Oct 22, 2014
1 IntroductionThank you for choosing our GLUCOCARD G+ meter.
• This manual contains operating procedures and precautions for the measurement of blood glucose levels using the GLUCOCARD G+ meter.
• Before using this test meter, thoroughly read this manual to ensure proper testing of blood glucose.
• In addition, carefully read the operating manual of the lancing device for blood sampling, and the package insert for the “G sensor”.
• It is recommended that you keep this manual nearby for use whenever necessary.
This product conforms to the Directive 98/79/EC.
This product may become infectious in the course of use. Discard the product in accordance with local regulations for biohazardous waste.
This product is suitable for self testing.Copyright ©2007 ARKRAY, Inc.
2Documentation ConventionsThis manual uses the following typographical conventions.
Indicates a potentially hazardous situation which, if not avoided, could result in the death of or transmission of serious disease to yourself or another person.
Indicates a potentially hazardous situation which, if not avoided, may result in property damage or minor to moderate injury to yourself or another person.
Indicates information to be kept in mind, such as the causes of malfunctions, notes that should be followed to obtain correct test results, or proper instructions for using functions.
Indicates additional explanations that help you make the best use of the test meter.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
WARNING
CAUTION
NOTE
TIP
3 ContentsIntroduction
Introduction.................................................... 1Documentation Conventions ......................... 2Contents ........................................................ 3Suggested Reading Order............................. 5
Precautions in UsagePrecautions in Usage .................................... 7
PreparationFeatures of the GLUCOCARD G+ Meter .... 15Understanding the Names of Parts ............. 16What's in the Package................................. 17Accessories Required for Blood Glucose Testing ................................ 18
G sensor .................................................. 19Lancing device......................................... 20
TestingTesting Your Blood Glucose........................ 21
Useful FunctionsReviewing Test Results .............................. 37Deleting All Test Results............................. 41Marking the Test Result .............................. 43Recording Meal Start Times ....................... 46Setting the Alarm Time ............................... 50Customizing the Test Meter........................ 54
Entering the Setup Mode “ ” ............. 55Display Order of the Setup Screens ....... 571. Beep tone............................................ 592. Date and time settings ........................ 603. Clock type ........................................... 634. Average test result indication.............. 645. Meal start time recording .................... 656. Alarm................................................... 667. Date and time indications.................... 688. End of the setup mode........................ 69Display when power is off ....................... 70
Replacing the Batteries............................... 71Clock Backup System ............................. 76
Checking the Test Meter (Control Solution Test)................................ 77
Contents 4
FAQFrequently Asked Questions....................... 87Icons and Indications That Appear on the Display ............................................. 99Troubleshooting ........................................ 116
Specifications............................................ 121Symbol Marks ........................................... 123Index ......................................................... 124
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
5 Suggested Reading OrderThis manual consists of the six sections shown on the right.
It is important to read through all the contents of the manual to ensure proper use of the GLUCOCARD G+ meter. However, you must be sure to read the “Precautions in usage”, “Preparation”, and “Testing” sections before using the test meter.
Product specifications and an index are found at the end of this manual. Read these sections for your reference.
Introduction
Precautionsin usage
Preparation
Testing
FAQ
Usefulfunctions
Suggested Reading Order 6
P.1 - 6Describes how to read this manual.
P.7 - 14Lists safety precautions. Be sure to read this section before using the GLUCOCARD G+ meter.
P.15 - 20Describes what you need for glucose testing. Be sure to read this section before testing your blood glucose.
P.21 - 36Describes the instructions for testing blood glucose levels. Be sure to read this section.
P.37 - 86Describes how to review test results, customize your test meter, and use other functions.
P.87 - 119Describes what you should do when the test meter does not work as you expect or when you encounter an error.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
7 Precautions in UsagePlease observe the cautionary notes below for safe use of the GLUCOCARD G+ meter.
Do not use the GLUCOCARD G+ meter while you are undergoing a xylose absorption test and your blood concentration is 8 mg/dL or higher. Such a condition makes blood glucose test results higher than actual levels.
G sensors that were used for blood glucose tests contain blood. To prevent exposure to pathogenic microbes, dispose of the used G sensors according to your doctor’s instructions, being careful not to contact any other person with them.
Keep all G sensors, lancets, and batteries out of the reach of children. If any of them is swallowed, consult a doctor immediately.
WARNING
WARNING
WARNING
Precautions in Usage 8
To prevent infection, be sure to dispose of used lancets. For more information, see the package insert of the lancet.
Used lancets are stained with blood. After using a lancet, cover it with a protective cap. Then, to prevent exposure to pathogenic microbes, dispose of the lancet, being careful not to contact any other person with it. For more information, see the package insert of the lancet.
WARNING IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
WARNING
9
Blood sampling from your forearm is available using the arm type lancing device. However, blood should be sampled from your finger under the following conditions.• You have symptoms of low blood glucose such as, (cold) sweat, a floating feeling, and
shivering.• You should immediately check if your blood glucose level is lower than the designated
level when you notice that your blood glucose has dropped.• There is a possibility that the blood glucose level has changed rapidly, such as after an
exercise.• You caught a cold or are in poor physical condition.
For people who feel sudden changes in their blood glucose level, it has been reported that forearm measurements are sometimes slower to reflect these changes than fingertip measurements.
WARNING
NOTE
Precautions in Usage 10
To prevent infection, be sure to disinfect the area to be lanced. If necessary, apply adhesive plaster to the puncture site to stop bleeding and protect the site. If the wound does not heal, disinfect the site frequently
Do not heat or burn to dispose of used batteries. Otherwise, they may explode and cause injury.
WARNING IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
WARNING
11
Before measurement, take the GLUCOCARD G+ meter and the G sensors out of the carrying case, and leave it out at least 20 to 30 minutes, where the surrounding temperature ranges from 10°C to 40°C and the relative humidity of 20 to 80%. If the test meter is transferred to a different place, a certain period of time will be required to acclimatize it to the ambient environment. The greater the difference in temperature between the test meter and the ambient environment, the longer period will be required for acclimatization. Insufficient acclimatization can produce inaccurate test results because of incorrect temperature compensation. In addition, do not expose the test meter to an area with rapid changes in temperature or with excessive humidity. This may cause moisture to condense inside the test meter, thereby you will get inaccurate test results. The test meter should be kept in storage within a temperature range between 0°C to 50°C.
CAUTION
Precautions in Usage 12
Do not grasp the test meter near the sensor inlet. A built-in thermo-sensor that compensates temperatures to minimize measurement errors is installed inside the inlet. If you hold this area, it will be warmed by your fingers, which will adversely affect test results.
To obtain accurate results and prevent malfunction of the instrument, be sure to keep the sensor inlet clean and prevent the infiltration of blood, moisture, dirt, or dust inside the test meter. When you clean the test meter, use soft cloth moistened with water. Note that this cloth must be thoroughly wrung out prior to use. Wiping the test meter aggressively with antiseptic solution may damage the display.
Do not use a test meter that has been dropped in water, or if any water has entered into the test meter. Even if you dry it, correct test results may not be obtained, and malfunction may result.
CAUTION IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
CAUTION
CAUTION
13
The test meter may perform incorrectly when you use it close to TV sets, microwave ovens, cellular phones, or low-frequency therapeutic devices, all of which radiate electromagnetic waves. To obtain accurate test results, turn off these power switches, or keep them more than 1 meter away.
Only use G sensors with the GLUCOCARD G+ meter. It is not possible to perform a test with a sensor of any other kind. In addition, do not forcibly insert any other sensor into the sensor inlet. Otherwise, the test meter may suffer a breakdown.
