrequired.
TheproceduredescribedbelowcanbeusedwiththeHCUboardinthe854servogauges,873SmartRadarand877FDI.
Forthe973SmartRadarLTwiththeICU_HPOboard,thisprocedurecanonlybeusedviatheHART®
communicationusingtheservicetoolEnsiteorEngauge.
Item Name Description
W1= Protectionlevel1 Enterprotectionlevel1
A5 Calibratecommand@4mA Withthiscommandacurrentofapproximately4mAwillflowthroughtheloop.ReadthecurrentindicatedbythemAmeter.
A3= Analogoutput@4mA Floatingpointformat;units:mA.EnterinthisitemthereadingfromthemAmeterundertheconditionofcommandA5(appr.4mA).
A6 Calibratecommand@18mA Withthiscommandacurrentofapproximately18mAwillflowthroughtheloop.ReadthecurrentindicatedbythemAmeter.
A4= Analogoutput@18mA Floatingpointformat;units:mA.EnterinthisitemthereadingfromthemAmeterundertheconditionofcommandA6(appr.18mA).
EX Exit Exitprotectionlevel.
FromICU_HPOsoftwareversionA1.1andhigher,the973SmartRadarLThasanalternativeitemforthecalibrationoftheanalogoutput.
Page 36
Analogleveloutput
Page36
Item Name Description
W1= Protectionlevel1 Enterprotectionlevel1
AE=+.40000000E+01 Withthisdataitemacurrentofapproximately4mAwillflowCalibrate@4mA throughtheloop.ReadthecurrentindicatedbythemAmeter.
A3= Analogoutput@4mA Floatingpointformat;units:mA.EnterinthisitemthereadingfromthemAmeterundertheconditionwithitemAEsetto4mA.
AE=+.18000000E+02 Withthisdataitemacurrentofapproximately18mAwillflowCalibrate@18mA throughtheloop.ReadthecurrentindicatedbythemAmeter.
A4= Analogoutput@18mA Floatingpointformat;units:mA.EnterinthisitemthereadingfromthemAmeterundertheconditionwithitemAEsetto18mA.
EX Exit Exitprotectionlevel.
Page 37
Analogleveloutput
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page37
5.2.4 Verifyoutputcurrent
Theanalogoutputcurrentcanbeverifiedbythesametestset-upasinfigure5.3.TheproceduredescribedbelowcanbeusedwiththeHCUboardinthe854servogauges,873SmartRadarand877FDI.Forthe973SmartRadarLTwiththeICU_HPOboard,thisprocedurecanonlybeusedviatheHART®
communicationusingtheservicetoolEnsiteorEngauge.
Item Name Description
W1= Protectionlevel1 Enterprotectionlevel1
AK= Analogoutputmode Setanalogoutputinoverwritemode:O
EX Exit Exitprotectionlevel.
Item Name Description
AF= Analogoutputoverwrite Floatingpointformat;units:mA.EnterinthisitemacurrentvaluewhichyouwanttoverifyonthemAmeter.Minimumvalue:3.5mA,maximumvalue:22mA.Thecurrentvalue,enteredinitemAF,shouldbeindicatedonthemAmeter.Itmaynotdiffermorethan12µA.
OneormorecurrentvaluescanbeverifiedbymeansofitemAF.Whenready,setitemAKbacktoitsoriginalsetting:
Item Name Description
W1= Protectionlevel1 Enterprotectionlevel1
AK= Analogoutputmode SetitemAKbacktoitsoriginalsetting(refertosection5.2.2).
EX Exit Exitprotectionlevel.
FromICU_HPOsoftwareversionA1.1andhigher,the973SmartRadarLThasanalternativeitemfortheverificationoftheanalogoutput.
Item Name Description
W1= Protectionlevel1 Enterprotectionlevel1
AE= Analogoutputoverwrite Floatingpointformat;units:mA.EnterinthisitemacurrentvaluewhichyouwanttoverifyonthemAmeter.Minimumvalue:3.5mA,maximumvalue:22mA.Thecurrentvalue,enteredinitemAE,shouldbeindicatedonthemAmeter.Itmaynotdiffermorethan12µA.TheoutputcurrentholdsthevaluetillanothervalueisgiveninitemAEortheverificationisendedbymeansofanexitcommand.
EX Exit Exitprotectionlevel.
Page 38
Analogleveloutput
Page38
5.3 Operation
5.3.1 Display
Foroperationofthedisplayandtheinformationonit,refertotheinstructionmanualoftheapplicablelevelgaugeor877FDI.Below,onlyanoverviewisgivenwhichdisplayformatgivesinformationabouttheanalogleveloutput.
