Page 1: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

CHEMISTRY 1210

Spring, 2019Kate Mazloumi

How to use the Spectronic 20

UV/Visible Spectrophotometer

Page 2: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

NOTE!

We have several models of Spectronic 20, so the

instrument that you use may look a bit

different, however, all have the same

components and the following general

instructions will apply.

Page 3: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

Plug in the instrument, then rotate the front left knob clockwise to turn on. Allow 15 minutes’ warm-up time.

APPLY POWER:

Knob in OFF position

Light is on when knob in on position

Page 4: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

SELECT THE WAVELENGTH

Adjust the wavelength control dial to the required setting, indicated on the wavelength scale.

You must set the W.E.B.S to catch the

result.W: Wavelength,

E: Empty, B: Blank, and S: Sample

Page 5: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

THE CUVETTE

Two cuvettes in a test tube rack will be found

beside each instrument.

One cuvette will be used for the “blank” while

the other will be used for the sample being

measured.

Page 6: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

The Cuvet te

*Every cuvette has a white line and dot that is

used to align it the same way each time in the

sample chamber.

*Never use a brush to clean a cuvette, as it

could scratch the inside of the glass causing

measurement errors.

*Rinse the cuvettes well with distilled water at

the end of the lab.

Page 7: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

Before you insert the cuvette into the

instrument, wipe away any fingerprints

or liquid with a clean Kimwipe!

Align the vertical mark with the mark in

the sample chamber of the instrument.

USING THE CUVETTE

Page 8: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

Empty compartment

With the sample compartment empty and the lid closed, adjust the power reading to 0.0% transmit tance using the left- hand knob.

ZEROING THE INSTRUMENT

Page 9: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

Standardize the light control: Rinse the cuvette well with the blank solution. Half fill the cuvette with the blank. Wipe the outside of the cuvette with a Kimwipe.

The blank can be distilled water or another reference liquid or solution.

Page 10: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

“BLANKING THE INSTRUMENT” (Adjusting to give 100% transmittance)

• Slide the cuvette gently into the sample compartment, then push it down until the bottom of the cuvette touches the bottom of the sample compartment.

• Flip the cover closed.• Adjust to 100% T using the knob on the

right-hand side at the front of the instrument.

Page 11: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

THE BLANK

Store the blank sample in the provided test tube rack. It will be reused throughout the experiment!

Page 12: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

Rinse the second cuvette three times with a small amount of the sample solution to be measured.

Fill cuvette to at least halfway with the sample.

Wipe the outside of the cuvette with a Kimwipe, then place the sample into the sample compartment.

Read and record the transmittance, estimating to one decimal place.

THE SAMPLE

Page 13: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

SUMMARY1- Apply power; Plug in, turn on (15 min warm up)

2- Select the desired wavelength.

3- Zero the instrument.

4- Standardize the light control.

5- Make a sample measurement. Note: step 4 must be repeated each time wavelength is changed.

Page 14: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

To create a calibration curve, you will be asked to prepare a numberof solutions with different known concentrations.

The transmittance of each must be measured. Dump the previoussample into the waste beaker, rinse the sample cuvette with a little ofthe new sample, then fill at least halfway. After wiping the cuvette,place it in the sample compartment and read the transmittance.

Check that the blank sample reads 100% transmittance betweenevery second sample. Adjust the right front knob if needed.

CALIBRATION CURVES

Page 15: How to use the Spectronic20 UV/Visible Spectrophotometer

Where thought meets action

*Reset your full WEB (in % T mode) if you change the wavelength.*Turn off the power switch and unplug the instrument. *Rinse the cuvettes well with distilled water and return them to the rack beside the instrument.

Convert transmittance to absorbance by using [2-log(%T)].