G sensors should not be left out of the bottle or held with fingers for a long period of time. Also, do not hold them with wet or dirty fingers. Otherwise correct test results may not be obtained.
Do not use a G sensor that has already made contact with blood or control solution. Otherwise correct test results may not be obtained.
CAUTION
CAUTION
CAUTION
CAUTION
Precautions in Usage 14
Do not use the G sensor after its expiration date. The G sensors contained in the bottle that was opened more than 3 months before now must not be used. Otherwise correct test results may not be obtained.
As soon as you take a G sensor out of the bottle, put the bottle cap on. Otherwise correct test results may not be obtained. Do not use a G sensor from a bottle that has been left uncapped. Use a new G sensor from a new bottle.
Store the control solution tightly capped. Otherwise correct test results may not be obtained.
Make sure that the batteries are loaded in the proper orientation. If it is positioned incorrectly, the test meter may not work correctly and the internal clock settings may be reset.
CAUTION IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
CAUTION
CAUTION
CAUTION
15 Features of the GLUCOCARD G+ MeterThe GLUCOCARD G+ meter is a blood glucose testing device designed for the selfmonitoring of blood glucose.
Features:• You can start testing without calibration.• Blood glucose can be accurately tested with a blood sample volume as low as 0.6 µ.• As soon as the blood is drawn, testing starts automatically. The test result is displayed in
5.5 seconds.• A large display makes it much easier to see test results.• The G sensor can be safely ejected from the test meter without having to contact it with
hands by using the sensor disposal lever.• 450 test results can be stored, and you can review the stored data as required.• By pressing the operation button, you can add a postprandial marker to the test result
obtained after a meal, or record the time when you start eating a meal and view the elapsed time on the display.
• The test meter can sound an alarm to inform you of when to test your blood glucose.
There are many other useful functions (see page 54).
16Understanding the Names of PartsNote the names of the parts below so that you will understand the instructions in this manual.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
Front
Sensor inlet
Display
Operation button
Rear
Battery cover
Sensor disposal lever
Connection terminal
17 What's in the PackageThe GLUCOCARD G+ meter package contains the items shown below. Check that the package contains a full set.
The GLUCOCARD G+ meter contains batteries, and the date and time have been preset before shipment.The date and time can be adjusted as needed (see page 60).
Carrying case
GLUCOCARD G+ meter Operating manual
TIP
TIP
18Accessories Required for Blood Glucose TestingIn addition to the test meter, you will also need the following accessories (sold separately). Prepare everything before starting.
Make sure you have G sensors and lancets that are within their expiration dates.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
G sensor Lancing device “Multi-Lancet S” Adhesive plaster
Sterile cotton
Lancet “Multilet”
NOTE
19
G sensorG sensors are exclusively used for the “GLUCOCARD G+ meter” and “GLUCOCARD G meter”. Before using a G sensor, carefully read the instructions on the package insert.
See the G sensor package insert for the measurement principle.
Bottle
Front
Front Back Back
Blood aspiration tip
Aspiration check window
Insertion portion (electrode)
G sensor
Expiration date
NOTE
Accessories Required for Blood Glucose Testing 20
Lancing device
For handling instructions for the lancing device and lancets, see the package insert and operating manual for the respective products.In addition to the Multi-Lancet S, many other lancing devices and lancets are available for use with the test meter.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
Lancing device “Multi-Lancet S”Lancet “Multilet”
Protective cap Prick button Set lever
Head cap
NOTE
TIP
21 Testing Your Blood GlucoseThe procedure for sampling blood from your finger and measuring the blood glucose level using the GLUCOCARD G+ meter is described here. Make sure you have the following items.
For handling instructions for the lancing device and lancets, see the package insert and operating manual for the respective products.
Log Book
G sensor
Lancing device “Multi-Lancet S”
Adhesive plasterLog book
Sterile cotton
Lancet “Multilet”
GLUCOCARD G+ meter
NOTE
Testing Your Blood Glucose 22
Taking out a G sensor IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• Uncap the bottle and take out a G sensor. As soon as you take it out, cap the bottle.
G sensors should not be left out of the bottle or held with fingers for a long period of time. Also, do not hold them with wet or dirty fingers. Otherwise incorrect test results may be obtained or an error may occur.
G sensors are humidity sensitive. As soon as you take out a G sensor, cap the bottle. Otherwise incorrect test results may be obtained or an error may occur.
Snap
Cap
G sensor
NOTE
NOTE
23
Inserting the G sensor• With the front side of the G sensor facing
you, completely insert the G sensor into the sensor inlet.
• The test meter will turn on with a short beep. Then, the full display will appear as shown at lower-left.
Do not apply blood to the G sensor while this indication is shown on the test meter. Otherwise, the test meter displays “E-8” and cannot start measurement.If you insert the G sensor backwards, “E-6” will be displayed with a long beep. If this occurs, remove the G sensor and insert it again in the correct orientation.
G sensor
Beep Front
NOTE
NOTE
Testing Your Blood Glucose 24
Checking the display IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• When a short beep sounds again, confirm that a blood icon “ ” and the last test result alternately appear.
If no test results are found in memory, (e.g., when you use the test meter for the first time), the blood icon “ ” and unit “ ” will be alternately displayed.
Blood icon
Last test result
Only “mg/dL” appears if no test result is found.
TIP
25
Disinfecting your finger and the lancing device• Disinfect the finger to be lanced and the tip
of the lancing device with sterile cotton. Then, dry them thoroughly.
To prevent infection, be sure to disinfect the relevant areas.
Thoroughly dry the disinfected areas. If blood is sampled before the disinfectant has dried, you will not get accurate test results.
WARNING
NOTE
Testing Your Blood Glucose 26
Sampling blood IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• Insert a lancet into the lancing device (for detailed instructions, see the operating manual for the lancing device).
• Place the tip of the lancing device on your finger and prick it. The lancet pricks your skin just for a moment and automatically retracts back into the lancing device.
To prevent infection, be sure to dispose of the used lancet.
WARNING
27
• Gently push around the puncture site to draw out a bead of blood.
If, usually, an insufficient amount of blood comes out, warm or massage the area to be lanced to improve the circulation of blood prior to sampling.
Reference for amount of blood: Approx. 1.5 mm in diameter
TIP
Testing Your Blood Glucose 28
Checking the display again IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• Check that the blood icon “ ” and the last test result alternately appear.
When 5 minutes have elapsed after the G sensor was inserted, the test meter turns itself off with a double beep. As a result, the indication on the display disappears. In such a case, remove the G sensor and insert it again to turn the test meter back on.
Blood icon
Last test result
TIP
29
Contacting the G sensor to the blood • Hold the GLUCOCARD G+ meter with the
tip of the G sensor downward.
• Bring the punctured finger under the tip of the G sensor, and contact the tip of the G sensor to the blood.
• Allow the G sensor to draw enough blood until the whole surface of the aspiration check window turns red (see the first NOTE on the next page).
• When a short beep sounds, remove your finger. The test meter will automatically start reading the G sensor.
Proper blood contact
Beep
Testing Your Blood Glucose 30Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
If the amount of blood is too low, a long beep will sound, “E-9” will be displayed, and the measurement will not start. In this case, insert a new G sensor into the test meter and reattempt aspiration of the required volume of blood.
Do not contact the G sensor to the blood with the tip of the G sensor upward. If an excessive amount of blood has exuded, it may penetrate the test meter through the sensor inlet and cause damage to the test meter.