Displayformat Displayedinformation
J Analogleveloutput
Note:ThereisnodisplayformatJonthe973SmartRadarLT,astheICU_HPOboardisnotconnectedtotheinternalIPCcommunicationbus.
5.3.2 Dataitems
Below,asummaryisgivenoftheavailabledataitemsanderrorcodes.
Item Description
AO
AQ
Measured/ Actualanalogoutputcurrent(floatingpointformat;units:mA)calculateddata ThisvalueisprecededbyonestatuscharacterofitemAQ
(onlywithHCUboardin854/873/877)Actualanalogoutputstatus(refertosection5.4.2)(onlywithHCUboardin854/873/877)
VV Verificationdata Refertodescriptionatsection5.4.3(onlywithHCUboardin854/873/877)
EA
OA
OF
FO
Diagnosticdata Erroranalogoutputrequest(onlywithHCUboardin854/873/877)(refertosection5.4.1)Communicationboard(ICU_HPO)actualstatus(onlyin973)(refertosection5.4.4)Communicationboard(ICU_HPO)lastencounteredfatalerror(onlyin973)(refertosection5.4.5)Communicationboard(ICU_HPO)fatalerrorcounter(onlyin973)
Page 39
Analogleveloutput
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page39
5.4 Troubleshooting
5.4.1 Analogoutputerrorrequest(itemEA)
ThisitemcontainsthemotsrecentanalogoutputerrorencounteredbytheHCUboard.
xx00 Noerror,valueatinitializationxx15 Notcalibrated ItemsA3and/orA4arenotsetcorrectly(default:+.00000000E+00)xx16 Upper/lowerrangeerror ItemsAMand/orANarenotsetcorrectly(bothequalvalue?)xx20 DACread-backerror Internalerror,trytoresolvewithRScommand(orpoweroff/on).Iferrorpersists,
changeHCUboardxx99 Analogoutputoptionboardnotinstalled ChangeHCUboardforthecorrecttype
xx : 22forMPUemulation30forHCUemulation
5.4.2 Analogoutputstatusrequest(itemAQ)
OnlyvalidwiththeHCUboardinthe854servogauge,873SmartRadaror877FDI.Theactualanalogoutputstatusisonecharacterwhichcanbedividedintotwodatafields:theoutputstatus(4bits:‘os’)andtheoutputvaluestatus(1bit:‘ov’)
AQ: b b b b b b b b7 6 5 4 3 2 1 0
0 1 0 ov os os os os
Outputvaluestatus(b ):40 : within4-20mArange1 : outof4-20mArange
Outputstatus(b ,b ,b ,b ):3 2 1 0
b b b b Analogoutputstatus3 2 1 0SettingofitemAK
Errorcodeondisplay
0 0 0 0 validproductlevelused L,M,A,B,I,J --0 0 0 1 lastvalidproductlevelused L,M LV0 0 1 0 invalidlevelused A,B IL0 0 1 1 novalidproductlevelavailable I,J NL0 1 0 0 nolastvalidlevelavailable L,M NL0 1 0 1 nolevelavailable A,B NL0 1 1 0 leveltime-out L,M,A,B,I,J TO0 1 1 1 invalidlevel I,J IL1 0 0 0 levelfailure A,B,I,J LF1 0 0 1 analogoutputgeneralfail -- FL1 0 1 0 analogoutputdisabled D DA1 0 1 1 overwritevalueused O,P OV1 1 0 0 nooverwritevalueavailable O,P NV1 1 0 1 calibrationmodeentered L,M,A,B,I,J CA
Page 40
Analogleveloutput
Page40
5.4.3 Analogoutputpointer(itemsVPandVV)
OnlyvalidwiththeHCUboardinthe854servogauge,873SmartRadaror877FDI.Bymeansofthevaluepointer(itemVP)avectorcanbeloadedtotheHCUoptionboard.Next,withitemVV,theselecteddataisreturned.ItemVPconsistsof4positions,inthemiddleseparatedbya‘.’or‘,’:vw.xy(orvw,xy).Thevaluesforthevaluepointerarelistedinthetabletogetherwiththeobtaineddata.
vw , xy Selecteddata Example/Dimension
00 , 00 HCU/ICU_HPIEmulation&Function VV=HCUHCAOST----
00 , 01 Configurationbootcode VV=HCUCONFIG:3F
00 , 02 HCU/ICU_HPIhardwareversion VV=HWVERSION:0000 , 03 Bootcodesoftwareversion VV=BOOTSWVERS:01
Emulation: MPU,HSU,HPU,HCUFunction: HC: HARTchannelinstalled
AO: analogoutputST: spottemperatureMT: VITOaveragetemperatureWS: externalwaterbottomprobeWT: VITOwaterbottomprobePR: pressuretransmitters
Salescodeoption: V bootcode: 10W 1AX 12Y 3F
Item Name Description
VP= Analogoutputvaluepointer Analogoutputpointer;format:2digits,separator,2digits (refertotableabove).
