7/26/2019 Guia de Instalacion 6850
1/48
2010-11-04 8H5320 1
Getting Started with the
DRYVIEW 6850 Laser Imager
Requirements to Perform these ProceduresTwo people will be needed to uncrate and move the Imager. One personcan complete the remaining steps.
The DRYVIEW 6850 Laser Imaging System must be installed by
qualified service personnel. Contact your Carestream Healthrepresentative for details.
Carestream Health, Inc. recommends that the service personnel whoinstall the Imager have the skills and knowledge to perform the following
tasks.
Tools You will need the following tools.
Task Recommended Skill or Knowledge
Uncrate the Imager. Ability to use common hand tools.
Use a personal computer to view the userdocumentation on the CD supplied with theImager.
Basic personal computer skills.
Check AC line voltage using a voltmeter. Basic knowledge of AC electricity and voltmeterfunctions.
Enter the network address of the Imager usingthe touch screen on the Imager.
Familiarity with TCP/IP network addressing conventions.
Configure the Imager to print from a modality. Familiarity with DICOM protocol.
Enter image quality parameters using the touchscreen on the Imager.
Familiarity with image quality adjustment. For example,adjusting image density.
Make a backup disk of the Imager configuration. Understand personal computer file name conventions.
Phillips screwdriver T20 TORX driver with magnetic tip
Pliers
Scissors
Voltmeter - capable of measuring ACvoltages up to 300 volts
16 mm open-end wrench oradjustable wrench that opensto 16 mm
T20 TORX L-key, 130 mm (5 in.) longor less
M6 (6 mm) hex wrench
7/26/2019 Guia de Instalacion 6850
2/48
2 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
Required Information You will need: IP Address and related network information for the Imager
Modality configuration information for each modality that will print tothe Imager
This information should have been previously collected and recorded inthe Site Readiness manual that was provided to your site for planning
purposes. In the Site Readiness manual, refer to the sections titledNetwork Requirements and Modality Configuration.
If you do not have this Site Readiness manual, contact your CarestreamHealth representative.
H210_0605DC-mod3
TORX dr iver
Torxdr iverShor t-handle
TORX L-key
Open-end wrench
Phillips screwdriver
Scissors
Pliers
Hex wrench
7/26/2019 Guia de Instalacion 6850
3/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 3
Completing the UncratingFollowing instructions on the crate, you removed the top of the crate to
access this manual, the Safety Manual, and a Publications Disc with
several user manuals.IMPORTANT: Before completing the uncrating, check that there is an
unobstructed space of 3 m (10 ft) at the front of the crate.The Imager will be unloaded from the front. Also, makesure you are uncrating on a level, horizontal surface.
IMPORTANT: Save all parts of the crate including the plastic and metalclips until the Imager is installed and operational.
1 . Use a Phillips screwdriver to pry out and remove the 11 plastic clipson the left side panel.
2 . Lift and remove the left side panel.
You will use this panel as a ramp later.
3 . Remove and save the 4 boxes that contain accessories.
H210_0599HC-mod2
Left side panel
Plastic clips
3 m
The Imager willbe un loaded fromthe front .
(10 f t)spaceAccessoriesboxes
7/26/2019 Guia de Instalacion 6850
4/48
4 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
4 . Bend the 4 metal clips open on the bottom of the front panel.
5 . Bend the 4 metal clips open on right edge of the front panel.
6 . Lift and remove the front panel.
7 . Bend the 4 metal clips open on the bottom of the rear panel.
8 . Bend the 3 metal clips open on the bottom of the right side panel.
9 . With 2 people, lift and remove the rear and right side panelstogether.
Front pan el
Clips
Clips
H210_0604HC
Right s ide p anel
Rear panel
ClipsClips
7/26/2019 Guia de Instalacion 6850
5/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 5
1 0 . With scissors, cut the foil bag all around the base of the Imager.
1 1 . Remove the foil bag.
NOTE: The bottom of the foil bag, under the Imager, will remain in place.
