  • Hybrid Brain-Computer Interface for Biomedical

    Cyber-Physical System Application using Wireless

    Embedded EEG Systems

    Rifai Chai*, Ganesh R. Naik, Sai Ho Ling, Hung T. Nguyen

    Centre for Health Technologies, Faculty of Engineering and Information Technology,

    University of Technology, Sydney, NSW 2007, Australia

    Email: [email protected]*, [email protected], [email protected], [email protected]

    * Corresponding Author

    Rifai Chai

    [email protected]

    Abstract

    Background: One of the key challenges of the biomedical cyber-physical system (CPS) is to

    combine cognitive neuroscience with the integration of physical systems to assist people with

    disabilities. Electroencephalography (EEG) has been explored as a non-invasive method of

    providing assistive technology by using brain electrical signals.

    Methods: This paper presents a unique prototype of a hybrid brain computer interface (BCI)

    which senses a combination classification of mental task, steady state visual evoked potential

    (SSVEP) and eyes closed detection using only two EEG channels. In addition, a

    microcontroller based head-mounted battery-operated wireless EEG sensor combined with a

    separate embedded system is used to enhance portability, convenience and cost effectiveness.

    This experiment has been conducted with five healthy participants and five patients with

    tetraplegia.

    Results: Generally, the results show comparable classification accuracies between healthy

    subjects and tetraplegia patients. For the offline artificial neural network classification for the

    target group of patients with tetraplegia, the hybrid BCI system combines three mental tasks,

    three SSVEP frequencies and eyes closed, with average classification accuracy at 74% and

    average information transfer rate (ITR) of the system of 27 bits per minute. For the real-time

    testing of the intentional signal on patients with tetraplegia, the average success rate of

    detection is 70% and the speed of detection varies from 2 to 4 seconds.

    Keywords: Brain-computer interface, cyber physical system, electroencephalography,

    artificial neural network, hybrid system, embedded system.

  • Background

    The new emerging frontier of technology modeling, cyber-physical systems (CPS), is

    related to the generation of new systems with the integration of the physical systems and real-

    time computational intelligence capabilities. This creates a fully intelligent system to tackle a

    variety of application fields such as robotics, intelligent buildings, biomedical and healthcare,

    energy, manufacturing control systems, and avionics [1, 2].

    In biomedical CPS fields, especially in cognitive neuroscience, there has been significant

    interest in creating assistive technologies to improve the quality of life of individuals with

    disability in areas such as system integration with motor prosthesis, wheelchair control,

    environmental control and other applications [3-5]. One of the important challenges in CPS is

    human-machine interfacing. An integrated cyber-physical system should meet the real-time

    operational requirements including portability, reliability and predictability. A dedicated

    embedded system together with wireless technology would be an ideal platform for such an

    integrated real-time CPS [2, 6, 7].

    In the application of wheelchair control for people with disabilities, various hands-free

    technologies have been used to replace the joystick including sip-and-puff [8], chin controller

    [9], muscle based system [10], voice recognition [11], tongue controller [12] and head

    movement systems [13]. These technologies have their own benefits and drawbacks. In

    practical situations, the user may feel uncomfortable with the operation of a sip-and-puff or a

    chin or tongue controller. Noisy environments can be problematic for voice recognition

    systems. Muscle and head movement technology are targeted for disabled individuals who

    are still able to provide the relevant body movement for real-time control. Certain

    neurological conditions such as amyotrophic lateral sclerosis (ALS), cervical spinal cord

    injury and brain stem stroke may lead to severe motor paralysis and mobility restriction

    referred to as a locked-in syndrome [14]. A brain computer interface (BCI) could be used as

    an alternative solution for these disabled individuals by converting brain activities to provide

    a means of control and communication [3, 5, 11, 15].

    The acquisition techniques which are available in BCI systems can basically be classified

    into invasive and non-invasive brain measurements. The invasive methods include intra-

    cortical recording using microelectrodes and electrocorticography (ECoG). Although these

    methods could provide a better temporal resolution and quality, they have drawbacks such as

    the risk of infection and scarring post-surgery [16]. The electroencephalography (EEG) as a

    non-invasive approach has good temporal resolution, portability and low cost compared to

    other non-invasive BCI methods, such as functional near infrared spectroscopy (fNIRS) and

    functional magnetic response imaging (fMRI) [17, 18].

    The applications of BCI–EEG as viewed from the aspect of mental strategies can be

    divided into either selective attention or spontaneous mental signal methods. The P300 [19,

    20] technique and the steady state visual evoked potential (SSVEP) [4, 21] technique are

    examples of the selective attention method. For these the user needs to concentrate on

    external stimuli that flash in succession (P300) or continuously in a certain frequency

    (SSVEP). The BCI speller system is a good example of an application using selective

    attention. BCI systems relying on spontaneous mental signals generated voluntarily by the

    user may include self-regulation of the slow cortical potential (SCP) [22], control of the

    sensory motor rhythm (SMR) [23, 24] and the event-related

    desynchronization/synchronization (ERD/ERS) [25, 26] which focuses on the motor imagery

  • area such as by imagining hand, foot or tongue movements.

    Although the motor imagery method provides a good option for the BCI applications, there

    is a possibility that individuals who have been paralyzed or are amputated for a number of

    years may not be able to perform motor imagery mental tasks very well [27, 28].

    Several researchers have used other non-motor imagery task BCI based on mental task

    imaginations. They used six EEG channels with electrodes placed on the scalp positions

    central (C3, C4), parietal (P3, P4) and occipital (O1, O2). The mental tasks chosen included

    baseline, multiplication, letter composing, figure 3-D rotation and counting [5, 27]. For a

    practical system, the number of channels needs to be reduced. For example some researchers

    have used different combinations two and three channels [5, 29]. Our previous study [5] has

    shown it is possible to reduce from 6 channels to 2 channels without losing accuracy. As a

    result, in this study, we are focusing on using only two EEG channels for portability and ease

    of use in a practical system. Looking from the brain function point of view, it has been found

    that the parietal lobe shows significant activity during mental arithmetic calculation, language

    and writing skills especially on the left hemisphere [30-32]. Another study showed mental

    figure rotation created activity in both left and right parietal lobes [33]. A visual task has been

    shown to produce activity more on the right occipital area [34]. Therefore parietal and

    occipital areas provide significant features for BCI purposes. Consequently, this paper

    develops and provides experiment results for a system using two channel wireless EEG

    placed on the left parietal (P3) and the right occipital (O2) lobes only.

