Helical Tomotherapy Quality Assurance Cheng Shi PhD Chengyu Shi, PhD Department of Radiation Oncology Di i i f M di l Ph i Division of Medical Physics Southwest Chapter of the AAPM, Oct 9, 2009
68

Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Mar 24, 2020

Download

Documents

dariahiddleston
Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Page 1: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical Tomotherapy Quality Assurance

Cheng Shi PhDChengyu Shi, PhDDepartment of Radiation Oncology

Di i i f M di l Ph iDivision of Medical Physics

Southwest Chapter of the AAPM, Oct 9, 2009

Page 2: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

ObjectivesObjectives• To introduce our five-year clinical experience

ith helical TomoTherapwith helical TomoTherapy• To introduce our five-year research experience

with helical TomoTherapy

There is no conflict of interestThere is no conflict of interest

Page 3: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

IGRT SolutionIGRT SolutionImaging

Ionizing Non-ionizing

CT Ultrasound MRIX-ray

PETSPECT RF Laser

OthersOthers

Page 4: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

IGRT QAIGRT QA• IGRT solution introduces precise targeting

• However, it may also introduce more uncertainties and increased risk to miss the targetg

• Therefore, comprehensive QA procedures are necessary

• Helical Tomotherapy has its own characteristics• Helical Tomotherapy has its own characteristics, and requires dedicated QA protocol

Page 5: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical TomoTherapy QAHelical TomoTherapy QA

QA

System PatientSystem Patient

Hardware Software Planning Delivery

Page 6: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical Tomotherapy ProcessHelical Tomotherapy ProcessR i d i

MVCTRegistered image

MVCT Guided

Image registration

Helical CT Delivery

Page 7: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical Tomotherapy SystemHelical Tomotherapy System

Hardware Software

Page 8: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical Tomotherapy QA DesignHelical Tomotherapy QA Design

Component-based: QA each component, h MLC d t t h tsuch as MLC, detector, couch etc.

System based: QA the whole systemSystem-based: QA the whole system performance

I am NOT good I am good

Page 9: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical Tomotherapy QAHelical Tomotherapy QA

Chester Ramsey, et al. "Mechanical and Dosimetric Quality Assurance for Helical Tomotherapy," International Journal of Radiation Oncology Biology Physics, 1 November 2007 (Vol. 69, Issue 3, Page S699)

Page 10: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical Tomotherapy QAHelical Tomotherapy QA

Chester Ramsey, et al. "Mechanical and Dosimetric Quality Assurance for Helical Tomotherapy," International Journal of Radiation Oncology Biology Physics, 1 November 2007 (Vol. 69, Issue 3, Page S699)

Page 11: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

TomoTherapy Daily QA PerformanceTomoTherapy Daily QA Performance

Page 12: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical Tomotherapy Target IssueHelical Tomotherapy Target Issue

Eroded target Rusted on-board ion chamber

Page 13: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical Tomotherapy Working ZonesHelical Tomotherapy Working Zones

Page 14: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

VPFN VIC Dose Rate RelationshipVPFN, VIC, Dose Rate Relationship

Page 15: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical Tomotherapy MLC QAHelical Tomotherapy MLC QA

Page 16: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Special MLC QA for Helical Tomotherapyp py

S h b d i dSeven test patterns have been designed

A. Box in BoxB. Modified Checkerboard. od ed C ec e boa dC. MLC Interleaf LeakageD IEC X gradientD. IEC-X gradientE. IEC-Y gradientF C l Fi ld A (2 t t )F. Complex Field A (2 tests)G. Complex Field B (2 tests)

Page 17: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Pattern A. Box in boxPattern A. Box in box

Page 18: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

A. Box in BoxPattern A Box in box. o oPattern A. Box in box

Page 19: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Pattern B. Modified checkerboxPattern B. Modified checkerbox• Can be thought as small box in box testg• Initially used for Mimic MLC test• Modified for tomotherapy MLC testpy

Page 20: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Pattern B Modified checkerboxPattern B. Modified checkerbox

Page 21: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Pattern C Inter leaf LeakagePattern C. Inter leaf Leakage

Page 22: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Pattern D. IEC-X GradientPattern D. IEC X GradientPNMW depthnormwedge PNMW ••=

PNMW ••= depthnormwedge PNMW ••=

Page 23: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Pattern E. IEC-Y GradientPattern E. IEC Y Gradient

PNMW depthnormwedge PNMW ••=

Page 24: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Pattern F. Complex APattern F. Complex A

Page 25: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Pattern G. Complex BPattern G. Complex B