Foil bag
7/26/2019 Guia de Instalacion 6850
6/48
6 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
Removing the Laser Imager from the Pallet1 . Remove shipping brackets on the left and right sides of the Imager.
Use an M6 (6 mm) hex wrench for the following steps.
a . Remove 4 screws that attach the left shipping bracket to thepallet.
b . Remove 4 screws that attach the left shipping bracket to theImager frame.
c . Remove the left shipping bracket.
d . Remove the right shipping bracket (repeat the above 3 steps).
Save the brackets and screws until you finish installing the Imager.
2 . Prepare the front panel of the shipping crate to be used as a ramp.
a . Open the two ramp supports to the upright position.
b . Use the hook and eye to lock each ramp support in the uprightposition.
4 screwsLeft shippingb r ack e t
4 screwsPallet
H200_0012BAA
Hook
(Ramp)
Ramp Supports
Front PanelEye
7/26/2019 Guia de Instalacion 6850
7/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 7
3 . Turn the ramp over.
4 . Place the ramp so the 2 bolts in the base of the ramp insert into theholes in the metal plate on the pallet.
5 . Check that the 4 stops are all the way up. If not, loosen the locknuton each stop and adjust the stops up as far as possible.
NOTE: If necessary, use a 16mm open-end wrench or adjustablewrench to loosen the locknut.
Bolts
Ramp
Metal platePallet
Locknut
Stop
7/26/2019 Guia de Instalacion 6850
8/48
8 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
CAUTION:Two people are required to move the Imager down the ramp.
Stand in front of the Imager as you control its movement downthe ramp. The force of the Imager down the ramp is about 48
kg (106 lbs). Be careful not to tip the Imager or allow it to rolloff the side of the ramp. The Imager weighs from 266 kg (587
lbs) to 287 kg (632 lbs) depending on whether or not there is asorter.
6 . With 2 people, move the Imager slowly down the ramp.
7 . Position the Imager so there is a space of at least 1 m (39 in.) at thefront, back, and sides of the Imager for removal of shipping
restraints.
7/26/2019 Guia de Instalacion 6850
9/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 9
Two Versions of the ImagerThere are 2 versions of the Imager:
Imager with the optional sorter
Basic Imager (without sorter)
The following figure shows both versions as they appear when removedfrom the shipping crate.
Unless otherwise stated, the procedures in this manual apply to eitherversion.
Im age r with o ptio nal so rte r Bas ic Image r
7/26/2019 Guia de Instalacion 6850
10/48
10 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
Checking AccessoriesCheck that the following items are included with the crate:
1 . A box containing electrical accessories:
4 power cords
2 network cables (long and short)
Set of 5 voltage matching plugs
2 . A box containing 2 preventive maintenance modules (PMM)
3 . A box containing 1 film tray and screws
4 . If the Imager includes the optional sorter, there is a box containing:
4 film trays
6 covers for the sorter mechanism
mounting screws
5 . If the Imager does not have the optional sorter, a box containing a
small plastic cover, screws, and hole plugs is included.NOTE: The User Documentation CD and 2 printed manuals were
removed when the top was removed from the crate. The CDcontains the following manuals: Site Readiness Checklist, UsersGuide, Safety Manual, Quality Control Guide for Mammography,and this manual.
IMPORTANT: You should have read the Safety manual by now. If not,please read it now.
Power CordsCarestream Health provides the following power cords.
Power cord with plug for use in North America
Power cord with plug for use in China Pwer cord with plug for use in the United Kingdom
Power cord with plug for use in Europe
If your Imager is located in another area, check to see if one of these
cords is compatible with the AC power outlets at your Imager site. If not,you must obtain a suitable power cord locally. You must use an
agency-approved power cord with a plug type suitable for your locationand Imager amperage. Contact your Carestream Health representative if
you need assistance.
