PLATES AND SCREWS: An Overview Presented by : Dr. REM KUMAR RAI
83

Implant screw plate

May 09, 2015

Download

Technology

Rem Kulung
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Implant screw plate

PLATES AND SCREWS:

An OverviewPresented by :

Dr. REM KUMAR RAI

Page 2: Implant screw plate

SCREW : BRIEF OVERVIEW

Page 3: Implant screw plate

SCREW: INTRODUCTION

An elementary machine to change the small applied rotational force into a large compression force

Function Holds the plate or other prosthesis to the

bone Fixes the # fragments ( Position screw) Achieves compression between the #

fragments (Lag screw)

Page 4: Implant screw plate

Screw: Parts

4 functional parts Head Shaft Thread Tip

shank

Page 5: Implant screw plate

Head: Function

1. Means for applying torque with a screwdriver

2. Acts as a stop (the undersurface) i.e. countersunk

Page 6: Implant screw plate

Head: Recess Types

1. Slotted 2. Cruciate 3. Philips 4. Hex/ Allen 5. Torx (eg Stardrive of Synthes)

Page 7: Implant screw plate

Head: Countersink

Undersurface of head Conical Hemispherical Morse-cone (steep): locking plates

Page 8: Implant screw plate

Screw: Shaft/ Shank

Smooth link Almost not present in standard

cortex screw Present in cortical SHAFT SCREW or

cancellous screw

Page 9: Implant screw plate

Screw: Run out

Transition between shaft and thread Site of most stress riser Screw break

Incorrectly centered hole Hole not perpendicular to the plate

Page 10: Implant screw plate

Screw: Thread

Inclined plane encircling the root Single thread May have two or more sets of

threads

V-thread profile: more stress at sharp corner

Buttress thread profile: less stress at the rounded corner

Page 11: Implant screw plate

Screw: Core Diameter

Narrowest diameter across the base of threads

Also the weakest part

Smaller root shear off

Torsional strength varies with the cube of its root diameter

Page 12: Implant screw plate

Screw: Pitch and Lead

Distance between the adjacent threads Cortex screw : small pitch 1.75mm Cancellous screw: large pitch

Pitch also determines the lead Lead :distance advanced in a complete

turn Equals pitch in single threaded screw Greater M.A. if smaller lead

Page 13: Implant screw plate

Screw: Thread Diameter

Diameter across the maximum thread width

Affects the pull out strength

Cancellous have larger thread diameter

Page 14: Implant screw plate

Screw: Tip Designs

1. Self-tapping tip: Flute Cuts threads in the bone over which

screw advances Cutting flutes chisel into the bone

and direct the cut chips away from the root

Page 15: Implant screw plate

Screw: 2.Non self tapping Lacks flutes Rounded tip Must be pre-cut in the pilot hole by

tap Pre-tapped threads help to achieve

greater effective torque and thus higher inter-fragmental compression

Better purchase

Page 16: Implant screw plate

Screw: 3.Corkscrew tip

Thread forming tips In Cancellous screws

which form own threads by compressing the thin walled trabecular bone

Inadequate for cortical bone

Page 17: Implant screw plate

Screw: 4.Trochar Tip

Like self tapping Displaces the bone as it

advances Malleolar screw Schanz screws Locking bolts for IMIL

Page 18: Implant screw plate

Screw: 5.Self drilling self tapping Like a drill bit In locked internal fixator plate hole Pre-drilling not required Good purchase in osteoporotic and

metaphyseal area

Page 19: Implant screw plate

Locking Screws vs Cortical Screws

Creates Fixed Angle Generates Friction/Compression

4.4mm Core Dia. 3.5mm Core Dia.

5.0 mm Locking Screw4.5 mm Cortical Screw

Page 20: Implant screw plate

Bending stiffness proportional to the core diameter

Pull out strength is proportional to the size of the thread

Cannulated screws have less bending stiffness

Page 21: Implant screw plate

Machine and Wood Screws Wood

Used in wood Large threads , usually tapered Pilot hole is small Elastic force from deformation of

wood Machine

Used in metals Pilot hole matches the size of the

screw core Tapped Elastic force from deformation of

the screw itself

Page 22: Implant screw plate

Tensile strength is directly proportional to the squared core diameter d2

Pull out strength is depends on the outer diameter

Shear strength is directly proportional to the cubed core diameter d3.