Example: VP=00.00:valuepointerloadedtorequesttheemulationandfunctions.
VV Analogoutputpointervalue Analogoutputpointervalue.ThisitemholdsthevaluerequestedbyitemVP(refertotableabove).Example: therequestedemulationandfunctions.
5.4.4 Communicationboard(ICU_HPO)actualstatus(itemOA)
OnlyvalidwiththeICU_HPOboardinthe973SmartRadarLT.ItemOAcontainstheactualstatusoftheICU_HPOboard,30secondsafterstartup.Thefirst30secondsafterinitialization(EX,orRScommandorpoweron)thisitemcontainstheICU_HPOsoftwareversion.
OK000 NoerrorencounteredER010 ICU(XPU-part)notfound BadornocommunicationbetweenICUboardandICU_HPOboard.CheckER011 ICU(XPU-part)notresponding connectorbetweenthoseboards,elsechangeeitheroftheboardsER012 ICU(XPU-part)answerwitherrorER020 DACerror Internalerror;trytoresolvewithRScommand(orpoweroff/on).Iferrorpersists,
changeICU_HPOboardER021 notcalibrated ItemsA3and/orA4arenotsetcorrectly(default:+.00000000E+00)ER030 NOVRAMchecksumerror InternalerroronICU_HPOboard.ChangeICU_HPOboardER031 NOVRAMaccesserrorER040 codechecksumerrorER050 RAMerror
Page 41
Analogleveloutput
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page41
5.4.5 Communicationboard(ICU_HPO)lastencounteredfatalerror(itemOF)
OnlyvalidwiththeICU_HPOboardinthe973SmartRadarLT.ThisitemholdsthelastencounteredfatalerrorcodeoftheICU_HPOboard.
0000 noerrorsencountered0001 cmxerrors Theseerrorsarefatalsoftwareerrorsandshouldnotoccur.0002 programerrors Iftheydooccurandarenotresolvedbyresettingtheinstrument,replacetheICU_HPOboard0003 interrupterrors (inclusiveICU_HPOEPROM).0004 watchdogerrors
Page 42
Appendix
Page42
AppendixA ASCIItable
Page 43
Appendix
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page43
Commandpushbutton
CautionIfthecommandselectionswitch(2)and/orthecommandpushbutton(1)areoperatedaccidentally,
orinthewrongway,thecalibrationdata(PVLRVorPVURV)arelost.
AppendixB Configurationprocedurefor964WaterScout
TheWaterScoutisacapacitivewaterprobe.Thesituationat0%waterlevelmeans:sensitivepartoftheWaterScouttotallyimmersedinproduct.ThemeasuredcapacitanceinthissituationmustcorrespondwiththePVLRVvalue.
Themeasuredcapacitancedepends,amongotherquantities,onthedielectricconstantoftheproduct.TheWaterScouthasbeenpre-calibratedinthefactory,howevernotintheproduct,butinair.Therefore,itisrequiredtodeterminethePVLRVvalueintheactualmeasuredproduct.
ThePVLRVcanbeprogrammedbytwoways:• 1 Bymeansofthecommandselectionswitchandcommandpushbutton• 2 BymeansoftheHART Communicatormodel275®
1 PVLRVsettingwithcommandpushbutton
MakesuretheWaterScoutisproperlyinstalled(refertoinstallationguide964WaterScout).CheckbymeansofahanddipthatnowaterisfoundatthesensitivepartoftheWaterScoutandtheproductlevelissuchthatitcoverstheentiresensitivepart.
Ifthereiswaterfoundwithinthesensitivepart,thenfollowtheprocedurewiththeHART communicator.Ifthis®
isnotpossible,theWaterScouthastobelifteduntilthesensitivepartisnotanymoreimmersedinwater.
Takecarethatthegroundconnectionisstillmaintained;thisisnotonlyrequiredforsafety,butalsorequiredforagoodmeasurement.
ApplypowertotheinstrumenttowhichtheWaterScoutisconnectedandwaitforapproximately2minutes.
