7/30/2019 Flowmeters Course
1/382
Flow Measurements
1
7/30/2019 Flowmeters Course
2/382
2
7/30/2019 Flowmeters Course
3/382
Definitions and UnitsFlow rate corrections
Differential Pressure Flow Transmitters
Differential Pressure Methods
Orifice Plates
Venturi Tubes
Flow Nozzles
Pitot Tubes
Vortex Type Flow Elements
Target Flowmeter
Turbine Flowmeter
Positive Displacement Flowmeter
Ultrasonic Flowmeter
Coriolis Flowmeter
3
7/30/2019 Flowmeters Course
4/382
4
7/30/2019 Flowmeters Course
5/382
Itis the art and science of:
1. applying instruments2. to sense a chemical or physical process
condition.
5
7/30/2019 Flowmeters Course
6/382
Measurement of a given quantity is an act or
the result of comparison between the
quantity and a predefined standard.
6
7/30/2019 Flowmeters Course
7/382
7
7/30/2019 Flowmeters Course
8/382
In order that the results are meaningful,
there are two basic requirements:
1. The standard used for comparison
purposes must be accurately defined andshould be commonly accepted.
2. The apparatus used and the methodadopted must be proved.
8
7/30/2019 Flowmeters Course
9/382
The advancement of science and technology
is dependent upon a parallel progress in
measurement techniques.
9
7/30/2019 Flowmeters Course
10/382
There are two major functions in all
branches of engineering:
1. Design of equipment and processes.
2. Proper operation and maintenance of
equipment and processes.
Both functions require measurements.
10
7/30/2019 Flowmeters Course
11/382
Direct Method: The unknown quantity is directly
compared against a standard.
Indirect Method: Measurement by direct methods
are not always possible, feasible and practicable.
Indirect methods in most of the cases areinaccurate because of human factors.
They are also less sensitive.11
7/30/2019 Flowmeters Course
12/382
In simple cases, an instrument consists of a
single unit which gives an output reading or
signal according to the unknown variable
applied to it.
In more complex situations, a measuring
instrument consists of several separateelements.
Instruments
12
7/30/2019 Flowmeters Course
13/382
These elements may consist of:
Transducer elements which convert the
measurand to an analogous form.
The analogous signal is then processed by
some intermediate means and then fed to
The end devices to present the results forthe purposes of display and or control.
13
7/30/2019 Flowmeters Course
14/382
These elements are:
A detector.
An intermediate transfer device.An indicator.
14
7/30/2019 Flowmeters Course
15/382
The history of development of instruments
encompasses three phases:
Mechanical.
Electrical.
Electronic.
15
7/30/2019 Flowmeters Course
16/382
Reaching corporate economic goals
Controlling a process
Maintaining safety
Providing product quality
Purpose of Process
Measurement
16
7/30/2019 Flowmeters Course
17/382
No matter how advanced or sophisticated the
distributed control system,
the control system is only as effective as the
process measurement instruments it isconnected to;
therefore, successful process control is
dependent on successful instrument
application.
17
7/30/2019 Flowmeters Course
18/382
To correctly apply instrumentation, an
engineer must clearly understand the
operations and limitations of the instrument,
as well as understanding the chemical and
physical properties of the process.
18
7/30/2019 Flowmeters Course
19/382
Fundamental to applying process
instrumentation is interpreting the instruments
performance envelope.
Every field measurement device has its own
distinct envelope that constitutes the process andenvironmental conditions it can perform to.
Likewise, every application has a characteristic
envelope that represents the application's
process and environmental conditions.
19
7/30/2019 Flowmeters Course
20/382
20
7/30/2019 Flowmeters Course
21/382
It is the science of measurement.
As a science, metrology uses terminology and
definitions that the process measurementengineer must be familiar with.
He must and have a clear understanding of,
because vendors may vary in the use of a term.
Metrology
21
7/30/2019 Flowmeters Course
22/382
The instrument engineer must consider the
following dynamic conditions that affect
process measurement:
Temperature Effects
Static Pressure Effects Interference
Instrumentation Response
Noise Damping and Digital Filtering
22
7/30/2019 Flowmeters Course
23/382
These dynamic conditions cause the presence ofuncertainty in measuring systems.
No measurement, however precise or repeated,
can ever completely eliminate this uncertainty.
The uncertainty of measuring systems is
exemplified in the effects temperature variationscan have on measurements.
23
7/30/2019 Flowmeters Course
24/382
Temperature influences can exhibit some of the
most severe effects on a process measurement,
both in the process media itself and the
measurement instrument.
Temperature Effects
24
7/30/2019 Flowmeters Course
25/382
Some obvious examples of severe temperatureinfluences include temperature-induced phase
transitions.
25
It would be hard to determine what would happen to an
orifice plate, differential pressure measurement if theprocess suddenly changed from a liquid to a solid or gas.
7/30/2019 Flowmeters Course
26/382
Other temperature induce dynamic changes
include:
Change in the dimensions of the measuring
element,
Modification of a resistance of a circuit, or
Temperature-induced change in the flux density
of a magnetic element.
26
7/30/2019 Flowmeters Course
27/382
Similar to temperature effects, pressure changes canalso trigger phase transitions, especially in gas
applications.
Pressure effects seen in differential pressure (DP)
devices are an example.
Because the differential pressure devices are used in
flow and level applications, the importance of pressure
effects should not be underestimated.
27
7/30/2019 Flowmeters Course
28/382
The goal is to minimize the total error that pressure
effects can cause.
To illustrate this, consider a differential pressure
instrument that is calibrated in a lab at zero static
pressure.
The transmitter is re-zeroed after installation by
opening an equalizing valve in the process under
pressure to eliminate zero shifts;
however, variations inline pressure are not accounted
for during normal operations.
28
7/30/2019 Flowmeters Course
29/382
Interference, in process measurement terms, refers toeither external power or electrical potential that can
interfere with the reception of a desired signal or the
disturbance of a process measurement signal.
29
7/30/2019 Flowmeters Course
30/382
The dynamic characteristic of instrumentation
response refers to how quickly a measuring instrument
reacts or responds to a measured variable.
An ideal, perfect instrument would have an
instantaneous response, which in effect, is called zero
lag.
In general, with modern electronic instrumentation,the response time is adequate for most applications.
Instrumentation Response
30
7/30/2019 Flowmeters Course
31/382
7/30/2019 Flowmeters Course
32/382
Noise is often described as a signal that doesnot represent actual process measurement
information.
Noise can originate internally within the process
measuring system or externally from the process
condition.
It makes up part of the total signal from which
the desired signal must be read.32
7/30/2019 Flowmeters Course
33/382
Damping is defined as the progressive reductionor suppression of oscillation in a device or system.
In more practical terms, damping describes theinstruments performance in the way a pointer or
indicator settles into a steady indication after a
change in the value of the measured quantity.
Damping and Digital Filtering
33
7/30/2019 Flowmeters Course
34/382
A response is not damped at all, oscillation
continues.
A response is underdamped or periodic, as is the
case when overshoot occurs.
A response is overdamped or aperiodic, when the
response is slower than an ideal or desired
condition.
A response is critically damped, when the
response represents an ideal or desired condition.
34
7/30/2019 Flowmeters Course
35/382
Range
It is defined as the region between the limits
within which a quantity is measured, received,or transmitted, expressed by stating the lower
and upper range values.
Measurement Terminology
35
7/30/2019 Flowmeters Course
36/382
Upper Range Value (URV) is defined as the highest
quantity that an instrument is adjusted to measure.
Lower Range Value (LRV) is defined as the lowest
quantity that an instrument is adjusted to measure.
Upper Range Limit (URL) is defined as the maximumacceptable value that a device can be adjusted to
measure.
Lower Range Limit (LRL) is defined as the minimum
acceptable value that a device can be adjusted to
measure.
36
7/30/2019 Flowmeters Course
37/382
37
7/30/2019 Flowmeters Course
38/382
Rangeability is the ratio of the maximum
measurable value to the minimum measurable
value.
Turndown is defined as the ratio of the normal
maximum measured variable through the
measuring device to the minimum controllablemeasured variable.
In a conventional differential pressure transmitter,if the maximum pressure is 7.45 kPa and the
minimum pressure is 1.24 kPa, the span turndown
is 6 to 1 (6:1).38
7/30/2019 Flowmeters Course
39/382
39
These terms are often interchanged, confused and
misunderstood.
7/30/2019 Flowmeters Course
40/382
40
7/30/2019 Flowmeters Course
41/382
41
7/30/2019 Flowmeters Course
42/382
Zero Elevation Range is defined as a range wherethe zero value of the measured variable is greater
than the lower range value.