7/26/2019 Guia de Instalacion 6850
11/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 11
Removing the Shipping Restraints1 . Unlock the shipping restraint in each film drawer.
a . Open the air intake (bottom) door.
b . Pull up and hold the latch release while you pull the upper draweropen.
H210_0525AC
Air intake d oor
Latch release
Upper d rawer
Middle drawer
Lower drawer
7/26/2019 Guia de Instalacion 6850
12/48
12 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
c . Loosen the screw on the heel lock.d . Move the heel lock to the right.
e . Tighten the screw.
f. Close the drawer.
g. Repeat for the middle and lower drawers if present.
H210_0533GC-mod
Heel lock Screw
7/26/2019 Guia de Instalacion 6850
13/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 13
2 . Remove the optics locking bracket and foam pads from the imagingassembly.
a . Pull up and hold the latch release while you pull the main dooropen.
Air intake d oor
Latch release
Main do or
Pu l l doorh e re
H210_0600BH210_0600BC
Optics Module
Foam Pads
Locking Bracket
7/26/2019 Guia de Instalacion 6850
14/48
14 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
b . From the front of the optics module, remove the foam pads.
c . From the rear of the optics module, remove the locking bracket.
3 . Install the preventive maintenance module (PMM).
a . Open the box containing the preventive maintenance module(PMM).
b . To install the PMM, push it in until it locks.
c . Close:
Main door
All drawers
PMM opening
7/26/2019 Guia de Instalacion 6850
15/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 15
4 . Remove shipping restraints inside the back of the Imager.
a . Remove and save 3 screws in the back panel and open the backpanel.
b . Remove 2 screws securing the processor to the frame.
Save these screws until you complete this installation procedure,
then discard.
H210_0821GC-mod
3 Screws
Back panel
Processor
2 screws
7/26/2019 Guia de Instalacion 6850
16/48
16 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
5 . Remove the optics shipping lock plate.
a . Remove and save 2 screws from the left panel.
b . Tilt the left panel back and lift to remove.
H210_0802GC-mod
2 screws
Left pan el
H210_0837GC-mod
Left pan el
7/26/2019 Guia de Instalacion 6850
17/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 17
c . Remove 8 screws from the shipping lock plate.
d . Remove the shipping lock plate.
Save the shipping lock plate and screws until you have completed
installation of the Imager, then discard. The screws may beneeded to install the sorter covers.
H210_084
8 Screws
H210_0841
Shipping Lock Plate
7/26/2019 Guia de Instalacion 6850
18/48
18 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
Verify Removal of Shipping Restraints
CAUTION:If all shipping restraints are not removed from the Imager as
instructed on the preceding pages, the Imager may bedamaged when power is applied.
1 . Check that you have moved the heel locks inside the film drawers tothe unlocked position. (See Page 12.)
2 . Check that you have removed the optics locking bracket and foampads. (See Page 13.)
3 . Check that you have removed the 2 screws that secured theprocessor to the frame. (See Page 15.)
4 . Check that you have removed the optics shipping lock plate. (SeePage 17.)
5 . Close the main door.
6 . Install the left panel. (See Page 16.)
7 . Close the back panel and install the 3 screws.
H210_0821GC-mod
3 Screws
Back panel
Left pan el
http://-/?-http://-/?-7/26/2019 Guia de Instalacion 6850
19/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 19
Installing the Film Trays and Sorter CoversThe Imager comes with either 1 or 5 film trays that you must install. Thenumber of trays depends on whether your Imager has the optional sorter.
If your Imager has the sorter, continue with the "Sorter Option"procedure below.
If your Imager does not have the optional sorter, continue with theNon-Sorter Option on page 28.
Sorter Option If the Imager has the optional sorter, you must install the 5 film trays and7 sorter covers. The following figure shows the items to be installed.
IMPORTANT: You must remove the back cover from the sorter andre-install the back cover later.
Film Tra ys
Front Cover
Back co ver
Back endcover
Top rightcover
Front idlercover
Left idlercover
Top left cover
Screws
Label7F2499
7F2500
Label on b ackCable Ties
7/26/2019 Guia de Instalacion 6850
20/48
20 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
The figure below shows the Imager with the trays and covers in place.Follow the procedure below to install the covers and trays.