    In addition, it has been known for EEG measurement that there is increased amplitude of

    alpha band (8-13 Hz) during eyes closed action of between 1 to 5 seconds. This eyes closed

    phenomenon can be found in normal person and individuals with disabilities and therefore it

    is reliable enough to be used as a mind switch to create a hand free control system [35].

    This study presents the application of biomedical CPS in the integration of the embedded

    system with the human neural physical system measured by wireless EEG. A hybrid

    classification system is proposed using a combination of wireless EEG sensing methods

    which include BCI based mental task, BCI based SSVEP and eyes closed detection. With the

    purpose of developing a practical real-time biomedical CPS application and for the rapid

    development of embedded system and wireless technology, this paper presents a

    microcontroller based CPS system consisting of a separate head-mounted battery-operated

    wireless EEG sensor communicating with a main embedded system to provide maximum

    portability, convenience and cost effectiveness. The developed EEG used in this study is a

    two channel only wireless EEG with the electrodes placed at the back of the head of parietal

    and occipital lobes.

    For the classification algorithm, the linear and non-linear methods have been explored in

    the EEG signal classification [36]. As an EEG signal is multi-dimensional, a non-linear

    method namely artificial neural network (ANN) is investigated. The ANN has been widely

    used in biomedical applications or other engineering applications particularly for

    classification algorithms [5, 37-39].

    Although many studies have reported the implementation of an EEG-based BCI system,

    there still many subjects unable to use particular mental strategy of EEG-based BCI, known

    as BCI illiteracy phenomena [40]. As a result, hybrid EEG-based BCI has been developed

    using the combination of different mental strategies [41]. Currently, most of the hybrid EEG-

    based BCIs use the combination of motor imagery with SSVEP [40] and P300 with SSVEP

  • [42-45].

    Essentially, this paper discusses the development of the novel hybrid BCI by combining

    mental task (non-motor imagery), SSVEP and eyes closed tasks that have not been explored

    in previous studies. The use of mental task (non-motor imagery) here is due to the selective

    motor imagery task defects in patients with severe disabilities. Also, individuals who have

    been paralyzed for several years may be unable to perform motor imagery tasks effectively.

    This proposed technique can provide an alternative solution for subjects who cannot use the

    other combination of hybrid BCI properly. In addition, compared to other hybrid BCI

    research that uses several EEG channels, this paper proposes the development of hybrid

    EEG-based BCI using only two EEG channels.

    Methods

    Prototype Hybrid BCI System

    The proposed hybrid BCI system for biomedical CPS application consists of a combination

    of mental task based-BCI, SSVEP based-BCI and eyes closed detection using a two channel

    only wireless EEG and embedded systems. The block diagram in Figure 1 shows three main

    modules of the prototype which uses embedded system controllers as functional blocks. The

    first module is a wireless EEG system which comprises of analog and digital components.

    The analog section consists of amplifiers and filter circuits. A combined microcontroller and

    2.4 GHz RF transceiver, (Nordic Semiconductor) handles the digital section and transfers

    data wirelessly. The same wireless microcontroller on the receiver captures and sends the

    data to a second module with the main microcontroller MCF5213 Coldfire® (Freescale™) for

    signal processing, feature extraction and classification. The third module, an LED stimulator

    box, consists of the Atmega128 (Atmel®) microcontroller to control three LEDs for SSVEP

    based BCI system.

    Wireless EEG, Main Controller and Stimulator Box

    The EEG amplifier specification requires a high common mode rejection ratio (CMRR)

    above 80 dB to tolerate the interfering noise into the system. The amplifier should also be

    able to detect EEG signals within the range of 5-300 µVolts [46].

    The proposed wireless EEG, as shown in Figure 2, is divided into analog and digital

    sections. The EEG is based on two channel bipolar EEG configuration. Each channel has

    non-inverting inputs (CH1+ and CH2+), inverting inputs (CH1- and CH2-) and a reference

    input electrode. The amplifier design is based on a DC coupled amplifier which consists of

    two stages of amplifiers. A precision current mode instrumentation amplifier (In-Amp)

    AD8553 is used at the first stage which internally contains a voltage to current amplifier,

    current to voltage amplifier and a high precision auto-zero amplifier. To accommodate the

    DC offset from the electrodes which could saturate the amplifier, the gain of In-Amp is set at

    a low value of 10. This is followed by a passive RC high pass filter circuit to remove the DC

    offset. The second stage amplifier also uses op-amp OPA333 to form a non-inverting

    amplifier circuit with an adjustable gain up to 1000 using a potentiometer.

    To minimize the radio frequency interference (RFI), a differential low pass filter circuit for

  • RF attenuation is attached. An additional filter is provided to the second stage of the amplifier

    which has an active low pass filter configuration. At the end of the analog block before

    connection to the microcontroller, an anti-aliasing filter is added. The bandwidth of these

    filters is 1.5 kHz. The total noise measurement is 3.5 µVolts referring to the input.

    The analog to digital converter (ADC) has 12-bit resolution and is configured in differential

    mode for an improved common mode rejection. For the further noise rejection, four ADC

    samples are taken and then averaged for each reading. This is particularly effective in

    removing internally generated noise by the microcontroller. The ADC is configured in

    differential mode for improved common mode with the least-significant bit of the ADC

    calculated as follows:

    𝐿𝑆𝐵 = 𝐹𝑆

    2𝑛 (1)

    where LSB is least- significant bit, FS is full scale analog input, n is the number of bit

    resolution of ADC, in this case n=12 . The ADC setting has configurations which include

    internal reference of 1.2V. In differential mode of ADC, the input range swings from –Vref/2

    to +Vref/2 which is equal to -0.6 to +0.6 V, so the total full scale (FS) is to 1.2 Volts. LSB

    can be calculated by using equation (1) which gives a result of 293 µVolts (1.2 V/ 4096). So

    theoretically, the total gain was 2000 so as to get the input resolution of 0.15 µV (293 µVolts

    / 2000) for 1 LSB. The actual measurement of total noise as shown in Table 1 for the

    developed wireless EEG was 3.5 µV. According to [46] the EEG signals are within the range

    of 5-300 µVolts, so the actual noise measured still meets the requirement for EEG signal

    measurement.