The zero value can be between the lower range
value and the upper range value, at the upper range
value, or above the upper range value.
42
7/30/2019 Flowmeters Course
43/382
7/30/2019 Flowmeters Course
44/382
44
7/30/2019 Flowmeters Course
45/382
Response Time is defined as the time taken for the
system output to rise from 0% to the first crossover
point of 100% of the final steady state value.
45
7/30/2019 Flowmeters Course
46/382
Accuracy is sometimes referred to as the maximum
uncertainty or limit of uncertainty.
In practical terms, accuracy qualitatively represents the
freedom from mistake or error.
In metrological terms, accuracy represents the degreeof conformity of an indicated value to an accepted
standard value, or ideal value.
46
7/30/2019 Flowmeters Course
47/382
Precision is confused with accuracy.
Precision, by definition, is the reproducibility withwhich repeated measurements of the same measured
variable can be made under identical conditions.
47
7/30/2019 Flowmeters Course
48/382
48
7/30/2019 Flowmeters Course
49/382
49
7/30/2019 Flowmeters Course
50/382
Reproducibility is the same as precision.
The close agreement among repeatedmeasurements of the output for the same value
input that are made under the same operating
conditions over a period of time, approaching from
both directions.
If the measuring instrument is given the same
inputs on a number of occasions and the results lieclosely together, the instrument is said to be of high
precision.
50
7/30/2019 Flowmeters Course
51/382
Repeatability
It is same as reproducibility except thatrepeatability represents the closeness of agreement
among a number of consecutive measurements of
the output for the same value of input under thesame operating conditions over a period of time
(approaching from the same direction).
51
7/30/2019 Flowmeters Course
52/382
Linearity is the closeness to which a curve
approximates a straight line.
Independent LinearityTerminal Linearity
Zero-based Linearity
52
7/30/2019 Flowmeters Course
53/382
Hysteresis
53
7/30/2019 Flowmeters Course
54/382
Deadband
54
7/30/2019 Flowmeters Course
55/382
Drift
It represents an undesired slow change or
amount of variation in the output signal over a
period of time (days, months, or years), with a
fixed reference input.
55
7/30/2019 Flowmeters Course
56/382
Zero Drift represents drift with zero input signal.
In practical terms, the zero of the measuringinstrument shifts.
In a mechanical instrument, it is usually caused by a
slipping linkage. The correction is to re-zero the
instrument.
In an electronic instrument, zero shift is usually causedby environmental changes. The correction is to re-zero
the electronic instrument.
56
7/30/2019 Flowmeters Course
57/382
Span Drift represents drift or gradual change in
calibration as the measurement moves up the scale
from zero.
In a mechanical instrument, it is usually caused by
changes in the spring constant of the instrument, or by
the linkage.
In a electronic instrument, span shift is usually caused
by changes in the characteristics of a component.
The correction can be to adjust the span of the display
element.
57
7/30/2019 Flowmeters Course
58/382
Partial Drift represents drift on only a portion of the
instruments span.
In a mechanical instrument, it is usually caused by an
overstressed part of the measuring instrument.
In an electronic instrument, partial shift is usually
caused by drift in an electronic component.
The correction is periodic inspection and calibration.
58
R li bili
7/30/2019 Flowmeters Course
59/382
Reliability
It represents a measuring devicesability to perform a
measurement function without failure over a specified
period of time or amount of use.
Usually reliability data is extrapolated.
Reliability is often expressed as (MTTF) specification.
After failure, repair must take place.
MTTF + MTTR = MTBF
59
7/30/2019 Flowmeters Course
60/382
Process measurement suppliers tend to follow severalrules when designing equipment to achieve reliability.
Keep the design simple,
Avoid using glass as a structural material,
Keep electronics cool as possible,
Provide easy serviceability.
Overview of Typical Design Criteria
60
7/30/2019 Flowmeters Course
61/382
Housing
Metals
Gasket Considerations
Seal Considerations Associated Hardware Options
Process Connections Options
Installation Orientation Effects of Vibration
Environment and Hardware Materials
61
7/30/2019 Flowmeters Course
62/382
Environment and Hardware Materials
Reliability Quality
Accuracy
Cost
Repeatability
Previous acceptance
Availability of spares
Compatibility with existing equipments
Flexibility of use
Compatibility with the environments Ease of maintenance
Ease of operation
Application suitability
62
l l d d
7/30/2019 Flowmeters Course
63/382
Electrical design and instrument
loop wiring considerations
Power Requirements
Power Consumption
Wiring Terminations Output Signal
RFI Effects
Grounding of Instruments Shielding Considerations
Lightning Protection
63
7/30/2019 Flowmeters Course
64/382
SAFETY CONSIDERATIONS
limiting the energy level
keeping sparks away from flammable mixtures
containing an explosion
diluting the gas level protecting against excessive temperature
Probability that a hazardous gas is present
Quantity of a hazardous gas
Nature of the gas (is it heavier or lighter than air) The amount of ventilation
The consequences of an explosion
64
7/30/2019 Flowmeters Course
65/382
65
7/30/2019 Flowmeters Course
66/382
66
7/30/2019 Flowmeters Course
67/382
Pressure values themselves are essential data
for monitoring.
Often, the values of process variables other
than pressure are derived from (inferred from)
the values that are measured for pressure.
67
P i f M i R l i P
7/30/2019 Flowmeters Course
68/382
Properties of Matter in Relation to Pressure
Measurement
68
7/30/2019 Flowmeters Course
69/382
Pressure Equation
Pressure is defined as the amount of force per unitarea.
P =F/A
where:
P = pressureF = force =ma
A = area
69
G Ab l t Diff ti l d V
7/30/2019 Flowmeters Course
70/382
Gauge, Absolute, Differential, and Vacuum
70
P M i D i
7/30/2019 Flowmeters Course
71/382
Pressure Measuring Devices
Categories of pressure measuring devices :
Gravitational gauges
Deformation sensors and switches
Transducers and transmitters
71
Gravitational Gauges
7/30/2019 Flowmeters Course
72/382
Gravitational Gauges
72
Deformation (Elastic) Sensors and Switches
7/30/2019 Flowmeters Course
73/382
Deformation (Elastic) Sensors and Switches
Bourdon Tube
73
7/30/2019 Flowmeters Course
74/382
74
7/30/2019 Flowmeters Course
75/382
Advantages
They are available in a wide variety of pressure ranges. They are proven and suitable for many pressure
applications.
They have good accuracy.
Disadvantages
Vibration and shock could be harmful to mechanicallinkage.
They are susceptible to hysteresis as they age.
75
7/30/2019 Flowmeters Course
76/382
Bellows
76
Diaphragm
7/30/2019 Flowmeters Course
77/382
Pressure Transducer
7/30/2019 Flowmeters Course
78/382
Pressure Transducer
It is a device that provides an electrical output signal that is
proportional to the applied process pressure.
The output signal is specified as either a volt, current, or
frequency output.
A pressure transducer always consists of two elements:
A force summing element, such as a diaphragm, converts
the unknown pressure into a measurable displacement or
force.
A sensor, such as a strain gauge, converts the displacement
or force into a usable, proportional output signal.
78
Strain Gauge
7/30/2019 Flowmeters Course
79/382
Strain Gauge
The sensor changes its electrical resistance when it
stretches or compresses.
79
Potentiometer Element
7/30/2019 Flowmeters Course
80/382
Potentiometer Element
80
Capacitive Sensor
7/30/2019 Flowmeters Course
81/382
Capacitive Sensor
81
Performance Advantages
7/30/2019 Flowmeters Course
82/382
Performance Advantages
They have good rangeability and response time.
They have very good accuracy.
Typical accuracies are about 0.1% of reading or 0.01 % of
full scale.
Typical transducers support a very wide pressure range.
High vacuum and low differential pressure ranges are
supported.
82
Inductance-Type Transducer
7/30/2019 Flowmeters Course
83/382
yp
Changing the spacing between two magnetic devices causes a
change in the reluctance.
The change in reluctance then represents the change in pressure.
One type of reluctance pressure transducer is the linear variable
differential transformer (LVDT).
83
Piezoelectric Gauge
7/30/2019 Flowmeters Course
84/382
Piezoelectric Gauge
Materials that create an electrical voltage when a force is
applied.
They measure rapidly changing pressures.
84
7/30/2019 Flowmeters Course
85/382
Performance Advantages
They provide a self generated output signal.
They have high speed of response.
They have good accuracy, about 1% of full
scale is typical.
