Top Banner
The world leader in serving science Eastern Analytical Symposium 2015 Implementing USP Chapters <232>, <233>: Be Prepared for the Changes to Come
41

Implementing USP Chapters , : Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters ,

May 24, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

The world leader in serving science

Eastern Analytical Symposium 2015

Implementing USP Chapters <232>, <233>: Be Prepared for the Changes to Come

Page 2: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

2

• Why Do We Test for Elemental Impurities?

• Shelf Life Concerns • Shortened shelf life • Tainted flavor

• Toxicity Concerns • Monitor for known toxic elements

• Quality Assurance and Control • Ensure proper product labeling • Flag contaminated products before

they are packaged and consumed

Elemental Impurities in Pharmaceutical Products

So, how do we do this testing?

Page 3: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

3

Overview of Current USP Guidelines

• How Do We Test for Elemental Impurities? • USP General Method for Heavy Metals <231>

• Colorimetric reaction for heavy metals (Pb, Hg, Bi, As, Sb, Sn, Cd, Ag, Cu, Mo)

• Reaction with thioacetamide at low pH

• Reaction produces a colored solution and an insoluble metal sulfide

• Colored solution visually compared to a standard solution containing 10 ppm Pb

Is this test fit for purpose?

Page 4: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

4

• Is This Test Suitable for Measuring Elemental Impurities? • Drawbacks to USP Method <231>

• Robustness − Reaction is complicated and time consuming; results may vary from day to day

and technician to technician − Sample prep involves ashing at high temperatures – loss of volatile elements

• Stability − Color changes occur rapidly − Solutions aren’t stable; results can’t be stored for future reference

• Variability − Color interpretation varies between technicians

• Sensitivity/Selectivity − 10 ppm for all sulfide-forming metals (no preferential reactions)

• Safety − Thioacetamide is a known carcinogen

• Interferences − Pb standard is not matrix-matched to the samples − Matrix components affect final color formation and metal recovery

Overview of Current USP Guidelines

How do we test for metal impurities?

Page 5: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

5

Overview of Upcoming USP Guidelines

• USP Chapters to Replace Chapter <231> • <232> Elemental Impurities – Limits • <233> Elemental Impurities – Procedure • <2232> Elemental Contaminants in Dietary Supplements

• Chapter <232> • Limits for 15 elements • “Big Four” – As, Cd, Hg, Pb (mandatory) • Remaining 11 elements (quantify if they might be present)

• Chapter <233> • Procedure 1 – ICP-OES • Procedure 2 – ICP-MS • Alternative procedure acceptance criteria

• Chapter <2232> • Information and guidance purposes • Limits for As, Cd, Hg, Pb • Impurity testing according to Chapter <233>

Page 6: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

6

• Permissible Daily Exposure (PDE) limits set for all 15 elements • Toxicity of each element dependent on its administration route

• Oral (solids and liquids) • Inhaled (aerosols and inhalers) • Parenteral (injectables and implants) • Large-volume parenteral (LVP)

USP <232> Elemental Impurities - Limits

PDE values based on an adult weighing 50 kg (110 lb) *Not considered a safety concern

Elements Oral daily dose PDE (µg/day)

Parenteral daily dose PDE (µg/day)

Inhalation daily dose PDE (µg/day)

LVP Component Limit (µg/g)

Cadmium 5.0 2.5 3.4 0.25 Lead 5.0 5.0 5.0 0.5

Inorganic arsenic 15 15 1.9 1.5 Inorganic mercury 15 1.5 1.2 0.15

Iridium 100 10 1.5 1.0 Osmium 100 10 1.5 1.0

Palladium 100 10 1.0 1.0 Platinum 100 10 1.5 1.0 Rhodium 100 10 1.5 1.0

Ruthenium 100 10 1.5 1.0 Chromium * * 2.9 *

Molybdenum 180 90 7.6 9.0 Nickel 600 60 6.0 6.0

Vanadium 120 12 1.2 1.2 Copper 1300 130 13 13

Elements Oral daily dose PDE (µg/day)

