Page 1
Simulation of Implant dentar personalizat BioMicron TransilvaniaDate: 7 iulie 2014Designer: SolidworksStudy name: Study 1Analysis type: Static
Table of ContentsDescription...............................................1
Assumptions.............................................2
Model Information.....................................2
Study Properties.......................................4
Units.........................................................4
Material Properties...................................5
Loads and Fixtures...................................6
Connector Definitions...............................7
Contact Information..................................8
Mesh Information....................................10
Sensor Details.........................................12
Resultant Forces.....................................12
Beams....................................................12
Study Results..........................................13
Conclusion..............................................17
DescriptionNo Data
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 1
Page 2
Assumptions
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 2
Page 3
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 3
Page 4
Model Information
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 4
Page 5
Model name: Implant dentar personalizat BioMicron Transilvania
Current Configuration: Default
Solid BodiesDocument Name and
ReferenceTreated As Volumetric Properties Document Path/Date
ModifiedCut-Revolve1
Solid Body
Mass:0.000311209 kgVolume:6.90041e-008
m^3Density:4510 kg/m^3Weight:0.00304984 N
C:\Users\A&C\Desktop\DIPLOMA
POGACEAN_DENTAL IMPLANT\SORIN
MATEI_ARHIVA 2\AU EI 4,1- 12.SLDPRT
Jul 06 19:15:27 2014
Cut-Extrude2
Solid Body
Mass:6.5389e-005 kgVolume:1.44987e-008
m^3Density:4510 kg/m^3
Weight:0.000640812 N
C:\Users\A&C\Desktop\DIPLOMA
POGACEAN_DENTAL IMPLANT\SORIN
MATEI_ARHIVA 2\AU EI 4,1.SLDPRT
Jul 06 15:36:32 2014
Cut-Revolve1
Solid Body
Mass:8.36985e-005 kgVolume:1.88988e-008
m^3Density:4428.78 kg/m^3Weight:0.000820246 N
C:\Users\A&C\Desktop\DIPLOMA
POGACEAN_DENTAL IMPLANT\SORIN
MATEI_ARHIVA 2\BM-
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 5
Page 6
Study PropertiesStudy name Study 1
Analysis type Static
Mesh type Solid Mesh
Thermal Effect: On
Thermal option Include temperature loads
Zero strain temperature 298 Kelvin
Include fluid pressure effects from SolidWorks Flow Simulation
Off
Solver type FFEPlus
Inplane Effect: Off
Soft Spring: Off
Inertial Relief: Off
Incompatible bonding options Automatic
Large displacement Off
Compute free body forces On
Friction Off
Use Adaptive Method: Off
Result folder SolidWorks document (C:\Users\A&C\Desktop\DIPLOMA POGACEAN_DENTAL IMPLANT\SORIN MATEI_ARHIVA 2)
UnitsUnit system: SI (MKS)
Length/Displacement mm
Temperature Kelvin
Angular velocity Rad/sec
Pressure/Stress N/m^2
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 6
Page 7
Material PropertiesModel Reference Properties Components
Name: Commercially Pure CP-Ti UNS R50700 Grade 4 (SS)
Model type: Linear Elastic Isotropic
Default failure criterion:
Max von Mises Stress
Yield strength: 5e+008 N/m^2Tensile strength: 5.5e+008 N/m^2Elastic modulus: 1.05e+011 N/m^2Poisson's ratio: 0.37
Mass density: 4510 kg/m^3Shear modulus: 4.5e+010 N/m^2
Thermal expansion
coefficient:
9e-006 /Kelvin
SolidBody 1(Cut-Revolve1)(AU EI 4,1- 12-1),SolidBody 1(Cut-Extrude2)(AU EI 4,1-1)
Curve Data:N/A
Name: Ti-6Al-4VSolution treated and aged (SS)
Model type: Linear Elastic Isotropic
Default failure criterion:
Max von Mises Stress
Yield strength: 8.27371e+008 N/m^2
Tensile strength: 1.05e+009 N/m^2Elastic modulus: 1.048e+011 N/m^2Poisson's ratio: 0.31
Mass density: 4428.78 kg/m^3Shear modulus: 4.10238e+010
N/m^2Thermal
expansion coefficient:
9e-006 /Kelvin
SolidBody 1(Cut-Revolve1)(BM-0056-1)
Curve Data:N/A
Name: Ti-6Al-4VSolution treated and aged (SS)
Model type: Linear Elastic Isotropic
Default failure criterion:
Unknown
Yield strength: 8.27371e+008 N/m^2
SolidBody 1(Chamfer1)(Bont 15 AU-VL 1-1)
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 7
Page 8
Tensile strength: 1.05e+009 N/m^2Elastic modulus: 1.048e+011 N/m^2Poisson's ratio: 0.31
Mass density: 4428.78 kg/m^3Shear modulus: 4.10238e+010
N/m^2Thermal
expansion coefficient:
9e-006 /Kelvin
Curve Data:N/A
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 8
