5 Chapter 2 General Screen Film Radiography and Its Limitations X-rays have been in use for over a century since the report of their discovery by Röentgen (1896). In the health and medical areas, x-rays have been used for both diagnosis and treatment of patients’ conditions. In diagnostic medical imaging (radiography), x-rays are used in a wide variety of examinations. Examples of medical imaging examinations can be found in Table 2.1. Table 2.1 Types of medical imaging examinations using x-rays as the energy source (Ballinger, 1991; Fauber, 2000; Gunn, 2002) general radiography – single projection images fluoroscopy – single projection imaging to display function or anatomical motion angiography – a rapid series of single projection images to capture motion of blood flow computed tomography (CT) – multiple projection imaging to display a cross-section of the anatomy bone mineral densitometry (BMD) – analysis of the bone strength. X-rays are electromagnetic (EM) radiation with characteristics of short wavelengths, very high frequencies and very high energy. High energy EM radiation exhibits characteristics of both waves and discrete bundles of energy called photons. X-ray photon energy is usually measured in kilo-electron volts (keV) and diagnostic x-ray energies are typically 10 to 150 keV. Such high energy EM radiation is ionising. Ionising radiation has potential harmful effects when it irradiates human tissue (Bushberg et al, 2002; Curry et al, 1990; Graham & Cloke, 2003). In diagnostic medical imaging, x-ray production and its attenuation in matter is well described by Bushberg et al, (2002), Bushong (2001), Curry et al (1990), Dowsett, Kenny & Johnston (1998), Graham & Cloke, (2003), Thompson et al (1994) and
20
Embed
General Screen Film Radiography and Its Limitations · the radiographic image. A polyenergetic x-ray beam produced within the x-ray tube anode has a low number of high energy photons
This document is posted to help you gain knowledge. Please leave a comment to let me know what you think about it! Share it to your friends and learn new things together.
Transcript
5
Chapter 2
General Screen Film Radiography and Its Limitations
X-rays have been in use for over a century since the report of their discovery by
Röentgen (1896). In the health and medical areas, x-rays have been used for both
diagnosis and treatment of patients’ conditions. In diagnostic medical imaging
(radiography), x-rays are used in a wide variety of examinations. Examples of
medical imaging examinations can be found in Table 2.1.
Table 2.1 Types of medical imaging examinations using x-rays as the
energy source (Ballinger, 1991; Fauber, 2000; Gunn, 2002)
general radiography – single projection images
fluoroscopy – single projection imaging to display function or anatomical
motion
angiography – a rapid series of single projection images to capture
motion of blood flow
computed tomography (CT) – multiple projection imaging to display a
cross-section of the anatomy
bone mineral densitometry (BMD) – analysis of the bone strength.
X-rays are electromagnetic (EM) radiation with characteristics of short wavelengths,
very high frequencies and very high energy. High energy EM radiation exhibits
characteristics of both waves and discrete bundles of energy called photons. X-ray
photon energy is usually measured in kilo-electron volts (keV) and diagnostic x-ray
energies are typically 10 to 150 keV. Such high energy EM radiation is ionising.
Ionising radiation has potential harmful effects when it irradiates human tissue
(Bushberg et al, 2002; Curry et al, 1990; Graham & Cloke, 2003).
In diagnostic medical imaging, x-ray production and its attenuation in matter is well
described by Bushberg et al, (2002), Bushong (2001), Curry et al (1990), Dowsett,
Kenny & Johnston (1998), Graham & Cloke, (2003), Thompson et al (1994) and
6
Webb (1988). An important consideration for this project is that a heterogeneous
intensity x-ray beam results from differential attenuation within the patient’s
anatomy.
The exit intensities of the x-ray photons will depend on the properties of the body
within the irradiated area and the characteristics of the x-ray beam. The mechanisms
that result in reduction of x-ray intensities are the types of interaction with matter,
such as the photoelectric effect and Compton scatter. Exit intensity of the x-ray beam
can be considered as a function of the probability that an x-ray photon will have an
attenuating event. Properties of the body that affect attenuation are effective atomic
number of the anatomical material, electron density of the anatomical material, and
thickness of the anatomical area or the distance the x-ray photons travel through
anatomy. As the effective atomic number of the tissue increases, the probability of an
interaction of the x-ray photon with the anatomy increases; as electron density of the
anatomy increases, the probability of an interaction of the x-ray photon with the
anatomy increases; and as the x-ray photon’s path distance through the body
increases, the probability of an interaction of the x-ray photon with the anatomy
increases. The degree of attenuation is also affected by the energy of the photons of
the entrance x-ray beam. As x-ray photon energy increases, the probability of an
interaction of the x-ray photons with the anatomy decreases (Bushberg et al, 2002;
Curry et al, 1990; Graham & Cloke, 2003).
X-ray beams are comprised of photons of many different energies, and as such are
polyenergetic. A characteristic of an x-ray beam that affects the rate of attenuation,
and hence the exit intensity of the beam, is its effective energy. Effective energy of
the x-ray beam can be considered as the weighted mean of the x-ray photon energies
within the beam (Bushberg et al, 2002; Carlsson & Carlsson, 1984). A common
means of measurement of the effective energy of the x-ray beam is the measurement
of the half value thickness (HVT) of the material that is irradiated. The HVT is the
thickness of the material irradiated that reduces the intensity of the exit x-ray beam to
half of its entrance intensity. Common materials used for the measurement of
effective energy of an x-ray beam are aluminium (Al) and copper (Cu). HVT, at
diagnostic x-ray energies, is typically stated in millimetres of Al or Cu (Bushberg et
al, 2002; Carlsson & Carlsson, 1984).