Procedure:
1 . Unpack the box containing a single film tray and screws.
2 . Unpack the box containing film trays and covers. The box contains:
4 trays
7 covers; the back cover will be installed on the sorter
Screws to install the trays and covers
3 . On the sorter, separate the 2 wires shown so that you can install ascrew in the screw hole, in a later step, without damaging the wires.
Back end
Front cover
Film trays
Top left cover( beh ind t r ays )
cover
Top rightBack cover
cover
Front end co ver
H210_0593HC
Sorter - Back
2 Wires
Screw hole
7/26/2019 Guia de Instalacion 6850
21/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 21
4 . Raise the sorter.
a . Lift the release latch.
b . Lift the left end of the sorter.
The sorter raises easily and remains up.
5 . Install the left idler cover.
a . Set the cover in place on the 2 pins.
b . Install 2 screws.
Release latch( b l u e )
H210_0547BCA
Left idlercover
Pi n
Screw
Slot
Screw
Pin
Hole
7/26/2019 Guia de Instalacion 6850
22/48
22 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
6 . Open the front end cover by pulling near the top of the cover.
7 . Install the front idler cover.
a . Install screw #1. Do not tighten.b . Position cover with the slot over screw #1.
c . Tighten screw #1.
d . Install and tighten screw #2.
e . Lower the sorter.
f. Close the front end cover.
Front end co ver
H210_0544BC
Screw
Slot
Scre w #2
# 1
Front idler cover
Screw# 1
Screw # 2
7/26/2019 Guia de Instalacion 6850
23/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 23
IMPORTANT: The film trays must be installed starting with the one on theleft and proceeding to the right.
8 . Install film trays.
a . Set the left film tray in place so that the 2 slots near the bottom ofthe tray slide over the 2 pins on the sorter.
b . Install 1 screw in the front and 1 in the back of the film tray.
c . Install each tray in order, from left to right.
9 . Raise the sorter.
H210_0554HCB-mod
Fron t screw
Pi n
Back screw
Left tray
Fron tRightLeft
Front
Slot
7/26/2019 Guia de Instalacion 6850
24/48
24 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
1 0 . Set the front cover in place and install 2 screws in the right end ofthe cover.
1 1 . Install 2 screws in the left end of the front cover.
exp0001a-mod
Front cover
Screw
Screw
Right end
exp0001b-mod2.eps
Screw Screw
Front co ver
7/26/2019 Guia de Instalacion 6850
25/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 25
1 2 . Install the top left cover, 7F2499.
a . Slide the cover into place so that the holes in the 2 brackets fitover the 2 pins.
b . Install 2 screws.
1 3 . Lower the sorter.
exp0002_a-mo
Top left co ver
Pi n
Pi n
2 Screws
Brackets
7/26/2019 Guia de Instalacion 6850
26/48
26 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
1 4 . Install the top r ight cover, 7F2500.
a . Slide the cover into position so the holes in the 2 brackets fit overthe 2 pins.
b . Install 2 screws.
c . Install 2 cable ties.
NOTE: 4 cable ties are supplied. Only 2 are required; the other 2 arespares.
1 5 . Raise the sorter.
H210_0542HC-mod
Pins
Brackets
2 screws
2 screws
End View
Top right co ver
cab leties cab le
7/26/2019 Guia de Instalacion 6850
27/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 27
1 6 . Set the back cover in place and install 4 screws.
1 7 . Set the back end cover in place and install 2 screws.
1 8 . Lower the sorter.
exp0003-mod.eps
Back cover
4 Screws
exp0004.eps
Back end cover
2 Screws
7/26/2019 Guia de Instalacion 6850
28/48
28 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
Non-Sorter Option If there is no sorter on the Imager, install a single film tray and a plasticcover.
The following figure shows the Imager with the tray and cover in place.
