Introduction
• Posterior chamber phakic intraocular lens (pIOL), effective for the correction of moderate to high ametropia
• The Visian implantable collamer lens (ICL; STAAR Surgical, Monrovia, CA)– Posterior chamber pIOL placed behind the iris
with the haptics resting on the ciliary sulcus
INSTITUTO DE OFTALMOLOGÍAFUNDACIÓN CONDE DE VALENCIANA
Alfonso JF, Baamonde B, Fernández-Vega L, et al. Posterior chamber collagen copolymer phakic intraocular lenses to correct myopia: five-year follow-up. J CataractRefractSurg. 2011
• The pIOL models have undergone improvements to minimize complications
• The newer V4c model is designed with a 0.36mm central hole to overcome acute pupillary block
INSTITUTO DE OFTALMOLOGÍAFUNDACIÓN CONDE DE VALENCIANA
Alfonso JF, Baamonde B, Fernández-Vega L, et al. Posterior chamber collagen copolymer phakic intraocular lenses to correct myopia: five-year follow-up. J CataractRefractSurg. 2011
Purpose
• To determine the incidence, etiology, risk factors and management of complications following implantation of phakic posterior chamber ICL V4 and its newer version ICL V4c.
INSTITUTO DE OFTALMOLOGÍAFUNDACIÓN CONDE DE VALENCIANA
Patients and Methods
• Retrospective review • Consecutive clinical case series. • Inclusion:• Patients who underwent either spheric or
toric ICL implantation between +8 to -24 D sphere and 0 to -6.5 cylinder
INSTITUTO DE OFTALMOLOGÍAFUNDACIÓN CONDE DE VALENCIANA
Results
• A total of 406 eyes were included, 349 ICL V4 and 57 ICL V4c model.
• 20 eyes presented complications (4.92%)• Mean follow-up of 47±31 months (3-127)– Myopic: n=204– Hyperopic: n=8– Toric: n=194
INSTITUTO DE OFTALMOLOGÍAFUNDACIÓN CONDE DE VALENCIANA
Measurement Preoperative Postoperative p Value
Sphere (D) -10.35 ±5.1D -0.09±1.06 <0.001
Cylinder (D) -2.63±1.44 -0.97±0.89 <0.001
SphEq (D) -11.6±5.12 -0.52±1.03 <0.001
UDVA(logMAR) 1.72±0.49 (20/1000)
0.23±0.22(20/30)
<0.001
CDVA (logMAR) 0.21±0.17(20/30)
0.12±0.13(20/25)
<0.001