85
Design of Pressure Transmitters
7/30/2019 Flowmeters Course
86/382
Design of Pressure Transmitters
86
Meter Body Designs
7/30/2019 Flowmeters Course
87/382
Meter Body Designs
87
Transmitter Process Locations
7/30/2019 Flowmeters Course
88/382
88
7/30/2019 Flowmeters Course
89/382
89
7/30/2019 Flowmeters Course
90/382
90
7/30/2019 Flowmeters Course
91/382
91
Purpose of Flow Measurement
7/30/2019 Flowmeters Course
92/382
Monitor and control the flow rates.
Develop material and energy balances.
Sustain the efficiency and to minimize
waste.
Purpose of Flow Measurement
92
Importance of Accurate
7/30/2019 Flowmeters Course
93/382
Material balances in separation processes.
Pumps and compressor operations.
Custody transfer operations.
Importance of Accurate
Measurement
93
Flowmeter Definition
7/30/2019 Flowmeters Course
94/382
A flowmeter is defined as A device that
measures the rate of flow or quantity of a
moving fluid in an open or closed conduit.
Flowmeter Definition
It usually consists of a primary device and asecondary device.
94
7/30/2019 Flowmeters Course
95/382
Primary Device
It is defined as The device mounted internally
or externally to the fluid conduit that produces a
signal with a defined relationship to the fluid
flow in accordance with known physical laws
relating the interaction of the fluid to the
presence of the primary device.
95
7/30/2019 Flowmeters Course
96/382
Secondary Device
It is defined as The device that responds to the
signal from the primary device and converts it to
a display or to an output signal that can betranslated relative to flow rate or quantity.
96
7/30/2019 Flowmeters Course
97/382
97
Some Drawing Symbols
7/30/2019 Flowmeters Course
98/382
Some Drawing Symbols
98
General Categories
7/30/2019 Flowmeters Course
99/382
Flow instrument categorization often varies.
1. Rate or quantity type.
2. Energy usage type.
General Categories
of Flow Instruments
99
1a Rate meters
7/30/2019 Flowmeters Course
100/382
They are the most common classification of
flowmeters.
Rate meters measure the process fluidsvelocity.
Because a pipes cross sectional area is known,
the velocity is then used to calculate the flow
rate.
1a. Rate meters
100
A t t ith i f th fl t
7/30/2019 Flowmeters Course
101/382
A rate meter can either infer the flow rate or
measure the velocity of the flowing fluid to
determine the flow rate.
In differential pressure flowmeter, the flow
rate is inferred from the measureddifferential pressure.
In turbine meter, the velocity of the fluidtimes the area is used to determine the flow
rate.
101
7/30/2019 Flowmeters Course
102/382
7/30/2019 Flowmeters Course
103/382
Meters that directly measure mass can alsobe considered either as
a quantity meter or as
a mass flow rate meter.
103
7/30/2019 Flowmeters Course
104/382
104
2. Energy Approach
7/30/2019 Flowmeters Course
105/382
A. Extractive Energy
Flowmeters take energy from the fluid
flow.
An orifice plate is an example of an
extractive-type.
gy pp
105
7/30/2019 Flowmeters Course
106/382
B. Additive Energy
Flowmeters introduce some energy into
the fluid flow.
A magnetic flowmeter is an example of an
additive type.
106
Volumetric Flow Rate
7/30/2019 Flowmeters Course
107/382
It represents the volume of fluid that passes ameasurement point over a period of time.
The calculation is based on the formula:
Q = A x v
where
Q = volumetric flow rate
A = cross-sectional area of the pipe
v = average flow velocity (flow rate)
107
Volumetric Flow Rate
Mass Flow Rate
7/30/2019 Flowmeters Course
108/382
It represents the amount of mass that passes a specific
point over a period of time.
The calculation is based on the formula:
W = Q x
whereW = mass flow rate
Q = volumetric flow rate
= density108
Mass Flow Rate
Units of Measure
7/30/2019 Flowmeters Course
109/382
109
Meter Run
7/30/2019 Flowmeters Course
110/382
It is defined as The upstream and downstream length
of pipe containing the orifice flanges and orifice plate or
orifice plate with or without quick change fittings.
No other pipe connections should be made within thenormal meter tube dimensions except for pressure taps
and thermowells.
The meter tube must create an acceptable flow pattern
(velocity profile) for the fluid when it reaches the orifice
plate.
110
Meter Run
7/30/2019 Flowmeters Course
111/382
111
Flow Straighteners (conditioners)
7/30/2019 Flowmeters Course
112/382
They help to provide accurate measurement when a
distorted flow pattern is expected.
They are installed in the upstream section of meter tube.
They reduce the upstream meter tube length
requirement.
112
Flow Straighteners (conditioners)
7/30/2019 Flowmeters Course
113/382
113
Compressible versus
7/30/2019 Flowmeters Course
114/382
Temperature and pressure changes cause the volume
of a fluid to change.
The change in volume is much more extreme in gasesthan in liquids.
For accurate gas flow measurements, the
compressibility factor is included in the measurement.
z =PV/nRT114
p
Incompressible Flow
7/30/2019 Flowmeters Course
115/382
115
Viscosity
7/30/2019 Flowmeters Course
116/382
116
Viscosity is frequently described as a fluids resistance
7/30/2019 Flowmeters Course
117/382
Viscosity is frequently described as a fluid s resistance
to flow.
It have a dramatic effect on the accuracy of flow
measurement.
Resistance to flow occurs because of internal friction
between layers in the fluid.
Water, for example, having low viscosity has less
resistance to flow.
117
When a fluid is in motion, layers of fluid are
7/30/2019 Flowmeters Course
118/382
subject to tangential shearing forces, causing the
fluid to deform.
Fluids low viscosity does not become an
influential property of the fluid upon flow
measurement.
However, when measuring the flow rate of a
fluid with high viscosity, the viscosity doesbecome an influential property in flow
measurement.118
Viscosity is often expressed in terms of the
7/30/2019 Flowmeters Course
119/382
Viscosity is often expressed in terms of the
following:
Dynamic viscosity
Kinematic viscosity
Viscosity index
Viscosity scales
119
Dynamic Viscosity (Absolute Viscosity)
7/30/2019 Flowmeters Course
120/382
It represents a fundamental viscosity measurement of a
fluid.
Density of fluid does not play a part in the viscosity
measurement.
Absolute viscosity is a ratio of applied shear stress to
resulting shear velocity.
The measurement units for dynamic (absolute) viscosity
are centipoise, Pascal-seconds, or lb/ft-second.
120
One method to measure viscosity is to rotate a
7/30/2019 Flowmeters Course
121/382
disk in the fluid at a particular rotational speed.
The rotational torque required to keep the disk
rotating divided by the speed of rotation and by
the disk contacting surface area is a measure of
absolute viscosity.
Another viscosity measurement that can be used
for liquids and gases is the falling sphereviscometer.
121
7/30/2019 Flowmeters Course
122/382
Rotational and Falling Sphere Viscometers122
Kinematic Viscosity (n)
7/30/2019 Flowmeters Course
123/382
It represents a ratio ofdynamic (absolute) viscosity to the
density of the fluid and is expressed in stokes (n = m / r).
In liquids, an increasing temperature usually results in
lowering the kinematic viscosity.
In gases, an increasing temperature increases the
kinematic viscosity.
The measurement units for kinematic viscosity are either
centistokes, meter2/second, or ft2/second.
123
7/30/2019 Flowmeters Course
124/382
124
Th th d f d t i i ki ti i it
7/30/2019 Flowmeters Course
125/382
The method for determining kinematic viscosity
involves measuring the time to drain a certain
volume of liquid by gravity out of a container
through a capillary tube or some type of restriction.
The time it takes to drain a liquid is directly relatedto viscosity.
The flow rate of fluids by gravity, which is the force
causing the flow, depends upon the density of the
fluids.
125
7/30/2019 Flowmeters Course
126/382
126
Ostwald Capillary Viscometer
Viscosity Index
7/30/2019 Flowmeters Course
127/382
It represents the change in viscosity with respectto temperature.
It is used with reference to petroleum products.
A high viscosity index number means that the
fluids viscosity does not change very much for a
given temperature, and vice versa.
127
Viscosity Scales
7/30/2019 Flowmeters Course
128/382
It represents viscosity measurements in time units.
Commonly used viscosity scales include the following:
oSaybolt Furol scales
oRedwood scalesoEngler scales
The three scales express kinematic viscosity in time
units rather than centistokes.
128
7/30/2019 Flowmeters Course
129/382
For example, if the kinematic viscosity of a fluid at 122
F is 900 centistokes, on the Saybolt Furol scale the
equivalent viscosity is expressed as 424.5 seconds
(centistokes x 0.4717).
Flow engineering reference manuals often provide
conversion formulas between centistokes and the
respective viscosity scale.