Follow the procedure below to install the tray and cover.
Procedure:
1 . Unpack the 2 boxes containing the following components:
H210_0800GC
Film Tray
Back end cover
Film Tray
2 screws
Back end cover
2 screws 4 Hole plugs
7/26/2019 Guia de Instalacion 6850
29/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 29
2 . Install 4 hole plugs.
3 . Open the front end cover by pulling near the top of the cover.
4 . Remove the front end cover (remove and save 2 screws).
H210_0576HC-mod-2
Hole plug
2 Screws
Front end co ver
7/26/2019 Guia de Instalacion 6850
30/48
30 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
5 . Insert the edge of the tray under the sheet metal guard and positionthe 2 slots on the tray over the 2 pins.
Sheet metal guard
Slot
Slot
Pin
Pin
7/26/2019 Guia de Instalacion 6850
31/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 31
NOTE: This view is tilted upward to show the detail below the tray.
6 . Rotate the tray upward and install the 2 screws in the above
drawing.7 . Install the front end cover, removed in step 4.
H210_0578HC-mod2
Screw
Screw
http://-/?-http://-/?-7/26/2019 Guia de Instalacion 6850
32/48
32 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
8 . Install the back end cover (2 screws).
Moving the Imager
1 . Move the Imager and accessories to the location where the Imagerwill be installed.
2 . Position the Imager temporarily so there is a space of at least 1 m(39 in.) at the back of the Imager for installation of the power cord
and network cable.
Back end cover
7/26/2019 Guia de Instalacion 6850
33/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 33
Connecting to Electrical Power1 . Open the box of electrical accessories. It contains:
4 power cords
Set of 5 voltage matching plugs
2 network cables, 2 m (6 ft.) and 10 m (33 ft.)
2 . Select a power cord.
7E9349 - for installations in North American
7F3990 - for installations in China
7F 6848 - for installations in Europe
7F6851 - for installations in the United Kingdom
Check that the power cord you select matches the AC power outletsat your site.
If your Imager is located outside the areas listed above, you may beable to use one of these power cords if the plug matches the power
outlets in your building.
If you cannot use any of the power cords provided, you must obtain
a suitable power cord locally. You must use an agency-approvedpower cord rated for adequate amperage with a plug type suitablefor your location. Contact your Carestream Health representative if
you need assistance.
3 . Use a voltmeter to measure the AC voltage at the power outlet thatwill be used for the Imager.
NOTE: It is very important that you verify the AC voltage.
H200_0032AAA
7/26/2019 Guia de Instalacion 6850
34/48
34 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
4 . From the bag of 5 voltage-selection plugs supplied in theaccessories package, select the plug that matches the measuredwall voltage.
5 . At the back of the Imager:
a . Remove and save 3 screws.
b . Open the back panel.
For Voltage of: Select the Plug with:
90 - 105 Vac Brown wires
106 - 130 Vac Red wires
180 - 210 Vac Orange wires
211 - 230 Vac Yellow wires
231 - 250 Vac Blue wires
H210_0841AA
Voltage Selection Plug
Colored wires
H210_0821GC-mod2
3 Screws
Back panel
7/26/2019 Guia de Instalacion 6850
35/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 35
6 . At the back of the Imager:
a . Loosen 2 screws.
b . Remove the electronics panel.
7 . Insert the correct voltage-selection plug into the connector on thepower supply.
The voltage selection plug is keyed so that it can only be insertedone way. Do not force the plug. When oriented correctly, it will insert
easily.8 . Keep the unused voltage-selection plugs until you have completed
installation of the Imager, then discard the plugs.
9 . Install the electronics panel removed in step 6.
1 0 . Close the back panel and install the 3 screws that fasten the backpanel.
2 screws
Electronics
Panel
210_0842BA
Voltage Selection Plug
http://-/?-http://-/?-7/26/2019 Guia de Instalacion 6850
36/48
36 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
1 1 . Connect the power cord to the AC power connector on the Imager.