Page 26: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

MVCT QAMVCT QA

3% contrast

1 6 mm resolution1.6 mm resolutionCT-density table consistency

Page 27: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Helical Tomotherapy TQATMHelical Tomotherapy TQA

Page 28: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Tomotherapy Patient-Specific QATomotherapy Patient Specific QA

QA Equipment

Generate QA procedure

Deliver QA procedure

Analysis QA procedure

Evaluate QA procedure

Page 29: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

QA EquipmentQA Equipment

Cheese Homo Phantom

EDR2 Film

A1SL Ion Chamber

Page 30: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Deliver QA procedure-PatientDeliver QA procedure Patient

PhantomRed lasers

Chamber

EDR2 film

Page 31: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Deliver QA procedure-CalibrationDeliver QA procedure Calibration

Page 32: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Analysis QA procedure-Absolute dosey p

Excel sheet for dose measurement

Page 33: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

What if the QA fails?Q

Find out why?

Phantom resolution? High dose gradient? Small thin PTV? Machine output? …

Repeat the QA procedure

Page 34: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Example-Overweightp g

•Patient with prostate cancer (weight 370 lb)•Tomotherapy couch can support ~440 lb.•Patient plan is homogeneous for PTVp g•Phantom size is very different from patient size resulting to non-homogenous dose at thesize, resulting to non-homogenous dose at the point of the chamber

Page 35: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Example-Overweight

25.90 cm

Page 36: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Example-Overweightp g

3 88% here

15.50 cm

3.88% here

6.22% here

Page 37: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Delta 4 Dosimeter for Tomo

• Purchased  September 2008TomoTherapy OptionRapidArc Option

Fi t f th• First few monthsResearchSBRT QASBRT QA

• Mid‐January 2009Fully clinical for tomotherapy DQA

• Absolute calibrate on T ThTomoTherapy

• Phantom MVCT image for planningplanning

Page 38: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Delta 4 Dosimeter ResultsDelta 4 Dosimeter Results• 140 DQAs  delivered Q

on Delta4 

• 34 SBRT patients with gamma index > 95% (3%/3mm) 

• Pass criteria3%/3mm

90%>90% gamma

• To date,Mean gamma: 97 1%Mean gamma: 97.1% (85.9‐100)

Page 39: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

MVCT VerificationMVCT Verification

Page 40: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

MVCT Verification

Page 41: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Setup UncertaintiesSe up U ce a es

Page 42: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

DiscussionDiscussion

T th i li bl l l ti lt d it h ld• Tomotherapy gives reliable calculation results, and it should be within 3% tolerance about dose measurement for most of casescases

• EDR2 is a good tool for relative and absolute dose measurementmeasurement

• Gamma factor is a good evaluation parameter

• Some QA cases need more analysis

• QA is a little hard to do in some cases such as spine andQA is a little hard to do in some cases such as spine and lung SBRT

Page 43: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Tips or Problems you may meetTips or Problems you may meet

• Scanning QA phantom with different resolution and extension

P t i h b i t h it i f th PTV• Put ion chamber into homogeneity region of the PTV

• Use high resolution phantom for smaller or thinner PTV

• You can extend EDR2 film range up to 600~700 cGy, however, that is for relative dose measurement only. For absolute dose measurement using film, you can do it under 400 cGy, and it should be good for normal treatment procedure.

• Delta 4 dosimeter is a good candidate for Tomo QA

Page 44: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

S R h T l D l tSome Research Tools Development

Page 45: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Design algorithmDesign algorithm

Hardware related• Target changeTarget change• Jaw• MLC• MLC…Software related• Composite plan• DVH…Clinical/Research need• Archive patientArchive patient…

Page 46: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Design languageDesign language

Excel VBA for clinicalMatlab for researchMatlab for researchVisual c++ for fast applicationXML, html

Page 47: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Tools we have developedTools we have developed

ToPinnacleTomoBinTomoBinTomoAnnualTomoGkPinnacleTomo-XML-DOMTomo-XML-DOMExcel sheets

Page 48: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

ToPinnacleToPinnacle

Need: Physicians need plan comparison between Tomotherapy and Pinnacle andbetween Tomotherapy and Pinnacle and composite plan

Design: Visual c++ program to transfer dose matrix data from Tomotherapy tomatrix data from Tomotherapy to Pinnacle

Page 49: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

ToPinnacleToPinnacle

Page 50: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

ToPinnacleToPinnacle

Page 51: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

TomoBin

Need: Tomo procedure needs sinogram d bi MLC d QAand binary MLC needs QA

Design: Matlab program to design tomotherapy MLC patterntomotherapy MLC pattern

Page 52: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

How were the patterns designed?How were the patterns designed?

