Top Banner
1 Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry Jonathan Williams Waters Corporation, Milford, MA, USA INTRODUCTION The global dispersion of hemoglobin variants through population migration has led to the requirement for their identification. An effective mass spectrometry- based procedure involves analysis of the intact globin chains in diluted blood to detect the variant through mass anomalies; this is followed by location of the variant amino acid residue by direct analysis of the enzymatically digested globins without prior chromatographic separation. 1 The objective of this part of the procedure is to identify the variant peptide, i.e. the peptide that contains the mutant amino acid, using the mass difference of the variant from the normal peptide obtained in the first step as a guide. Generally, this is straightforward, since there are very few cases of interference between the ions in the mass spectrum. Once observed, the variant peptide can be sequenced by tandem MS in order to positively identify or confirm the mutation. On occasion, however, the variant peptide ion occurs in a part of the mass spectrum where there is interference from another ion, which makes detection and identification difficult or impossible. Chromatographically separating the tryptic peptides using reversed phase LC/MS would potentially offer a solution. Another solution is to use ion mobility coupled with mass spectrometry (IM-MS) to separate the variant peptide ion from the interfering ion. This would allow the variant peptides to be observed and sequenced by collision induced dissociation following IM. 2 In this application note, the doubly-charged tryptic peptide from a variant of low abundance occurred at the same m/z value as a singly-charged interfering ion. Here, we demonstrate fragmentation of the triply-charged tryptic peptide ions using electron transfer dissociation (ETD), (as opposed to collision induced dissociation (CID) of the triply-charged ions) for sequence information. This technique is shown for the identification of an alpha-chain variant having normal clinical presentation, namely Hb Riccarton. WATERS SOLUTIONS SYNAPT ® G2 HDMS™ T-Wave™ KEY WORDS Hemoglobin, ETD, Riccarton, hemoglobin APPLICATION BENEFITS Occasionally, there are instances where a hemoglobin variant peptide ion occurs in a part of the digest spectrum where an interference from another ion makes identification difficult. If variant ions with three or more charges are detected, Electron Transfer Dissociation (ETD) of these ions has the potential to aid in the identification of human Hb variants, and thus provide simple and easy to interpret tandem mass spectra.
5

Identification of Human Hemoglobin Protein Variants Using ... · Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass

Oct 11, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Identification of Human Hemoglobin Protein Variants Using ... · Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass

1

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass SpectrometryJonathan WilliamsWaters Corporation, Milford, MA, USA

IN T RO DU C T IO N

The global dispersion of hemoglobin variants through population migration has

led to the requirement for their identification. An effective mass spectrometry-

based procedure involves analysis of the intact globin chains in diluted blood

to detect the variant through mass anomalies; this is followed by location of

the variant amino acid residue by direct analysis of the enzymatically digested

globins without prior chromatographic separation.1 The objective of this part of

the procedure is to identify the variant peptide, i.e. the peptide that contains the

mutant amino acid, using the mass difference of the variant from the normal

peptide obtained in the first step as a guide. Generally, this is straightforward,

since there are very few cases of interference between the ions in the mass

spectrum. Once observed, the variant peptide can be sequenced by tandem

MS in order to positively identify or confirm the mutation.

On occasion, however, the variant peptide ion occurs in a part of the mass

spectrum where there is interference from another ion, which makes detection

and identification difficult or impossible. Chromatographically separating the

tryptic peptides using reversed phase LC/MS would potentially offer a solution.

Another solution is to use ion mobility coupled with mass spectrometry (IM-MS)

to separate the variant peptide ion from the interfering ion. This would allow the

variant peptides to be observed and sequenced by collision induced dissociation

following IM.2

In this application note, the doubly-charged tryptic peptide from a variant of

low abundance occurred at the same m/z value as a singly-charged interfering

ion. Here, we demonstrate fragmentation of the triply-charged tryptic peptide

ions using electron transfer dissociation (ETD), (as opposed to collision induced

dissociation (CID) of the triply-charged ions) for sequence information. This

technique is shown for the identification of an alpha-chain variant having normal

clinical presentation, namely Hb Riccarton.WAT E R S SO LU T IO NS

SYNAPT® G2 HDMS™

T-Wave™

K E Y W O R D S

Hemoglobin, ETD, Riccarton, hemoglobin

A P P L I C AT IO N B E N E F I T S

Occasionally, there are instances where a

hemoglobin variant peptide ion occurs in a part

of the digest spectrum where an interference

from another ion makes identification difficult.