Page 23: Implant screw plate

AO/ASIF Screws: Types

Cortical Fully threaded Shaft screw 1.5:phalanx *drill bit 1.1 mm 2.7: mc and phalanx *bit:2.0 3.5: Radius/ Ulna/ Fibula/

Clavicle*bit:2.5 4.5: Humerus/ Tibia/ femur *bit:3.2

2:phalanx

Page 24: Implant screw plate

Cancellous Fully threaded

Cannulated or Non- cannulated Partially threaded

16mm or 32 mm Cannulated or Non-cannulated

4.0, drill bit 2.5mm humeral condyle

6.5 drill bit 3.2mm tibial and femoral condyle

Page 25: Implant screw plate

Cannulated screws 3.0 4.0 4.5 6.5 7.0 7.3

Page 26: Implant screw plate

Locking screws synthes 2.4mm 3.5mm 4.0mm 5.0mm 7.3mm

Page 27: Implant screw plate

AO/ASIF Screws

• Cancellous screws:– a wood type– core diameter is less– the large threads– Higher pitch– Greater surface are for purchase– Untaped pilot hole– Pilot hole equals the core diameter– lag effect option with partially threaded

screws – theoretically allows better fixation in soft

cancellous bone.

Page 28: Implant screw plate

AO/ASIF Screws

• Cortical screws:– a machine type– Smaller threads– Lower pitch– Large core diameter– Smaller pitch higher

holding power– greater surface area of

exposed thread for any given length

– better hold in cortical bone

Page 29: Implant screw plate

Special Screws

Herbert Screw Dynamic Hip Screw Malleolar Screw Locking bolt Interference screw Suture anchor Acutrak screw Pedicle screw

Page 30: Implant screw plate

Herbert Screw Specialized to achieve

interfragmentary compression Headless Threads at both ends Pitch differential between the

leading and trailing threads Compression by the difference

in thread pitch Coarser pitch moves a greater

distance with each turn than does the finer pitch

Page 31: Implant screw plate

Clinical application:Lag Screw Used to compress fracture fragments

Use to hold plates on bone

Threads only engage far cortex

Can be achieve with: - Partially threaded screw - Fully threaded with over drilling near cortex

Page 32: Implant screw plate

Clinical Application: Positional Screw

Page 33: Implant screw plate

Lag Screw

Partially Threaded cancellous screwThreads must be completely across the

fracture to achieve compression and purchase far cortex if possible

Page 34: Implant screw plate
Page 35: Implant screw plate

PLATES

Introduction : Bone plates are like internal

splints holding together the fractured ends of a bone.

Page 36: Implant screw plate

Mechanical functions of plate1. Transmit forces from one end to

another, bypassing and thus protecting the area of fractures.

2. Holds the fracture ends together in alignment throughout the healing process.

Page 37: Implant screw plate

Names of plates.

1. Shape (Semitubular, 1/3rd tubular)2. Width of plate (Small, Narrow,

Broad)3. Shape of screw holes. (Round, Oval)4. Surface contact characteristics. (LC,

PC)5. Intended site of application

(Condylar Plate)6. According to the function

Page 38: Implant screw plate

Type of plate – Functional Regardless of their length, thickness,

geometry, configuration and types of hole, all plates may be classified in to 4 groups according to their function.

1. Neutralization plate.2. Compression plates.3. Buttress plate.4. Tension band plates.

Page 39: Implant screw plate

Standard Plates

Narrow DCP-4.5 mm Broad DCP – 4.5 mm,3.5 mm DCP LC-DCP 3.5 & 4.5mm Reconstruction plate 3.5 & 4.5mm 1/3 tubular plate 2.7, 3.5 & 4.5 mm

Page 40: Implant screw plate

Special Plates

T Plates T&L Buttress plates Lateral Tibial head buttress plates Condylar buttress plate Narrow lenthening plates Broad Lengthening plate Spoon plate Clover leaf plate

Page 41: Implant screw plate

NEUTRALIZATION PLATE• Acts as a ""bridge””

protection

• No compression at the fracture site

• neutralization plate is to protect the screw fixation of • a short oblique fracture

• a butterfly fragment

• a mildly comminuted fracture of a long bone

• fixation of a segmental bone defect in combination with bone grafting.

