EP CASE REPORT ............................................................................................................................................................................. Haemoptysis and pulmonary haemorrhage associated with cryoballoon ablation Tolga Aksu*, Sukriye Ebru Golcuk, and Kivanc Yalin Kocaeli Derince Education and Research Hospital, Derince, Kocaeli, Turkey * Corresponding author. E-mail address: [email protected] A 55-year-old male patient underwent pulmonary vein isolation by cryoballoon of symptomatic medication-resistant paroxysmal atrial fibrillation. During cryoablation application at 2508C, sudden haemoptysis was developed. Fluoroscopy demonstrated wedge-shaped consolidation of the left middle zone that may be consistent with pulmonary infarction. The computed tomography scan of the chest confirmed the consolidation in the left lower lobe suggestive of a pulmonary haemorrhage. Background Pulmonary vein isolation (PVI) by cryoballoon has evolved over the past years as an alternative for the treatment of paroxysmal atrial fibrillation (PAF) refractory to anti-arrhythmic drug treatment. Pulmonary haemorrhage is a very rare complication after cryoballoon application. Case A 55-year-old male patient was referred to our hospital for PVI by cryoballoon of drug-resistant PAF. Two PAF episodes per day were observed on the 24 h ambulatory Holter monitoring. Pulmonary vein isolation was performed with a 28 mm cryoballoon (Arctic Front Advanced TM , Medtronic Cryocath). The patient was anti-coagulated with iv heparin and the activated clotting time (ACT) was kept 350 s after a transseptal puncture. Firstly, the left superior pulmonary vein (LSPV) was targeted. After inflation of the cryoballoon within the ostium of LSPV, 2488C tem- peratures were reached during the first application. During the second application at 2508C, sudden haemoptysis was developed ( 100 cc in 15 s). Cryoablation was stopped and the balloon was deflated. Activated clotting time was checked and was 400 s. Patient desaturated to 85% and fluoroscopy demonstrated wedge-shaped consolidation of the left middle zone that may be consistent with pulmonary infarction (Hampton hump-like image) (Figure 1, left panel). Nasal oxygen therapy increased the saturation and haemoptysis was discontinued. The consolidation area did not enlarge on fluoroscopic examination. Other pulmonary veins were isolated successfully within optimal ACT values. The computed tomography (CT) scan of the chest (Figure 1, right panel) confirmed the consolidation in the left lower lobe at the superior boundary suggestive of a pulmonary haemorrhage. Warfarin was started at the third day after the procedure and haemoptysis did not re-occur. Six months after the ablation, no haemoptysis re-occurred and the patient has not any AF episodes since. The chest X-ray abnormality was resolved completely. Published on behalf of the European Society of Cardiology. All rights reserved. & The Author 2015. For permissions please email: [email protected]. Pulmonary hemorrhage after PV Cryo-ballon ablation Cs Cryo-balloon catheter Figure 1 Pulmonary haemorrhage after PV cryoballoon ablation.
2