    The nRF24LE1™ microcontroller uses the real time clock (RTC) serves as the main

    ticking of the 256 Hz sampling rate of the system and provides wake up functionality for

    power saving mode. Most of the routines are handled by an interrupt service routine to create

    a real-time data acquisition. When the timer interrupt has elapsed, it activates an ADC

    interrupt. The raw ADC value is transferred via RF 2.4 GHz after attaching it to the transfer

    protocol. As soon as the wireless receiver detects any incoming data from the interrupt RF

    routine and the protocol is matched, data is forwarded to a UART for connection to a PC via

    RS-232 serial line and simultaneously to a serial peripheral bus (SPI) port connected to the

    main embedded system. The main embedded system using 32 bit MCF5213 Coldfire® from

    FreescaleTM

    microcontroller which implements the µC/OS real-time operating system

    (RTOS) which runs several application tasks simultaneously.

    The LED stimulation box has three LEDs that flicker at frequencies 6Hz, 13Hz and 16Hz.

    A microcontroller Atmega128 (Atmel®) controls the three frequencies stimulus using three

    separate interrupt timer routines. In BCI, the good frequency stimulator response is between

    5Hz and 20Hz [47]. The chosen three frequencies are inside this range, excluding the alpha

    band (8-12Hz).

    Computational Intelligence

    In the main microcontroller, a data pre-processor extracts the previous 1s window of EEG

    channel readings every quarter second for processing by the features extraction algorithm. As

    a result, 1s window data set with some overlap between data sets is obtained every 0.25 sec.

    An overall 10 second data window is used resulting in 37 overlapping segments. To further

    improve signal quality, the data sets are then passed through two digital signal processing

    (DSP) filters: first, a moving average filter of 3 data samples width to smooth the signal then

  • a second order Butterworth band-pass filter with the bandwidth 0.5Hz to 100Hz.

    A power spectral density (PSD) analysis is applied for the features extraction method for

    mental task based BCI, SSVEP based BCI and eyes closed task. This is done by squaring a

    256 point fast Fourier transform (FFT) to every 1s width data set and converting into the PSD

    value in frequency bands of EEG: δ (0-3Hz), θ (4-7Hz), α (8-13Hz) and β (14-30Hz). The δ

    rhythm is not used due to the low frequency noise such as noise generated from ocular

    artifact and the γ rhythm is also discarded. As a result, only 27 PSD value (θ, α, β) are formed

    in each channel and 54 PSD value for two channels.

    This study utilizes a multi-layer feed forward back-propagation neural network with one

    hidden layer network as shown in Figure 3. A log-sigmoid function was assigned as the

    activation function which provides data values between one and zero. As a result, prior to the

    ANN the feature data value needs to be scaled to within the range of zero to one as follows:

    𝑥∗ = (𝑥 − 𝑥𝑚𝑖𝑛)/(𝑥𝑚𝑎𝑥 − 𝑥𝑚𝑖𝑛) (2)

    where x is the input EEG features (in this case PSD values) before normalization, x* is the

    input EEG features after normalization, xmin is the minimum value of the input EEG feature

    and xmax is the maximum value of the input EEG features.

    The output vector z and the k-th component zk are computed as follows:

    𝒛 = 𝑓(𝐖𝐱∗ + 𝑏) (3)

    𝑧𝑘(𝑥, 𝑤) = 𝑓1 (𝑏𝑘 + ∑ 𝑤𝑘𝑗 𝑓2 (𝑏𝑗 + ∑ 𝑤𝑗𝑖 𝑥𝑛∗

    𝑛

    𝑖=1

    )

    𝑚

    𝑗=1

    ) (4)

    where x* represents the input EEG feature vector, or in this case, PSD values from two EEG

    channels, W is the weight matrix vector, b is the scalar bias, z is the output vector or the

    three-class classification of hybrid BCI (three mental tasks or three SSVEP tasks), f1 and f2,

    are the activation functions for each of the two layers used in the ANN and log-sigmoid

    function used in this paper, n is the number of input nodes, m is the number of hidden nodes,

    k is the number of output nodes, wji is the weight to the hidden unit yj from input unit xi, wkj

    represents the weight to output unit zk from hidden unit yj. The biases are represented by bj and bk respectively.

    To accelerate the convergence of the error back propagation learning method, the ANN is

    trained by the Levenberg-Marquardt algorithm. The data is divided into three datasets:

    training, validation and testing set. The validation set is used as an early stopping method to

    make sure the ANN does not over-train.

    For the ANN training and classification, the total dataset for mental tasks, SSVEP tasks,

    eyes closed action for each task per subject is 222 units (37 overlapping segments × 6

    sessions), which is divided into training, validation and testing sets of 74 units each. The

    number of hidden neurons is varied between 4 to 20 neurons until the best classification was

    obtained with the minimum mean square error (MSE). A validation set is used for ANN

    training early stop determination.

    One of the error cycles performance is shown in Figure 10. Note that the error (MSE) of the

    training set decreased smoothly. The validation set dropped from the beginning up to 8

    epochs, then started to increase continuously. The training of the network was stopped at 8

    epochs because the validation performance started to increase continuously. This is used to

  • prevent over-training of the network. For the performance measurement, classification

    accuracy for multi-class classification is used in this paper as evaluation criteria as follows:

    𝐴𝑐𝑐𝑢𝑟𝑎𝑐𝑦 = 𝑇𝑃 + 𝑇𝑁

    𝑇𝑃 + 𝑇𝑁 + 𝐹𝑃 + 𝐹𝑁 (5)

    where TP (true positive) refers to a true detection of intentional control (IC) of BCI task

    being correctly classified as IC event, TN refer to no control (NC) of BCI task being

    classified as NC event. FP refers to false detected event or a NC of BCI task being classified

    as an IC event, FN is the IC of BCI task being classified as NC event.