Page 42: Implant screw plate

The Neutralization Plate

Lag screws: compression and

initial stability

Plate: protects the

screws from bending and torsional loads

Page 43: Implant screw plate

NEUTRALIZATION PLATE

Page 44: Implant screw plate

COMPRESSION PLATE• produces a locking force across

a fracture site

• Newton's Third Law (action and reaction are equal opposite)

• plate is attached to a bone fragment then pulled across the fracture site by a device, producing tension in the plate

• direction of the compression force is parallel to the plate

Page 45: Implant screw plate

COMPRESSION

Static: does not change with time

Dynamic: periodic partial loading & unloading due to functional activity

1. Tension Band wiring2. Tension Band Plating

Page 46: Implant screw plate

BONE UNDER COMPRESSION

• Superior stability – Utilization of physiological forces

• Improved milieu for bone healing

• Early mobilization

Page 47: Implant screw plate

DCP (Dynamic Compression Plate):Principle :

- a self compression plate due to the special geometry of screw holes which allow the axial compression.

Page 48: Implant screw plate

Dynamic compression principle: screw head slides down the inclined plate hole as it is tightened, with the head forcing the bone-screw to move towards the fracture, thereby compressing the fracture

Page 49: Implant screw plate

• Screw hole and the spherical gliding principle.

• Axial compression result from the an interplay between screw hole geometry and eccentric placement of the screw in the screw hole.

Page 50: Implant screw plate

The shape of the holes of the dynamic compression plate allows inclination of the screws in a transverse

direction of +7° and in a longitudinal direction of 25°.

Page 51: Implant screw plate

Advantage of DCP :

1. Inclined insertion 25°longitudinal and 7° sideways

2. Placement of a screw in neutral position without the danger of distraction of fragments

3. Insertion of a lag screw for the compression4. Usage of two lag screws in the main

fragments for axial compression5. Compression of several fragments individually

in comminuted fractures6. Application as a buttress plate in articular area

Page 52: Implant screw plate

Shortcomings of DCP :

1. Flat under surface.2. Inclination upto 25°3. Plate hole distribution (extended

middle segment)

Page 53: Implant screw plate

The structure of a limited-contact dynamic

compression plate

1.Structured undersurface

2.Undercut screw holes3.Trapezoid cross

section

LC-DCP

Page 54: Implant screw plate

In the DCP (A), the area at the plate holes is less stiff than the area between them so while bending, the plate

tends to bend only in the areas of the hole.

The LC-DCP(B) has an even stiffness without the risk of buckling at the screw holes.

Page 55: Implant screw plate

The LC-DCP offers additional advantage

Improve blood circulation by minimizing plate-bone contact

More evenly distribution of stiffness through the plate

Allows small bone bridge beneath the plate

Page 56: Implant screw plate

The trapezoid cross section of the plate results in a smaller contact area between plate and bone.

The plate holes are uniformaly spaced, which permits easy positioning of the plate.

Undercuts plate holes; undercut at each end of the plate hole allows 40 tilting of screws both ways along the long axis of the plate. Lag screw fixation of short oblique fractures is thereby possible.