Haemoptysis and pulmonary haemorrhage associated with cryoballoon ablation

Oct 17, 2022

Download

Documents

Welcome message from author
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
Haemoptysis and pulmonary haemorrhage associated with cryoballoon ablation* Corresponding author. E-mail address: [email protected]
A 55-year-old male patient underwent pulmonary vein isolation by cryoballoon of symptomatic medication-resistant paroxysmal atrial fibrillation. During cryoablation application at 2508C, sudden haemoptysis was developed. Fluoroscopy demonstrated wedge-shaped consolidation of the left middle zone that may be consistent with pulmonary infarction. The computed tomography scan of the chest confirmed the consolidation in the left lower lobe suggestive of a pulmonary haemorrhage.
Background
Pulmonary vein isolation (PVI) by cryoballoon has evolved over the past years as an alternative for the treatment of paroxysmal atrial fibrillation (PAF) refractory to anti-arrhythmic drug treatment. Pulmonary haemorrhage is a very rare complication after cryoballoon application.
Case
A 55-year-old male patient was referred to our hospital for PVI by cryoballoon of drug-resistant PAF. Two PAF episodes per day were observed on the 24 h ambulatory Holter monitoring. Pulmonary vein isolation was performed with a 28 mm cryoballoon (Arctic Front AdvancedTM, Medtronic Cryocath). The patient was anti-coagulated with iv heparin and the activated clotting time (ACT) was kept 350 s after a transseptal puncture.
Firstly, the left superior pulmonary vein (LSPV) was targeted. After inflation of the cryoballoon within the ostium of LSPV, 2488C tem- peratures were reached during the first application. During the second application at 2508C, sudden haemoptysis was developed (100 cc in 15 s). Cryoablation was stopped and the balloon was deflated. Activated clotting time was checked and was 400 s. Patient desaturated to 85% and fluoroscopy demonstrated wedge-shaped consolidation of the left middle zone that may be consistent with pulmonary infarction (Hampton hump-like image) (Figure 1, left panel). Nasal oxygen therapy increased the saturation and haemoptysis was discontinued. The consolidation area did not enlarge on fluoroscopic examination. Other pulmonary veins were isolated successfully within optimal ACT values.
The computed tomography (CT) scanof thechest (Figure1, rightpanel) confirmed theconsolidation in the left lower lobeat the superior boundary suggestive of a pulmonary haemorrhage. Warfarin was started at the third day after the procedure and haemoptysis did not re-occur. Six months after the ablation, no haemoptysis re-occurred and the patient has not any AF episodes since. The chest X-ray abnormality was resolved completely.
Published on behalf of the European Society of Cardiology. All rights reserved. & The Author 2015. For permissions please email: [email protected].
Pulmonary hemorrhage after PV Cryo-ballon ablation
Cs
Discussion
Haemoptysis as a complication of cryoballoon ablation for AF is an uncommon side effect seen in 2.1% of patients.1 In our experience (in 74 cryoballoon applications), this was the first haemoptysis case. Unfortunately, the exact cause of this complication is uncertain and could be due to catheter trauma. Direct damage to the tissue surrounding the pulmonary vein or deeper inside the lung could be caused by cath- eterizing it with the guide wire or by distal inflation of the balloon. In our case, this mechanism could be involved, because causing a vascular rupture through instrumentation, bleeding should be expected to occur immediately. Another possible mechanism might be that a complete isolation with very low-freezing temperatures causes cryo-injury to the adjacent tissues and vasculature. Haemoptysis after the cryoablation procedure was reported in three cases.2,3 Two of these cases were evaluated by the CT scan of the chest. The aetiology was pulmonary infarction2 and pulmonary vein intramural haematoma,3 respectively, in these cases. To our knowledge, our case is the first report of pulmonary haemorrhage as a cause of haemoptysis during cryoablation. We doubt that freezing was the cause, as the bleeding should have occurred following the rewarming.
Conflict of interest: none declared.