RISK OF ELECTRIC SHOCK:Dangerous voltage!
1 2 . Connect the power cord to the building AC power outlet.
AC power con nector
7/26/2019 Guia de Instalacion 6850
37/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 37
Connecting to the NetworkThe Imager receives digital images from medical imaging devices(modalities) over a 10/100Base-T or 1000Base-T Ethernet network. For
10/100Base-T Ethernet networks, 2 CAT 5 UTP cables are provided inthe accessories package for the Imager: 2 m (6 ft.) and 10 m (33 ft.) long.
For 1000Base-T Ethernet networks, the user must provide a CAT 6 UTPcable of sufficient length.
1 . Select one of the network cables.
2 . Connect the network cable to the network connector on the Imager.
3 . Connect the other end of the network cable to the nearest buildingnetwork connector.
Networkcable
Network connector
Networkconnector
7/26/2019 Guia de Instalacion 6850
38/48
38 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
Securing the Imager in its Operating Position1 . Position the Imager where it will normally be used.
2 . Secure the Imager in position.
Rotate the stops clockwise until they contact the floor.
Tighten the lock nuts finger-tight against the base of the Imager.
NOTE: There are 4 stops, one located at each corner of the Imager.
Stop
Locknut
7/26/2019 Guia de Instalacion 6850
39/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 39
Applying Power1 . Check that all doors and panels are closed and secured.
2 . Remove the protective plastic cover from the touch screen.
3 . Press the power switch to the ON (|) position.
The indicator light on the local panel flashes green momentarily.
4 . Press the power button on the local panel.
The touch screen displays a series of start-up screens, then
displays the Change Language screen.
NOTE: If the Imager does not power up, refer to Troubleshooting onpage 45.
Touchscreen
Powerswitch
Powerbutton
Localp an e l
Indicatorlight
7/26/2019 Guia de Instalacion 6850
40/48
40 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
Configuring the ImagerIn the following procedure, you will set up the Imager to suit yourrequirements. After the initial setup steps, an Installation Assistant on
the touch screen will guide you through the configuration procedure.
Most screens in this procedure contain aHelpbutton in the upper rightcorner. Locate this Helpbutton and use it if you need assistance.
Help button
Initial Setup NOTE: If you are asked for a passcode during the procedure, enter thepasscode 123456.
1 . On the Change Language screen, select your language.
2 . Touch OK.
The Change Service Phone screen appears.
3 . Select the service phone number for your country.
If the number for your service center is not in the list, or if it isdifferent than the one in the list, do the following:
a . Touch Enter Number.
A keypad appears.
b . Use the keypad to enter the appropriate phone number.
c . Touch OK.
4 . Touch OK.
The Language and Phone verification screen appears.
7/26/2019 Guia de Instalacion 6850
41/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 41
5 . Are the Language and Service Phone Number correct?
6 . Touch OK.
Follow the instructions on the touch screen to configure the Imager
and complete the installation process. The following stepssupplement the instructions that appear on the touch screen.
During this process you enter the network information for the Imagerand configuration information for each modality that will print to the
Imager.NOTE: For more information on modality setup and image optimization,
see section 4 in the Users Guide, 2G8647.
7 . On the Welcome screen, touch OK.
8 . On the next 4 screens, enter time zone, current date and format, andverify. Touch OKon each screen to continue.
9 . On the Setup -Insert Film screen, touch Unlock, and wait 5seconds.Open a film supply drawer, load a film cartridge, close the drawer,and touch OK.
Network Setup 1 0 . On the Network Configuration screens, enter network information,and touch Saveon the last screen to save the information.
1 1 . If a message states that a reboot is required after entering thenetwork information, select Yesto reboot now. Follow the
instructions to turn off the Imager power.
1 2 . Press the power button on the local panel to restart the Imager.
A network information screen appears after the Imager starts.
1 3 . Verify the network information and touch OK.
1 4 . On the Ping Test screen, touch OK.