If variant ions with three or more charges are

detected, Electron Transfer Dissociation (ETD)

of these ions has the potential to aid in the

identification of human Hb variants, and thus

provide simple and easy to interpret tandem

mass spectra.

Page 2: Identification of Human Hemoglobin Protein Variants Using ... · Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass

2Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

E X P E R IM E N TA L

ETD-MS conditions

MS system: SYNAPT G2

Ionization mode: ESI positive

Capillary voltage: 3.5 kV (for positive)

Cone voltage: 30 V

Desolvation temp.: 200 ˚C

Source temp.: 120 ˚C

Ionization mode: Glow discharge

negative

Glow discharge: 55 µA (for negative)

Reagent: para-nitrotoluene

(m/z 137).

Sample preparation

A blood sample was submitted for investigation by mass spectrometry because

an abnormality had been detected during a routine screen in a hematology

laboratory. For mass spectral analysis, the sample was prepared and analyzed

as described previously.1 The sample was diluted 50-fold with water, denatured,

and digested with trypsin for 30 minutes. The resulting mixture of peptides was

diluted a further 10-fold (final solution in 50.0% aqueous acetonitrile containing

0.2% formic acid) for direct infusion into the ESI source of the mass spectrometer.

Mass spectrometry

ESI-MS was performed on a SYNAPT G2 hybrid quadrupole / ion mobility /

oa-ToF mass spectrometer fitted with ETD functionality. In brief, the instrument

comprises three consecutive, gas filled, travelling wave (T-Wave) RF stacked

ring ion guides prior to the ToF mass analyzer. For ETD type fragmentation, a

sub-ambient pressure (~2 mbar) glow discharge anion source was used to fill

the Trap T-Wave cell with quadrupole mass selected ETD reagent anions formed

from para-nitrotoluene (m/z 137). During acquisition, the source polarity and

quadrupole set mass were switched to allow triply-charged peptides formed

from ESI to interact with stored reagent anions in the Trap T-Wave. For efficient

ETD within the Trap T-Wave region, the helium bath gas was set to a pressure of

5 x 10-2 mbar. The Transfer T-Wave cell was pressurized to 5 x 10-3 mbar with

argon. The T-Wave speed and amplitude were set to 300 m/sec and 0.2 V

respectively. Solutions were infused at a flow rate of 5 µL/min.

R E SU LT S A N D D IS C U S S IO N

A low abundance hemoglobin variant was detected during a routine hospital

screening by cation exchange-HPLC and it was submitted for identification by MS.

Analysis of the intact chains showed that the variant was 30 Da heavier than the

normal α-chain. There are five mutations that, on genetic grounds, are likely to

increase the mass by 30 Da; either Ala→Thr, Arg→Trp, Gly→Ser, Thr→Met,

or Val→Glu.

Page 3: Identification of Human Hemoglobin Protein Variants Using ... · Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass

3Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

Figure 1. Part spectra from 30-minute tryptic digests of a normal Hb (top) and the variant containing Hb (bottom).

Figure 1 shows the analysis of the 30-minute digests of a normal control (top spectrum) and the sample

containing the variant (bottom spectrum). The analysis revealed that the αT62+ ion at m/z 932.5 in the variant

spectrum had an interfering ion (normal βT2+ ion) at the expected m/z of the variant αT62+ ion. This peak,

m/z 15 higher than the αT62+ ion, suggested two possible mutations, either α51Gly→Ser or α53Ala→Thr,

as shown in Figure 1.