    Hybrid system algorithms

    The algorithm used for the operation of the hybrid BCI system on the main embedded

    system controller as shown in Figure 4. The data packets from the first-in-first-out (FIFO)

    buffer are passed to the signal processing and classification methods. This is followed by the

    moving 1s window data, band-pass Butterworth filter and moving average filter.

    At the very first start of the routine as normal condition mode, eyes closed action for a

    window of 1s is used to decide whether the BCI control mode is used or not. Here, a window

    of 1s data is compared to a threshold for the eyes closed detection. Figure 5 provides the

    example of the threshold value for one of the subjects used. In this figure the time domain

    EEG signal of eyes closed swings from 2350 to 2800 of the amplitude or ADC values while

    the baseline (eyes open) signal has the value between 2400 to 2600 of the ADC value. A safe

    threshold value at 2700 can be applied for this case so that when the EEG signal is above the

    threshold, eyes closed signal is detected. If the EEG signal is below the threshold, eyes open

    is detected.

    After entering the BCI control mode, the eyes closed classification is used again to enter

    the hybrid BCI selection mode to select different BCI mental strategies. The FFT-PSD

    algorithm, normalization algorithm and ANN classification are performed to provide the

    classification of eyes closed task or opened task. The FFT radix-2 of 256 point was written in

    an assembler language in order to access the special accumulator register for a faster

    transform operation. The result of eyes closed classification is checked for the duration of the

    eyes closed action. If the eyes closed is detected for 1s duration then the system will enter to

    mental task-based BCI mode which enables the system to classify three mental tasks. In each

    selection of the BCI type, FFT-PSD, feature scaling/normalization algorithm and ANN is

    performed to provide the classification of selected BCI. The results of the BCI classification

    are mapped to the relevant commands. If the eyes closed is detected for 2s duration then the

    system will enter to SSVEP-based BCI mode which enables the system to classify three

    SSVEP frequencies. If the eyes closed is detected for more that 2s duration, it will turn the

    system off. Figure 6 shows information of modes for hybrid system algorithm which includes

    the BCI mode, Hybrid selection mode, mental tasks-based BCI mode and SSVEP-based BCI

    mode.

    Experimental Data Collection

    Prototype Testing

    An initial testing used an EEG simulator (MinSim300 -Netech) to inject a sinusoidal signal

    with an adjustable frequency and amplitude. The result shows that the wireless EEG has been

  • able to detect the simulation signal with minimum test signal at 10µV in different

    frequencies.

    The wireless EEG system uses a small square shaped printed circuit board (PCB) with an

    area of 36 mm2 shown in Table 1. There are two EEG channels which include two bipolar

    montages with five electrodes; two electrodes are used in each channel and one electrode is

    used for the reference. The experimental CMRR measurement was undertaken by connecting

    the input of the wireless EEG to a signal generator and the output to the oscilloscope. The

    signal generator peak to peak amplitude was set to full common mode voltage. The result of

    CMRR measurement has an average value of about 95 dB at 50 Hz. The gain can be set up to

    104. The current consumption of the wireless EEG with the power saving mode programming

    was 5 mA if the system was operated continuously. The noise measurement was 3.5 µV

    referring to the input. This current consumption of the wireless EEG is low enough to allow

    power to be supplied with just a coin cell battery for a battery life of up to 45 hours.

    The FFT assembly routine on the MCF5213 microcontroller has an execution time of 520

    µs and showed a comparable result with the FFT Matlab function in Fig. 7 with ± 3%

    tolerance.

    Data Collection

    This study was approved by the Human Research Ethics Committee at the University of

    Technology, Sydney. EEG signals generated from mental tasks, SSVEP tasks and eyes closed

    action were collected using the developed BCI prototype from five healthy participants and

    five patients with high level spinal cord injury (SCI). Healthy participants have ages between

    25 and 35 years. The patients with tetraplegia have ages between 45 and 80 years and have

    levels of cervical SCI impairment at C3, C4, C5 and C6 with the details shown in Table 2.

    The relevant tasks used are as follows:

    Mental arithmetic: Participants imagined mentally solving a non-trivial multiplication problem.

    Mental figure rotation: Participants were asked to imagine a cube being rolled forward.

    Mental letter composing: Participants composed a simple word in their mind.

    Mental visual counting: Participants performed mentally to visualize appearing and disappearing a number being counting upward on a blackboard.

    SSVEP 6 Hz: Participants were asked to concentrate on an LED that continuously flickers at 6 Hz.

    SSVEP 13 Hz: Participants were asked to concentrate an LED that continuously flickers at 13 Hz.

    SSVEP 16 Hz: Participants were asked to concentrate an LED that continuously flickers at 16 Hz.

    Eyes closed: This task is used to measure alpha wave production. Participants were asked to relax while performing eyes closed action.

    Baseline: Participants were asked to be relaxed with open eyes, thinking of nothing in particular.

    The placement of EEG electrodes is based on the international 10-20 system. For each

    subject gold electrodes were positioned using a bipolar montage at P3-T4 for channel 1

    (CH1) and O2-T3 for channel 2 (CH2) as shown in Figure 7. The location CZ was used for

  • the reference electrode. EEG gel was applied to keep the impedance level low and a better

    electrical contact. Unnecessary movements and eye blinks were kept as minimal as possible

    during data collection in each session. Each task was measured for a total of 6 sessions with

    each session lasting 13 seconds. The first 3 seconds of data were discarded for preparation

    time, the remaining 10 seconds were used for the further signal processing.

    Results

    There are distinct differences of the data features in terms of amplitude and frequency

    between SSVEP based BCI and eyes closed action. The SSVEP BCI in the experiment uses

    three stimulation frequencies: 6Hz, 13Hz and 16Hz respectively. Figure 8 shows the plotting

    of the PSD feature for the three frequencies in which the relevant target frequency can be

    easily identified from both channels at the highest peak in the power spectrum.