1 OpencoverofWaterScouttransmitterbox
2 Setcommandselectionswitch(2)inposition1(mostleftposition)
3 Presscommandpushbutton(1)atleastfor1second
4 Setcommandselectionswitchbackinposition4(mostrightposition)
5 ClosecoverfromWaterScouttransmitterbox.
Page 44
% water water level 25 mm offset
sensitive length× 100 %
% water 125 25 0
500× 100 20 %
PV LRV PV URV PV URV PV100 % water
× 100
PV LRV 1000 1000 840100 20
× 100 800 pF
Appendix
Page44
Manualmeasurementofwaterlevel
2 PVLRVsettingwithHART Communicator®
MakesuretheWaterScoutisinstalledproperly(refertoinstallationguide964WaterScout).
Checkbymeansofahanddipthewaterlevel.
Ifthereisnowaterfound,orthewaterlevelisbelowthesensitivepartoftheWaterScout,themeasuredcapacitance(PV)isusedasthevalueforPVLRV.
IfthereiswaterfoundwithinthesensitivepartoftheWaterScout,aninterpolationisappliedonthemeasuredcapacitancetofindthecapacitancevalueatzerowaterlevel.
Refertofigure:manualmeasurementofwaterlevel.
Theinterpolationofthecapacitancevalueisdoneasfollows:
Example:
TocalculatethePVLRVvalue,theactualPV(actualmeasuredcapacitance)andthePVURVcapacitancewith100%waterlevel)shouldbeknown.ThesetwovaluescanbereadbytheHART communicator.®
AnotherwayistorequestforitemPVattheinstrument.ThisitemholdthemeasuredcapacitanceinpF.ThePVURVisalsoprintedonthetestformsuppliedwiththeWaterScout.
ThePVLRVisthencalculatedasfollows:
Example:
Proceedasfollows:
ApplypowertotheinstrumenttowhichtheWaterScoutisconnectedandwaitforapproximately2minutes.
1 Enterprotectionlevel2oftheinstrument.Note:
Ifthe847PETisused,itisadvisedtodisablethekeyboardtime-outfunctionofthePET.
Page 45
Appendix
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page45
HARTCommunicator1→Offline2 Online3 FrequencyDevice4 Utility
HARTCommunicatorOnline1→4
Generic:Ignorenext50occurrencesofstatus
YES NO
Generic:Analogoutput1anditsdigitalrepresentativeareinfixedmode,andnotresponsivetoinputchanges
NEXT
Generic:Online(Generic)1→Devicesetup2 PV 840.000 pF3 PVAO 4.000mA4 PVLRV 720.000pF5 PVURV1000.000pF
Generic:1→PVLRV2 PVURV
2 Issuethe‘StopHARTrequest’command(itemSR).ThiscommandwillaborttheHARTcommunicationschedular.Note:
ThisisnecessaryastheoptionalHARTboarddoesn’tallowasecondmasterontheHARTline.
3 SwitchontheHARTcommunicator.Waitfortheselftestmessagethatnodeviceisfound.ThenpressOK<F4>.
4 ConnecttheHARTcommunicatortotheHARTcommunicationline(totheWaterScout).
5 TheHARTcommunicatorshowsthefollowingmessageonthedisplay:
Select“Online”bypressingthe“⇓”keyandthe“⇒”key.
6 Afterthescanningiscompleted,thefollowingmessagewillbeshown:Note:
IfmoreHART devicesareconnected,theiraddressesareshowntoo.®
Selecttheprobewithaddress4bypressingthe(“⇓”keyand)“⇒”key.
7 Thefollowingtwomessagesaredisplayedintermittently.AnsweralwayswithNEXT<F3>andYES<F1>.
8 Thefollowingmenuwillbeshown:
Onthismenuthemeasuredcapacitanceisdisplayed(example:PV840pF)aswellasthefactoryprogrammedPVURVvalue(example:PVURV1000pF).Notethesetwovaluesandusetheminthecalculations(ifrequired)asgiveninthebeginningofthissection.
Select4(PVLRV)bypressingthe“⇓”keythreetimesandthe“⇒”key.
9 Select1(PVLRV)fromthenextmenubypressingthe“⇒”key.Enterthe(calculated)valueforPVLRVonthenumericalkeypathandpressENTER<F4>.