129
7/30/2019 Flowmeters Course
130/382
130
Basic Hydraulic Equations
7/30/2019 Flowmeters Course
131/382
131
7/30/2019 Flowmeters Course
132/382
132
Bernoulli Equation
7/30/2019 Flowmeters Course
133/382
P = Static Pressure (pounds force per sq. ft)
r = Density (rho) (pounds mass per cubic ft)
v = Velocity (feet per second)
g = Acceleration of Gravity (feet per second2)
Z = Elevation Head Above a Reference Datum (feet)
133
7/30/2019 Flowmeters Course
134/382
134
Continuity Equation
7/30/2019 Flowmeters Course
135/382
The Equation of Continuity states that the volumetric
flow rate can be calculated by multiplying the cross
sectional area of the pipe at a given point by the
average velocity at that point.
Q = A x v
where
Q = volume flow rate (cubic feet per minute)A = pipe cross-sectional area (square feet)
v = average fluid velocity (feet per minute)
135
Reynolds Number
7/30/2019 Flowmeters Course
136/382
It is a major distinctive quality of fluid flow
as
The ratio of Inertial Forces to Viscous
Forces.
136
7/30/2019 Flowmeters Course
137/382
137
7/30/2019 Flowmeters Course
138/382
138
7/30/2019 Flowmeters Course
139/382
Laminar flow is defined by low Reynoldsnumbers with the largest flowing fluid
moving coherently without intermixing.
Turbulent flow is defined by high Reynolds
numbers with much mixing.
139
Turbulent flow is best when high heat transfer is
wanted,
7/30/2019 Flowmeters Course
140/382
while laminar flow is best when flowing fluid is to bedelivered through a pipe with low friction losses.
Flow is considered laminar when the Reynolds number
is below 2,000.
Turbulent flow occurs when the Reynolds number is
above 4,000.
Between these numbers, the flow characteristics have
not been defined.140
7/30/2019 Flowmeters Course
141/382
141
7/30/2019 Flowmeters Course
142/382
142
Newtonian versus
N t i Fl id
7/30/2019 Flowmeters Course
143/382
In Newtonian fluids, the resistance to deformation
when subjected to shear (consistency of fluid) is
constant if temperature and pressure are fixed.
Whereas in a non-Newtonian fluid, resistance to
deformation is dependent on shear stress even
though the pressure and temperature are fixed.
143
non-Newtonian Fluids
7/30/2019 Flowmeters Course
144/382
Rheograms
7/30/2019 Flowmeters Course
145/382
It can be used to determine the characteristics of any fluid.
Rheograms evolved from the science of rheology, which
studies flow.
(Rheo, derived from the Greek language, means a
flowing.)
Rheograms are useful as an aid to interpret viscositymeasurements.
145
7/30/2019 Flowmeters Course
146/382
Non-Newtonian Fluids
Fluids that do not show a constant ratio of shear stress to
7/30/2019 Flowmeters Course
147/382
Fluids that do not show a constant ratio of shear stress to
shear rate are defined as non-Newtonian fluids.
Fluids exhibit different viscosity at different shear rates.
In non-Newtonian fluids, there is a nonlinear relation between
the magnitude of applied shear stress and the rate of angulardeformation.
Non-Newtonian fluids, which have different classifications,
tend to be liquid mixtures of suspended particles.
Thick hydrocarbon fluids are considered non-Newtonian
fluids.147
7/30/2019 Flowmeters Course
148/382
148
7/30/2019 Flowmeters Course
149/382
149
FLOW MEASURING DEVICE
SELECTION CRITERIA
7/30/2019 Flowmeters Course
150/382
SELECTION CRITERIA
Application fundamentals
Specifications
Safety considerations
Metallurgy Installation considerations
Maintenance and calibration
Compatibility with existing process instrumentation
Custody transfer concerns
Economic considerations
Technical direction150
7/30/2019 Flowmeters Course
151/382
Application Fundamentals Flowchart151
7/30/2019 Flowmeters Course
152/382
Flowmeter Applications
7/30/2019 Flowmeters Course
153/382
153
Flowmeter Applications (Continued)
7/30/2019 Flowmeters Course
154/382
154
7/30/2019 Flowmeters Course
155/382
Strainers are used to protect meters from debris in a
liquid stream
7/30/2019 Flowmeters Course
156/382
liquid stream.
Strainers are not intended for filtering a liquid.
Strainers should be carefully selected to ensure that they
have a low pressure drop when used with high velocityflowmeters.
156
Deaerators are air elimination devices that protect the
f i i l l f i
7/30/2019 Flowmeters Course
157/382
meter from receiving a large slug of air.
The air elimination device separates that air from the
liquid through the use of special baffles.
In the case of some positive displacement meters, a largeslug of air can completely damage the meter.
In the case of a turbine meter, air may not cause
damage, but will cause errors in readings (registrations).
157
Isolation Valves are typically provided at a meter inlet to
permit meter repairability without shutting down the
7/30/2019 Flowmeters Course
158/382
permit meter repairability without shutting down the
process.
Block and Bleed Valves are used in meter runs to
provide a means for calibration. These valves divert the
flow to the meter prover loop.
Control Valves provide a means of controlling flowrate
and/or back pressure.
For example, flowrate control is necessary to prevent a
positive displacement meter from over-speeding.
158
Typical Maintenance Concerns by Flowmeter Type
7/30/2019 Flowmeters Course
159/382
159
Accuracy Reference
7/30/2019 Flowmeters Course
160/382
Accuracy is measured in terms of maximum positive
and negative deviation observed in testing a device
under a specified condition and specified procedure.
The accuracy rating includes the total effect ofconformity, repeatability, dead-band, and hysteresis
errors.
An accuracy reference of simply 2% is incomplete.
160
Percent of Rate Accuracy: It applies to meters such as turbine meters, DC
magnetic meters, vortex meters, and Coriolis meters.
7/30/2019 Flowmeters Course
161/382
Percent of Full Scale Accuracy: It refers to the accuracy of primary meters such
as rotameters and AC magnetic meters.
Percent of Maximum Differential Pressure: It applies to differential pressure
flow transmitters.
161
Totalization
7/30/2019 Flowmeters Course
162/382
It represents the process of counting the amount of
fluid that has passed through a flowmeter.
Its purpose is to have periodic (daily or monthly)
readings of the material usage or production.
The totalization data is used for billings for material
usage or production.
162
Multivariable Transmitters
7/30/2019 Flowmeters Course
163/382
163
In measuring flow, temperature is required to
compensate for changes in density.
7/30/2019 Flowmeters Course
164/382
A multivariable transmitter is essentially fourtransmitters in one package.
A multivariable transmitter measures differential
pressure, absolute pressure, and process temperature.
The multivariable transmitter also calculates the
compensated flow.
Traditionally, three separate transmitters and flow
calculation were required for this measurement.164
The multivariable transmitter incorporates
microprocessor based technology which provides the
7/30/2019 Flowmeters Course
165/382
advantages of better readability and tighter
integration.
Additionally, the multivariable transmitter reduces
installation costs, spares inventories, andcommissioning times.
The transmitter has the flexibility to be used in
applications such as custody transfer, energy andmaterial balances, and advanced control and
optimization.
165
7/30/2019 Flowmeters Course
166/382
166
Custody Transfer
7/30/2019 Flowmeters Course
167/382
167
7/30/2019 Flowmeters Course
168/382
168
Flow measurement for custody transfer, where
ownership of a product transfers is on occasion
7/30/2019 Flowmeters Course
169/382
ownership of a product transfers, is on occasion
regarded as a separate flow measurement topic.
There are two types of custody transfer in flow
measurement:
1. Legal, which falls under weight and measure
requirements.
2. Contract, which is a mutual agreement between
seller and buyer.
169
In process control applications, the accuracy
i t b l t
7/30/2019 Flowmeters Course
170/382
170
requirement may be several percent,
but for custody transfer operations the accuracy
requirement may be in tenths of a percent.
Reasons for metering hydrocarbons
Custody Transfer Concerns
7/30/2019 Flowmeters Course
171/382
Reasons for metering hydrocarbons.
Classifications of custody transfer
measurements.
Meter provers required.
171
Reasons for Metering Hydrocarbons
In typical oil processing plants liquid hydrocarbons are
7/30/2019 Flowmeters Course
172/382
In typical oil processing plants, liquid hydrocarbons are
metered at each custody transfer point and often at pointswhere custody does not change.
Several reasons for the metering are:
Corporate accounting requires data.
Billing is dependent upon accurate measurements.
Losses are detectable.
Business decisions are based on the measurement data. Assist negotiations, if necessary
Provide auditable, historical records.