Page 57: Implant screw plate

Sizes of DCPName of plate Small Narrow Broad

Width 11 mm 13.5 mm 17.5mm

Profile 4 mm 5.4 mm 5.4 mm

Screw 2.7 , 3.5 cortex screw and 4 mm cancellous screw

4.5 mm cortex screw & 6.5mm canellous screw

4.5 mm cortex screw & 6.5mm canellous screw

Sizes of LCDCPName of plate Small Narrow Broad

Width 11 mm 13.5 mm 17.5mm

Profile 4 mm 5.4 mm 5.4 mm

Screw 2.7 , 3.5 and 4 mm cancellous screw

4.5 mm & 6.5mm canellous screw

4.5 mm & 6.5mm canellous screw

Name of plate Small Narrow Broad

Width 11 mm 13.5 mm 17.5mm

Profile 4 mm 5.0 mm 5.0 mm

Screw 4 mm locking screw 5 mm locking screw 5 mm locking screw

Sizes of LCP

Page 58: Implant screw plate

BUTTRESS PLATE• is to strengthen

(buttress) a weakened area of cortex

• The plate prevents the bone from collapsing during the healing process.

• A buttress plate applied a force to the bone which is perpendicular (normal) to the flat surface of the plate.

Page 59: Implant screw plate

• The fixation to the bone should begin in the middle of the plate, i.e. closest to the fracture site on the shaft. The screws should then be applied in an orderly fashion, one after the other, towards both ends of the plate.

• A representative clinical example of a buttress plate is the T-plate used for the fixation of fractures of the distal radius and the tibial plateau.

BUTTRESS PLATE

Page 60: Implant screw plate

Bridge Plating :Bridge Plating for

comminuted fracture

-instead of individually fixing each fragment

-minimal disruption to blood supply

-reduction is performed indirectly

- compression is only sometimes possible

Page 61: Implant screw plate

Wave Plating :

Wave Plating for non union

Page 62: Implant screw plate

ADDITIONAL PRINCIPLES OF PLATE FIXATION

The engineering principle of the tension band is widely used in fracture fixation. It applies to the conversion of tensile forces to compression forces on the convex side of an eccentrically loaded bone.

Page 63: Implant screw plate

Reconstruction Plates :

Can be bent and twisted in two dimensions.

Decrease stiffness than DCP.

Should not be bent more than 15°.

Used were the exact and complex contouring is required. eg. Pelvis, Distal Humerus, Clavicle.

Page 64: Implant screw plate

Reconstruction plates are thicker than third tubular plates but not quite as thick as dynamic compression plates. Designed with deep notches between the holes, they can be contoured in 3 planes to fit

complex surfaces, as around the pelvis and acetabulum. Reconstruction plates are provided in straight and slightly thicker and stiffer precurved lengths. As with tubular plates, they have oval screw

holes, allowing potential for limited compression.

Page 65: Implant screw plate

One Third Tubular Plates : Plates have the form of one third of the

circumference of a cylinder. Low rigidity (1mm thick). Oval holes – Axial compression can be

achieved. Uses – Lateral malleolus, distal ulna,

metatarsals.

Page 66: Implant screw plate

limited stability. The thin design allows for easy shaping and is primarily used on the lateral malleolus and distal

ulna. The oval holes allow for limited fracture compression with eccentric screw placement.

Page 67: Implant screw plate

LOCKING COMPRESSION PLATE (LCP)Principle :

Angular-stability whereas stability of conventional plates is friction between the plate and bone

Screw locking principle

Provides the relative stability Healing by callus formation (Secondary Healing)

Page 68: Implant screw plate

LCP: internal external fixator

Page 69: Implant screw plate

Stability under load

By locking the screws to the plate, the axial force is transmitted over the length of the plate secondary loss of the intraoperative reduction is reduced

Blood supply to the bone

No additional compression after locking

Periosteal blood supply will be preserved

Page 70: Implant screw plate

Unicortical Fixation

Conventional Plating

Small Small LoadLoad

Small Small LoadLoad

Screws have single point of

fixation

Screws have two points of fixationLocked Plating

Page 71: Implant screw plate

Principle of internal fixation using LCP :

1. 1st reduced the # as anatomical as possible2. Cortical screw should be used 1st in a fracture

fragment3. If the locking screw have been put, use of the

cortical screw in the same fragment without loosening and retightening of the locking screw is not recommended

