Automation in Heart-Lung Machine

Prepared by

R.ABDUL MUNEESRENJITH JOHN ABRAHAM

M.VIGNESH

Page 2: Heart Lung Mechine

IntroductionThe Heart-lung machine is a device which is used to provide blood circulation and oxygenation while the heart is stopped during the open heart surgery.

Heart-lung machine consists of three main functional units, the pump, the oxygenator and heat exchanger.

To function, the heart-lung machine must be connected to the patient in a way that allows blood to be removed, processed, and returned to the body.

Page 3: Heart Lung Mechine

HistoryThe researchers searched mainly for a means to provide oxygenated blood to organ preparations. They filtered the blood through various screens and membranes and even pumped it through the lungs of dogs or monkeys, but their problem was not to be solved, though this may be considered the beginning of research into a heart-lung device.

The first successful open heart procedure on a human utilizing the heart lung machine was performed by John Gibbon in 1953. He repaired an atrial septal defect in an 18-year-old woman.

Page 4: Heart Lung Mechine

Purpose of Heart-lung machine:-The heart-lung machine provides the benefit of a

motionless heart during open heart surgery. The heart-lung machine is valuable during this time since

the patient is unable to maintain blood flow to the lungs or the body.

Page 5: Heart Lung Mechine

Surgical procedures in which Heart-lung machine is used:-Coronary artery bypasses surgery.Cardiac valve repair and/or replacement

(aortic valve, mitral valve, tricuspid valve, pulmonic valve)

Repair of large septum defects (atrial septum defect, ventricular septum defect, atrioventricular septum defect)

Transplantation (heart transplantation, lung transplantation, heart–lung transplantation)

Implantation of heart.

Page 6: Heart Lung Mechine

Block diagram:-

Page 7: Heart Lung Mechine

Components of Heart-lung Machine:-CardioplegiaCannulaBlood reservoirHeparin PumpRoller pumps/Centrifugal pumpOxygenatorsHeat Exchanger

Page 8: Heart Lung Mechine

CardioplegiaHeart-lung machine circuit consists of a separate circuit

for infusing a solution into the heart itself to produce cardioplegia to stop the heart from beating, and to provide myocardial protection (to prevent death of heart tissue).

Page 9: Heart Lung Mechine

Cannula

A venous cannula removes oxygen deprived blood from a patient's body. An arterial cannula is sewn into a patient's body and is used to infuse oxygen-rich blood.

A cardioplegia cannula is sewn into the heart to deliver a cardioplegia solution to cause the heart to stop beating

Page 10: Heart Lung Mechine

Blood reservoirThe blood reservoir serves as a receiving chamber for

venous return, facilitates a venous bubble trap, provides a convenient place to add drugs, fluids, or blood, and adds storage capacity for the perfusion system.

Reservoirs may be rigid (hard) plastic or soft, collapsible plastic bags.

Page 11: Heart Lung Mechine

Heparin Pump

Once using a heart-lung machine during open-heart surgery, A powerful anticoagulant, Heparin should be given to the patient in order to reduce the blood's ability to clot, reducing the risk of clots forming in the heart-lung machine.

Page 12: Heart Lung Mechine

Pumping system1.Roller pumps Roller pumps compress a segment of the

blood filled tubing. Blood is then pushed ahead of the moving roller.

Roller pumps consist tubing, which is compressed by two rollers 180° apart. Forward flow is generated by roller compression and flow rate depends upon the diameter of the tubing, rate of rotation.

Excessive compression of the tubing in a roller pump increases the risk of hemolysis.

Page 13: Heart Lung Mechine

Conti

Fig: Diagrams of blood pumps. (A) Roller pumps with two rollers, 180 degrees apart. The compression of the rollers against the raceway is adjustable. Blood is propelled in the direction of rotation. (B) The impeller pump uses vanes mounted on a rotating central shaft. (C) The centrifugal pump uses three rapidly rotated, concentric cones to propel blood forward by centrifugal force

Page 14: Heart Lung Mechine

2 Centrifugal pumpCentrifugal pump have replaced roller

pumps at many institutions; mainly because they cannot over pressurize the heart-lung machine system.

It consists of smooth plastic cones, which when rotated rapidly, propel blood by centrifugal force. An arterial flow meter is required to determine forward blood flow, which varies with the speed of rotation and the after load of the arterial line.

Page 15: Heart Lung Mechine

Oxygenators Oxygenators not only supply vital oxygen for the blood,

but also transport carbon dioxide, anaesthetics and other gases into and out of the circulation.

There are three types of oxygenator 1. Membrane oxygenators2. Bubble oxygenator3. Film oxygenator

Page 16: Heart Lung Mechine

1.Membrane oxygenators Membrane oxygenators imitate the natural lung by a thin membrane of either micro porous polypropylene or silicone rubber between the gas and blood phases.

Diagram of a hollow fiber membrane oxygenator and heat exchanger unit.

Oxygen enters one end of the bundles of hollow fibers and exits at the opposite end. The hollow fiber bundles are potted at each end to separate the blood and gas compartments. Oxygen and carbon dioxide diffuse in opposite directions across the aggregate large surface of the hollow fibers.

Page 17: Heart Lung Mechine

2.Bubble oxygenator In bubble oxygenators, venous blood drains directly into a chamber into which oxygen is infused through a diffusion plate. The diffusion plate produces thousands of small oxygen bubbles within blood. Gas exchange occurs across a thin

film at the blood-gas interface around each bubble. Carbon dioxide diffuses into the bubble and oxygen diffuses outward into blood.

Page 18: Heart Lung Mechine

Heat ExchangerControl body temperature by heating or cooling blood passing through the perfusion circuit.In heat exchanger, blood flows through

spiralling coils made of stainless steel. The inner walls of the coils are coated with polymers to limit blood-surface interactions. The circulating water is chilled to nearly 0 0C in an ice bath and heated by an electric resistance coil to an absolute maximum of 42 0C.

Page 19: Heart Lung Mechine

Stopping of Heart-lung machineOnce operation is completed, The surgeon

gradually occludes the arterial pressure and

pulse, and central venous pressure are monitored and adjusted. When cardiac performance is satisfactory and stable, protamine is given to reverse cardioplegia and blood return from the surgical field is discontinued.

Page 20: Heart Lung Mechine

ComplicationsHaemolysisCapillary leak syndromeClotting of blood in the circuit – can block the

circuit or send a clot into the patient.Air embolismLeakage – lose blood perfusion of tissue if a line

becomes disconnected.1.5% of patients that undergo Heart-lung machine

are at risk of developing Acute Respiratory Distress Syndrome.

Page 21: Heart Lung Mechine

Why Automation in the Heart-lung Machine needed?During open heart surgery, Per fusionists take care of the

connection procedures and maintain the performance of machine as per requirement at the time of surgeries; sometimes there occur faults due to carelessness. To overcome that Automation of Heart-Lung machine is a need.

During surgery, Pressure in the arterial line is monitored continuously to instantly detect any increased resistance to arterial inflow into the patient.

Page 22: Heart Lung Mechine

Computerized CPB SystemThe computerized CPB system consists of an

80286 microcomputer. a reservoir blood level sensor, a collapse sensor, and temperature sensor.

Page 23: Heart Lung Mechine

Safety mechanismsThe computer regularly verifies all samples

with their prior values and detects not only artifacts but also hardware failures.

On detecting the improper sensor output, the computer maintains both the flow rates constant and displays a warning message with estimated locations of the trouble.

The pump rotation speeds are also monitored and matched with the computer's command.

In case the computer breaks down, a perfusionist can take over the pump control by turning an automatic/manual switch to the manual mode.