172
Classification of Custody Transfer Measurements
7/30/2019 Flowmeters Course
173/382
For a custody transfer measurement of a liquid
hydrocarbon, a contract requires a volumetric measurement
at standard conditions of temperature and pressure.
The techniques to do this are broadly categorized as static
and dynamic.
Static measurements are accomplished through automatic
tank gauging.
Dynamic measurements are accomplished through liquid
metering methods.
173
7/30/2019 Flowmeters Course
174/382
174
7/30/2019 Flowmeters Course
175/382
175
7/30/2019 Flowmeters Course
176/382
176
Meter Provers Required
7/30/2019 Flowmeters Course
177/382
Any flowmeters indication of a volumerepresents an unknown volume unless the
volume can be compared to a known
volume.
The known volumes are called
meter provers
177
7/30/2019 Flowmeters Course
178/382
For a meter to be considered accurate, the
meter must be proved at the same
conditions of flowrate, temperature
pressure, and product viscosity.
178
7/30/2019 Flowmeters Course
179/382
179
FLOW METER CALIBRATION:
IMPORTANCE AND TECHNIQUES
7/30/2019 Flowmeters Course
180/382
Calibration is typically performed in a laboratorysetting at several different flow rates, and uses
conditions such as changing densities, pressure,
and temperatures.
Proving differs from calibration in that it is done
in the field, typically under a single set of
conditions.
180
7/30/2019 Flowmeters Course
181/382
The calibration can be defined as the
comparison of a measuring instrument with
specified tolerance but an undetermined
accuracy, to a measurement standard with
known accuracy
181
The use of non-calibrated instruments creates
potentially incorrect measurement and erroneous
7/30/2019 Flowmeters Course
182/382
p y
conclusions and decisions.
It is calibration that:
provides assurance and confidence in
measurement.
maintains product in specified ranges.
182
Calibration can be a simple dimensional check to
d t t t i bl
7/30/2019 Flowmeters Course
183/382
detect measurement variables.
Before starting calibration, a decision must be
made for the following:
Which variables should be measured.
What accuracy must be maintained.
183
S l t f i t i ll t
7/30/2019 Flowmeters Course
184/382
Some element of error exists in all measurements
no matter how carefully they are conducted.
The magnitude of the error can never be easily
determined by experiments;
the possible value of the error can be calculated.
184
Method of Calibrations
I l th fl t d i lib t d
7/30/2019 Flowmeters Course
185/382
In general the flow measurement devices are calibrated
by three methods:
Wet calibration uses the actual fluid flow.
Dry calibration uses flow simulation by means of an
electronic or mechanical signal.
A measurement check of the physical dimensions anduse of empirical tables relating flow rate to these
dimensions is another form of calibration.
185
Wet Calibration
It t l fl id fl
7/30/2019 Flowmeters Course
186/382
It uses actual fluid flow.
Generally it provides high accuracy for a flowmeter
and is used when accuracy is a prime concern.
Precision flowmeters are usually wet calibrated at
the time of manufacture.
Wet calibration for flowmeters is usually performed
with water, air, or hydrocarbon fuels.
186
Dry Calibration
I i f d fl i h h
7/30/2019 Flowmeters Course
187/382
It is performed on a flowmeter without the
presence of a fluid medium.
The input signal is Hz, mV, or P.
It is much more uncertain than wet calibration.
The overall accuracy of the flow device isinferred because the flow transducer is bypassed.
187
The input signal for a dry calibration must
7/30/2019 Flowmeters Course
188/382
p g y
be provided by a measurement standard.
The value of the output signal requires use
of other measurement standard.
Follow the manufacturers guideline and
procedures for dry calibration.
188
Provers
The proving operation verifies the meters
7/30/2019 Flowmeters Course
189/382
The proving operation verifies the meter s
performance and assurance.
The necessity for proving depends on how accurate
the measurement must be for the product being
handled.
Prover is considered part of the metering stations
cost and is permanently installed at the facilities.
For low value products, portable provers are used.189
Methods of Meter Proving
Pipe provers are one of the most common
7/30/2019 Flowmeters Course
190/382
Pipe provers are one of the most common
types of provers in industry today.
The process does not have to be shut down
when proving a meter.
Two types of pipe provers:
Unidirectional prover,Bidirectional prover.
190
Unidirectional Provers
7/30/2019 Flowmeters Course
191/382
191
It displaces a known volume by means of a
displacer traveling in only one direction inside
the prover.
The displacers travel is detected by detector
switches within the prover.
7/30/2019 Flowmeters Course
192/382
192
Bidirectional Provers
7/30/2019 Flowmeters Course
193/382
It requires a displacer to travel in both directionsto complete one prover run.
After stabilizing pressure and temperature, the
displacer is put into the system.
It will slow down flow in the system for a time
until the displacer picks up speed.
193
7/30/2019 Flowmeters Course
194/382
194
Small Volume Provers
7/30/2019 Flowmeters Course
195/382
They can accommodate a wide range of flow
rates.
They are compact in size and have less volume
than conventional unidirectional and
bidirectional pipe provers.
The time to obtain a meter factor is significantlydecreased.
195
Master Meter Method
7/30/2019 Flowmeters Course
196/382
It is used when a pipe prover is unavailable.
The master meter method uses a known reliable
meter configured in series with the meter to beproved.
The meter measurements are then compared.
196
Weight and Volume Methods
7/30/2019 Flowmeters Course
197/382
Weight and Volume Methods
Static calibration
Dynamic calibration
197
Static Calibration
The flow is quickly started to begin the test held
7/30/2019 Flowmeters Course
198/382
The flow is quickly started to begin the test, held
constant during the test, and then shut off at theend of the test.
The totalized flow reading from the flowmeters iscompared with the weight or volume collected and
the performance of the meter is calculated.
The static calibration system operates best withflowmeters that have low sensitivity to low flow
rates.198
Dynamic Calibration
The flow is kept at a constant rate before the
7/30/2019 Flowmeters Course
199/382
The flow is kept at a constant rate before the
beginning of the test.
The flow reading from the flow meter and initial
weight or volume are read together to start the testand after the desired collection period to end the
test.
The dynamic calibration systems are limited by the
meters speed of the response.
199
7/30/2019 Flowmeters Course
200/382
200
7/30/2019 Flowmeters Course
201/382
201
7/30/2019 Flowmeters Course
202/382
202
Basic Equations
7/30/2019 Flowmeters Course
203/382
203
7/30/2019 Flowmeters Course
204/382
204
As long as the fluid speed is sufficiently subsonic
(V < mach 0.3),
the incompressible Bernoulli's equationdescribes the flow reasonably well.
7/30/2019 Flowmeters Course
205/382
205
It is recommended that location 1 be positioned
one pipe diameter upstream of the orifice, and
location 2 be positioned one-half pipe diameter
7/30/2019 Flowmeters Course
206/382
p p p
downstream of the orifice.
206
7/30/2019 Flowmeters Course
207/382
For flow moving from 1 to 2, the pressure at 1
will be higher than the pressure at 2;
the pressure difference as defined will be a
positive quantity.
207
From continuity, the velocities can be replaced
by cross-sectional areas of the flow and the
volumetric flowrate Q,
7/30/2019 Flowmeters Course
208/382
,
208
Solving for the volumetric flowrate Q gives,
7/30/2019 Flowmeters Course
209/382
209
For real flows (such as water or air), viscosity
and turbulence are present and act to convert
kinetic flow energy into heat.
7/30/2019 Flowmeters Course
210/382
gy
To account for this effect, a discharge coefficient
Cd is introduced into the above equation to
marginally reduce the flowrate Q,
210
Since the actual flow profile at location 2
downstream of the orifice is quite complex,
7/30/2019 Flowmeters Course
211/382
thereby making the effective value of A2uncertain, the following substitution introducing
a flow coefficient Cf is made,
where Ao is the area of the orifice.
211
As a result, the volumetric flowrate Q for real
flows is given by the equation,
7/30/2019 Flowmeters Course
212/382
212
The flow coefficient Cf is found from experiments
and is tabulated in reference books;
7/30/2019 Flowmeters Course
213/382
It ranges from 0.6 to 0.9 for most orifices.
Since it depends on the orifice and pipe diameters
(as well as the Reynolds Number), one will oftenfind Cf tabulated versus the ratio of orifice
diameter to inlet diameter, sometimes defined as,
213
Most Common P flowmeters
Orifice plates
7/30/2019 Flowmeters Course
214/382
Venturi
Flow nozzles
pitot tube / annubar
Elbow or wedge meter
214
7/30/2019 Flowmeters Course
215/382
215
Meter Tube Assembly Example
Orifice Plate
It is the main element within an orifice meter
7/30/2019 Flowmeters Course
216/382
tube.