    In practice, the peak frequency during eyes closed action can be used to check and test

    instrumentation wireless EEG. If the EEG electrodes are properly attached to the scalp, the

    system should be able to detect the peak alpha frequency during eyes closed action. The time

    plots in Figure 9 show a larger amplitude signal from eyes closed compared to eyes opened

    ones. The dominant alpha frequency between 8-13Hz during eyes closed action on the

    occipital lobe (O2) also clearly showed in the frequency plots.

    The training of the classifier has been developed offline with the training, validation and

    testing sets taken from five healthy participants and five patients with tetraplegia. Due to the

    large differences in EEG based-BCI known as inter-subject variability [48] which could

    affect the performance, in this study, the training of the ANN was done offline with all

    subjects contributed portions of data for training and testing datasets. In the case of adaptive

    learning for new user, the system can collect a small sample of data from a new user and re-

    training the system to enable the classifier to adapt to the data generalization from the new

    user which was previously unseen by the classifier. This new user adaptive BCI approach

    will be our future study.

    One of the error cycles performance of the ANN training is shown in Figure 10. Note that

    the error (MSE) of the training set decreased smoothly. The validation set dropped from the

    beginning up to 8 epochs, then started to increase continuously. The training of the network

    was stopped at 8 epochs because the validation performance started to increase continuously.

    This is used to prevent over-training of the network.

    The results of mental task classification for five healthy subjects are shown in Table 3 with

    different classification accuracy result for each subject. Table 3 shows each combination of

    two mental tasks with an accuracy of above 80%. Moreover, the combination of three mental

    tasks for most of the subjects has accuracy of 72±1%. Additional four mental tasks

    classification is also shown with the result of 64%. The result of the SSVEP task combination

    shows a higher classification accuracy compared to the mental task. The average

    classification on each frequency (6Hz, 13Hz, and 16Hz) vs. baseline task resulted in accuracy

    of 85±4%, while three frequencies combined classification resulted in accuracy of 80%. The

    eye closed vs. baseline task has classification accuracy of 97%.

    The classification results for patients with tetraplegia are shown in Table 4. The results

    show comparable classification accuracies to the healthy group. The two mental task

    classifications still have accuracy of 80% and above. The three mental task classifications

  • yield accuracies of 70±2%, almost the same as in healthy group. The four mental task

    classifications have accuracy of 62% which is also comparable to the healthy group. The

    accuracies for SSVEP (6Hz, 13Hz, and 16Hz) vs. baseline task have value between 82% and

    84%. The three frequency classifications yield accuracy of 71%. The eye closed vs. baseline

    task classification accuracy is still remaining high of 93%.

    Due to the limitation of the recruitment for this study, in which 5 healthy participants (25-

    35 years old) and 5 patients with tetraplegia (45-80 years old) were available for the

    experiment. However, the comparison between the healthy group and the patient group on

    Table 5 shows in relation for combination three mental tasks classification resulted in p-value

    > 0.2. This means that there were no significant difference between the 5 healthy subjects and

    the 5 patients with tetraplegia.

    Discussions

    It should be noted that the baseline task could sometimes reaches higher amplitude and

    smaller peak in the alpha bands (8-12Hz) especially when the subjects are relaxed. Moreover,

    there is also a dominant alpha wave associated with the eyes closed task. This is the reason

    why the SSVEP frequencies are chosen not in the range of the alpha band frequency to ensure

    there is no overlapping during relaxation and during eyes closed action.

    The idea of the proposed hybrid system is to combine the mental tasks based BCI, SSVEP

    based BCI and eyes closed task together with the measurement system just using only a

    wireless two channel EEG with electrode positions at parietal (P3) and occipital (O2) lobes.

    The three mental tasks combination is the chosen for BCI applications, for example in

    wheelchair navigation using three commands (forward, left and right). The SSVEP based BCI

    with three frequencies combination could be applied to the other applications such as BCI

    speller system, motor prostheses, environmental control and any other task in the scenario

    without the subject concentrating on two tasks at the same time. The system is turned on or

    off by using the eyes closed task with duration of above 2s. When the system is turned off,

    the user could perform normal mental activities naturally without the worry of false

    activation. Eyes closed task with different duration periods is used as well to select between a

    mental task and SSVEP type of BCI.

    Hybrid BCI algorithm (Figure. 4) is applied by using the combination of eyes closed task,

    mental task based BCI and SSVEP based BCI together with the offline classification result is

    shown in Table 6 for using ANN classifier. The average ANN classification accuracy of

    combinations of eyes closed, three mental tasks and three frequencies SSVEP is 74%.

    Another performance measurement technique could also be used, the information transfer

    rate (ITR) or bit rate as the amount of reliable information received by this hybrid system

    using the following bit rate function:

    𝐵 = 𝑉 [log2 𝑁 + 𝑃 log2 𝑃 + (1 − 𝑃) log2((1 − 𝑃)/(𝑁 − 1))] (5)

    where B is bit rate (bits/min), V is the application speed in trials/mins. N is the number of

    possible tasks and P is to the classification accuracy [49]. In general, the ITR depends on the

    number of classification tasks used, the time of the system for the classification and the

    classification accuracy. The ITR values previously reported for BCI systems have reached at

    least 25 bits/min [50].

  • In this experiment the system speed is set as variable which could be adjusted. For example

    Figure 11 shows the ITR in bits per trial with the speed of the system configured to provide a

    classification every second (V = 60 trials/min). In this prototype experiment, the mean value

    of the bit rate is between 21 bits/min until 39 bits/min or 27 bits/min in average value for five

    patients with tetraplegia.

    For comparison purposes, the classification accuracy using other classifiers which include

    linear discriminant analysis (LDA) [51, 52], support vector machine (SVM)[52, 53] as shown

    in Table 7, together with the result of ANN. The average classification accuracy between the

    combinations of eyes closed, three mental tasks and three frequencies shows that the LDA

    classifier gives a lower average accuracy of 63.7% compared to SVM and ANN. The SVM

    classifier is comparable to ANN with accuracy of 73.2%. The ANN classifier with the

    average accuracy of 74% is the algorithm being implemented in the embedded system for

    real-time classification.