Page 9
Loads and FixturesFixture name
Fixture Image Fixture Details
Fixed-1
Entities: 1 edge(s), 23 face(s)Type: Fixed Geometry
Resultant ForcesComponents X Y Z Resultant
Reaction force(N) 0.0163343 0.0214571 -349.872 349.872Reaction
Moment(N·m)0 0 0 0
Fixed-2
Entities: 11 face(s)Type: Fixed Geometry
Resultant ForcesComponents X Y Z Resultant
Reaction force(N) 0.000126843 0.000350551 -0.000837069 0.000916329Reaction
Moment(N·m)0 0 0 0
Load name Load Image Load Details
Gravity-1
Reference: Top PlaneValues: 0 0 -9.81
Units: SI
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 9
Page 10
Load name Load Image Load Details
Force-1
Entities: 1 face(s), 1 plane(s)Type: Apply normal force
Value: 350 N
Connector DefinitionsNo Data
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 10
Page 11
Contact Information
Contact Contact Image Contact Properties
Contact Set-1
Type: Bonded contact pair
Entites: 2 face(s)
Contact Set-2
Type: Bonded contact pair
Entites: 2 face(s)
Contact Set-4
Type: Bonded contact pair
Entites: 2 face(s)
Contact Set-5
Type: Bonded contact pair
Entites: 2 face(s)
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 11
Page 12
Contact Contact Image Contact Properties
Contact Set-6
Type: Bonded contact pair
Entites: 2 face(s)
Contact Set-7
Type: Bonded contact pair
Entites: 2 face(s)
Contact Set-8
Type: Bonded contact pair
Entites: 2 face(s)
Global Contact
Type: BondedComponents: 1
component(s)Options: Compatible
mesh
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 12
Page 13
Mesh InformationMesh type Solid Mesh
Mesher Used: Curvature based mesh
Jacobian points 4 Points
Maximum element size 0.592353 mm
Minimum element size 0.118471 mm
Mesh Quality High
Remesh failed parts with incompatible mesh
Off
Mesh Information - DetailsTotal Nodes 1018738
Total Elements 703194
Maximum Aspect Ratio 193.36
% of elements with Aspect Ratio < 3 99
% of elements with Aspect Ratio > 10 0.244
% of distorted elements(Jacobian) 0.00356
Time to complete mesh(hh;mm;ss): 00:00:51
Computer name: AC
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 13
Page 14
Mesh Control Information:
Mesh Control Name Mesh Control Image Mesh Control Details
Control-1
Entities: 1 Solid Body (s)Units: mmSize: 0.141211
Ratio: 1.5
Control-2
Entities: 1 Solid Body (s)Units: mmSize: 0.151543
Ratio: 1.5
Control-3
Entities: 1 Solid Body (s)Units: mmSize: 0.151543
Ratio: 1.5
Control-4
Entities: 1 Solid Body (s)Units: mmSize: 0.10543
Ratio: 1.5
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 14
Page 15
Sensor DetailsNo Data
Resultant Forces
Reaction ForcesSelection set
Units Sum XSum Y Sum Z Resultant
Entire Model N 0.0164535 0.0216554 -349.872 349.872
Reaction MomentsSelection set
Units Sum XSum Y Sum Z Resultant
Entire Model N·m 0 0 0 0
BeamsNo Data
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 15
Page 16
Study Results
Name Type Min MaxStress1 VON: von Mises Stress 0.00151846 N/m^2
Node: 433032.40191e+009 N/m^2Node: 516497
Implant dentar personalizat BioMicron Transilvania-Study 1-Stress-Stress1
Name Type Min MaxDisplacement1 URES: Resultant Displacement 0 mm
Node: 250.0937302 mmNode: 438126
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 16
Page 17
Name Type Min Max
Implant dentar personalizat BioMicron Transilvania-Study 1-Displacement-Displacement1
Name Type Min MaxStrain1 ESTRN: Equivalent Strain 6.62202e-012
Element: 1374420.0163425Element: 295982
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 17
Page 18
Name Type Min Max
Implant dentar personalizat BioMicron Transilvania-Study 1-Strain-Strain1
Name TypeDisplacement1{1} Deformed Shape
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 18
Page 19
Implant dentar personalizat BioMicron Transilvania-Study 1-Displacement-Displacement1{1}
Name Type Min MaxStress2 VON: von Mises Stress 0.00151846 N/m^2
Node: 433032.40191e+009 N/m^2Node: 516497
Implant dentar personalizat BioMicron Transilvania-Study 1-Stress-Stress2
Name Type Min MaxStress3 VON: von Mises Stress 0.00151846 N/m^2
Node: 433032.40191e+009 N/m^2Node: 516497
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 19
Page 20
Name Type Min Max
Implant dentar personalizat BioMicron Transilvania-Study 1-Stress-Stress3
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 20
Page 21
Image-1
Conclusion
Analyzed with SolidWorks SimulationSimulation of Implant dentar personalizat
BioMicron Transilvania 21