Yes No
Touch OK.
The Welcome
screen appears.
a . Touch the Language Edit button or theService Phone Editbutton.
b . Select the correct language or enter thecorrect service phone number, then touchOK.
c . Touch OK.
The Welcome screen appears.
7/26/2019 Guia de Instalacion 6850
42/48
42 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
Remote ManagementServices Setup
1 5 . At Do you want to configure Remote Management Services(RMS)?, touch Yes.
NOTE: CARESTREAM Remote Management Services allows Technical
Support immediate access to the Imager and is a valuable timesavings tool when you need support. To use RemoteManagement Services at your site, you must be able to accessthe Internet using the same subnet that the Imager uses.
1 6 . Enter the name and address location on the next screens, andtouch OK.
1 7 . Touch Registerto start Remote Management Services, and touchOKat RMS registration in progress.
Modality Setup 1 8 . When Setup Send Image appears, send an image from one ofthe modalities you want to set up with the Imager.
NOTE: The default DICOM port number of the Imager is 5040; the
default AE Title is DV6850.1 9 . When A new modality has printed to this Imager. Be sure to
configure it later appears, touch OK.
2 0 . At Setup Send Image, touch OKand wait for the Modality Resetscreen.
2 1 . Touch the name of the modality you printed from and touch Select.
2 2 . On the Sorter Bin screen, select one of the 5 sorter bins to receiveprints from this modality and touch OK.
If your Imager does not have a sorter, the only choice is sorter bin 1.
NOTE: When a modality prints to the Imager for the first time, the Imagersearches for the modality in the Modality Preferences Database(MPDB) in the Imager. The MPDB contains typical image quality
settings for many modalities.
2 3 . This message indicates the results of the search in the MPDB: Thismodality is recognized as below, for which recommended image
quality parameters are available. Do you want to use them?
A second part of the message indicates if the modality was found in
the MPDB.
2 4 . On the next 4 screens, Select Manufacturer, Modality Type, Model,confirm the selections, and touch Apply new defaults.
2 5 . When This will reset all modality parameters! Are you sure?appears, select Yes.
Touch Yes Touch No
If the above message correctlylists the modality type,manufacturerand modelforthe modality you are
configuring.This indicates that the modalityhas been found in the MPDB.
Advance to step 2 5 .
If the above message lists thefollowing parameter names:
RB Defaults Modality Type, RBDefaults Manufacture, RB
Defaults Model.The modality has not beenfound in the MPDB.
Continue with step 2 4 .
http://-/?-http://-/?-http://-/?-http://-/?-7/26/2019 Guia de Instalacion 6850
43/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 43
2 6 . When Modality configuration reset appears, touch OK.
2 7 . At Send another image to the printer from the same modality. TouchOK when the image is printed, print another image from the
modality you are configuring.
NOTE: If you changed the image quality settings in step 24, the imageprints using the new image quality settings.
2 8 . When the new image is printed:
a . Compare the first and second images from the modality.Determine if the second print is acceptable.
b . Touch OK.
2 9 . At If the image quality on the most recent print is not acceptable, . .. ., touch OK.
NOTE: If you have determined that the second print from the modality isnot acceptable, you can adjust the image quality settings later -after you have configured all of the modalities that will print to theImager. Refer to step 35.
3 0 . Do you have another modality to set up?
3 1 . At the Modality Reset screen, touch Exit.
3 2 . At Exit installation, touch Yes.
3 3 . At Back up Current configuration, touch Yes.
The backup operation can take several minutes.3 4 . On the Backup screen, read the instructions and touch OK.
This completes the standard configuration procedure.
NOTE: If the image quality of prints from any of the modalities is notacceptable, manually adjust the individual image qualityparameters for a modality. If you need to make image qualityadjustments, continue with the next step.
3 5 . For additional modality configuration selections, touch the followingbuttons, starting on the Main Screen:Utilities>Configuration>Configure Modality. Refer to the UsersGuide on CD or the General Help About the Imager on page 44formore information.