10 PressSEND<F2>andpressOK<F4>twice.
11 PressHOME<F3>.ThisconcludesthesettingofthePVLRVvalue.TheHARTcommunicatorcanbeswitchedoffanddisconnected.
12 Giveexitcommand(itemEX)onthe847PET(orbyservicetool),whichthenre-startstheinstrumentandHARTcommunication.
Page 46
% water water level 25 mm offset
sensitive length× 100 %
% water 125 25 0
500× 100 20 %
VR VT VT MX
100 % water× 100 VR 1000
1000 840100 20
× 100 800 pF
Appendix
Page46
Manualmeasurementofwaterlevel
AppendixC Configurationprocedurefor765/766VITOwaterprobe
MakesuretheVITOwater(temperature)probeisinstalledproperly(refertoinstallationguideVITO)
Checkbymeansofahanddipthewaterlevel.
Ifthereisnowaterfound,orthewaterlevelisbelowthesensitivepartoftheVITOwaterprobe,themeasuredcapacitanceisusedasthevalueforitemVR.
IfthereiswaterfoundwithinthesensitivepartoftheVITOwaterprobe,aninterpolationisappliedonthemeasuredcapacitancetofindthecapacitancevalueatzerowaterlevel.
Refertofigure:manualmeasurementofwaterlevel.
Theinterpolationofthecapacitancevalueisdoneasfollows:
Example:
TocalculatethevalueforitemVR(minimumcapacitance),theactualmeasuredcapacitance(itemMX)andthemaximumcapacitance(itemVT)shouldbeknown.ThesetwovaluescanbereadbythePETorviatheserviceprogramfromtheinstrument.
Theminimumcapacitancevalueisthencalculatedasfollows:Example:
Proceedasfollows:Applypowertotheinstrumenttowhichthe762VITOInterfaceisconnectedandwaitforapproximately2minutes.
Item Name Description
W2= Protectionlevel2 Enterprotectionlevel2
MX Actualmeasuredcapacitance Readtheactualmeasuredcapacitance(floatingpointformat).
VT Maximumcapacitance Readthemaximumcapacitance(floatingpointformat).
VR= Minimumcapacitance Programtheminimumcapacitance;eitherthevaluefromitemMXorthecalculatedminimumcapacitancevalue(floatingpointformat).
EX Exit Exitprotectionlevel.
Page 47
PV URVnew
PV URVorg. PV LRV
cos
PV LRV
PV URVnew 258 12
0.866 12 296 pF
Appendix
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page47
Relation: probelength-measuredwaterlevel
AppendixD Configurationofsidemountedwaterprobes
Thesidemountedwaterprobeworksonacapacitiveprinciple.Thesituationat0%waterlevelmeans:sensitivepartofprobetotallyimmersedintheproduct.ThecapacitancemeasuredinthissituationmustcorrespondwiththePVLRVvalue.
Themeasuredcapacitancedepends,amongotherquantities,onthedielectricconstantoftheproduct.Thewaterprobehasbeenpre-calibratedinthefactory,howevernotintheproduct,butinair.Therefore,itisrequiredtodeterminethePVLRVvalueintheactualmeasuredproduct.Checkbymeansofamanualdipthatnowaterispresentinthesensitivepartoftheprobe.Theproductlevelmustbesuchhighthatitcoversthesensitivepartofthewaterprobe.OnlyundertheseconditionsthePVLRVcanbedetermined.
ThePVURVvalueisthemeasuredcapacitancewhenthesensitivepartoftheprobeiscompletelyimmersedinwater.Thisisalsopre-programmedinthefactory.
TheSVvalue(SecondaryVariable)isrequestedinthecommunicationbetweenthewaterprobeandtheinstrument.TheSVisthemeasuredwaterlevelinmetres.
Asthesidemountedwaterprobeisinstalledunderanangleαwiththevertical,theimmersedprobelengthdoesnotcorrespondwiththeactualwaterlevel(refertofigureattheright).Thesolutionforthisistodecreasetheprobelengthvalue(programmedinSVURV)bythefactorcosa.However,thiscan’tbedonebytheHARTcommunicator.
Therefore,anothermethodisused.ThespanofthePV(PVURV-PVLRV)willbeincreasedbythefactorcosα.Usethefollowingmethod:
Note:ItisrequiredtohavethePVLRVvalueadjustedtothecorrectvalue,priortouseitintheformulaabove.
Example:PVLRV: 12pF(correctedfor0%water)PVURV: 258pFangleα: 30°
Proceedasfollows:
Applypowertotheinstrumenttowhichthewaterprobeisconnectedandwaitforapproximately2minutes.
1 Enterprotectionlevel2oftheinstrument.Note:
Ifthe847PETisused,itisadvisedtodisablethekeyboardtime-outfunctionofthePET.