Parenteral daily dose PDE (µg/day)

Inhalation daily dose PDE (µg/day)

LVP Component Limit (µg/g)

Cadmium 5.0 2.5 3.4 0.25 Lead 5.0 5.0 5.0 0.5

Inorganic arsenic 15 15 1.9 1.5 Organic mercury 15 1.5 1.2 0.15

Iridium 100 10 1.5 1.0 Osmium 100 10 1.5 1.0

Palladium 100 10 1.0 1.0 Platinum 100 10 1.5 1.0 Rhodium 100 10 1.5 1.0

Ruthenium 100 10 1.5 1.0 Chromium * * 2.9 *

Molybdenum 180 90 7.6 9.0 Nickel 600 60 6.0 6.0

Vanadium 120 12 1.2 1.2 Copper 1300 130 13 13

Page 7: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

7

Calibration Requirements • Definition of J:

• Concentration (w/w) of the Target (PDE) Limit, appropriately diluted to the working range of the instrument

• Calculate J Values for All Analytes of Interest

• Calibrate the Instrument • Calibrate for each element at concentrations equal to 0.5J and 2J

USP <233> Elemental Impurities – Procedure

PDE

(Max. Daily Dose) X (Dil. Factor) J =

As Cd Hg Pb

Target value [µg/day] 15 5 15 5

Target limit [µg/g] 1.5 0.5 1.5 0.5

Dilution factor for ICP-MS 1000 1000 1000 1000

J [ng/g] 1.5 0.5 1.5 0.5

0.5 J [ng/g] 0.75 0.25 0.75 0.25

2.0 J [ng/g] 3.0 1.0 3.0 1.0

Assume: Oral administration Dose: 10g/day

Page 8: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

8

Validation Requirements – Accuracy • Control Samples

• Solutions spiked with the elements of interest at concentrations equal to 0.5J, 1J and 1.5J, where J is the indicated limit

• Test Samples

• Unknown samples must be spiked with the elements of interest at concentrations equal to 0.5J, 1J and 1.5J, where J is the indicated limit. Each sample must be prepared in triplicate

• Spike Recoveries • Results for the test samples described above must be within 80-150% of

the spiked levels

USP <233> Elemental Impurities – Procedure

Page 9: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

9

Validation Requirements – Precision • Short-term precision

• Six independent samples of the material under test spiked with the elements of interest

• The relative standard deviation of the samples must be lower than 20%

• Intermediate precision • Analysis of the precision samples on either a different instrument, a different

day or by a different analyst • The relative standard deviation of these and precision samples combined

must be less than 25%

USP <233> Elemental Impurities – Procedure

Page 10: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

10

• For USP <233> three sample preparation options:

1. Direct Aqueous • Dissolution in an aqueous matrix

• Not all excipients soluble e.g. TiO2

2. Direct Organic • Dissolution in an organic matrix

• Not all excipients soluble e.g. Magnesium stearate

• Example with DMSO and ICP-OES

3. Indirect Solution • Closed vessel digestion

• See ICP-MS section for an example

• Most universal method

• A typical drug can be described as an API (active pharmaceutical ingredient) + an excipient.

• Common excipients are:

• Binders e.g. xanthan gum

• Glidants and lubricants e.g magnesium stearate

• Disintegrants e.g. crospolividone (E1202)

• Sweeteners e.g. sucrose

• Flavourings e.g. fruit

• Pigments e.g. titanium dioxide

• Preservatives e.g. methylparaben

• Coating e.g. shellac or gelatine

Sample Preparation

Page 11: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

11

Pharmaceutical Analysis – the Scope of Our Elemental Offering

AAS

ICP-OES

ICP-MS

SF-ICP-MS

Features include: • Single- or multi-element analysis • Measurement at concentrations from

parts per trillion to percent • Low-cost, high-throughput systems • Space-saving ergonomic design with

low gas consumption

Products include: • Atomic absorption spectrometers • Atomic emission spectrometers • Mass spectrometers