7
Attenuation of polyenergetic x-ray beams can then be considered as a function of
their effective energy or HVT. As the effective energy of x-ray beams increases, the
probability of interactions of the x-ray photons decreases. A higher effective energy
x-ray beam will result in a higher exit intensity than when the effective energy of the
entrance beam is lower.
Subject contrast is the differences in exit intensities of the x-ray beam, within the
irradiated field, that result from different rates of attenuation within the irradiated
body. Large subject contrast results from large differences in attenuation rates. Large
subject contrast will typically result from a low effective energy beam as well as
large differences in attenuation (Bushong, 2001; Fauber, 2000; Gunn, 2002).
Subject contrast produces differences in optical densities that can be visualised
within the resulting image. Optical density differences are known as radiographic or
image contrast. The degree of radiographic contrast depends on the subject contrast
and also depends on factors within the image receptor (Fauber, 2000; Gunn, 2002;
Thompson et al, 1994).
2.1 Filtration and Shaping of the X-ray Beam
Polyenergetic x-ray beams used in medical imaging have a maximum photon energy,
which corresponds to the peak kilovoltage (kVp) applied across the x-ray tube. The
kVp is determined by the radiographer, depending upon the desired characteristics of
the radiographic image. A polyenergetic x-ray beam produced within the x-ray tube
anode has a low number of high energy photons and there are an increasing number
of photons at each energy level as the photon energy decreases from the maximum
energy. This is depicted graphically as the dashed line in Figure 2.1 (Curry et al,
1990). In Figure 2.1, the maximum photon energy is 150 keV which results from
setting 150 kVp across the x-ray tube.
8
Figure 2.1 Typical energy spectrum of a 150kVp x-ray beam (Curry et al, 1990,
p.33)
Filtration is the removal of x-ray photons from the beam by attenuation when the
beam is passed through a medium (Bushberg et al, 2002; Curry et al, 1990;
Thompson et al, 1994). There are two types of general filtration: inherent and added
filtration.
X-ray photons are produced by interaction of an accelerated electron with a target
atom. The atoms producing the x-ray photons are often at depth within the target. As
the photons travel within the target itself, attenuation can occur. As the photons
travel further through the x-ray tube, attenuation can again occur within the glass of
the x-ray tube, within the cooling oil that surrounds the glass x-ray tube and within
other parts of the x-ray tube housing and collimator. This is the process of inherent
filtration. Lower energy photons have a high probability of attenuation. Photons with
energies below 15 keV are fully attenuated by the x-ray tube’s inherent filtration
(Bushberg et al, 2002; Curry et al, 1990; Graham & Cloke, 2003; Thompson et al,
1994). The solid line in Figure 2.1 depicts a typical shape of the x-ray spectrum
following inherent filtration of the Bremsstrahlung and characteristic radiation.
9
Added filtration is the addition of other material in the path of the x-ray beam prior to
the x-ray beam entering the patient’s anatomy. The prime purpose of added filtration
is the further removal of low energy photons from the x-ray beam. The effect of
added filtration on the shape of the x-ray spectrum can be seen in Figure 2.2 (Curry
et al, 1990). Low energy photons, if allowed to enter the patient’s body, would have
a high probability of being fully attenuated by the anatomy and hence increasing the
absorbed dose to the patient without contributing to the image (Bushberg et al, 2002;
Curry et al, 1990; Thompson et al, 1994).
Figure 2.2 X-ray spectra at 90kVp with added filtration (filtered) and without
added filtration (unfiltered) (Curry et al, 1990, p.89)
Added filtration will remove more lower energy photons from the beam than higher
energy photons. Added filtration assists in lowering the patient absorbed dose.
Another effect of placing added filtration in the path of the x-ray beam is that of
shifting the effective energy of the beam. In Figure 2.2 it can be seen that the
addition of filtration to the beam has increased the effective energy of the beam.
Added filtration can be any material that is placed in the path of the x-ray beam. It is
usually measured in equivalent thickness of aluminium. When the prime purpose is
10
dose reduction, added filtration material is of uniform thickness across the entire
x-ray field. Minimum amounts of added filtration are usually regulated by statutory
authorities. In Australia, these standards are regulated by various state Acts and
Regulations such as the New South Wales Radiation Control Act (NSW) (1990), the
Radiation Control Regulation (NSW) (2003) and the Radiation Guidelines 6 (2004).
Aluminium and copper are the most common added filtration material (Bushberg et
al, 2002; Curry et al, 1990; Thompson et al, 1994). Typical thicknesses of added
filtration material in diagnostic radiography are 2 – 3 mm of Al or 0.1 mm of Cu
(Carlsson & Carlsson, 1996). The use of erbium filter material has been reported by
Chakera et al (1982), Shrimpton et al (1988) and Cranage et al (1992). Other authors