2 Issuethe‘StopHARTrequest’command(itemSR).ThiscommandwillaborttheHARTcommunicationschedular.Note:
ThisisnecessaryastheoptionalHARTboarddoesn’tallowasecondmasterontheHARTline.
Page 48
Appendix
Page48
HARTCommunicator1→Offline2 Online3 FrequencyDevice4 Utility
HARTCommunicatorOnline1→4
Generic:Ignorenext50occurrencesofstatus
YES NO
Generic:Analogoutput1anditsdigitalrepresentativeareinfixedmode,andnotresponsivetoinputchanges
NEXT
Generic:Online(Generic)1→Devicesetup2 PV 12.100pF3 PVAO 4.000mA4 PVLRV 15.000pF5 PVURV258.000pF
Generic:1→PVLRV2 PVURV
3 SwitchontheHARTcommunicator.Waitfortheselftestmessagethatnodeviceisfound.ThenpressOK<F4>.
4 ConnecttheHARTcommunicatortotheHARTcommunicationline(tothewaterprobe).
5 TheHARTcommunicatorshowsthefollowingmessageonthedisplay:
Select“Online”bypressingthe“⇓”keyandthe“⇒”key.
6 Afterthescanningiscompleted,thefollowingmessagewillbeshown:Note:
IfmoreHART devicesareconnected,theiraddressesareshowntoo.®
Selecttheprobewithaddress4bypressingthe(“⇓”keyand)“⇒”key.
7 Thefollowingtwomessagesaredisplayedintermittently.AnsweralwayswithNEXT<F3>andYES<F1>.
8 Thefollowingmenuwillbeshown:Onthismenuthemeasuredcapacitanceisdisplayed(example:12.1pF)aswellasthefactoryprogrammedvaluesforPVLRV(example:15pF)andPVURV(example:258pF).ThemeasuredPVhastobeprogrammedinthePVLRV(assumedcondition:sensitivepartofprobenotinwater).WiththeprogrammedPVLRVvalue,thecalculationisdoneforthePVURV
Select4(PVLRV)bypressingthe“⇓”keythreetimesandthe“⇒”key.
9 Select1(PVLRV)fromthenextmenubypressingthe“⇒”key.EnterthevalueforPVLRVonthenumericalkeypathandpressENTER<F4>.
Select2(PVURV)formthemenubypressingthe“⇓”and“⇒”key.Enterthere-calculatedvalueforPVURVonthenumericalpathandpressENTER<F4>.
10 PressSEND<F2>andpressOK<F4>twice.
11 PressHOME<F3>.ThisconcludesthesettingofthePVLRVandPVURVvalues.TheHARTcommunicatorcanbeswitchedoffanddisconnected.
12 Giveexitcommand(itemEX)onthe847PET(orbyservicetool),whichthenre-startstheinstrumentandHARTcommunication.
Page 49
Appendix
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page49
762VITOInterface
AppendixE Partnumbers
ForVITOfamily:
Refertofigure:762VITOInterface.
Item Description Partno.
1 ProtectivecoveroverHARTterminals 0186.901
2 Printedcircuitboard762VITOInterface 0762.550
3 O-ring,190.18x2.62,NBR70 2132.991
4 Cover762VITOInterface 0186.798
For854ATG/854XTG/873SmartRadar/877FDI:
HCUboardfunctionality Partno.
SpottemperaturePt100 0854.932
Temperatureand/orwaterprobe 0854.930
Temperatureand/orwaterprobe+HARTdevice(s) 0854.931
SpottemperaturePt100+HARTdevice(s) 0854.933
Analogleveloutput 0854.935
Analogleveloutput+VITOtemperatureand/orwaterprobe 0854.936
Analogleveloutput+VITOtemperatureprobe 0854.937
Analogleveloutput+SpottemperaturePt100+VITOtemperatureand/orwaterprobe+HARTdevice(s)
0854.934
For973SmartRadarLT:
ICU_HPIboardfunctionality Partno.