References 1. Kumar N, Timmermans C, Das M, Dassen W, Philippens S, Maessen J et al. Hemoptysis after cryoablation for atrial fibrillation: truth or just a myth? Chest 2014;146:e173–5. 2. BhagwandienR,VanBelle Y,de GrootN, JordaensL. Hemoptysis afterpulmonary vein isolationwith acryoballoon: ananalysisof thepotential etiology. J CardiovascElectrophysiol
2011;22:1067–9. 3. Conte G, Chierchia GB, Casado-Arroyo R, Ilsen B, Brugada P. Pulmonary vein intramural hematoma as a complication of cryoballoon ablation of paroxysmal atrial fibrillation.
J Cardiovasc Electrophysiol 2013;24:830–1.
<< /ASCII85EncodePages false /AllowTransparency false /AutoPositionEPSFiles true /AutoRotatePages /PageByPage /Binding /Left /CalGrayProfile () /CalRGBProfile (sRGB IEC61966-2.1) /CalCMYKProfile (U.S. Web Coated \050SWOP\051 v2) /sRGBProfile (sRGB IEC61966-2.1) /CannotEmbedFontPolicy /Warning /CompatibilityLevel 1.5 /CompressObjects /Off /CompressPages true /ConvertImagesToIndexed true /PassThroughJPEGImages false /CreateJobTicket false /DefaultRenderingIntent /Default /DetectBlends true /DetectCurves 0.1000 /ColorConversionStrategy /LeaveColorUnchanged /DoThumbnails false /EmbedAllFonts true /EmbedOpenType false /ParseICCProfilesInComments true /EmbedJobOptions true /DSCReportingLevel 0 /EmitDSCWarnings false /EndPage -1 /ImageMemory 524288 /LockDistillerParams false /MaxSubsetPct 100 /Optimize true /OPM 1 /ParseDSCComments true /ParseDSCCommentsForDocInfo false /PreserveCopyPage true /PreserveDICMYKValues true /PreserveEPSInfo true /PreserveFlatness true /PreserveHalftoneInfo false /PreserveOPIComments true /PreserveOverprintSettings false /StartPage 1 /SubsetFonts true /TransferFunctionInfo /Preserve /UCRandBGInfo /Remove /UsePrologue false /ColorSettingsFile () /AlwaysEmbed [ true ] /NeverEmbed [ true /Courier /Courier-Bold /Courier-BoldOblique /Courier-Oblique /Helvetica /Helvetica-Bold /Helvetica-BoldOblique /Helvetica-Oblique /Symbol /Times-Bold /Times-BoldItalic /Times-Italic /Times-Roman /ZapfDingbats ] /AntiAliasColorImages false /CropColorImages true /ColorImageMinResolution 150 /ColorImageMinResolutionPolicy /OK /DownsampleColorImages true /ColorImageDownsampleType /Bicubic /ColorImageResolution 175 /ColorImageDepth -1 /ColorImageMinDownsampleDepth 1 /ColorImageDownsampleThreshold 1.50286 /EncodeColorImages true /ColorImageFilter /DCTEncode /AutoFilterColorImages false /ColorImageAutoFilterStrategy /JPEG2000 /ColorACSImageDict << /QFactor 0.40 /HSamples [1 1 1 1] /VSamples [1 1 1 1] >> /ColorImageDict << /QFactor 0.40 /HSamples [1 1 1 1] /VSamples [1 1 1 1] >> /JPEG2000ColorACSImageDict << /TileWidth 256 /TileHeight 256 /Quality 20 >> /JPEG2000ColorImageDict << /TileWidth 256 /TileHeight 256 /Quality 15 >> /AntiAliasGrayImages false /CropGrayImages true /GrayImageMinResolution 150 /GrayImageMinResolutionPolicy /OK /DownsampleGrayImages true /GrayImageDownsampleType /Bicubic /GrayImageResolution 175 /GrayImageDepth -1 /GrayImageMinDownsampleDepth 2 /GrayImageDownsampleThreshold 1.50286 /EncodeGrayImages true /GrayImageFilter /DCTEncode /AutoFilterGrayImages false /GrayImageAutoFilterStrategy /JPEG2000 /GrayACSImageDict << /QFactor 0.40 /HSamples [1 1 1 1] /VSamples [1 1 1 1] >> /GrayImageDict << /QFactor 0.40 /HSamples [1 1 1 1] /VSamples [1 1 1 1] >> /JPEG2000GrayACSImageDict << /TileWidth 256 /TileHeight 256 /Quality 20 >> /JPEG2000GrayImageDict << /TileWidth 256 /TileHeight 256 /Quality 15 >> /AntiAliasMonoImages true /CropMonoImages true /MonoImageMinResolution 1200 /MonoImageMinResolutionPolicy /OK /DownsampleMonoImages true /MonoImageDownsampleType /Bicubic /MonoImageResolution 175 /MonoImageDepth 4 /MonoImageDownsampleThreshold 1.50286 /EncodeMonoImages true /MonoImageFilter /CCITTFaxEncode /MonoImageDict << /K -1 >> /AllowPSXObjects true /CheckCompliance [ /None ] /PDFX1aCheck false /PDFX3Check false /PDFXCompliantPDFOnly false /PDFXNoTrimBoxError true /PDFXTrimBoxToMediaBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXSetBleedBoxToMediaBox true /PDFXBleedBoxToTrimBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXOutputIntentProfile (None) /PDFXOutputConditionIdentifier () /PDFXOutputCondition () /PDFXRegistryName () /PDFXTrapped /False /CreateJDFFile false /Description << /ENU () >> >> setdistillerparams << /HWResolution [600 600] /PageSize [612.000 792.000] >> setpagedevice