m/z914 916 918 920 922 924 926 928 930 932 934 936 938 940 942

%

0

100

m/z914 916 918 920 922 924 926 928 930 932 934 936 938 940 942

%

0

100 917.9917.5

932.5

918.4

918.9928.9928.4

919.5921.5

929.4929.9

933.5

934.5 936.4935.5

936.9939.4 939.9

918.0917.5

932.5

918.5

932.5919.0

928.4

919.5922.5

928.9929.4

930.0

933.5

934.5936.9936.4

935.5 937.4 939.9 940.4

NormalαT62+

NormalαT62+

Normal αT6 peptideα 41 56

T Y F P H F D L S H G S A Q V K

S TNormalβT2+

Variant αT62+

βT2+

As a consequence of the interference from the intense singly-charged ion, detection and subsequent

identification of the variant through MS/MS of the doubly-charged ion would have been difficult. In order to

distinguish these two possibilities, the normal and variant αT63+ ions were sequenced by ETD since there were

no interferences observed for these ions. Accurate mass alone would not be sufficient to determine if the

mutation was Gly→Ser or Ala→Thr from the digest mass spectrum since the accurate mass difference between

these two possibilities is 30.0106 Da in both instances; hence, the need for tandem mass spectrometry.

Page 4: Identification of Human Hemoglobin Protein Variants Using ... · Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass

4Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

Figure 2 (top) shows the ETD mass spectrum of the peak at m/z 612, which is consistent with the sequence for

the normal αT63+ ion. Figure 2 (bottom) shows the ETD mass spectrum for the peak at m/z 622 for the putative

variant αT63+ ion. The 30 Da mass-increase between the normal and variant at z'6 and c"11 and all subsequent

z' and c" ions identifies the mutation as α51Gly→Ser, Hb Riccarton.

m/z400 600 800 1000 1200 1400 1600 1800

%

0

100

m/z400 600 800 1000 1200 1400 1600 1800

%

0

100 918.0

612.3612.0

z'5516.3

c''3z'4

429.3z'3

358.2

359.2430.3z'6

573.3

917.5

612.6

614.2

c''5;663.3 c''6810.4

z'9910.5

918.5

1835.9

919.0

1833.9

919.5 c''111319.6

z'111172.6c''7

925.4

z'101025.5 1554.8

c''12z'13

1406.71478.7 1818.91705.8

1836.9

1837.9

1838.9

933.0

622.3622.0

z'3358.2

z'6603.3

z'5516.3

359.2 430.3 517.3

932.5

622.6

c''5663.3

c''6810.4

664.3

c''7925.4

924.5

1865.9933.5

934.01863.9

934.5 z'101055.5941.5

1848.91584.8c''10

1262.6z'11

1202.6 c''131507.7

c''111349.6 1735.91600.8

1866.9

1867.9

1869.0

c"101262.6

Precursor ionNormalαT63+

Precursor ionVariantαT63+

Normal αT6 peptideα 41 56

T Y F P H F D L S H G S A Q V K

S

*

*

*

*

Δ +30.0 Da

*

*

*

[M+2H]2+/[M+3H]2+

[M+H]+/[M+2H]+ /[M+3H]+

•••

Figure 2. ETD product ion spectra from normal (top) and variant (bottom) αT63+ ions. The 30 Da mass-increase at z'6 and c"11 in identifies the mutation as α51Gly→Ser, Hb Riccarton.

Page 5: Identification of Human Hemoglobin Protein Variants Using ... · Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass

Waters Corporation34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com

CO N C LU S IO NS■■ It has been demonstrated that when isobaric interferences are

observed and precursor ion selection for CID is difficult, the

selection of higher charge states for sequence information

using ETD has significant potential for the identification of

hemoglobin variants.

■■ The mass spectra obtained by selection of triply-charged ions

for fragmentation using ETD can often be less complex and

easier to interpret than the corresponding CID spectra of the

same charge state.

Waters and SYNAPT are registered trademarks of Waters Corporation. T-Wave, HDMS, and T he Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.

©2012 Waters Corporation. Produced in the U.S.A. April 2012 720004291en AG-PDF

References

1. Wild BJ, Green BN, Cooper EK, Lalloz MRA, Stephens AD, Layton DM. Blood Cells Mol. Dis. 2001; 27, 691.

2. Williams JP, Giles K, Green BN, Scrivens JH, Bateman RH. Rapid Commun. Mass Spectrom. 2008; 22: 3179.