SpottemperaturePt100 0973.922
Temperatureand/orwaterprobe 0973.920
Temperatureand/orwaterprobe+HARTdevice(s) 0973.921
SpottemperaturePt100+HARTdevice(s) 0973.923
SpottemperaturePt100+VITOtemperatureand/orwaterprobe+HARTdevice(s)
0973.924
Page 50
Appendix
Page50
AppendixF Relateddocuments
Instructionmanualseries854ATGlevelgaugeInstructionmanualseries854XTGlevelgaugeInstructionmanual873SmartRadarInstructionmanual877FieldDisplay&InterfaceInstructionmanual973SmartRadarLT
InstructionmanualHIMS/HTGandvapourpressure(P3)measurement
Installationguide762VITOInterface&764,765or766VITOtemperatureand/orwaterbottomprobesInstallationguide864MTTMultiThermosensorThermometer
ItemdocumentationforEnraf854,873,877&973gauges
Page 51
Index
InstructionmanualTemperature,WaterbottomandAnalogoutputoptions Page51
Index
Activelength . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Actualmeasuredcapacitance . . . . . . . . 24,28,46Analog20mAlevel . . . . . . . . . . . . . . . . . . . . . . . 34Analog4mAlevel . . . . . . . . . . . . . . . . . . . . . . . . 34Analogoutput H0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
@18mA . . . . . . . . . . . . . . . . . . . . . . . . . 35,36@4mA . . . . . . . . . . . . . . . . . . . . . . . . . . 35,36current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38errorrequest . . . . . . . . . . . . . . . . . . . . . . . . . 39mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34,37overwrite . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37statusrequest . . . . . . . . . . . . . . . . . . . . . 38,39
Angle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26,47Averagegastemperature . . . . . . . . . . . . . . . . . . 20Averageproducttemperature . . . . . . . . . . . . . . . 20Calibratecommand@18mA . . . . . . . . . . . 35,36Calibratecommand@4mA . . . . . . . . . . . . 35,36Calibration MU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12,20
sidemountedwaterprobe . . . . . . . . . . . . . . . 47waterprobe . . . . . . . . . . . . . . . . . . . . 43,44,46
Caution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Communicationboard(ICU_HPO) OA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38,40
actualstatus . . . . . . . . . . . . . . . . . . . . . . 38,40fatalerrorcounter . . . . . . . . . . . . . . . . . . . . . 38lastencounteredfatalerror . . . . . . . . . . . 38,41
Elementconfiguration . . . . . . . . . . . . . . . . . . . . . 10Enabletemperaturetransmission . . . . . . . . . 11,19Error TF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11,19
analogoutputrequest . . . . . . . . . . . . . . . . . . 38temperaturerequest . . . . . . . . . . . . . . . . 12,20waterrequest . . . . . . . . . . . . . . . . . . . . . . . . 28
FatalHCU/ICU_HPIerrors . . . . . . . . . . 12,20,28Gasimmersiondepth . . . . . . . . . . . . . . . . . . . . . 18HARTcommunicationerrors . . . . . . . . . . . . 28,29HCUboard . . . . . . . . . . . . . 6,7,35,37,39,40,49ICUboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33ICU_HPIboard . . . . . . . . . . . . . . . . . . . . . . 6,7,49ICU_HPOboard . . . . . . . . . 6,31,33-35,37,40,41Indicatormode . . . . . . . . . . . . . . . . . . . . . . . . . . 11Items VV(spotelement) . . . . . . . . . . . . . . . . . . 12,14
A3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35,36A4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35,36A5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35A6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35AE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36,37AF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37AG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20AK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34,37AM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34AN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34AO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38AP . . . . . . . . . . . . . . . . . . . . . . . . . . . 10,12,20AQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38,39EA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38,39EG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11,19
Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3EM(averagetemperature) . . . . . . . . . . . 20,21
EM(spotelement) . . . . . . . . . . . . . . . . . . 12,13EW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28,29FH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28FM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12,20FO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
HE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28,29IM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11LH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28,29LW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28,29M0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12,20MG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18MI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10,18MK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18MN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10MO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10,18MP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18MQ . . . . . . . . . . . . . . . . . . . . . . . . 12,13,20,21
MW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19MX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24,46N4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