The test meter cannot detect the blood if you drop the blood from above the aspiration check window. Be sure to contact the tip of the G sensor to the blood.
NOTEFilled
Aspiration check window
Not filled
NOTE
NOTE
31
During measurement...• The test meter will count down from 4
seconds to indicate the remaining time until it completes the measurement.
Do not remove the G sensor until the test result is displayed.
Counting down
NOTE
Testing Your Blood Glucose 32
Checking the test result IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• 5.5 seconds later, the test result will appear on the display with a short beep. The GLUCOCARD G+ meter will automatically store the test result, date, and time in memory.
If “ ” or “ ” appears, retest your blood glucose using a new G sensor. If one of these indications is displayed again, your condition may be severely aggravated. Immediately consult your doctor.A postprandial marker “ ” or a delete marker “ ” can be added to the obtained test result (see page 43).The stored test results can be reviewed later on the display (see page 37).
Beep
Example: Blood glucose level 118 mg/dL Measured at 7:15 pm on May 22nd
NOTE
NOTE
TIP
33
Disinfecting the puncture site• Disinfect the puncture site using sterile
cotton.
To prevent infection, be sure to disinfect the puncture site. If necessary, apply adhesive plaster to the site to stop bleeding or protect the site. If the wound does not heal, disinfect the site frequently
WARNING
Testing Your Blood Glucose 34
Recording the test result IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• Make a record of the test result in your log book.
Report the test results to your doctor when consulting about future medical treatment.
NOTE
35
Removing the G sensor• Slide the sensor disposal lever to eject the
used G sensor. The GLUCOCARD G+ meter will turn itself off.
To prevent exposure to pathogenic microbes, dispose of the blood-stained G sensor, being careful not to contact any other person with it.
Sensor disposal leverWARNING
Testing Your Blood Glucose 36
Removing the lancet IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• Put the protective cap on the lancet and remove the lancet from the lancing device.
To prevent exposure to pathogenic microbes, a used lancet should not be touched by any other person. Be sure to cover the used lancet with the protective cap and dispose of it in accordance with your doctor’s instructions.
For handling instructions for the lancing device and lancets, see the package insert and operating manual of the respective products.
WARNING
NOTE
37 Reviewing Test ResultsThe GLUCOCARD G+ meter automatically stores up to 450 test results, including blood glucose test results, control test results, and meal start times (see page 46). You can review these test results starting from the most recent one.
If the number of tests exceeds 450, the oldest test result will be deleted.TIP
Reviewing Test Results 38Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
1. Check that the test meter is off. Press the operation button and immediately release it. The GLUCOCARD G+ meter will turn on with a short beep and full display.
Holding down the operation button will put the test meter in a different mode.
2. As soon as “ ” and the current date and time appear, immediately press the operation button again.
The power will turn off in two seconds unless the operation button is pressed soon after “ ” appears.
1.
2.
Power off
Full display
Date and time
Beep
NOTE
NOTE
39
3. The last data appears. The bottom of the display shows “1”, and then changes to the date of the test.
The average test result appears with an “ ” icon when the average test result indication has been set to on (see page 64). Press the operation button again to view the last test result.“ ” will appear following the current date and time indications (see step 2) if no data is found in memory.The figures at left show an example of a blood glucose test result (for display images of other kinds of data, see pages 106 - 108).
3.
Example: Blood glucose test result
Blood glucose level
Test date
Data order
TIP
TIP
TIP
Reviewing Test Results 40Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
4. To view the second most recent test result, press the operation button again.
Each time the operation button is pressed, one older result will appear.Hold down the operation button to move through the results quickly.
5. While the oldest (earliest) test result is being displayed, pressing the operation button proceeds to “ ”. Pressing the operation button again returns the display to the last test result.
To exit the test result review, wait for 3 minutes until the test meter turns itself off with a double-beep.
4.
5.
Next test result
End of test result reviewing
TIP
TIP
TIP
41 Deleting All Test Results
You can delete all test results stored in the GLUCOCARD G+ meter.
Deleted test results cannot be recovered.
1. Review the test results and go to “ ” (see page 40).
2. Hold down the operation button for 5 seconds or more, and release it as soon as the “ ” and “ ” indications alternately appear.
1.
2. 5 sec. or more
Displayed alternately
NOTE
Deleting All Test Results 42Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
3. Hold down the operation button for 5 seconds or more again.
4. “ ” will appear. All of the test results have now been deleted.
After 3 minutes, the test meter will turn itself off with a double-beep.
3.
4.
5 sec. or more
TIP
43 Marking the Test Result
You can add a postprandial marker “ ” or a delete marker “ ” to the obtained test result before storing the result in the GLUCOCARD G+ meter. Such marking is useful for distinguishing the type of test results when you review the results later (see page 37).
: Indicates that the blood glucose levels were measured after a meal.
: Indicates that the value was abnormal. Test results with this marker will also be stored in memory, but not included in the calculation of the average test result (see page 64.)
Postprandial marker
Example:
Delete marker
Marking the Test Result 44Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
1. After a test result is obtained from a blood glucose test (see page 32), leave the G sensor inserted in the test meter.
2. Press the operation button one or more times to select the marker you want.
2.
Unmarked
Postprandial marker Delete marker
45
3. Remove the G sensor from the test meter. The test result will be stored in the GLUCOCARD G+ meter along with the selected marker. Then, the test meter will turn itself off.
Once you remove the GLUCOCARD G sensor from the test meter, it is not possible to change the marker selection.
NOTE
46Recording Meal Start TimesIntroduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
You can record the time when you start eating a meal and view the elapsed time on the display. This is useful towards managing blood glucose levels obtained after meals. The recorded times can be reviewed later on the display (see page 37).
Observe the instructions of your doctor for meal-start-time recording.
Meal start time recording must be turned on before you can record times (see page 65).
Meal start time recording is set to off when you use the test meter for the first time.
Example: Recorded meal start time
Meal start time recording: ON
NOTE
TIP
47
1. Check that the test meter is off, and then hold down the operation button for 2 seconds or more. The test meter will turn on with a beep and full display. Then ,the blinking mealtime icon “ ” will appear.
2. While “ ” is blinking, press the operation button again.
If 5 minutes have passed, with “ ” blinking, before the operation button is pressed, the test meter will turn itself off without storing the meal start time. (When the alarm is on, the test meter will proceed to the alarm time setup screen instead of powering off.)
1.
2.
Power-off
Beep
2 sec. or more
NOTE
Recording Meal Start Times 48Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
As soon as the current date and time appears, they will be stored in memory and the test meter will turn itself off.
3. The time elapsed (in minutes) since the recorded meal start time appears while the power is off.
The elapsed time is indicated in minutes.
The elapsed time is displayed for 3 hours (180 min.) from the meal start time.
When the date and time indications are set to on, the current time will appear next to the elapsed time (see page 101).
3.
Postprandial marker
Power-off
Elapsed time (e.g. 30 min.)
NOTE
TIP
TIP
49
4. When you test your blood glucose level within 3 hours (180 min.) from the last recorded meal start time, the postprandial marker “ ” and the elapsed time will appear below the displayed result.
The elapsed time is indicated in minutes.
The postprandial marker and elapsed time (in minutes) will also appear while reviewing the test result later (see page 37).
4.
Postprandial marker
Test result
Example:Blood glucose level 260 mg/dLTested at 8:15 p.m.Elapsed 60 min. since meal start time
NOTE
NOTE
50Setting the Alarm TimeIntroduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
The test meter can sound an alarm at a specific time every day to inform you of when to test your blood glucose.