    The final weights from ANN training giving the best classification accuracy of the tasks are

    transferred as parameters to the main embedded system for real-time application. Further

    testing of the real-time embedded hybrid BCI system has been conducted for patients with

    tetraplegia to prove that the hybrid task is adequate to be used for the intentional signal to

    control the wheelchair. The three-task classifications from Table 6 are mapped to the three

    wheelchair movements (left, forward and right). The eyes-closed task is used to switch the

    system on/off and for hybrid system selection. The set-up experiment of patients with

    tetraplegia is shown in Figure 12.

    The LCD displays the output of the command recognition as shown in Figure 13 which is

    related to the hybrid BCI tasks. The embedded system turns into a waiting state for any

    command activation. As soon as a command is activated, the LCD displays the result of the

    classification on the first line of LCD. If the on/off command is activated, LCD displays ‘O/F

    On/Off’. If the turn left command is activated, LCD displays ‘L Left’. For move

    forward command, LCD displays ‘F Forward’ and for turn right command, LCD display

    ‘R Right’. The second line of the LCD displays the same information: ‘L F O/F R’ to

    represent four intentional signals of wheelchair control.

    The result of the real-time testing with the five patients with tetraplegia is shown in Table

    8. Essentially, there are four basic commands used for the wheelchair which includes: on/off,

    left, forward and right. Eyes closed task is used to switch the system on and off. The chosen

    best classification of three mental tasks or three frequencies SSVEP (Table 6) is used to

    provide three wheelchair steering states: turn left, move forward and turn right. Each

    particular task related to each command is repeated 25 times for the testing experiment. The

    correctly detected commands and speed of execution are identified.

    For the on and off command using eyes closed task, the success rates among five patients

    with tetraplegia are still high enough at between 80 and 96%. This is relevant to the ANN

    classification result of the eyes closed task. Testing for the ‘turn left’ command indicates

    success rates of between 60 and 72%. The success rates detection for the ‘move forward’ and

    turn right commands are between 60 and 68%. The success rates of these three wheelchair

    steering movements are relevant to the ANN classification result of the three mental tasks or

    3 SSVEP frequency classifications. The average success rate of the detection from five

    patients with tetraplegia is 70%.The speed of execution each task is varied from 2 to 4

    seconds.

  • Note that this study used data from 5 healthy participant and 5 patients with tetraplegia.

    The limitation of current algorithm may not effectively work for new user. In the case of

    adaptive learning for new user, the system can collect small sample for re-training to enable

    the generalization of the classifier. This new user adaptive BCI study will be our future study.

    Conclusion

    The prototype of hybrid brain computer interface for the biomedical cyber-physical system

    application has been presented here which has a small size two channel wireless EEG,

    microcontroller based system and accompanying low power embedded processing system

    with portability, convenience and cost effectiveness compared to other PC based systems.

    The hybrid system is successfully able to combine the mental task classification using three

    mental task combinations, SSVEP based BCI with three frequencies classification and eyes

    closed detection by using only two EEG channels with active position at parietal (P3) an

    occipital (O2) lobes. This experiment is involved with five healthy subjects and five patients

    with high level injury SCI (tetraplegia).

    The developed hybrid BCI system provides more flexibility for the user to choose the

    suitable task. The results show comparable offline classification accuracies between healthy

    subjects and patients with tetraplegia. For patients with tetraplegia, the average offline ANN

    classification accuracy of the combination of mental task based-BCI, SSVEP based BCI and

    eyes closed is 74% and the average information transfer rate is 27 bits per minute. For the

    real-time testing of the intentional signal on patients with tetraplegia, the average success rate

    of detection is 70% and the speed of detection varies from 2 to 4 seconds. Future direction of

    this research is the real-time brain controlled wheelchair and the use of the sensor for obstacle

    avoidance for the safety to create a semi-autonomous brain controlled wheelchair system.

    List of Abbreviations

    ADC: Analog to digital converter; ALS: Amyotrophic lateral sclerosis; ANN: Artificial

    neural network; BCI: Brain computer interface; CPS: Cyber-physical system; CMRR:

    Common mode rejection ratio; EEG: Electroencephalography; ERD: Event-related

    desynchronization; ERS: Event-related synchronization; FFT: Fast Fourier transform; FIFO:

    First-in-first out; In-Amp: Instrumentation amplifier; ITR: Information transfer rate; MSE:

    Mean square error; PCB: Printed circuit board; PSD: Power spectral density; RFI: Radio

    frequency interference; RTC: Real time clock; RTOS: Real-time operating system; SCI:

    Spinal cord injury; SMR: Sensory motor rhythm; SPI: Serial peripheral bus; SSVEP: Steady

    state visual evoked potential; UART: Universal asynchronous receiver/transmitter.

    Acknowledgements

    Not applicable.

    Funding

    This work was supported by Australian Research Council under Discovery Grant

    DP0666942.

  • Data Availability

    Please contact author for data requests.

    Authors’ contributions

    RC conducted the experiments; performed all data analysis and wrote the manuscript. GN

    and SL advised the analysis edited the manuscript. HN supervised the study; advised the

    analysis and edited the manuscript.

    Competing interests

    The authors declare that they have no competing interests.

    Ethics Approval and Consent to Participate

    All participants provided written informed consent prior to participating in the study. The

    study was approved by the Human Research Ethics Committee at the University of

    Technology Sydney with reference number: UTS HREC 2011-217A.

    Author details

    Centre for Health Technologies, Faculty of Engineering and Information Technology,

    University of Technology, Sydney, NSW 2007, Australia.

    References

    1. Broy M, Cengarle MV, Geisberger E: Cyber-physical systems: imminent challenges. In Monterey workshop. Springer; 2012:1-28.

    2. Suh SC, Tanik UJ, Carbone JN, Eroglu A: Applied cyber-physical systems. Springer 2014, 2:27. 3. Thompson DE, Blain-Moraes S, Huggins JE: Performance assessment in brain-computer

    interface-based augmentative and alternative communication. BioMedical Engineering OnLine 2013, 12(1):1-23.