Consider dose and mechanical factorsSeven patterns (9 tests) have been designeddesignedA software platform is also designed to

t d l th ttgenerate and analyze the patternsThe patterns can be transferred easilyThe patterns can be transferred easily to other centers with a helical tomotherapy unittomotherapy unit

Page 53: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

TomoAnnualTomoAnnual

Need: How to compare commission data with annual QA data? How to predict ptarget failure?

Design: Matlab program to use commissionDesign: Matlab program to use commission data as baseline for monthly and annual

fprofile comparisonC Shi S Stathakis N Papanikolaou “Development and application of aC Shi, S Stathakis, N Papanikolaou, Development and application of a platform for analyzing helical tomotherapy annual and monthly profiles,” the XVth International Conference on the Use of Computers in Radiation Therapy Toronto Canada June 4 7 abstract # 6102 (2007)Therapy, Toronto, Canada, June 4-7, abstract # 6102 (2007).

Page 54: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

TomoAnnualTomoAnnual

Page 55: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

TomoAnnualTomoAnnual

Profiles collected for theprevious target a-d and after thetarget replacement etarget replacement e.

The angle difference between two profiles for d) is around 1 degree.

Page 56: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

TomoGkPinnacleTomoGkPinnacle

Need: for research use; how to transfer data from one planning station to another?p g

D i M tl b t t dDesign: Matlab program to convert dose matrix between different TPS

C Shi, J Penagaricano, Y Yan, N Papanikolaou, “Application of the Post-Processing Dose Tool (PPD) to Dosimetrically Compare Gamma Knife andProcessing Dose Tool (PPD) to Dosimetrically Compare Gamma Knife and Hi Art Tomotherapy,” SU-FF-T-101, AAPM 48th Annual Meeting (2006)

Page 57: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

TomoGkPinnacleTomoGkPinnacle

Page 58: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

TomoGkPinnacle->APT4DTomoGkPinnacle >APT4D

Lan Lin, Chengyu Shi, Gregory Swanson, Nikos Papanikolaou, "Development of a Novel Post-processing Treatment Planning Platform for 4D Radiotherapy," Technology in Cancer Research and Treatment (April 2008)Technology in Cancer Research and Treatment (April 2008)

Page 59: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Tomo-XML-DOMTomo XML DOM

Need: how to extract necessary information from tomotherapy plan to perform a 2nd check?p

Design: XML language to extract archived patient informationarchived patient information

T Roland CY Shi N Papanikolaou "Development and application of aT Roland, CY Shi, N Papanikolaou, Development and application of a robust toolkit, Tomo-XML-DOM, for helical tomotherapy system," Submitted to 50th AAPM Annual Meeting (2008)

Page 60: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Tomo-XML-DOMTomo XML DOM

Page 61: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Tomo Patient QA 2nd CheckTomo Patient QA 2 Check

Page 62: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Excel sheetsExcel sheets

Need: daily, monthly, annual QA, patient setup, QA…p,

D i l VBADesign: excel VBA

Page 63: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Excel sheets-Annual QAExcel sheets Annual QA

Page 64: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

AAPM TG 148 in-progressAAPM TG 148, in progress

Page 65: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

Where do we go?Where do we go?

• System-based QA vs. component-based QA

• Internet QA with intra/inter institution comparison

• Risk estimation for failure rate and prediction

S lf QA hi d QA• System self QA vs. third-part QA

Page 66: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

SummarySummary

• Current Helical Tomotherapy QA Status is component-based QA and varies from center to p Qcenter

• Future implementation of Helical Tomotherapy QA maybe system-based, internet-based, risk-Q y y , ,estimation, and system-self QA

• AAPM TG 148 will provide recommendations for QA tests and frequency of testingQ q y g

Page 67: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

AcknowledgementsAcknowledgements

• Dr. Papanikolaou• Dr GutierrezDr. Gutierrez• Dr. Lan Lin• Dr. Fan-Chi (Frances) Su• Dr Vikren SarkarDr. Vikren Sarkar• Roland Teboh• Weihong He• Luis Vazquez Quino• Luis Vazquez Quino

Page 68: Helical Tomotherapy Quality Assurancechapter.aapm.org/swaapm/Past/Fall2009/2009_Fall_Papers/Z_Shi_Tomo-QA.pdfHelical Tomotherapy Quality Assurance Cheng Shi PhDChengyu Shi, PhD Department

E il hi @ h dEmail: [email protected]