The alarm sounds for 30 seconds. Press the operation button to silence it.
The alarm must be turned on before you can set the alarm time (see page 66).
The “ ” icon appears while power is off when the alarm is on. The alarm is set to off when you use the test meter for the first time.
The alarm does not sound if the beep tone setting is set to off.
When the alarm goes off:
Alarm: ON
Double- beeps
NOTE
TIP
51
1. Check that the test meter is off, and then press and hold the operation button for 2 seconds or more. The test meter will turn on with a beep and full display. Then, “ ” will blink.
The “ ” icon will blink instead if meal start time recording is on. This is not an error. Go to step 2.
2. After 5 seconds, you will be prompted to set the alarm time.
1.
2.
Power off
Alarm time setup
Beep
2 sec. or more
After 5 sec.
NOTE
Setting the Alarm Time 52Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
3. Hour setting: With the hour setting blinking, each time you press the operation button, the hour advances by one. 5 seconds after making the setting, the minute setting will start blinking.
4. Minute setting: With the minute setting blinking, each time you press the operation button, the minute advances by one.
In steps 3 and 4, holding down the operation button quickly advances the hour or minute setting.
5. Five seconds after you last press the operation button, your settings will be stored and the test meter will turn itself off.
3.
4.
5.
Power-off
Minute setting
Hour setting
5 sec. later
TIP
53
MEMO
54Customizing the Test MeterThe test meter provides many selectable options. Customize the test meter as your needs.
Function Description DefaultBeep tone Mutes beeps or changes the tone of the beeps (see page
59).High tone
Date and time settings
Sets the internal calendar and clock (see page 60). (Preset)
Clock type Switches between the 12-hour and 24-hour formats of the internal clock (see page 63).
12-hour format
Average test result indication
Turns on/off the average test result indication that appears at the beginning of test result reviewing (see page 64).
Off
Meal start time recording
Turns on/off meal start time recording (see page 65). Off
Alarm Turns on/off the alarm (see page 66). Off
Date and time indications
Turns on/off the date and time indications that appear on the display while the test meter is off (see page 68).
Off
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
55
Entering the Setup Mode “ ”Use the following procedure to enter the setup mode.
1. Check that the power is off, and hold down the operation button for 5 seconds or more.
2. The test meter will turn on with a beep and full display. Then, “ ” will appear, indicating the test meter is now in the setup mode.
Soon after that, the first setup screen for beep tone setting ( ) will appear with a double-beep. Then, the setup screen changes every 5 seconds. Make changes to each setup screen as you need (see page 57).
1.
2.
Power-off
Setup mode
5 sec. or more
Beep TIP
Customizing the Test Meter 56Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
When meal start time recording and/or alarm is on, use the following procedure to enter the setup mode.
1. Check that the power is off, and hold down the operation button for 2 seconds or more.
2. The test meter turn will turn on with a beep and full display. Make sure the blinking “ ” or “ ” icon appears, and hold down the operation button for 5 seconds or more.
3. A beep will sound again and the full display will appear. Then, “ ” will appear, indicating the test meter is now in the setup mode.
NOTE
1.
2.
3.
Power-off
Meal start time recording: ON
Alarm: ON
Setup mode
2 sec. or more
5 sec. or more
57
Display Order of the Setup ScreensOnce the test meter has been in the setup mode ( ), it automatically displays a different setup screen every 5 seconds in the order shown below.
Setup mode Page 55
Year Month
Beep tone Page 59
Date and time settings: Year Page 60
Date and time settings: Month Page 60
Power-off Page 70
End of setup mode Page 69
Date and time indications
Page 68
Press operation button
Customizing the Test Meter 58
The setup screen does not change while you are operating it with the operation button.To exit the setup mode and turn off the test meter, wait for “ ” to appear without pressing the operation button.
TIPTIP
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
Day Hour Minute
Date and time settings: Day Page 60
Date and time settings: Hour Page 60
Date and time settings: Minute Page 60
Alarm Page 66
Meal start time recording
Page 65
Average test result indication
Page 64
Clock type Page 63
59
Beep toneBeeps that sound during operation can be selected from “high tone”, “low tone”, and “off (no sounds)”.
1. Just after “ ” is displayed (see page 55), a double-beep of the currently-selected tone will sound and the first setup screen with “ ” at the lower-left will appear.
2. With “ ” displayed, press the operation button one or more times until the desired option appears.
To exit the setup mode, wait for “ ” to appear. The test meter will then turn itself off.
2.
High tone
Muted Low tone
Double -beep 1.
TIP
Customizing the Test Meter 60
Date and time settings IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
You may need to set the internal calendar and clock, for example, after replacing the batteries.
1. After “ ” is displayed (see page 55), wait for “ ” to appear at the lower-left of the display.
2. Year setting: With the year setting blinking, each time you press the operation button, the year advances by one. 5 seconds after making the setting, you will be prompted to set the month.
When the year reaches “2034”, pressing the operation button brings it back to “2005”.In steps 2, 5 and 6, holding down the operation button quickly forwards the year, hour, or minute setting.
1.
2.
Date/time setting
Year setting
TIP
TIP
61
3. Month setting: With the month setting blinking, each time you press the operation button, the month advances by one. 5 seconds after making the setting, the day setting will start blinking.
4. Day setting: With the day setting blinking, each time you press the operation button, the day advances by one. 5 seconds after making the setting, the hour setting will start blinking.
3.
4.
Day setting
Month setting
Customizing the Test Meter 62Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
5. Hour setting: With the hour setting blinking, each time you press the operation button, the hour advances by one. 5 seconds after making the setting, the minute setting will start blinking.
6. Minute setting: With the minute setting blinking, each time you press the operation button, the minute advances by one.
To exit the setup mode, wait for “ ” to appear. The test meter will then turn itself off.
5.
6.
Minute setting
Hour setting
TIP
63
Clock typeThe clock type can be selected between 12-hour (am/pm) and 24-hour formats.
1. After “ ” is displayed (see page 55), wait for “ ” or “ ” to blink at the lower-right of the display.
2. Press the operation button to select your desired option.
To exit the setup mode, wait for “ ” to appear. The test meter will then turn itself off.
2.
12-hour format
24-hour format
1.
TIP
Customizing the Test Meter 64
Average test result indication IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
Your average test result for the past 14 or 30 days can be displayed at the beginning of test result reviewing (see page 37).
1. After “ ” is displayed (see page 55), wait for “ ” to appear at the lower-left of the display.
2. Press the operation button one or more times to select your desired option.
To exit the setup mode, wait for “ ” to appear. The test meter will then turn itself off.
2.
Not displayed
14-day average 30-day average
1.
TIP
65
Meal start time recordingWhen you want to store meal start times in the GLUCOCARD G+ meter (see page 46), first of all, turn on the meal start time recording as explained in the procedure below.
1. After “ ” is displayed (see page 55), wait for “ ” to appear at the lower-left of the display.
2. Press the operation button to select your desired option.
To exit the setup mode, wait for “ ” to appear. The test meter will then turn itself off.
2.
OFF
ON
1.
TIP
Customizing the Test Meter 66
Alarm IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
To have the test meter sound an alarm at a specific time every day, first of all, turn on the alarm using the following procedure. Then, set the alarm time (see page 50).
1. After “ ” is displayed (see page 55), wait for “ ” to appear at the lower-left of the display.
2. Press the operation button to select your desired option.
To exit the setup mode, wait for “ ” to appear. The test meter will then turn itself off.
2.
OFF
ON
1.
TIP
67
When the alarm is on:
While the test meter is off, it displays “ ” indicating the alarm is on.