    4. Tong J, Zhu D: Multi-phase cycle coding for SSVEP based brain-computer interfaces. BioMedical Engineering OnLine 2015, 14(1):1-13.

    5. Chai R, Ling SH, Hunter GP, Tran Y, Nguyen HT: Brain-Computer Interface Classifier for Wheelchair Commands Using Neural Network With Fuzzy Particle Swarm Optimization. IEEE Journal of Biomedical and Health Informatics 2014, 18(5):1614-1624.

    6. Facchinetti T, Della Vedova ML: Real-time modeling for direct load control in cyber-physical power systems. IEEE Transactions on Industrial Informatics 2011, 7(4):689-698.

    7. Malinowski A, Yu H: Comparison of embedded system design for industrial applications. IEEE Transactions on Industrial Informatics 2011, 7(2):244-254.

  • 8. Rojhani S, Stiens SA, Recio AC: Independent sailing with high tetraplegia using sip and puff controls: integration into a community sailing center. The Journal of Spinal Cord Medicine 2016:1-10.

    9. Guo S, Cooper RA, Boninger ML, Kwarciak A, Ammer B: Development of power wheelchair chin-operated force-sensing joystick. In Engineering in Medicine and Biology, 2002 24th Annual Conference and the Annual Fall Meeting of the Biomedical Engineering Society EMBS/BMES Conference, 2002 Proceedings of the Second Joint; 23-26 Oct. 2002. 2002:2373-2374 vol.2373.

    10. Hamedi M, Salleh S-H, Astaraki M, Noor AM: EMG-based facial gesture recognition through versatile elliptic basis function neural network. BioMedical Engineering OnLine 2013, 12(1):1-22.

    11. Pinheiro CG, Naves EL, Pino P, Losson E, Andrade AO, Bourhis G: Alternative communication systems for people with severe motor disabilities: a survey. BioMedical Engineering OnLine 2011, 10(1):1-28.

    12. Huo X, Ghovanloo M: Evaluation of a wireless wearable tongue–computer interface by individuals with high-level spinal cord injuries. Journal of Neural Engineering 2010, 7(2):026008.

    13. Nguyen H, King L, Knight G: Real-time head movement system and embedded linux implementation for the control of power wheelchairs. In Engineering in Medicine and Biology Society, 2004 IEMBS'04 26th Annual International Conference of the IEEE. IEEE; 2004:4892-4895.

    14. Chaudhary U, Birbaumer N, Ramos-Murguialday A: Brain–computer interfaces in the completely locked-in state and chronic stroke. Progress in Brain Research 2016, 228:131-161.

    15. Wolpaw JR: Brain–computer interfaces as new brain output pathways. The Journal of physiology 2007, 579(3):613-619.

    16. Mason S, Bashashati A, Fatourechi M, Navarro K, Birch G: A comprehensive survey of brain interface technology designs. Annals of biomedical engineering 2007, 35(2):137-169.

    17. Wolpaw JR: Brain-computer interfaces. Handbook of clinical neurology 2012, 110:67-74. 18. He B, Gao S, Yuan H, Wolpaw JR: Brain–computer interfaces. In Neural Engineering. Springer;

    2013: 87-151. 19. Sellers EW, Donchin E: A P300-based brain–computer interface: initial tests by ALS patients.

    Clinical neurophysiology 2006, 117(3):538-548. 20. Turnip A, Hong K-S, Jeong M-Y: Real-time feature extraction of P300 component using adaptive

    nonlinear principal component analysis. BioMedical Engineering OnLine 2011, 10(1):1-20. 21. Allison BZ, McFarland DJ, Schalk G, Zheng SD, Jackson MM, Wolpaw JR: Towards an

    independent brain–computer interface using steady state visual evoked potentials. Clinical neurophysiology 2008, 119(2):399-408.

    22. Birbaumer N, Hinterberger T, Kubler A, Neumann N: The thought-translation device (TTD): neurobehavioral mechanisms and clinical outcome. IEEE transactions on Neural Systems and rehabilitation engineering 2003, 11(2):120-123.

    23. Kübler A, Nijboer F, Mellinger J, Vaughan TM, Pawelzik H, Schalk G, McFarland DJ, Birbaumer N, Wolpaw JR: Patients with ALS can use sensorimotor rhythms to operate a brain-computer interface. Neurology 2005, 64(10):1775-1777.

    24. Choi D, Ryu Y, Lee Y, Lee M: Performance evaluation of a motor-imagery-based EEG-Brain computer interface using a combined cue with heterogeneous training data in BCI-Naive subjects. BioMedical Engineering OnLine 2011, 10(1):1-12.

    25. Pfurtscheller G, Brunner C, Schlögl A, Da Silva FL: Mu rhythm (de) synchronization and EEG single-trial classification of different motor imagery tasks. Neuroimage 2006, 31(1):153-159.

    26. Ortega J, Asensio-Cubero J, Gan JQ, Ortiz A: Classification of motor imagery tasks for BCI with multiresolution analysis and multiobjective feature selection. BioMedical Engineering OnLine 2016, 15(1):149-164.

    27. Curran E, Sykacek P, Stokes M, Roberts SJ, Penny W, Johnsrude I, Owen AM: Cognitive tasks for driving a brain-computer interfacing system: a pilot study. IEEE Transactions on Neural Systems and Rehabilitation Engineering 2004, 12(1):48-54.

  • 28. Birch GE, Bozorgzadeh Z, Mason SG: Initial on-line evaluations of the LF-ASD brain-computer interface with able-bodied and spinal-cord subjects using imagined voluntary motor potentials. IEEE Transactions on Neural Systems and Rehabilitation Engineering 2002, 10(4):219-224.

    29. Craig D, Nguyen H, Burchey H: Two channel EEG thought pattern classifier. In Engineering in Medicine and Biology Society, 2006 EMBS'06 28th Annual International Conference of the IEEE. IEEE; 2006:1291-1294.

    30. Osaka M: Peak alpha frequency of EEG during a mental task: Task difficulty and hemispheric differences. Psychophysiology 1984, 21(1):101-105.

    31. Menon V, Desmond J: Left superior parietal cortex involvement in writing: integrating fMRI with lesion evidence. Cognitive brain research 2001, 12(2):337-340.