It is the simplest and most economical type of
all differential pressure flowmeters.
It is constructed as a thin, concentric, flat metal
plate.
The plate has an opening or orifice.216
7/30/2019 Flowmeters Course
217/382
An orifice plate is installed perpendicular to the
fluid flow between the two flanges of a pipe.
As the fluid passes through the orifice, the
restriction causes an increase in fluid velocity and
a decrease in pressure.
217
The potential energy (static pressure) is
7/30/2019 Flowmeters Course
218/382
The potential energy (static pressure) is
converted into kinetic energy (velocity).
As the fluid leaves the orifice, fluid velocity
decreases and pressure increases as kineticenergy is converted back into potential energy
(static pressure).
218
7/30/2019 Flowmeters Course
219/382
Orifice plates always experience some energy
loss that is, a permanent pressure loss caused
by the friction in the plate.
219
7/30/2019 Flowmeters Course
220/382
The Beta ratio is defined as the ratio of thediameter of orifice bore to internal pipe
diameter.
< 1
220
The most common holding system for an orifice
7/30/2019 Flowmeters Course
221/382
plate is a pair of flanges, upstream anddownstream piping, and a pressure tap.
The pressure taps are located either on orifice
flanges or upstream and downstream of the pipe
from the orifice plate.
221
For precise measurement, various types of
fittings are used:
7/30/2019 Flowmeters Course
222/382
junior fittings,
senior fittings, and
simplex fittings.
222
The fittings provide:
7/30/2019 Flowmeters Course
223/382
easy installation of an orifice plate,
removal of the plate for changes in flow rate
services, and
convenient removal for inspection and
maintenance.
223
Senior Orifice Fitting
It is a dual-chamber device that reigns as the
7/30/2019 Flowmeters Course
224/382
224
most widely used means of measurement fornatural gas.
Simplex Orifice Plate Holder
It is a single-chamber fittings that house and
7/30/2019 Flowmeters Course
225/382
225
accurately position an orifice plate for differentialpressure measurement.
Junior Orifice Fitting
It is a single-chamber fitting, engineered and
7/30/2019 Flowmeters Course
226/382
226
g g g
manufactured to make orifice plate changing
quick and easy at installations where line
movement from flange spreading is undesirable.
7/30/2019 Flowmeters Course
227/382
227
7/30/2019 Flowmeters Course
228/382
228
7/30/2019 Flowmeters Course
229/382
Limitations of orifice plates include a high
irrecoverable pressure and a deterioration in
accuracy and long term repeatability because ofedge wear.
229
Two types of orifice plates designs areavailable:
7/30/2019 Flowmeters Course
230/382
Paddle type and
Universal type.
230
The paddle type orifice plate
It is used with an orifice flange, has a handle for
7/30/2019 Flowmeters Course
231/382
easy installation between flanges.
On the paddle type plate, the orifice bore,
pressure rating (flange rating), bore diameter,Beta ratio, and nominal line size are stamped on
the upstream face of the plate.
The outside diameter of a paddle plate varies
with the ANSI pressure rating of the flanges.231
The universal orifice plate
It is designed for use in quick change fittings.
7/30/2019 Flowmeters Course
232/382
The universal plate is placed in a plate holder,
the outside diameter is the same for all pressure
ratings for any given size.
When using orifice fittings, the internal
diameter of the meter tube must be specified
because the orifice plate is held in an orifice
plate sealing unit.
232
7/30/2019 Flowmeters Course
233/382
Weep Hole
7/30/2019 Flowmeters Course
234/382
Some orifice plates have a small hole in theorifice plate besides an orifice bore either
above the center of the plate, or below the center of the plate.
234
7/30/2019 Flowmeters Course
235/382
The area of the weep hole must be considered
7/30/2019 Flowmeters Course
236/382
when sizing an orifice plate.
An orifice plate with a weep hole should not be
used when accurate measurement is required in a
flow measurement application, such as in gas
sales service.
236
7/30/2019 Flowmeters Course
237/382
7/30/2019 Flowmeters Course
238/382
The orifice plate, although a relatively simple
element, is a precision measuring instrument and
should be treated accordingly.
238
Critical items considered when evaluating orifice
plates are the following:
7/30/2019 Flowmeters Course
239/382
Flatness, smoothness, and cleanliness ofthe orifice plate.
The sharpness of the upstream orificeedge.
The bore diameter and thickness of the
orifice plate.
239
Orifice Plate Dimensions
7/30/2019 Flowmeters Course
240/382
d represents the bore of the orifice plate.
D represents the pipe inside diameter.
Dam height represents the difference of pipe inner diameter and diameter of bore
divided by 2.
T represents the thickness of the plate.
e represents the orifice plate bore thickness which is 1/2 T
is called orifice plate bevel angle. It is 45 , +20 0.
240
7/30/2019 Flowmeters Course
241/382
7/30/2019 Flowmeters Course
242/382
Types of Orifice Plates
242
Concentric Plates
The concentric orifice bore plates are used in
7/30/2019 Flowmeters Course
243/382
general flow measurement applications.
The concentric orifice plate has an orifice bore
in the center of the plate.
The concentric bore plate is used for clean fluid
services, as well as for applications requiring
accurate flow measurement.
243
The center of bore is either
7/30/2019 Flowmeters Course
244/382
beveled or
straight.
The beta ratio for the concentric plate is
between 0.1 to 0.75.
244
Eccentric Plates
It is similar to a concentric plate, but the
7/30/2019 Flowmeters Course
245/382
eccentric plate has the bore in an offset position.
The eccentric orifice plate is used when dirty
fluids are measured, to avoid the tendency ofhole plugging if a concentric plate were used.
Flow coefficient data is limited for eccentric
orifices; therefore, it provides less accurate
measurement.245
7/30/2019 Flowmeters Course
246/382
246
In an eccentric orifice plate, the hole is bored
tangent to the inside wall of the pipe or, more
l l
7/30/2019 Flowmeters Course
247/382
commonly, tangent to a more concentric circlewith a diameter not smaller than 98% of the
pipes internal diameter.
When lacking specific process data for the
eccentric orifice plate, the concentric orifice plate
data may be applied as long as accuracy is not a
major issue.
247
Make sure that flanges or gaskets do not
interfere with the plate hole.
h l f
7/30/2019 Flowmeters Course
248/382
The line size ranges from 4 to 14.
It can be made smaller than a 4 as long as the
orifice bore does not require a beveling edge.
Beta ratio is limited between 0.3 to 0.8.
Flange taps are recommended for eccentric
orifice plate installations.248
Segmental Plates
It looks like a segment of a circle with
d i l h l i ff f h l
7/30/2019 Flowmeters Course
249/382
segmented circle hole in offset from the platescenter.
The orifice hole is bored tangent to the insidewall of the pipe or tangent to a more concentric
circle with a diameter not smaller than 98% of
the pipe internal diameter.
Installation is similar to eccentric type.249
Quadrant Edge Plate
I i d f l i R ld b
7/30/2019 Flowmeters Course
250/382
It is used for lower pipe Reynolds numberswhere flow coefficients for sharp-edge orifice
plates are highly variable.
It is used for viscous clean liquid applications.
Nominal pipe size ranges between 1 to 6.
250
Orifice Plate Parameters
(1) Orifice flow rate.
(2) Pi li i d i
7/30/2019 Flowmeters Course
251/382
(2) Pipe line size and pressure rating.(3) Thickness of orifice plate.
(4) Orifice Bore (d).
(5)Orifice plate holders: The orifice plate holder includes
orifice flanges, orifice fittings.(6) Beta Ratio.
(7) Differential Pressure (P).
(8) Temperature.
(9) Reynolds Number (Re).
(10) Pressure taps.
251
Pressure Taps
7/30/2019 Flowmeters Course
252/382
252
7/30/2019 Flowmeters Course
253/382
253
Flange Taps
Holes drilled into a pair of flanges.
7/30/2019 Flowmeters Course
254/382
Flange tap holes are not recommended when
the pipe size is below 2 inches.
254
Pipe Taps
Pi t l t d t 2 5 D t d 8 D
7/30/2019 Flowmeters Course
255/382
Pipe taps are located at 2.5 D upstream and 8 Ddownstream from the orifice plate.
Exact location of the taps is not critical.
However, the effect of pipe roughness and
dimensional inconsistencies can be severe.
255
The uncertainty of measurement is 50 % greater
ith f ll fl t th ith t l t th
7/30/2019 Flowmeters Course
256/382
with full flow taps than with taps close to the
orifice.