When it comes to the set time for the alarm, the test meter sounds continuous double-beeps for 30 seconds. Press the operation button to silence it.
For the first 3 hours (180 min.) after the meal start time was recorded, the time elapsed since recording started appears instead of “ ”. However, the alarm works in the same way as illustrated at left.
When it comes to the set time:
During power-off
Double-beeps
TIP
Customizing the Test Meter 68
Date and time indications IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
While the test meter is off, it can display the current date and time.
The setup screen for the date and time indication does not appear when the alarm has been set to on (see page 66).
1. After “ ” is displayed (see page 55), wait for “ ” to appear at the lower-left of the display.
2. Press the operation button to select your desired option.
To exit the setup mode, wait for “ ” to appear. The test meter will then turn itself off.
2.
OFF
ON
1.
TIP
TIP
69
End of the setup modeWhen you have made all of your settings, leave the test meter as it is without pressing the operation button. The test meter will finally display “ ”, and turn itself off.
This completes the settings in the setup mode.
While “ ” is being displayed, pressing the operation button takes you back to the beep tone setting (see page 59).
Power-off
TIP
Customizing the Test Meter 70
Display when power is offWhile the test meter is off, the display will look like the following according to the settings.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• Standard display(Blank)
• Alarm: ON
• Date and time indications
The current date and time appear.
• Meal start time recording: ONWithin 3 hours (180 min.) since recording started
Date and time indications: OFF
• Meal start time recording: ONWithin 3 hours (180 min.) since recording started
Date and time indications: ON
71 Replacing the Batteries
When the batteries are weak, the battery icon “ ” appears on the display. Immediately replace the batteries with new ones. Prepare two lithium batteries, CR2032. You can purchase them in a convenience store or an electronics store.
If the battery icon is displayed, the test meter may suddenly shut down while measurement is in progress.The battery replacement should be completed within 5 minutes (see page 76.)Data stored in the GLUCOCARD G+ meter will not be deleted after the battery replacement.
Lithium batteries CR2032 × 2
Battery icon
NOTE
TIP
TIP
Replacing the Batteries 72Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
• Remove the battery holder from the rear side of the GLUCOCARD G+ meter. You can detach the holder by sliding it out toward you while pushing down on the portion marked with “ ” on the holder.
Do not touch the metal portion, which appears after the battery holder is removed, with your fingers or any metal.
Battery holder
CAUTION
73
• Remove the current batteries from the battery holder.
Replacing the Batteries 74Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
• Load new batteries in the proper position.
Face the “+” side down when loading batteries.
Make sure that the batteries are loaded in the proper orientation. Incorrect battery loading may keep the test meter from operating. Moreover, the internal calendar and clock may be reset.
Battery orientation NOTE
CAUTION
75
• Slide the battery holder back into the test meter. You will hear a click when the holder is set in place.
Keep all batteries out of the reach of children. If a battery is swallowed, consult a doctor immediately.
Do not heat or burn to dispose of used batteries. Otherwise, they may explode and cause injury.
Batteries must be discarded according to your local environmental regulations.
WARNING
WARNING
NOTE
Replacing the Batteries 76
Clock Backup SystemThe GLUCOCARD G+ meter has a clock backup system that retains the date and time settings for the internal clock during the battery replacement. The clock can be backed up for approximately 5 minutes. If you replace the batteries within 5 minutes, the clock setting will not be required. Note, however, that the clock may be reset under any of the following conditions, even if the batteries are replaced within this period.
• The batteries were not replaced immediately after the battery icon “ ” appeared.• More than 5 minutes has passed since the batteries were loaded in the incorrect
orientation.
If the clock setting has been reset, either the blinking “ ” or “ ” indication appears when you insert a G sensor into the test meter. Correct the clock setting (see page 60).
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
NOTE
77 Checking the Test Meter (Control Solution Test)To ensure the GLUCOCARD G+ meter reads accurate blood glucose levels, perform a control test using the control solution of the specified concentration. You need not perform this control test every time blood glucose is tested. Perform the control test only when:
• You want to check the performance of the GLUCOCARD G+ meter and a G sensor before testing blood glucose.
• Either the GLUCOCARD G+ meter or G sensor appears to be functioning improperly.• You feel unsure because the test result does not correspond to your subjective
symptom.• The GLUCOCARD G+ meter received a shock, for instance, by dropping it on the floor.
The control solution is not included in the GLUCOCARD G+ meter package. Please contact your distributor.Control test results are not included in the calculation of the average test result (see page 64.)
NOTE
NOTE
Checking the Test Meter (Control Solution Test) 78
Items Required
The control solution must be tested within a temperature range of 20°C and 30°C.Read the package insert of the control solution carefully.There are three kinds of control solutions of differing concentrations: L (low range), N (normal range), and H (high range).
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
Log Book
G sensor
Control solution
Log bookGLUCOCARD G+ meter
NOTENOTETIP
79
Taking out a G sensor• Uncap the bottle and take out a G sensor. As
soon as you take it out, cap the bottle
G sensors should not be left out of the bottle or held with fingers for a long period of time. Also, do not hold them with wet or dirty fingers. Otherwise incorrect test results may be obtained or an error may occur.
G sensors are humidity sensitive. As soon as you take out a G sensor, cap the bottle. Otherwise incorrect test results may be obtained or an error may occur.
Snap
Cap
G sensor
NOTE
NOTE
Checking the Test Meter (Control Solution Test) 80
Inserting the G sensor IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• With the front side of the G sensor facing you, completely insert the G sensor into the sensor inlet.
• The test meter will turn on with a short beep. Then, the full display will appear as shown at left.
Do not apply the control solution to the G sensor while this indication is shown on the test meter. Otherwise, the test meter displays “E-8” and cannot start measurement.
G sensor
Beep Front
NOTE
81
Checking the display• When a short beep sounds again, confirm
that a blood icon “ ” and the last test result alternately appear .
Blood icon
Latest test result
Checking the Test Meter (Control Solution Test) 82
Entering the Control Mode IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• With “ ” and the last test result alternately displayed, hold down the operation button for 2 seconds or more.
• The blinking “ ” indication and the “ ” marker will appear. This indicates the test meter is now in the control mode.
Be sure to enter the control mode prior to a control test. If you attempt a control test before entering the control mode, you will obtain an inaccurate test result.
Control mode
2 sec. NOTE
83
Applying the control solution to the G sensor• Place the test meter on a table.
• While the blinking “ ” is displayed, apply the control solution to the tip of the G sensor.
• Allow the G sensor to draw enough blood until the whole surface of the aspiration check window turns red (see the first “NOTE” on the next page).
• When a short beep sounds, remove the control solution bottle. The test meter will then automatically start reading the G sensor.
Control solution
Checking the Test Meter (Control Solution Test) 84Introduction
Precautionsin usage
PreparationTesting
Useful
functionsFA
Q
Draw enough blood until the whole surface of the aspiration check window of the G sensor turns red.
Be sure to cap the control solution tightly after use. If you leave the solution poorly capped, the concentration of the solution will change.
Do not contact the G sensor to the control solution with the tip of the G sensor upward. If an excessive amount of solution has exuded from the bottle, it may penetrate the test meter through the sensor inlet and cause damage to the test meter.
There is no danger of infection from touching the control solution.
NOTE
Filled
Aspiration check window
Not filled
NOTE
NOTE
TIP
85
Judging the test result• After 5.5 seconds, the control test result will
appear with a beep. If the result is within the acceptable range printed on the package insert of the solution, the GLUCOCARD G+ meter and G sensor are working correctly.