OB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11,19OF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38,41PV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28SR . . . . . . . . . . . . . . . . . . . . . . . . . . . 28,45,47TD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10,18
TU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10,18U0-UF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20V0-VF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20VP(analogoutput) . . . . . . . . . . . . . . . . . . . . 40VP(averagetemperature) . . . . . . . . . . . . 21,22VP(spotelement) . . . . . . . . . . . . . . . . . . . . . 14VP(waterlevel) . . . . . . . . . . . . . . . . . . . . . . . 30VR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24,46VT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24,46VV(analogoutput) . . . . . . . . . . . . . . . . . 38,40VV(averagetemperature) . . . . . . . . . . . . . 20-22
VV(waterlevel) . . . . . . . . . . . . . . . . . . . . 28,30WB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25-27WF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27WH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27WL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27WP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26,27WS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
LastfatalHCU/ICU_HPIerror . . . . . . . 12,20,28Maximumcapacitance . . . . . . . . . . . . . . . . . 24,46Minimumcapacitance . . . . . . . . . . . . . . . . . . 24,46MIR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8MIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8MRT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8MTT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8,15,16
wiringconnection . . . . . . . . . . . . . . . . . . . . . . 19
Nozzleheight . . . . . . . . . . . . . . . . . . . . . . . . 25,26
Page 52
Index
Page52
O-ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Optionalboard . . . . . . . . . . . . . . . . . . . . . . . 11,19
HCU . . . . . . . . . . . . . . . . 6,7,35,37,39,40,49ICU_HPI . . . . . . . . . . . . . . . . . . . . . . . . 6,7,49ICU_HPO . . . . . . . . . . . 6,31,33-35,37,40,41
Pointervalue request . . . . . . . . . . . . . . . . . . . . . . . . . . 10,12analogoutput . . . . . . . . . . . . . . . . . . . . . . 38,40temperature(average) . . . . . . . . . . . . . . . . . 22temperature(spot) . . . . . . . . . . . . . . . . . . . . . 14waterbottom . . . . . . . . . . . . . . . . . . . . . . . . . 30
PrimaryVariable . . . . . . . . . . . . . . . . . . . . . . 24,28UpperRangeValue . . . . . . . . . . . . . . . . . . . . 24
PrintVITOInterface . . . . . . . . . . . . . . . . . . . . . . 49Productimmersiondepth . . . . . . . . . . . . . . . . . . 18Productlevel . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Protectivecover . . . . . . . . . . . . . . . . . . . . . . . . . 49PVLRV . . . . . . . . . . . . . . . . . . . . . . . . . . 24,34,47PVURV . . . . . . . . . . . . . . . . . . . . . . . . . 24,34,47Rearcover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49RTD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10SecondaryVariable . . . . . . . . . . . . . . . . . . . . . . 47Sensitivelengthtemperatureprobe . . . . . . . . . . 18Serialnumberexternalwaterprobe . . . . . . . . . . 28Sidemountedwaterprobe . . . . . . . . . . . . . . 23,27Spotelement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Spotpositions . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Spottemperatures . . . . . . . . . . . . . . . . . . . . . . . 20StopHARTrequest . . . . . . . . . . . . . . . . 28,45,47SV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47SVURV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Switchhysteresis . . . . . . . . . . . . . . . . . . . . . 10,18
Waterullage . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Temperaturedimension . . . . . . . . . . . . . . . . . . . . . . . . 10,18elementoffset . . . . . . . . . . . . . . . . . . . . . . . . 18errorrequest(averagetemperature) . . . . . . . 21errorrequest(spotelement) . . . . . . . . . . . . . 13
sourceselection . . . . . . . . . . . . . . . . . . . 11,19status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10statusconversion . . . . . . . . . . . . . . . . . . 10,18statusrequest(averagetemperature) . . . 20,21statusrequest(spotelement) . . . . . . . . . 12,13
Testresistance . . . . . . . . . . . . . . . . . . . . . . . 12,20Totallength . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Valuepointer
analogoutput . . . . . . . . . . . . . . . . . . . . . . . . . 40temperature(average) . . . . . . . . . . . . . . . . . 22temperature(spot) . . . . . . . . . . . . . . . . . . . . . 14waterbottom . . . . . . . . . . . . . . . . . . . . . . . . . 30
VITO . . . . . . . . . . . . . . . . . . . . 8,15,16,23,27,49Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Wateralarmhysteresis . . . . . . . . . . . . . . . . . . . . 27Watererrorrequest . . . . . . . . . . . . . . . . . . . . . . 29Waterhighalarm . . . . . . . . . . . . . . . . . . . . . . . . 27Waterlevel . . . . . . . . . . . . . . . . . . . . . . . . . . 25,28Waterlevelfieldcommunication . . . . . . . . . . . . . 27Waterlevelstatus . . . . . . . . . . . . . . . . . . . . . . . . 29Waterlowalarm . . . . . . . . . . . . . . . . . . . . . . . . . 27Waterprobe
bottomposition . . . . . . . . . . . . . . . . . . . . . 25-27calibration . . . . . . . . . . . . . . . . . . . . . 43,44,46length . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26,27
Waterprobebottomposition . . . . . . . . . . . . . . . 27
WaterScout . . . . . . . . . . . . . . . . . . . . . . . . . . 23,27