After the installation process is complete, it is recommended that youreview the General Help information about the Imager (see General
Help About the Imager on page 44).
Yes No
a . Print an image from themodality to the Imager.
b . Repeat the modality setupprocedure starting with
step 19on Page 42.
Continue with step 3 1 .
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-7/26/2019 Guia de Instalacion 6850
44/48
44 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
Help SystemMost screens contain a Helpbutton located in the upper right corner ofthe screen.
General Help About theImager
Information about the functions and features of the Imager is available by
touching Helpon the Main Screen. The following screen appears.
Each tab on the screen contains Help topics relative to the title on thetab. To view information about a particular topic do the following.
1 . Select a:
Tab
Topic
2 . Touch View Topic.
A help page displays with information about the selected topic.
Touch Screen Help Touching the Helpbutton displays a screen that provides informationand/or instructions concerning the screen on which the button wastouched. Closing the Help screen returns you to the original screen.
Tabs
Topics
7/26/2019 Guia de Instalacion 6850
45/48
Getting Started with the DRYVIEW 6850 Laser Imager
2010-11-04 8H5320 45
Discard Excess MaterialWhen you have completed all installation steps and the Imager isoperational, discard the following material.
Shipping crate, pallet, plastic and metal clips, and packagingmaterial
2 L-brackets and 16 screws that secured the Imager to the pallet
4 unused voltage-selection plugs
3 unused power cords
1 unused network cable
2 screws that secured the processor for shipping
Optics locking bracket and foam pads under optics assembly
Troubleshooting
Laser Imager Does NotStart Up
CAUTION:Never pull on the cord to disconnect the plug from the walloutlet. Grasp the plug and pull to disconnect.
1 . Press the power button on the local panel.
If this does not solve the problem, continue with step 2.
2 . Press the power switch, at the bottom of the Imager, to OFF (O).
3 . Disconnect the power cord from the wall outlet.
4 . Check that the power cord is properly connected to the AC powerconnector on the Imager.
5 . Plug the power cord into the wall outlet.
6 . Press the power switch to the ON (|) position.
7 . If the above steps do not start the Imager,
a . Verify that the wall outlet is energized. If voltage is present, go tostep b.
b . Remove the electronics panel and check that the voltageselection plug is seated properly. See step 6on Page 35and step7on Page 35.
8 . If the Imager still does not start up, contact your Carestream Healthrepresentative.
NOTE: When power is turned ON, the Imager goes through a learningperiod and then displays Ready.
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-7/26/2019 Guia de Instalacion 6850
46/48
46 8H5320 2010-11-04
Getting Started with the DRYVIEW 6850 Laser Imager
Imager Does Not Display"Ready" on the TouchScreen
1 . Review and perform the instructions on the touch screen.
2 . If error code 2x-118 or 2x-119 appears on the touch screen, checkthat the heel locks in the film drawers are in the correct position. See
steps c, dand eon Page 12.3 . If the error code 30-932, 30-935, 30-936, 30-937 or 30-938 appears
on the touch screen, check that the film trays and sorter covers have
been installed correctly. Review Installing the Film Trays and SorterCovers on page 19.
If the problem persists, contact your Carestream Healthrepresentative.
Additional Trouble-shooting Information
For more troubleshooting tips, see:
Troubleshooting Help on the touch screen:
1 . Touch theHelpbutton.
2 . Touch Troubleshooting.A list of trouble shooting topics that you can select appears.
Users Guide on the documentation CD supplied with the Imager.
Section 3 in the Users Guidecontains troubleshooting information.
7/26/2019 Guia de Instalacion 6850
47/48
PublicationHistory
Revision Date Reason for Change
A 2010-11-04 First release.
150 Verona Street
Rochester, New York 14608
Carestream Health, Inc., 2010
CARESTREAM and DRYVIEW are trademarks of Carestream Health, Inc.
8H5320Rev. A
7/26/2019 Guia de Instalacion 6850