If the test result is outside the range, there may be a problem with the test meter or G sensor (see page 92).Record the control test result in your log book for management of the instrument accuracy. The control test result will be automatically stored in the GLUCOCARD G+ meter.Add a “ ” marker to any test result that appears to be abnormal by pressing the operation button (see page 43).
Beep
Example: Test result 100 mg/dL
NOTE
TIP
TIP
TIP
Checking the Test Meter (Control Solution Test) 86
Removing the G sensor IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
• Slide the sensor disposal lever to eject the used G sensor. The GLUCOCARD G+ meter will turn itself off.
There is no danger of infection from touching the control solution.
Sensor disposal lever
TIP
87 Frequently Asked QuestionsSome typical questions about the GLUCOCARD G+ meter will be answered here. Read this section when you are not sure about the operation or if the test meter seems to be defective. If the problem is still not solved or you feel unsure, contact your distributor.
Question 1: Why doesn’t measurement begin after the G sensor comes in contact with the blood?.......................................................................Page 89
Question 2: How should I respond to abnormal test results? ...............................Page 90Question 3: Why doesn’t an indication appear or it appears only for an instant,
after a G sensor was inserted? .........................................................Page 91Question 4: How should I respond when the control test result is not within
the acceptable range?........................................................................Page 92Question 5: How should I handle a situation when “ ” or “ ” appears as
a result of the test?.............................................................................Page 93Question 6: What is the remedy for an error (e.g. E-1)?........................................Page 93Question 7: What does the blinking clock indication,
“ ” or “ ”, mean?..................................................................Page 93Question 8: What does the battery icon “ ” mean? .......................................Page 94Question 9: What is the procedure to reset the temperature error icon “ ”?........Page 94
Frequently Asked Questions 88
Question 10: Why does the beep not sound?......................................................... Page 95Question 11: Why isn’t the average test result displayed? ..................................... Page 95Question 12: Why can’t meal start times be recorded? .......................................... Page 96Question 13: Why doesn't the elapsed time from the meal start time appear
on the display? .................................................................................. Page 97Question 14: Why does the alarm not sound?........................................................ Page 98Question 15: Why aren't the date and time indications displayed
while the power is off? ....................................................................... Page 98
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
89
Question 1: Why doesn’t measurement begin after the G sensor comes in contact with the blood?
Check the following points:• Are you using a new G sensor? Do not reuse G sensors that have made contact with
blood or control solution before.• Did you insert the G sensor completely into the sensor inlet? See page 23.• Did you apply the blood to the G sensor while the blood icon “ ” and the last test result
were alternately displayed? See page 24.• Was the blood properly applied to the tip of the G sensor? See pages 29 - 30.
Frequently Asked Questions 90
Question 2: How should I respond to abnormal test results?
Check the following points:• Are you using a new G sensor? Do not reuse G sensors that have made contact with
blood or control solution before.• Is the G sensor within its expiration date?• Are you using a G sensor from a bottle that was opened within the past 3 months?• Was the bottle of G sensors tightly capped?• Is the GLUCOCARD G+ meter working without displaying the temperature error icon
“ ”? See page 94.• Do the GLUCOCARD G+ meter and the G sensor conform to the ambient temperature/
humidity?• Was the blood properly applied to the tip of the G sensor? See pages 29 - 30.• Perform a control test to see if the test meter and G sensor are working correctly. See
pages 77 - 86.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
91
Question 3: Why doesn’t an indication appear or it appears only for an instant, after a G sensor was inserted?
Check the following points:• Did you insert the G sensor completely into the sensor inlet? See pages 23.• Have the batteries been properly loaded? See page 74.• The batteries may be weak. Replace them with new ones. See pages 71 - 75.
Frequently Asked Questions 92
Question 4: How should I respond when the control test result is not within the acceptable range?
Check the following points:• Was the test performed in the control mode? See page 82.• Are you using a new G sensor? Do not reuse G sensors that have made contact with
blood or control solution before.• Are the G sensor and control solution within their expiration dates?• Are you using a G sensor from a bottle that was opened within the past 3 months?• Was the bottle of G sensors tightly capped?• Was the bottle of the control solution tightly capped?• Was the control solution measured within the temperature range between 20 and 30°C? For more information, see the package insert of the control solution.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
93
Question 5: How should I handle a situation when “ ” or “ ” appears as a result of the test?
• The test meter indicates “ ” if the blood glucose level is higher than 600 mg/dL, and “ ” if it is lower than 10 mg/dL. Retest your blood glucose level using a new G sensor. If you still get the same indication, your condition may have been severely aggravated. Immediately consult your doctor.
Question 6: What is the remedy for an error (e.g. “E-1”)?
• For information on individual error codes, see pages 116 - 119.
Question 7: What does the blinking clock indication, “ ” or “ ”, mean?
• The internal clock has been reset. Correct the date and time settings. See pages 60 - 62.
Frequently Asked Questions 94
Question 8: What does the battery icon “ ” mean?
• The batteries are weak. Immediately replace them with new ones (CR2032). See pages 71 - 75.
Question 9: What is the procedure to reset the temperature error icon “ ”?
• The temperature inside the test meter is between 0°C and 10°C. Use the test results obtained in such environments for reference purposes only because the test meter reading is less accurate than that in the acceptable temperature range.
• To obtain accurate test results, move the GLUCOCARD G+ meter and the G sensor to an area with a temperature range between 10°C and 40°C. Then, leave them for 20-30 minutes or longer. Make sure that the temperature error icon “ ” has disappeared, retest your blood glucose level. See page 11.
If the test is performed with the temperature error icon “ ” displayed, the obtained test result is marked with that icon before being stored in memory. However, this test result is not included in the calculation of the average test result.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
NOTE
95
Question 10: Why does the beep not sound?
• Has the buzzer been turned off? Set the beep tone to “ (high tone)” or “ (low tone)” if you want the test meter to beep. See page 59.
Question 11: Why isn’t the average test result displayed?
• Has the average test result indication been turned off? (It is set to off at the time of purchase of the GLUCOCARD G+ meter.) Set the average test result indication to “ ” or “ ” to view the average test result. See page 64.
Frequently Asked Questions 96
Question 12: Why can’t meal start times be recorded?
• Has the internal clock been reset? Correct the date and time settings before recording a meal start time. See pages 60 - 62.
• Has meal start time recording been turned off? (It is set to off at the time of purchase of the GLUCOCARD G+ meter.) Turn on meal start time recording to record meal start times. See page 65.
Meal start time recording must be turned on before you can record times (see page 46).
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
TIP
97
Question 13: Why doesn't the elapsed time from the meal start time appear on the display?
• Have more than 3 hours passed since the meal start time was recorded? The elapsed time is displayed only for 3 hours after recording starts. See pages 46 - 49.
• Has the internal clock been reset? Correct the date and time settings before recording a meal start time. See pages 60 - 62.
• Were the date and time settings changed after the meal start time was recorded? Making changes to these settings within 3 hours from when the meal start time is recorded disables the elapsed time indication.
• Were the test results deleted from memory after the meal start time was recorded? Deleting the results within 3 hours from when the meal start time is recorded disables the elapsed time indication.
Frequently Asked Questions 98
Question 14: Why does the alarm not sound?
• Has the internal clock been reset? Correct the date and time settings before setting the alarm. See pages 60 - 62.
• Has the alarm been turned off ? (It is set to off at the time of purchase of the GLUCOCARD G+ meter.) Turn on the alarm if you want to use it. See page 66.