4. If locking screw is used first avoid spinning of plate

5. Unicortical screws causes no loss of stability

Page 72: Implant screw plate
Page 73: Implant screw plate

6. Osteoporotic bones bicortical screws should be used.

7. In the comminuted # screw holes close to the fracture should be used to reduce stain.

8. In the fracture with small or no gap the immediate screw holes should be left unfilled to reduced the strain.

Page 74: Implant screw plate
Page 75: Implant screw plate

Indications :1. Osteoporotic #2. Periprosthetic #3. Multifragmentry #4. Delayed change from external fixation to internal

fixation.Advantages :1. Angular stability 2. Axial stability3. Plate contouring not required4. Less damage to the blood supply of bone5. Decrease infection because of submuscular

technique6. Less soft tissue damage

Page 76: Implant screw plate

HOW MANY SCREWS ? Hands-on experience suggests that, in the

humerus, screws grip seven cortices on each side of the fracture ; in the radius and the ulna, five; in the tibia, six, and in the femur, seven.Bones No. of

CorticesNo. of Holes

Type of Plate

Forearm 5 to 6 Cortex 6 holes Small 3.5

Humerus

7 to 8 Cortex 8 holes Narrow 4.5

Tibia 7 to 8 Cortex 7 holes Narrow 4.5

Femur 7 to 8 Cortex 8 holes Narrow 4.5

Clavicle 5 to 6 Cortex 6 holes` Small 3.5

Page 77: Implant screw plate

HOW CLOSE TO THE FRACTURE SITE?

A screw, as a result, should not be placed closer than one centimeter from the fracture line.

Page 78: Implant screw plate

Timing of Plate Removal, Recommendations for removal of plates in the lower limb :

Bone / Fracture Time after implantation in months Malleolar fractures 8-12 The tibial pilon 12-18 The tibial shaft 12-18 The tibial head 12-18

Page 79: Implant screw plate

The femoral condyles 12-24 The femoral shaft: Single plate, Double Plate 24-36 From month 18, in 2 steps ( Interval 06 months) Pertrochanteric and femoral neck fractures Upper

extremity 12-18 Optional Shaft of radius / ulna 24-28 Distal radius 8-12 Metacarpals 4-6

Page 80: Implant screw plate
Page 81: Implant screw plate

DIFFERENT AO SCREWS LARGE STANDARD SCREWS. 4.5 mm Cortex Screw

6.5 mm Cancellous Screw

Malleolar Screw 4.5

CANNULATED SCREW SYSTEM

6.5 Cannulated Screw

4.0 mm Cannulated Screw

3.5 Cannulated Screw

SMALL FRAGMENT SCREW 3.5 mm Cortical Screw

4.0 Canceleous Screw-Partially Threaded.-Fully Threaded

MINI SCREW 2.7 mm Cortex Screw

2.0 mm Cortex Screw

1.5 mm Cortex Screw

Page 82: Implant screw plate

Screw Core diameter

Thread

diameter

Pitch Drill bit for

gliding

hole

Drill bit for

thread

hole

Tap diameter

Large Standard Screws

7mm Cancellous Screw

4.5mm

7mm 2.75mm

4.5mm

7mm

6.5mm cancellous screw

3.5mm

6.5mm

2.7mm

3.2mm

4.5mm

6.5mm

4.5mm cancellous screw

4.5mm

3.1mm

1.75mm

3.2mm

4.5mm

4.5mm cortical

3mm 4.5 mm

1.75mm

4.5mm

3.2mm

4.5mm

Page 83: Implant screw plate

Small Fragment Screws

3.5mm cancellous screw

2.5mm

3.5mm

1.25mm

2.7mm

3.5mm

4mm Cancellous screw

1.9mm

4mm 1.75mm

2.5mm

4mm

3.5mm Cortex Screw

2.4mm

3.5mm

1.25mm

3.5mm

2.5mm

3.5mm

Mini Fragment Screws

2.7mm Cortex Screw

1.9mm

2.7mm

1mm 2.7mm

2mm 2.7mm

2mm Cortex Screw

1.3mm

2mm 0.6mm

2mm 1.5mm

2mm

1.5mm Cortex Screw

1mm 1.5mm

0.5mm

1.5mm

1.1mm

1.5mm