• Single quad, sector field, multi-collector • Ion chromatography for speciation

IC-ICP-MS

Page 12: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

12

Elemental Analysis Portfolio

Trace Elemental Products

AA Flame AA Furnace ICP-OES Q-ICP-MS HR-ICP-MS

Analytical capabilities (detection limits, sensitivity, analysis speed)

Initial investment

Single-Element Techniques Lower Sample Throughput

Multi-Element Techniques Higher Sample Throughput

Page 13: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

13

Preferred Instrumentation for USP <233>

ICP-OES Procedure 1

ICP-MS Procedure 2

Page 14: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

14

• Ion vs. atom lines • Fewer ion lines seen in radial view • Consider ionization buffer (Ex. CsCl)

• Spectral interferences • Background correction to deal with partial overlaps • Inter Element Correction (IEC) for direct overlap of lines

• Sensitivity and dynamic range • Use of primary, secondary and tertiary wavelengths to extend the

dynamic range • Purged Optical Pathway to measure in low UV

ICP-OES: Some Mind Keepers

Page 15: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

15

• Plasma Robustness: • Oxide ratio, doubly charged ions • Free running frequency-matched RF generator for organic sample matrix

• Interference Removal • Single mode analysis using kinetic energy discrimination, low mass cut-off

• (High) Matrix Content: • Autodilution • Argon Gas Dilution

ICP-MS: More Mind Keepers

Page 16: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

16

Os - Highly volatile in oxidized form

• Os solutions need to be stabilized in order to maintain original content of Os

Loss of Os in sample solutions after storage

Potential issue when using closed vessel microwave digestion

As – Attention to ionization effects

• Increased ionization of As in presence of carbon

Potential overestimation of the real As content

Potential issue when using direct dissolution methods

“Difficult“ Analytes

* Venzago et al., J. Anal. At. Spectrom. 28 (2013), 1125-1129

Solution: Stabilize solutions prior to analysis

Solution: Matrix matched calibration or microwave digestion

Larsen et al., J. Anal. At. Spectrom. 9 (1994), 1099-1105

Page 17: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

The world leader in serving science

Analysis by ICP-OES

Application Example: Over-the-Counter Drugs

Page 18: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

18

Samples: Anti-inflammatory and Antihistamine • Preparing samples in DMSO

• DMSO (dimethyl sulfoxide) is a very strong solvent

• Less toxic than DMF (dimethylformamide)

• High-boiling point

• Drawbacks of using DMSO • Require silicone pump tubing • O-rings on spray-chamber require changing more often • Will not dissolve all excipients

• For example: silica, titanium dioxide

Analysis of Over-the-Counter Medicine

Page 19: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

19

Sample Preparation and Analysis • 0.5 g of dehydrated sample was dissolved in 25 g of DMSO

• J defined as the w/w concentration of analyte at Target Limit after dilution • Target Limit > MDL; recoveries tested at the 0.5J and 1.5J

Analysis of Over the Counter Medicine

0

20

40

60

80

100

120

140

160

Drug 1 + 0.5 J - 1

Drug 1 + 0.5 J - 2

Drug 1 + 0.5 J- 3

Drug 2 + 0.5 J - 1

Drug 2 + 0.5 J - 2

Drug 2 + 0.5 J - 3

Elements 0.5 J (µg/kg)

Cadmium 25

Lead 25

Inorganic As 75

Inorganic Hg 75

Iridium 500

Osmium 500

Palladium 500

Platinum 500

Rhodium 500

Ruthenium 500

Molybdenum 900

Nickel 3000

Vanadium 600

Copper 6500

Page 20: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

20

Sample Analysis and Spike Recovery Data • Precision

• Determined by analyzing six individual samples • Samples spiked at 1J • USP acceptance criteria < 20%

Analysis of Over the Counter Medicine

Elements Measured Conc Results for Drug 1 (µg/L)