Pipe taps are not normally used unless it isrequired to install the orifice meter on a existing
pipe, or other taps cannot be used.
256
Corner Taps
Corner taps are a style of flange taps.
7/30/2019 Flowmeters Course
257/382
The only difference between corner and flange
taps is that the pressure is measured at the
corner between the orifice plate and the pipewall.
Corner taps are used when the pipe size is 2 or
less.
257
Vena Contracta Taps
When an orifice plate is inserted into the
7/30/2019 Flowmeters Course
258/382
flowline, it creates an increase in flow velocity
and a decrease in pressure.
The location of the vena contracta point isbetween 0.35 to 0.85 of pipe diameters
downstream of the plate, depending on the beta
ratio and Reynolds number.
258
7/30/2019 Flowmeters Course
259/382
Pressure and Flow Profile
259
Vena contracta taps are located 1D upstream and at the
Vena contracta location downstream.
Vena contracta Taps are the optimum location for
7/30/2019 Flowmeters Course
260/382
Vena contracta Taps are the optimum location formeasurement accuracy.
They are not used for pipes less than 6 in diameter.
260
Liquid Service
Tap Locations The pressure tap location in liquid
service orifice meters should be located to prevent
7/30/2019 Flowmeters Course
261/382
service orifice meters should be located to preventaccumulation of gas or vapor in the connection
between the pipe and the differential pressure
instrument.
The differential pressure instrument should be close
to the pressure taps or connected through downward
sloping connecting pipe of sufficient diameter toallow gas bubbles to flow back into the line.
261
Transmitter Installation The installation of
differential pressure transmitters should be
located below the pipe and sloping upwards
toward the pipe to prevent the collection of gas
7/30/2019 Flowmeters Course
262/382
toward the pipe to prevent the collection of gasbubbles in the impulse tubing.
Vent Holes are required for venting of any gas ina liquid service.
Location of the vent hole in a liquid service is atthe top of a pipe, above the center line.
262
Gas Services
Tap Locations Pressure tap locations in a gas
service must be installed in the top of the line
7/30/2019 Flowmeters Course
263/382
service must be installed in the top of the linewith upward sloping connections towards a pipe.
The differential pressure measuring instrumentmay be close-coupled to the pressure taps in the
side of the lines or connected through upward
sloping connecting pipe of sufficient diameter to
prevent liquid from accumulating in the line.
263
Transmitter Installation The installation of
differential pressure transmitters should be
located above the pipe with the impulse tubing
sloping downward towards the pipe so that any
7/30/2019 Flowmeters Course
264/382
sloping downward towards the pipe so that anycondensate drains into the pipe.
Drain Holes A drain hole is required for drainingof any liquid in a gas service.
Location of the drain hole is below the center line
of the pipe.
264
Steam Services
Tap Locations require the use of condensing
chambers in steam or vapor applications because
7/30/2019 Flowmeters Course
265/382
chambers in steam or vapor applications becausecondensate can occur at ambient temperatures.
Generally, the pressure tap connection has adownward sloping connection from the side of
the pipe to the measuring device.
265
Transmitter Installation The installation of
differential pressure transmitters should be
located above the pipe with the impulse tubing
sloping downward towards the pipe so that any
7/30/2019 Flowmeters Course
266/382
sloping downward towards the pipe so that anycondensate drains into the pipe.
Drain Holes A drain hole is required for drainingof any condensate liquid in a steam service.
The location of a drain hole is below the centerline of the pipe.
266
Standard Flow
Flow measurement of a fluid stated in volume units at base
(standard) conditions of P and T is called standard flow.
7/30/2019 Flowmeters Course
267/382
For crude petroleum and its liquid products, the vapor pressure is
atmospheric
pressure at base temperature, the base pressure is called
equilibrium vapor pressure.
The base condition for natural gases is defined as a pressure of
14.73 psia (101.56 kPa) at a temperature of 60 F (15.56C).
267
Compensated Flow
Compensated flow represents a flow under fluid conditions that
may vary.
7/30/2019 Flowmeters Course
268/382
The conditions are measured and used along with flowmeter
signal to compute the true flow rate from the flowmeter.
The output signal from a flowmeter represents the true flow rate
value under specified fluid conditions.
For a liquid service, variations in density or viscosity can change
the meters accuracy.
For gas services, a change in temperature, pressure, and
molecular weight can ruin the accuracy of the meter.
268
7/30/2019 Flowmeters Course
269/382
269
Computer Programs for Sizing Orifice Plates
ORICALC-2,
EA-25
7/30/2019 Flowmeters Course
270/382
EA 25,
ORSPEC,
FLOWEL,
INSTRUCALC,
ORIFICE2, and
FLOW CONSTANT270
http://www.pipeflowcalculations.com/orifice/
7/30/2019 Flowmeters Course
271/382
271
Common primary element errors:
Beta ratio is too large for the meter run
Orifice plate is not flat, it is concave or convex
Orifice does not have sharp edges
Orifice plate is installed backwards
7/30/2019 Flowmeters Course
272/382
Orifice plate is installed backwards Orifice plate is damaged through poor handling
An incorrect size is used for the orifice meter tube or plate
Orifice plate is not centered in the line
Orifice meter tube is corroded
Tap locations are incorrect Contaminants build up on orifice plate
Contaminants build up on meter run
Hydrates build up on meter run and orifice plate
Flow conditioners are dislodged and move closer to plate
Leaks occur around orifice plate Pressure tap or thermowell installed upstream of meter
Welding meter supports distorts meter run
272
Common secondary element errors:
Gauge lines are too small
Gauge lines are too long
Gauge lines leak
7/30/2019 Flowmeters Course
273/382
Gauge lines leak
Gauge lines have sags or loops that collect condensates
Gauge line slopes are not correct
Incorrect ranges are used on secondary instruments
Differential pressure transmitter was not zeroed properly Excessive dampening is used in secondary instrument
273
Other Differential Pressure Flowmeters
7/30/2019 Flowmeters Course
274/382
274
Flow Nozzles
The flow nozzle is another type of differential-
producing device that follows Bernoullis
7/30/2019 Flowmeters Course
275/382
producing device that follows Bernoulli stheorem
The permanent pressure loss produced by theflow-nozzle device is approximately the same as
the permanent pressure loss produced by the
orifice plates.
275
The flow nozzle can handle dirty and abrasive
fluids better than can an orifice plate
7/30/2019 Flowmeters Course
276/382
fluids better than can an orifice plate.
In a flow nozzle with the same line size, flow
rate, and beta ratio as an orifice meter, thedifferential pressure is lower, and the permanent
pressure loss is less.
276
Performance and Applications
Changing a flow nozzle is more difficult than
7/30/2019 Flowmeters Course
277/382
Changing a flow nozzle is more difficult thanchanging an orifice plate when there is a change
in flow rate requirements.
Flow nozzles are used for steam, high velocity,
nonviscous, erosive fluids, fluids with some
solids, wet gases, and similar materials.
277
The flow nozzles pass 60% more flow than the
orifice plate of the same diameter and
7/30/2019 Flowmeters Course
278/382
orifice plate of the same diameter and
differential pressure.
A flow nozzles inaccuracy of 1% of rate isstandard with 0.25% of rate flow calibrated.
278
7/30/2019 Flowmeters Course
279/382
279
Typical Nozzle Installations
Venturi Meter
A venturi design can be described as a restriction
with a long passage with smooth entry and exit
7/30/2019 Flowmeters Course
280/382
with a long passage with smooth entry and exit.
Venturi tubes produce less permanent pressure
loss and more pressure recovery than the other
meters.
It is one of the more expensive head meters.
Low pressure drops for non-viscous fluids.280
7/30/2019 Flowmeters Course
281/382
281
7/30/2019 Flowmeters Course
282/382
282
Venturi Designs
Performance Advantages:
The long form venturi develops up to 89%f 0 75 b t ti d
7/30/2019 Flowmeters Course
283/382
The long form venturi develops up to 89%pressure recovery for a 0.75 beta ratio and
decreases to 86% recovery for a 0.25 beta ratio.
The short form venturi develops up to 85%
recovery at 0.75 beta ratio and decreases to 7 %
at 0.25 beta ratio.
283
A venturi meter has a low permanent pressure
loss and high recovery at higher beta ratios.
7/30/2019 Flowmeters Course
284/382
A venturi meter can be used for dirty fluids and
slurries.
Higher accuracy (better than orifice).
284
Performance Disadvantages:
A venturi meter is a very expensive measuring
7/30/2019 Flowmeters Course
285/382
A venturi meter is a very expensive measuring
device to use.
A venturi meter has limited rangeability and isonly installed when flow rates rangeability is less
than 3 to 1.