The alarm does not sound if the buzzer tone setting is set to off (see page 59).
Question 15: Why aren't the date and time indications displayed while the power is off?
• Has the internal clock been reset? Correct the date and time settings. See pages 60 - 62.• Have the date and time indications been turned off? (It is set to off at the time of
purchase of the GLUCOCARD G+ meter.) Turn on the date and time indications to view the current date and time.
• Has the alarm been turned on? The date and time indications are disabled while the alarm is on.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
TIP
99 Icons and Indications That Appear on the DisplayAll indications and icons you may see on the GLUCOCARD G+ meter are described here.
Memory icon
Delete marker Control marker
Unit icons
Battery icon
Blood icon
Postprandial marker/ Mealtime icon
Temperature error icon
Morning icon Afternoon icon
Icons and Indications That Appear on the Display 100
At power-on
During power-off
Display DescriptionFull display: This initial display appears when the test meter is turned on by inserting a G sensor into the sensor inlet or by pressing the operation button. An automatic check is performed during this indication.
Display DescriptionNo indication: In the default settings, nothing appears on the display while the test meter is off.
Date and time indications: When the date and time indications are turned on, the current date and time appear on the display while the test meter is off. See page 68.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
101
Display DescriptionAlarm icon: When the alarm is turned on (see page 66), the “ ” icon appears on the display while the test meter is off. When it comes to the set time, continuous double-beeps sound for 30 seconds. Press the operation button to silence it.
Elapsed time indication: After a meal start time is recorded and the test meter turns itself off, the elapsed time (in minutes) since recording started is displayed for 3 hours (180 min.).
Elapsed time/date-and-time indications: In addition to the indication shown in “Elapsed time indication”, the current time appears at the right of the elapsed time if the date and time indications have been set to on (see page 68).
Icons and Indications That Appear on the Display 102
Blood glucose testing (normal): See pages 21 - 36.Display Description
Alternating display of the last test result and blood icon: These indications tell you that it is time to apply your blood to the G sensor. If no test results exist in the memory, the blood icon “ ” and unit “ ” are alternately displayed.
Countdown: While testing is underway, the test meter counts down from 4 seconds to indicate the remaining time until it completes the measurement.
Test result: The obtained blood glucose level is displayed.
The indications at left are an example (blood glucose level: 118 mg/dL).
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
TIP
103
Display DescriptionTest result: If your blood glucose level was tested within 3 hours (180 min.) from the last recorded meal start time, the postprandial marker and the elapsed time since recording started appear below the obtained test result.
The indications at left are an example (blood glucose level: 260 mg/dL, elapsed time: 60 min, and test time: 8:15 p.m.).
TIP
Icons and Indications That Appear on the Display 104
Blood glucose testing (abnormal)Display Description
: This indication appears when the blood glucose level is higher than 600 mg/dL. Retest your blood using a new G sensor. If the same indication still appears, immediately consult your doctor.
: This indication appears when the blood glucose level is lower than 10 mg/dL. Retest your blood using a new G sensor. If the same indication still appears, immediately consult your doctor.
Battery icon: The batteries are weak. Immediately replace them with new CR2032 batteries (see pages 71 - 75).
If the battery icon is displayed, the test meter may suddenly shut down while measurement is in progress.
Temperature error icon: This indicates the temperature inside the test meter is too low, making the obtained test result less accurate. To obtain accurate test results, acclimatize the test meter and G sensor in an adequate environment (see page 94).
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
NOTE
105
Test result markers: See pages 43 - 45.
During meal start time recording: See page 46.
Display DescriptionPostprandial marker: This marker appears when you press the operation button while the test result is being displayed. Add this marker to a test result obtained after a meal.
Delete marker: This marker appears when you press the operation button twice while the test result is being displayed. Add this marker to a test result that appears to be abnormal. Note that test results with this marker are not included in the calculation of the average test result.
Display DescriptionMealtime icon: If meal start time recording has been turned on (see page 65), the blinking mealtime icon will appear when the operation button is held down for 2 seconds or more while power is off. While the icon is blinking, pressing the operation button again stores the current time in memory. Record times to start meals in order to manage your blood glucose levels.
Icons and Indications That Appear on the Display 106
Test results review and deletion: See pages 37 - 42.Display Description
Current time: The current date and time are shown when you press the operation button for 2 seconds while the test meter is off. Following this indication, the last test result will appear.
Average test result indication: When the average test result indication has been set to on (either “14d” or “30d”), the average for the past 14 or 30 days appears prior to the last test result.
The indications at left are an example (30-day average result : 123 mg/dL).
Average test result indication (no data): This indicates no test results with dates within the past 14 or 30 days are found in memory although the average test result indication has been set to on (either “14d” or “30d”).
Test result (blood glucose level and order of data precedence): The test result obtained before is shown. The bottom of the display indicates the order of precedence used for data reviewing. “1” means the last test result.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
TIP
107
Display DescriptionTest result (test result and test date): Soon after the test result appears with the order of data precedence, the bottom of the display is replaced with the test date and time.
The indications at left are an example (blood glucose level: 101 mg/dL).
Test result (elapsed time): The postprandial marker and the elapsed time since the last recorded meal start time appear below the test result that was obtained within 3 hours (180 min.) from when the meal start time was recorded.
The indications at left are an example (blood glucose level: 260 mg/dL, elapsed time: 60 min, and test time: 8:15 p.m.).
Test result (postprandial marker): The postprandial marker “ ” appears with a test result that was obtained after a meal and to which this marker is manually added before the result is stored in memory.
TIP
TIP
Icons and Indications That Appear on the Display 108
Display DescriptionTest result (delete marker): The delete marker “ ” appears with a test result that was stored in memory as an abnormal value. This test result is not included in the calculation of the average test result.
Test result (control marker): The control marker “ ” appears with the control test result. This test result is not included in the calculation of the average test result.
Meal start times: Meal start time data contains the mealtime icon “ ” ane the time recorded at the start of a meal.
“ ” : This indication appears following the oldest (earliest) test result to inform you that all data reviewing (up to 450 results) has been completed.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
109
Display DescriptionTest result deletion: “ ” and “ ” are displayed alternately. When you press the operation button for 5 seconds while these indications are being displayed, all of the test results will be deleted from the memory before “dEL” is displayed the next time.
Icons and Indications That Appear on the Display 110
In the setup mode: See pages 54 - 70.Display Description
Setup mode “ ” : The test meter has just entered the setup mode (see pages 55 - 56). From this indication, a different setup screen appears every 5 seconds.
Beep tone setting: Change the beep tone or turn off the beeps (see page 59). A selection can be made each time you press the operation button. The available options include “ (no beeps)”, “ (low tone)”, and “ (high tone)”.
The beep tone is set to “ (high tone)” at the time of purchase of the GLUCOCARD G+ meter.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
TIP
111
Display DescriptionYear setting: Set the year (see page 60). Each time you press the operation button, the number increases by one. The range is from “2005” to “2034”.
Holding down the operation button quickly forwards the setting.
Month, day, hour, and minute settings: Set the date (see pages 61 - 62). Each time you press the operation button, the blinking number increases by one. You will be prompted to make a different setting every 5 seconds.
Holding down the operation button quickly forwards the setting.
Clock type setting: Set the clock type as either 12-hour format (AM/PM) or 24-hour format (see page 63). Each time you press the operation button, the indication is switched between “12:00” and “24:00”.