Rep 1 Rep 2 Rep 3 Rep 4 Rep 5 Rep 6 % RSD

Cd 232.4 232.7 234.7 239.1 235.6 229.9 1.4 Pb 45.9 45.2 44.6 47 46.6 43 3.2 As 12.1 12.7 12.8 14 12.9 11.4 6.9 Hg 130.7 130.8 132,5 136.5 131.8 127.4 2.3 Ir 944.5 941.3 948.2 963.7 950.9 924.5 1.4

Os 954.8 952.7 959 974.9 960.5 940 1.2 Pd 918.8 914.7 914.6 928.6 929.4 890.6 1.5 Pt 924.4 917.6 931.5 949.9 934.6 910.7 1.5 Rh 921.5 907.2 907.5 917.6 915.8 874.9 1.9 Ru 955.5 966.5 953.6 972.8 967.5 932.7 1.5 Mo 956.8 952 959.6 974 959.5 937.7 1.2 Ni 4669 4666 4706 4787 4718 4610 1.3 V 962.5 952.9 945.5 960.1 961.7 928.9 1.4

Cu 9680 9590 9522 9666 9668 9318 1.5

Page 21: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

The world leader in serving science

Analysis by ICP-MS

Application Example: Over-the-Counter Drugs

Page 22: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

22

• Four over-the-counter products were locally sourced • Two samples of each were weighed into 15 ml disposable glass vials • 3 ml of conc. HNO3 was added to each vial • System was closed, and pressurized with N2 at 40 bar • Microwave digestion procedure:

• When the solution was cooled to <60 ˚C, the digest was transferred to a polypropylene vial and made up to 50 ml with 1% HCl

• Samples were further diluted before analysis (with high purity 2% HNO3) to give total dilution factors of between 100 and 1000

Analysis of Over-the-Counter Medicine

Step Time (min) Temperature (˚C) Power (kW)

1 15 200 1.5 2 10 200 1.5

Page 23: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

23

Results: Spike Recoveries at 0.5J

Analysis of Over the Counter Medicine

0%

20%

40%

60%

80%

100%

120%

140%

160%

180%

200%

51V 60Ni 63Cu 75As 98Mo 101Ru 103Rh 105Pd 111Cd 189Os 193Ir 195Pt 202Hg 208Pb

Pharma 1-1

Pharma 1-2

Pharma 2-1

Pharma 2-2

Page 24: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

24

ICP-MS Detection Limits Compared to Daily Dose

Analysis of Over the Counter Medicine

Element Instrument Detection Limit (ng/mL)

Method Detection Limit (µg/g)

Concentration Limit Max. Daily Dose of

≤10 g/day (µg/g) Cd 0.0001 0.0001 0.5 Pb 0.0005 0.0005 0.5 As 0.0005 0.0005 1.5 Hg 0.003 0.003 1.5 Ir 0.002 0.002 10

Os 0.0006 0.0006 10 Pd 0.0008 0.0008 10 Pt 0.0005 0.0005 10 Rh 0.0007 0.0007 10 Ru 0.001 0.001 10 Mo 0.003 0.003 18 Ni 0.003 0.003 60 V 0.006 0.006 12

Cu 0.009 0.009 130

Page 25: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

The world leader in serving science

Simplified Workflow and Reporting

Page 26: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

26

Thermo Scientific™ Qtegra™ Intelligent Scientific Data Solution™ Software

A Simple Workflow to Quality Results

Dashboard

“Get Ready”

LabBook

“Create LabBook”

Sample Details

LIMS Manual

Template

Experiment method and QA/QC protocol

LIMS LabBook

Results and report

1.

2.

3.

