Horiba LASER DIFFRACTION PARTICLE SIZE ANALYSER LA-960 The LA-960 features a Wide Measurement range from 10nm - 5000um to measure every application. NIST-traceable size standards verify that the LA-960 accurately measures peaks as fine as 30 nanometers. Guaranteed Accuracy of ± 0.6%and Precision of ± 0.1% for NIST Standards Suitable for Suspensions, Emulsions, Powders, Pastes, Gels, and Creams 60 - Second Measurement cycle, even in wet mode boost productivity Long life solid state laser source better than Gas Filled He-Ne Lasers Fully compliant with ISO 13320 recommendations regarding the measurement of materials on D10, D50 and D90. Advance Scientific Equipment PVT LTD Innovative Solutions. Quality Services Horiba LA-960 is the Highest Performance laser diffraction analyzer available. The ability to measure real world samples down to lowest particle size of 30 nm and up to 5,000 μm is unique and extremely valuable for labs looking for flexibility. FEATURES AT A GLANCE TECHNOLOGY
2
Embed
Horiba LASER DIFFRACTION PARTICLE SIZE ANALYSER LA …advscientific.com/wp-content/uploads/2015/07/HORIBA-LA-960... · Horiba LASER DIFFRACTION PARTICLE SIZE ANALYSER LA-960 The LA-960
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Horiba LA-960 is the Highest Performance laser diffraction analyzer available. The ability to measure real world samples down to lowest particle size of 30 nm and up to 5,000 µm is unique and extremely valuable for labs looking for flexibility.
POWDERJET DRY FEEDER ACCESSORY Dispersion Method Compressed air dispersion using Venturi nozzle Sample Delivery Vibrating feeder Sample Disposal Vacuum-‐driven evacuation Measurement range
100nm – 5000 μm
Controls Communication:Serial cable to LA-‐960 main unit Measurement: Vibrating Feeder controlled automatically via feedback or manually by user, Vacuum Auto/OFF, Compressed air Auto/OFF, Air pressure is adjustable from 0-‐0.4Mpa in 40 steps
Measurement Time
Typical measurement takes 2 seconds or longer.
Dimensions 332 (W) mm x 321 (D) mm x 244 (H) mm (not including dimensions of projections and LA-‐960 measurement unit)
Power for PowderJet Operation
AC 100 V/120 V/230V, 50/60 Hz, 1500 VA (Including vacuum; not including LA-‐960 measurement unit)
Compressed Air Supply Pressure
Compressed air supply origin pressure: 0.4 -‐0.8 MPa Compressed air controlling range:0.01 -‐0.4 MPa
Compressed Air Connection
Quick connector for resin tube with 6 mm outer diameter (Compressed air supply equipment must be provided separately)
Remarks Vaccum is equipped as standard
Measurement Method
Mie Scattering and Fraunhofer diffraction
Measurement Range
10nm to 5000* μm (*Upto 3000 micron in wet mode)
Measurement Time
Typical measurement takes 60 seconds from liquid filling, sampling and measurement to rinsing.
Measurement Method
Circulation measurement or fraction cell measurement (Fraction cell is optional)
Sample Quantity Approximately 10 mg to 5 g (Depends on particle size, distribution and density)
Dispersing Volume
Approximately 180 mL for standard pumping system. 5/10/15 mL for Fraction Cell accessory. Manual filling: 35mL, Automatic filling : 40 mL for MiniFlow accessory .Approximately 1L of LiterFlow option.
Available carrier fluid
Aqua* (A Type), Organic solvent (S type) (*Ethanol can be used as a dispersing additive)
Communication USB 2.0 Light Source Red solid state 5mW laser diode(650nm), Blue solid state 3mW LED