    32. Rivera SM, Reiss A, Eckert MA, Menon V: Developmental changes in mental arithmetic: evidence for increased functional specialization in the left inferior parietal cortex. Cerebral cortex 2005, 15(11):1779-1790.

    33. Milivojevic B, Hamm JP, Corballis MC: Hemispheric dominance for mental rotation: it is a matter of time. Neuroreport 2009, 20(17):1507-1512.

    34. Vanni S, Revonsuo A, Saarinen J, Hari R: Visual awareness of objects correlates with activity of right occipital cortex. NeuroReport 1996, 8(1):183-186.

    35. Craig A, Moses P, Tran Y, McIsaac P, Kirkup L: The effectiveness of a hands-free environmental control system for the profoundly disabled. Archives of physical medicine and rehabilitation 2002, 83(10):1455-1458.

    36. Lotte F, Congedo M, Lécuyer A, Lamarche F, Arnaldi B: A review of classification algorithms for EEG-based brain–computer interfaces. Journal of neural engineering 2007, 4(2):R1.

    37. Nguyen HT: Intelligent technologies for real-time biomedical engineering applications. International Journal of Automation and Control 2008, 2(2-3):274-285.

    38. Chai R, Ling SH, Hunter GP, Nguyen HT: Mental non-motor imagery tasks classifications of brain computer interface for wheelchair commands using genetic algorithm-based neural network. In The 2012 International Joint Conference on Neural Networks (IJCNN). IEEE; 2012:1-7.

    39. Åberg MC, Wessberg J: Evolutionary optimization of classifiers and features for single-trial EEG Discrimination. BioMedical Engineering OnLine 2007, 6(1):1-8.

    40. Allison BZ, Neuper C: Could anyone use a BCI? In Brain-computer interfaces. Springer; 2010: 35-54.

    41. Allison BZ, Wolpaw EW, Wolpaw JR: Brain–computer interface systems: progress and prospects. Expert review of medical devices 2007, 4(4):463-474.

    42. Li Y, Pan J, Wang F, Yu Z: A Hybrid BCI System Combining P300 and SSVEP and Its Application to Wheelchair Control. IEEE Transactions on Biomedical Engineering 2013, 60(11):3156-3166.

    43. Xu M, Qi H, Wan B, Yin T, Liu Z, Ming D: A hybrid BCI speller paradigm combining P300 potential and the SSVEP blocking feature. Journal of neural engineering 2013, 10(2):026001.

    44. Yin E, Zhou Z, Jiang J, Chen F, Liu Y, Hu D: A Speedy Hybrid BCI Spelling Approach Combining P300 and SSVEP. IEEE Transactions on Biomedical Engineering 2014, 61(2):473-483.

    45. Wang M, Daly I, Allison BZ, Jin J, Zhang Y, Chen L, Wang X: A new hybrid BCI paradigm based on P300 and SSVEP. Journal of Neuroscience Methods 2015, 244:16-25.

    46. Webster J: Medical instrumentation: application and design. John Wiley & Sons; 2009. 47. Friman O, Volosyak I, Graser A: Multiple channel detection of steady-state visual evoked

    potentials for brain-computer interfaces. IEEE Transactions on Biomedical Engineering 2007, 54(4):742-750.

    48. Grosse-Wentrup M, Schölkopf B: A review of performance variations in SMR-based brain− computer interfaces (BCIs). In Brain-Computer Interface Research. Springer; 2013: 39-51.

    49. McFarland DJ, Sarnacki WA, Wolpaw JR: Brain–computer interface (BCI) operation: optimizing information transfer rates. Biological psychology 2003, 63(3):237-251.

    50. Wolpaw JR, Birbaumer N, McFarland DJ, Pfurtscheller G, Vaughan TM: Brain–computer interfaces for communication and control. Clinical neurophysiology 2002, 113(6):767-791.

  • 51. Lu J, Plataniotis KN, Venetsanopoulos AN: Face recognition using LDA-based algorithms. IEEE Transactions on Neural networks 2003, 14(1):195-200.

    52. Franc V, Hlavác V: Statistical pattern recognition toolbox for Matlab. Prague, Czech: Center for Machine Perception, Czech Technical University 2004.

    53. Guler I, Ubeyli ED: Multiclass support vector machines for EEG-signals classification. IEEE Transactions on Information Technology in Biomedicine 2007, 11(2):117-126.

  • Table 1. Wireless EEG Specifications

    Table 2. Information of Healthy Subjects and Patients with Tetraplegia

    Table 3. Results of Classification Accuracy for Five Healthy Subjects (S1-S5)

    Table 4. Accuracies for Five Patients with Tetraplegia (SCI1-SCI5)

    Table 5. Comparison of healthy subjects and patients with tetraplegia to classify three

    mental tasks combinations

    Table 6. Results of Accuracy of Patients with Tetraplegia (SCI1-SCI5) with Chosen

    Task Combinations for Hybrid BCI Algorithm

    Table 7. Comparison Accuracy of the classifiers (LDA, SVM and ANN) of Patients with

    Tetraplegia (SCI1-SCI5) with Chosen Task Combinations for Hybrid BCI

    Table 8. Results of Real-time Testing for 5 Patients with Tetraplegia with the Repeated

    Command of Intentional Signals

    Figure 1. Hybrid BCI system, EEG based CPS module

    Figure 2. Wireless EEG module

    Figure 3. ANN architecture

    Figure 4. Pseudo-code for hybrid BCI system detection algorithm

    Figure 5. The threshold value of eyes closed signal

    Figure 6. Modes selection in the Hybrid BCI algorithm

    Figure 7. Two channels bipolar EEG electrodes placement

    Figure 8. SSVEP-BCI features plotting

    Figure 9. Alpha eyes closed detection with wireless EEG

    Figure 10. Training and validation error cycle

    Figure 11. Plotting information transfer rate (ITR) or bit rate in bits/min

    Figure 12. Experiment of Hybrid BCI system with patients with tetraplegia

    Figure 13. Real-time BCI detection, mapped to commands and LCD output