Bar / QR Code

Page 27: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

27

A Simple Workflow to Quality Results

Page 28: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

28

• Method

• QA/QC Protocol

• Sample List

• Results

• Report

• Audit trail

A Simple Workflow to Quality Results

• LabBook: Method Setup to Data Reporting in Just 5 Clicks

Page 29: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

29

• Flexible tool for online generation of printable records • Can be printed online with data acquisition • Password protected .pdf files can be generated

• Tables are completely customizable • Information about Instrument, LabBook, Method Parameters

• Filtering options for reporting of specific samples • Sample name, Sample type, Analyst etc. • Individual tables for subsamples in a bigger batch • Filtering across all acquired data sets is possible to generate e.g. history

records of QA/QC samples

A Simple Workflow to Quality Results

Page 30: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

30

Data Reporting – Example 1

LabBook information: Created by, Acquired by, Last changed by,...

Customized table: Direct overview on precision test for Sample XYZ, Average Recovery and RSD are automatically calculated

Page 31: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

31

Data Reporting – Example 2

Report Intermediate Precision Test: Results obtained by two operators are summarized in one table

Average value and RSD are calculated

Every individual LabBook as datasource can be indentified and history can be displayed

Page 32: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

32

Data Reporting – Example 3

Verification of all installed files Checksums to verify consistency Verification of data evaluation algorithms Executable whenever required

Page 33: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

The world leader in serving science

Qualification

Page 34: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

34

Validation, Software and FDA Compliance

• System qualification includes: • Specification Qualification (SQ) • Design qualification (DQ) • Installation and operational qualification (IQ/OQ) • Performance qualification (PQ) • Periodic Requalification (RQ)

• System qualification covers: • Instrument Hardware and Software

• Compliance with federal regulations is critical • Part 11 in Title 21 of the US code of Federal Regulations (21 CFR Part 11)

• Governs food and drugs in the United States • Includes Federal guidelines for storing and protecting electronic records • Must contain electronic signatures etc.

Page 35: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

35

• Step-by-step install/operation procedures are signed off at each step by certified field service engineers

The Qualification Process

System Qualification

Installation Qualification (IQ)

Performed by Engineer at Installation

Performed by Engineer at Preventive

Maintenace Visit

Operation Qualification (OQ)

Performed by analyst at intervals

prescribed by internal QA or

protocol compliance

Performance Qualification (PQ)

Analyst performed and maintained

schedule of testing. Includes method development and

verification; Development of local

SOP´s etc..

Page 36: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

36

Full 21 CFR Part 11 Compliant Tool Set

• Thermo Scientific™ iCAP™ Qualification Kit • Available for ICP-OES and ICP-MS • Contains all necessary documentation for IQ/OQ ....And more!

Description of all necessary steps for fast and reliable implementation of a system in your lab Manuals and Application Notes for a quick start Documentation for IQ/OQ, Preventive Maintenance and Requalification (RQ) carried out with a service engineer Best Practices Guide, useful information provided by Application Chemists for new users

Page 37: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

37

Full 21 CFR Part 11 Compliant Tool Set

Supported 21 CFR 11 Sections 1.10 Controls for closed systems 11.30 Controls for open systems 11.50 Signatures manifestations 11.70 Signature / record linking 11.100 General requirements for electronic signatures 11.200 Electronic signatures components and controls 11.300 Controls for identification codes / passwords

Page 38: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

38

Typical Instrument Implementation Process

Page 39: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

39

Elemental Speciation

Page 40: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

40

Check out our resources on www.thermoscientific.com

• Social media: Facebook, Twitter, YouTube

• Analyte Guru Blog www.analyteguru.com

• Thermo Scientific AppsLab

• Community pages

• Knowledge base

• Subscribe to our blog and community pages to receive information updates that are relevant to your application!

Thank You For Your Attention

Page 41: Implementing USP Chapters <232>, <233>: Be Prepared for ...tools.thermofisher.com/content/sfs/brochures/PP-TEA-USP-232-233... · Implementing USP Chapters <232>,

41

Check out our pharma-specific resources on http://www.thermoscientific.com/en/community/pharma-biopharma.html

• Columns/Consumables Guide • USP Primer • Webinar • Blogs

Thank You For Your Attention