285
Pitot Tubes
The previously discussed primary differential
pressure flow metering devices utilized the
7/30/2019 Flowmeters Course
286/382
pressure flow metering devices utilized thedifference in static pressure perpendicular to the
direction of flow as a basis for inferring velocity.
The actual velocity was not measured, but was
calculated after many experimental laboratory
measurements and correlations.
286
The Pitot tube measures a fluid velocity by
converting the kinetic energy of the flow into
potential energy.
7/30/2019 Flowmeters Course
287/382
The conversion takes place at the stagnation
point, located at the Pitot tube entrance.
287
7/30/2019 Flowmeters Course
288/382
288
A pressure higher than the free-stream (i.e.
dynamic) pressure results from the kinematic to
potential conversion.
7/30/2019 Flowmeters Course
289/382
This "static" pressure is measured by comparing
it to the flow's dynamic pressure with a
differential manometer
289
7/30/2019 Flowmeters Course
290/382
290
7/30/2019 Flowmeters Course
291/382
291
Performance Advantages:
It creates very little permanent pressure drop and, as a
result, is less expensive to operate.
7/30/2019 Flowmeters Course
292/382
A pitot tube can be installed on 4 and more.
Performance of the pitot tube is historically proven.
A pitot tubes installation and operation costs are low.
A pitot tube can be a standard differential producingdevice for all pipe sizes.
292
Performance Disadvantages:
Point-type pitot tubes require traversing the
flow stream for average velocity.
7/30/2019 Flowmeters Course
293/382
g y
Poor rangeability.
Nonlinear square root characteristic.
Difficulty of use in dirty flow streams.
293
Annubars
The sensing points are arrayed along perpendicular
diameters with the number of points in each traverse
based upon the duct size.
7/30/2019 Flowmeters Course
294/382
p
294
7/30/2019 Flowmeters Course
295/382
295 Annubar Design
Performance
The diamond shape annubar has long term
accuracy.
7/30/2019 Flowmeters Course
296/382
y
The annubar has an accuracy of 1% of actual
flow and 0.1 repeatability of the actual value.
The annubar has low installation costs; a system
shutdown is not required to install the device.
296
The annubar produces a repeatable signal evenwhen the run requirements are not met.
The annubar flow sensor can handle a widerange of flow conditions with two measuring
7/30/2019 Flowmeters Course
297/382
range of flow conditions with two measuring
instruments.
The annubar should not be used if the viscosity
approaches 50 centipoise.
The annubar can be used on two phase flowmeasurements.
297
Applications
The annubar can be used for liquid and gas flow
measurement services.
7/30/2019 Flowmeters Course
298/382
Generally, the annubar is used in clean liquid
services to avoid plugging.
The annubar can be installed for low and
medium pressure applications without shutting
down the system.
298
Wedge Type Flowmeter
The basic system consists of a cylindrical pressure
vessel into which a constriction "wedge" is
fabricated thereby leaving a open segment of a
7/30/2019 Flowmeters Course
299/382
299
y g p g
known height.
Pressure taps which receive the sensors on either sideof the "wedge" provide the differential signal to the
Flow Transmitter which is then related, by formula, to
the rate of flow occurring through the open segment.
7/30/2019 Flowmeters Course
300/382
300
Elbow Type Flowmeter
A differential pressure
exists when a flowing fluidchanges direction due to a
7/30/2019 Flowmeters Course
301/382
301
changes direction due to a
pipe turn.
The pressure difference
results from the centrifugal
force.
Since pipe elbows exist inplants, the cost for these
meters is very low.
However the accuracy is
7/30/2019 Flowmeters Course
302/382
302
However, the accuracy is
very poor.
They are only applied
when reproducibility is
sufficient and other flow
measurements would bevery costly.
7/30/2019 Flowmeters Course
303/382
303
7/30/2019 Flowmeters Course
304/382
304
7/30/2019 Flowmeters Course
305/382
305
7/30/2019 Flowmeters Course
306/382
306
7/30/2019 Flowmeters Course
307/382
307
7/30/2019 Flowmeters Course
308/382
308
Flowing fluid forces the turbine wheels to rotate
at a speed proportional to the velocity of the
Turbine Meters
7/30/2019 Flowmeters Course
309/382
p p p y
fluid.
309
For each revolution of the turbine wheel, a
pulse is generated.
The rotational speed of shaft and frequency ofthe pulse corresponds to the volumetric flow
7/30/2019 Flowmeters Course
310/382
the pulse corresponds to the volumetric flow
rate through the meter.
310
K-factor
It is the number of pulses per unit of measurement
generated by the rotor as it turns inside the turbine.
It is usually indicted as Pulses per Gallon
7/30/2019 Flowmeters Course
311/382
311
It is usually indicted as Pulses per Gallon
7/30/2019 Flowmeters Course
312/382
312
7/30/2019 Flowmeters Course
313/382
313Turbine Meter
7/30/2019 Flowmeters Course
314/382
314
Insertion Type Turbine Meter
Performance Advantages
Excellent accuracy and good rangeability over the
full linear range of a meter.
7/30/2019 Flowmeters Course
315/382
Low flow rate designs are available.
Some versions do not require electrical power.
Overall meter cost is not high.
Output signal from the meter is at a high
resolution rate, which helps reduce meter proving.315
Performance Disadvantages
Sensitive to a fluids increasing viscosity.
T h fl id t bl
7/30/2019 Flowmeters Course
316/382
Two phase fluids can create usage problems.
Straight upstream piping or straighteningvanes are required in a turbine meter installation
to eliminate the flow turbulence into the meter.
316
ElectromagneticFlowmeters
7/30/2019 Flowmeters Course
317/382
317
Faradays Law states that emf is created when a
conductive fluid moves through a magnetic
field.
7/30/2019 Flowmeters Course
318/382
318
The axis of the conductive fluid flows at a right angle tothe magnetic field. Fluid flowing in this manner causes
a voltage that is proportional to the flow rate.
7/30/2019 Flowmeters Course
319/382
319 Magnetic Flowmeter Principles
The voltage developed at the electrodes has an
extremely low level signal.
A signal conditioner must amplify the signal.
7/30/2019 Flowmeters Course
320/382
There are two types of magnetic flowmeters:
AC excitation, and
DC pulse excitation.
320
AC Excitation
In an AC type magnetic flowmeter, line voltage (120 or 240
V AC) is applied directly to the magnetic coils.
This generates a magnetic field in the outer body that varies
7/30/2019 Flowmeters Course
321/382
with the frequency of the applied voltage.
An AC meters signal has a sine wave pattern.
The magnitude of the sine wave is directly proportional to
the flow velocity.
The system produces an accurate, reliable, fast responding
meter.321
DC Pulse Excitation
In a DC type magnetic flowmeter, line voltage is the
main source of power, but instead of applying it directlyto the coils, it is first applied to a magnet driver circuit.
7/30/2019 Flowmeters Course
322/382
The magnet driver circuit sends low frequency pulses to
the coils to generate a magnetic field.
The DC pulse system eliminates the zero shift problem
that occurs in an AC system.
322
7/30/2019 Flowmeters Course
323/382
323
Performance Advantages
It is non-obstructive and has no moving parts.
Pressure drop is very little.
DC pulse type power can be as low as 15 to 20 watts
7/30/2019 Flowmeters Course
324/382
DC pulse-type power can be as low as 15 to 20 watts.
Suitable for acid, bases, water, and aqueous solutions.
Lining materials provide good electric insulation and
corrosion resistance.
The magnetic meter can handle extremely low flow.
It can be used for bidirectional flow measurements..
324
Performance Disadvantages
The meters only measure conductive fluid flows.
(Hydrocarbons, gases, and pure substances cannot be measured)
7/30/2019 Flowmeters Course
325/382
( y , g , p )
Proper electrical installation care is required.
Conventional meters are heavy and larger in size.
Meters are expensive.
325
Installation
Proper magnetic flow meter operation is very
dependent upon the installation.
7/30/2019 Flowmeters Course
326/382
Installation considerations for a magnetic flowmeter
primarily involve the following:
Meter orientation
Minimum piping requirement
Grounding
326
7/30/2019 Flowmeters Course
327/382
327
Magnetic Flowmeter Installation Practices
Applications
It is suited for measurement of slurries and dirty fluids
because magnetic flowmeters do not have sensors that
enter the flowing stream of fluids.
7/30/2019 Flowmeters Course
328/382
Magnetic flowmeters are not affected by viscosity or the
consistency of Newtonian or non-Newtonian fluids.
The resulting change in flow profile caused by a change
in Reynolds number or upstream configuration piping
does no