TIP
TIP
Icons and Indications That Appear on the Display 112
Display DescriptionAverage test result indication setting: Turn on/off the average test result indication (see page 64). Each time you press the operation button, the setting changes to “oFF (no indication)”, “14d (14-day average)” or “30d (30-day average)”. When “14d” or “30d” is selected, you can view the average test result of the past 14 or 30 days at the beginning of test result reviewing.
The average test result is set to “off” at the time of purchase of the GLUCOCARD G+ meter.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
TIP
113
Display DescriptionMeal start time recording setting: Turn on/off meal start time recording (see page 65). Each time you press the operation button, the indication is switched between “on (activated)” and “off (inactivated)”.
Meal-start-time recording is set to “off” at the time of purchase of the GLUCOCARD G+ meter.
Alarm setting: Turn on/off the alarm (see page 66). Each time you press the operation button, the indication is switched between “on (activated)” and “off (inactivated)”.
The alarm is set to “off” at the time of purchase of the GLUCOCARD G+ meter.
TIP
TIP
Icons and Indications That Appear on the Display 114
Display DescriptionDate and time indications setting: Turn on/off the date and time indication that appears on the display while the test meter is off (see page 68). Each time you press the operation button, the indication is switched between “on (activated)” and “off (inactivated)”.
The default setting is “oFF”.
“ ” : This indicates the setup mode is about to end (see page 69). After a while, the test meter automatically turns itself off.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
TIP
115
Control solution testing: See pages 77 - 86.Display Description
Control mode: The test meter is ready for a control test. Apply the control solution to the G sensor.
Control test result: This shows the result of the control test. This value will be stored in memory however it will not be included in the calculation of the average test result.
116Troubleshooting
Display Cause RemedyThe GLUCOCARD G+ meter is faulty.
Contact your distributor.
The temperature is outside of the acceptable range.
Move the GLUCOCARD G+ meter and G sensor to an area with the temperature range between 10°C and 40°C. After they have been left out for 20 to 30 minutes or longer, retest your blood glucose. If “E-2” still appears, contact your distributor.
The GLUCOCARD G+ meter is faulty.
Contact your distributor.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
117
Display Cause Remedy1. Condensation may have
occurred in the GLUCOCARD G+ meter.
2. The GLUCOCARD G+ meter is faulty.
Move the GLUCOCARD G+ meter and the G sensor to an area with the temperature range between 10°C and 40°C and the relative humidity range between 20% and 80%. After they have been left out for 20 to 30 minutes or longer, retest your blood glucose. If the same indication still appears, contact your distributor.
1. The G sensor has been inserted in the incorrect orientation.
2. The G sensor’s insertion area (electrode) may be contaminated with water or alcohol.
1. Check the front and back of the G sensor. Then, insert it into the test meter in the correct orientation.
2. Retest your blood glucose level using a new G sensor.
Troubleshooting 118
Display Cause Remedy1. The inserted sensor is not
appropriate.2. The G sensor’s insertion area
(electrode) may be contaminated with water or alcohol.
1. Insert a new type of G sensor only. Do not use any other commercial sensors.
2. Retest your blood glucose level using a new G sensor.
This error code also appears when an older spec. G sensor has been inserted. Use only new spec. G sensors.
1. The G sensor inserted into the sensor inlet has been stained with blood or control solution.
2. Blood was drawn before the blood icon “ ” and the last test result are alternately displayed.
1. Retest your blood glucose level using a new G sensor.
2. Eject the G sensor and insert it back in. Retest your blood glucose level.
IntroductionPrecautions
in usagePreparation
TestingU
sefulfunctions
FAQ
NOTE
119
Display Cause Remedy1. The amount of blood aspirated is
low.2. Blood did not contact the tip of
the G sensor.
Apply the blood to the tip of a new G sensor and allow the G sensor to draw enough blood to test it. See page 29.
Troubleshooting 120
MEMO
121 Specifications
Name Blood glucose test meter, GLUCOCARD G+ meter
Model GT-1820
Test Glucose in whole blood
Sample volume 0.6 µL
Test strip G sensor (new type only)
Measuring range 10 - 600 mg/dL
Measuring time 5.5 seconds
Temperature compensation
Automatic compensation by using a built-in thermo-sensor
External input/output
A connection terminal for communication
Power source Two 3-volt lithium batteries CR2032 or DL2032
Battery life Approximately 2000 tests The number of available test times may be decreased depending on the usage.
Low battery indication
Steadily lit battery icon
NOTE
Specifications 122
Batteries have been loaded in the GLUCOCARD G+ meter before shipment, thus it is possible to start a test as soon as the test meter is purchased. However, the batteries loaded before shipment of the test meter may not satisfy the battery life stated on the previous page.The specifications and appearance of the product may subject to change for improvements, without prior notice.
Power consumption
0.03 W (max)
Memory 450 test results
Clock accuracy ±80 sec/month
Operating environment
Temperature range 10-40°C (50-104°F), Humidity 20-80% RH(No dew condensation)
Dimensions 53 x 88 x 18 mm
Weight Approximately 50 g
Accessories Operating manual (1 copy), Carrying case (1 pc), Lithium batteries CR2032 (2 pc) loaded
NOTE
NOTE
123 Symbol Marks
Symbol Description Symbol DescriptionIn Vitro Diagnostic Medical Device
Catalogue number
This product conforms to the Directive 98/79/EC.
Caution, consult accompanying documents
Manufacturer Use By
Suitable for self testing Batch code
Serial number Storage temperature limitation
Consult Instructions for Use Authorized Representative in the European Community
Biological Risks
0
50
124IndexIndications
............................................................ 63 ........................................................... 63
....................................................... 41, 109 ............................................... 51, 66, 101 ............................................................. 64 ............................................................. 59
.......................................................... 82, 115 .............................................................. 68
.......................................................... 82 ....................................................... 41, 109 - .................................................... 116 .............................................................. 65
......................................... 40, 69, 108, 114 ............................................................... 104 ............................................................... 104 ....................................................... 55, 110
............................................................. 60
AAlarm ..................................................... 66, 113Average test result indication ......... 64, 106, 112
BBattery cover ..................................................16Battery icon “ ” ............................... 71, 104Beep tone ............................................... 59, 110Blood glucose .................................................21Blood icon “ ” ....................................... 24, 102
CChecking the test meter ......................... 77, 115Clock backup system .....................................76Clock type .............................................. 63, 111Connection terminal .......................................16Control marker “ ” .............................108Control mode .......................................... 82, 115Control solution test ............................... 77, 115Customizing the test meter ............................54
DDate and time indications ....................... 68, 114Date and time settings ........................... 60, 111Delete marker “ ” ...................43, 105, 108Deleting all test results ........................... 41, 106Display ..................................................... 16, 99Display when power is off ..............................70
125
GG sensor ......................................................... 19
LLancet ............................................................ 20Lancing device ............................................... 20
MMarking test results ................................ 43, 105Meal start time recording ........................ 65, 113Mealtime icon “ ” .................. 47, 105, 108Multi-Lancet S ................................................ 20Multilet ............................................................ 20
OOperation button ............................................ 16
PPostprandial marker “ ” ........ 43, 105, 107Power-off ...................................................... 100Power-on ...................................................... 100
RRecording meal start times .................... 46, 105
Replacing the batteries .................................. 71Reviewing test results ............................ 37, 106
SSensor disposal lever .................................... 16Sensor inlet .................................................... 16Setting the alarm time .................................... 50Setup mode ........................................... 55, 110Specifications .............................................. 121
TTemperature error icon “ ” ......................... 104Test result ...................... 32, 102, 106, 107, 115Testing blood glucose .................... 21, 102, 104Troubleshooting ........................................... 116